Abstract
A data modeling and processing method for electron paramagnetic resonance (EPR)-based pH spectroscopy is presented. The proposed method models the EPR spectrum of a pH-sensitive probe in both protonated and unprotonated forms. Under slow-exchange conditions, the EPR spectrum of a sample with an unknown pH value can be accurately represented by a weighted sum of the two models, with the pH value completely determined by their relative weights. Unlike traditional pH spectroscopy, which relies on locating resonance peaks, the proposed modeling-based approach utilizes the information from the entire scan and hence leads to more accurate estimation of pH for a given acquisition time. By employing the proposed methodology, we expect a reduction in the pH estimation error by more than a factor of three, which represents an order of magnitude reduction in acquisition time compared to the traditional method.
Keywords: EPR, pH, acidosis, modeling, nitroxide, spectroscopy
1. Introduction
Perturbation in pH homeostasis is associated with several pathological states, including renal disease [1], inflammation [2], ischemia [3], chronic lung disorder [4], intrauterine abnormalities [5], and stomach ulceration [6]. Also, certain stress conditions–such as biochemical shock, interruption of normal blood supply, and high exercise levels–may suppress the body’s ability to self-regulate local pH values [7, 8]. More importantly, local pH values may interfere with delivery, absorption, and pharmacological effectiveness of drugs, and with efficacy of different treatment regimens [9, 10]. For example, in tumors, extracellular pH is lower than in normal tissue and can be correlated with the disease progression, prognosis, and response to radiotherapy and chemotherapy [11, 12]. Therefore, accurate measurement of region-specific pH values using noninvasive methods carries considerable biomedical and clinical relevance.
Electron paramagnetic resonance (EPR) is a branch of spectroscopy in which electrons with unpaired spins, when placed in a magnetic field, absorb electromagnetic radiation to transition from a low-energy level to a high-energy level. Environment-dependent spectral changes of paramagnetic materials have led to widespread biological application of EPR, including measurement of oxygen [13], pH [14], perfusion [15], redox metabolism [16], and detection of short-lived radicals [17].
Measurement of pH using EPR relies on quantifiable and reversible changes in the spectra of exogenous pH probes [18]. The imidazoline and imidazolidine nitroxide radicals (NRs) are the most commonly used pH-sensitive EPR probes [19]. Figure 1 shows 4-amino-2,2,5,5-tetramethyl-3-imidazoline-1-yloxy (ATI) [20], a commonly used imidazolidine NR. The protonation of the N-3 atom of ATI leads to a decrease in unpaired electron density at the N-1 atom and is manifested in the EPR spectrum as a decrease in the hyperfine coupling constant.
Figure 1.

Protonation of 4-amino-2,2,5,5-tetramethyl-3-imidazoline-1-yloxy (ATI), an imidazoline nitroxide radical commonly used for EPR-based pH measurements. The pKa value of ATI has been reported to be 6.1.
Figure 2 shows the EPR spectrum of ATI, with three resonance peaks, measured using a continuous-wave (CW) L-band (1.2 GHz) EPR spectrometer at two different pH values, one on either side of pKa. The sample at pH = 4.00 is primarily comprised of the protonated form (RH+), while the sample at pH = 8.30 is primarily comprised of the unprotonated form (R) of the NR. A noticeable difference between the spectra of RH+ and R is the relative location of the resonance peaks. Traditionally, pH is inferred from one of two spectral characteristics: either location of a resonance peak with respect to a fixed reference point or the separation between two resonance peaks. In either case, the quantity measured from spectrum is compared to a calibration curve to yield a pH estimate.
Figure 2.
Three-peak EPR spectrum of ATI collected at two different pH values using a CW L-band spectrometer. The spectra were collected with 0.036 mT field modulation amplitude and 5.01 mT field sweep. The vertical lines provide a visual comparison of the peak locations. Here, ‘o’ represents zero-crossings of the resonance peaks.
In CW EPR, where the first field-modulation harmonic of the absorption spectrum is observed, the locations of resonance peaks are generally identified by measuring the zero-crossings of the spectrum, as shown in Fig. 2. In some studies, peak locations in the integrated spectrum have been used in lieu of zero-crossings [14]. In general, the location of the highest-field peak displays the greatest difference between R and RH+, and hence is the most sensitive indicator of pH.
EPR is inherently more sensitive than nuclear magnetic resonance due to the larger magnetic moment of an electron compared to that of a proton. EPR is also superior in terms of functional specificity due to the absence of overlapping EPR signals from endogenous species. Recent development of low-frequency CW and pulsed EPR systems [21, 22] along with development of sensitive, stable, and nontoxic pH probes [23, 24], has broadened the applicability of EPR-based techniques. In fact, this technique has already been used to measure the stomach pH in live murine subjects [25], monitor pH of tumor tissue in live mice [26], characterize release processes in biodegradable polymers [27], and study the influence of a drug on the microacidity of skin in humans and rats [28].
Despite recent efforts, EPR data acquisition for pH spectroscopy and imaging remains a time-consuming proposition, especially when high measurement accuracy is desired. For in vivo applications, where conditions may change rapidly, it is desirable to accelerate the acquisition process of pH spectroscopy without sacrificing measurement accuracy. A major shortcoming of the conventional EPR-based pH measurement method, called hyperfine coupling method (HCM), is its reliance on a small fragment, e.g., zero-crossing, of the noisy measured spectrum. In this work, we propose and verify a modeling-based approach, called parametric modeling method (PMM), that utilizes the information from the entire measured spectrum to improve the measurement accuracy.
The remainder of the paper is organized as follows: Section 2 describes parametric modeling of EPR spectra and the related pH estimation procedure; Section 3 discusses material and methods for L-band spectroscopy experiments; Section 4 presents results; Section 5 includes discussion; and Section 6 summarizes the conclusions.
2. Spectral modeling and pH estimation
For a probe with known pKa value, pH is defined by
| (1) |
where α and β define the concentrations of RH+ and R, respectively, and p represents the pH value.
In the case of slow-exchange, any measured spectrum with arbitrary pH value can be accurately represented by a superposition of RH+ and R spectra [29]. Once the RH+ and R spectra have been modeled, the relative contributions (α/β) of RH+ and R spectra, for a given sample with unknown pH, can be estimated by fitting the measured spectrum with a weighted sum of the two modeled spectra. The pH value can then be inferred using Eq. 1. In this section, we describe the parametric modeling of RH+ and R spectra; we then present an estimator for the unknown pH value.
2.1. Spectral modeling
The first step is to parameterize the EPR spectrum of a pH-sensitive EPR probe. To ensure the adequacy of a small magnetic field sweep, the parametric modeling of the spectrum has been restricted to the highest-field resonance peak. Also, we use multiple Voigt functions to describe the spectra. The use of multiple functions provides a more accurate representation of complex spectra, e.g., ones with 13C satellite peaks. Using multiple Voigt functions, a parameterized spectrum for RH+, denoted by l+ (B), can be expressed as
| (2) |
where K represents the total number of Voigt functions used in the modeling;
represents the intensity of the kth Voigt function;
stands for a first-derivative Lorentzian function; B represents the external magnetic field; Γ+ denotes half-width at half-maximum linewidth of
;
represents a Gaussian function centered at
with standard deviation
; and * represents a convolution in the first variable.
In CW EPR, signal-to-noise ratio (SNR) can be improved by applying overmodulation [30]. To incorporate overmodulation in our modeling, we replace the first derivative Lorentzian function,
, with the first harmonic of an overmodulated Lorentzian function,
, defined by Robinson et al. [31], yielding
| (3) |
where Bm is the field modulation amplitude.
Likewise, the spectra of R can be modeled using a combination of over-modulated Voigt functions to generate l− (B), yielding
| (4) |
For a given probe, l+ (B) and l− (B) are constructed by fitting measured spectra of RH+ and R with Eqs. 3 and 4, respectively, and estimating the parameters accented with ‘+’ and ‘−’ superscripts on the right sides of Eqs. 3 and 4.
2.2. pH estimation
Once both l+ (B) and l− (B) have been separately constructed, a spectrum, l(B), with an arbitrary pH value, under the assumption of slow-exchange [29], can be written as
| (5) |
For a measured noisy spectrum, Y(B), the values of α and β can be estimated using least-squares fitting, yielding
| (6) |
where Y and l are vector representations of sampled spectra Y(B) and l(B), respectively. The unknown pH can be computed from α̂ and β̂ using Eq. 1.
3. Material and Methods
In this section, we describe the procedures used in preparing pH-sensitive samples. We also briefly mention the protocols used for EPR data collection.
3.1. Sample preparation
The sample was prepared by dissolving 1 mM ATI in 10 mM phosphate buffer solution. A total of ten ATI samples, with pH values ranging from 4.00 to 8.30, were prepared. The pH was adjusted by adding 0.1 M HCl and NaOH solutions to the sample. After the pH was adjusted and measured using a pH meter, a small volume (0.25 ml) of each sample was poured into a separate 0.5 ml Eppendorf PCR tube. Figure 3 shows the pH values of all ten samples.
Figure 3.
A total of ten ATI samples, each with a different pH value, were prepared. A small volume (0.25 ml) of each sample was placed in a separate 0.5 ml Eppendorf PCR tube. The samples with the lowest pH value, s+, and the highest pH value, s−, were used to construct parametric models l+ (B) and l− (B), respectively, while the other eight samples, s1, s2, ···, s8, with intermediate pH values were used for validation.
The s+ and s− samples, with pH values more than two units away from pKa, were assumed to be comprised of purely protonated and unprotonated forms and were used to build l+ (B) and l− (B), respectively. The other eight samples, s1, s2, ···, s8, were used for validation. After the preparation and before the EPR measurements, the samples were stored in a refrigerator at 4°C.
3.2. Data collection
The data were collected on an L-band CW EPR spectrometer under room-air conditions. The spectrometer was fitted with automatic tuning and coupling mechanisms and an 8 mm diameter surface-coil resonator to hold the samples. Two datasets were collected for each sample: one at low modulation amplitude (0.036 mT) and one at high modulation amplitude (0.18 mT). Only the highest-field peak was scanned. Each dataset had ten scans collected under identical conditions. The other acquisition parameters were: magnetic field sweep width=1.6 mT, center of magnetic field sweep=48.78 mT, acquisition time per scan=8 s, microwave power=0.5 mW, and number of data samples per scan=1024.
4. Results
In this section, we verify the proposed method using experimental data. Also, we draw a comparison between PMM and HCM in terms of measurement accuracy or, equivalently, data acquisition time.
The parametric models, l+ (B) and l− (B), were constructed from the measured spectra of s+ and s− samples, respectively. First, all ten scans from s+ and s− were independently averaged to improve SNR, and the averaged spectra were then modeled using Eqs. 3 and 4 to generate l+ (B) and l− (B), respectively. The process of constructing l+ (B) and l− (B) was repeated for both low-modulation and high-modulation datasets.
Three overmodulated Voigt functions (K = 3 in Eqs. 3 and 4) were used to describe the spectra for both the datasets. One Voigt function was used to describe the visibly dominant resonance peak, while the other two Voigt functions were used to describe the small 13C satellite peaks that reside on either side of the dominant peak. The amplitudes (ρ) of the 13C satellite peaks were only 3% to 5% of the dominant peak. For the low-modulation dataset, peak-to-peak linewidths (ΔBpp) of l+ (B) and l− (B) were observed to be 0.1024 mT and 0.0878 mT, respectively. For the high-modulation dataset, ΔBpp values of l+ (B) and l− (B) were observed to be 0.1585 mT and 0.1537 mT, respectively. Figure 4 shows the modeling results for both the datasets.
Figure 4.
Spectral modeling of the highest-field resonance peak of ATI sample at two extreme pH values. The modeling of s+ spectrum with p = 4.00 yielded l+ (B), while the modeling of s− spectrum with p = 8.30 yielded l− (B). The modeling was performed using Eqs. 3 and 4 for two different field modulation amplitudes: 0.036 mT (a) and 0.18 mT (b).
For each dataset, measurements from s1, s2, ···, s8 as well as from s+ and s− were fitted with a weighted sum of l+ (B) and l− (B) to estimate α and β and hence the pH values. Figures 5 and 6 show examples of fitting results for the s3, s5, and s7 samples. Figures 7 and 8 display the mean and standard deviation of pH values estimated using PMM and HCM. Table 1 compares the average standard deviation of estimated pH for the two methods for s1, s2, ···, s8 samples.
Figure 5.
Fitting results of the low-modulation (0.036 mT) dataset. Three scans–one for s3, one for s5, and one for s7–are depticted (a). The spectra were fitted with a weighted sum of previously constructed l+ (B) and l− (B) shown in Fig. 4a. The resulting components, l+ (B) and l− (B), are also shown (b).
Figure 6.
The fitting process shown in Fig. 5 is repeated for high-modulation (0.18 mT) dataset.
Figure 7.
Estimated pH values plotted against the calibration curves (solid gray line) for low-modulation dataset for both HMC (a) and PMM (b). Error bar, for each sample, indicates ±5 standard deviations of estimated pH values. Here, h represents the separation between measured zero-crossing of the highest-field resonance peak and a fixed reference point, which in this case was arbitrarily chosen at 48.74 mT.
Figure 8.
Estimation of pH values, similar to the one shown in Fig. 7, for the high-modulation dataset.
Table 1.
pH measurement accuracy of traditional HCM and proposed PMM. Here, indicates the standard deviation of the estimated pH values averaged across the eight samples, s1, s2, ···, s8.
| HCM (p̃) | PMM (p̂) | |||
|---|---|---|---|---|
| low-modulation dataset (Bm = 0.036 mT) | ||||
| high-modulation dataset (Bm = 0.180 mT) |
The calibration curves, THCM and TPMM, shown in Figs. 7 and 8 are defined as,
| (7) |
where the parameters cl and ck represent the pKa value of the probe and the other parameters (ca, cd, and cs) define the shape of the sigmoid function.
For THCM, all four parameters were estimated by fitting the THCM function to the observed h vs. p data, with h being the measured location of the highest-field resonance peak from an arbitrarily selected fixed reference point. Traditionally, the sigmoid function for THCM has been defined in terms of three parameters, with the fourth parameter, cs, fixed at one. The motivation behind including cs as an additional parameter is to generate a tight fit between THCM and h vs. p curve for a wide range of Bm. The shape of h vs. p curve changes significantly with Bm, with smaller Bm generating larger growth rates and thus requiring larger values of cs. For low-modulation dataset (Fig. 7), the values of ca, cd, ck, and cs were estimated to be 76.9 μT, 1.7 μT, 6.241, and 1.785, respectively. For high-modulation dataset (Fig. 8), the values of ca, cd, ck, and cs were estimated to be 78.3 μT, 0.85 μT, 6.296, and 1.276, respectively. After estimating all four parameters of the THCM function, the pH values were determined by
| (8) |
where p̃ is the pH estimate based on HCM.
For PMM, the value of cl was determined by fitting TPMM function to β̂/(α̂ + β̂) vs. p data, with α̂ and β̂ determined using Eq. 6. The estimated values of cl were found to be 6.289 and 6.291 for the low-modulation and high-modulation datasets, respectively. Once the value of cl was estimated, the pH values of the samples were calculated from the measured data by
| (9) |
where p̂ is the pH estimate based on PMM.
5. Discussion
EPR-based pH spectroscopy relies on reversible changes in the EPR spectrum as a function of pH. The conventional method, HCM, quantifies the unknown pH value by comparing the location of a zero-crossing of one of the resonance peaks against a previously determined calibration curve. Although, this approach has been extensively used for pH measurements, it is inefficient as only a small number of data points of the measured spectrum are actually utilized in calculating pH. The proposed modeling-based approach, PMM, on the other hand, models the entire resonance peak (or peaks) and hence utilizes all the measured spectrum for computing pH.
In PMM, the complexity of the model (Eqs. 3 and 4) in itself does not impact the accuracy of the method as long as the spectra are accurately modeled. For ATI, a combination of three overmodulated Voigt functions was adequate to accurately represent spectra of RH+ and R at two different modulation amplitudes. Using more than three Voigt functions to model l+ (B) and l− (B) spectra (results not shown) yielded no improvement in the pH measurement accuracy. Other NRs, with more complex spectra, may require more than three Voigt functions for accurate modeling. On the other hand, the spectral differences between RH+ and R can impact the accuracy of both HCM and PMM significantly. For example, a probe with large hyperfine coupling differences between the spectra of RH+ and R is more likely to produce accurate results as opposed to a probe with small hyperfine coupling differences.
For HCM, it is a common practice to use a sigmoid function to approximate observed variations in hyperfine coupling as a function of pH. However, depending on the spectral shape of the probe and the experimental parameters, the approximation may not be reasonable as there is no physical reason for the observed hyperfine coupling to follow a sigmoid function exactly. For example, if EPR spectra of R and RH+ exhibit different linewidths, which is generally the case, the h vs. p curve no longer possesses the odd-symmetry of a sigmoid function. Therefore, for such cases, estimating pH using THCM may lead to systematic errors in the estimation of pKa. In fact, for HCM, we observed a Bm dependent error in the estimation of pKa. For the low-modulation dataset, HCM underestimated pKa by approximately 0.06 units. When the lowest-field peak was considered (data not shown) instead of the highest-field peak, HCM underestimated pKa by approximately 0.1 units for the low-modulation dataset. The trend of underestimating pKa even held for the traditional calibration curve, which entails fixing the value of cs to one. This Bm dependence of estimated pKa can be attributed to asymmetries originating from the lineshape differences between R and RH+, which are more pronounced at smaller Bm values as evident from the mismatch between ΔBpp values of l+ (B) and l− (B). The previously reported pKa of ATI, when measured at X-band using HCM, is 6.1. It is likely that this reported pKa value is an underestimation of the true value for the reasons discussed here. For the high-modulation dataset, the pKa values determined by the two methods, HCM and PMM, were in close agreement.
The β/(α + β) vs. p curve by definition follows the sigmoid function of TPMM, enabling PMM to estimate pKa more accurately and without any systematic bias. The pKa values reported by PMM were consistent, within 6.90 ± 0.005, for both the modulation amplitudes and for both the highest-field and lowest-field peaks (data not shown).
The proposed method is based on the assumption of slow-exchange, which ensures that any spectrum can be modeled as a weighted sum of l+ (B) and l− (B). It has been previously demonstrated by Khramtsov et al. and others that the assumption of slow-exchange can be ensured by limiting concentrations of NR and accompanying buffer as well as by selecting 3 < pKa < 11 [32, 33, 19]. Generally, concentrations of 1 mM or less for NR and 10 mM or less for buffer are adequate to ensure slow-exchange [34, 35].
The structure in the residual, especially for the s5 sample with p ≈ pKa (Fig. 5), is probably due to a deviation from the slow-exchange assumption. This small deviation, however, does not exhibit significant impact on the pH estimation. The accuracy of results (Figs. 7 and 8), which explicitly rely on the slow-exchange assumption, provides validation of the assumption for the reported experimental conditions. Note the extracellular buffer concentrations encountered in vivo are even lower that what we have used in our experiments. Therefore, by limiting the concentration of ATI to 1 mM, we can ensure slow-exchange to be a reasonable assumption under in vivo conditions. For a given NR, ex vivo calibration studies, similar to one presented here, should be conducted to establish the conditions required for slow-exchange, including limits on the concentrations of NR and buffer.
In this work, we have assumed that s+ and s− consist of purely RH+ and R, respectively. This is a reasonable assumption as s+ and s− contained less than 1% contamination from R and RH+, respectively. If spectra corresponding to sufficiently high and low pH values are unavailable, any other pair of spectra with sufficiently different pH values can be used instead as long the precise pH values of the samples are accurately known. For example, spectra from s3 (p = 5.70) and s7 (p = 6.71) can be modeled using Eqs. 3 and 4 and later used to fit the spectrum from a sample with unknown pH. However, using samples with pH values close to pKa may require more than three Voigt functions for accurate modeling (see Fig. 5).
Traditionally, multiple resonance peaks are scanned, and the estimation of pH is based on the separation between two peaks. Scanning multiple peaks suppresses the impact of unintended spectral shifts due to excitation frequency drift that may arise due to animal motion. However, this approach can lead to longer acquisition times due to wider scans. Since we are proposing scanning a single resonance peak, the collected data must be corrected for a frequency drift. Although we did not observe any frequency drift for the spectroscopy experiments, a drift, if present, must be accounted for to generate accurate, reproducible results. We suggest using either of two procedures to tackle the issue of frequency drift during in vivo measurements: (i) Use wider scans to build l+ (B) and l− (B) and record the corresponding excitation frequency using a frequency counter. For a given modulation amplitude and pH-sensitive probe, this step needs to be performed only once under ex vivo conditions. During in vivo data acquisition, record the excitation frequency immediately before or after each scan. The sweep width of in vivo scans should be smaller than the sweep width used for l+ (B) and l− (B) scans. For in vivo scans, compute shifts in measured spectra, in the units of magnetic field, based on the frequency difference of each scan with respect to l+ (B) and l− (B) scans. Before fitting each scan using Eq. 5, digitally shift l+ (B) and l− (B) to match the observed shift in the scan and then truncate the shifted l+ (B) and l− (B) to match the narrower sweep width of the scan. (ii) Record the excitation frequency for scans used to build l+ (B) and l− (B). Design a hardware feedback loop that on-the-fly changes the center of sweep width in proportion to the frequency drift from the previously recorded value for l+ (B) and l− (B) scans. Measure and process the in vivo data as if there were no frequency drift. Also, a small uncertainty in the location of the resonance peak, if present despite the proposed measures, can be tolerated by including it into the model as we have previously demonstrated for oximetry application [36].
Figures 7 and 8 display the experimental results against the ideal calibration curves defined in Eq. 7. As reported in Table 1, PMM exhibits high estimation accuracy for both the low-modulation and high-modulation datasets. For the low-modulation dataset, PMM provides sixfold improvement while for the high-modulation dataset, PMM provides over threefold improvement compared to HCM. Since the standard deviations of p̂ or p̃ are proportional to SNR, which, in turn, is proportional to the square-root of scan time, the data acquisition speedup offered by PMM is a factor of 36 for the low-modulation dataset and a factor of 13 for the high-modulation dataset.
Under high SNR conditions, the measured spectra of RH+ and R can be effectively used in lieu of l+ (B) and l− (B), respectively. A parametric modeling of RH+ and R spectra, however, carries two major advantages. First, parametric modeling of EPR spectra would allow optimization of acquisition parameters, including field modulation amplitude and sweep width, via statistical sensitivity analysis [37]. Second, the parametric modeling would enable developing forward models, which relate the EPR projection data to spatial distribution of α and β, and thus potentially reducing the data acquisition time for pH imaging. A similar concept has already been applied to EPR imaging of oxygen, providing a factor of forty reduction in the acquisition time. [38].
6. Conclusions
We have presented a data modeling and processing approach to reduce the acquisition time of EPR-based pH spectroscopy by over an order of magnitude. The method is based on parametric modeling of the EPR spectrum and relies on slow-exchange, which is a reasonable assumption for the experimental setup adopted in this work and is expected to hold valid under in vivo conditions. The method is verified using measurements from a commonly used pH-sensitive probe for a physiologically relevant pH range.
A data modeling and processing method for EPR-based pH spectroscopy is presented.
The method is based on parametric modeling of EPR spectra under slow-exchange conditions.
The proposed method offers an order of magnitude reduction in acquisition time.
For validation, L-band spectroscopy studies are reported.
Acknowledgments
We thank Andrey Bobko for preparing the ATI samples. This work was supported by NIH grant EB008836.
References
- 1.Raghunand N, Howison C, Sherry A, Zhang S, Gillies R. Renal and systemic pH imaging by contrast-enhanced MRI. Magn Reson Med. 2003;49:249–257. doi: 10.1002/mrm.10347. [DOI] [PubMed] [Google Scholar]
- 2.Gautam M, Benson C, Sluka K. Increased response of muscle sensory neurons to decreases in pH after muscle inflammation. Neuroscience. 2010;170:893–900. doi: 10.1016/j.neuroscience.2010.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Vogt S, Troitzsch D, Moosdorf R. Epicardial measurement of alterations in extracellular pH and electrolytes during ischemia and reperfusion in cardiac surgery. Biomed Tech (Berl) 2009;54:315–321. doi: 10.1515/BMT.2009.038. [DOI] [PubMed] [Google Scholar]
- 4.Koukoulis G, Caldwell R, Inokawa H, Button B, Sevala M, Lyles J, Takashima S, Blackwell J, Randell S, Egan T. Trends in lung pH and PO2 after circulatory arrest: implications for non-heart-beating donors and cell culture models of lung ischemia-reperfusion injury. J Heart Lung Transplant. 2005;24:2218–2225. doi: 10.1016/j.healun.2005.06.009. [DOI] [PubMed] [Google Scholar]
- 5.Pinkney J, Vernon P, Carstensen E, Gillies S, Phillips D, Yudkin J. Intracellular pH, intrauterine growth and the insulin resistance syndrome. Clin Physiol. 2001;21:640–647. doi: 10.1046/j.1365-2281.2001.00340.x. [DOI] [PubMed] [Google Scholar]
- 6.Katz P. Review article: intragastric and oesophageal pH monitoring in patients with gastro-oesophageal reflux disease. Aliment Pharmacol Ther. 2006;23:3–11. doi: 10.1111/j.1365-2036.2006.02801.x. [DOI] [PubMed] [Google Scholar]
- 7.Chrusch C, Bautista E, Jacobs H, Light R, Bose D, Duke K, Mink S. Blood pH level modulates organ metabolism of lactate in septic shock in dogs. J Crit Care. 2002;17:188–202. doi: 10.1053/jcrc.2002.35816. [DOI] [PubMed] [Google Scholar]
- 8.Riediker M, Danuser B. Exhaled breath condensate pH is increased after moderate exercise. J Aerosol Med. 2007;20:13–18. doi: 10.1089/jam.2006.0567. [DOI] [PubMed] [Google Scholar]
- 9.Chan Y, Wong T, Byrne F, Kavallaris M, Bulmus V. Acid-labile core cross-linked micelles for pH-triggered release of antitumor drugs. Biomacromolecules. 2008;9:1826–1836. doi: 10.1021/bm800043n. [DOI] [PubMed] [Google Scholar]
- 10.Lieberman R, D’Aquino J, Ringe D, Petsko G. Effects of pH and iminosugar pharmacological chaperones on lysosomal glycosidase structure and stability. Biochemistry. 2009;48:4816–4827. doi: 10.1021/bi9002265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Lindner D, Raghavan D. Intra-tumoural extra-cellular pH: a useful parameter of response to chemotherapy in syngeneic tumour lines. Br J Cancer. 2009;100:1287–1291. doi: 10.1038/sj.bjc.6605022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Park H, Lee S, Chung H, Rhee Y, Lim B, Ha S, Griffin R, Lee H, Song C, Choi E. Influence of environmental pH on G2-phase arrest caused by ionizing radiation. Radiat Res. 2003;159:86–93. doi: 10.1667/0033-7587(2003)159[0086:ioepog]2.0.co;2. [DOI] [PubMed] [Google Scholar]
- 13.Ahmad R, Kuppusamy P. Theory, instrumentation, and applications of electron paramagnetic resonance oximetry. Chem Rev. 2010;110:3212–3236. doi: 10.1021/cr900396q. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Gallez B, Mader K, Swartz H. Noninvasive measurement of the pH inside the gut by using pH-sensitive nitroxides. An in vivo EPR study. Magn Reson Med. 1996;36:694–697. doi: 10.1002/mrm.1910360507. [DOI] [PubMed] [Google Scholar]
- 15.Gallez B, Bacic G, Goda F, Jiang J, O’Hara J, Dunn J, Swartz H. Use of nitroxides for assessing perfusion, oxygenation, and viability of tissues: in vivo EPR and MRI studies. Magn Reson Med. 1996;35:97–106. doi: 10.1002/mrm.1910350113. [DOI] [PubMed] [Google Scholar]
- 16.Khan N, Das D. Application of in vivo EPR for tissue pO2 and redox measurements. Methods Mol Biol. 2009;559:131–139. doi: 10.1007/978-1-60327-017-5_10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Caldwell S, Quin C, Edge R, Hartley R. A dual sensor spin trap for use with EPR spectroscopy. Org Lett. 2007;9:3499–3502. doi: 10.1021/ol071285o. [DOI] [PubMed] [Google Scholar]
- 18.Khramtsov V. Biological Imaging and Spectroscopy of pH. Current Organic Chem. 2005;9:909–923. [Google Scholar]
- 19.Khramtsov V, Volodarskii L. Use of Imidazoline Nitroxides in Studies of Chemical Reactions: ESR Measurements of the Concentration and Reactivity of Protons, Thiols, and Nitric Oxide. In: Berliner L, editor. Biological Magnetic Resonance. Springer; New York: 2002. pp. 109–180. [Google Scholar]
- 20.Khramtsov V. Functional EPR Spectroscopy and Imaging of Nitroxides. In: Pifat-Mrzljak G, editor. Supramolecular Structure and Function. Vol. 9. Springer; Dordrecht: 2007. pp. 181–208. [Google Scholar]
- 21.Deng Y, He G, Petryakov S, Kuppusamy P, Zweier J. Fast EPR imaging at 300 MHz using spinning magnetic field gradients. J Magn Reson. 2004;168:220–227. doi: 10.1016/j.jmr.2004.02.012. [DOI] [PubMed] [Google Scholar]
- 22.Epel B, Sundramoorthy S, Mailer C, Halpern H. A Versatile High Speed 250 MHz Pulse Imager for Biomedical Applications. Concepts Magn Reson B Magn Reson Eng. 2008;33B:163–176. doi: 10.1002/cmr.b.20119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Yoshitomi T, Suzuki R, Mamiya T, Matsui H, Hirayama A, Nagasaki Y. pH-sensitive radical-containing-nanoparticle (RNP) for the L-band-EPR imaging of low pH circumstances. Bioconjug Chem. 2009;20:1792–1798. doi: 10.1021/bc900214f. [DOI] [PubMed] [Google Scholar]
- 24.Bobko A, Eubank T, Voorhees J, Efimova O, Kirilyuk I, Petryakov S, Trofimiov D, Marsh C, Zweier J, Grigor’ev I, Samouilov A, Khramtsov V. In Vivo Monitoring of pH, Redox Status, and Glutathione Using L-band EPR for Assessment of Therapeutic Effectiveness in Solid Tumors. Magn Reson Med. 2012 doi: 10.1002/mrm.23196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Potapenko D, Foster M, Lurie D, Kirilyuk I, Hutchison J, Grigor’ev I, Bagryanskaya E, Khramtsov V. Real-time monitoring of drug-induced changes in the stomach acidity of living rats using improved pH-sensitive nitroxides and low-field EPR techniques. J Magn Reson. 2006;182:1–11. doi: 10.1016/j.jmr.2006.06.002. [DOI] [PubMed] [Google Scholar]
- 26.Bobko A, Kirilyuk I, Eubank T, Voorhees J, Marsh C, Grigor’ev I, Khramtsov V. In Vivo Extracellular pH Monitoring in Tumor Tissues of PyMT mice: Effect of GM-CSF Treament. Free Radic Biol Med. 2009;47:S166. [Google Scholar]
- 27.Mader K, Gallez B, Liu K, Swartz H. Non-invasive in vivo characterization of release processes in biodegradable polymers by low-frequency electron paramagnetic resonance spectroscopy. Biomaterials. 1996;17:457–461. doi: 10.1016/0142-9612(96)89664-5. [DOI] [PubMed] [Google Scholar]
- 28.Kroll C, Hermann W, Stosser R, Borchert H, Mader K. Influence of drug treatment on the microacidity in rat and human skin–an in vitro electron spin resonance imaging study. Pharm Res. 2001;18:525–530. doi: 10.1023/a:1011066613621. [DOI] [PubMed] [Google Scholar]
- 29.Weil J, Bolton J. Relaxation Times, Linewidths and Spin Kinetic Phenomena. Ch 10. John Wiley and Sons; Hoboken: 2007. pp. 301–356. [Google Scholar]
- 30.Deng Y, Pandian R, Ahmad R, Kuppusamy P, Zweier J. Application of magnetic field over-modulation for improved EPR linewidth measurements using probes with Lorentzian lineshape. J Magn Reson. 2006;181:254–261. doi: 10.1016/j.jmr.2006.05.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Robinson B, Mailer C, Reese A. Linewidth analysis of spin labels in liquids. I. Theory and data analysis. J Magn Reson. 1999;138:199–209. doi: 10.1006/jmre.1999.1737. [DOI] [PubMed] [Google Scholar]
- 32. Molin Y. Spin exchange: Principles and applications in chemistry and biology. Springer-Verlag; New York: 1980. [Google Scholar]
- 33.Khramtsov V, Weiner L, Eremenko S, Belchenko O, Schastnev P, Grigor’ev I, Reznikov V. Proton exchange in stable nitroxyl radicals of the imidazoline and imidazolidine. J Magn Reson. 1985;61:397–408. [Google Scholar]
- 34.Glazachev Y, Grigor’ev I, Reijerse E, Khramtsov V. EPR Studies of 15N- and 2H-Substituted pH-Sensitive Spin Probes of Imidazoline and Imidazolidine Types. Appl Magn Reson. 2001;20:489–505. [Google Scholar]
- 35.Hyodo F, Matsumoto S, Devasahayam N, Dharmaraj C, Subramanian S, Mitchell J, Krishna M. Pulsed EPR imaging of nitroxides in mice. J Magn Reson. 2009;197:181–185. doi: 10.1016/j.jmr.2008.12.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Ahmad R, Caia G, Potter L, Petryakov S, Kuppusamy P, Zweier J. In vivo multisite oximetry using EPR-NMR coimaging. J Magn Reson. 2010;207:69–77. doi: 10.1016/j.jmr.2010.08.011. [DOI] [PMC free article] [PubMed] [Google Scholar]