Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2013 Apr 1.
Published in final edited form as: Int J Radiat Oncol Biol Phys. 2012 Apr 1;82(5):e709–e716. doi: 10.1016/j.ijrobp.2011.05.042

Mitigating Errors in External Respiratory Surrogate-Based Models of Tumor Position

Kathleen T Malinowski 1,2, Thomas J McAvoy 2,3, Rohini George 1, Sonja Dieterich 4, Warren D D’Souza 1,2
PMCID: PMC3346954  NIHMSID: NIHMS360512  PMID: 22429333

Abstract

Purpose

To investigate the impact of tumor site, measurement precision, tumor-surrogate correlation, training data selection, model design, and inter-patient and inter-fraction variations on the accuracy of external-marker based models of tumor position.

Methods and Materials

Cyberknife Synchrony system log files comprising synchronously acquired positions of external markers and the tumor from 167 treatment fractions were analyzed. The accuracies of Synchrony, ordinary-least-squares (OLS) regression, and partial-least-squares (PLS) regression models for predicting tumor position from external markers were evaluated. Quantity and timing of data used to build the predictive model were varied. Effects of tumor-surrogate correlation and of the precision in both tumor and external surrogate position measurements were explored by adding noise to the data.

Results

Tumor position prediction errors increased over the duration of a fraction. Increasing training data quantities did not always lead to more accurate models. Adding uncorrelated noise to the external marker-based inputs degraded tumor-surrogate correlation models by 16% for PLS and 57% for OLS. External marker and tumor position measurement errors led to tumor position prediction changes 0.3 to 3.6 times the magnitude of measurement errors, varying widely with model algorithm. Tumor position prediction errors were significantly associated with patient index but not with fraction index or tumor site. PLS was as accurate as Synchrony and more accurate than OLS.

Conclusions

The accuracy of surrogate-based inferential models of tumor position was affected by all of the investigated factors except tumor site and fraction index.

Keywords: intra-fraction motion, tumor tracking, tumor-surrogate relationship, partial-least-squares regression, tumor-surrogate predictive models

I. Introduction

Respiration-induced tumor motion degrades radiation therapy targeting accuracy.13 While 4DCT and brief fluoroscopic movies can be used to measure tumor motion over a few respiratory cycles, accurately predicting the extent of tumor motion during treatment often cannot be accomplished because of day-to-day and breath-to-breath variations.4 Furthermore, a transient event such as a deep breath or a cough can cause the tumor to move out of its usual range of positions.59 As a result, motion management technologies correcting for real-time tumor motion must accurately track the tumor.811

Direct tracking systems, which measure the position of the tumor itself, either continuously image or electromagnetically localize the tumor. Fluoroscopically imaging over the entire treatment fraction12 imparts unnecessary ionizing radiation to the healthy tissue. While electromagnetic signals can be used to track implanted fiducial markers without ionizing radiation,13 at the time of this writing the technology has only been approved for use in the prostate.

Indirect tracking systems rely on external surrogates of respiratory motion, such as optical markers affixed to the torso, to predict the position of the tumor.810,1417 Tumor motion models based on the positions of multiple markers affixed to the torso show particular promise as alternatives to direct tracking systems.910,1416

Most studies to date of surrogate-based tumor localization do not explore potential variations in the model or explain sources of error. One notable exception is the study by Yan et al.,16 which considers both single and multiple marker models and quantifies the impact of shifting input data in time to account for marker-tumor phase offsets. While the literature has suggested that such issues as signal-to-noise ratio (SNR)18 and variations in the tumor-surrogate relationship7,1921 might lead to indirect tumor tracking system errors, to our knowledge no study to date has provided a systematic investigation of factors that may contribute to the accuracy of surrogate-based predictive models of tumor position. This information would make it possible to more accurately localize the tumor in real time.

In this study, we evaluate potential sources of error in tumor position prediction models. The primary purpose of this study was to investigate the impacts of: (1) tumor site, (2) tumor and external surrogate measurement precision, (3) tumor-surrogate correlation, (4) training data selection, (5) model design, (6) inter-patient variations, and (7) inter-fraction variations on the accuracy of external marker-based models of tumor position in the context of three modeling algorithms: ordinary-least-squares (OLS) regression, the Cyberknife Synchrony algorithm, and partial-least-squares (PLS) regression. The secondary goal of this study was to evaluate PLS regression for modeling tumor motion from external surrogates. Preliminary studies10 have indicated that PLS may be more accurate than computationally simpler multiple-marker linear algorithms such as OLS and Synchrony because of the collinearity in external markers’ motion.22

II. Methods and Materials

II.A. Position Data

Data consisted of Synchrony (Accuray Inc., Sunnyvale, CA) system log files comprising 113, 10, and 44 treatment fractions from 61, 5, and 22 lung, liver, and pancreas patients, respectively, who underwent stereotactic body radiation therapy at Georgetown University Hospital, Washington, DC. These data were obtained under an Institutional Review Board protocol. The files included 3D positions of: (1) the centroid of a set of 2–3 fiducials implanted in the tumor that were measured via stereoscopic x-rays, henceforth referred to as the tumor position; and (2) three markers affixed to the patient’s form-fitting vest that were tracked optically. The position of the tumor, identified as the centroid of the set of implanted fiducials, was aligned in time with those of the external markers according to the timestamps in the system log files. Each dataset included 40–112 (mean=62) concurrent tumor and marker localizations spaced at a mean interval of 66 sec. The log files also included the continuous real-time output of the Synchrony inferential model.

II.B. Applying and Testing Inferential Models

For each fraction, a series of models were tested. First, a model was fit (Section II.C) to a training dataset of concurrent tumor and external marker positions. This model was used to predict tumor positions from other external marker measurements in the same treatment fraction. Each predicted tumor position, derived from marker measurements alone, was validated against the tumor position measured from stereoscopic x-rays. This process was repeated for training datasets of each set of N consecutive measurements in a treatment fraction, each time testing the model against the tumor localizations that were acquired during that fraction after the training dataset. Prediction error was defined as the magnitude of the vector displacement between the stereoscopic x-ray-measured tumor position and the modeled tumor position.

II.C. Model Design

II.C.1. Input Data Pre-processing

In addition to the raw external marker positions, an alternative input was created by projecting marker displacements onto a single dimension of motion, defined as ri (Figure 1). The motivation for comparing raw 3D inputs to projected, 1D inputs was to better understand the impact of this data processing technique on the Cyberknife Synchrony algorithm’s performance. As described by Sayeh et al.,23 the Cyberknife Synchrony system uses ri as the inputs to its tumor position models.

Figure 1.

Figure 1

Example of (a) 3D surrogate marker motion data (solid) projected onto its first principle component (arrow) line (dashed) to obtain (b) a 1D representation of respiration-induced surrogate marker motion.

To calculate ri(t) for a marker i, the projection line was determined from the three-column matrix of marker positions acquired during training data collection. The direction of the line was given by the first principal component vector of the marker positions, which is defined as the first eigenvector of the marker data covariance matrix.24 The projection line also was designed to intersect the average 3D marker position, mi, of the marker data. Each 3D marker position was then projected orthogonally onto the projection line. For each position (xi(t), yi(t), zi(t)), ri(t) was defined as the distance between the projected position of the sample on the line and mi.

II.C.2. Model Algorithms

Three model fitting algorithms were considered: OLS regression, PLS regression, and the Synchrony algorithm. OLS and PLS are mutlilinear models that operate according to Ŷ = X · B, where X is matrix of inputs (respiratory surrogate positions) with m measurements and n inputs, B is an n × 3 matrix of regression coefficients, and Ŷ is a matrix of m × 3 outputs (predicted tumor positions).

OLS defines B as the set of regression coefficients that minimizes the expression ||YX · B||2. The OLS solution is given by B = X + · Y, where X + is the Moore-Penrose pseudo-inverse of X. For full-rank, real X, X + is defined as X + = (X T · X)−1 · XT. For rank-deficient, real X, X + = V · Σ+ · UT, where U, Σ, and V are the matrices in the singular value decomposition of X given by X =U · Σ · V. The terms of Σ+ are given by i,j+=(i,j)1 for Σi, j ≠ 0 and i,j+=0 for Σi, j =0.

PLS regression finds the regression coefficients that minimize the expression 12BT·XTX·B(XTY)T·B.25 To accomplish this task, PLS regression first compresses the input data matrix, X, into an m × A matrix of scores, T = [t1, t2, …, tA], where A is typically less than the number of variables in the raw input data. The scores, ti, are orthogonal, linear combinations of the columns of X. The optimal number of scores, A, for each dataset was determined through a cross-validation process, in which models based on various numbers of scores were created, each time leaving one sample out of the training data and using it to evaluate the model. In most cases 1–3 scores were utilized in the PLS model. The SIMPLS algorithm26 was applied to determine the PLS regression coefficients. The scores of Y, ui for i = 1 to A, were linear combinations of the columns of Y that are chosen so as to maximize the covariance between ti and ui. The first X score, t1, was given by t1 = X · XT · Y 0/norm(X · XT · Y), where Y0 is mean-centered Y. The first X weight, r1, and the first X basis, v1, were both unity. The Y loadings and scores were calculated as qi=Y0T·ti and ui = Y0 · qi, respectively. The X basis was updated with each iteration as vi=vi1Vi1·(Vi1T·(tT·X)T), where Vi1 = [v1, v2, …, vi1]. Subsequent X weights and scores were calculated as ri=(Si1vi·(viT·Si1))·qi and ti =X · ri. The regression coefficient matrix, B, was given by B =R · QT, where R = [r1, r2, …, rA] and Q = [ q1, q2, …, qA].

The inferential model in the Synchrony system was reproduced for this study based on the description by Sayeh et al.23 so that the training data and timing could be controlled for comparison to OLS and PLS results. The algorithm can be stated briefly as follows. Initially, separate linear, quadratic, or hybrid models were developed for each of the three markers. The linear and quadratic models were defined by Ŷi = A1,iri + A0,i and Y^i=A2,iri2+A1.iri+A0,i, respectively. For quadratic and hybrid models, separate coefficients were calculated for inhalation and exhalation. The final estimated tumor position was given by the average of the outputs of models using the three markers. The system selects between linear or quadratic models through a modified standard error function that selects for increased accuracy and reduced computational complexity.23 In the clinical Cyberknife Synchrony system, the operator selects both the number of images used to train the model and the frequency of images acquired during treatment. Our implementation of the Synchrony model, SYNr, based on 15 tumor localizations (the typical training data length in the clinic from which our data originated) was validated against the actual logged 15-sample model output. In the SI, ML and AP directions, respectively, the mean (± standard deviation) distance between SYNr and actual Synchrony outputs were 0.5 ± 1.2 mm, 0.5 ± 1.7 mm, and 0.3 ± 0.5 mm. For this study, only the linear form was considered, because more than 70% of cases in the database did not utilize quadratic models when the modified standard error function was applied.

Both the OLSr and the SYNr algorithms utilize an ordinary linear regression step. The SYNr algorithm differs from OLSr in that SYNr creates separate linear models for each of the 3 markers and averages their outputs to predict the tumor position. In contrast, the OLSr model regresses on one matrix comprising all of the marker data, leading to a different set of regression coefficients.

II.C.3. Summary of Models Evaluated

Five tumor position prediction models were considered: OLSxyz, OLSr, PLSxyz, PLSr, and SYNr. OLSxyz and PLSxyz models utilized the 3D marker data. OLSr, PLSr, and SYNr models utilized the projected marker data.

II.D. Training Data Selection

II.D.1. Quantity of Training Data

Training datasets of N = 6 to N = 35 samples of measured tumor and external surrogate positions were evaluated for their accuracy in inferring tumor positions that occurred subsequent to the training data. Mean and 95th percentile errors over 20 minutes of data were evaluated.

II.D.2. Time Since Training Data Acquisition

Errors for models in which 6 samples of concurrent external marker and tumor positions comprised the training data were binned by time elapsed since the end of training data.

II.D.3. Tumor-Surrogate Correlation

The impact of uncorrelated external surrogate inputs on tumor position prediction model accuracy was explored in OLSr, PLSr, and SYNr models. For each model type, models with 6 samples of concurrent external marker and tumor positions were created. Then new models were trained on the same training data (6 samples × 3 inputs) plus an additional input vector of Gaussian noise (6 random values × 1 input) for a total of 4 inputs. The accuracy of models with and without the additional noise input were compared.

II.E. Inter-Patient and Inter-Fraction Variations

Each patient underwent 1–5 treatment fractions. The mean tumor-surrogate model prediction error for each treatment fraction was determined. Using the 46 patients for whom more than one treatment fraction of data was available, the Kruskal-Wallis one-way analysis of variance test was applied to determine whether the mean error of a fraction was significantly associated with either patient index or fraction index.

II.F. Measurement Precision

Tumor-surrogate models created from the training data were compared to models with simulated noise in either input (external marker) or output (tumor) measurements. In each case, models based on 6 training samples were evaluated on the testing dataset.

To simulate radiographic tumor localization uncertainties, noise from a Gaussian distribution was added to the radiographic tumor localizations. Noise levels were varied by randomly selecting the standard deviation of the Gaussian noise in each trial. Another set of models was then created from training data with noise and used to determine new predicted tumor positions in the testing dataset. This process was repeated for noise added to external marker localizations.

III. Results

III.A. Model Design

For OLSxyz, PLSxyz, OLSr, PLSr, and SYNr models trained on 6 samples, 22.8%, 12.7%, 19.0%, 9.1%, and 9.5% of the predictions, respectively, exceeded 0.5 cm (Figure 2). The OLSxyz, PLSxyz, and OLSr distributions peaked in the 0.1 cm to 0.2 cm range, and the PLSxyz, PLSr, and SYNr distributions peaked in the 0–0.1 cm range. PLSr and SYNr results did not differ significantly for any quantity of training points.

Figure 2.

Figure 2

Histogram of radial errors from the first minute of testing samples for models based on 6 training samples. The last bin comprises all errors greater than 0.5 cm.

III.B. Training Data Selection

III.B.1. Quantity of Training Data

The 95th percentile and mean tumor position prediction errors for each number of training samples are shown in Figure 3. The OLSxyz error peaked to an average of 7.7 cm at N = 10. OLSxyz, and OLSr 95th percentile errors exceeded 1 cm for some values of N, but PLSxyz, PLSr, and SYNr 95th percentile errors did not exceed 1 cm in the tested range of N. Mean errors for the six models converged to approximately 0.2 cm for N > 20.

Figure 3.

Figure 3

Mean and 95th percentile tumor position model prediction errors in the test data acquired over 20 minutes for (a) OLSxyz, (b) PLSxyz, (c) OLSr, (d) PLSr, and (e) SYNr models. For clarity, the y-axis in (a) is scaled differently than that of (b–e).

III.B.2. Time Since Training Data Acquisition

Average errors for each model type increased as time elapsed after the end of the training data (Figure 4). Mean errors in 0–2 min and in 10–20 min of data increased from 0.3 to 1.3 cm for OLSxyz, 0.2 to 0.5 cm for PLSxyz, 0.2 to 0.6 cm for OLSr, 0.2 to 0.4 cm for PLSr, and 0.2 cm to 0.4 cm for SYNr.

Figure 4.

Figure 4

Mean and standard deviation (error bars) of tumor position model prediction errors, binned by time elapsed since the end of training data collection.

* The standard deviation of the OLSxyz error bar in the >20 min bin is 2.3 cm.

III.C. Tumor-Surrogate Correlation

The uncorrelated (Gaussian noise) input degraded prediction models (Figure 5). Mean errors in the 0–2 min bin increased by 57% for OLSr and by 16% for PLSr.

Figure 5.

Figure 5

The impact of an uncorrelated (Gaussian noise) input variable in OLSr and PLSr models. Mean and standard deviation of model errors indicate that the additional input increased errors. In the 0–2 min bin, OLSr errors increased by 57%, and PLSr errors increased by 16%.

III.D. Tumor Site

For each modeling algorithm, error distributions were comparable across the three tumor sites (Figure 6). Lung, liver, and pancreas mean errors were 2.2 mm, 1.9 mm, and 2.3 mm, respectively, and further details are described in Table I. Of the three sites, only the pancreas errors differed significantly (Kruskal-Wallis one-way ANOVA, p < 0.05) from errors of the other sites. However, it can be argued that the 0.1 mm mean error difference between lung and pancreas cases is not clinically significant. It is not clear whether liver results would differ significantly from lung results with a larger sample size.

Figure 6.

Figure 6

Histograms of mean error in each fraction by tumor site for (a) OLSxyz, (b) PLSxyz, (c) OLSr, (d) PLSr, and (e) SYNr models, each based on 6 training data points.

Table I.

Errors for models based on 6 training points, stratified by tumor site

Mean ± standard deviation of error (mm)
Model Lung Liver Pancreas
OLSxyz 3.0 ± 4.5 2.7 ± 2.7 3.3 ± 4.6
PLSxyz 1.9 ± 2.0 1.7 ± 1.3 2.0 ± 2.0
OLSr 2.5 ± 3.4 2.1 ± 2.6 2.6 ± 2.8
PLSr 1.8 ± 1.6 1.6 ± 1.1 1.8 ± 1.5
Synchrony 1.7 ± 1.6 1.6 ± 1.0 1.8 ± 1.5

Pooled results 2.2 ± 2.9 1.9 ± 1.9 2.3 ± 2.8

III.E. Inter-Patient and Inter-Fraction Variations

For each of the model types, the model error in a single fraction was significantly (p < 0.05) associated with the patient from whom the fraction was recorded. This pattern is evident in Figure 7, in which the mean errors of the four fractions of patient 2, for instance, cluster around a value that differs from the value associated with patient 8. The mean model error over a fraction was not significantly (p > 0.05) associated with the treatment fraction index.

Figure 7.

Figure 7

Mean and standard deviation of 9-training-datapoint PLSr in each fraction for 9 patients. Errors represent samples acquired within 20 minutes of the training dataset. For clarity, the figure is limited to the 9 patients with >3 treatment fractions available for analysis; the results for these patients are consistent with the dataset as a whole.

III.F. Measurement Precision

When noise was added to measurements of radiographic tumor positions in the training dataset (Figure 8), the resulting average error in the testing dataset varied as 1.1, 0.3, and 0.4 times the average noise for OLSr, PLSr, and SYNr, respectively; noise-to-error correlations ranged from 0.44 to 0.53. When noise was added to the measurement of the external surrogate position data, the resulting average error in the testing dataset varied as 3.7, 0.6, and 0.6 times the average noise for OLSr, PLSr, and SYNr, respectively, and noise-to-error correlations ranged from 0.09 to 0.32.

Figure 8.

Figure 8

(a) Tumor and (b) external marker localization noise (in the training data used to build the model that predict tumor position) versus tumor position prediction error. Best fit lines represent least-squared-error linear regression fits with an assumption of zero error for zero noise. OLSr regression lines have slopes of (a) 1.1 and (b) 3.7. The PLSr line overlies the SYNr line in (b).

IV. Discussion

In this study, a series of factors impacting the accuracy of respiratory surrogate models of tumor motion were explored. Model accuracy was affected by inter-patient variations, tumor and external surrogate measurement precision, tumor-surrogate correlation, training data selection, tumor-surrogate correlation, and model design. Tumor site and fraction index were not predictive of model accuracy. These results provide the reader with a framework for designing and evaluating a surrogate-based tumor position prediction model. In addition, PLS models were more accurate than OLS models and were as accurate as Synchrony models.

Many studies reporting high tumor-surrogate correlations have evaluated only a few minutes of data.21,27 The results of these studies may not be representative of the behavior of the model over entire treatment fractions. Analyses of datasets capturing motion over the duration of a treatment fraction have reported variations in the tumor-surrogate relationship.78,19 In this study, model errors increased over time; we attribute this increasing error to tumor-surrogate relationship changes. In practice, this effect could be overcome by updating the model during the treatment fraction. Intra-fraction model updates have been utilized by tracking systems like Cyberknife Synchrony, but model updates are not common in gated treatment protocols.

In this work, projecting external marker motion onto a single dimension improved accuracy by increasing the correlation between surrogate signals and tumor motion. Shifting the surrogate signal in time relative to the tumor signal is a common method for increasing tumor-surrogate correlation.16,19,2829 However, because of changes in the tumor-surrogate phase offset, these shifted surrogates are not likely to maintain their improved correlation over time.7,19

PLSr and SYNr results did not differ significantly. We investigated other model designs (not described in detail here) but were unable to reduce model errors below those of the PLSr model. Because the continuous external marker motion in the data log files appears to indicate a high signal-noise ratio (SNR), we hypothesize that the accuracy of any model based on this dataset is limited by precision of the gold-standard radiographic tumor localizations. PLSr and SYNr represent the state-of-the-art in indirect tumor localization algorithms. In addition to their accuracy, each requires only milliseconds to derive tumor position from surrogate data. Subject to the limitations of SNR and the other factors investigated in this study, both PLSr and SYNr are candidates for real-time applications.

Decreased measurement precision in either gold-standard tumor localizations or external surrogate measurements was found to have considerable impact on model accuracy. For OLSr models, the errors in tumor position prediction induced by measurement precision limitations were 1–4 times as large as the measurement errors themselves. For PLSr and Synchrony models, the errors in tumor position prediction were 0.3–0.6 times as large as the measurement errors. Many optical (surrogate) tracking systems are capable of achieving sub-millimeter accuracy,3031 but infrared tracking accuracies exceeding one millimeter have been reported in clinically available devices.32 Typical errors of x-ray-based localizations of bony anatomy or fiducial tracking are 0.6–2 mm.32

Surprisingly, utilizing a larger quantity of training data, whether in dimensionality of marker motion or in number of training data points, did not improve the accuracy of tumor position predictions. In the case of OLS models, the error peaked at the transition between underdetermined and overdetermined systems. For any modeling algorithm, the optimal number of training points will also vary as a function of spacing between training samples. Training data captured over a period of minutes may encompass tumor-surrogate relationship changes that would not be present in training data captured in a shorter period of time.

PLS was consistently more accurate than OLS, a conclusion in agreement with the diaphragm tracking work of Qiu et al.10 OLS regression coefficients are volatile when inputs are collinear, tending to change considerably for different training samples from the same dataset.22 As a result, PLS has been described a superior alternative to OLS when the inputs are highly collinear.33

Yan et al. has shown that multiple-surrogate models of tumor motion can be more accurate than single-marker models16 by overcoming the location effect, in which tumor-surrogate correlation varies with marker placement.15 Nevertheless, our results support the conclusion that incorporating additional surrogate-based inputs that are uncorrelated with tumor motion may actually degrade the model. Fortunately, respiration-induced tumor motion is correlated with such respiratory surrogates as markers affixed to the torso, spirometry, and bellows systems.7,16,34

Finally, patient-specific, fraction-specific, and site-specific results were evaluated. Results did not differ between lung, liver, and pancreas cancers. Furthermore, the model error was not found to be significantly associated with fraction index. Patient index, on the other hand, was significantly associated with model accuracy. The practical implication of these results is that the design of a study to evaluate tumor motion models should use a large enough group of patients to obtain statistically significant results; multi-site and inter-fraction data are less important unless a model will be applied without revision on multiple treatment days.

V. Conclusions

The accuracy of tumor position prediction models using external surrogates was affected by inter-patient variations, measurement precision, tumor-surrogate correlation, training data selection, and model design, but tumor site and fraction index were not predictive of model accuracy. PLS regression models were more accurate OLS models and as accurate as Synchrony models.

Footnotes

Conflicts of Interest Notification: Supported in part by grant CA124766 from the NIH/NCI and by the Achievement Rewards for College Scientists scholarship.

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Keall PJ, Mageras GS, Balter JM, et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys. 2006;33:3874–3900. doi: 10.1118/1.2349696. [DOI] [PubMed] [Google Scholar]
  • 2.Wu QJ, Thongphiew D, Wang Z, et al. The impact of respiratory motion and treatment technique on stereotactic body radiation therapy for liver cancer. Med Phys. 2008;35:1440–1451. doi: 10.1118/1.2839095. [DOI] [PubMed] [Google Scholar]
  • 3.Chaudhari SR, Goddu SM, Rangaraj D, et al. Dosimetric variances anticipated from breathing- induced tumor motion during tomotherapy treatment delivery. Phys Med Biol. 2009;54:2541–2555. doi: 10.1088/0031-9155/54/8/019. [DOI] [PubMed] [Google Scholar]
  • 4.Nelson C, Starkschall G, Balter P, et al. Assessment of lung tumor motion and setup uncertainties using implanted fiducials. IJROBP. 2007;67:915–923. doi: 10.1016/j.ijrobp.2006.10.033. [DOI] [PubMed] [Google Scholar]
  • 5.Suh Y, Dieterich S, Cho B, et al. An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients. Phys Med Biol. 2008;53:3623–3640. doi: 10.1088/0031-9155/53/13/016. [DOI] [PubMed] [Google Scholar]
  • 6.Nehmeh SA, Erdi YE, Pan T, et al. Quantitation of respiratory motion during 4D-PET/CT acquisition. Med Phys. 2004;31:1333–1338. doi: 10.1118/1.1739671. [DOI] [PubMed] [Google Scholar]
  • 7.Ozhasoglu C, Murphy M. Issues in respiratory motion compensation during external-beam radiotherapy. IJROBP. 2002;52:1389–1399. doi: 10.1016/s0360-3016(01)02789-4. [DOI] [PubMed] [Google Scholar]
  • 8.Berbeco RI, Nishioka S, Shirato H, et al. Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates. Phys Med Biol. 2005;50:3655–3667. doi: 10.1088/0031-9155/50/16/001. [DOI] [PubMed] [Google Scholar]
  • 9.Seppenwoolde Y, Berbeco RI, Nishioka S, et al. Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: a simulation study. Med Phys. 2007;34:2774–2784. doi: 10.1118/1.2739811. [DOI] [PubMed] [Google Scholar]
  • 10.Qiu P, D’Souza WD, McAvoy TJ, et al. Inferential modeling and predictive feedback control in real-time motion compensation using the treatment couch during radiotherapy. Phys Med Biol. 2007;52:5831–5854. doi: 10.1088/0031-9155/52/19/007. [DOI] [PubMed] [Google Scholar]
  • 11.Vedam SS, Keall PJ, Kini VR, et al. Determining parameters for respiration-gated radiotherapy. Med Phys. 2001;28:2139–2146. doi: 10.1118/1.1406524. [DOI] [PubMed] [Google Scholar]
  • 12.Shirato H, Shimizu S, Kitamura K, et al. Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor. IJROBP. 2000;48:435–442. doi: 10.1016/s0360-3016(00)00625-8. [DOI] [PubMed] [Google Scholar]
  • 13.Santanam L, Malinowski K, Hubenshmidt J, et al. Fiducial-based translational localization accuracy of electromagnetic tracking system and on-board kilovoltage imaging system. IJROBP. 2008;70:892–899. doi: 10.1016/j.ijrobp.2007.10.005. [DOI] [PubMed] [Google Scholar]
  • 14.George R, Suh Y, Murphy M, et al. On the accuracy of a moving average algorithm for target tracking during radiation therapy treatment delivery. Med Phys. 2008;35:2356–1265. doi: 10.1118/1.2921131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Yan H, Zhu G, Yang J, et al. Investigation of the location effect of external markers in respiratory-gated radiotherapy. J App Clin Med Phys. 2008;9:57–68. doi: 10.1120/jacmp.v9i2.2758. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Yan H, Yin F, Zhu G, et al. The correlation evaluation of a tumor tracking system using multiple external markers. Med Phys. 2007;33:4073–4084. doi: 10.1118/1.2358830. [DOI] [PubMed] [Google Scholar]
  • 17.Yang D, Lu W, Low DA, et al. 4D-CT motion estimation using deformable image registration and 5D respiratory motion modeling. Med Phys. 2008;35:4577–4590. doi: 10.1118/1.2977828. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ionascu D, Jiang SB, Nishioka S, et al. Internal-external correlation investigations of respiratory induced motion of lung tumors. Med Phys. 2007;34:3893–3903. doi: 10.1118/1.2779941. [DOI] [PubMed] [Google Scholar]
  • 19.Hoisak JDP, Sixel KE, Tirona R, et al. Correlation of lung tumor motion with external surrogate indicators of respiration. IJROBP. 2004;60:1298–1306. doi: 10.1016/j.ijrobp.2004.07.681. [DOI] [PubMed] [Google Scholar]
  • 20.Hugo G, Vargas C, Liang J, et al. Changes in the respiratory pattern during radiotherapy for cancer in the lung. Radiother Oncol. 2006;78:326–31. doi: 10.1016/j.radonc.2006.02.015. [DOI] [PubMed] [Google Scholar]
  • 21.Gierga DP, Brewer J, Sharp GC, et al. The correlation between internal and external markers for abdominal tumors: implications for respiratory gating. IJROBP. 2005;61:1551–1558. doi: 10.1016/j.ijrobp.2004.12.013. [DOI] [PubMed] [Google Scholar]
  • 22.Allen MP. Understanding Regression Analysis. New York: Plenum Press; 1997. pp. 99–100. [Google Scholar]
  • 23.Sayeh S, Wang J, Main WT, et al. Respiratory Motion Tracking for Robotic Radiosurgery. In: Urschel HC, editor. Treating Tumors that Move with Respiration. New York: Springer; 2007. pp. 15–29. [Google Scholar]
  • 24.Wise BM, Gallagher NB. The process chemometrics approach to process monitoring and fault detection. J Proc Cont. 1996;6:329–348. [Google Scholar]
  • 25.Rosipal R, Krämer N. Overview and recent advances in partial least squares. In: Saunders C, Grobelnik M, Gunn S, et al., editors. Subspace, Latent Structure and Feature Selection. New York: Springer; 2006. pp. 34–51. [Google Scholar]
  • 26.de Jong S. SIMPLS: an alternative approach to partial least squares regression. Chemomet Intel Lab Sys. 1993;18:251–263. [Google Scholar]
  • 27.Beddar AS, Kainz K, Briere TM, et al. Correlation between internal fiducial tumor motion and external marker motion for liver tumors imaged with 4D CT. IJROBP. 2007;67:630–638. doi: 10.1016/j.ijrobp.2006.10.007. [DOI] [PubMed] [Google Scholar]
  • 28.Cerviño LI, Chao AK, Sandhu A, Jiang SB. The diaphragm as an anatomic surrogate for lung tumor motion. Phys Med Biol. 2009;54:3529–3541. doi: 10.1088/0031-9155/54/11/017. [DOI] [PubMed] [Google Scholar]
  • 29.Vedam SS, Kini VR, Keall PJ, et al. Quantifying the predictability of diaphragm motion during respiration with a noninvasive external marker. Med Phys. 2003;30:505–513. doi: 10.1118/1.1558675. [DOI] [PubMed] [Google Scholar]
  • 30.Meeks SL, Bova FJ, Wagner TH, et al. Image localization for frameless stereotactic radiotherapy. IJROBP. 2000;46:1291–1299. doi: 10.1016/s0360-3016(99)00536-2. [DOI] [PubMed] [Google Scholar]
  • 31.Phillips MH, Singer K, Miller E, et al. Commissioning an image-guided localization system for radiotherapy. IJROBP. 2000;48:267–276. doi: 10.1016/s0360-3016(00)00581-2. [DOI] [PubMed] [Google Scholar]
  • 32.Willoughby TR, Forbes AR, Buchholz D, et al. Evaluation of an infrared camera and X-ray system using implanted fiducials in patients with lung tumors for gated radiation therapy. IJROBP. 2006;66:568–75. doi: 10.1016/j.ijrobp.2006.05.029. [DOI] [PubMed] [Google Scholar]
  • 33.Abdi H. Partial least squares regression (PLS-regression) In: Lewis-Beck M, Bryman A, Futing T, editors. Encyclopedia for research methods for the social sciences. Thousand Oaks, CA: Sage; 2003. pp. 792–795. [Google Scholar]
  • 34.Werner R, White B, Handels H, et al. Technical note: development of a tidal volume surrogate that replaces spirometry for physiological breathing monitoring in 4D CT. Med Phys. 2010;37:615–9. doi: 10.1118/1.3284282. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES