Abstract
Background. The toxicity of bound solutes could be better evaluated if we could adjust the clearance of such solutes independent of unbound solutes. This study assessed whether bound solute clearances can be increased while maintaining urea clearance constant during the extended hours of nocturnal dialysis.
Methods. Nine patients on thrice-weekly nocturnal dialysis underwent two experimental dialysis treatments 1 week apart. The experimental treatments were designed to provide the same urea clearance while providing widely different bound solute clearance. One treatment employed a large dialyzer and high dialyzate flow rate (Qd) of 800 mL/min while blood flow (Qb) was 270 mL/min. The other treatment employed a smaller dialyzer and Qd of 300 mL/min while Qb was 350 mL/min.
Results. Treatment with the large dialyzer and higher Qd greatly increased the clearances of the bound solutes p-cresol sulfate (PCS: 27 ± 9 versus 14 ± 6 mL/min) and indoxyl sulfate (IS: 26 ± 8 versus 14 ± 5 mL/min) without altering the clearance of urea (204 ± 20 versus 193 ± 16 mL/min). Increasing PCS and IS clearances increased the removal of these solutes (PCS: 375 ± 200 versus 207 ± 86 mg/session; IS: 201 ± 137 versus 153 ± 74 mg/session), while urea removal was not different.
Conclusions. The removal of bound solutes can thus be increased by raising the dialyzate flow and dialyzer size above the low levels sufficient to achieve target Kt/Vurea during extended treatment. Selectively increasing the clearance of bound solutes provides a potential means to test their toxicity.
Keywords: bound solutes, hemodialysis, urea modeling, uremia
Introduction
Solutes that bind to plasma proteins may be important uremic toxins [1–3]. The clinical importance of such solutes could be better evaluated if we could selectively adjust their clearance by hemodialysis. The current study evaluated a means to increase the clearance of bound solutes independent of the urea clearance during nocturnal dialysis. The extended hours of nocturnal dialysis allow target Kt/Vurea values to be obtained with lower dialyzate flows and smaller dialyzers than are commonly used for conventional thrice-weekly treatment. Most reports of nocturnal dialysis thus describe treatment with blood flows from 200 to 300 mL/min, dialyzate flows (Qd) from 300 to 500 mL/min and small dialyzers [4–8]. Mathematical models predict that lowering the ratios of the dialyzate flow and dialyzer mass transfer area coefficient to the blood flow (Qb) will reduce the clearance of bound solutes more than the clearance of urea [9, 10]. This suggests that the removal of bound solutes is limited when low dialyzate flows and small dialyzers are employed during the extended hours of nocturnal dialysis. The current study tested the hypothesis that bound solute clearances can be increased without affecting urea clearance by increasing the dialyzate flow and mass transfer area coefficient (KoA) relative to the blood flow during nocturnal dialysis sessions lasting 8 h.
Materials and methods
Studies were performed in nine stable hemodialysis patients receiving in-center nocturnal hemodialysis thrice weekly. Two experimental hemodialysis prescriptions were devised using a previously described mathematical model [9]. The High KoA-Qd prescription included a Rexeed 25S dialyzer (Asahi Kasei Kuraray, Tokyo, Japan), Qd 800 mL/min and Qb 270 mL/min, while the Low KoA-Qd prescription included an F160NR dialyzer (Fresenius Medical Care, Waltham, MA), Qd 300 mL/min and Qb 350 mL/min. The KoAurea values provided by the manufacturers are 1881 mL/min for the Rexeed 25S and 1064 mL/min for the F160NR and both are classified as high-flux dialyzers. Each subject was dialyzed once with each of the two experimental prescriptions in random order at midweek sessions 1 week apart. The dialysis time, ultrafiltration target, heparin therapy and dialyzate remained the same as for the patients’ standard treatment. Plasma samples were obtained pre-, mid-, post- and 1 h post-dialysis. The mid- and post-dialysis plasma samples were collected from the arterial line using the low-blood-flow technique (Qb 50 mL/min) as recommended by Kidney Disease Outcomes Quality Initiative guidelines [11]. Spent dialyzate from the first and second halves of each treatment was collected in separate 208-L barrels. The dialyzate volumes were determined by weight and dialyzate was mixed thoroughly prior to collection of samples for chemical analysis.
Concentrations of p-cresol sulfate (PCS), indoxyl sulfate (IS) and hippurate in plasma and dialyzate were measured by high-performance liquid chromatography as previously described [12]. Dialyzate PCS values were below the limit of quantification in two patients and PCS values for these patients were not included in the analysis. Free unbound solute concentrations were measured in pre-dialysis plasma ultrafiltrates obtained using Nanosep 30K Omega separators (Pall, Ann Arbor, MI) and the unbound fraction was calculated as the ultrafiltrate concentration divided by the total concentration. Urea in plasma and dialyzate was measured using a commercial kit (1770-500; Thermo Electron Corp., Melbourne, Australia). Potassium was measured using an iSTAT-1 (Abbott, Princeton, NJ). Plasma albumin, phosphorus and amino acids were measured in the clinical laboratory.
Clearance values were calculated as the amount collected in the dialyzate divided by the log mean of the total plasma concentrations before and after the dialyzate collection. Reduction ratios were calculated as the difference between the initial and final plasma levels divided by the initial plasma level. Values are expressed as mean ± SD throughout. Statistical comparisons between the high KoA-Qd and low KoA-Qd sessions and between each bound solute and urea were performed using the Wilcoxon signed-rank sum test using SPSS software.
Results
The characteristics of the nine patients are summarized in Table 1. The patients had been maintained on in-center nocturnal hemodialysis for an average of 13 ± 8 months. Their routine prescriptions included blood flow 300 mL/min, dialyzate flow 500 mL/min and F160 or F180 dialyzers (Fresenius Medical Care). The treatment time was 480 min in eight patients and 420 min in one patient and the average monthly urea reduction ratio was 84 ± 4%.
Table 1.
Dialysis vintage (years) | 5 ± 4 |
Nocturnal dialysis vintage (months) | 13 ± 8 |
Qb (mL/min) | 300 |
Qd (mL/min) | 500 |
Dialyzer | F160 or F180 |
Treatment time (min) | 473 ± 20 |
URR (%) | 84 ± 4 |
BMI (kg/m2) | 28 ± 9 |
The treatment parameters for the two experimental hemodialysis sessions (designated High KoA-Qd and Low KoA-Qd) performed 1 week apart in each patient are summarized in Table 2. The High KoA-Qd session was designed to enhance protein-bound solute clearance by employing a greater Qd and larger dialyzer as compared to the Low KoA-Qd session. The dialyzate composition and treatment time remained the same for the two sessions and the values for pre-dialysis weight and ultrafiltration rate were similar.
Table 2.
High KoA-Qd | Low KoA-Qd | |
Qb (mL/min) | 270 | 350 |
Qd (mL/min) | 800 | 300 |
Dialyzer | Rexeed 25S | F160NR |
Treatment time (min) | 473 ± 20 | 473 ± 20 |
Pre-dialysis weight (kg) | 89 ± 35 | 89 ± 35 |
Ultrafiltration rate (mL/h/kg) | 2.4 ± 2.0 | 3.0 ± 2.6 |
Qb, blood flow; Qd, dialyzate flow.
The clearances, reduction ratios, pre-dialysis plasma levels and amounts removed in the dialyzate for each solute are summarized in Table 3. The urea clearances were not significantly different during the High KoA-Qd session and the Low KoA-Qd session (204 ± 20 versus 193 ± 16 mL/min). The urea reduction ratio was slightly higher during the High KoA-Qd session (86 ± 5 versus 82 ± 5%) and the amount of urea removed in the dialyzate was the same during both sessions.
Table 3.
High KoA-Qd | Low KoA-Qd | |
UreaN | ||
Clearance (mL/min) | 204 ± 20 | 193 ± 16 |
Reduction ratio (%) | 86 ± 5a | 82 ± 5 |
Plasma pre (mg/dL) | 43 ± 7 | 47 ± 13 |
Removed in dialyzate (g) | 19 ± 5 | 20 ± 3 |
PCS | ||
Clearance (mL/min) | 27 ± 9a,b | 14 ± 6b |
% Free pre-dialysis | 7.9 ± 3.9 | 7.5 ± 4.7 |
Reduction ratio (%) | 59 ± 8a,b | 41 ± 11b |
Plasma pre (mg/dL) | 4.7 ± 2.1 | 4.4 ± 2.0 |
Removed in dialyzate (mg) | 375 ± 200a | 207 ± 86 |
IS | ||
Clearance (mL/min) | 26 ± 8a,b | 14 ± 5b |
% Free pre-dialysis | 7.1 ± 3.5 | 7.0 ± 3.9 |
Reduction ratio (%) | 66 ± 6a,b | 46 ± 9b |
Plasma pre (mg/dL) | 2.6 ± 1.0 | 2.8 ± 1.3 |
Removed in dialyzate (mg) | 201 ± 137a | 153 ± 74 |
Hippurate | ||
Clearance (mL/min) | 121 ± 14a,b | 83 ± 13b |
% Free pre-dialysis | 52 ± 13 | 52 ± 12 |
Reduction ratio (%) | 89 ± 7a,b | 81 ± 7 |
Plasma pre (mg/dL) | 6.8 ± 6.1 | 6.5 ± 4.7 |
Removed in dialyzate (mg) | 1464 ± 1332 | 1236 ± 913 |
aP < 0.05, High KoA-Qd versus Low KoA-Qd.
bP < 0.05, clearance and reduction ratios for other solutes compared to urea.
As expected, clearances of the tightly bound solutes PCS and IS were much lower than the clearance of urea. And for these solutes, the High KoA-Qd prescription provided clearances that were much greater than those provided by the Low KoA-Qd prescription. For PCS, the average clearance of 27 ± 9 mL/min during the High KoA-Qd session was approximately twice the clearance of 14 ± 6 mL/min during the Low KoA-Qd session and the average ratio of the clearance values in individual subjects was 2.0 ± 0.3. The greater clearance during the High KoA-Qd session was not attributable to lesser protein binding, as the free fractions of PCS were similar for the two sessions (7.9 ± 3.9 and 7.5 ± 4.7%, respectively). The greater PCS clearance during the High KoA-Qd session resulted in a greater PCS reduction ratio. And because the pre-dialysis plasma levels were similar for the two sessions, substantially more PCS was removed with the higher clearance provided by the High KoA-Qd session than with the lower clearance provided by the Low KoA-Qd prescription (375 ± 200 versus 207 ± 86 mg).
The values for IS closely paralleled those for PCS. For IS, the average clearance was 26 ± 8 mL/min during the High KoA-Qd session as compared to 14 ± 5 mL/min during the Low KoA-Qd session and the average ratio of the clearance values in individual subjects was 1.9 ± 0.4. The greater clearance during the High KoA-Qd session was again not attributable to lesser protein binding (free fraction 7.1 ± 3.5 versus 7.0 ± 3.9%) and resulted in a greater IS reduction ratio. As was the case with PCS, average pre-dialysis plasma levels were similar for the two sessions and more IS was removed during the High KoA-Qd session than the Low KoA-Qd session (201 ± 137 versus 153 ± 74 mg). The increases in clearance achieved with the High KoA-Qd session were nearly the same for IS and PCS but the increase in amount of solute removed was less for IS, due largely to small differences in pre-dialysis concentrations of the two solutes at the two sessions.
Values for hippurate, which was ∼50% protein bound, were between those for urea and the more tightly bound solutes PCS and IS. The average clearance of 121 ± 14 mL/min during the High KoA-Qd session was significantly greater than the clearance of 83 ± 13 mL/min during the Low KoA-Qd session and the average ratio of the clearance values in individual subjects was 1.5 ± 0.3. The reduction ratio was 89 ± 7% for the High KoA-Qd session and 81 ± 7% for the Low KoA-Qd session. But while the average pre-dialysis plasma levels of hippurate were similar for both sessions, the variability in levels among individual subjects was such that the increase in clearance and modest increase in reduction ratio achieved during the High KoA-Qd session were not accompanied by a significant increase in hippurate removal.
Solute concentration profiles during the experimental sessions are further depicted in Figure 1. Because the Low KoA-Qd session reduced the levels of PCS and IS by <50%, the reduction ratio for these bound solutes could be increased by the higher clearance provided by the High KoA-Qd session. There was little change in solute levels between the end of treatment and 1 h post-treatment. Statistically significant rebound was observed only for urea after the High KoA-Qd and Low KoA-Qd sessions and for hippurate after the High KoA-Qd session (2 ± 1, 3 ± 2 and 4 ± 4% of the pre-dialysis concentration, respectively). Values for clearance and concentration reduction ratio during the first and second halves of the sessions are summarized in Table 4. For the tightly bound solutes, the clearances were nearly the same during the first and second halves of each session but the concentration reduction ratios were lower during the second half of the session.
Table 4.
High KoA-Qd | Low KoA-Qd | |||
First half | Second half | First half | Second half | |
Urea | ||||
Clearance (mL/min) | 194 ± 21 | 213 ± 30 | 187 ± 18 | 200 ± 20 |
Reduction ratio (%) | 58 ± 7 | 67 ± 7a | 58 ± 4 | 58 ± 11 |
PCS | ||||
Clearance (mL/min) | 28 ± 9 | 26 ± 9 | 14 ± 6 | 14 ± 6 |
Reduction ratio (%) | 44 ± 8 | 28 ± 6a | 27 ± 8 | 19 ± 7a |
IS | ||||
Clearance (mL/min) | 27 ± 9 | 24 ± 8 | 14 ± 5 | 14 ± 6 |
Reduction ratio (%) | 48 ± 8 | 35 ± 7a | 33 ± 6 | 19 ± 7a |
Hippurate | ||||
Clearance (mL/min) | 116 ± 12 | 126 ± 22 | 84 ± 14 | 83 ± 14 |
Reduction ratio (%) | 73 ± 15 | 61 ± 7a | 59 ± 9 | 54 ± 14 |
aP < 0.05 first half versus second half.
Routine plasma chemistry values pre- and post-dialysis are summarized in Table 5. Despite an average net ultrafiltration of ∼2 L, plasma albumin levels did not increase significantly post-dialysis. The reduction in plasma potassium and phosphorus levels was similar for the two sessions. The reduction in total plasma amino acid levels and the amount of amino acids removed in the dialyzate were also similar for the two sessions.
Table 5.
High KoA-Qd | Low KoA-Qd | |
Albumin | ||
Plasma pre (g/dL) | 3.7 ± 0.4 | 3.7 ± 0.3 |
Plasma post (g/dL) | 3.7 ± 0.4 | 3.7 ± 0.4 |
Percent increase (%) | 0 ± 8 | 1 ± 8 |
Potassium | ||
Plasma pre (mmol/L) | 4.4 ± 0.5 | 4.5 ± 0.4 |
Plasma post (mmol/L) | 3.2 ± 0.5 | 3.1 ± 0.4 |
Reduction ratio (%) | 26 ± 13 | 29 ± 8 |
Phosphorus | ||
Plasma pre (mg/dL) | 5.7 ± 1.3 | 6.2 ± 1.3 |
Plasma post (mg/dL) | 1.9 ± 0.6 | 2.3 ± 0.6 |
Reduction ratio (%) | 66 ± 7 | 62 ± 5 |
Total amino acids | ||
Plasma pre (mg/dL) | 30 ± 9 | 33 ± 7 |
Plasma post (mg/dL) | 18 ± 6 | 23 ± 9 |
Reduction ratio (%) | 38 ± 17 | 29 ± 17 |
Removed in dialyzate (g) | 13 ± 4 | 14 ± 4 |
Discussion
The extended treatment time used for nocturnal hemodialysis provides several potential benefits [13]. Early studies showed that home treatments performed five to six nights per week for a total of 30–48 h improved blood pressure control and reduced left ventricular mass [14–16]. These beneficial effects were attributed to improved fluid management. In addition, home nocturnal treatment improved phosphorus control and reduced levels of the low-molecular-weight protein β2-microglobulin [14, 16, 17]. The recently published randomized control trial of frequent nocturnal hemodialysis confirmed these improvements in blood pressure and phosphorus control [18]. More recent studies have shown that in-center nocturnal treatment performed thrice weekly provides similar benefits [4–7, 19].
Because standard Kt/Vurea target values are easily achieved during extended treatment, most nocturnal hemodialysis programs have prescribed lower values for Qb, Qd and KoA than those used for conventional in-center treatment [5–8]. Thus, Lacson et al. [7] reported that treatments with an average Qb of 306 ± 58 mL/min, Qd of 496 ± 118 mL/min and dialyzer surface area of 1.6 ± 0.1 m2 provided an average eKt/Vurea of 2.2 ± 0.6 in patients receiving in-center nocturnal dialysis from a large provider. Most other reports of in-center nocturnal dialysis also describe dialyzate flows and dialyzer sizes ranging from 300 to 500 mL/min and 1.2 to 1.5 m2 though dialyzate flows of 600–750 mL/min and larger dialyzers have been used in some centers [4].
As long as Kt/Vurea is used as a measure of adequacy, it is reasonable to employ submaximal flows and small dialyzers for extended treatments. Little if any benefit was obtained by increasing eKt/Vurea from 1.2 to 1.5 in the HEMO study, and common nocturnal prescriptions provide eKt/V values well >1.5 even though they employ low dialyzate flows and small dialyzers [20]. Moreover, urea reduction ratios during most nocturnal treatments are so high that increasing urea clearance and Kt/Vurea would not cause a notable reduction in the plasma urea levels.
There are numerous solutes, however, that accumulate in kidney failure but do not behave like urea [21–23]. One potentially important group of such solutes includes small solutes that bind to plasma proteins [1, 24]. The dialytic clearance of these solutes is restricted because only the free portion is available for diffusion across the dialysis membrane. Dialysis modeling predicts that the clearance of bound solutes can be increased by increasing KoA and Qd relative to Qb. Increasing the Qd maintains the transmembrane concentration gradient by keeping the dialyzate solute concentration below the free solute concentration in the plasma [10]. Increasing KoA increases transmembrane solute diffusion while the free solute concentration in the plasma limits the transmembrane concentration gradient. These measures can increase clearance for bound solutes like IS and PCS because as free solute diffuses across the membrane, solute is released from albumin to maintain equilibrium between the free and bound portions as governed by the association constant. This tends to maintain the free solute level and allows the solute clearance to rise to values which are much higher than the product of the plasma flow and the free solute concentration at the dialyzer blood inlet. It should be noted that increasing Qd would not be expected to increase the clearance of solutes of large size, for which clearance by conventional dialysis is determined almost exclusively by KoA [25].
During short dialysis sessions, a high Qb is required to achieve target urea clearances and Kt/Vurea values for patients of average size. When Qb is high, the increases in KoA and Qd required to greatly increase bound solute clearances can be obtained only by using two machines to increase Qd above the standard single machine capacity of 800 mL/min and by using two dialyzers to increase KoA [12]. The current study showed that extended treatment provides an opportunity to increase bound solute clearances independent of urea clearance using standard equipment. We used the model described by Walther et al. [9] to develop dialysis prescriptions that would provide the same urea clearance but greatly different bound solute clearances. The High KoA-Qd prescription employed the maximum KoA and Qd values available to us. To keep urea clearance nearly the same, Qb was increased slightly above the level of the patients’ regular prescription for the Low KoA-Qd prescription and reduced slightly below this level for the High KoA-Qd prescription.
The effect of the experimental prescriptions on solute clearances conformed to the predictions of the model. Compared to the Low KoA-Qd prescription, the High KoA-Qd prescription increased clearances of the tightly bound solutes PCS and IS by 105 ± 34 and 88 ± 35%, respectively. The clearance of the less tightly bound solute hippurate increased by 48 ± 31% and the clearance of urea did not increase significantly. Both prescriptions provided more than adequate dialysis by conventional standards.
Increasing solute clearances can provide benefit only if it increases solute removal. For urea, the clearances provided by both our experimental prescriptions and the patients’ usual nocturnal prescriptions achieved a >80% reduction ratio. We could have increased urea clearance by increasing Qb along with KoA and Qd but this would not have notably increased the removal of urea. The hippurate clearance provided by our Low KoA-Qd prescription also achieved a >80% reduction ratio. Even a larger increase in clearance than that provided by our High KoA-Qd prescription could thus not have greatly increased the removal of hippurate. The results for PCS and IS, however, were very different. With the Low KoA-Qd prescription, the reduction ratios for these solutes averaged <50% and the increases in clearance provided by the High KoA-Qd prescription therefore increased solute removal significantly. Of note, the dependence of bound solute clearance on dialyzate flow could account for the recent finding that in contrast to the removal of urea, phosphorus and β2-microglobulin, the removal of bound solutes was not increased when dialyzate flow was reduced while dialysis duration was extended [26, 27].
The amount of a solute removed by dialysis also depends on the solute’s distribution among various body compartments. If PCS and IS are modeled as being removed from a single compartment of constant volume, we would obtain average volumes of distribution for these solutes of 0.13–0.15 L/kg body weight. Meijers et al. [28] found, however, that though clearances of PCS and IS remained constant during extended dialysis sessions, the plasma concentrations of these solutes were reduced by lesser fractions during the second half than the first half of the sessions. The current study confirmed this previous finding, which is inconsistent with solute removal from a single compartment. Further experiments will be required to elucidate the compartmental behavior of PCS and IS.
Increasing clearances could potentially increase the removal of valuable solutes as well as uremic wastes. We restricted the clearance of small unbound solutes by adjusting Qb downward when KoA and Qd were increased. This prevented increased loss of amino acids and presumably of other valuable unbound solutes during the High KoA-Qd session. The High KoA-Qd session could also increase the loss of valuable bound solutes. Valuable bound solutes such as vitamin B12, thyroxine and testosterone are, however, very tightly bound to specific carrier proteins. Significant amounts of these solutes would be removed only if KoA and Qd were increased to much higher values than we achieved.
There is considerable, albeit inconclusive, evidence that protein-bound uremic solutes are toxic [1]. Particularly, strong evidence has been assembled for the toxicity of IS and PCS [2, 3, 29–32]. The current study shows that during extended treatment, the clearances of these solutes can be increased without altering urea clearance and using standard equipment. A particularly large increase in bound solute clearances can be obtained during nocturnal dialysis because target Kt/Vurea values are achieved with relatively low KoA and Qd values. But bound solute clearances could also be increased in other settings in which submaximal values of KoA and Qd are now employed. Submaximal dialyzate flows are regularly prescribed because increasing Qd to >1.2–1.5 times Qb increases the cost of treatment without improving its effectiveness as judged by Kt/Vurea [33, 34]. Many centers outside the USA employ treatments lasting 4 h or longer and obtain target Kt/Vurea values with Qd values ∼500 mL/min. We would predict that increasing Qd from 500 mL/min to 800 mL/min would increase the clearances of PCS and IS by ∼20% with commonly used dialyzers, while increasing the urea clearance by only ∼5%.
Bound solute clearances can also be increased by combining convective with diffusive clearance in hemodiafiltration treatments [35, 36]. High convective flows are required, however, and clearances of unbound solutes are increased along with clearances of bound solutes. The present results offer a means to increase the clearance of bound solutes while holding constant the clearance of urea and other small unbound compounds. This approach could provide a means for further testing the contribution of protein-bound solutes to morbidity in end-stage renal disease.
Acknowledgments
The authors wish to acknowledge the assistance of the staff of the Satellite Redwood City dialysis unit and Asahi for supplying the Rexeed 25S dialyzers. T.L.S. was supported by a National Institutes of Health (NIH) Nephrology Research Training Grant (T32 DK7357). F.J.-G.L. was supported by a National Kidney Foundation Fellowship. Other support was provided by the NIH (R21 DK077326 to T.H.H. and R33 DK071251 and R21DK84439 to T.W.M.).
Conflict of interest statement. None declared.
(See related article by Basile and Lomonte. Kt/V urea does not tell it all. Nephrol Dial Transplant 2012; 27: 1284–1287.)
References
- 1.Jourde-Chiche N, Dou L, Cerini C, et al. Protein-bound toxins—update 2009. Semin Dial. 2009;22:334–339. doi: 10.1111/j.1525-139X.2009.00576.x. [DOI] [PubMed] [Google Scholar]
- 2.Niwa T. Indoxyl sulfate is a nephro-vascular toxin. J Ren Nutr. 2010;20(5 Suppl):S2–S6. doi: 10.1053/j.jrn.2010.05.002. [DOI] [PubMed] [Google Scholar]
- 3.Meijers BK, Van Kerckhoven S, Verbeke K, et al. The uremic retention solute p-cresyl sulfate and markers of endothelial damage. Am J Kidney Dis. 2009;54:891–901. doi: 10.1053/j.ajkd.2009.04.022. [DOI] [PubMed] [Google Scholar]
- 4.Troidle L, Finkelstein F, Hotchkiss M, et al. Enhanced solute removal with intermittent, in-center, 8-hour nocturnal hemodialysis. Hemodial Int. 2009;13:487–491. doi: 10.1111/j.1542-4758.2009.00383.x. [DOI] [PubMed] [Google Scholar]
- 5.Cravedi P, Ruggenenti P, Mingardi G, et al. Thrice-weekly in-center nocturnal hemodialysis: an effective strategy to optimize chronic dialysis therapy. Int J Artif Organs. 2009;32:12–19. doi: 10.1177/039139880903200102. [DOI] [PubMed] [Google Scholar]
- 6.Bugeja A, Dacouris N, Thomas A, et al. In-center nocturnal hemodialysis: another option in the management of chronic kidney disease. Clin J Am Soc Nephrol. 2009;4:778–783. doi: 10.2215/CJN.05221008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Lacson E, Jr., Wang W, Lester K, et al. Outcomes associated with in-center nocturnal hemodialysis from a large multicenter program. Clin J Am Soc Nephrol. 2010;5:220–226. doi: 10.2215/CJN.06070809. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.David S, Kumpers P, Eisenbach GM, et al. Prospective evaluation of an in-centre conversion from conventional haemodialysis to an intensified nocturnal strategy. Nephrol Dial Transplant. 2009;24:2232–2240. doi: 10.1093/ndt/gfp029. [DOI] [PubMed] [Google Scholar]
- 9.Walther JL, Bartlett DW, Chew W, et al. Downloadable computer models for renal replacement therapy. Kidney Int. 2006;69:1056–1063. doi: 10.1038/sj.ki.5000196. [DOI] [PubMed] [Google Scholar]
- 10.Meyer TW, Leeper EC, Bartlett DW, et al. Increasing dialysate flow and dialyzer mass transfer area coefficient to increase the clearance of protein-bound solutes. J Am Soc Nephrol. 2004;15:1927–1935. doi: 10.1097/01.asn.0000131521.62256.f0. [DOI] [PubMed] [Google Scholar]
- 11.Hemodialysis Adequacy 2006 Work Group. Clinical practice guidelines for hemodialysis adequacy, update 2006. Am J Kidney Dis. 2006;48(Suppl 1):S2–S90. doi: 10.1053/j.ajkd.2006.03.051. [DOI] [PubMed] [Google Scholar]
- 12.Luo FJ, Patel KP, Marquez IO, et al. Effect of increasing dialyzer mass transfer area coefficient and dialysate flow on clearance of protein-bound solutes: a pilot crossover trial. Am J Kidney Dis. 2009;53:1042–1049. doi: 10.1053/j.ajkd.2009.01.265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kliger AS. More intensive hemodialysis. Clin J Am Soc Nephrol. 2009;4(Suppl 1):S121–S124. doi: 10.2215/CJN.02920509. [DOI] [PubMed] [Google Scholar]
- 14.Culleton BF, Walsh M, Klarenbach SW, et al. Effect of frequent nocturnal hemodialysis vs conventional hemodialysis on left ventricular mass and quality of life: a randomized controlled trial. JAMA. 2007;298:1291–1299. doi: 10.1001/jama.298.11.1291. [DOI] [PubMed] [Google Scholar]
- 15.Nesrallah G, Suri R, Moist L, et al. Volume control and blood pressure management in patients undergoing quotidian hemodialysis. Am J Kidney Dis. 2003;42(1 Suppl):13–17. doi: 10.1016/s0272-6386(03)00532-8. [DOI] [PubMed] [Google Scholar]
- 16.Bayliss G, Danziger J. Nocturnal versus conventional haemodialysis: some current issues. Nephrol Dial Transplant. 2009;24:3612–3617. doi: 10.1093/ndt/gfp491. [DOI] [PubMed] [Google Scholar]
- 17.Pierratos A. Daily nocturnal home hemodialysis. Kidney Int. 2004;65:1975–1986. doi: 10.1111/j.1523-1755.2004.00603.x. [DOI] [PubMed] [Google Scholar]
- 18.Rocco MV, Lockridge RS, Jr, Beck GJ, et al. The effects of frequent nocturnal home hemodialysis: the Frequent Hemodialysis Network Nocturnal Trial. Kidney Int. 2011; 80: 1080--1091 doi: 10.1038/ki.2011.213. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Troidle L, Hotchkiss M, Finkelstein F. A thrice weekly in-center nocturnal hemodialysis program. Adv Chronic Kidney Dis. 2007;14:244–248. doi: 10.1053/j.ackd.2007.03.002. [DOI] [PubMed] [Google Scholar]
- 20.Eknoyan G, Beck GJ, Cheung AK, et al. Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med. 2002;347:2010–2019. doi: 10.1056/NEJMoa021583. [DOI] [PubMed] [Google Scholar]
- 21.Vanholder R, De Smet R, Glorieux G, et al. Review on uremic toxins: classification, concentration, and interindividual variability. Kidney Int. 2003;63:1934–1943. doi: 10.1046/j.1523-1755.2003.00924.x. [DOI] [PubMed] [Google Scholar]
- 22.Glorieux G, Vanholder R. New uremic toxins—which solutes should be removed? Contrib Nephrol. 2011;168:117–128. doi: 10.1159/000321750. [DOI] [PubMed] [Google Scholar]
- 23.Niwa T. Update of uremic toxin research by mass spectrometry. Mass Spectrom Rev. 2011;30:510–521. doi: 10.1002/mas.20323. [DOI] [PubMed] [Google Scholar]
- 24.Meijers BK, Bammens B, Verbeke K, et al. A review of albumin binding in CKD. Am J Kidney Dis. 2008;51:839–850. doi: 10.1053/j.ajkd.2007.12.035. [DOI] [PubMed] [Google Scholar]
- 25.Bhimani JP, Ouseph R, Ward RA. Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and beta2-microglobulin during clinical haemodialysis. Nephrol Dial Transplant. 2010;25:3990–3995. doi: 10.1093/ndt/gfq326. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Eloot S, Van Biesen W, Dhondt A, et al. Impact of hemodialysis duration on the removal of uremic retention solutes. Kidney Int. 2008;73:765–770. doi: 10.1038/sj.ki.5002750. [DOI] [PubMed] [Google Scholar]
- 27.Basile C, Libutti P, Di Turo AL, et al. Removal of uraemic retention solutes in standard bicarbonate haemodialysis and long-hour slow-flow bicarbonate haemodialysis. Nephrol Dial Transplant. 2011;26:1296–1303. doi: 10.1093/ndt/gfq543. [DOI] [PubMed] [Google Scholar]
- 28.Meijers B, Toussaint ND, Meyer T, et al. Reduction in protein-bound solutes unacceptable as marker of dialysis efficacy during alternate-night nocturnal hemodialysis. Am J Nephrol. 2011;34:226–232. doi: 10.1159/000330176. [DOI] [PubMed] [Google Scholar]
- 29.Schepers E, Meert N, Glorieux G, et al. P-cresylsulphate, the main in vivo metabolite of p-cresol, activates leucocyte free radical production. Nephrol Dial Transplant. 2007;22:592–596. doi: 10.1093/ndt/gfl584. [DOI] [PubMed] [Google Scholar]
- 30.Sirich T, Meyer TW. Indoxyl sulfate: long suspected but not yet proven guilty. Clin J Am Soc Nephrol. 2011;6:3–4. doi: 10.2215/CJN.10141110. [DOI] [PubMed] [Google Scholar]
- 31.Barreto FC, Barreto DV, Liabeuf S, et al. Serum indoxyl sulfate is associated with vascular disease and mortality in chronic kidney disease patients. Clin J Am Soc Nephrol. 2009;4:1551–1558. doi: 10.2215/CJN.03980609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Liabeuf S, Barreto DV, Barreto FC, et al. Free p-cresylsulphate is a predictor of mortality in patients at different stages of chronic kidney disease. Nephrol Dial Transplant. 2010;25:1183–1191. doi: 10.1093/ndt/gfp592. [DOI] [PubMed] [Google Scholar]
- 33.Kult J, Stapf E. Changing emphasis in modern hemodialysis therapies: cost-effectiveness of delivering higher doses of dialysis. Int J Artif Organs. 2007;30:577–582. doi: 10.1177/039139880703000704. [DOI] [PubMed] [Google Scholar]
- 34.Bhimani JP, Ouseph R, Ward RA. Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and beta2-microglobulin during clinical haemodialysis. Nephrol Dial Transplant. 2010;25:3990–3995. doi: 10.1093/ndt/gfq326. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bammens B, Evenepoel P, Verbeke K, et al. Removal of the protein-bound solute p-cresol by convective transport: a randomized crossover study. Am J Kidney Dis. 2004;44:278–285. doi: 10.1053/j.ajkd.2004.04.033. [DOI] [PubMed] [Google Scholar]
- 36.Meert N, Eloot S, Waterloos MA, et al. Effective removal of protein-bound uraemic solutes by different convective strategies: a prospective trial. Nephrol Dial Transplant. 2009;24:562–570. doi: 10.1093/ndt/gfn522. [DOI] [PubMed] [Google Scholar]