Pseudo-Continuous Arterial Spin Labeling Quantifies Relative Cerebral Blood Flow in Acute Stroke

Daymara A Hernandez 1, Reinoud PH Bokkers 2, Raymond V Mirasol 1,4, Marie Luby 1, Erica C Henning 1, José G Merino 1,3, Steven Warach 1, Lawrence L Latour 1
PMCID: PMC3299538  NIHMSID: NIHMS350131  PMID: 22343640

Abstract

Background and Purpose

The aim of this study was to test whether arterial spin labeling (ASL) can detect significant differences in relative cerebral blood flow (rCBF) in the core, mismatch, and reverse-mismatch regions, and whether rCBF values measured by ASL in those areas differ from values obtained using DSC MRI.

Methods

Acute stroke patients were imaged with diffusion (DWI) and perfusion (ASL and DSC) MRI. An expert reader segmented the ischemic lesion on DWI and the DSC time-to-peak (TTP) maps. Three regions were defined: core (DWI+, TTP+), mismatch (DWI−, TTP+) and reverse-mismatch (DWI+, TTP−). For both ASL and DSC, rCBF maps were created with commercially available software, and the ratio was calculated as the mean signal intensity measured on the side of the lesion to that of the homologous region in the contralateral hemisphere. Values obtained from core, mismatch, and reverse-mismatch were used for paired comparison.

Results

Twenty-eight patients were included in the study. The mean age was 65.6 (16.9) years with a median baseline NIHSS of 10 (IQR 4-17). Median time from last known normal to MRI was 5.7 hours (IQR 2.9-22.6). Mean rCBF ratios were significantly higher in the mismatch 0.53 (0.23) versus the core 0.39 (0.33) and reverse-mismatch 0.68 (0.49) versus the core 0.38 (0.35). Differences in rCBF measured with DSC and ASL was not significant.

Conclusions

ASL allows for the measurement of rCBF in the core and mismatch regions. Values in the mismatch were significantly higher than in the core, suggesting there is potential salvageable tissue.

Keywords: acute stroke, cerebral blood flow, perfusion quantification

INTRODUCTION

The primary therapeutic target in patients with acute ischemic stroke is decreased cerebral perfusion, and these areas can be identified with perfusion-weighted (PWI) sequence (1). Diffusion-weighted imaging (DWI) is the most sensitive magnetic resonance imaging (MRI) technique for identifying the acute ischemic core (2). When used together, PWI and DWI can identify brain tissue with hypoperfusion but normal diffusion (i.e. there is PWI-DWI mismatch): this tissue is ischemic but potentially salvageable and therefore a target for reperfusion therapies (2-4).

Dynamic susceptibility contrast (DSC) imaging is the MRI perfusion technique most often used in current clinical practice. DSC MRI tracks the passage of a bolus of gadolinium contrast agent through the brain by acquiring a time-series of fast T2*-weighted images (5). Because it takes hours to clear the contrast agent, repetitive injections to monitor changes in blood flow in response to therapeutic intervention is technically impractical. In addition, gadolinium is contraindicated in patients with renal insufficiency and there are limits to the amount that can be safely administered (6).

Arterial spin labeling (ASL) is a noninvasive alternative for imaging whole brain cerebral perfusion that uses radiofrequency pulses instead of gadolinium to label inflowing arterial blood (7-8). As a complementary method to DWI and DSC MRI, ASL may be used to repetitively and quantitatively monitor changes in cerebral blood flow in regions of ischemic core and mismatch, it can be used to gain new insights into cerebrovascular pathophysiology and the response to therapy.

Recent studies show that ASL depicts perfusion deficits and can qualitatively identify areas of PWI-DWI mismatch comparable to DSC MRI (9-11). Before ASL can be used routinely to monitor longitudinal changes in cerebral blood flow over time, it is important to demonstrate that it can detect regional differences that correspond to the ischemic lesion at a single time-point.

The aim of this study was to test whether ASL could be used to detect significant differences in relative cerebral blood flow (rCBF) in the core, mismatch, and reverse-mismatch regions, and whether rCBF values measured by ASL in those areas differ from values obtained using DSC MRI.

METHODS

Patients

This is a retrospective analysis which is part of an ongoing quality improvement process to provide perfusion imaging to patients in renal failure. The data collection and analysis were done in compliance with the NIH and study site institutional ethics requirements. Patients were included in this study if they had a baseline MRI with interpretable baseline ASL and DSC, a detectable perfusion deficit on ASL as identified by two stroke neurologists blinded to patient information, and a discharge diagnosis of ischemic stroke. If at the time of the initial evaluation the patients were eligible for treatment with standard intravenous tissue plasminogen activator (IV-tPA), the baseline MRI had to be completed within 4.5 hours from the time last known well and before any acute treatment was started. Patients were excluded if they had a carotid occlusion, a posterior circulation infarct, or if they were unable to have a DSC study because their GFR was less than 30 mL/min.

Imaging

All patients were imaged on a clinical 3 Tesla MRI scanner (Achieva, Philips Medical Systems, The Netherlands) equipped with an eight-channel coil and locally developed ASL perfusion imaging software. Imaging was done as part of the standard baseline evaluation of acute stroke and included DWI, ASL and DSC perfusion imaging.

ASL perfusion imaging was performed with a pseudo-continuous labeling sequence according to a previously published protocol (12). The parameters of the pseudo-continuous ASL sequence were: 2.5 minutes duration, 3×3×7 mm resolution, TR/TE = 4000/14 ms, pairs of control/label 12; 20 slices with a 7.0 mm thickness with whole brain coverage, and background suppression which allows to reduce the acquisition time while improving the signal-to-noise ratio (SNR) (13). DSC and DWI were acquired with the vendor’s standard commercially available sequences. The parameters for the DSC sequence were: 1.7 minutes in duration, 3×3×7 mm resolution, gradient echo sequence with a TR/TE 1000/25 ms; 20 slices with a slice thickness of 7.0 mm with a single dose of gadolinium at 5ml/sec. The parameters for the DWI sequence were: 3.0 minutes in duration; diffusion tensor acquisition 15 direction with post-acquisition registration prior to calculation of the trace-weighted DWI, TR/TE 4500/62.1 ms; SENSE factor 1.75; FOV 240×240 mm; 40 slices with a continuous slice thickness of 3.5 mm.

ASL-CBF perfusion maps were calculated from the labeled and unlabeled ASL images with Matlab (The MathWorks, Mass, version 7.5) according to a previously published model (14). The T2* transverse relaxation rate and T1 of arterial blood at 3T were assumed to be 50 ms and 1680 ms, respectively (15-16). The water content of blood was assumed to be 0.76% (10). The labeling efficiency was assumed to be 85% based on numerical simulations of the labeling process comparable to simulations by Wu et al (17).

DSC time-to-peak (TTP) maps were calculated from the DSC sequences with the vendor’s standard perfusion software (Advanced Brain Perfusion, Philips Healthcare, Cleveland, OH) on the MRI console. DSC-CBF perfusion maps were generated in Perfscape™ (Olea Medical). Automated noise reduction was applied to the perfusion images. An automatic arterial input function (AIF) was produced by sampling multiple points, not including the hypoperfused area. The mean AIF was used along with circular deconvolution. The upper limit threshold of the intensity values was reduced to 30000 from 32000 with the commercially available imaging analysis software (MIPAV, Medical Image Processing, Analysis and Visualization, version 4.4, National Institutes of Health, Bethesda, MD) for multiple cases.

Qualitative Image Analysis

Two experienced stroke neurologists blinded to clinical information read the images. Each reader assessed each DWI, ASL-CBF and DSC-TTP by itself in a random order to determine the presence of a lesion on DWI and a perfusion deficit on ASL-CBF and DSC-TTP. In addition, they rated image quality as excellent, good, fair, poor and uninterpretable. A consensus reading was held for disagreements about the presence of perfusion deficits on ASL-CBF. The readers used commercially available software to view the images (MIPAV) and were able to adjust for contrast, color scheme and size.

Quantitative Image Analysis

DWI and DSC-TTP were used to identify the areas of the core (the area where there is diffusion and perfusion deficit [DWI+, TTP+]); mismatch (the area where there is perfusion but no diffusion deficit [DWI−, TTP+]); and reverse-mismatch (the area where there is diffusion but no perfusion deficit [DWI+, TTP−]). The images were co-registered using the following parameters: optimized automatic registration, 6-rigid degrees of freedom, trilinear interpolation, and a correlation ratio cost function. DWI sequences were resampled to the slice thickness of the DSC-TTP and ASL to co-localize them to 20 slices. Mid-sagittal alignment was performed on trace-weighted DWI and used to co-register DSC-TTP, DSC-CBF, and ASL-CBF series data.

Following co-registration an expert reader segmented the ischemic lesion on DWI and the perfusion deficit on DSC-TTP. These areas were overlapped to create 3 volumes of interest (VOI): core, mismatch and reverse-mismatch (Figure 1). Once combined, the 3 VOIs were converted to binary-masks to generate the ipsilateral mask. To generate the contralateral mask, the ipsilateral was cloned and mirrored across midline into the contralateral hemisphere to identify a homologous region of healthy tissue as a control. A binary brain mask was generated for each patient DWI, to reduce contamination in the perfusion data from signal outside of the brain. Then the ipsilateral and contralateral masks were combined with the brain mask to remove regions overlapping the ventricles, sulci, or cerebrospinal fluid, resulting in the final mask with 6 sub-regions. The final mask which contained the 6 sub-regions (core, mismatch and reverse-mismatch on both hemispheres) was converted to the final VOIs. Following segmentation, ASL-CBF images were reviewed and the presence of hyperintensitiy in the sulci, suggestive of delayed arrival of the ASL tag in the arteries, was noted.

Figure 1.

Figure 1

The acute diffusion and perfusion deficits were segmented on DWI and DSC-TTP. The volumes of interest were superimposed to identify the core, mismatch and reverse-mismatch. The regions of interest were flipped across midline to the contralateral hemisphere to identify a homologous region of healthy tissue as a control (a) DWI+, (b) DSC-TTP+, (c, d) core - green (the area where there is diffusion and perfusion deficit [DWI+, TTP+]); mismatch – blue (the area where there is perfusion but no diffusion deficit [DWI−, TTP+]); and reverse-mismatch – red (the area where there is diffusion but no perfusion deficit [DWI+, TTP−]).

The final VOIs were copied to the ASL-CBF and DSC-CBF where the number of voxels, volume, average voxel intensity, standard deviation of intensity and median intensity were measured. The rCBF was calculated as the ratio of the mean signal intensity measured on the side of the lesion to that of the homologous region in the contralateral hemisphere. Because small volumes are prone to measurement error, the quantitative analysis was limited to patients who had core, reverse-mismatch and mismatch volumes ≥2 mL.

Statistical Analysis

SPSS (SPSS Inc., Chicago, Illinois, version 18.0.0) was used for all statistical analyses. Patient demographics and signal intensity values were analyzed with the descriptive statistical analysis tool. Mismatch, reverse-mismatch and core VOIs≥2cc were used for comparisons among the sub-regions in ASL and DSC. In order to determine if the rCBF measured by ASL in the core, mismatch and reverse-mismatch differ from the values obtained using DSC the Wilcoxon Signed Rank test and Paired Sample t-Test were used. All p-values <0.05 were considered statistically significant. The values are expressed as mean (standard deviation [SD]) and median (interquartile range [IQR]) according to the normality distribution values of the Kolmogorov-Smirnov.

RESULTS

Over the seven months of the study (June 2009 - January 2010), 105 consecutive patients had a baseline MRI that included ASL and DSC sequences, and 28 met the inclusion and exclusion criteria. The mean age was 65.6 (16.9), 15 were women, and the median baseline NIHSS score was 10 (IQR 4-17). The median time from last seen normal to MRI was 5.7 hours (IQR 2.9-22.6). Eleven patients (39%) were treated with IV-tPA; in them, the median baseline NIHSS was 11 (IQR 4-15.5); and the median time from last seen normal to baseline MRI was 2.3 hours (IQR 2.1-3.7).

The quality of the ASL-CBF and DSC-TTP images varied. Of the ASL-CBF, 82.1% were good to excellent, 7.1% were fair, and 10.7% were of poor quality. The respective values for the DSC-TTP were 78.6%, 7.1%, and 14.3%. Intraluminal high intensity was observed in 15 patients: seen in regions of reverse-mismatch in 2, mismatch in 12; and core in 1.

Table 1 shows the median diffusion and perfusion lesion volumes on DWI and DSC-TTP maps. Regions of mismatch were found in all 28 patients, core regions were found in 25, and reverse mismatch in 26. The median volumes on ASL for the ipsilateral hemisphere are: reverse-mismatch 2.4 mL; mismatch 42.1 mL; and core 6.0 mL. A total of 23 patients were included in the core and mismatch, 17 patients in the core and reverse-mismatch, and 18 patients in the mismatch and reverse-mismatch analyses for core and mismatch, core and reverse-mismatch, and mismatch and reverse-mismatch ≥2 mL respectively.

Table 1.

Median Lesion Volumes
(mL)
All Patients
(n = 28)
tPA Patients
(n= 11)
DWI 10.8 (2.1-16.2) 12.0 (1.4-21.6)
TTP 55.6 (15.0-133.2) 58.0 (17.1-128.3)
Core 7.0 (1.9-21.0) 9.7 (0.9-23.4)
Mismatch 49.8 (12.7-105.8) 49.3 (17.0-113.0)
Reverse-mismatch 2.4 (0.4-6.7) 2.3 (0.7-6.3)

The rCBF ratios using ASL and DSC are shown in Figure 2 and Figure 3. In the core/mismatch analysis, using ASL the rCBF ratio was 0.39 (0.33) in the area of the core and 0.53 (0.23) in the area of mismatch (p=0.023; mean difference 14.2% [95%, C.I. 0.02-0.26]). Using DSC, the CBF ratio was 0.4 (0.21) in the core and 0.63 (0.18) in the region of mismatch (p<0.001; mean difference 23.6% [95%, C.I. 0.16 to 0.31]).

Figure 2.

Figure 2

ASL and DSC relative cerebral blood flow values of the core versus reverse-mismatch and mismatch respectively shows a close correspondence, and in most part there is roughly a linear relationship between the two techniques.

Figure 3.

Figure 3

ASL and DSC relative cerebral blood flow values of the core, reverse-mismatch and mismatch respectively. The agreement between ASL and DSC is moderate due to motion artifact, small lesions, mis-registration or random error. There are 4 patients in the reverse-mismatch that had a higher rCBF value on ASL than DSC out of which 1 had a very small volume, 1 had poor image quality and 2 may have had a mis-registration or random error. There is 1 patient on the mismatch that showed a higher rCBF value on ASL than DSC because it had a very small volume.

In the core/reverse-mismatch analysis, using ASL the rCBF ratio was 0.38(0.35) in the area of the core and 0.68(0.49) in the area of reverse-mismatch (p = 0.001; mean difference 30% [95%, CI 14.2 to 45.7%]). Using DSC, the CBF ratio was 0.36 (0.22) in the core and 0.60 (0.22) in the region of reverse-mismatch (p = 0.001, mean difference 24.4% [95%, C.I. 0.11 to 0.38]).

In the mismatch/reverse-mismatch analysis, using ASL the rCBF ratio was 0.51 (0.19) in the area of the mismatch and 0.71 (0.54) in the area of reverse-mismatch (p=0.155; mean difference 20.5% [95%, C.I. −0.09-0.50]). Using DSC, the rCBF ratio was 0.64 (0.18) in the mismatch and 0.64 (0.34) in the region of reverse-mismatch (p=0.971; mean difference 0.3% [95%, C.I. −0.18 to 0.19]).

For the 28 patients studied, no significant difference could be detected between rCBF values obtained using ASL from those obtained using DSC in all sub-regions; core (p = 0.545), mismatch (p = 0.711) and reverse-mismatch (p = 0.620).

DISCUSSION

This study is the first to demonstrate that the ischemic core and areas of mismatch and reverse-mismatch can be differentiated quantitatively with ASL; using rCBF values obtained during a baseline routine clinical MRI study that includes a 2.5 minute pseudo-continuous ASL sequence. Our findings suggest that ASL may be used to quantify areas of salvageable tissue and to dynamically monitor regional blood flow responses to a therapeutic intervention. Because ASL does not require the use of an exogenous contrast agent and compliments DWI and DSC MRI, it may be used to repeatedly image acute stroke patients to learn more about stroke pathophysiology and evaluate new treatments.

Validation studies addressing CBF quantification have previously compared ASL to PET, SPECT and DSC in a wide range of neurological and oncological diseases including stroke (18-21). ASL has been compared with DSC to detect perfusion deficits and perfusion/diffusion mismatch in patients with acute stroke (9-11). Using a 6-minute continuous labeling sequence, Chalela et al showed that the quantitative regional CBF values were decreased in the affected hemisphere in comparison to the contralateral tissue in 15 ischemic stroke patients imaged within 24 hours of symptom onset, only one of whom received IV-tPA treatment (10). However, tissue viability depends on the duration of blood flow decline, regional differences in rCBF related to collateral flow, and other factors. For ASL to be practical in the acute setting, the imaging protocol must be brief yet allow quantification of rCBF to identify regions of potentially salvageable tissue. Furthermore, validation studies about the success rate of ASL have obtained equivalent results to those obtained in this study, where only ±10.7% of the images were of poor quality (22-23). In this study, we have demonstrated that it is possible to identify regional differences in rCBF values that correspond to core, mismatch and reverse-mismatch using a brief ASL sequence in patients prior and after administration of IV-tPA.

The effect of reperfusion therapies such as tPA can be monitored with perfusion imaging (24-25). Because ASL is a non-invasive perfusion technique that does not use contrast agents, repetitive perfusion measurements may be performed without the limitation of gadolinium clearance or risk of nephrogenic systemic fibrosis in patients with poor renal function (6). Measuring reperfusion at consecutive time points could allow us to measure the speed and duration of thrombolytic activity or ischemia progression to broaden our understanding of these hemodynamic processes (26). Unfortunately, in comparison to DSC MRI, the perfusion deficit is not as conspicuous on ASL and smaller lesions may be missed (Bokkers RPH et al., unpublished data, 2011). However, ASL can be used in a complimentary fashion to DWI and DSC MRI as we have shown here. Lesion and mismatch may be best defined using DWI and DSC, with ASL providing the mechanism for quantification of rCBF to assess therapeutic response in repetitive measures.

A potential limitation of this study is that the VOIs were defined on DSC-TTP, but used to assess CBF values for both DSC and ASL. Because ASL and DSC are acquired at different times during the exam and later co-registered, a potential exists for mis-registration and differences between localization of the perfusion deficits between the two acquisition techniques. Our results using the TTP maps for lesion depiction show, however, a close correspondence between both techniques within the depicted perfusion deficits, i.e. core and mismatch. Conversely, as a result of the differences between these two modalities, the rCBF values of the reverse-mismatch on ASL show a greater range than on DSC as illustrated in Figure 2. Since reverse-mismatch may represent regions of spontaneous reperfusion and high reactive hyperemia it is possible to obtain values that exceed a value of unity as was the case on ASL but not on DSC. Furthermore, the volumes in the reverse-mismatch VOIs were much smaller than those in the areas of mismatch and core; the volume of reverse-mismatch in our patients was approximately one seventeenth of the mismatch. The regions of reverse-mismatch, therefore, may have greater variability owing to a smaller volume of tissue, contributing to the mean values averaged in the VOI. This results in greater susceptibility of the values to random and systematic error due to mis-registration. A potential solution would be to use a transit time sensitive ASL imaging technique based on multi read-outs after the labeling bolus (27), however it is uncertain with the natural T1 decay of the arterial blood bolus whether these techniques are sensitive to delayed arrival in the flow territory distal to the blockage as blood is recruited through collateral vessels. Additionally, intraluminal high intensity was present in a greater fraction of regions of mismatch versus the reverse-mismatch, possibly due to the effects of delayed transit time. This could be attributed to collateral blood flow trying compensating hypoperfused areas and lead to artifactually high rCBF in ASL of those regions.

A second limitation of this study is that ASL is relatively insensitive for measuring white matter perfusion due to the low perfusion signal, which may hamper CBF quantification within the perfusion deficits as defined with DSC-TTP. To minimize this effect, we employed an ASL method combining pseudo-continuous ASL labeling pulses, background suppression, and an MRI field strength of 3.0 Tesla, which has been shown to be sensitive for white matter perfusion and perfusion deficits (28). A third limitation of this study is the sample size which could have contributed an inability to detect significant differences in rCBF values between ASL and DSC. A fourth limitation of the study is that there were no significant differences between the areas of mismatch and reverse-mismatch. Therefore, even though reverse-mismatch appears normal on TTP there is still a decrease in rCBF that is almost the same as mismatch, and consequently areas of mismatch may not be much lower than normally appearing tissue.

This study demonstrates that pseudo-continuous ASL can be used to quantify CBF values in the ischemic core and in areas with perfusion/diffusion mismatch in patients with hyperacute stroke. There were no rCBF differences between ASL and DSC MRI. As ASL is a quick and non-invasive perfusion technique, it may potentially be used in the future to determine patient eligibility for tPA treatment, clinical trial enrollment, and therapy response.

ACKNOWLEDGEMENTS

The authors thank the NIH Stroke Program at the Washington Hospital Center Stroke Center for their support of the study.

SOURCES OF FUNDING This research was supported by the Division of Intramural Research of the National Institute of Neurological Disorders and Stroke, National Institutes of Health.

Footnotes

CONFLICTS OF INTEREST The authors have no disclosures to report.

Part of this work was presented at the 2011 International Stroke Conference, Los Angeles, CA, USA 2011.

This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

REFERENCES

  • 1.Latchaw RE, Alberts MJ, Lev MH, Connors JJ, Harbaugh RE, Higashida RT, et al. Recommendations for imaging of acute ischemic stroke: A scientific statement from the american heart association. Stroke. 2009;40:3646–3678. doi: 10.1161/STROKEAHA.108.192616. [DOI] [PubMed] [Google Scholar]
  • 2.Warach S, Dashe JF, Edelman RR. Clinical outcome in ischemic stroke predicted by early diffusion-weighted and perfusion magnetic resonance imaging: A preliminary analysis. J Cereb Blood Flow Metab. 1996;16:53–59. doi: 10.1097/00004647-199601000-00006. [DOI] [PubMed] [Google Scholar]
  • 3.Heiss WD, Thiel A, Grond M, Graf R. Which targets are relevant for therapy of acute ischemic stroke? Stroke. 1999;30:1486–1489. doi: 10.1161/01.str.30.7.1486. [DOI] [PubMed] [Google Scholar]
  • 4.Kidwell CS, Alger JR, Saver JL. Evolving paradigms in neuroimaging of the ischemic penumbra. Stroke. 2004;35:2662–2665. doi: 10.1161/01.STR.0000143222.13069.70. [DOI] [PubMed] [Google Scholar]
  • 5.Ostergaard L, Sorensen AG, Kwong KK, Weisskoff RM, Gyldensted C, Rosen BR. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part ii: Experimental comparison and preliminary results. Magn Reson Med. 1996;36:726–736. doi: 10.1002/mrm.1910360511. [DOI] [PubMed] [Google Scholar]
  • 6.ACR Committee on Drugs and Contrast Media: ACR Manual on Contrast Media [Accessed July 14, 2011];American College of Radiology web site. (Version 7). 2010 http://www.acr.org/SecondaryMainMenuCategories/quality_safety/contrast_manual/F ullManual.aspx.
  • 7.Williams DS, Detre JA, Leigh JS, Koretsky AP. Magnetic-resonance-imaging of perfusion using spin inversion of arterial water. Proceedings of the National Academy of Sciences of the United States of America. 1992;89:212–216. doi: 10.1073/pnas.89.1.212. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Detre JA, Leigh JS, Williams DS, Koretsky AP. Perfusion imaging. Magn Reson Med. 1992;23:37–45. doi: 10.1002/mrm.1910230106. [DOI] [PubMed] [Google Scholar]
  • 9.Siewert B, Schlaug G, Edelman RR, Warach S. Comparison of epistar and t2*-weighted gadolinium-enhanced perfusion imaging in patients with acute cerebral ischemia. Neurology. 1997;48:673–679. doi: 10.1212/wnl.48.3.673. [DOI] [PubMed] [Google Scholar]
  • 10.Chalela JA, Alsop DC, Gonzalez-Atavales JB, Maldjian JA, Kasner SE, Detre JA. Magnetic resonance perfusion imaging in acute ischemic stroke using continuous arterial spin labeling. Stroke. 2000;31:680–687. doi: 10.1161/01.str.31.3.680. [DOI] [PubMed] [Google Scholar]
  • 11.Bokkers RP, Warach S, van Osch MJ, Duyn J, Latour LL. Whole-brain arterial spin labeling perfusion mr imaging in patients with acute stroke. Stroke. 2010;41:E273–E273. doi: 10.1161/STROKEAHA.110.589234. Abstract. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Dai W, Garcia D, de Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields. Magn Reson Med. 2008;60:1488–1497. doi: 10.1002/mrm.21790. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Fernandez-Seara MA, Edlow BL, Hoang A, Wang J, Feinberg DA, Detre JA. Minimizing acquisition time of arterial spin labeling at 3t. Magn Reson Med. 2008;59:1467–1471. doi: 10.1002/mrm.21633. [DOI] [PubMed] [Google Scholar]
  • 14.Alsop DC, Detre JA. Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow. J Cereb Blood Flow Metab. 1996;16:1236–1249. doi: 10.1097/00004647-199611000-00019. [DOI] [PubMed] [Google Scholar]
  • 15.Golay X, Petersen ET, Hui F. Pulsed star labeling of arterial regions (pulsar): A robust regional perfusion technique for high field imaging. Magn Reson Med. 2005;53:15–21. doi: 10.1002/mrm.20338. [DOI] [PubMed] [Google Scholar]
  • 16.Lawrence KSS, Wang J. Effects of the apparent transverse relaxation time on cerebral blood flow measurements obtained by arterial spin labeling. Magnetic Resonance in Medicine. 2005;53:425–433. doi: 10.1002/mrm.20364. [DOI] [PubMed] [Google Scholar]
  • 17.Wu WC, Fernandez-Seara M, Detre JA, Wehrli FW, Wang J. A theoretical and experimental investigation of the tagging efficiency of pseudocontinuous arterial spin labeling. Magn Reson Med. 2007;58:1020–1027. doi: 10.1002/mrm.21403. [DOI] [PubMed] [Google Scholar]
  • 18.Noguchi T, Kawashima M, Irie H, Ootsuka T, Nishihara M, Matsushima T, et al. [Accessed July 14, 2011];Arterial spin-labeling mr imaging in moyamoya disease compared with spect imaging. Eur J Radiol. 2011 doi: 10.1016/j.ejrad.2011.01.016. [published online ahead of print February 18, 2011] http://www.sciencedirect.com/science/article/pii/S0720048X11000453. [DOI] [PubMed]
  • 19.Kimura H, Kado H, Koshimoto Y, Tsuchida T, Yonekura Y, Itoh H. Multislice continuous arterial spin-labeled perfusion mri in patients with chronic occlusive cerebrovascular disease: A correlative study with co2 pet validation. J Magn Reson Imaging. 2005;22:189–198. doi: 10.1002/jmri.20382. [DOI] [PubMed] [Google Scholar]
  • 20.Bokkers RP, Bremmer JP, van Berckel BN, Lammertsma AA, Hendrikse J, Pluim JP, et al. Arterial spin labeling perfusion mri at multiple delay times: A correlative study with h(2)(15)o positron emission tomography in patients with symptomatic carotid artery occlusion. J Cereb Blood Flow Metab. 2010;30:222–229. doi: 10.1038/jcbfm.2009.204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Warmuth C, Gunther M, Zimmer C. Quantification of blood flow in brain tumors: Comparison of arterial spin labeling and dynamic susceptibility-weighted contrast-enhanced mr imaging. Radiology. 2003;228:523–532. doi: 10.1148/radiol.2282020409. [DOI] [PubMed] [Google Scholar]
  • 22.Chng SM, Petersen ET, Simine I, Sitoh YY, Lim CCT, Golay X. Territorial arterial spin labeling in the assessment of collateral circulation: Comparison with digital subtraction angiography. Stroke. 2008;39:3248–3254. doi: 10.1161/STROKEAHA.108.520593. [DOI] [PubMed] [Google Scholar]
  • 23.Hendrikse J, Petersen ET, Cheze A, Chng SM, Venketasubramanian N, Golay X. Stroke. 2009;40:1617–1622. doi: 10.1161/STROKEAHA.108.539866. [DOI] [PubMed] [Google Scholar]
  • 24.Kidwell CS, Saver JL, Mattiello J, Starkman S, Vinuela F, Duckwiler G, et al. Thrombolytic reversal of acute human cerebral ischemic injury shown by diffusion/perfusion magnetic resonance imaging. Ann Neurol. 2000;47:462–469. [PubMed] [Google Scholar]
  • 25.Marks MP, Tong DC, Beaulieu C, Albers GW, de Crespigny A, Moseley ME. Evaluation of early reperfusion and iv tpa therapy using diffusion- and perfusion-weighted mri. Neurology. 1999;52:1792–1798. doi: 10.1212/wnl.52.9.1792. [DOI] [PubMed] [Google Scholar]
  • 26.Merino JG, Latour LL, An L, Hsia AW, Kang DW, Warach S. Reperfusion half-life: A novel pharmacodynamic measure of thrombolytic activity. Stroke. 2008;39:2148–2150. doi: 10.1161/STROKEAHA.107.510818. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.MacIntosh BJ, Lindsay AC, Kylintireas I, Kuber W, Günther M, Robson MD, et al. Multiple Inflow Pulsed Arterial Spin-Labeling Reveals Delays in the Arterial Arrival Time in Minor Stroke and Transient Ischemic Attack. AJNR Am J Neuroradiol. 2010;31:1892–1894. doi: 10.3174/ajnr.A2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.van Osch MJ, Teeuwisse WM, van Walderveen MA, Hendrikse J, Kies DA, van Buchem MA. Can arterial spin labeling detect white matter perfusion signal? Magn Reson Med. 2009;62:165–173. doi: 10.1002/mrm.22002. [DOI] [PubMed] [Google Scholar]

RESOURCES