Abstract
This paper describes the advantages of using single impacts of large cluster projectiles (e.g. C60 and Au400) for surface mapping and characterization. The analysis of co-emitted time-resolved photon spectra, electron distributions and characteristic secondary ions shows that they can be used as surface fingerprints for target composition, morphology and structure. Photon, electron and secondary ion emission increases with the projectile cluster size and energy. The observed, high abundant secondary ion emission makes cluster projectiles good candidates for surface mapping of atomic and fragment ions (e.g., yield >1 per nominal mass) and molecular ions (e.g., few tens of percent in the 500 < m/z < 1500 range).
Keywords: photon emission, electron emission, secondary ion yield, cluster-SIMS
The performance of SIMS for molecular analysis has been notably enhanced with large cluster projectiles, e.g. C60 and Au400.[1–3] However for surface imaging purposes, the use of cluster beams has some disadvantages when compared to mono-atomic and poly-atomic beams (e.g., Cs, Au1–3, Bi1–3, etc.). For example, a C60 beam is constrained by limitations in source brightness; in the case of massive clusters, e.g. Au400, from liquid metal ion sources (Au-LMIS), their energy and angular emission are incompatible with tight focusing. In the present paper, we describe the potentialities of surface mapping by using spatially and temporally isolated events of individual projectile impacts for the case of C60 or Au400 projectiles. In particular, we present their advantage for surface molecule characterization, via electron emission microscopy (EEM), combined with the detection of secondary ions, SIs, and of photons emitted from a single cluster impact.
An experimental setup that comprises a cluster primary ion beam, an electron emission microscope, a photon detector and a ToF mass spectrometer was used for this study (Figure 1). Two primary ion beams were used: i) massive gold projectiles (e.g., Au400+4) from a Au-LMISource installed in a100 kV Pegase Platform and ii) C60+1,2 from a 15kV in-house-built effusion source.[4, 5] The primary ion projectiles are mass-selected using a Wien filter and focused into the analysis chamber. To achieve the single event analysis mode, the primary ions were pulsed and/or collimated so that experiments were performed at <500 Hz. Emitted photons, electrons and secondary ions were collected per projectile impact. A photomultiplier (PMT, R4220P model from Hamamatsu Photonics), with an active window from 185 to 710 nm and a maximum 22% detection efficiency at 410 nm, was positioned behind the target for photon detection. Electrons emitted per impact were accelerated and then deflected using a weak magnetic field toward an electron emission microscope. Electron images were recorded per projectile impact using a position sensitive detector/fast digital camera, and after processing the images the x-y coordinates of the impact were determined.[6] The secondary ions were detected using a microchannel-plate based multi-anode detector and were stored on a multi-channel time-to-digital converter (TDC). Materials for neat targets were obtained from Sigma Aldrich (St. Louis, MO). Surface target homogeneity was achieved by using vapor and electro-sprayed deposition techniques. For electro-sprayed depositions, samples were dissolved in 50% water/methanol solution. For photon experiments, all samples were deposited onto a 70–100 Ω/sq indium tin oxide coated glass (ITO/glass) substrate from Sigma Aldrich (St. Louis, MO). A sagittal rat brain section was used for secondary ion yield comparison purposes (courtesy of Dr Amina S. Wood, NIDA, National Institute of Health, Baltimore, MD); details on the rat brain section preparation and extraction can be found elsewhere.[7]
Figure 1.
a) Schematics of the instrumental setup used for the detection of co-emitted photons, electrons and secondary ions from a single projectile impact. b) Image of the Pegase 120qkV Platform (Au-LMIS) coupled to the analysis chamber. In b, top) and bottom) insets of electron images as seen in the EEM phosphor screen for a continuous Au400 beam are shown.
Experimental results show that the impact of an energetic C60 or Au400 projectile on a surface target induces a strong electronic excitation of the impacted volume (~103 nm 3) and ionized ejecta. At the single impact level, photon, electron and secondary ion co-emission is observed, where their abundances increase with the projectile energy.[8] Time-resolved photon spectra of a series of targets (e.g., ionic salts, small dyes, small fluorophores and fluorophore labeled proteins) show that in the case of cluster projectiles the photon emission can be described by a narrow Gaussian distribution (few ns width) with a very short time decay in contrast to longer decays observed with monatomic and small polyatomic projectiles.[9] A comparison of the time-resolved photon spectra obtained from C60 and Au400 projectile is shown in Figure 2 for the case of a CsI/ITO target. In the case of C60, we attribute the two fast components to the hollow nature of the projectile, which dissociates after penetrating a few layers and induces collision cascades involving projectile constituents and surface atoms targets. A single component, fast photon emission is observed under massive gold projectile impacts associated to the projectile coherent motion. For the cases studied to date, the photon emission is target specific and increases with the projectile energy.
Figure 2.
Time-resolved photon spectra from a CsI/ITO target bombarded with left) C60 and right) Au400 projectiles.
Multiple electron emissions are observed from a single impact, which enables the determination of the x-y coordinate of the impact site. Electron emission depends on the target composition, morphology and structure. For example, we have recently shown that the number of electrons emitted per C60 impact increases from organics to semiconductors to metals to ionic salts.[5] Moreover, we have also observed differences on the number of electrons from targets of similar composition (e.g., between bulk Al oxide, 50 nm Al oxide particles and 2nm Al bohemite whiskers) and from single layer vs multi-layer nano-structures.[10] The cluster size and energy also affects the number of electrons emitted. As an example, electron distributions from a Glycine target bombarded by individual C60 and Au400 cluster projectiles are shown in Figure 3. The electron distributions follow a Poisson distribution, analogous to those observed from atomic ion bombardment.[11, 12] A significant feature is that in the case of C60 and Au400 cluster bombardment, electron emission is observed where kinetic electron emission from comparable velocity atomic projectiles does not occur.[13, 14] Moreover, due to C60 and Au400 charge states (q=1,2 and q=1–4, respectively), the phenomenon cannot be attributed to potential electron emission mechanism. That is, the unexpected electron emission under individual cluster bombardment is related to a collective effect during the projectile impact. Our results suggest that electronic excitation at the impact site are responsible for the coincidental electron and photon emission, being both characteristic of the target.
Figure 3.
Electron distributions from a Glycine target bombarded by individual C60 and Au400 cluster projectiles.
The unique feature of positional mass spectrometry is the ability to identify ions co-emitted from a single projectile impact. This information is spatially resolved in the size of the area of co-emission. As an example, total secondary ion and specific ion maps are shown in Figure 4 for a CsI coated 400 mesh grid. An important feature of surface mapping is that the number of secondary ion per impact has to be significant. For the case of atomic and poly-atomic ions, due to lower secondary ion yields, this approach is mainly limited to low mass secondary ions (e.g., atomic ions and small fragments). Nevertheless, due to the high desorption yields observed with cluster ions, surface mapping of molecular ions can be attainable from single impacts and from non-damage areas.
Figure 4.
a)Total secondary ion and b) ion specific (CsI)I− maps obtained from a CsI coated 400 mesh grid.
A comparison of secondary ion yields from model and native biological targets can be found in Table 1 for: i) 50 keV Au3+ (equivalent to commercially available 50 keV Bi3+2), ii) C60 and iii) massive Au400+4 projectiles. As a general trend, higher secondary ion emission and reduced fragmentation are observed for C60 and Au400+4 relative to Au3+1 projectiles. Previous studies have shown that for small gold cluster projectiles (e.g., Au3+), the secondary ion emission from organic samples reaches a maximum around 30–40 keV/atom.[15] The analysis of native biological samples shows that 130 keV Au3+ secondary ion yields are one and two orders lowers when compared to 130 keV Au9+ and 520 keV Au400+4, respectively. A similar trend has been observed from biomolecular solids.[16] In the case of cluster projectiles, C60, we have previously shown that the secondary ion yield increases with the projectile energy (15–43 keV).[8] For massive 300–520 keV Au400+4 projectiles, multiple atomic and small fragment ions are observed per projectile impact (yield >1 per nominal mass); moreover, molecular ion yields of few tens of percent are obtained from model targets (single component) and few percents from native biological surfaces in the 500 < m/z < 1500 range.[17, 18]
Table 1.
Absolute secondary ion yields as a function of the projectile size and energy for a series of model biological surfaces (single component) and native biological surfaces (rat brain sections). Data are not corrected for detection efficiency; they are only corrected for transmission efficiency.
Target | m/z | Au3+ 50 keV | C60+2 43 keV | Au400 +4 520keV |
---|---|---|---|---|
CsI target | 127 | 3.84E-01 | 6.56E-01 | 14.59 |
CsI target | 387 | 1.61E-01 | 7.35E-01 | 12.54 |
Model Biological Surfaces | ||||
Glycine CN− | 26 | 1.08E-01 | 2.12E-01 | 4.55 |
Glycine M− | 75 | 8.87E-01 | 4.57E-01 | 3.04 |
PG18:0–18:1 PO3− | 79 | 1.56E-01 | 2.75E-01 | 4.08 |
PG18:0–18:1 M− | 775 | 5.23E-03 | 3.14E-02 | 2.53E-01 |
YGGFL CN− | 26 | 6.64E-02 | - | 5.26 |
YGGFL M− | 554 | 1.08E-02 | - | 5.74E-01 |
Ang III CN− | 26 | 7.85E-02 | - | 4.59 |
Ang III M− | 929 | 1.06E-03 | - | 1.54 |
Native Biological Surfaces | ||||
CN− | 26 | 2.76E-02* | 7.89E-02 | 3.67 |
PO3− | 79 | 2.77E-01* | 2.55E-01 | 3.11 |
Lipid 1 PI38:4/ST24:1 | 885.5 | 7.27E-03* | 7.05E-03 | 1.41E-01 |
Lipid 2 ST24:0(OH) | 906.6 | 4.17E-03* | 3.27E-03 | 1.50E-01 |
data correspond to Au3+ at 130 keV.
These results show the advantage of individual cluster impacts for surface characterization in comparison with atomic and poly-atomic beam techniques. The event-by-event bombardment-detection localization mode offers as a unique feature the combined characterization of a target surface using co-emitted photons, electrons and secondary ions; this information can be later used as a surface fingerprint for target composition, morphology and structure. This approach is promising for molecular ion mapping at the nanoscale level with large applications in materials science, as well as biological and nanotechnology studies.
Acknowledgments
This work was supported by the National Science Foundation (Grant CHE-0750377). F. A. F-L acknowledges the National Institute of Health support (Grant No. 1K99RR030188-01). We will also like to thank Dr. Amina S. Wood for her insights and support.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Winograd N, Postawa Z, Cheng J, Skazal C, Kozole J, Garrison BJ. Improvements in SIMS continue: Is the end in sight? Appl Surf Sci. 2006;252:6836. [Google Scholar]
- 2.Verkhoturov SV, Rickman RD, Guillermier C, Hager GJ, Locklear JE, Schweikert EA. Molecular ion emission from single large cluster impacts. Applied Surface Science. 2006;252:6490. [Google Scholar]
- 3.Tempez A, Schultz JA, Della-Negra S, Depauw J, Jacquet D, Novikov A, Lebeyec Y, Pautrat M, Caroff M, Ugarov M, Bensaoula H, Gonin M, Fuhrer K, Woods A. Orthogonal time-of-flight secondary ion mass spectrometric analysis of peptides using large gold clusters as primary ions Rapid Commun. Mass Spectrom. 2004;18:371. doi: 10.1002/rcm.1342. [DOI] [PubMed] [Google Scholar]
- 4.Della-Negra S, Arianer J, Depauw J, Verkhoturov SV, Schweikert EA. The Pegase project, a new solid surface probe: focused massive cluster ion beams. Surface and Interface Analysis. 2011;43:66. [Google Scholar]
- 5.Verkhoturov SV, Eller MJ, Rickman RD, Della-Negra S, Schweikert EA. Single Impacts of C60 on Solids: Emission of Electrons, Ions and Prospects for Surface Mapping. The Journal of Physical Chemistry C. 2009;114:5637. [Google Scholar]
- 6.Eller MJ, Verkhoturov SV, Della-Negra S, Rickman RD, Schweikert EA. Real-time localization of single C60 impacts with correlated secondary ion detection. Surface and Interface Analysis. 2011;43:484. [Google Scholar]
- 7.Colsch B, Woods AS. Localization and imaging of sialylated glycosphingolipids in brain tissue sections by MALDI mass spectrometry. Glycobiology. 2010;20:661. doi: 10.1093/glycob/cwq031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Fernandez-Lima FA, Eller MJ, Verkhoturov SV, Della-Negra S, Schweikert EA. Photon, Electron, and Secondary Ion Emission from Single C60 keV Impacts. The Journal of Physical Chemistry Letters. 2010;1:3510. doi: 10.1021/jz1014345. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Fernandez-Lima FA, Pinnick VT, Della-Negra S, Schweikert EA. Photon emission from massive projectile impacts on solids. Surface and Interface Analysis. 2011;43:53. doi: 10.1002/sia.3592. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Eller MJ, Verkhoturov SV, Della-Negra S, Schweikert EA. Electron Emission from Hypervelocity C60 Impacts. The Journal of Physical Chemistry C. 2010;114:17191. doi: 10.1021/jz1014345. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Krebs KH. Secondary electron release at metal surfaces due to positive molecular ions. Annalen der Physik. 1964;13:97. [Google Scholar]
- 12.Winter HP, Vana M, Lemell C, Aumayr F. Interaction of slow multicharged ions with solid surfaces: current concepts and new information on slow electron emission. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 1996;115:224. [Google Scholar]
- 13.Brunelle A, Chaurand P, Della-Negra S, Beyec YL, Parilis E. Secondary Electron Emission Yields from a CsI Surface Under Impacts of Large Molecules at Low Velocities (5×103–7×104 ms−1) Rapid Communications in Mass Spectrometry. 1997;11:353. [Google Scholar]
- 14.Aumayr F, Betz G, Mark TD, Scheier P, Winter HP. Electron emission from a clean gold surface bombarded by slow multiply charged fullerenes. Int J Mass Spectrometry. 1998;174:317. [Google Scholar]
- 15.Brunelle A, Della-Negra S, Depauw J, Jacquet D, Le Beyec Y, Pautrat M, Baudin K, Andersen HH. Enhanced secondary-ion emission under gold-cluster bombardment with energies from keV to MeV per atom. Physical Review A. 2001;63:022902. [Google Scholar]
- 16.Novikov A, Caroff M, Della-Negra S, Depauw J, Fallavier M, Le Beyec Y, Pautrat M, Schultz JA, Tempez A, Woods AS. The Aun cluster probe in secondary ion mass spectrometry: Influence of the projectile size and energy on the desorption/ionization rate from biomolecular solids. Rapid Communications in Mass Spectrometry. 2005;19:1851. doi: 10.1002/rcm.1995. [DOI] [PubMed] [Google Scholar]
- 17.Fernandez-Lima FA, DeBord JD, Eller MJ, Verkhoturov SV, Della-Negra S, Schweikert EA. Enhanced Molecular Ion emission using a 100kV Massive Gold Cluster source: Glycerophospholipids and Sphingolipids. International Journal of Mass Spectrometry. (to be submitted) [Google Scholar]
- 18.Fernandez-Lima FA, Post J, DeBord JD, Eller MJ, Verkhoturov SV, Della-Negra S, Woods AS, Schweikert EA. Analysis of native biological surfaces using a 100kV Massive Gold Cluster source. Journal of the American Society for Mass Spectrometry. doi: 10.1021/ac201481r. (to be submitted) [DOI] [PMC free article] [PubMed] [Google Scholar]