Abstract
Purpose
Quantification of myocardial blood flow (MBF) with generator-produced 82Rb is an attractive alternative for centres without an on-site cyclotron. Our aim was to validate 82Rb-measured MBF in relation to that measured using 15O-water, as a tracer 100% of which can be extracted from the circulation even at high flow rates, in healthy control subject and patients with mild coronary artery disease (CAD).
Methods
MBF was measured at rest and during adenosine-induced hyperaemia with 82Rb and 15O-water PET in 33 participants (22 control subjects, aged 30 ± 13 years; 11 CAD patients without transmural infarction, aged 60 ± 13 years). A one-tissue compartment 82Rb model with ventricular spillover correction was used. The 82Rb flow-dependent extraction rate was derived from 15O-water measurements in a subset of 11 control subjects. Myocardial flow reserve (MFR) was defined as the hyperaemic/rest MBF. Pearson’s correlation r, Bland-Altman 95% limits of agreement (LoA), and Lin’s concordance correlation ρ c (measuring both precision and accuracy) were used.
Results
Over the entire MBF range (0.66–4.7 ml/min/g), concordance was excellent for MBF (r = 0.90, [82Rb–15O-water] mean difference ± SD = 0.04 ± 0.66 ml/min/g, LoA = −1.26 to 1.33 ml/min/g, ρ c = 0.88) and MFR (range 1.79–5.81, r = 0.83, mean difference = 0.14 ± 0.58, LoA = −0.99 to 1.28, ρ c = 0.82). Hyperaemic MBF was reduced in CAD patients compared with the subset of 11 control subjects (2.53 ± 0.74 vs. 3.62 ± 0.68 ml/min/g, p = 0.002, for 15O-water; 2.53 ± 1.01 vs. 3.82 ± 1.21 ml/min/g, p = 0.013, for 82Rb) and this was paralleled by a lower MFR (2.65 ± 0.62 vs. 3.79 ± 0.98, p = 0.004, for 15O-water; 2.85 ± 0.91 vs. 3.88 ± 0.91, p = 0.012, for 82Rb). Myocardial perfusion was homogeneous in 1,114 of 1,122 segments (99.3%) and there were no differences in MBF among the coronary artery territories (p > 0.31).
Conclusion
Quantification of MBF with 82Rb with a newly derived correction for the nonlinear extraction function was validated against MBF measured using 15O-water in control subjects and patients with mild CAD, where it was found to be accurate at high flow rates. 82Rb-derived MBF estimates seem robust for clinical research, advancing a step further towards its implementation in clinical routine.
Keywords: Myocardial blood flow, Positron emission tomography, Rubidium-82, Healthy subjects, Coronary artery disease
Introduction
With the advent of hybrid PET/CT driven by oncology imaging, there are a growing number of centres using cardiac PET with 82Rb. This short-lived radioisotope (half-life 76.4 s) represents an attractive alternative to clinical imaging, as there is no need for a cyclotron [1–3]. 82Rb allows clinical imaging with short protocols (20–30 min in total) and a high patient throughput, providing better image quality and overall sensitivity in the diagnosis of coronary artery disease (CAD) as compared to myocardial scintigraphy with 201Tl or 99mTc-based radiotracers [2, 4, 5]. It is currently commercially available in the US, and more recently in Europe [6].
Over the last two decades, investigators have developed methods for using 82Rb dynamic PET for deriving absolute quantification of myocardial blood flow (MBF) [7–14]. However, most centres do not take full advantage of the PET information, as quantification of MBF is not yet implemented in daily clinical routine, although it shows prominent advantages over myocardial scintigraphy [15]. Indeed, MBF quantification is very relevant to diffusely abnormal MBF in conditions such as severe or balanced three-vessel disease and dysfunction of the coronary microcirculation [16, 17], and even in the absence of angiographic lesions in primary or secondary cardiomyopathies [18, 19]. There is growing evidence that important diagnostic or prognostic information can be gained from quantification of MBF or MBF reserve (MFR) in addition to myocardial perfusion imaging [20–26]. The clinical utility of quantification has gained importance for the diagnosis, prognosis and quantification of the various disorders of myocardial perfusion [3, 15, 24, 26–28].
Therefore, it is of great importance to be able to quantify MBF with 82Rb accurately. The quantification of 82Rb MBF has been compared to 13N-ammonia- or microsphere-derived MBF in a number of human and animal studies [14, 29, 30]. MBF quantification might be inaccurate during maximal hyperaemia, due to the reduced extraction of 82Rb at high flow rates leading to an important, potentially noisy correction [15, 28]. It is therefore highly relevant to verify that quantification of hyperaemic MBF is accurate. Indeed, it was recently found that the extent of impairment of maximum hyperaemic MBF is prognostically informative, as decreased MFR can be due to abnormally increased resting MBF in the presence of conserved hyperaemic MBF [31]. Thus, a single measurement of MBF during vasodilator stress might be sufficient to identify myocardium supplied by an artery with haemodynamically significant stenosis and therefore at risk [20, 31].
82Rb-derived MBF quantification has not yet been compared directly with the gold standard for measuring MBF in humans, 15O-water, a radiotracer 100% of which can be extracted from the circulation, even at very high flow rates [32, 33]. Therefore, validation of 82Rb-derived MBF against MBF measured using 15O-water is an important step, as quantification using 82Rb is ready to enter daily clinical practice. Whether MBF quantification with 82Rb is accurate enough even at high flow rates as compared to quantification with 15O-water has not been determined in humans in a clinical setting with modern PET/CT. To this aim, we wanted to derive an appropriate 82Rb extraction function as compared to the 100%-extractible 15O-water and determine whether a one-tissue compartment model could provide adequate MBF quantification in a group of healthy control subjects and patients with mild, stable CAD.
Materials and methods
Study population
Enrolled in the study were 33 subjects (17 men, 16 women) comprising 22 healthy control subjects (aged 30 ± 13 years) with no coronary risk factors and a low probability of CAD (<5%) based on the absence of cardiac symptoms [34], and 11 patients with known, stable CAD (60 ± 13 years) presently free of any cardiac symptoms (typical or atypical chest pain or angina pectoris equivalent) and without a history of previous transmural myocardial infarction (this to avoid biais due to different radiotracer-specific behaviour in infarcts). Participants underwent a medical history, physical examination, blood pressure measurements and ECG; fasting blood was collected for routine testing (chemistry and lipid panel). None of the control subjects was taking any medication. Pregnancy was excluded in all women of child-bearing age. The study was performed at two locations (82Rb in Lausanne and 15O-water in Zurich). Both local Ethics Committees, as well as the Swiss regulatory authorities approved the study protocol. All participants gave written informed consent before enrolment.
Study protocol
Each participant underwent rest and stress imaging with 82Rb and 15O-water within 4 weeks (median time interval 4 days, interquartile range 2–8 days). They refrained from taking caffeine-containing substances for at least 24 h and were asked to fast at least 6 h prior to the PET studies. MBF was measured at rest and during adenosine-induced pharmacological vasodilation; the two measurements were separated by 15 min. Vasodilation was induced with adenosine (140 μg/kg/min) for 6 min with 82Rb or 15O-water administration 2.5 min after the start of the infusion. 82RbCl (1,100 MBq) was obtained from a 82Sr/82Rb generator (CardioGen-82; Bracco, Princeton, NJ) and injected by an automated infusion system (RbM Services, LLC, Oakdale, TN) using a dedicated left-arm venous line (separate from the line used for adenosine) in 10–20 ml of 0.9% NaCl/water solution over 10–20 s. 15O-water (700 MBq) was produced by the University of Zurich Hospital cyclotron and injected intravenously over 20 s at a rate of 24 ml/min [35]. In all participants, 12-lead ECG, heart rate and blood pressure were recorded at 2-min intervals and averaged over the first 2 min of data acquisition to derive the rate–pressure product (RPP = heart rate × systolic blood pressure) as an index of cardiac work and coronary vascular resistance (CVR = mean arterial blood pressure/MBF).
Image acquisition
Dynamic cardiac PET was performed with 82Rb in Lausanne (Discovery LS; GE Medical System, Milwaukee, WI) using a 10-min acquisition starting immediately after the start of the 82Rb infusion (21 frames, 35 slices each, 2-D mode, 14.5-cm axial field of view; 12 × 8, 5 × 12, 1 × 30, 1 × 60, 1 × 120 and 1 × 240 s). Cardiac 15O-water PET was performed in Zurich (Discovery ST RX; GE Medical System, Milwaukee, WI) with the acquisition of a background frame immediately before the infusion of 15O-water and a 5-min dynamic acquisition (24 frames, 47 slices each, 2-D mode; 15.2-cm axial field of view; 14 × 5, 3 × 10, 3 × 20, 4 × 30 s). First a scout CT acquisition (120 kV, 10 mA, table speed 10 cm/s) was performed, followed by a low-dose CT scan for attenuation correction (120 kV, 10 mA, 1.0 s per rotation, pitch 0.75) just before the rest 82Rb or 15O-water acquisition. The low-dose CT scan was repeated immediately after the hyperaemic 82Rb acquisition for attenuation correction. The radiation dose for each participant was estimated to be 2 × 1.3 mSv for rest and stress 82Rb [36], 2 × 0.8 mSv for 15O-water, and 3 × 0.2 mSv for the low-dose attenuation correction CT scan [37].
Image processing
Sinograms were corrected for attenuation and reconstructed on dedicated workstations (GE Medical System, Milwaukee, WI) using standard iterative reconstruction algorithms (CHUV comprising OSEM with two iterations and 28 subsets, 3.27-mm FWHM post-filter, 2.34-mm loop filter, and 128 × 128-pixel matrix size; USZ comprising OSEM with two iterations and 28 subsets, 3.49-mm post-filter, 2.50-mm loop filter, and 128 × 128-pixel matrix size). Dynamic images were analysed for 82Rb and 15O-water in a blinded manner using PMOD 2.85 cardiac PET analysis software (PMOD Technologies, Zurich, Switzerland). Myocardial images were generated from the 82Rb and 15O-water studies and reoriented along short-axis, and vertical and horizontal long-axis views.
For the 82Rb studies, myocardial images were generated by averaging the images from 90–600 s. Myocardial volumes of interest (VOI) were drawn on at least 12 consecutive slices. Two VOIs were drawn in the left ventricle (LV) and right ventricle (RV) on four to six consecutive basal slices for LV and RV input functions, respectively. For the 15O-water studies, factor images were generated and used for drawing VOIs. The LV myocardium surface was subdivided using the standard 17-segment model of the American Heart Association (AHA). Time–activity curves were generated for each of the segments and for the blood in the LV and RV and fitted to a kinetic model to calculate absolute regional MBF in millilitres per minute per gram. Hyperaemic and rest MBF for each segment were divided to obtain the regional MFR. Global MBF and MFR were computed from the 17-segment average. Regional values were derived in each of the coronary artery territories by averaging the corresponding segments over the left anterior descending coronary artery (AHA segments 1, 2, 7, 8, 13, 14 and 17), the left circumflex artery (AHA segments 3, 4, 9, 10 and 15) and the right coronary artery (AHA segments 5, 6, 11, 12 and 16) [38].
MBF estimation
A one-tissue compartment model (sometimes called a two-compartment model) including spillover correction was employed for fitting the 82Rb time–activity data from the different myocardial segments [14]. The activity concentration in the myocardium C m(t) was calculated as , where K 1 and k 2 denote the uptake and washout rate constants of the one-tissue compartment model and C LV(t) the activity concentration of the blood in the LV, and ⊗ is the convolution operation. The operational equation which was fitted to the measurements was given by , where V LV and V RV are the fractional contributions of the LV and RV blood activity to the PET signal measured in the myocardial segments. K 1, k 2 and V LV were always fitted, whereas V RV was only fitted for the septal segments and otherwise fixed at a value of zero.
To account for the flow-dependent extraction of 82Rb, a generalized Renkin-Crone function was used to recover MBF from K 1, as follows: , where F is the flow and E is the first-pass flow-dependent extraction fraction [14, 39]. The a and b factors were determined from the best fit of the Renkin-Crone function to K 1 as a function of F. F was measured with 15O-water using rest and hyperaemic MBF from a subset of half of the control subjects (n = 11), who were chosen by sampling the 15O-water hyperaemic MBF values according to quintiles (to obtain uniform sampling across all hyperaemic MBF values) and by choosing the two corresponding K 1 measurements closest to the median of K 1 across the corresponding quintile.
For 15O-water, a standard one-compartment model was used [35]. For both radioisotopes, the LV and RV time–activity curves were used to correct for spillover of all LV segments and only the septal segments for the RV. Artefacts due to late starting of the scan relative to the 15O-water infusion precluded image analysis in two hyperaemic 15O-water studies.
Statistical analysis
The results are presented as means ± SD. Comparisons across groups were performed using the t-test or the t-test after logarithm normalization for skewed data distributions (only high-sensitivity C-reactive protein in our data). The a and b parameters for the nonlinear extraction function to compute 82Rb K 1 from 15O-water-derived flow F were derived from a nonlinear least-squares algorithm with 95% confidence bounds derived from the covariance matrix [40].
The correspondence between the 82Rb and 15O-water measurement methods was assessed by the Pearson’s correlation coefficient r and the Bland-Altman 95% limit of agreement methods. In addition, we used the concordance correlation as proposed by Lin [41], which is essentially equivalent to the well-known kappa coefficient but is applicable to continuous data. This concordance correlation coefficient ρ c evaluates both accuracy and precision, indicating how far the measurement pairs fall from the line of identity, ranging from +1 (perfect agreement) to 0 (no agreement) to −1 (perfect inverse agreement) [41–43]. For graphical representation, we used the reduced major axis regression line, which is a useful summary of the data and is defined as the line going through the intersection of the means with a slope given by the sign of the Pearson’s correlation r and the ratio of the respective standard deviations. All tests were two-sided and statistical analyses were performed with Stata 10.1 software (Stata Corporation, College Station, TX) using a p value of <0.05 as the significance level.
Results
Clinical and laboratory findings
The characteristics of the study groups are listed in Table 1. CAD patients were older than the control subjects and had higher mean BMI and systolic blood pressure, although these mean values were still within the normal ranges. All patients with hypertension or dyslipidaemia were adequately treated by medications lowering blood pressure or cholesterol. Mean fasting glucose was normal in CAD patients, although it was higher than in control subjects. Total, HDL and LDL cholesterol were similar between the groups, but the HDL/total cholesterol ratio was higher in CAD patients than in control subjects. The high-sensitivity CRP was significantly higher in CAD patients after logarithm normalization because of a non-normal, skewed distribution. All the participating patients had stable CAD without previous transmural myocardial infarction (three patients with previously known non-transmural myocardial infarction) and were presently free of chest pain or angina pectoris equivalent. All previously documented significant coronary stenoses (six patients) had been revascularized (by stenting or coronary artery bypass surgery).
Table 1.
Variable | Controls (n = 22) | CAD patients (n = 11) | P |
---|---|---|---|
Age (years) | 30 ± 13 | 60 ± 13 | <0.001 |
Body mass index (kg/m2) | 22.5 ± 2.3 | 26.7 ± 3.4 | 0.001 |
Blood pressure (mmHg) | 111 ± 11/63 ± 10 | 129 ± 19/66 ± 11 | 0.004/0.40 |
Fasting plasma glucose (mmol/l) | 4.12 ± 0.75 | 5.11 ± 0.65 | 0.002 |
Total cholesterol (mmol/l) | 4.55 ± 0.66 | 4.77 ± 1.00 | 0.48 |
HDL cholesterol (mmol/l) | 1.65 ± 0.46 | 1.37 ± 0.34 | 0.059 |
LDL cholesterol (mmol/l) | 2.74 ± 1.01 | 2.52 ± 0.55 | 0.44 |
HDL/total cholesterol index | 2.88 ± 0.67 | 3.74 ± 1.55 | 0.043 |
Triglyceride level (mmol/l) | 0.86 ± 0.32 | 1.42 ± 0.58 | 0.002 |
High-sensitivity CRP (mg/l) | 2.2 ± 2.3 | 13 ± 26 | 0.009 |
HOMA-IR | 1.9 ± 0.7 | 2.7 ± 0.6 | 0.042 |
Hypertension | 0 (0%) | 5 (45%) | 0.002 |
Dyslipidaemia | 0 (0%) | 7 (64%) | <0.001 |
Obesity | 0 (0%) | 2 (18%) | 0.10 |
Diabetes | 0 (0%) | 1 ( 9%) | 0.33 |
Smoking | 0 (0%) | 2 (18%) | 0.10 |
Familial history of early CAD | 0 (0%) | 1 ( 9%) | 0.33 |
Previous revascularization | 0 (0%) | 6 (55%) | <0.001 |
Left anterior descending artery | – | 5 (45%) | – |
Left circumflex artery | – | 3 (27%) | – |
Right coronary artery | – | 2 (18%) | – |
Previous nontransmural infarction | 0 (0%) | 3 (27%) | 0.030 |
Left anterior descending artery | – | 1 (9%) | – |
Left circumflex artery | – | 1 (9%) | – |
Right coronary artery | – | 1 (9%) | – |
Typical or atypical angina | 0 (0%) | 0 (0%) | 1.0 |
Shortness of breath | 0 (0%) | 0 (0%) | 1.0 |
HOMA-IR homeostasis model assessment of insulin resistance.
Myocardial perfusion imaging
Visual and semiquantitative analyses of all rest and stress 82Rb myocardial activity images revealed normal uniform perfusion in the majority of the segments (1,114/1,122 or 99.3%). Myocardial uptake was normal in the 22 control subjects and 7 patients (summed stress score, SSS = 0), while four patients had slightly abnormal myocardial stress polar maps (all SSS ≤ 2, in one or two segments). One patient with previous nontransmural myocardial infarction had a fixed stress and rest defect in two segments corresponding to the revascularized left circumflex artery (SSS = 2; summed rest score, SRS = 2; summed difference score, SDS = 0); two revascularized patients (one with a previously known nontransmural myocardial infarction in the right coronary artery) had minor reversible perfusion defects in one segment (SSS = 1, SRS = 0, SDS = 1), and one unrevascularized patient had a minor reversible perfusion defect in two segments (SSS = 2, SRS = 0, SDS = 2). During adenosine infusion, no ECG ST-segment changes or unexpected side effects were observed. All daily quality controls of the 82Rb elution were passed with 82Sr and 85Sr breakthrough values well below maximal allowed limits (<6% of the allowed 82Sr/82Rb limit of 20 × 10–6 and <1.7% of the allowed 85Sr/82Rb limit of 200 × 10–6).
Haemodynamics
The haemodynamic conditions during both 15O-water and 82Rb studies were similar, as indicated by equivalent RPP at rest (7.6 ± 1.8 vs. 7.4 ± 1.7 × 103·min-1·mmHg, respectively; p = 0.28) and during hyperaemia (10.0 ± 2.1 vs. 9.9 ± 2.0 × 103·min-1·mmHg, respectively; p = 0.48). One control subject declined repeated adenosine administration due to unpleasant symptoms experienced during his first study (flush), which were however within the range of commonly observed side effects of adenosine.
Extraction function of 82Rb
The generalized Renkin-Crone function was fitted to the 82Rb K 1 values as a function of the 15O-water MBF in half of the control subjects (n = 11) to derive the a and b constants. The fit was excellent (a = 0.80, b = 0.59 ml/min/g, R 2 = 0.97, RMSE = 0.165) allowing conversion of the K 1 rate constant estimates to MBF (Fig. 1).
MBF findings
A scatter plot showing the concordance between the 82Rb and 15O-water MBF is shown in Fig. 2a. Over the whole MBF range (0.66–4.7 ml/min/g), the reduced major axis was close to the line of perfect concordance. Pearson’s correlation r was 0.89, and Lin’s concordance correlation ρ c was 0.88 (95% CI 0.82–0.94). A Bland-Altman plot (Fig. 2b) showed good agreement (95% limits of agreement = ±1.96 × SD = 1.30, or −1.26 to 1.34 ml/min/g) and no mean difference (0.04 ± 0.66 ml/min/g, or 2 ± 33%, p = 0.69) in the MBF measurements. When performed separately, the mean and 95% limits of agreement for rest MBF and stress MBF were −0.02 ml/min/g (±1.96 × SD = 0.46, or −0.048 to 0.044 ml/min/g) and −0.10 ml/min/g (±1.96 × SD = 1.79, or −1.69 to 0.1.89 ml/min/g), respectively.
A scatter plot for MFR also showed good concordance over the whole MFR range (1.79–5.81) with a reduced major axis almost equal to the line of identity (Fig. 3a). The corresponding Pearson’s correlation r was 0.83 and Lin’s concordance correlation ρ c was 0.82 (95% CI 0.68–0.96). A Bland-Altman plot (Fig. 3b) also showed excellent concordance (95% limits of agreement = ±1.96 × SD = 1.14, or −0.99 to 1.28) and no significant mean difference (0.14 ± 0.58, or 4 ± 18%, p = 0.26).
Values of MBF at rest and during stress, MFR and CVR in the CAD patients and control subjects are shown in Table 2 for both 82Rb and 15O-water cardiac PET., there was no significant MBF group difference at rest, while both hyperaemic MBF and MFR were significantly decreased in CAD patients as compared to control subjects, with a corresponding higher hyperaemic CVR in CAD patients.
Table 2.
Cardiac PET tracer | Variable | Control subjects (n = 11) | CAD patients (n = 11) | p value |
---|---|---|---|---|
15O-water | Rest MBF (ml/min/g) | 1.00 ± 0.24 | 0.95 ± 0.16 | 0.64 |
Stress MBF (ml/min/g) | 3.62 ± 0.68 | 2.53 ± 0.74 | 0.002 | |
MFR | 3.79 ± 0.98 | 2.65 ± 0.62 | 0.004 | |
Stress CVR (mmHg/ml·min·g) | 23 ± 6 | 37 ± 13 | 0.005 | |
82Rb | Rest MBF (ml/min/g) | 1.03 ± 0.42 | 0.88 ± 0.21 | 0.33 |
Stress MBF (ml/min/g) | 3.82 ± 1.21 | 2.53 ± 1.01 | 0.013 | |
MFR | 3.88 ± 0.91 | 2.85 ± 0.86 | 0.012 | |
Stress CVR (mmHg/ml·min·g) | 23 ± 9 | 38 ± 17 | 0.017 |
Discussion
Quantification of MBF with 82Rb cardiac PET is of increasing clinical and prognostic value, but its accuracy against quantification using 15O-water, a tracer 100% of which can be extracted from the circulation, has not been validated at very high MBF rates such as during hyperaemia. The extraction function of 82Rb was derived in normal control subjects from comparison with 15O-water allowed the K 1 parameter to be accurately converted with an excellent fit to blood flow, even at high flow rates. Using this function in a second, different group of control subjects and CAD patients, excellent concordance and agreement was obtained between MBF or MFR derived from 15O-water and 82Rb without a systematic mean difference over the whole range of rest and hyperaemic flows. Finally, in our group of patients with mild CAD with homogeneous myocardial perfusion in the majority of segments, MBF quantification with either 82Rb or 15O-water revealed significantly decreased MBF, MFR and CVR as compared to control subjects.
The a and b parameters found in our study (a = 0.80, b = 0.59 ml/min/g) are close to the parameters obtained by Lortie et al. using a similar process with 13N-ammonia cardiac PET in 14 healthy subjects (a = 0.77, b = 0.63 ml/min/g) and the factors determined by Schelbert [39] based on a study by Glatting et al. [44] using the argon inert gas method (a = 0.73, b = 0.59 ml/min/g). These a and b parameters are known to represent the flow-dependent permeability surface product, PS (PS = ln(1/a) × F + b, i.e. 0.22 × F + 0.59 in our study) due to increased capillary recruitment that is related to the extraction fraction (E = 1 − e−PS/F) [8, 45]. The differences among these three extraction curves can be appreciated in Fig. 4, showing that parameters derived in this study lead to slightly larger corrections at higher flows than both other methods. To verify the validity of our 82Rb extraction function derivation from 15O-water, we performed the same fit using the other half of the control subjects (n = 11) and obtained almost identical fit parameters (a = 0.80, b = 0.58 ml/min/g, R 2 = 0.96, RMSE = 0.186). This shows robustness of the proposed sampling approach across quintiles of 15O-water hyperaemic flows to get unbiased estimates of the 82Rb extraction function.
Comparisons were made with the Lin’s concordance correlation coefficient, as it is the most appropriate test for assessing the equivalence of two methods of measurements [41–43]. It offers more insight into the observed differences, as ρ c equals the product of the standard Pearson’s correlation r (a measure of imprecision) and the bias coefficient c b (a measure of inaccuracy). Among well-known approaches to compare data from two measurement methods, Lin’s concordance correlation offers many advantages. Indeed, when comparing data obtained with two measurement methods, the Pearson’s correlation coefficient alone fails to detect a departure from the 45° line through the origin, thus ignoring inaccuracy. Using the conventional paired t-test between corresponding data from each method could lead to rejecting a well reproducible method due to a very small residual error. With the least squares approach, one would fail to detect departure from slope = 1 and intercept = 0 if the data were very scattered. Finally, the coefficient of variation and the intraclass coefficient of correlation would not distinguish bias from imprecision.
In our study, most MBF and MFR differences were due to imprecision (intrasample variations), as the bias correction factors were very close to 1 (c b = ρ c/r = 0.99 for both MBF and MFR), indicating an excellent accuracy between 82Rb and 15O-water. The Lin’s concordance correlation was slightly higher for MBF than for MFR, but the difference was not statistically significant, as shown by overlapping confidence intervals. The use of correlations is only meaningful when comparisons are performed over the whole clinical range of potential MBF or MFR, since agreement over a small range cannot be extrapolated to a wider range. This is the reason why rest and stress measurements for both patients and control subjects were considered together for this analysis. However, similar findings were found when control subjects and patients were analysed separately, with ρ c for MBF of 0.86 in control subjects and 0.90 in patients, respectively. The corresponding values for MFR were of 0.87 in control subjects and 0.56 in patients), respectively. Interestingly, this last value for MFR in patients was slightly but not significantly lower than that in control subjects, due to a significantly lower range of MFR in patients than in control subjects. Although this was not the purpose of our study, good concordances were found in each coronary artery territory, and no significant differences in regional MBF were observed among coronary artery territories regarding concordance of 82Rb vs. 15O-water measurements (ρ c = 0.89, 95% CI 0.83–0.95, for the left anterior descending artery: ρ c = 0.80, 95% CI 0.69–0.90, for the left circumflex artery; ρ c = 0.77, 95% CI 0.66–0.88, for the right coronary artery), although the left anterior descending and left circumflex territories values were slightly but not significantly better than the right coronary territory values. Finally, we wondered if 82Rb measurements across coronary territories were subject to wider variations than 15O-water measurements, but using the variance ratio test no differences in the standard deviations of the measurements were observed (SD = 1.43 for all 82Rb measurements, SD = 1.20 for all 15O-water measurements, p = 0.17).
The Bland-Altman limit of agreement method showed a good to excellent agreement for MBF and MFR, with only a few measurement pairs outside the limits of agreement, confirming a normal distribution of the measurements. There was a variability of pharmacologically induced hyperaemia, as previously reported, with differences up to 12 ± 12% in stress MBF [46, 47]. Thus, the observed differences between the 82Rb and 15O-water studies (−2 ± 33%) may be attributable largely to physiological variability as opposed to methodological biases due to differences in location, time or sequence order of the 82Rb and 15O-water measurements. It is worth noting that we found similar limits of agreements for rest MBF (−0.02 ± 0.46, mean ± 1.96 × SD) to those found by Yoshinaga et al. (−0.05 ± 0.49, mean ± 1.96 × SD) [16].
The MBF and MFR means and ranges obtained with 82Rb in healthy control subjects and patients with CAD are comparable to previously published values in these groups [14, 48, 49]. Interestingly, the group difference was even more significant when considering the CVR that takes into account the effect of arterial pressure. Decreases in MBF and MFR in CAD patients in relation to control subjects were seen with 82Rb as well as with the gold standard 15O-water. Thus, it should be possible to perform clinical studies comparing MBF or MFR among groups using 82Rb-based quantitation.
The quantitative aspect of cardiac PET may actually reveal more extensive disease than myocardial scintigraphy or nonquantitative PET [5, 50], which frequently shows decreased perfusion only in the myocardial area supplied by the artery with the most severe stenosis [51]. In keeping with these results, Sampson at al. have recently found that 45% of patients with multivessel CAD on coronary angiography had stress perfusion defects in only one territory on nonquantitative visual analysis of 82Rb cardiac PET/CT and 21% of the patients with one-vessel CAD on coronary angiography had stress PET abnormalities in more than one territory [5].
In our study, the myocardium was affected similarly in most territories in most of the CAD patients, with only four patients with regionally decreased MBF and MFR of very limited extent (one or two segments). Moreover, only a few patients showed mildly reduced MFR (<2.0–2.5), in agreement with the absence of previous transmural myocardial infarction and the presence of only mild diffuse coronary artery stenosis or effective previous revascularization. This suggests that patients with multiple coronary risk factors have a uniformly decreased response to pharmacological vasodilation due to diffuse (micro- or macrovascular) endothelial dysfunction without necessarily having obstructive CAD [16, 17], which may be underestimated when measuring stenosis severity alone, as previously shown [52, 53]. Importantly, nonquantitative 82Rb cardiac PET assessment has been shown to be valuable for risk stratification of future cardiovascular events, even after nondiagnostic myocardial SPECT scintigraphy [38]. Recently the additional prognostic power has been shown to be enhanced by quantitative MFR measurements [21, 24–26]. Whether this can be further enhanced by absolute measurement of hyperaemic MBF alone remains to be determined [54].
Cardiac PET and PET/CT myocardial perfusion imaging is becoming a clinically mature, cost-effective method [2, 4–6, 35, 52]. Further benefits over conventional myocardial SPECT scintigraphy include higher specificity, sensitivity and accuracy [4], lower patient and technologist/physician radiation doses [55], shorter examination duration, and the capacity to quantitate MBF and MFR. It was recently demonstrated to be cost-effective in the clinical setting [56]. Generator-produced 82Rb certainly has several key advantages over other PET radiopharmaceuticals such as 15O-water or 13N-ammonia: its use obviates the need for a cyclotron and, compared to the use of 13N-ammonia, allows a higher patient throughput [57]. Notably, all 82Rb cardiac PET examinations performed in our study led to successful MBF quantitation, further indicating its clinical feasibility. However, special attention to technical details is necessary, especially to possible misregistration artefacts, which may present as shifts between attenuation correction and the PET images [58]. MBF quantitation processing, challenging in the past, is becoming easier and faster with the advent of software dedicated to cardiac processing with automated reorientation and myocardial VOI drawing [59, 60].
This study had a few limitations which should be mentioned. The fact that MBF was not measured during the same session so that differences in time, location and possibly hyperaemic flow may have increased the observed random intrasample variations leading to an underestimation of true concordance. We excluded patients with previous myocardial infarction so as not to introduce additional MBF bias from known differences in radiotracer behaviour in myocardial scar due to tissue fibrosis [61, 62]. The abnormal myocardial perfusion seen in four CAD patients may have affected our comparisons based on MBF averaged over the 17 LV segments, although the defects were moderate in severity, limited in extent, and affected similar segments on the 82Rb and 15O-water studies; furthermore, no significant differences were observed among coronary artery territories or when these four patients were excluded from the analysis. Great care was taken in the execution of the PET/CT studies to avoid any misregistration between the attenuation-correction CT image and the PET image, notably by repeating the attenuation-correction CT scan immediately after the pharmacological stress. Although alignment between PET and CT images was excellent with no displacement greater than 5 mm, it is nevertheless possible that smaller image shifts may still have influenced MBF quantification, resulting in a slight degradation of concordance and agreement.
Quantification of MBF with 82Rb with a newly derived correction for the nonlinear extraction function was validated against 15O-water in control subjects and patients with mild CAD. It was found to be accurate event at high flow rates using modern PET/CT and commercially available software. Thus, 82Rb-derived MBF estimates are robust in the clinical setting even at higher flow rates, advancing a step further towards its routine implementation in clinical practice.
Acknowledgments
The authors are indebted to Mr. Jérôme Malterre and Luca Modolo (in Lausanne), Mr. Ratko Milovanovic and Dr. Tobias Hoefflinghaus (in Zurich) for assisting in the studies, and to Rolf R. Hesselmann, PhD (Cyclotron Unit, Zurich) for 15O-water radioisotope production. They also would like to thank Drs. Kathrin Marx, MD, for logistic support and David W. Crook, MD, for editorial assistance.
This work was supported by research grants from the Swiss National Science Foundation (grant no. 320000-109986), the Michel Tossizza Foundation (Lausanne, Switzerland), and the Société Académique Vaudoise (Lausanne, Switzerland). P.A.K. was supported by the Swiss National Science Foundation (grant no. PP00A-114706). J.O.P. was a recipient of an Academic Research Award from the Leenaards Foundation (Lausanne, Switzerland).
Conflicts of interest
None.
Open Access
This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
References
- 1.Gould KL, Goldstein RA, Mullani NA, Kirkeeide RL, Wong WH, Tewson TJ, et al. Noninvasive assessment of coronary stenoses by myocardial perfusion imaging during pharmacologic coronary vasodilation. VIII. Clinical feasibility of positron cardiac imaging without a cyclotron using generator-produced rubidium-82. J Am Coll Cardiol. 1986;7:775–789. doi: 10.1016/S0735-1097(86)80336-9. [DOI] [PubMed] [Google Scholar]
- 2.Machac J. Cardiac positron emission tomography imaging. Semin Nucl Med. 2005;35:17–36. doi: 10.1053/j.semnuclmed.2004.09.002. [DOI] [PubMed] [Google Scholar]
- 3.Schelbert HR. Quantification of myocardial blood flow: what is the clinical role? Cardiol Clin. 2009;27:277–289. doi: 10.1016/j.ccl.2008.12.009. [DOI] [PubMed] [Google Scholar]
- 4.Bateman TM, Heller GV, McGhie AI, Friedman JD, Case JA, Bryngelson JR, et al. Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. J Nucl Cardiol. 2006;13:24–33. doi: 10.1016/j.nuclcard.2005.12.004. [DOI] [PubMed] [Google Scholar]
- 5.Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. J Am Coll Cardiol. 2007;49:1052–1058. doi: 10.1016/j.jacc.2006.12.015. [DOI] [PubMed] [Google Scholar]
- 6.Groves AM, Speechly-Dick ME, Dickson JC, Kayani I, Endozo R, Blanchard P, et al. Cardiac (82)rubidium PET/CT: initial European experience. Eur J Nucl Med Mol Imaging. 2007;34:1965–1972. doi: 10.1007/s00259-007-0537-8. [DOI] [PubMed] [Google Scholar]
- 7.Herrero P, Markham J, Shelton ME, Weinheimer CJ, Bergmann SR. Noninvasive quantification of regional myocardial perfusion with rubidium-82 and positron emission tomography. Exploration of a mathematical model. Circulation. 1990;82:1377–1386. doi: 10.1161/01.CIR.82.4.1377. [DOI] [PubMed] [Google Scholar]
- 8.Yoshida K, Mullani N, Gould KL. Coronary flow and flow reserve by PET simplified for clinical applications using rubidium-82 or nitrogen-13-ammonia. J Nucl Med. 1996;37:1701–1712. [PubMed] [Google Scholar]
- 9.Huang SC, Williams BA, Krivokapich J, Araujo L, Phelps ME, Schelbert HR. Rabbit myocardial 82Rb kinetics and a compartmental model for blood flow estimation. Am J Physiol. 1989;256:H1156–H1164. doi: 10.1152/ajpheart.1989.256.4.H1156. [DOI] [PubMed] [Google Scholar]
- 10.Herrero P, Markham J, Shelton ME, Bergmann SR. Implementation and evaluation of a two-compartment model for quantification of myocardial perfusion with rubidium-82 and positron emission tomography. Circ Res. 1992;70:496–507. doi: 10.1161/01.res.70.3.496. [DOI] [PubMed] [Google Scholar]
- 11.Coxson PG, Huesman RH, Borland L. Consequences of using a simplified kinetic model for dynamic PET data. J Nucl Med. 1997;38:660–667. [PubMed] [Google Scholar]
- 12.Lin JW, Sciacca RR, Chou RL, Laine AF, Bergmann SR. Quantification of myocardial perfusion in human subjects using 82Rb and wavelet-based noise reduction. J Nucl Med. 2001;42:201–208. [PubMed] [Google Scholar]
- 13.El Fakhri G, Sitek A, Guerin B, Kijewski MF, Di Carli MF, Moore SC. Quantitative dynamic cardiac 82Rb PET using generalized factor and compartment analyses. J Nucl Med. 2005;46:1264–1271. [PubMed] [Google Scholar]
- 14.Lortie M, Beanlands RS, Yoshinaga K, Klein R, Dasilva JN, Dekemp RA. Quantification of myocardial blood flow with (82)Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging. 2007;34:1765–1774. doi: 10.1007/s00259-007-0478-2. [DOI] [PubMed] [Google Scholar]
- 15.Knuuti J, Kajander S, Maki M, Ukkonen H. Quantification of myocardial blood flow will reform the detection of CAD. J Nucl Cardiol. 2009;16:497–506. doi: 10.1007/s12350-009-9101-1. [DOI] [PubMed] [Google Scholar]
- 16.Yoshinaga K, Manabe O, Katoh C, Chen L, Klein R, Naya M, et al. Quantitative analysis of coronary endothelial function with generator-produced 82Rb PET: comparison with 15O-labelled water PET. Eur J Nucl Med Mol Imaging. 2010;37:2233–2241. doi: 10.1007/s00259-010-1541-y. [DOI] [PubMed] [Google Scholar]
- 17.Yoshinaga K, Manabe O, Tamaki N. Assessment of coronary endothelial function using PET. J Nucl Cardiol. 2011;18:486–500. doi: 10.1007/s12350-011-9370-3. [DOI] [PubMed] [Google Scholar]
- 18.Beanlands RS, Muzik O, Melon P, Sutor R, Sawada S, Muller D, et al. Noninvasive quantification of regional myocardial flow reserve in patients with coronary atherosclerosis using nitrogen-13 ammonia positron emission tomography. Determination of extent of altered vascular reactivity. J Am Coll Cardiol. 1995;26:1465–1475. doi: 10.1016/0735-1097(95)00359-2. [DOI] [PubMed] [Google Scholar]
- 19.Camici PG, Crea F. Coronary microvascular dysfunction. N Engl J Med. 2007;356:830–840. doi: 10.1056/NEJMra061889. [DOI] [PubMed] [Google Scholar]
- 20.Hajjiri MM, Leavitt MB, Zheng H, Spooner AE, Fischman AJ, Gewirtz H. Comparison of positron emission tomography measurement of adenosine-stimulated absolute myocardial blood flow versus relative myocardial tracer content for physiological assessment of coronary artery stenosis severity and location. JACC Cardiovasc Imaging. 2009;2:751–758. doi: 10.1016/j.jcmg.2009.04.004. [DOI] [PubMed] [Google Scholar]
- 21.Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150–156. doi: 10.1016/j.jacc.2009.02.069. [DOI] [PubMed] [Google Scholar]
- 22.Knaapen P, Camici PG, Marques KM, Nijveldt R, Bax JJ, Westerhof N, et al. Coronary microvascular resistance: methods for its quantification in humans. Basic Res Cardiol. 2009;104:485–498. doi: 10.1007/s00395-009-0037-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Neglia D, L'Abbate A. Myocardial perfusion reserve in ischemic heart disease. J Nucl Med. 2009;50:175–177. doi: 10.2967/jnumed.108.056317. [DOI] [PubMed] [Google Scholar]
- 24.Tio RA, Dabeshlim A, Siebelink HM, de Sutter J, Hillege HL, Zeebregts CJ, et al. Comparison between the prognostic value of left ventricular function and myocardial perfusion reserve in patients with ischemic heart disease. J Nucl Med. 2009;50:214–219. doi: 10.2967/jnumed.108.054395. [DOI] [PubMed] [Google Scholar]
- 25.Fukushima K, Javadi MS, Higuchi T, Lautamaki R, Merrill J, Nekolla SG, et al. Prediction of short-term cardiovascular events using quantification of global myocardial flow reserve in patients referred for clinical 82Rb PET perfusion imaging. J Nucl Med. 2011;52:726–732. doi: 10.2967/jnumed.110.081828. [DOI] [PubMed] [Google Scholar]
- 26.Ziadi MC, Dekemp RA, Williams KA, Guo A, Chow BJ, Renaud JM, et al. Impaired myocardial flow reserve on rubidium-82 positron emission tomography imaging predicts adverse outcomes in patients assessed for myocardial ischemia. J Am Coll Cardiol. 2011;58:740–748. doi: 10.1016/j.jacc.2011.01.065. [DOI] [PubMed] [Google Scholar]
- 27.Beanlands RS, Ziadi MC, Williams K. Quantification of myocardial flow reserve using positron emission imaging the journey to clinical use. J Am Coll Cardiol. 2009;54:157–159. doi: 10.1016/j.jacc.2009.03.049. [DOI] [PubMed] [Google Scholar]
- 28.Camici PG, Rimoldi OE. The clinical value of myocardial blood flow measurement. J Nucl Med. 2009;50:1076–1087. doi: 10.2967/jnumed.108.054478. [DOI] [PubMed] [Google Scholar]
- 29.Lautamaki R, George RT, Kitagawa K, Higuchi T, Merrill J, Voicu C, et al. Rubidium-82 PET-CT for quantitative assessment of myocardial blood flow: validation in a canine model of coronary artery stenosis. Eur J Nucl Med Mol Imaging. 2009;36:576–586. doi: 10.1007/s00259-008-0972-1. [DOI] [PubMed] [Google Scholar]
- 30.El Fakhri G, Kardan A, Sitek A, Dorbala S, Abi-Hatem N, Lahoud Y, et al. Reproducibility and accuracy of quantitative myocardial blood flow assessment with 82Rb PET: comparison with 13N-ammonia PET. J Nucl Med. 2009;50:1062–1071. doi: 10.2967/jnumed.104.007831. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Camici PG. Absolute figures are better than percentages. JACC Cardiovasc Imaging. 2009;2:759–760. doi: 10.1016/j.jcmg.2009.04.005. [DOI] [PubMed] [Google Scholar]
- 32.Bergmann SR, Fox KA, Rand AL, McElvany KD, Welch MJ, Markham J, et al. Quantification of regional myocardial blood flow in vivo with H215O. Circulation. 1984;70:724–733. doi: 10.1161/01.CIR.70.4.724. [DOI] [PubMed] [Google Scholar]
- 33.Bol A, Melin JA, Vanoverschelde JL, Baudhuin T, Vogelaers D, De Pauw M, et al. Direct comparison of [13N]ammonia and [15O]water estimates of perfusion with quantification of regional myocardial blood flow by microspheres. Circulation. 1993;87:512–525. doi: 10.1161/01.cir.87.2.512. [DOI] [PubMed] [Google Scholar]
- 34.Diamond GA, Forrester JS. Analysis of probability as an aid in the clinical diagnosis of coronary artery disease. N Engl J Med. 1979;300:1350–1358. doi: 10.1056/NEJM197906143002402. [DOI] [PubMed] [Google Scholar]
- 35.Adachi I, Gaemperli O, Valenta I, Schepis T, Siegrist PT, Treyer V, et al. Assessment of myocardial perfusion by dynamic O-15-labeled water PET imaging: validation of a new fast factor analysis. J Nucl Cardiol. 2007;14:698–705. doi: 10.1016/j.nuclcard.2007.05.012. [DOI] [PubMed] [Google Scholar]
- 36.Senthamizhchelvan S, Bravo PE, Esaias C, Lodge MA, Merrill J, Hobbs RF, et al. Human biodistribution and radiation dosimetry of 82Rb. J Nucl Med. 2010;51:1592–1599. doi: 10.2967/jnumed.110.077669. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007;116:1290–1305. doi: 10.1161/CIRCULATIONAHA.107.688101. [DOI] [PubMed] [Google Scholar]
- 38.Yoshinaga K, Chow BJ, Williams K, Chen L, de Kemp RA, Garrard L, et al. What is the prognostic value of myocardial perfusion imaging using rubidium-82 positron emission tomography? J Am Coll Cardiol. 2006;48:1029–1039. doi: 10.1016/j.jacc.2006.06.025. [DOI] [PubMed] [Google Scholar]
- 39.Schelbert HR. Positron emission tomography of the heart: methodology, findings in the normal and the diseased heart, and clinical applications. In: Phelps ME, editor. Molecular imaging and its biological applications. 1. New York: Springer; 2004. pp. 389–508. [Google Scholar]
- 40.Royston P. sg1.2: nonlinear regression command. Stata Tech Bull. 1992;7:11–18. [Google Scholar]
- 41.Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45:255–268. doi: 10.2307/2532051. [DOI] [PubMed] [Google Scholar]
- 42.Lin LIK. Assay validation using the concordance correlation coefficient. Biometrics. 1992;48:599–604. doi: 10.2307/2532314. [DOI] [Google Scholar]
- 43.Correction: A Note on the Concordance Correlation Coefficient. Biometrics 2000;56:324–5.
- 44.Glatting G, Bergmann KP, Stollfuß JC, Weismüller P, Kochs M, Hombach V, et al. Myocardial Rb Extraction Fraction: Determination in Humans. Journal of the American College of Cardiology 1995;25:364A–5A.
- 45.Schelbert HR, Phelps ME, Huang SC, MacDonald NS, Hansen H, Selin C, et al. N-13 ammonia as an indicator of myocardial blood flow. Circulation. 1981;63:1259–1272. doi: 10.1161/01.CIR.63.6.1259. [DOI] [PubMed] [Google Scholar]
- 46.Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG. Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med. 1999;40:1848–1856. [PubMed] [Google Scholar]
- 47.Nagamachi S, Czernin J, Kim AS, Sun KT, Bottcher M, Phelps ME, et al. Reproducibility of measurements of regional resting and hyperemic myocardial blood flow assessed with PET. J Nucl Med. 1996;37:1626–1631. [PubMed] [Google Scholar]
- 48.Manabe O, Yoshinaga K, Katoh C, Naya M, De Kemp RA, Tamaki N. Repeatability of rest and hyperemic myocardial blood flow measurements with 82Rb dynamic PET. J Nucl Med. 2009;50:68–71. doi: 10.2967/jnumed.108.055673. [DOI] [PubMed] [Google Scholar]
- 49.Sdringola S, Johnson NP, Kirkeeide RL, Cid E, Gould KL. Impact of unexpected factors on quantitative myocardial perfusion and coronary flow reserve in young, asymptomatic volunteers. JACC Cardiovasc Imaging. 2011;4:402–412. doi: 10.1016/j.jcmg.2011.02.008. [DOI] [PubMed] [Google Scholar]
- 50.Yoshinaga K, Katoh C, Noriyasu K, Iwado Y, Furuyama H, Ito Y, et al. Reduction of coronary flow reserve in areas with and without ischemia on stress perfusion imaging in patients with coronary artery disease: a study using oxygen 15-labeled water PET. J Nucl Cardiol. 2003;10:275–283. doi: 10.1016/S1071-3581(02)43243-6. [DOI] [PubMed] [Google Scholar]
- 51.Uren NG, Crake T, Lefroy DC, de Silva R, Davies GJ, Maseri A. Reduced coronary vasodilator function in infarcted and normal myocardium after myocardial infarction. N Engl J Med. 1994;331:222–227. doi: 10.1056/NEJM199407283310402. [DOI] [PubMed] [Google Scholar]
- 52.Gould KL. Positron emission tomography in coronary artery disease. Curr Opin Cardiol. 2007;22:422–428. doi: 10.1097/HCO.0b013e3281fc9835. [DOI] [PubMed] [Google Scholar]
- 53.Sdringola S, Loghin C, Boccalandro F, Gould KL. Mechanisms of progression and regression of coronary artery disease by PET related to treatment intensity and clinical events at long-term follow-up. J Nucl Med. 2006;47:59–67. [PubMed] [Google Scholar]
- 54.Bengel FM. Leaving relativity behind: quantitative clinical perfusion imaging. J Am Coll Cardiol. 2011;58:749–751. doi: 10.1016/j.jacc.2011.02.068. [DOI] [PubMed] [Google Scholar]
- 55.Schleipman AR, Castronovo FP, Jr, Di Carli MF, Dorbala S. Occupational radiation dose associated with Rb-82 myocardial perfusion positron emission tomography imaging. J Nucl Cardiol. 2006;13:378–384. doi: 10.1016/j.nuclcard.2006.03.001. [DOI] [PubMed] [Google Scholar]
- 56.Siegrist PT, Husmann L, Knabenhans M, Gaemperli O, Valenta I, Hoefflinghaus T, et al. (13)N-ammonia myocardial perfusion imaging with a PET/CT scanner: impact on clinical decision making and cost-effectiveness. Eur J Nucl Med Mol Imaging. 2008;35:889–895. doi: 10.1007/s00259-007-0647-3. [DOI] [PubMed] [Google Scholar]
- 57.Kaufmann PA. 82-Rubidium – the dawn of cardiac PET in Europe? Eur J Nucl Med Mol Imaging. 2007;34:1963–1964. doi: 10.1007/s00259-007-0632-x. [DOI] [PubMed] [Google Scholar]
- 58.Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med. 2007;48:1112–1121. doi: 10.2967/jnumed.107.039792. [DOI] [PubMed] [Google Scholar]
- 59.PMOD Technologies. User's guide: PMOD Cardiac Modeling (PCARD). http://www.pmod.ch/technologies/pdf/doc/PCARD.pdf. Accessed 10 Feb 2012.
- 60.Klein R, Renaud JM, Ziadi MC, Thorn SL, Adler A, Beanlands RS, et al. Intra- and inter-operator repeatability of myocardial blood flow and myocardial flow reserve measurements using rubidium-82 pet and a highly automated analysis program. J Nucl Cardiol. 2010;17:600–616. doi: 10.1007/s12350-010-9225-3. [DOI] [PubMed] [Google Scholar]
- 61.Nitzsche EU, Choi Y, Czernin J, Hoh CK, Huang SC, Schelbert HR. Noninvasive quantification of myocardial blood flow in humans. A direct comparison of the [13N]ammonia and the [15O]water techniques. Circulation. 1996;93:2000–2006. doi: 10.1161/01.cir.93.11.2000. [DOI] [PubMed] [Google Scholar]
- 62.Gerber BL, Wijns W, Vanoverschelde JL, Heyndrickx GR, De Bruyne B, Bartunek J, et al. Myocardial perfusion and oxygen consumption in reperfused noninfarcted dysfunctional myocardium after unstable angina: direct evidence for myocardial stunning in humans. J Am Coll Cardiol. 1999;34:1939–1946. doi: 10.1016/S0735-1097(99)00451-9. [DOI] [PubMed] [Google Scholar]