Abstract
Conventionally, the assessments of endothelial function and arterial stiffness require different sets of equipment, making the inclusion of both tests impractical for clinical and epidemiological studies. Pulse wave analysis (PWA) provides useful information regarding the mechanical properties of the arterial tree and can also be used to assess endothelial function. PWA is a simple, valid, reliable, and inexpensive technique, offering great clinical and epidemiological potential. The current paper will outline how to measure arterial stiffness and endothelial function using this technique and include discussion of validity and reliability.
1. Introduction
Cardiovascular disease (CVD), the leading cause of mortality in the Western world [1], has a very long asymptomatic phase of development, starting as early as the first decade of life [2]. It is imperative, therefore, that clinical scientists and epidemiologists have at their disposal simple, valid, and reliable techniques to assess and track the progression of CVD. Noninvasive assessment techniques fall under two broad categories: those that assess endothelial health and those that assess arterial stiffness. Assessment of endothelial function indicates the functional health of the vascular system, whereas arterial stiffness assesses structural characteristics. Together, these techniques may provide complimentary indices of CVD risk.
Conventionally, assessments of endothelial function and arterial stiffness require different sets of equipment, making the inclusion of both tests impractical for clinical and epidemiological studies. Pulse wave analysis (PWA) is a simple and noninvasive technique that has been widely used in epidemiological [3] and interventional studies [4]. PWA provides useful information regarding the mechanical properties of the arterial tree and the ventricular-vascular interaction [5] and can also be used to assess endothelial function [6]. PWA is a simple, valid, reliable, and inexpensive technique, offering great clinical and epidemiological potential. The current review will outline how to measure arterial stiffness and endothelial function using this technique and include discussion of validity and reliability.
2. Arterial Stiffness
Arterial stiffness is a general term that collectively describes distensiblility, compliance, and elastic modulus of the arterial vascular system. These properties are not homogenous along the arterial tree and muscular and elastic vessels differ. Arterial stiffness can be measured systemically, regionally, or locally. Local measurements provide important physiological information and are more quantitative and sensitive than systemic indices. However, these measurements give no indication of how the artery of interest interacts with central function (i.e., the heart) as part of an integrative system. Regional arterial stiffness is measured at arterial sites of major physiologic importance such as the aorta where the arterial buffering function is principally expressed, or a particular limb. Systemic arterial stiffness affects the global buffering properties of the arterial system, just as arterial blood pressure can be considered as a global value of hemodynamic load, systemic arterial stiffness reflects the overall opposition of large arteries to the pulsatile effects of ventricular ejection.
A number of methodologies have been applied to the in vivo assessment of arterial stiffness. These methodologies fall into three broad groups: (1) relating change in area of an artery to distending pressure, that is, local arterial stiffness, (2) measuring pulse wave velocity, that is, regional arterial stiffness, and (3) pulse wave analysis, that is, systemic arterial stiffness. Ultrasound and magnetic resonance imaging (MRI) are capable of measuring local arterial stiffness [21–23] as well as pulse wave velocity [24–30], but these methodologies require expensive equipment (especially in the case of MRI) and a high level of technical expertise and are often impractical within the clinical or epidemiological setting. PWV can also be assessed using dedicated equipment, including oscillometric [31–34], tonometric [32, 35–37], volume plethysmographic [38, 39], and photo plethysmographic [40–43] devices. These devices either measure the pulse wave at two peripheral sites or record the electrocardiogram and measure the pulse wave at a peripheral site, to estimate the regional arterial stiffness. Alternatively, a number of devices are also available to estimate systemic arterial stiffness using PWA.
2.1. Measurement
A number of commercial devices are available to automate PWA assessments, including Compilor (tonometric device) [44, 45], Sphygmocor (tonometric device) [44, 46–49], PulsePen (tonometric device) [50, 51], ARCSolver [47, 52] (oscillometric device), Arteriograph [33, 53–57] (oscillometric device), Omron (oscillometric device) [55, 58], PulseCore [59, 60] (oscillometric device), Viscorder [32, 34, 49] (oscillometric device), and PulseTrace [6, 42, 48, 61] (photoplethysmographic device). Applanation tonometry is considered the “gold standard” and is the most widely used technique [62]. A probe is conventionally placed on the skin overlying the radial artery, and pressure is applied to distort or applanate (flatten) the artery, creating a signal which approximates arterial pressure. The peak and trough of the radial pulse wave correspond, respectively, to systolic and diastolic blood pressure measured conventionally on the brachial artery, since blood pressure is practically identical in brachial and radial arteries [63]. Mean blood pressure is determined by integration of the radial wave. A generalized transfer factor is then used to generate the corresponding central arterial waveform [64–67].
Figure 1 shows typical features of the aortic pulse pressure waveform, from which can be derived augmentation pressure (AP), augmentation index (AIx), and arrival time of reflected waves at the central aorta (Tr). Tr represents the time from the onset of the ejected pulse waveform to the onset of the reflected wave and reflects aortic pulse wave velocity [68]. AP is the additional aortic systolic pressure generated by the return of the reflected waves at the central aorta, expressed in absolute terms [62]. AIx is the AP as a percentage of central pulse pressure and is a composite measure of aortic wave reflection and systemic arterial stiffness [62, 69]. Although the timing of the arrival of the reflected wave at the proximal aorta is largely determined by large artery PWV, AIx is not interchangeable with PWV. It is influenced by vasoactive drugs independently of PWV [70], suggesting that it is also determined by the intensity of wave reflection, which, in turn, is determined by the diameter and elasticity of small arteries and arterioles. A number of variables are known to influence AIx. AIx increases with MAP [71] and is inversely related to body height [72] and heart rate [73, 74], with a 10 bpm increase in heart rate resulting in a 4% reduction in AIx [73]. AIx should be normalized for a heart rate of 75 beats per minute (AIx@HR75). One of the most widely used devices, SphygmoCor, automatically adjusts the AIx at an inverse rate of 4.8% for each 10 bpm increment. The AIx@HR75 is only calculated when the patient's heart rate is between 40 and 110 bpm. Outside of this range the software will display an N/C indicating that no calculation was possible.
Figure 1.

Aortic pulse pressure waveform. Systolic and diastolic pressures are the peak and trough of the waveform. Augmentation pressure is the additional pressure added to the forward wave by the reflected wave. Augmentation index is defined as the augmentation pressure as a percentage of pulse pressure. The dicrotic notch represents closure of the aortic valve and is used to calculate ejection duration. Time to reflection is calculated as the time at the onset of the ejected pulse waveform to the onset of the reflected wave.
Using the SphygmoCor to illustrate the procedure (a typical setup is shown in Figure 2), PWA takes approximately 20 minutes to complete. Following at least 10 minutes supine rest, brachial artery systolic and diastolic blood pressures are measured in the nondominant arm and used to calibrate the PWA measurements taken on the radial artery. Radial artery waveforms are then recorded in duplicate. A high-fidelity tonometer is used to obtain pressure waveforms by applying gentle pressure over the nondominant radial artery and repositioning the device until the greatest pulse signal is detected. Data is collected directly into a personal computer, and recordings are assessed visually to ensure that the best possible recording is obtained. After 20 sequential waveforms are acquired, an averaged peripheral waveform is generated and a corresponding aortic waveform is derived (see Figure 3). When consecutive AIx@HR75 readings differ by more than 4%, a third reading is obtained, and the mean of the closest two readings is taken.
Figure 2.

Electronics module (SphygmoCor device, AtCor Medical, Sydney, Australia).
Figure 3.

Radial artery applanation tonometry recording. The upper long panel shows the radial pressure waveform above the derived central pressure waveform. The upper right panel shows the overlaid radial waveforms, including the operator index, and the middle panel shows the quality control indices. The bottom left panel demonstrates a magnified radial arterial waveform. Systolic and diastolic pressures are 145/92 mmHg. The bottom right panel provides a magnified derived central pressure waveform. Central pressure is 137/93 mmHg.
2.2. Validity
There is evidence that increased aortic wave reflections have adverse effects on ventricular afterload and coronary perfusion, and their pathological role has been demonstrated in several diseases [75–78]. Furthermore, increased central arterial wave reflections have been shown to independently predict cardiovascular risk and mortality [5, 79, 80]. Increased amplitude and the earlier return of the reflected wave within the cardiac cycle augments the central systolic blood pressure, resulting in increased wave reflections [69]. The amplitude and timing of reflected pressure waves are determined primarily by vascular elasticity, peripheral vascular resistance, heart rate, and left ventricle function [81].
2.3. Reliability
Any valid technique utilised for the measurement of physiological variables must be reproducible [82]. A high intra- and interobserver reproducibility of baseline AIx has been observed in healthy controls and patients with cardiovascular disease and renal dysfunction [7, 38, 82–94]. Good reproducibility of baseline time to reflection (Tr), an alternative index to AIx, has been reported in several studies [88, 91, 94]. However, the majority of these trials examined the reproducibility using simple Bland-Altman analysis, whereas studies using more definitive intra-class correlation coefficient (ICC) and the contribution of variance components are limited [81, 82, 91, 93, 94]. ICC values for repeated measurements taken at hourly or weekly intervals have been reported to be 0.72–0.90 for AIx [91, 94], 0.90 for AP [81], and 0.43–0.84 for Tr [81, 91, 94]. In addition, few trials have reported the reproducibility of heart rate corrected AIx [82, 91, 93], although it is frequently used in wave reflection studies. While this technique potentially offers a valid and reliable marker of CVD risk, further study is required to determine sample size recommendations for AIx normalized to heart rate.
2.4. Recommendations
The PWA technique is particularly suitable for incorporation into clinical trials. Strengths of this technique include simplicity of assessment, relatively short training requirements for investigators, low time commitment for subjects, noninvasiveness, portability, and cost-effectiveness. Both AIx and AP confer similar reproducibility and have been reported as more reliable than Tr. Since AIx is strongly influenced by heart rate, both AIx and AP should be corrected for this confounding factor [73]. Further study is warranted to determine whether the three indices of arterial stiffness provide additive prognostic value.
3. Endothelial Health
Functionally, the endothelium is a large autocrine, paracrine, and endocrine organ that plays a key role in vascular homeostasis [95]. Endothelial dysfunction is a pivotal, yet potentially reversible, step that has been shown to precede and predict overt CVD [96]. The endothelium has been recognized for the important role it plays in regulating vascular reactivity via the release of dilator mediators, including nitric oxide (NO) [97–100], prostaglandins [101], and endothelial-derived hyperpolarizing factor [102, 103]. The capacity of the endothelium to regulate vascular tone (reactivity) is used to confirm the health of the endothelium.
Established methodologies for evaluating peripheral endothelial function include strain-gauge venous occlusion forearm plethysmography [104, 105], ultrasound-measured flow-mediated dilation (FMD) [105–107], peripheral arterial tonometry (e.g., using the EndoPat device) [105], and laser Doppler flowmetry [105]. These techniques assess the vasodilator responses to endothelium-dependent stimuli such as acetylcholine and increased shear stress and to endothelium-independent stimuli, including sodium nitroprusside and glyceryl trinitrate (GTN). The two most commonly used techniques are strain-gauge plethysmography and FMD, with FMD being considered the “gold standard” for assessing endothelial function. FMD is a noninvasive, valid [10–13], and moderately reliable [14, 15] technique but is expensive and highly technical (Table 1). Strain-gauge plethysmography also offers acceptable reliability [17–20] but is an invasive technique when coupled with intra-arterial infusion of vasodilators (Table 1). As such, these techniques are often impractical for use in clinical trials or epidemiological studies. Alternatively, a limited number of studies have used PWA to assess endothelial function.
Table 1.
Comparison of noninvasive techniques for assessing endothelial function.
| Technique | Equipment | Cost | Skill level | Test time | Validity | Reliability | Ref |
|---|---|---|---|---|---|---|---|
| PWA | Applanation tonometry | Medium US$15,000 |
Low | 35 mins | Medium* | Medium-high* d = 0.9–2.3% |
[6–9] |
| FMD | (i) Ultrasound (ii) Tourniquet |
High >US$50,000 |
High | 20 mins | High | Medium CV: 14–50% |
[10–16] |
| Plethysmography | Strain-gauge | Low US$10,000 |
Med. | 30 mins | Medium | Medium CV: 8–27% |
[17–20] |
FMD: flow-mediated dilation; CV: coefficient of variation; d: sample bias (mean difference); PWA: pulse wave analysis. *Further study is needed to corroborate these findings.
3.1. Measurement
Endothelial function can be assessed noninvasively by evaluating the effects of inhaled salbutamol on the AIx [6–8]. AIx is a measure of systemic arterial stiffness [62]. Notably, NO, considered the central molecule governing endothelial function [97–100], is a key modulator of arterial stiffness [108]. Chowienczyk et al. [6] demonstrated that salbutamol, a β2 agonist and endothelium-dependent vasodilator, in part reduces wave reflection by activation of the L-arginine-NO pathway.
The test takes approximately 35 minutes to complete; following at least 10 minutes supine rest, baseline PWA is recorded as described above, and PWA recordings are then repeated after 5, 10, 15, and 20 minutes after the administration of 400 μg inhaled salbutamol. The maximal decrease in AIx from baseline following a salbutamol challenge is used as an index of endothelial function. Endothelium-independent function can also be assessed using PWA by measurement of the reduction in AIx following sublingual GTN administration.
3.2. Validity
Wilkinson et al. [7] and Hayward et al. [8] evaluated the effects of inhaled salbutamol and sublingual GTN on the AIx. Investigators observed a significant correlation between the salbutamol-mediated reductions in AIx and increase in forearm blood flow during infusion of acetylcholine that was abolished by coadministration of NG-monomethyl-L-arginine (L-NMMA), an endothelial NO synthase inhibitor, suggesting that this may represent a valid approach for assessing endothelial function. Conversely, coadministration of intravenous L-NMMA has no influence on sublingual GTN-mediated reductions in AIx, consistent with an endothelium-independent effect [37, 74]. To date, relatively few clinical studies have employed PWA to assess endothelial function. However, preliminary validation studies have demonstrated blunted AIx responses to salbutamol in subjects exhibiting diabetes [6], hypercholesterolaemia [7], coronary artery disease [8], and peripheral vascular disease [9] among others.
3.3. Reliability
Studies examining the reproducibility of salbutamol-mediated effects on AIx are limited [7, 8, 82]. Hayward et al. [8] assessed reliability in healthy subjects using Bland-Altman analysis and reported excellent between-day mean difference (d = 0.9 ± 2%). Wilkinson et al. [7] also assessed reliability in healthy subjects using Bland-Altman analysis and reported similar findings; between-day mean difference in the AIx response was −2.3 ± 3.0 for salbutamol-mediated changes and 0.2 ± 6.5 for glyceryl trinitrate (GTN)-mediated changes. More recently, Paul et al. [82] produced discrepant findings when they assessed within-day reliability using the more definitive ICC method [109]. Relatively low reliability was reported for control (ICC: 0.18) and chronic heart failure groups (CHF) (ICC: 0.04). However, relatively low reliability was also reported for GTN-mediated changes in AIx for both control (ICC: 0.58) and CHF groups (ICC: 0.17). Endothelial function assessments were repeated at hourly intervals, and the authors suggest that the low ICC values may reflect diurnal variability or a carryover effect from the first assessment. Further study is warranted to determine the reliability and sample size requirements for this promising technique.
3.4. Recommendations
Salbutamol-mediated effects on AIx can be used to estimate endothelial function. This technique offers a number of advantages, namely simplicity of assessment, relatively short training requirements for investigators, low time requirement, portability, cost-effectiveness, ease of use, good reliability indicated by preliminary studies, and the salbutamol challenge isolating the NO pathway. However, more clinical studies are required to further validate this test. Also, while NO is unarguably an important molecule governing endothelial function, the endothelium can release additional complimentary or compensatory molecules [110], most notably prostacyclin [101] and endothelial-derived hyperpolarizing factor [102, 103]. The relative importance of these molecules varies by vascular bed and between individuals, particularly between individuals exhibiting a number of diseased states [111–120]. Therefore, while this test can provide a snapshot of endothelial function it may not provide a complete picture. Further study is warranted to determine whether baseline AIx (systemic arterial stiffness) and AIx following salbutamol inhalation confer additive prognostic value. Further study is also required to clarify sample size recommendations for this test.
4. Conclusion
PWA is a simple technique capable of assessing systemic arterial stiffness and endothelial function. This test is particularly suitable for clinical and epidemiological studies. Further study is required to determine: (1) sample size requirements, and (2) whether baseline AIx (systemic arterial stiffness) and AIx following salbutamol inhalation confer additive prognostic value.
References
- 1.American Heart Association. eart Disease and Stroke Statistics—2006 Update. American Heart Association; 2006. [Google Scholar]
- 2.Berenson GS, Srinivasan SR, Bao W, Newman WP, III, Tracyand RE, Wattigney WA. Association between multiple cardiovascular risk factors and atherosclerosis in children and young adults. The Bogalusa Heart Study. The New England Journal of Medicine. 1998;338(3):1650–1656. doi: 10.1056/NEJM199806043382302. [DOI] [PubMed] [Google Scholar]
- 3.Vlachopoulos C, Aznaouridis K, O’Rourke MF, Safar ME, Baou K, Stefanadis C. Prediction of cardiovascular events and all-cause mortality with central haemodynamics: a systematic review and meta-analysis. European Heart Journal. 2010;31(15):1865–1871. doi: 10.1093/eurheartj/ehq024. [DOI] [PubMed] [Google Scholar]
- 4.Williams B, Lacy PS, Thom SM, et al. Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: principal results of the Conduit Artery Function Evaluation (CAFE) study. Circulation. 2006;113(9):1213–1225. doi: 10.1161/CIRCULATIONAHA.105.595496. [DOI] [PubMed] [Google Scholar]
- 5.Vlachopoulosand C, O'Rourke M. Genesis of the normal and abnormal arterial pulse. Current Problems in Cardiology. 2000;25(5):303–367. doi: 10.1067/mcd.2000.104057. [DOI] [PubMed] [Google Scholar]
- 6.Chowienczyk PJ, Kelly RP, MacCallum H, et al. Photoplethysmographic assessment of pulse wave reflection: blunted response to endothelium-dependent beta2-adrenergic vasodilation in type II diabetes mellitus. Journal of the American College of Cardiology. 1999;34(7):2007–2014. doi: 10.1016/s0735-1097(99)00441-6. [DOI] [PubMed] [Google Scholar]
- 7.Wilkinson IB, Hall IR, MacCallum H, et al. Pulse-wave analysis: clinical evaluation of a noninvasive, widely applicable method for assessing endothelial function. Arteriosclerosis, Thrombosis, and Vascular Biology. 2002;22(1):147–152. doi: 10.1161/hq0102.101770. [DOI] [PubMed] [Google Scholar]
- 8.Hayward CS, Kraidly M, Webband CM, Collins P. Assessment of endothelial function using peripheral waveform analysis: a clinical application. Journal of the American College of Cardiology. 2002;40(3):521–528. doi: 10.1016/s0735-1097(02)01991-5. [DOI] [PubMed] [Google Scholar]
- 9.Kals J, Kampus P, Kals M, Pulges A, Teesalu R, Zilmer M. Effects of stimulation of nitric oxide synthesis on large artery stiffness in patients with peripheral arterial disease. Atherosclerosis. 2006;185(2):368–374. doi: 10.1016/j.atherosclerosis.2005.06.014. [DOI] [PubMed] [Google Scholar]
- 10.Schroeder S, Enderle MD, Ossen R, et al. Noninvasive determination of endothelium-mediated vasodilation as a screening test for coronary artery disease: pilot study to assess the predictive value in comparison with angina pectoris, exercise electrocardiography, and myocardial perfusion imaging. American Heart Journal. 1999;138(4):731–739. doi: 10.1016/s0002-8703(99)70189-4. [DOI] [PubMed] [Google Scholar]
- 11.Heitzer T, Schlinzig T, Krohn K, Meinertz T, Münzel T. Endothelial dysfunction, oxidative stress, and risk of cardiovascular events in patients with coronary artery disease. Circulation. 2001;104(22):2673–2678. doi: 10.1161/hc4601.099485. [DOI] [PubMed] [Google Scholar]
- 12.Neunteufl T, Heher S, Katzenschlager R, et al. Late prognostic value of flow-mediated dilation in the brachial artery of patients with chest pain. American Journal of Cardiology. 2000;86(2):207–210. doi: 10.1016/s0002-9149(00)00857-2. [DOI] [PubMed] [Google Scholar]
- 13.Inaba Y, Chen JA, Bergmann SR. Prediction of future cardiovascular outcomes by flow-mediated vasodilatation of brachial artery: a meta-analysis. International Journal of Cardiovascular Imaging. 2010;26(6):631–640. doi: 10.1007/s10554-010-9616-1. [DOI] [PubMed] [Google Scholar]
- 14.Hijmering ML, Stroes ESG, Pasterkamp G, Sierevogel M, Banga JD, Rabelink TJ. Variability of flow mediated dilation: consequences for clinical application. Atherosclerosis. 2001;157(2):369–373. doi: 10.1016/s0021-9150(00)00748-6. [DOI] [PubMed] [Google Scholar]
- 15.Woodman RJ, Playford DA, Watts GF, et al. Improved analysis of brachial artery ultrasound using a novel edge-detection software system. Journal of Applied Physiology. 2001;91(2):929–937. doi: 10.1152/jappl.2001.91.2.929. [DOI] [PubMed] [Google Scholar]
- 16.De Roos NM, Bots ML, Schouten EG, Katan MB. Within-subject variability of flow-mediated vasodilation of the brachial artery in healthy men and women: implications for experimental studies. Ultrasound in Medicine and Biology. 2003;29(3):401–406. doi: 10.1016/s0301-5629(02)00709-3. [DOI] [PubMed] [Google Scholar]
- 17.Walker HA, Jackson G, Ritter JM, Chowienczyk PJ. Assessment of forearm vasodilator responses to acetylcholine and albuterol by strain gauge plethysmography: reproducibility and influence of strain gauge placement. British Journal of Clinical Pharmacology. 2001;51(3):225–229. doi: 10.1046/j.1365-2125.2001.00330.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Thijssen DHJ, Bleeker MWP, Smits P, Hopman MTE. Reproducibility of blood flow and post-occlusive reactive hyperaemia as measured by venous occlusion plethysmography. Clinical Science. 2005;108(2):151–157. doi: 10.1042/CS20040177. [DOI] [PubMed] [Google Scholar]
- 19.Alomari MA, Solomito A, Reyes R, Khalil SM, Wood RH, Welsch MA. Measurements of vascular function using strain-gauge plethysmography: technical considerations, standardization, and physiological findings. American Journal of Physiology. 2004;286(1):H99–H107. doi: 10.1152/ajpheart.00529.2003. [DOI] [PubMed] [Google Scholar]
- 20.Wilkinsonand IB, Webb DJ. Venous occlusion plethysmography in cardiovascular research: methodology and clinical applications. British Journal of Clinical Pharmacology. 2001;52(6):631–646. doi: 10.1046/j.1365-2125.2001.01495.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Kullo IJ, Malik AR. Arterial ultrasonography and tonometry as adjuncts to cardiovascular risk stratification. Journal of the American College of Cardiology. 2007;49(13):1413–1426. doi: 10.1016/j.jacc.2006.11.039. [DOI] [PubMed] [Google Scholar]
- 22.Nelson AJ, Worthley SG, Cameron JD, et al. Cardiovascular magnetic resonance-derived aortic distensibility: validation and observed regional differences in the elderly. Journal of Hypertension. 2009;27(3):535–542. doi: 10.1097/hjh.0b013e32831e4599. [DOI] [PubMed] [Google Scholar]
- 23.Ohayon J, Gharib AM, Garcia A, et al. Is arterial wall-strain stiffening an additional process responsible for atherosclerosis in coronary bifurcations?: an in vivo study based on dynamic CT and MRI. American Journal of Physiology. 2011;301(3):H1097–H1106. doi: 10.1152/ajpheart.01120.2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Calabia J, Torguet P, Garcia M, et al. Doppler ultrasound in the measurement of pulse wave velocity: agreement with the Complior method. Cardiovascular Ultrasound. 2011;9(1, article 13) doi: 10.1186/1476-7120-9-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Grotenhuis HB, Westenberg JJM, Steendijk P, et al. Validation and reproducibility of aortic pulse wave velocity as assessed with velocity-encoded MRI. Journal of Magnetic Resonance Imaging. 2009;30(3):521–526. doi: 10.1002/jmri.21886. [DOI] [PubMed] [Google Scholar]
- 26.Jiang B, Liu B, McNeill KL, Chowienczyk PJ. Measurement of pulse wave velocity using pulse wave Doppler ultrasound: comparison with arterial tonometry. Ultrasound in Medicine and Biology. 2008;34(3):509–512. doi: 10.1016/j.ultrasmedbio.2007.09.008. [DOI] [PubMed] [Google Scholar]
- 27.Benetos A, Joly L, Perret-Guillaume C, et al. Pulse wave velocity assessment by external noninvasive devices and phase-contrast magnetic resonance imaging in the obese. Hypertension. 2009;54(2):421–426. doi: 10.1161/HYPERTENSIONAHA.109.133645. [DOI] [PubMed] [Google Scholar]
- 28.Koivistoinen T, Virtanen M, Hutri-Kahonen N, et al. Arterial pulse wave velocity in relation to carotid intima-media thickness, brachial flow-mediated dilation and carotid artery distensibility: the Cardiovascular Risk in Young Finns Study and the Health 2000 Survey. Atherosclerosis. 2012;220(2):387–393. doi: 10.1016/j.atherosclerosis.2011.08.007. [DOI] [PubMed] [Google Scholar]
- 29.Laffon E, Marthan R, Montaudon M, Latrabe V, Laurent F, Ducassou D. Feasibility of aortic pulse pressure and pressure wave velocity MRI measurement in young adults. Journal of Magnetic Resonance Imaging. 2005;21(1):53–58. doi: 10.1002/jmri.20227. [DOI] [PubMed] [Google Scholar]
- 30.Tillin T, Chambers J, Malik I, et al. Measurement of pulse wave velocity: site matters. Journal of Hypertension. 2007;25(2):383–389. doi: 10.1097/HJH.0b013e3280115bea. [DOI] [PubMed] [Google Scholar]
- 31.Naidu MUR, Reddy BM, Yashmaina S, Patnaik AN, Rani PU. Validity and reproducibility of arterial pulse wave velocity measurement using new device with oscillometric technique: a pilot study. BioMedical Engineering Online. 2005;4, article 49 doi: 10.1186/1475-925X-4-49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.van Leeuwen-Segarceanu EM, Tromp WF, Bos WJW, Vogels OJM, Groothoff JW, Van Der Lee JH. Comparison of two instruments measuring carotid-femoral pulse wave velocity: vicorder versus SphygmoCor. Journal of Hypertension. 2010;28(8):1687–1691. doi: 10.1097/HJH.0b013e32833a8b83. [DOI] [PubMed] [Google Scholar]
- 33.Horvath IG, Nemeth A, Lenkey Z, et al. Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity. Journal of Hypertension. 2010;28(10):2068–2075. doi: 10.1097/HJH.0b013e32833c8a1a. [DOI] [PubMed] [Google Scholar]
- 34.Hickson SS, Butlin M, Broad J, Avolio AP, Wilkinson IB, McEniery CM. Validity and repeatability of the Vicorder apparatus: a comparison with the SphygmoCor device. Hypertension Research. 2009;32(12):1079–1085. doi: 10.1038/hr.2009.154. [DOI] [PubMed] [Google Scholar]
- 35.Salvi P, Lio G, Labat C, Ricci E, Pannier B, Benetos A. Validation of a new non-invasive portable tonometer for determining arterial pressure wave and pulse wave velocity: the PulsePen device. Journal of Hypertension. 2004;22(12):2285–2293. doi: 10.1097/00004872-200412000-00010. [DOI] [PubMed] [Google Scholar]
- 36.Sugawara J, Hayashi K, Yokoi T, Tanaka H. Carotid-femoral pulse wave velocity: impact of different arterial path length measurements. Artery Research. 2010;4(1):27–31. doi: 10.1016/j.artres.2009.11.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Weber T, Ammer M, Rammer M, et al. Noninvasive determination of carotid-femoral pulse wave velocity depends critically on assessment of travel distance: a comparison with invasive measurement. Journal of Hypertension. 2009;27(8):1624–1630. doi: 10.1097/HJH.0b013e32832cb04e. [DOI] [PubMed] [Google Scholar]
- 38.Matsui Y, Kario K, Ishikawa J, Eguchi K, Hoshide S, Shimada K. Reproducibility of arterial stiffness indices (pulse wave velocity and augmentation index) simultaneously assessed by automated pulse wave analysis and their associated risk factors in essential hypertensive patients. Hypertension Research. 2004;27(11):851–857. doi: 10.1291/hypres.27.851. [DOI] [PubMed] [Google Scholar]
- 39.Munakata M, Ito N, Nunokawa T, Yoshinaga K. Utility of automated brachial ankle pulse wave velocity measurements in hypertensive patients. American Journal of Hypertension. 2003;16(8):653–657. doi: 10.1016/s0895-7061(03)00918-x. [DOI] [PubMed] [Google Scholar]
- 40.Liu AB, Hsu PC, Chenand ZL, Wu HT. Measuring pulse wave velocity using ECG and photoplethysmography. Journal of Medical Systems. 2011;35(5):771–777. doi: 10.1007/s10916-010-9469-0. [DOI] [PubMed] [Google Scholar]
- 41.Wang MC, Wu AB, Cheng MF, Chen JY, Ho CS, Tsai WC. Association of arterial stiffness indexes, determined from digital volume pulse measurement and cardiovascular risk factors in chronic kidney disease. American Journal of Hypertension. 2011;24(5):544–549. doi: 10.1038/ajh.2010.266. [DOI] [PubMed] [Google Scholar]
- 42.Padilla JM, Berjano EJ, Saiz J, Rodriguezand R, Facila L. Pulse wave velocity and digital volume pulse as indirect estimators of blood pressure: pilot study on healthy volunteers. Cardiovascular Engineering. 2009;9(3):104–112. doi: 10.1007/s10558-009-9080-5. [DOI] [PubMed] [Google Scholar]
- 43.Alty SR, Angarita-Jaimes N, Millasseau SC, Chowienczyk PJ. Predicting arterial stiffness from the digital volume pulse waveform. IEEE Transactions on Biomedical Engineering. 2007;54(12):2268–2275. doi: 10.1109/tbme.2007.897805. [DOI] [PubMed] [Google Scholar]
- 44.Jatoi NA, Mahmud A, Bennett K, Feely J. Assessment of arterial stiffness in hypertension: comparison of oscillometric (Arteriograph), piezoelectronic (Complior) and tonometric (SphygmoCor) techniques. Journal of Hypertension. 2009;27(11):2186–2191. doi: 10.1097/HJH.0b013e32833057e8. [DOI] [PubMed] [Google Scholar]
- 45.Baulmann J, Schillings U, Rickert S, et al. A new oscillometric method for assessment of arterial stiffness: comparison with tonometric and piezo-electronic methods. Journal of Hypertension. 2008;26(3):523–528. doi: 10.1097/HJH.0b013e3282f314f7. [DOI] [PubMed] [Google Scholar]
- 46.Hope SA, Meredith IT, Cameron JD. Arterial transfer functions and the reconstruction of central aortic waveforms: myths, controversies and misconceptions. Journal of Hypertension. 2008;26(1):4–7. doi: 10.1097/HJH.0b013e3282f0c9f5. [DOI] [PubMed] [Google Scholar]
- 47.Wassertheurer S, Kropf J, Weber T, et al. A new oscillometric method for pulse wave analysis: comparison with a common tonometric method. Journal of Human Hypertension. 2010;24(8):498–504. doi: 10.1038/jhh.2010.27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Millasseau SC, Kelly RP, Ritter JM, Chowienczyk PJ. Determination of age-related increases in large artery stiffness by digital pulse contour analysis. Clinical Science. 2002;103(4):371–377. doi: 10.1042/cs1030371. [DOI] [PubMed] [Google Scholar]
- 49.Kracht D, Shroff R, Baig S. Validating a new oscillometric device for aortic pulse wave velocity measurements in children and adolescents. American Journal of Hypertension. 2011;24(12):1294–1299. doi: 10.1038/ajh.2011.147. [DOI] [PubMed] [Google Scholar]
- 50.Palombo C, Kozakova M, Morizzo C, et al. Circulating endothelial progenitor cells and large artery structure and function in young subjects with uncomplicated type 1 diabetes. Cardiovascular Diabetology. 2011;10, article 88 doi: 10.1186/1475-2840-10-88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Salvi P, Safar ME, Labat C, Borghi C, Lacolley P, Benetos A. Heart disease and changes in pulse wave velocity and pulse pressure amplification in the elderly over 80 years: the PARTAGE Study. Journal of Hypertension. 2010;28(10):2127–2133. doi: 10.1097/HJH.0b013e32833c48de. [DOI] [PubMed] [Google Scholar]
- 52.Weber T, Wassertheurer S, Rammer M, et al. Validation of a brachial cuff-based method for estimating central systolic blood pressure. Hypertension. 2011;58(5):825–832. doi: 10.1161/HYPERTENSIONAHA.111.176313. [DOI] [PubMed] [Google Scholar]
- 53.Gaszner B, Lenkey Z, Illyes M, et al. Comparison of aortic and carotid arterial stiffness parameters in patients with verified coronary artery disease. Clinical Cardiology. 2012;35(1):26–31. doi: 10.1002/clc.20999. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Gavaller H, Sepp R, Csanady M, Forsterand T, Nemes A. Hypertrophic cardiomyopathy is associated with abnormal echocardiographic aortic elastic properties and arteriograph-derived pulse-wave velocity. Echocardiography. 2011;28(8):848–852. doi: 10.1111/j.1540-8175.2011.01469.x. [DOI] [PubMed] [Google Scholar]
- 55.Rezai MR, Goudot G, Winters C, Finn JD, Wuand FC, Cruickshank JK. Calibration mode influences central blood pressure differences between SphygmoCor and two newer devices, the Arteriograph and Omron HEM-9000. Hypertension Research. 2011;34(9):1046–1051. doi: 10.1038/hr.2011.75. [DOI] [PubMed] [Google Scholar]
- 56.Nemes A, Takacs R, Gavaller H, et al. Correlations between aortic stiffness and parasympathetic autonomic function in healthy volunteers. Canadian Journal of Physiology and Pharmacology. 2010;88(12):1166–1171. doi: 10.1139/Y10-095. [DOI] [PubMed] [Google Scholar]
- 57.Nemes A, Takacs R, Gavaller H. Correlations between Arteriograph-derived pulse wave velocity and aortic elastic properties by echocardiography. Clinical Physiology and Functional Imaging. 2011;31(1):61–65. doi: 10.1111/j.1475-097X.2010.00980.x. [DOI] [PubMed] [Google Scholar]
- 58.Seibert F, Behrendt C, Schmidt S, van der Giet M, Zidek W, Westhoff TH. Differential effects of cyclosporine and tacrolimus on arterial function. Transplant International. 2011;24(7):708–715. doi: 10.1111/j.1432-2277.2011.01265.x. [DOI] [PubMed] [Google Scholar]
- 59.Climie RE, Schultz MG, Nikolic SB, Ahuja KD, Felland JW, Sharman JE. Validity and reliability of central blood pressure estimated by upper arm oscillometric cuff pressure. American Journal of Hypertension. 2012;25(4):414–420. doi: 10.1038/ajh.2011.238. [DOI] [PubMed] [Google Scholar]
- 60.Lowe A, Harrison W, El-Aklouk E, Ruygrok P, Al-Jumaily AM. Non-invasive model-based estimation of aortic pulse pressure using suprasystolic brachial pressure waveforms. Journal of Biomechanics. 2009;42(13):2111–2115. doi: 10.1016/j.jbiomech.2009.05.029. [DOI] [PubMed] [Google Scholar]
- 61.Millasseau SC, Guigui FG, Kelly RP, et al. Noninvasive assessment of the digital volume pulse: comparison with the peripheral pressure pulse. Hypertension. 2000;36(6):952–956. doi: 10.1161/01.hyp.36.6.952. [DOI] [PubMed] [Google Scholar]
- 62.O’Rourke MF, Jiang APXJ. Pulse wave analysis. British Journal of Clinical Pharmacology. 2001;51(6):507–522. doi: 10.1046/j.0306-5251.2001.01400.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Nicholsand WW, O'Rourke MF. Ascending aortic pressure waves. In: Nichols WW, O'Rourke MF, editors. McDonald's Blood Flow in Arteries Theoretical, Experimental and Clinical Principles. London, UK: Edward Arnold; 1998. [Google Scholar]
- 64.Karamanoglu M, O’Rourke MF, Avolio AP, Kelly RP. An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man. European Heart Journal. 1993;14(2):160–167. doi: 10.1093/eurheartj/14.2.160. [DOI] [PubMed] [Google Scholar]
- 65.Chen CH, Nevo E, Fetics B, et al. Estimation of Central aortic pressure waveform by mathematical transformation of radial tonometry pressure: validation of generalized transfer function. Circulation. 1997;95(7):1827–1836. doi: 10.1161/01.cir.95.7.1827. [DOI] [PubMed] [Google Scholar]
- 66.Takazawa K, O’Rourke MF, Fujita M, et al. Estimation of ascending aortic pressure from radial arterial pressure using a generalised transfer function. Zeitschrift fur Kardiologie. 1996;85(3):137–139. [PubMed] [Google Scholar]
- 67.Fetics B, Nevo E, Chen CH, Kass DA. Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry. IEEE Transactions on Biomedical Engineering. 1999;46(6):698–706. doi: 10.1109/10.764946. [DOI] [PubMed] [Google Scholar]
- 68.Marchais SJ, Guerin AP, Pannier BM, Levy BI, Safar ME, London GM. Wave reflections and cardiac hypertrophy in chronic uremia: influence of body size. Hypertension. 1993;22(6):876–883. doi: 10.1161/01.hyp.22.6.876. [DOI] [PubMed] [Google Scholar]
- 69.Oliverand JJ, Webb DJ. Noninvasive assessment of arterial stiffness and risk of atherosclerotic events. Arteriosclerosis, Thrombosis, and Vascular Biology. 2003;23(4):554–566. doi: 10.1161/01.ATV.0000060460.52916.D6. [DOI] [PubMed] [Google Scholar]
- 70.Kelly RP, Millasseau SC, Ritter JM, Chowienczyk PJ. Vasoactive drugs influence aortic augmentation index independently of pulse-wave velocity in healthy men. Hypertension. 2001;37(6):1429–1433. doi: 10.1161/01.hyp.37.6.1429. [DOI] [PubMed] [Google Scholar]
- 71.Wilkinson IB, MacCallum H, Hupperetz PC, van Thoor CJ, Cockcroft JR, Webb DJ. Changes in the derived central pressure waveform and pulse pressure in response to angiotensin II and noradrenaline in man. Journal of Physiology. 2001;530(3):541–550. doi: 10.1111/j.1469-7793.2001.0541k.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Smulyan H, Marchais SJ, Pannier B, Guerin AP, Safar ME, London GM. Influence of body height on pulsatile arterial hemodynamic data. Journal of the American College of Cardiology. 1998;31(5):1103–1109. doi: 10.1016/s0735-1097(98)00056-4. [DOI] [PubMed] [Google Scholar]
- 73.Wilkinson IB, MacCallum H, Flint L, Cockcroft JR, Newby DE, Webb DJ. The influence of heart rate on augmentation index and central arterial pressure in humans. Journal of Physiology. 2000;525(1):263–270. doi: 10.1111/j.1469-7793.2000.t01-1-00263.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Gatzka CD, Cameron JD, Dart AM, et al. Correction of carotid augmentation index for heart rate in elderly essential hypertensives. ANBP2 Investigators. Australian Comparative Outcome Trial of Angiotensin-Converting Enzyme Inhibitor- and Diuretic-Based Treatment of Hypertension in the Elderly. American Journal of Hypertension. 2001;14(6, part 1):573–577. doi: 10.1016/s0895-7061(00)01320-0. [DOI] [PubMed] [Google Scholar]
- 75.London GM, Blacher J, Pannier B, Guérin AP, Marchais SJ, Safar ME. Arterial wave reflections and survival in end-stage renal failure. Hypertension. 2001;38(3):434–438. doi: 10.1161/01.hyp.38.3.434. [DOI] [PubMed] [Google Scholar]
- 76.Riggio S, Mandraffino G, Sardo MA, et al. Pulse wave velocity and augmentation index, but not intima-media thickness, are early indicators of vascular damage in hypercholesterolemic children. European Journal of Clinical Investigation. 2010;40(3):250–257. doi: 10.1111/j.1365-2362.2010.02260.x. [DOI] [PubMed] [Google Scholar]
- 77.Russo C, Jin Z, Takei Y, et al. Arterial wave reflection and subclinical left ventricular systolic dysfunction. Journal of Hypertension. 2011;29(3):574–582. doi: 10.1097/HJH.0b013e328342ca56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Chen JY, Chou CH, Lee YL, et al. Association of central aortic pressures indexes with development of diabetes mellitus in essential hypertension. American Journal of Hypertension. 2010;23(10):1069–1073. doi: 10.1038/ajh.2010.145. [DOI] [PubMed] [Google Scholar]
- 79.Weber T, O’Rourke MF, Lassnig E, et al. Pulse waveform characteristics predict cardiovascular events and mortality in patients undergoing coronary angiography. Journal of Hypertension. 2010;28(4):797–805. doi: 10.1097/HJH.0b013e328336c8e9. [DOI] [PubMed] [Google Scholar]
- 80.Wang KL, Cheng HM, Sung SH, et al. Wave reflection and arterial stiffness in the prediction of 15-year all-cause and cardiovascular mortalities: a community-based study. Hypertension. 2010;55(3):799–805. doi: 10.1161/HYPERTENSIONAHA.109.139964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Papaioannou TG, Stamatelopoulos KS, Georgiopoulos G, et al. Arterial wave reflections during the menstrual cycle of healthy women: a reproducibility study. Hypertension. 2009;54(5):1021–1027. doi: 10.1161/HYPERTENSIONAHA.109.137703. [DOI] [PubMed] [Google Scholar]
- 82.Paul B, Hewitson CL, Woodman RJ, Mangoni AA. Analysis of short-term reproducibility of arterial vasoreactivity by pulse-wave analysis after pharmacological challenge. Clinical and Experimental Pharmacology and Physiology. 2009;36(1):49–54. doi: 10.1111/j.1440-1681.2008.05033.x. [DOI] [PubMed] [Google Scholar]
- 83.Wilkinson IB, Fuchs SA, Jansen IM, et al. Reproducibility of pulse wave velocity and augmentation index measured by pulse wave analysis. Journal of Hypertension. 1998;16(12, part 2):2079–2084. doi: 10.1097/00004872-199816121-00033. [DOI] [PubMed] [Google Scholar]
- 84.Liang YL, Teede H, Kotsopoulos D, et al. Non-invasive measurements of arterial structure and function: repeatability, interrelationships and trial sample size. Clinical Science. 1998;95(6):669–679. doi: 10.1042/cs0950669. [DOI] [PubMed] [Google Scholar]
- 85.Siebenhofer A, Kemp CRW, Sutton AJ, Williams B. The reproducibility of central aortic blood pressure measurements in healthy subjects using applanation tonometry and sphygmocardiography. Journal of Human Hypertension. 1999;13(9):625–629. doi: 10.1038/sj.jhh.1000887. [DOI] [PubMed] [Google Scholar]
- 86.Filipovsky J, Svobodovaand V, Pecen L. Reproducibility of radial pulse wave analysis in healthy subjects. Journal of Hypertension. 2000;18(8):1033–1040. doi: 10.1097/00004872-200018080-00007. [DOI] [PubMed] [Google Scholar]
- 87.Savage MT, Ferro CJ, Pinder SJ, Tomson CRV. Reproducibility of derived central arterial waveforms in patients with chronic renal failure. Clinical Science. 2002;103(1):59–65. doi: 10.1042/cs1030059. [DOI] [PubMed] [Google Scholar]
- 88.Papaioannou TG, Stamatelopoulos KS, Gialafos E, et al. Monitoring of arterial stiffness indices by applanation tonometry and pulse wave analysis: reproducibility at low blood pressures. Journal of Clinical Monitoring and Computing. 2004;18(2):137–144. doi: 10.1023/b:jocm.0000032809.71793.b8. [DOI] [PubMed] [Google Scholar]
- 89.Weber T, Auer J, O’Rourke MF, et al. Arterial stiffness, wave reflections, and the risk of coronary artery disease. Circulation. 2004;109(2):184–189. doi: 10.1161/01.CIR.0000105767.94169.E3. [DOI] [PubMed] [Google Scholar]
- 90.ter Avest E, Holewijn S, Stalenhoef AFH, De Graaf J. Variation in non-invasive measurements of vascular function in healthy volunteers during daytime. Clinical Science. 2005;108(5):425–431. doi: 10.1042/CS20040300. [DOI] [PubMed] [Google Scholar]
- 91.Papaioannou TG, Karatzis EN, Karatzi KN, et al. Hour-to-hour and week-to-week variability and reproducibility of wave reflection indices derived by aortic pulse wave analysis: implications for studies with repeated measurements. Journal of Hypertension. 2007;25(8):1678–1686. doi: 10.1097/HJH.0b013e3281ab6c58. [DOI] [PubMed] [Google Scholar]
- 92.Crilly M, Coch C, Clark H, Bruceand M, Williams D. Repeatability of the measurement of augmentation index in the clinical assessment of arterial stiffness using radial applanation tonometry. Scandinavian Journal of Clinical & Laboratory. 2007;67(4):413–422. doi: 10.1080/00365510601131946. [DOI] [PubMed] [Google Scholar]
- 93.Crilly M, Coch C, Bruce M, Clark H, Williams D. Indices of cardiovascular dunction derived from peripheral pulse wave analysis using radial applanation tonometry: a measurement repeatability study. Vascular Medicine. 2007;12(3):189–197. doi: 10.1177/1358863X07081134. [DOI] [PubMed] [Google Scholar]
- 94.Frimodt-Møller M, Nielsen AH, Kamper AL, Strandgaard S. Reproducibility of pulse-wave analysis and pulse-wave velocity determination in chronic kidney disease. Nephrology Dialysis Transplantation. 2008;23(2):594–600. doi: 10.1093/ndt/gfm470. [DOI] [PubMed] [Google Scholar]
- 95.O’Rourke MF, Kelly RP. Wave reflection in the systemic circulation and its implications in ventricular function. Journal of Hypertension. 1993;11(4):327–337. doi: 10.1097/00004872-199304000-00001. [DOI] [PubMed] [Google Scholar]
- 96.Ross R. Atherosclerosis—an inflammatory disease. The New England Journal of Medicine. 1999;340(2):115–126. doi: 10.1056/NEJM199901143400207. [DOI] [PubMed] [Google Scholar]
- 97.Gao Y. The multiple actions of NO. Pflugers Archiv European Journal of Physiology. 2010;459(6):829–839. doi: 10.1007/s00424-009-0773-9. [DOI] [PubMed] [Google Scholar]
- 98.Michel T, Vanhoutte PM. Cellular signaling and NO production. Pflugers Archiv European Journal of Physiology. 2010;459(6):807–816. doi: 10.1007/s00424-009-0765-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Moncadaand S, Higgs EA. Nitric oxide and the vascular endothelium. Handbook of Experimental Pharmacology. 2006;176, part 1:213–254. doi: 10.1007/3-540-32967-6_7. [DOI] [PubMed] [Google Scholar]
- 100.Napoli C, de Nigris F, Williams-Ignarro S, Pignalosa O, Sica V, Ignarro LJ. Nitric oxide and atherosclerosis: an update. Nitric Oxide. 2006;15(4):265–279. doi: 10.1016/j.niox.2006.03.011. [DOI] [PubMed] [Google Scholar]
- 101.Koller A, Sun D, Kaley G. Role of shear stress and endothelial prostaglandins in flow- and viscosity-induced dilation of arterioles in vitro. Circulation Research. 1993;72(6):1276–1284. doi: 10.1161/01.res.72.6.1276. [DOI] [PubMed] [Google Scholar]
- 102.Bryan RM, Jr., You J, Golding EM, Marrelli SP. Endothelium-derived hyperpolarizing factor: a cousin to nitric oxide and prostacyclin. Anesthesiology. 2005;102(6):1261–1277. doi: 10.1097/00000542-200506000-00028. [DOI] [PubMed] [Google Scholar]
- 103.Taylor SG, Weston AH. Endothelium-derived hyperpolarizing factor: a new endogenous inhibitor from the vascular endothelium. Trends in Pharmacological Sciences. 1988;9(8):272–274. doi: 10.1016/0165-6147(88)90003-x. [DOI] [PubMed] [Google Scholar]
- 104.Joannides R, Bellienand J, Thuillez C. Clinical methods for the evaluation of endothelial function—a focus on resistance arteries. Fundamental and Clinical Pharmacology. 2006;20(3):311–320. doi: 10.1111/j.1472-8206.2006.00406.x. [DOI] [PubMed] [Google Scholar]
- 105.Lekakis J, Abraham P, Balbarini A, et al. Methods for evaluating endothelial function: a position statement from the European Society of Cardiology Working Group on Peripheral Circulation. European Journal of Cardiovascular Prevention and Rehabilitation. 2011;18(6):775–789. doi: 10.1177/1741826711398179. [DOI] [PubMed] [Google Scholar]
- 106.Celermajer DS, Sorensen KE, Gooch VM, et al. Non-invasive detection of endothelial dysfunction in children and adults at risk of atherosclerosis. The Lancet. 1992;340(8828):1111–1115. doi: 10.1016/0140-6736(92)93147-f. [DOI] [PubMed] [Google Scholar]
- 107.Stonerand L, Sabatier MJ. Assessments of endothelial function using ultrasound. In: Ainslie P, editor. Ultrasound. Rijeka, Croatia: InTech; 2012. [Google Scholar]
- 108.Wilkinson IB, Qasem A, McEniery CM, Webb DJ, Avolio AP, Cockcroft JR. Nitric oxide regulates local arterial distensibility in vivo. Circulation. 2002;105(2):213–217. doi: 10.1161/hc0202.101970. [DOI] [PubMed] [Google Scholar]
- 109.Fleiss JL. The Design and Analysis of Clinical Experiments. New York, NY, USA: John Wiley & Sons; 1986. [Google Scholar]
- 110.Stoner ML, Erickson M, Young JM, Fryerand M, Sabatier MJ. There’s more to flow-mediated dilation than nitric oxide. doi: 10.5551/jat.11973. Journal of Atherosclerosis and Thrombosis. In press. [DOI] [PubMed] [Google Scholar]
- 111.McLean PG, Aston D, Sarkar D, Ahluwalia A. Protease-activated receptor-2 activation causes EDHF-like coronary vasodilation: selective preservation in ischemia/reperfusion injury: involvement of lipoxygenase products, VR1 receptors, and C-fibers. Circulation Research. 2002;90(4):465–472. doi: 10.1161/hh0402.105372. [DOI] [PubMed] [Google Scholar]
- 112.Golding EM, You J, Robertson CS, Bryan RM., Jr. Potentiated endothelium-derived hyperpolarizing factor-mediated dilations in cerebral arteries following mild head injury. Journal of Neurotrauma. 2001;18(7):691–697. doi: 10.1089/089771501750357636. [DOI] [PubMed] [Google Scholar]
- 113.Malmsjo M, Bergdahl A, Zhao XH, et al. Enhanced acetylcholine and P2Y-receptor stimulated vascular EDHF-dilatation in congestive heart failure. Cardiovascular Research. 1999;43(1):200–209. doi: 10.1016/s0008-6363(99)00062-0. [DOI] [PubMed] [Google Scholar]
- 114.Brandes RP, Behra A, Lebherz C, et al. N(G)-nitro-L-arginine- and indomethacin-resistant endothelium-dependent relaxation in the rabbit renal artery: effect of hypercholesterolemia. Atherosclerosis. 1997;135(1):49–55. doi: 10.1016/s0021-9150(97)00145-7. [DOI] [PubMed] [Google Scholar]
- 115.Thollon C, Fournet-Bourguignon MP, Saboureau D, et al. Consequences of reduced production of NO on vascular reactivity of porcine coronary arteries after angioplasty: importance of EDHF. British Journal of Pharmacology. 2002;136(8):1153–1161. doi: 10.1038/sj.bjp.0704828. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Miura H, Wachtel RE, Liu Y, et al. Flow-induced dilation of human coronary arterioles important role of Ca2+-activated K+ channels. Circulation. 2001;103(15):1992–1998. doi: 10.1161/01.cir.103.15.1992. [DOI] [PubMed] [Google Scholar]
- 117.FitzGerald GA, Smith B, Pedersen AK, Brash AR. Increased prostacyclin biosynthesis in patients with severe atherosclerosis and platelet activation. The New England Journal of Medicine. 1984;310(17):1065–1068. doi: 10.1056/NEJM198404263101701. [DOI] [PubMed] [Google Scholar]
- 118.Mayahi L, Mason L, Bleasdale-Barr K, et al. Endothelial, sympathetic, and cardiac function in inherited (6R)-L-Erythro-5,6,7,8-Tetrahydro-L-Biopterin deficiency. Circulation. 2010;3(6):513–522. doi: 10.1161/CIRCGENETICS.110.957605. [DOI] [PubMed] [Google Scholar]
- 119.Hornig B, Maier V, Drexler H. Physical training improves endothelial function in patients with chronic heart failure. Circulation. 1996;93(2):210–214. doi: 10.1161/01.cir.93.2.210. [DOI] [PubMed] [Google Scholar]
- 120.Ghiadoni L, Versari D, Magagna A, et al. Ramipril dose-dependently increases nitric oxide availability in the radial artery of essential hypertension patients. Journal of Hypertension. 2007;25(2):361–366. doi: 10.1097/HJH.0b013e3280115901. [DOI] [PubMed] [Google Scholar]