Abstract
Diffusion coefficient measurements are important for many biological and material investigations, such as studies of particle dynamics and kinetics, and size determinations. Among current measurement methods, single particle tracking (SPT) offers the unique ability to simultaneously obtain location and diffusion information about a molecule while using only femtomoles of sample. However, the temporal resolution of SPT is limited to seconds for single-color-labeled samples. By directly imaging three-dimensional diffusing fluorescent proteins and studying the widths of their intensity profiles, we were able to determine the proteins' diffusion coefficients using single protein images of submillisecond exposure times. This simple method improves the temporal resolution of diffusion coefficient measurements to submilliseconds, and can be readily applied to a range of particle sizes in SPT investigations and applications in which diffusion coefficient measurements are needed, such as reaction kinetics and particle size determinations.
Introduction
It is important to determine the diffusion coefficients of particles for many biological and material applications, such as single-molecule dynamics studies (1–3), biochemical and pharmaceutical reaction kinetics studies (4,5), and particle size and shape determinations (6). Among current methods for measuring diffusion coefficients, such as NMR (7), dynamic light scattering (8), fluorescence correlation spectroscopy (FCS) (9–11), and fluorescence recovery after photobleaching (FRAP) (12), the technique known as single-particle tracking (SPT) offers the unique ability to determine location and diffusion coefficients simultaneously. This is essential for molecular mechanism investigations in heterogeneous environments such as inside a cell's cytoplasm (13) or flagella (14), a membrane in vivo (15), and on a DNA molecule (1) in vitro. Because of this capability, and the additional advantage that SPT experiments require less than femtomoles of sample, SPT can be a powerful tool for measuring diffusion coefficients in a large number of biological investigations (in vitro and in vivo) in which supplies are scarce.
However, the drawback of using SPT for diffusion coefficient measurements is the low temporal resolution. In single-molecule fluorescence imaging studies, stationary or slowly moving (relative to the data-acquisition timescales) single-molecule intensity profiles are called point spread functions (PSFs), and are fit to Gaussian functions to determine the molecules' localization information. The centroid of the Gaussian function determines the lateral location of the molecule at the time of imaging, and the standard deviation (SD) determines the axial location. In SPT diffusion coefficient measurements, consecutive locations of a single fluorophore are measured, and diffusion coefficients are obtained from mean-square displacement analysis of the particle's single trajectories (1,13,16). This method requires at least 20 consecutive location measurements for each single trajectory. With the current single-photon camera imaging rate of ∼100 frames/s for a finite-sized imaging area, 0.2 s is required, and three-dimensional (3D) diffusion coefficient (D3D) measurements up to order 105 nm2/s have been reported (17). However, this requirement of 0.2 s is too long for diffusion coefficient measurements of fast-moving molecules, such as nanometer-sized proteins that diffuse beyond the typical imaging depth of ∼400 nm of single-molecule imaging microscope setups in <1 ms (a typical 5 nm protein has D3D ≈ 108 nm2/s and diffuses nm in 1 ms). A recently developed SPT method measures D3D up to 1.7 × 107 nm2/s by labeling the particles with two colors (18); however, multicolor labeling may not be feasible for many biological particles of interest, which restricts the applicability of the method.
A SPT method that can determine 3D diffusion coefficients of single-colored nanometer-sized biological entities in their native environment is highly desirable for in vivo and in vitro studies. For the molecule to be captured within the microscope's imaging depth, the imaging time must be <1 ms. Here we report a novel (to our knowledge) method that can be used to determine the diffusion coefficient of nanometer-sized Brownian molecules from the SD values of the molecules' intensity profiles using submillisecond exposure times. This single-image molecular analysis (SIMA) study of dynamic molecules is an extension of our previous stationary molecule investigations (19). In this study, we used enhanced green fluorescent protein (eGFP) as the nanometer-sized fluorescent molecule for measurements and analyses.
Because the imaging times in our method are <1 ms, the temporal resolution of the diffusion coefficient measurements is improved by at least 1000-fold over the minutes-long FCS method (multiple measurements, each ∼20 s long), 200-fold over the 0.2-s-long centroid SPT method, 50-fold over the typically 50-ms-long FRAP method, and 10-fold over the two-color SPT method. Furthermore, the improvement in temporal resolution is achieved without compromising the precision of the D3D measurements, and the single-image nature of the method avoids the photobleaching and limited lifetime photon problems associated with single-molecule fluorescence imaging studies. Below, we describe our measurement method, which relates the SD of a 3D freely diffusing protein's intensity profile to its diffusion coefficient D3D. In a previous study, Schuster et al. (20) used a similar concept to relate slow 2D diffusion coefficients (up to 1.1 × 106 nm2/s with a temporal resolution > 25 ms) to a fluorophore's spot sizes. Here, we extend that study to fast 3D diffusions (D3D up to >108 nm2/s and temporal resolutions < 1 ms). By providing the 2D to 3D modification, and the explicit conditions for measuring particles of different sizes (i.e., the appropriate exposure time for a particular particle size; see Appendix S1 and Appendix S2 in the Supporting Material), our study allows for D3D determination of different-sized particles in their native solvents.
Materials and Methods
Sample preparation and imaging
eGFP molecules (4999-100; BioVision, Mountain View, CA) were diluted in 0.5× TBE buffer (45 mM Tris, 45 mM boric acid, 1 mM EDTA, pH 8.0) to 0.03 nM. For stationary eGFP studies, manufacturer-precleaned fused-silica chips (6W675-575 20C; Hoya, San Jose, CA) were used, and isolated eGFP molecules were adsorbed to surfaces at low concentration. For diffusing eGFP studies, the manufacturer-precleaned fused-silica chips were treated with oxygen plasma for 3 min, rendering them hydrophilic to prevent eGFP adsorption (21). The hydrophilic fused-silica surface can be considered ballistic for the diffusing eGFP molecules in our experiments and simulations. For both studies, a protein solution of 5 μL was sandwiched between the fused-silica surface and an oxygen-plasma-cleaned coverslip (2.2 × 2.2 cm2), resulting in a 10.5-μm-thick water layer. Because the oxygen-plasma-treated fused-silica surface is hydrophilic, the buffer quickly wetted the surface and bubbles were rarely observed. The coverslip edges were then sealed with nail polish to prevent possible stray flow of the buffer due to evaporation.
Single-molecule imaging was performed on a Nikon Eclipse TE2000-S inverted microscope (Nikon, Melville, NY) in combination with a Nikon 100× objective (1.49 N.A., oil immersion). The samples were excited by prism-type total internal reflection fluorescence (TIRF) microscopy with a linearly polarized 488 nm laser line (I70C-SPECTRUM argon/krypton laser; Coherent, Santa Clara, CA) focused on a 40 × 20 μm2 region. The 488 nm line was filtered from the multiline laser emission with the use of polychromatic acousto-optic filters (48062 PCAOM model; NEOS Technologies, Melbourne, FL). The laser excitation was pulsed with an illumination interval of 30 ms for the stationary eGFP molecules shown in Fig. 1 and Fig. S2, and between 0.3 and 1 ms for the diffusing eGFP molecules. The excitation intensities were 2.7 and 3.2 kW/cm2 for the respective stationary eGFP molecules, and 37.5 kW/cm2 for the diffusing molecules. Images were captured by an iXon back-illuminated electron multiplying charge coupled device (EMCCD) camera (DV897ECS-BV; Andor Technology, Belfast, Northern Ireland). An additional 2× expansion lens was placed before the EMCCD, producing a pixel size of 79 nm. The excitation filter was 488 nm/10 nm, and the emission filter was 525 nm/50 nm.
Figure 1.

Comparison of stationary and diffusing eGFP molecules. (A) An image of stationary eGFP molecules adsorbed on a fused-silica surface. Five of the seven molecules have SNR > 2.5. (B) Intensity profiles of the stationary eGFP molecules in panel A in photon counts. (C) Intensity profile (dots) and Gaussian fit (mesh) to the stationary eGFP molecule denoted by arrow in A and B. For this molecule, the SNR is 9.8, sx = 107.2 nm, and sy = 107.9 nm. (D) Diffusing eGFP molecules near a reflective hydrophilic fused-silica surface at 1 ms exposure time. Six of the eight molecules have a SNR > 2.5. The scale bars for A and D are 2 μm. (E) Intensity profiles of the diffusing eGFP molecules in D. (F) Intensity profile (dots) and Gaussian fit (mesh) to the diffusing eGFP molecule denoted by the arrows in D and E. For this molecule, the SNR is 3.5, sx = 202.2 nm, and sy = 192.4 nm. It is clear that the intensity profiles of diffusing molecules are wider (or have larger SDs) than those of stationary molecules.
Data acquisition and selection
We obtained movies by synchronizing the onset of camera exposure with laser illumination for different intervals. The maximum gain level of the camera was used and the data acquisition rate was 1 MHz pixels/s (≈3.3 frames/s). We checked the single-molecule images to ensure that there were no saturations in the intensity profiles. For the defocusing analysis of stationary eGFP molecules, we selected 21 × 21 pixel boxes centered at the molecule by hand using ImageJ (NIH, Bethesda, MD), and used the intensity values for 2D Gaussian fitting. For the diffusing eGFP molecule movies, we selected all visible diffusing eGFP intensity profiles in the peak laser excitation region of 10 × 10 μm2 by hand using 39 × 39 pixel boxes centered at the molecule. The center 25 × 25 pixels of the boxes were used for 2D Gaussian fitting, and the peripheral pixels were used for experimental background analysis.
Before performing the analysis, we converted the camera's intensity count at each pixel in an image into the photon count by using the camera-to-photon count conversion factor calibrated the same day of the measurement, as described in our previous article (22). We obtained the number of detected photons in an image by subtracting the total photon count of the background from the total photon count of the image. The eGFP intensity profiles were fit to a 2D Gaussian function to obtain the SD values of the molecule:
| (1) |
where f0 is the multiplication factor; sx and sy are SDs in the x and y directions, respectively; x0 and y0 are the centroid location of the molecule; and is the mean background offset in photons.
For the defocusing eGFP analysis, we selected 17 adsorbed eGFP molecules with a minimum photon count of 229 and signal/noise ratios (SNRs, ) > 3.75, where I0 is the peak PSF photon count (after subtracting the mean background offset ) and is the background variance in photons. For the diffusing eGFP molecules, we used an SNR of 2.5 as a selection criterion. We did not use PSFs with photon counts of <50 in the analysis. At each exposure time, we acquired 1600 data points from four movies (two acquired at different regions of an imaging chip on the same day, and two acquired from different chips on other days). The numbers of diffusing eGFP data used for the experimental analysis that satisfied the SNR criteria were 419–1066 for the 0.3–1 ms exposure times, respectively.
Diffusing eGFP simulations
We simulated 3D Brownian diffusion eGFP trajectories at a range of exposure times using FCS-determined eGFP D3D = 8.86 × 107 nm2/s and triplet-state statistics. The starting locations of the trajectories followed the distribution function described in Appendix S6. The step sizes in the x, y, and z directions were randomly selected from a Gaussian distribution with a mean of zero and SD of with a step time t0 = 1 μs. Because of the reflective fused-silica-water interface, the simulated z-values were maintained above zero. The number of steps in a simulation was t / t0. At each x, y location in a trajectory, when the molecule was not in a triplet dark state, a Poisson distributed number of photons (Appendix S5) was drawn from a Gaussian PSF spatial distribution with a mean of zero and the corresponding SD value for the axial location (Appendix S4). We added this relative displacement of the photons to the simulated x, y location of the molecule, generating the actual x, y location of the emitted photons at the simulation step.
The simulated photons of each trajectory were binned into 50 × 50 pixels with a pixel size of 79 nm. We then converted the photon count of each pixel into the modified camera count using Eq. 4 of DeSantis et al. (22), with the photon multiplication factor of the camera set at M = 1 to include the camera count variance effect. We generated random background photons at each pixel using the corresponding experimental background distribution functions for the exposure time (22). The final intensity profiles were fit to a 2D Gaussian function to obtain the two SD values for the image. For each SD datum of diffusing eGFP molecules shown in Fig. 5, 1000 independent trajectories were simulated.
Figure 5.

Comparing sx and D3D results. (A) Experimental (circles), simulation (disks), and theoretical calculation (squares) measurements of diffusing eGFP intensity profiles' mean sx versus t. In the experimental and simulation results, the error bars are the SDs of the sx distributions. (B) Experimental D3D calculated from Eq. 5. The error bars are ΔD3D calculated using Eq. 6; the dashed line is the FCS-determined eGFP D3D of 8.86 × 107 nm2/s for comparison.
Results
Figs. 1 and 2 illustrate the principle of this method. In a finite exposure time, the intensity profile of a moving molecule is wider (or more blurry) than that of an immobile molecule. Fig. 1 A shows a 30-ms frame image of stationary eGFP molecules adsorbed on a fused-silica surface, and Fig. 1 D shows a 1-ms frame image of diffusing eGFP molecules near a hydrophilic fused-silica surface (21). These figures clearly show that the diffusing-molecule images are blurry compared with the immobile-molecule images. In Fig. 1, B and E, the intensity profiles of the stationary and diffusing eGFP molecules are plotted, and in Fig. 1, C and F, the respective selected intensity profiles are fitted to Gaussian functions. Although both intensity profiles fit well to a Gaussian function, the width (or SD) of the diffusing protein's intensity profile is larger than that of the stationary protein.
Figure 2.

Simulated image formation and analysis process of a diffusing eGFP molecule. (A) Trajectory of a diffusing eGFP molecule in free solution under TIRF evanescent excitation at the exposure time of 0.6 ms. The data are grayscaled to correspond to the particle's axial locations (Appendix S5). (B) The emitted photons from the trajectory form an intensity profile (top, colored plot), which is then projected onto a 2D camera screen (bottom, black and white image). (C) Gaussian fit (mesh) to the intensity profile of the diffusing eGFP (dots), where sx = 119.4 nm, and sy = 142.2 nm.
In general, the final image of a diffusing molecule, such as those in Fig. 1, is the sum of the emitted photons along its diffusion trajectory projected onto a 2D imaging screen. Fig. 2 A shows a simulated eGFP diffusion trajectory at 0.6-ms exposure time using 0.005 ms steps for clarity. The data are grayscaled to correspond to the particle's axial locations (Appendix S5). The emitted photons, after photon-to-camera count conversion, were projected onto a 2D imaging screen and binned into our camera pixels (each 79 × 79 nm2 in size; Fig. 2 B, bottom, gray image), and the corresponding diffusing eGFP PSF intensity profile was formed in the colored image above. The total photon count of this image was 414. The 2D Gaussian fit to the diffusing eGFP intensity profile is shown in Fig. 2 C, yielding SD values in the x and y directions. The -values presented in this article are the results from fitting to these experimental and simulated PSF data, and were used to quantify the blur of diffusing eGFP molecules and consequently the diffusion coefficient D3D.
To determine D3D from diffusing fluorophore images, we performed experimental measurements, analytical calculations, and simulations. Below, we show that when we checked the experimental results against the theoretical calculation and numerical simulation results, we obtained good agreement, which validates our method of measuring nanometer-sized fluorophore diffusion coefficients.
Fig. 3 A shows representative eGFP images (chosen such that the molecule's respective sx-values were within ±5 nm of the means to the respective diffusing eGFP intensity profile SD distributions in Fig. 3 B) acquired at 0.3, 0.7, and 1 ms exposure times in experimental measurements. As expected, the SD values of these respective single diffusing eGFP molecules increase from 136.4 to 160.9 and 175.5 nm, validating the notion that the SD provides a quantitative measure of the motion-induced blurriness of single fluorophore images.
Figure 3.

Diffusing eGFP images and intensity profile SD distributions at different exposure times. (A) Three representative images showing diffusing eGFP molecules at exposure times of 0.3, 0.7, and 1 ms. The intensity profile SD values increase with the exposure time. The scale bar is 1 μm. (B) EGFP intensity profile SD distributions (normalized by counts for comparison) at the three aforementioned exposure times, showing increasing values of 136.8 ± 27.7 (mean ± SD), 159.0 ± 32.2, and 172.1 ± 34.8 nm, respectively.
In analytical calculations, we deduce an expression relating a diffusing eGFP's SD to D3D. We first project the eGFP PSFs at all focal depths onto a 2D imaging screen, forming an axial-direction-projected PSF f(x,y), and then convolve this projected PSF with the lateral location distribution of the molecule in a trajectory, which we define as a pathway distribution function (PWDFx,y) in the lateral directions g(x,y):
| (2) |
Below, we decompose an eGFP's 3D diffusion process into two components for the sx and D3D calculation: a 1D diffusion along the axial direction, and a 2D diffusion in the lateral direction.
It is known that as the defocusing distance between the fluorophore and the focal plane increases, so does the SD of the PSF. Consequently, to calculate the intensity profile, one must integrate over all axial locations to which the molecule may have traveled during the exposure time to obtain an axial-direction-projected PSF, f(x,y). Because diffusion values in the lateral and axial dimensions are statistically independent of each other, we choose to perform this integration before convolving the resulting PSF with PWDFx,y in the lateral dimensions to obtain the final projected 2D intensity profile of the 3D diffusing molecule on an imaging screen.
In the axial direction, we compute the axial-direction-projected PSF by numerically integrating defocused PSFs through z for all pixelated x, y-values:
| (3) |
where nm2/s is the FCS-determined eGFP diffusion coefficient (Appendix S3); C(z) and are the amplitude and SDs of our imaged, defocused eGFP Gaussian PSFs (Appendix S5), respectively; z0 and are the mean and variance of the diffusing eGFPs' Gaussian PWDFzs (Appendix S7); describes the decaying TIRF evanescent excitation intensity; and the range for the z integration is the imaging depth of 0–400 nm measured from the focal point at the fused-silica surface. The resulting axial-direction-projected PSF f(x,y) remains Gaussian, and the SD is a function of the exposure time t as nm, where 111 nm and 0.0634 are fitted values.
In the lateral directions, we numerically calculate g(x,y) of a freely diffusing eGFP particle by simulations. Fig. 4 A shows nine random PWDFxs at exposure time t = 0.6 ms. Six of the nine PWDFxs have one peak (unipeaked or unimodal) and can be fitted to a Gaussian function with R2 > 0.8. Fig. 4 B shows the SD distribution of PWDFxs, combining the Gaussian fitted SD values for the unipeaked PWDFxs and the numerical particle location distribution SD values for the double-peaked PWDFxs (mean = 96.8 nm). Fig. 4 C shows that when the nine PWDFxs in Fig. 4 A are convolved with single-eGFP PSFs at focus with s0 = 108.2 nm, all convolved PWDFxs fit well to a Gaussian function, and the mean of the SD distribution is 147.1 nm. Therefore, although not all PWDFxs are unipeaked, taken over all, we can view PWDFxs as Gaussian functions with an average t-dependent SD value of . For the 0.6 ms exposure time data, Ax = 0.0882. We found Ax to be insensitive to exposure times < 1 ms (mean Ax = 0.0926).
Figure 4.

Study of the eGFP lateral PWDFxs and their convolution with PSFs. (A) Nine random eGFP PWDFxs at 0.6 ms exposure time and Gaussian fits to the unimodal distributions with R2 > 0.8. (B) The distribution of 1000 PWDFx SDs, fitted with a Gaussian. (C) The nine PWDFxs in (A) convolved with eGFP PSFs at focus with s0 = 108.2 nm. (D) The SD distribution of 1000 PWDFx convolved eGFP PSFs at focus and its Gaussian fit.
Given that f(x,y) (at focus and the axial-direction-projected) and g(x,y) are both Gaussian functions, in the lateral directions their convolution can be described by another Gaussian function with a variance equal to the sum of the two variances. Using the focused eGFP PSFs with s0 = 108.2 nm and PWDFx at 0.6 ms, the lateral direction SD value nm = 145.2 nm, which is very close to the mean SD value of the above PSF-convolved-PWDFxs of 147.1 nm.
Because we have observed that both the axial-direction-projected PSFs and the lateral PWDFx,ys are Gaussian, the final projected intensity profiles' SD of diffusing molecules is
| (4) |
where nm is the SD of the axial-direction-projected PSFs for our experimental parameters, and is the variance of PWDFx,ys with . This relation enables one to determine D3D from the SD of a single-molecule's intensity profile and the exposure time as
| (5) |
When the particle is diffusing in the 2D xy plane or along the 1D x or y axis, the 0.0634 term should be removed.
Using FCS-determined eGFP nm2/s in Eq. 4, we plot the analytical eGFP sx results in Fig. 5 A and compare them with the experimental mean eGFP SD values obtained from Fig. 3 B, where the error bars are the SDs of the eGFP intensity profile SD distributions in Fig. 3 B. The analytical and experimental results show excellent agreement within 0.7 ms. Note that sx starts to deviate from the experimental results at t > 0.8 ms; this is because the exposure time begins to approach the diffraction-limit-determined value for eGFP for this study (Appendix S1).
In simulations of diffusing eGFP intensity profiles (as shown in Fig. 2), we used the FCS-determined D3D. Fig. 5 A juxtaposes the simulated diffusing eGFP SD results with the experimental results; the two mean values and error bars agree at all exposure times (Fig. S4 compares the results at t = 0.6 ms).
To determine the precision of the measured D3D from single eGFP images, we performed an error propagation analysis of using Eq. 5:
| (6) |
where is the SD measurement precision of the single fluorophore's intensity profile (i.e., the experimental error bars in Fig. 5 A) (22). Fig. 5 B compares the experimentally determined D3D and ΔD3D from single diffusing eGFP image SD measurements with the FCS-determined eGFP D3D = 8.86 nm2/s, showing agreement.
At 0.7 ms, ΔD3D = 5.2 × 107 nm2/s for a single eGFP image using both the statistically independent mean sx- and sy-values of = 162.1 nm and = 39.2 nm. It is 57% of the eGFP D3D of 8.86 × 107 nm2/s. Because there are ∼30 molecules in a typical frame image of <1 ms exposure time, the precision of the D3D measurement further improves by times to 10%, which is comparable to the precision of the FCS D3D measurements (11). In spatially restrictive situations, such as in vivo imaging in typically micron-sized cells, where only one image can be obtained at a time, repeated single-image measurements will enable a precise determination of D3D.
Discussion
Although in this study we focused on fast diffusion of nanometer-sized proteins in free solution with nm2/s, our methodology applies to 3D diffusion at all rates. When diffusion coefficients are low for large particles, in a crowded environment, or in viscous solvents (such as in cells (23) or glycerol), the molecule's intensity profile will be more localized. Consequently, one should use longer exposure times to observe noticeable changes in the SD from the stationary values. Appendix S1 and Appendix S2 explain the procedure used to determine the appropriate exposure times for a particle of unknown D3D.
In anisotropic environments where D1D-values along the x and y axes differ, our separate D1D measurements along the two lateral directions allow for such differentiation. If the diffusion coefficient differs along the axial direction, recalculation of Eq. 3 with new Dz- and Az-values will still allow for determination of Dx and Dy using Eq. 5.
In certain cases, the molecules' movements can deviate from 3D unbiased Brownian motion (e.g., directional motion or diffusion with a drift). In future extension studies of these alternative motions using our single-image-based method, investigators should determine and PWDFx,y,z before convolving the axial-direction-projected PSF with PWDFx,y for the final intensity profile. As long as the mean numerical SD of locations in the molecule's trajectory is less than half of the diffraction limit at the exposure time, the projected convolved image of the molecule will be a unimodal intensity profile that can be fitted to a Gaussian function, and the resulting will provide information about the molecule's dynamics.
In summary, we have presented a new (to our knowledge) single-molecule fluorescence image analysis method that measures fast diffusion coefficients with high precision. The experimental setup and data analysis are simple to use with standard microscopy imaging systems, and the method is applicable to a wide range of diffusion coefficient measurements with greatly improved temporal resolution. Applications to basic research and pharmaceutical investigations, such as fast drug screening, can be envisioned.
Supporting Material
References
- 1.Wang Y.M., Austin R.H., Cox E.C. Single molecule measurements of repressor protein 1D diffusion on DNA. Phys. Rev. Lett. 2006;97:048302. doi: 10.1103/PhysRevLett.97.048302. [DOI] [PubMed] [Google Scholar]
- 2.Gorman J., Greene E.C. Visualizing one-dimensional diffusion of proteins along DNA. Nat. Struct. Mol. Biol. 2008;15:768–774. doi: 10.1038/nsmb.1441. [DOI] [PubMed] [Google Scholar]
- 3.Thompson M.A., Lew M.D., Moerner W.E. Localizing and tracking single nanoscale emitters in three dimensions with high spatiotemporal resolution using a double-helix point spread function. Nano Lett. 2010;10:211–218. doi: 10.1021/nl903295p. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Auer M., Moore K.J., Stoeckli K.A. Fluorescence correlation spectroscopy: lead discovery by miniaturized HTS. Drug Discov. Today. 1998;3:457–465. [Google Scholar]
- 5.Ganser A., Roth G., Brock R. Diffusion-driven device for a high-resolution dose-response profiling of combination chemotherapy. Anal. Chem. 2009;81:5233–5240. doi: 10.1021/ac900415s. [DOI] [PubMed] [Google Scholar]
- 6.He L., Niemeyer B. A novel correlation for protein diffusion coefficients based on molecular weight and radius of gyration. Biotechnol. Prog. 2003;19:544–548. doi: 10.1021/bp0256059. [DOI] [PubMed] [Google Scholar]
- 7.Webb G.A. Vol. 34. Royal Society of Chemistry; Cambridge, UK: 2005. (Nuclear Magnetic Resonance). [Google Scholar]
- 8.Berne B., Pecora R. General Publishing Company; Toronto, Ontario, Canada: 2000. Dynamic Light Scattering: With Application to Chemistry, Biology and Physics. [Google Scholar]
- 9.Magde D., Elson E., Webb W.W. Thermodynamic fluctuations in a reacting system - measurement by fluorescence correlation spectroscopy. Phys. Rev. Lett. 1972;29:705–708. [Google Scholar]
- 10.Schenk A., Ivanchenko S., Nienhaus G.U. Photodynamics of red fluorescent proteins studied by fluorescence correlation spectroscopy. Biophys. J. 2004;86:384–394. doi: 10.1016/S0006-3495(04)74114-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Petrásek Z., Schwille P. Precise measurement of diffusion coefficients using scanning fluorescence correlation spectroscopy. Biophys. J. 2008;94:1437–1448. doi: 10.1529/biophysj.107.108811. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Potma E.O., de Boeij W.P., Wiersma D.A. Reduced protein diffusion rate by cytoskeleton in vegetative and polarized dictyostelium cells. Biophys. J. 2001;81:2010–2019. doi: 10.1016/s0006-3495(01)75851-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Blainey P.C., van Oijen A.M., Xie X.S. A base-excision DNA-repair protein finds intrahelical lesion bases by fast sliding in contact with DNA. Proc. Natl. Acad. Sci. USA. 2006;103:5752–5757. doi: 10.1073/pnas.0509723103. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Nachury M.V., Seeley E.S., Jin H. Trafficking to the ciliary membrane: how to get across the periciliary diffusion barrier? Annu. Rev. Cell Dev. Biol. 2010;26:59–87. doi: 10.1146/annurev.cellbio.042308.113337. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Ritchie K., Spector J. Single molecule studies of molecular diffusion in cellular membranes: determining membrane structure. Biopolymers. 2007;87:95–101. doi: 10.1002/bip.20801. [DOI] [PubMed] [Google Scholar]
- 16.Qian H., Sheetz M.P., Elson E.L. Single particle tracking. Analysis of diffusion and flow in two-dimensional systems. Biophys. J. 1991;60:910–921. doi: 10.1016/S0006-3495(91)82125-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Levi V., Ruan Q., Gratton E. 3-D particle tracking in a two-photon microscope: application to the study of molecular dynamics in cells. Biophys. J. 2005;88:2919–2928. doi: 10.1529/biophysj.104.044230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Semrau S., Pezzarossa A., Schmidt T. Microsecond single-molecule tracking (μsSMT) Biophys. J. 2011;100:L19–L21. doi: 10.1016/j.bpj.2010.12.3721. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.DeCenzo S.H., DeSantis M.C., Wang Y.M. Single-image separation measurements of two unresolved fluorophores. Opt. Express. 2010;18:16628–16639. doi: 10.1364/OE.18.016628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Schuster J., Cichos F., von Borczyskowski C. Diffusion measurements by single-molecule spot-size analysis. J. Phys. Chem. A. 2002;106:5403–5406. [Google Scholar]
- 21.Zareh S.K., Wang Y.M. Single-molecule imaging of protein adsorption mechanisms to surfaces. Microsc. Res. Tech. 2011;74:682–687. doi: 10.1002/jemt.20954. [DOI] [PubMed] [Google Scholar]
- 22.DeSantis M.C., DeCenzo S.H., Wang Y.M. Precision analysis for standard deviation measurements of immobile single fluorescent molecule images. Opt. Express. 2010;18:6563–6576. doi: 10.1364/OE.18.006563. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Swaminathan R., Hoang C.P., Verkman A.S. Photobleaching recovery and anisotropy decay of green fluorescent protein GFP-S65T in solution and cells: cytoplasmic viscosity probed by green fluorescent protein translational and rotational diffusion. Biophys. J. 1997;72:1900–1907. doi: 10.1016/S0006-3495(97)78835-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Weisshart K., Jüngel V., Briddon S.J. The LSM 510 META-ConfoCor 2 system: an integrated imaging and spectroscopic platform for single-molecule detection. Curr. Pharm. Biotechnol. 2004;5:135–154. doi: 10.2174/1389201043376913. [DOI] [PubMed] [Google Scholar]
- 25.Willig K.I., Kellner R.R., Hell S.W. Nanoscale resolution in GFP-based microscopy. Nat. Methods. 2006;3:721–723. doi: 10.1038/nmeth922. [DOI] [PubMed] [Google Scholar]
- 26.Egner A., Hell S.W. Equivalence of the Huygens–Fresnel and Debye approach for the calculation of high aperture point-spread functions in the presence of refractive index mismatch. J. Microsc. 1999;193:244–249. [Google Scholar]
- 27.Huang B., Wang W., Zhuang X. Three-dimensional super-resolution imaging by stochastic optical reconstruction microscopy. Science. 2008;319:810–813. doi: 10.1126/science.1153529. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Yildiz A., Forkey J.N., Selvin P.R. Myosin V walks hand-over-hand: single fluorophore imaging with 1.5-nm localization. Science. 2003;300:2061–2065. doi: 10.1126/science.1084398. [DOI] [PubMed] [Google Scholar]
- 29.Yildiz A., Tomishige M., Selvin P.R. Kinesin walks hand-over-hand. Science. 2004;303:676–678. doi: 10.1126/science.1093753. [DOI] [PubMed] [Google Scholar]
- 30.Axelrod D. Total internal reflection fluorescence microscopy. Methods Cell Biol. 1989;30:245–270. doi: 10.1016/s0091-679x(08)60982-6. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
