Abstract
Purpose: To make the planning of volumetric modulated arc therapy (VMAT) faster and to explore the tradeoffs between planning objectives and delivery efficiency.
Methods: A convex multicriteria dose optimization problem is solved for an angular grid of 180 equi-spaced beams. This allows the planner to navigate the ideal dose distribution Pareto surface and select a plan of desired target coverage versus organ at risk sparing. The selected plan is then made VMAT deliverable by a fluence map merging and sequencing algorithm, which combines neighboring fluence maps based on a similarity score and then delivers the merged maps together, simplifying delivery. Successive merges are made as long as the dose distribution quality is maintained. The complete algorithm is called VMERGE.
Results:VMERGE is applied to three cases: a prostate, a pancreas, and a brain. In each case, the selected Pareto-optimal plan is matched almost exactly with the VMAT merging routine, resulting in a high quality plan delivered with a single arc in less than 5 min on average.
Conclusions:VMERGE offers significant improvements over existing VMAT algorithms. The first is the multicriteria planning aspect, which greatly speeds up planning time and allows the user to select the plan, which represents the most desirable compromise between target coverage and organ at risk sparing. The second is the user-chosen epsilon-optimality guarantee of the final VMAT plan. Finally, the user can explore the tradeoff between delivery time and plan quality, which is a fundamental aspect of VMAT that cannot be easily investigated with current commercial planning systems.
Keywords: VMAT, optimization, convex, multi-criteria, pareto optimal
INTRODUCTION
In the late 1990s, intensity-modulated radiation therapy (IMRT) entered the clinical scene and quickly rose to a dominant position in radiation treatment. The first IMRT treatment was delivered as slice-by-slice rotational arc therapy using a special-purpose collimator.1 Soon after that, IMRT delivered from a small number (around ten) of fixed gantry angle positions using a general-purpose multileaf collimator (MLC) took over as the dominant IMRT delivery technique and has been in that role ever since. Throughout this paper, we will use the term “IMRT” to denote this specific delivery mode. The reader should keep in mind, however, that IMRT is a broader term that includes various types of delivery, including arc therapies.
Rotational arc therapy in the form of volumetric intensity-modulated arc therapy (IMAT), delivered using an MLC, was proposed in 1995 (Ref. 2), but has been dormant for a long time until it recently re-emerged as volumetric modulated arc therapy (VMAT) with additional degrees of freedom, i.e., gantry speed and dose rate. One reason why it took so long for IMAT or VMAT to become a viable treatment option is that VMAT planning poses a larger optimization problem than IMRT because it delivers radiation from every angle around the patient, and therefore dose computations need to be done for far more angles than IMRT. An even bigger hurdle presented by VMAT optimization is due to the coupling between adjacent angles. More specifically, to ensure an efficient VMAT delivery, one should not move the MLC leaves more than needed between neighboring angles. If VMAT plans are to be optimized with delivery time in mind, leaf positions need to be taken into account, which results in a large-scale nonconvex optimization problem. Therefore, VMAT planning remains a great challenge and can be much more time consuming than IMRT planning.3, 4
We have previously introduced interactive multicriteria-optimization (MCO) to facilitate IMRT planning.5 The basic idea of interactive MCO is that a representative set of “Pareto-optimal” treatment plans are precalculated and stored in a database. Pareto-optimal treatment plans are those for which improving one criterion value is not possible unless some other criterion value deteriorates. The user can then navigate interactively through this database by exploring convex combinations of the database plans to obtain one that yields the desired tradeoff between different criteria. Therefore, MCO eliminates the time-consuming trial-and-error aspect of the planning process and offers a more streamlined and effective approach to treatment plan optimization.6 Since VMAT poses a more challenging treatment planning problem, MCO has the potential for great benefit in VMAT planning. This has been our main motivation in developing this work. The goal of this paper is to develop and test a multicriteria approach to VMAT planning that allows for exploring tradeoffs between planning objectives, as well as the dose quality and the delivery efficiency. We will achieve this by developing a two-stage method. In the first stage, MCO is used to navigate to a desired plan. In the second stage, the obtained plan is made VMAT deliverable, where the delivery efficiency is iteratively improved while maintaining the dose quality.
It is possible to blindly write down the VMAT optimization problem and then apply various optimization algorithms to try to solve it. However, due to the complexity of the problem, we feel it is more useful to first clearly understand the physical basis (hardware, treatment dose parameters, etc.) of the VMAT problem. The issues of fraction size, dose rate, leaf speed, and gantry speed are intertwined in VMAT. Larger fraction sizes mean more dose delivered per degree, which means, all else being equal, the gantry will rotate more slowly. This then implies that the leaves have relatively more time to move across the field to create modulated fields. But at the outset, based on a certain fraction size and a set of machine parameters, should we expect to see generally small fields or large fields? Can we be sure that when the gantry slows down or the dose rate decreases during a VMAT delivery that this is beneficial behavior and not simply an artifact of the optimization? To address these questions, it is useful to consider a specific scenario.
Assuming a 2 Gy fraction to the entire target can be delivered via a single 10 cm × 10 cm field, this can be done in 200 monitor units (this is a typical MU/Gy calibration). With a maximum dose rate of 600 MU/min, the 2 Gy could be delivered in 20 s. At a maximum gantry speed of 6 degrees per second, a single revolution takes 60 s. Therefore, for a single revolution VMAT plan at maximal dose rate and gantry speed, in order to preserve the integral dose one will see average segment sizes on the rough order of 20/60 = 1/3 the size of a 10 cm × 10 cm field.34 If one is willing to have the beam slow down to an average of half speed, completing the single arc in 120 s, one would see on average segments with a further 50% reduction in size. This provides some intuition on when to expect small segments in a VMAT delivery.
With a maximum leaf speed of 2.5 cm/s, leaves can travel across a 10 cm field in 4 s, in which time the gantry can rotate up to 24°. To deliver highly modulated fields that are spaced close together, the gantry may need to slow down. In general, in a VMAT delivery, the gantry needs to slow down when it takes longer to deliver the fluence pattern (required over some arc portion) than can be done at top gantry speed. It is useful to break this into two cases that represent the two causes for gantry slowing. The first case is when the fluence map has fluence levels that exceed the maximal fluence level that can be delivered at top gantry speed over the given arc portion. This clearly requires the gantry to slow down. The second case is when the field is modulated so much that it takes more time than available at top gantry speed to deliver all the isolated humps of the fluence map. The reason to distinguish these two cases is the following: in the first case, assuming the fluence profile is flat but at a large value, leaf speed is not the limiting factor because the gantry needs to slow down just to get enough dose in. For the second case, however, the modulated fields might be able to be delivered without slowing down the gantry if the leaves could move fast enough.
A useful relationship here is that the delivery time for a fluence map to be delivered by a left-to-right leaf sweep across the field is equal to the time it takes for the leaves to cross the field at top leaf speed plus the sum-of-positive-gradients (SPG) term [see Eq. 3]. The SPG for an IMRT field is a measure of the “ups and downs” of the field. SPG can be minimized exactly in a convex optimization framework,7 whereas leaf travel distance, if incorporated up-front in the optimization, results in a nonconvex problem. In our approach, we handle leaf travel issues in a VMAT-customized fluence map merging-and-sequencing routine which explicitly ensures that the dose distribution quality is maintained, while the delivery efficiency is successively improved. Our algorithm is designed to solve one of the key design issues of VMAT planning: where to optimally slow down the gantry. By merging like neighboring fluence maps and validating that the dose distribution after the merge is still good, we eliminate unnecessary gantry slow downs which arise from “overdelivery” of fluence maps. With our approach, the leaves travel back and forth at a high frequency only when needed and likewise the beam slows down only when necessitated by leaf travel requirements or SPG requirements.
VMAT treatments are currently delivered with Elekta8 and Varian9 equipment, and VMAT-like deliveries have been recently reported using Siemens equipment.10 The treatment delivery systems deployed by the different manufacturers have different designs, and thus impose different delivery constraints in treatment planning. The Elekta and Varian linacs both allow dynamic machine parameter changes during the irradiation, whereas for Siemens, the delivery proceeds via a burst mode in a step and shoot fashion. A full review of VMAT optimization techniques is provided by Yu and Tang.11 Here, we review two algorithms which show the two general approaches used for VMAT planning. In 2007, Varian adopted a one step algorithm for single-arc VMAT, reported by Otto,9 under the tradename Rapid Arc (Varian Medical Systems, Palo Alto, CA). The method first optimizes the MLC motions for a coarse sampling of static points. Finer sampling is achieved by iteratively adding samples interpolated between existing static points until the desired sampling frequency is reached. Throughout, leaf positions are modified by local random search. By permuting the machine parameters directly, the Otto et al. solution to VMAT must solve a highly nonconvex problem. An alternative algorithm described in Bzdusek et al.,12 that became commercialized under the name Smart Arc (Philips Radiation Oncology Systems, Fitchburg, WI), uses a two-step approach to the problem. The algorithm initially optimizes a set of fluence maps at a coarse angular sampling around an arc, followed by the sequencing of these maps and a direct parameter optimization of the resulting leaf positions. The initial fluence optimization step is a convex problem (if convex objective functions are used) and so can benefit from fast reliable optimization methods. Similar two-step approaches are found in Refs. 13, 14.
The method proposed herein uses a two-step approach similar to SmartArc: a fluence based optimization followed by leaf sequencing. However, our algorithm begins with fine beam spacing and leaf sequencing is accomplished using a unidirectional sequencing algorithm. This initial finely sampled plan represents the best possible treatment given unrestricted time to deliver it. After obtaining this initial plan, neighboring fluence maps are iteratively merged to increase gantry speed and decrease delivery time. In this way, we work from the ideal solution toward one that is epsilon close (user-chosen) to dose optimality but has greatly increased delivery efficiency. We call the complete algorithm VMERGE.
METHODS
We begin by solving a 180 equi-spaced beam IMRT problem. We solve a multicriteria version of the IMRT optimization problem, which allows the planner to explore the tradeoffs between target coverage and healthy organ sparing, finally choosing a best-compromise solution.6, 15, 16 Such a solution represents an ideal dosimetric plan, where treatment time is ignored. To actually deliver this solution, one would deliver the full IMRT fluence maps at every 2°, which would be time consuming. Instead, we successively coarsen this 180 beam fluence map solution such that the delivery is made faster, while the dose quality is kept within user selected bounds. Thus, in the sequencing step, we allow the user to explore the tradeoff between dose quality and delivery time.
In Secs. 2A, 2B, and 2C, we describe the details of each of these components of our VMAT planning approach.
180 beam IMRT solution and Pareto surface plan selection
We consider the following multicriteria IMRT problem:
(1) |
Here, d is the vector of voxel doses, D is the dose-influence matrix, and f is a concatenation of all the fluence maps into a single beamlet fluence vector. The constraint set C is a convex set of dose constraints. This can include, for example, bounds on mean structure doses, and minimum and maximum doses to individual voxels. Note, the constraints in this set should be known to be simultaneously achievable (i.e., they are easy constraints to meet). More difficult dosimetric goals, and mutually conflicting goals, should be cast as objectives, in order to eliminate the possibility of infeasibility in this stage 1 optimization problem.
The objective functions are g1(d),…,gN(d) where N is the number of objectives defined. The optimization objectives can be any of the following: minimize the maximum structure dose, maximize the minimum structure dose, or minimize or maximize the mean structure dose. In general, any convex functions would be permissible.17 For our optimization, we only consider these ones since they can be handled with a linear solver, and since in the multicriteria planning context, they are typically sufficient to create high quality treatment plans.18, 19
We solve this problem multiple times, approximating the Pareto surface, by following the methods detailed in Ref. 20. Briefly, this method uses a feasibility projection solver that iteratively projects onto violated constraints until all constraints are satisfied. Objectives are turned into constraints with initially loose bounds which are gradually tightened until they are within user specified tolerance of optimality. After the projection solver runs for the N objectives and some mixed objective plans (for mean dose objectives, a mixed plan would be obtained by optimizing a weighted combination of the mean structure doses; for more sophisticated strategies, see Refs. 20, 21, 22, 23), the user navigates the solution space, which amounts to choosing the most preferable convex combination of the calculated Pareto surface plans. The best choice plan is user dependent. Indeed, this is the entire point of MCO: if optimal plan selection criteria could be determined a priori then navigating a Pareto surface would be unnecessary. For the cases studied herein, we selected plans with conformal dose distributions to the targets and which emphasized large mean dose sparing of the most important critical structures. The selected plan, which we consider the ideal dosimetric plan, is then passed to the leaf sequencing and merging routine, described in Secs. 2B, 2C.
Fluence map smoothing
In the VMERGE algorithm, fluence maps will be combined based on similarity. Smoothing the fluence maps during optimization will reduce the noise level in the fluence maps, therefore making the similarity comparisons more meaningful. Fluence map smoothing also results in faster final delivery time, as the results will show. We consider two smoothing methods. The first constrains all fluence levels to be below a certain value. We call this max beamlet smoothing, and it is attractive because it is simple to implement. The second smoothing method uses an SPG smoother during the solver’s projection steps. During the projection iterations, an SPG smoothing step is periodically called. For each fluence map, this step identifies the single row with the largest SPG and redistributes the fluences by reducing the peak fluences of that row by a factor of 0.99, and then adding the 1% to the neighboring adjacent beamlets. This is a heuristic approach to controlling the SPG with a projection solver, inspired by smoothing kernels in projection-based image reconstruction.24 Note, both max beamlet smoothing and SPG smoothing are convex in the sense that averaging two or more plans that have been smoothed will result in a plan that is as good as or better than (in terms of the smoothing metric) the average of the smoothing metric of the base plans.
Unidirectional leaf sequencing
In leaf sequencing, the task is to create a set of MLC leaf trajectories, which produce the desired fluence map, while the gantry rotates over the arc portion allotted to that map. To be deliverable, the leaf trajectories must not have leaf velocities greater than a given maximum value, either within the delivery of a given fluence map or between the delivery of one map and the next. A simple way of ensuring that this condition is met is to sequence the trajectories as an alternating sequence of left to right and right to left dynamic MLC (dMLC) leaf sweeps. All leaves are aligned at one edge of the field at the beginning of the arc portion delivery and align at the opposite edge of the field at the end of the arc portion, ready to commence the next arc portion with the leaves moving in the opposite direction. By aligning the leaves at the edge of the field at the beginning and end of the arc portion, the gap between closed leaves can be hidden underneath a collimator jaw, thus reducing the undesirable leakage radiation. However, the effects of minimum leaf gaps and leaf leakage are not explicitly considered in this work.
The dMLC leaf sweep trajectory, also known as “sliding window,” is calculated using the equations provided by Refs. 25, 26, 27, which give the leaf velocity of the leading (νlead) and trailing (νtrail) leaves in terms of the maximum leaf velocity (νmax) and the local fluence gradient . Equations 2 give the leaf velocities and require a constant dose rate (r) over the arc portion.
(2) |
The time for all leaf pairs to traverse the field is given by Eq. 3 and is governed by the width of the field WF and the SPG of the fluence map (which is equal to the SPG of the leaf row with highest SPG) divided by the dose rate r
(3) |
The (·)+ operator is shorthand for max(·, 0). That is, the gradient term is included in the sum only if it is positive.
Each fluence map is locked to a given portion of the gantry rotation arc, so that if the gantry rotation time over the arc portion is less than the leaf travel time required, the gantry speed is reduced. In a similar manner, if the required leaf travel time is less than the gantry rotation time, then the leaf travel time is increased by reducing the dose rate over the arc portion. It should be noted that currently a continuously variable dose rate is assumed, but also that there is no strict requirement that the leaves take the full duration of the time available to complete the fluence modulation. However, it is dosimetrically favorable to reduce the dose rate if this can be accomplished without an effect on delivery time, as this will lead to larger beam apertures.35 All else being equal, larger apertures are preferred due to smaller output factor variations at larger field sizes, making leaf end calibration less critical.
The leaf velocities of the leading and trailing leaves are then assigned to the right and left leaves respectively. The next fluence map is processed in the opposite direction (right to left) and the leading and trailing leaf trajectories assigned to the left and right leaves respectively. This process then repeats for all the fluence maps in the VMAT arc.
Merging neighboring fluence maps
The purpose of the merging algorithm is to lower the beam-on treatment time by reducing the number of distinct fluence maps that need to be delivered. To deliver each fluence map, the leaves must make a full unidirectional sweep across the aperture over the arc portion that the fluence map is specified for. VMAT solutions with a large number of distinct fluence maps thus require the gantry to move slowly in order to give the leaves sufficient time to move across the fields. Our merging algorithm iteratively merges neighboring fluence maps, reducing the total leaf travel distance and allowing the gantry to move more quickly around the full arc.
We begin with 180 optimized fluence maps, which are delivered over the ranges [0,2°], [2°, 4°], …[358°, 360°]. The initial solution is a high-quality treatment plan, and we seek to merge fluence maps in a way that preserves this optimized dose distribution. Our merging strategy is based on the observations that (1) merging fluence maps with the greatest degree of similarity will have the least effect on the final dose distribution, and (2) merging fluence maps with smaller arc portions will have less of an effect on the dose distribution than merging fluence maps defined over longer arc portions.
These two observations allow us to define a similarity metric between any two neighboring fluence maps f1 and f2, with arc portion lengths of θ1 and θ2. The similarity metric δ is defined as the Frobenius norm of the difference between the maps (normalized by their arc portion lengths to make them comparable), and scaled by the combined arc portion length θ1 + θ2
(4) |
Here, i and j are indices over the rows and columns of the fluence maps. We incorporate this similarity metric into a greedy algorithm that merges a single pair of fluence maps with every iteration, such that after n iterations the number of fluence maps is 180 – n. At each step, the neighboring pair with the lowest δ score is selected for merging. The merged fluence map is defined as the sum of the two neighboring fluence maps, with a new arc portion equal to the union of the initial two arc portions. In this merging strategy, each merge effectively averages two fluence maps over their combined arc length.
The merged maps are then sequenced using a left-to-right sweep (or right-to-left, depending on which side the leaves begin on). As the gantry rotates through the arc and the leaves sweep across the field, the fluence from the leaf opening is deposited onto the nearest original 2° fluence map. In this way, we accurately (within 2°) simulate the fluence as it is delivered by the moving gantry and the moving MLCs.
The stopping criterion for the greedy search will depend on the planner’s desired balance between treatment time and plan quality. It is expected that this tradeoff should lean heavily in favor of plan quality, so that the final merged dose distribution should be nearly identical to the original 180 beam VMAT plan. The final plan will be more time efficient than the 180 beam solution but retain optimality to within a small factor epsilon, which can be chosen by the user.
VMAT settings
We use the following VMAT delivery parameters: maximum gantry speed = 1 rotation/min, maximum leaf speed = 2.5 cm/s, maximum dose rate = 600 MU/min. For each case, we design a 2 Gy fraction plan. It is important to note that unlike step and shoot IMRT optimization, where fraction dose scaling does not fundamentally affect the plan, here fraction dose is important since it is linked to dose rate, gantry speed, and leaf speed. For display purposes, we scale the dose-volume-histograms (DVHs) up to the total dose delivered from all the fractions. We also display the optimization formulations for the total dose.
Clinical cases
We apply VMERGE to three different clinical cases: a prostate case (voxel size: 3 × 3 × 2.5 mm3, 1 cm × l cm beamlets) planned to 77 Gy, a pancreas case (voxel size: 2.6 × 2.6 × 2.5 mm3, 1 cm × l cm beamlets) to 50.4 Gy, and a brain case (voxel size: 1.35 × 1.35 × 1.25 mm3, 0.5 cm × 0.5 cm beamlets) to 32 Gy.
We use CERR 3.0 beta 3 (Ref. 28) quadratic infinite beam (QIB) dose computation to calculate the 180 beam dose influence matrix, which takes about 5 min for the prostate case, 10 min for the pancreas case, and 20 min for the brain case. With the dose matrix calculated, dose distribution computation for a given fluence map takes seconds. The 180 beam optimization runs take between 1 and 5 min, depending on the treatment plan complexity and the size of the dose influence matrix.
In prostate treatment, the main dosimetric tradeoff is between the rectum dose and the target coverage. However, for our analysis, we hold the prostate coverage fixed at the prescription level, 77 Gy, and consider the tradeoff between mean dose to the rectum (more precisely, the anterior rectum as contoured by the physician), the bladder, and the unclassified tissue.
Pancreas is a challenging site for VMAT for the same reason that it is challenging for beam angle optimization: the target volume is surrounded by kidneys, stomach, and liver, and the optimal radiation entry directions are not obvious.29
We select the brain case due to the potential of VMAT to treat isolated metastatic lesions in a single gantry revolution. Minimizing treatment time in these cases is important because of the precise set up required for such treatments. In this case there are two adjacent lesions, one of them being inside the brainstem, and the prescription to both is 32 Gy. For the optimization, we consider the targets as a single combined structure.
Prostate formulation
(5) |
where u.t. stands for unclassified tissue: all the voxels not belonging to any other structure.
Pancreas formulation
(6) |
The shell is a 0.7 cm band around the target used to promote dose conformality to the target.
Brain formulation
(7) |
Sensitivity to algorithm settings
For the prostate case, we investigate some variations to the algorithm. The first is smoothing, discussed in Sec. 2A1, where we compare no smoothing, max beamlet smoothing, and SPG smoothing. We also investigate how the final VMAT solution depends on beamlet size and beam angle spacing, in order to show that our solution technique yields a fundamentally correct VMAT plan and not one that is sensitive to parameter choices. Since most commercial VMAT solutions calculate the final dose on a 2° gantry spacing, we choose to use this as our baseline angular spacing grid. We examine the nature of the VMAT solution that arises when this grid is coarsened to 4° (thus, we start by solving a 90 beam IMRT problem). We then take this angular grid and further investigate shrinking the beamlet size by a factor of 2 in the leaf travel direction (creating 0.5 cm × l cm beamlets).
RESULTS
Prostate case
We navigate to a solution with (mean rectum dose, mean bladder dose, mean u.t.) = (0.50, 0.50, 0.12)*77 Gy. Plan quality is assessed using the mean dose to the anterior rectum, femoral heads, and bladder, the standard error from the prescription level of the dose to the prostate, and the volume of the prostate receiving the prescription dose.
Figure 1 shows the results of VMERGE for the prostate case. The plan does not degrade until after 145 iterations, where we begin to lose target coverage. Therefore, we select a plan requiring 35 arc portions and a treatment time of 311 s (original time of 906 s with 180 arc portions) as having the optimal tradeoff between treatment time and plan quality. This tradeoff between quality and time is shown for both the target [Fig. 1a] and three key organs-at-risk [Fig. 1b]. The DVH comparing the original plan (solid line) to the simplified plan (dashed line) is shown in Fig. 1c. The simplified plan has (mean rectum dose, mean bladder dose) = (0.50, 0.50)*77 Gy, with 95.3% of the prostate volume receiving the full dose (V77 Gy). Femoral head constraints are easily maintained. The plot of the arc portions in Fig. 1d shows that the gantry speed varies during the course of the single arc. Final arc portion sizes range from 4° to 22°. Smaller arc portions require the gantry to slow down to allow the leaves time to traverse the aperture and represent the fluence maps that are most dissimilar to their neighbors.
Sensitivity: Fluence map smoothing
For the prostate case, the final treatment delivery time for no smoothing, max beamlet smoothing, and SPG smoothing, are 413, 339, and 311 s, respectively. Figure 2a shows the DVH plots for the final merged plans for these three smoothing methods. Each plan is created by merging from an initial 180 beam solution and represents the point on the tradeoff curve with the best compromise between plan quality and treatment time. The three plans are highly similar, differing only slightly in the dose to the femoral heads, which is simply reflective of the original plan (since all plans easily satisfy the femoral head constraint of 45 Gy, feasible plans exist with different femoral head doses; in this case the initial 180 beam “no smoothing” plan happened to have a larger femoral head dose). Because SPG smoothing offers a significant improvement on the delivery time, we use it for all of the nominal cases (Figs. 1,3, and 4).
Sensitivity: Discretization settings
Figure 2b shows DVH plots for final plans created by iterative merging from an initial 180 beam solution (1 cm × l cm beamlet), 90 beam solution (1 cm × l cm), and 90 beam solution with reduced beamlet sizes (0.5 cm × l cm). The final plans have similar treatment times and target coverage. In the optimization with the smaller beamlets, we reduce the rectum dose down as much as possible. As expected, finer beamlet resolution yields better dose distribution shaping (the mean anterior rectum dose is reduced from 50% to 44% of the prescription dose), which indicates the value of reducing the beamlet size (for a theoretical analysis of the effect of beamlet size, see Ref. 30). But the stability and usefulness of the algorithm is unchanged by beamlet size. To ease the computational burden we remain with the 1 cm2 beamlets for the pancreas case. We switch to 0.5 cm2 beamlets for the brain case due to its overall smaller target. This beamlet resolution is available in, for example, the Varian Millennium 120 MLC.
Pancreas case
For the pancreas case, we navigate to a solution with (mean shell dose, mean kidneys dose, mean liver dose, mean stomach dose) = (0.93, 0.24, 0.22, 0.21)*50.4 Gy.
Figure 3 shows the VMERGE results. Plan quality begins to degrade after 160 merges, after which the mean dose to the kidneys, liver and stomach begin to increase. The selected plan requires 20 fluence maps and 202 s (initial time of 859 s) and reflects nearly the same DVH as the original 180 beam solution. Final arc portion sizes range from 8° to 28°. The selected plan has (mean kidneys dose, mean liver dose, mean stomach dose) = (0.24, 0.22, 0.21)*50.4 Gy, with 98.3% of the tumor volume receiving the full dose (V50.4Gy).
Brain case
For the brain case, we navigate to a solution with (mean chiasm dose, mean brainstem dose, mean u.t. dose) = (0.15, 0.9, 0.44)*32 Gy and all target voxels receiving 32 Gy or higher.
Figure 4 shows the VMERGE results. Plan quality begins to degrade after 155 merges, after which the mean dose to the optic chiasm begins to rise. The selected plan requires 25 fluence maps and 228 s (initial time of 522 s) and reflects nearly the same DVH as the original 180 beam solution. The selected plan has (mean chasm dose, mean brainstem dose) = (0.17, 0.90)*32 Gy, with 97.6% of the target volume receiving the full dose (V32 Gy). Figure 4d shows that from the anterior and posterior angles, finer sampling is required to shape the dose around the optic chiasm. The smallest arc portion sizes are 4°. Maximal gantry speeds are achieved near the lateral angles, where the radiation does not have to conform around an OAR, and the largest arc portion is 50°.
DISCUSSION AND CONCLUSIONS
VMERGE is designed to intrinsically address two closely related VMAT issues: (1) how much fluence modulation is needed as the gantry proceeds around the patient, and (2) with what frequency do the fluence maps need to be delivered. The second point is an important one in the context of minimizing delivery time in the finite leaf speed setting. Finite leaf speed is the single parameter that makes the single arc coplanar VMAT optimization problem so challenging. As leaf speed approaches infinity, the complexity and delivery time of a plan are governed solely by the SPG of the fluence maps, and this quantity can be minimized exactly in a convex optimization setting.7 However, in the finite leaf speed setting, for fields even with low SPG the gantry may have to slow down just to give the leaves time to travel across the field. This information cannot be represented in a convex optimization setting. Similar to the decomposition of IMRT planning into a convex fluence map optimization step and then a leaf sequencing step, we show that we can decompose the VMAT problem into two sequential stages, with the additional challenge arising from the continuous motion of the leaves and the gantry.
We adopt an approach of starting with a fine solution and gradually coarsening it. By successively coarsening, we decrease the delivery time while maintaining a good dose distribution. Given current computing capacity and tailored algorithms to solve the IMRT problem (e.g., Refs. 20, 31), it is not difficult to solve a 180 beam IMRT optimization problem. Our method of starting from the fine ideal solution and then making it deliverable contrasts with all of the other VMAT approaches which start from a coarse solution and then add segments to improve the dose quality.9, 11 Another way to view the difference is: we start at the global minimum of a relaxed convex problem and then modify that solution to make it deliverable while staying as close as desired to the ideal plan, while the coarse-to-fine approaches first employ a global search and then use local refinements to optimize further, but do not yield any guarantees on optimality. Such guarantees are important both for high quality patient care and for proper comparisons of treatment modalities, comparisons which can easily be muddled when using planning systems with no optimality guarantees.
We study two approaches to create initially smoother 180 beam IMRT fluence maps. Neither of the two methods has a dramatic effect on the final dose distribution, but both offer improvements of the final treatment time. Max beamlet smoothing reduces treatment time from 413 s to 339 s (i.e., by 74 s, an 18% reduction), and SPG smoothing by another 28 s. Even without smoothing, the merging routine combines neighboring fluence maps, so noise in the maps that is not dosimetrically meaningful or useful will tend to get washed out. Still, having initially smooth maps allows the merging routine to merge down to a solution with improved delivery time. For practical implementation, max beamlet smoothing is trivial to implement in the projection solver. SPG smoothing on the other hand, while able to be written as a convex minimization problem, is implemented heuristically as described above. An exact minimization of SPG can be done by introducing auxiliary variables,7 but this has not been pursued at this time. However, due to the usefulness of SPG smoothing, which is even greater for higher dose-per-fraction plans (where the SPG term in Eq. 3 becomes more dominant), it is prudent to continue to investigate how to efficiently minimize SPG for large IMRT problem instances.
Using a finely spaced angular grid of beams and an IMRT solver to compute optimal fluence maps, one may observe that neighboring maps are highly similar. Indeed, in symmetric cases like a donut-shaped target with a central circular organ at risk, the optimal maps from each angle are identical.36 And yet, even with 180 identical fluence maps, one does not know from this information how frequently those modulated fields need to be formed by the MLC leaves. VMERGE resolves this by successively merging similar neighboring maps for as long as the plan quality is maintained. Even with our simple greedy selection strategy (merging the current two neighboring fluence maps with the best similarity score and employing no backtracking), we see that we can merge many of the maps and not change the plan perceptibly. A somewhat surprising result of our analyses, however, is the large number of adjacent merges that can be done and still yield a good dose distribution. Examining Figs. 1d, 4d, one can observe that the largest angular sweeps are generally the sweeps which avoid the critical structures. In that case, any fluence modulation is due to target coverage rather than OAR sparing, so the relevant voxels of the dose-influence matrix become the target voxels. Since target voxels are generally closer to isocenter than OAR voxels, what can be considered a “small angle” is relatively larger for target voxels, which offers some justification for the large arc portions. For the pancreas case, Fig. 3, the OARs are spread evenly around the target, so the final arc portions are more uniform in size. It remains to be investigated if something other than the greedy merging strategy, and/or merging based on dose distribution similarities rather than fluence map similarities, could be more successful.
VMERGE is a two-stage method where the user first explores the dosimetric tradeoffs exactly as done for IMRT MCO, and then either presses a button which does automatic merging to a prespecified epsilon dose deviation threshold or uses the merging routine to interactively explore the tradeoff between dose quality and delivery time, i.e., the information displayed in Figs. 1, 3, and 4. To create these figures, we run the merging algorithm until the gantry is moving at maximum speed for the entire rotation, in order to show the entire plan quality-delivery time tradeoff curve. This merging takes on the order of 5 min for each of the cases (this includes computing the dose and plan evaluation metrics after each merge; if you knew a priori how many merges you wanted, the merging would take on the order of 10 s). The planning time needed for a complete VMERGE planning session can be estimated as follows: dose-influence matrix calculation (10 min) + computation of Pareto surface plans (15 min) + navigation (5 min) + merging and sequencing (5 min), which totals to 35 min. Dose-influence matrix calculation and Pareto surface computation can each be parallelized, which would then speed the whole procedure to around 15 min. While a direct comparison with commercial VMAT planning time is beyond the scope of this report, the literature indicates times in the range of 20–60 min.32, 3
Although we used fairly simple formulations for the initial 180 beam IMRT solution, the method proposed herein does not preclude the use of more sophisticated formulations. For example, quadratic penalty formulations, dose-volume constraints, equivalent uniform dose, and biological objective functions could all be used (provided that memory and time efficient implementations exist for solving these other formulations). Also, there is flexibility in the sequencing/merging step, and any number and type of user-defined cutoff values could be defined to determine what solution along the delivery time/plan quality curve should be delivered. We have also used the simple pencil beam decomposition approach for the IMRT and VMAT problem. The dose distribution produced by such “Dij” approaches, in naive implementations, can degrade significantly when final dose calculations—which include output factor corrections and leaf transmissions—are performed. However, we have proceeded with the Dij approach due to its mathematical tractability and because it is possible to successfully include delivery effects into Dij-based approaches.33
Because we allow the gantry to slow down as much as possible to deliver the required fluence patterns, there can be no dosimetric advantages to double arc (or more) solutions. The only possible advantage for multiple arcs is treatment time. Double arc solutions will be treatment time superior when most of the fluence maps are highly modulated large fields which can be delivered quicker in two sweeps, with the leaves reset on the second sweep to be in favorable positions. For example, a double hump fluence map, with the humps separated by a wide zero fluence section, would be faster with a two arc approach if the first arc delivered the first hump and the leaves could be positioned correctly during the second arc to deliver the second hump. To make a double arc plan overall faster in treatment delivery, one would need a large number of the fluence maps to be of this nature. We speculate that such situations will not arise often clinically, and that therefore single arc solutions, with good optimizers, will typically be the right choice. On the other hand, a partial arc solution could save treatment time by allowing the gantry to spend more time at angles that are more beneficial for radiation delivery. A partial arc solution could be formed by prescribing an arc based on user experience, or by observing the original 180 beam solution and deciding to eliminate an angle sector which has generally low fluence. Once an arc is chosen, the multicriteria dose optimization and subsequent merging steps would proceed exactly as specified in this report.
Other extensions to the proposed approach include collimator rotation, couch rotation, and non unidirectional leaf sequencing. There could be a reduction in treatment time by aligning the collimator to the fluence maps in such a way to minimize fluence map delivery time [Eq. 3]. The collimator angle trajectory could be determined by examining the fluence maps of the 180 beam IMRT solution with this metric, and therefore seems not an overly difficult extension of the approach, although one would need to implement a method for forming fluence maps with a rotating MLC collimator. Couch rotation on the other hand, which leads to noncoplanar arcs, leads to a much larger optimization space and is accordingly a much more challenging problem. Finally, we have considered only unidirectional leaf sequencing in this paper. It is possible that more general leaf sequencing would allow one to deliver the merged maps more quickly. However, for fluence rows with multiple ups-and-downs, one always needs to sweep both leaves across almost the entire row to achieve inner row modulations. For this reason, we do not believe there will be dramatic gains in switching to more general leaf sequencing, but we do plan on investigating this.
While interesting algorithmic challenges remain for VMAT (non coplanar arcs, dynamic collimator rotations, optimal partial arc creation, and integrating the MCO dose exploration with the delivery efficiency tradeoff), we have introduced a method for single arc coplanar VMAT that guarantees delivery of an epsilon-optimal dose distribution. Since VMERGE starts with 180 equi-spaced beams, and then makes it VMAT-deliverable, the method guarantees a provably optimal (up to an arbitrarily small user-specified tolerance) treatment plan, which is something no commercial VMAT planning system currently does.
ACKNOWLEDGMENTS
The project described was supported by Award No. R01CA103904 from the National Cancer Institute. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Cancer Institute or the National Institutes of Health. Thanks to Brian Winey for providing the brain case.
References
- Carol M. P., “Peacock: A system for planning and rotational delivery of intensity modulated fields,” Int. J. Imaging Syst. Technol. 6, 56–61 (1995). 10.1002/ima.v6:1 [DOI] [Google Scholar]
- Yu C., “Intensity-modulated arc therapy with dynamic multileaf collimation: An alternative to tomotherapy,” Phys. Med. Biol. 40(9), 1435–1449 (1995). 10.1088/0031-9155/40/9/004 [DOI] [PubMed] [Google Scholar]
- Oliver M., Ansbacher W., and Beckham W., “Comparing planning time, delivery time and plan quality for IMRT, rapidarc and tomotherapy,” J. Appl. Clin. Med. Phys. 10(4), 117–131 (2009). 10.1120/jacmp.v10i4.3068 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rao M., Yang W., Chen F., Sheng K., Ye J., Mehta V., Shepard D., and Cao D., “Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy,” Med. Phys. 37(3), 1350–1359 (2010). 10.1118/1.3326965 [DOI] [PubMed] [Google Scholar]
- K.-H. Küfer, Scherrer A., Monz M., Alonso F., Trinkaus H., Bortfeld T., and Thieke C., “Intensity modulated radiotherapy—A large scale multi-criteria programming problem,” OR Spectrum 25, 223–249 (2003). 10.1007/s00291-003-0125-7 [DOI] [Google Scholar]
- Craft D., Hong T., Shih H., and Bortfeld T., “Improved planning time and plan quality through multicriteria optimization for intensity-modulated radiotherapy,” Int. J. Radiat. Oncol. Biol. Phys. 82(1), 83–90 (2012). 10.1016/j.ijrobp.2010.12.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Craft D., Süss P., and Bortfeld T., “The tradeoff between treatment plan quality and required number of monitor units in IMRT,” Int. J. Radiat. Oncol. Biol. Phys. 67(5), 1596–1605 (2007). 10.1016/j.ijrobp.2006.11.034 [DOI] [PubMed] [Google Scholar]
- Bedford J. and Warrington A., “Commissioning of volumetric modulated arc therapy (VMAT),” Int. J. Radiat. Oncol. Biol. Phys. 73, 537–545 (2008). 10.1016/j.ijrobp.2008.08.055 [DOI] [PubMed] [Google Scholar]
- Otto K., “Volumetric modulated arc therapy: IMRT in a single gantry arc,” Med. Phys. 35(1), 310–317 (2008). 10.1118/1.2818738 [DOI] [PubMed] [Google Scholar]
- Salter B., Sarkar V., Wang B., Shukla H., Szegedi M., and Rassiah-Szegedi P., “Rotational IMRT delivery using a digital linear accelerator in very high dose rate ‘burst mode’,” Phys. Med. Biol. 56(7), 1931–1946 (2011). 10.1088/0031-9155/56/7/002 [DOI] [PubMed] [Google Scholar]
- Yu C. and Tang G., “Intensity-modulated arc therapy: Principles, technologies and clinical implementation,” Phys. Med. Biol. 56(5), R31–R54 (2011). 10.1088/0031-9155/56/12/001 [DOI] [PubMed] [Google Scholar]
- Bzdusek K., Friberger H., Eriksson K., Hrdemark B., Robinson D., and Kaus M., “Development and evaluation of an efficient approach to volumetric arc therapy planning,” Med. Phys. 36(6), 2328–2339 (2009). 10.1118/1.3132234 [DOI] [PubMed] [Google Scholar]
- Wang C., Luan S., Tang G., Chen D., Earl M., and Yu C., “Arc-modulated radiation therapy (AMRT): A single-arc form of intensity-modulated arc therapy,” Phys. Med. Biol. 53(22), 6291–6303 (2008). 10.1088/0031-9155/53/22/002 [DOI] [PubMed] [Google Scholar]
- Luan S., Wang C., Cao D., Chen D., Shepard D., and Yu C., “Leaf-sequencing for intensity-modulated arc therapy using graph algorithms,” Med. Phys. 35(1), 61–69 (2008). 10.1118/1.2818731 [DOI] [PubMed] [Google Scholar]
- Monz M., Küfer K.-H., Bortfeld T., and Thieke C., “Pareto navigation: Algorithmic foundation of interactive multi-criteria IMRT planning,” Phys. Med. Biol. 53(4), 985–998 (2008). 10.1088/0031-9155/53/4/011 [DOI] [PubMed] [Google Scholar]
- Thieke C, Küfer K.-H., Mon M., Scherrer A., Alonso F., Oelfke U., Huber P., Debus J., and Bortfeld T., “A new concept for interactive radiotherapy planning with multicriteria optimization: First clinical evaluation,” Radiother. Oncol. 85(2), 292–298 (2007). 10.1016/j.radonc.2007.06.020 [DOI] [PubMed] [Google Scholar]
- Romeijn H., Dempsey J., and Li J., “A unifying framework for multi-criteria fluence map optimization models,” Phys. Med. Biol. 49, 1991–2013 (2004). 10.1088/0031-9155/49/10/011 [DOI] [PubMed] [Google Scholar]
- Craft D., Halabi T., and Bortfeld T., “Exploration of tradeoffs in intensity-modulated radiotherapy,” Phys. Med. Biol. 50(24), 5857–5868 (2005). 10.1088/0031-9155/50/24/007 [DOI] [PubMed] [Google Scholar]
- Hong T., Craft D., Carlsson F., and Bortfeld T., “Multicriteria optimization in intensity-modulated radiation therapy treatment planning for locally advanced cancer of the pancreatic head,” Int. J. Radiat. Oncol. Biol. Phys. 72(4), 1208–1214 (2008). 10.1016/j.ijrobp.2008.07.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chen W., Craft D., Madden T., Zhang K., Kooy H., and Herman G., “A fast optimization algorithm for multi-criteria intensity modulated proton therapy planning,” Med. Phys. 37(9), 4938–4935 (2010). 10.1118/1.3481566 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bokrantz R. and Forsgren A., Royal Institute of Technology, Technical Report No. TRITA-MAT-2011-OS3, 2011.
- Rennen G., Van Dam E., and Den Hertog D., “Enhancement of sandwich algorithms for approximating higher dimensional convex Pareto sets,” INFORMS Journal on Computing 23, 493–517 (2011). [Google Scholar]
- Craft D., Halabi T., Shih H., and Bortfeld T., “Approximating convex Pareto surfaces in multi-objective radiotherapy planning,” Med. Phys. 33(9), 3399–3407 (2006). 10.1118/1.2335486 [DOI] [PubMed] [Google Scholar]
- Herman G., Fundamentals of Computerized Tomography: Image Reconstruction from Projections (Springer-Verlag, London, 2009). [Google Scholar]
- Svensson R., Kallman P., and Brahme A., “An analytical solution for the dynamic control of multileaf collimators,” Phys. Med. Biol. 39, 37–61 (1994). 10.1088/0031-9155/39/1/003 [DOI] [PubMed] [Google Scholar]
- Stein J., Bortfeld T., B. Dörschel, and Schlegel W., “Dynamic x-ray compensation for conformal radiotherapy by means of multi-leaf collimation,” Radiother. Oncol. 32, 163–173 (1994). 10.1016/0167-8140(94)90103-1 [DOI] [PubMed] [Google Scholar]
- Spirou S. V. and Chui C. S., “Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimators,” Med. Phys. 21(7), 1031–1041 (1994). 10.1118/1.597345 [DOI] [PubMed] [Google Scholar]
- Deasy J., Blanco A., and Clark V., “Cerr: A computational environment for radiotherapy research,” Med. Phys. 30(5), 979–985 (2003). 10.1118/1.1568978 [DOI] [PubMed] [Google Scholar]
- Woudstra E., “Beam orientation selection in radiotherapy treatment planning,” Ph.D. thesis, Technical University Delft, 2006. [Google Scholar]
- Bortfeld T., Oelfke U., and Nill S., “What is the optimum leaf width of a multileaf collimator?,” Med. Phys. 27(11), 2494–2502 (2000). 10.1118/1.1319524 [DOI] [PubMed] [Google Scholar]
- Men C., Gu X., Choi D., Majumdar A., Zheng Z., Mueller K., and Jiang S., “Gpu-based ultrafast IMRT plan optimization,” Phys. Med. Biol. 54(21), 6565–6573 (2009). 10.1088/0031-9155/54/21/008 [DOI] [PubMed] [Google Scholar]
- Zhang P., Happersett L., Hunt M., Jackson A., Zelefsky M., and Mageras G., “Volumetric modulated arc therapy: Planning and evaluation for prostate cancer cases,” Int. J. Radiat. Oncol. Biol. Phys. 76(5), 1456–1462 (2009). 10.1016/j.ijrobp.2009.03.033 [DOI] [PubMed] [Google Scholar]
- Jelen U., Sohn M., and Alber M., “A finite size pencil beam for IMRT dose optimization,” Phys. Med. Biol. 50(8), 1747–1766 (2005). 10.1088/0031-9155/50/8/009 [DOI] [PubMed] [Google Scholar]
- In clinically implemented VMAT treatment planning systems, small segments are avoided by delivering treatment with a variable dose rate, which is typically much less than the maximum.
- Consider a uniform field. To deliver this using the sliding window technique when the dose rate is halved, the trailing leaf will wait twice as long before departing, resulting in a larger average distance between the leading and trailing leaves. In general, the fact that smaller dose rates lead to larger average leaf opening sizes can be shown by using Eq. 2 to compute the leaf trajectory curves, and noting that the average opening size is equal to the area between these curves divided by the optimal delivery time, given in Eq. 3.
- In general, adjacent fluence maps of the 180 beam IMRT solution will be similar due to the underlying similarity of the dose-influence matrices, which can be explained by the small angle approximation. More precisely, for dose-influence similarity to lead to fluence map similarity one requires that the optimizer uses the dose-influence matrix in a consistent way. Gradient algorithms, linear programming, and projection solvers are examples of such methods.