Abstract
A very simple, selective, and fast flow injection spectrophotometeric method is developed for determination of nickel using furildioxime as complexing agent. Micellar solution of brij-35 is employed to solubilize the sparingly soluble complex of Ni-furildioxime in buffered aqueous system (pH-9.00). Under optimized conditions, absorbance is linear from 0.02 to 10 μg mL−1 using 500 μL sample volume and from 10 to 30 μg mL−1 using 50 μL sample volume of nickel at 480 nm, with R 2 = 0.9971 and 0.9916, respectively. The molar absorption coefficient and Sandell's sensitivity were 6.0 × 103 L mol−1 cm−1 and 0.01 ng cm−2, respectively. The sample throughput of the method is 120 samples per hour with RSD of 0.01–0.2% for 0.02 to 10 μg mL−1 nickel (n = 5), indicating that the method is highly precise and reproducible. Interference from cobalt is removed by Nitroso R-salt-modified XAD-16. The developed method is validated by analysing certified reference materials and is applied to assess nickel content of commercially available cigarettes.
1. Introduction
Interest in the determination of nickel has increased over the last few years because of the influence of this metal on humans. Adverse effects of inorganic, water-soluble nickel species occur after contact with skin, causing nickel dermatitis. After inhalation, respiratory tract irritation and asthma can also result. Besides carcinogenensis, bioassays in animals showed that certain nickel compounds are potent carcinogens when administered via parenteral routes [1].
In rapidly expanding analytical fields, such as environmental, biological, and material monitoring of trace metals, there is an increasing need to develop simple, sensitive, selective, and reproducible analytical techniques that do not use expensive or complicated test equipments. Many sensitive instruments, such as spectrofluorimetry, X-ray fluorescence spectrometry, neutron activation analysis, atomic absorption spectrophotometery, and chemiluminescence have been widely applied for the determination of nickel [2–8]. However, spectrophotometric methods still have the advantages of simplicity and require no expensive or complicated test equipment. Also, there are various automated spectrophotometric methods (flow injection based) for the determination of nickel, but due to extraction of complex in organic solvents, complicated FIA setups are used [9]. A flow Injection visible diffused reflectance method by precipitating nickel with furildioxime was reported but reproducibility of peak heights is not satisfactory [10].
Spectrophotometrically, nickel is determined by using complexing agents, such as bis(acetylacetone)ethylenediimine, 2-(2-Quinolylazo)-5-diethylaminoaniline, dimethylglyoxime, and furildioxime [9, 11–13].
Oximes are group of compounds, which reacts with limited number of metal ions, hence come under the umbrella of selective reagents. These form water-insoluble complexes extractable into organic solvents, but the extraction is not 100% due to partial solubility of complexes. Furildioxime forms complex with nickel that is absorbed in chloroform at 438 nm [12]. Inspite of high selectivity of this reagent toward nickel, its use is limited due to sparingly soluble precipitates and solubilizing of Ni-furildioxime [14]. Recently, dissolution of precipitate formed by Ni-furildioxime is tried by ultrasonic irradiation but quick transfer from ultasnoic bath to spectrophotometer is the condition for analysis as turbidity reappears [15].
The ability of surfactants to solubilize the water-insoluble complexes in homogenous environment can overcome this problem. There is only one report that addresses the flow injection determination of nickel using pyridylazoresorcinol in TX-100 solution but it is laboratory exercise published by undergraduate students and proper optimization is not worked out [16]. Micellar solutions in FI have unique feature of sensitivity and speed of analysis with less waste for treatment. Micellar solutions obviate the need of phase separator in FIA systems which resulted in simple setups with more reproducible assays [14, 17, 18].
This study is aimed at developing simple, rapid, selective, and sensitive assay for nickel. Furildioxime is selected as complexing reagent due to its selective behavior toward nickel. Micellar solutions are used as microheterogonous system to carry out complexation reaction, and flow injection is employed to automate the system and to provide high-throughput determination. Assay is made selective by removing cobalt, major interfering metal ion, by using NitrosoR-salt modified XAD-16.
2. Experimental
2.1. Apparatus
A four-channel peristaltic pump (Gilson Minipuls 3) was used to propel the carrier stream, equipped with PVC pump tubes with i.d 0.5 mm (anachem). Samples were injected via rotary PTFE valve (Rheodyne 5020). PTFE 0.5 mm i.d. tubing was used throughout the remainder system. The detector used was Spectronic 20 at 488 nm equipped with 100-μL flow through cell. Yew strip chart recorder was used to record the signal. Shimadzu UV-365 spectrophotometer was used to record the spectra of nickel-furildioxime system. A Perkin-Elmer (Lamda-2, Germany) double-beam spectrophotometer was used for the comparison of the results. A digital pH meter (WTW, Inolab pH level 1, Germany) was used to measure the pH of the solution.
2.2. Reagents
All chemicals used were of analytical reagent grade or highest purity available. Double-distilled deionized water was used throughout.
2.3. Furildioxime Solution
A 5% solution was prepared by dissolving the 5.0 g of furildioxime (Merck, Darmstadt, Germany) in 100 mL of methanol (Merck).
2.4. Nickel(II) Standard Solution
A 100 mL stock solution (1 mg mL−1) of nickel was prepared by dissolving appropriate amount of nickel sulfate (Fluka, Switzerland) in double-distilled deionized water containing 1 mL of 1 + 1 nitric acid. More diluted solutions were prepared from this stock solution freshly when required in double-distilled deionized water.
2.5. Brij-35 Stock Solution (5%)
A 500 mL stock solution of Brij-35 (Merck) was prepared by dissolving 50 g of Brij-35 in 250–300 mL of water and sonicated for 15 minutes and was diluted to mark with double-distilled deionized water when become transparent.
2.6. Other Solutions
Solutions of the large number of anions, cations, and complexing agents were prepared from their analaR grade or equivalent grade water-soluble salts. All glassware was kept in nitric acid (1 + 1) for at least a day and then washed with deionized water. Stock solutions and environmental water sample were kept in polypropylene bottles containing 1 mL concentrated nitric acid.
2.7. Procedure
2.7.1. Preparation of Reagent for FIA
A single-channel FIA manifold was used throughout. Reagent “R” was prepared by taking 0.05 mL of Brij-35 (5%) and 5.0 mL of buffer (NH4Cl/NaOH buffer, pH = 9.00, and I = 1) was mixed and 0.4 mL 5% furildioxime was added and mixed well, then diluted to 25.0 mL with deionized water. The reagent was pumped at the flow rate of 2.0 mL min−1. The sample was injected at position “I” into flowing reagent stream. The sample reacts at point reaction coil “RC” of 100 cm and is detected at 488 nm (Figure 1).
Figure 1.

Block diagram for the FIA setup used for determination of nickel using furildioxime.
2.7.2. Nitroso-R Salt-Modified XAD-16
Amberlite XAD-16 (5 g) and 50 mL of aqueous 0.1% nitroso-R salt were stirred with the help of magnetic stirrer. Absorbance of nitroso-R salt in solution was monitored continuously using UV-Visible spectrophotometer. When the absorbance of nitroso-R salt became constant, the resin was filtered and dried. The amount of chelating reagent required for the modification of surface of XAD-16 was optimized by varying the concentration and volumes of nitroso-R salt. Recovery (%) of nickel was found to be quantitative using 50 mL of 0.1% nitroso-R salt.
3. Results and Discussion
Various chemical and FIA parameters were studied to get optimum signal (absorbance). Studied range, and selected parameters are shown in Table 1.
Table 1.
Selected chemical and FIA parameters obtained with optimization experiments.
| Parameter | Studied range | Selected value |
|---|---|---|
| Size of sample loop (μL) | 50–500 | 500 |
| Reagent flow rate (mL min−1) | 1.0–4.2 | 1.2 |
| Length of reaction coil (cm) | 50–300 | 100 |
| pH | 3.5–12.0 | 9.0 |
| Concentration of reagent (%) | 0.04–0.25 | 0.08 |
| Brij-35 (%) | 0.01–2.0 | 0.01 |
| NH4Cl/NaOH (M) | 0.04–0.4 | 0.2 |
3.1. Factors Affecting Absorbance of Nickel-Furildioxime Complex
3.1.1. Absorption Spectra
The absorption spectra of nickel-furildioxime in Brij-35 at pH 9.00 was recorded using Shimadzu UV-365 spectrophotometer, Japan. The absorption spectrum of nickel-furildioxime shows the maximum absorbance at 480 nm. In all instances, measurements were made at 480 nm against reagent blank.
3.1.2. Effect of Surfactant Type and Concentration
Various types of surfactants; anionic (sodium dodecyl sulfate), cationic (cetyl trimethyl ammonium bromide), and nonionic (Brij-35) were studied; Brij-35 was found to be the best surfactant for the system. Ionic surfactants were rejected for not providing adequate solubility and high background signal which hampers the determination of metal ions at low concentration. In 0.01% Brij-35 medium, however, maximum absorbance was observed; hence 0.01% Brij-35 was used in the determination procedure. It was observed that very little concentration is enough to solublize the complex, while on increasing concentration, the absorbance decreases and remains constant after 0.2% Brij-35. This may be due to micellar dilution effect and also the viscosity of the system increases with surfactant concentration which in turn decreases absorbance [19].
3.1.3. Effect of pH
Effect of pH was observed by varying pH of reagent from 3.5–12.0 using HCl/NaOH at room temperature. The absorbance was maximum at pH 9.0, using final concentration of buffer 0.2 M at room temperature 25 ± 5°C. Outside this range, the absorbance decreased (Figure 2). This is due to deprotonation of furildioxime at higher pH which favors the formation of complex with positively charged Ni2+. The same trend is reported for extraction systems and complexation in aqueous medium, this shows that pH-dependant chemistries are not altered in micellar medium [20]. For all subsequent measurements, 0.2 M buffer (NH4Cl/NaOH) at pH 9.0 was used.
Figure 2.

Effect of pH on the absorbance of nickel-fruildiome complex.
3.1.4. Effect of Reagent Concentration
Different volumes of stock 5% methanolic solution of furildioxime were added to prepare reagent “R” for FIA, and fixed concentration of metal ion was injected (50 μL). It was observed that on increasing concentration, signal increases up to 0.1% and then starts decreasing. Low reagent concentration may lead to incomplete complexation, but at higher concentration, the background signal increases and reduces net signal so, reagent concentration of 0.1% was used for further studies.
3.1.5. Effect of FIA Parameters
Flow parameters like pump speed (flow rate), reaction coil length, and sample volume were studied by varying pump speed, home-made reaction coils from Teflon tubes, and loops of different injection volumes, respectively. There is very small effect on the signal on increasing pump speed. Decrease in absorbance on increasing reaction coil length was observed, this shows that the reaction in micellar medium is spontaneous and delaying the reactants in coil will not shift the equilibrium to produce more products. So, reaction coil length of 100 cm and pump speed 12 rpm (corresponds to 2.1 mL min−1) was selected. Although, increasing sample volume increased absorbance from 0.2–0.5 for 50–500 μLs. Large sample volume was used to construct calibration graphs at lower concentrations while 50-μL loop was used at higher concentrations which produce sharp peaks.
3.1.6. Analytical Figures of Merit
The absorbance was linear from 0.02 to 10 μg mL−1 using 500 μL sample volume (y = 0.0927 + 0.0331) and 10–30 μg mL−1 using 50-μL sample volume (y = 0.0428 + 0.0764) of nickel at 480 nm, with R 2 = 0.9971 and 0.9916, respectively. The molar absorption coefficient and Sandell's sensitivity were 6.0 × 103 L mol−1 cm−1 and 0.01 ng cm−2, respectively.
The sample throughput of the method is 120 samples per hour. The RSD (n = 5) is 0.01–0.2% for 0.02 to 10 μg mL−1 nickel, indicating that the method is highly precise and reproducible. Salient features of proposed method are summarized in Table 2.
Table 2.
Salient features of the proposed method.
| Parameter | Optimized value |
|---|---|
| pH | 9.0 |
| Linear range (μg mL−1) | 0.02–30 |
| Detection Limit (ng mL−1) | 10 |
| Dispersion Coefficient | 4.0 |
| Reproducibility | 0.0–0.2 |
| Sample throughput | 120 samples per hour |
3.1.7. Effect of Foreign Ions
The effect of over 40 cations, anions, and complexing agents were studied on the determination of only 1 μg mL−1 of nickel. The criterion for interference was an absorbance value varying for more than 5% for the expected value of nickel alone [21]. No interference was found from the following; 1000-fold amount of sulfate, sulfite, nitrate, bromide, chloride, iodide, Na, K, Ca, N H 4 +, tartarate, citrate, thiosulfate, hydrogen phosphate, and azide. 50-fold amount of Fe2+, Cd2+, Cr3+, MoVI, CeIV, Sb3+, CrVI, SeIV, Fe3+, Ag+, Ba2+, As3+, Pb2+, In3+, Tl+, Mn2+, Pd2+, VV, Be2+, Sn2+, Bi2+, Zn2+, Hg+, and Hg2+. Cobalt, iron, and copper are interfering ions in spectrophotometeric determination of nickel using furildioxime [15]. Here, using nonionic micellar solutions, no interference was observed from iron and copper while cobalt interference remains problem and was selectively removed using modified XAD-16.
Sample solution (5 mL) containing different metal ions including Ni and cobalt was passed through column packed with 0.2 g of modified XAD-16. Nitroso R salt complexes of both Ni and cobalt were retained on column. Therefore, in order to recover retained Ni, column was washed with 2.5 mL of 1 N HNO3 [12]. Cobalt-furildioxime complex is stable at lower pH (1 N HNO3) thus remain attached with modified resin while Ni-furildioxime decomposes and gets eluted with acid. The pH of eluent was neutralized with NaOH and 50 or 500 μLs injected in the flowing stream of reagent as mentioned above.
3.2. Applications
3.2.1. Determination of Nickel in Synthetic Mixtures
Several synthetic mixtures of varying compositions containing nickel and diverse ions of known concentration were determined by the present method. The results are shown in Table 3.
Table 3.
Determination of nickel in synthetic mixtures.
| Sample | Composition of mixture | Nickel in μg mL−1 | ||
|---|---|---|---|---|
| Added | Found | % Recovery | ||
| A | Nickel | 2.0 | 2.0 | 100 |
| B | A + VV (25 fold) | 2.0 | 1.8 | 90 |
| C | B + Zn2+ (25 fold) + Ca (25 fold) |
2.0 | 1.8 | 90 |
| D | C + Fe (25 fold) + CrVI (25 fold) |
2.0 | 2.2 | 110 |
3.2.2. Determination of Nickel in Cigarette Samples
The proposed method was applied to the determination of nickel tobacco samples. 5 g of each tobacco sample was placed in a breaker of 100 mL capacity and after adding 2 mL of concentrated HNO3, the sample was heated almost to dryness on a hot plate at 110°C, then 2 mL of concentrated HNO3 was again added to the residue and the solution was heated at 150°C for 2 h. After adding a further 2 mL of concentrated HNO3 and 1 mL of 60% HClO4, the solution was heated at 150°C for 4 h until white fumes appeared. This procedure was repeated twice. Solution was cooled, filtered, and diluted with deionized water. The nickel content was assayed by standard addition method. Results are summarized in Table 4.
Table 4.
Determination of nickel in tobacco samples.
| Sample | Amount of nickel μg g−1 | RSD (%), n = 3 |
|---|---|---|
| Sample-1 | 0.760 | 1.3 |
| Sample-2 | 0.110 | 2.2 |
| Sample-3 | 0.75 | 1.4 |
| Sample-4 | 0.50 | 2.4 |
| Sample-5 | 0.00 | 0.00 |
| Sample-6 | 0.125 | 3.4 |
| Sample-7 | 0.60 | 1.4 |
| Sample-8 | 0.70 | 2.1 |
| Sample-9 | 0.150 | 2.9 |
3.2.3. Validation of Method
Some of the reference materials were analysed by proposed method and % recovery was calculated. The results are shown in Table 5.
Table 5.
Determination of nickel in Certified Reference Materials.
| CRM | Composition of CRM | Nickel found | % recovery |
|---|---|---|---|
| Alloy | 50% nickel, 50% aluminium |
49.99 | 99.9 |
|
| |||
| BAS (69b) stainless steel | 18.6% Cr, 9.35% nickel |
9.32 | 99.6 |
|
| |||
| BCS 261, stainless steel | 0.083% C, 0.39% Si, 17.2% Cr, 13.08% Ni, 0.71% Ta & Nb, 0.66% Mn |
19.0 | 105 |
4. Conclusion
A highly selective, sensitive, simple, and rapid method for the determination of nickel using furildioxime in nonionic micellar medium has been developed. Use of micelles eliminates the toxic organic solvents, otherwise, necessary. Using homogenous environment of organized assemblies, the system was coupled with FIA, which increases sample throughput and minimizes sample consumption. The method was successfully applied for the determination of nickel in tobacco samples and complex synthetic mixtures.
Acknowledgment
All the authors of the manuscript declare that they do not have a direct financial relation with any commercial identity mentioned in the paper that might lead to a conflict of interest for any of the authors.
References
- 1.Miró M, Estela JM, Cerdà VC. Application of flowing stream techniques to water analysis: part III. Metal ions: alkaline and alkaline-earth metals, elemental and harmful transition metals, and multielemental analysis. Talanta. 2004;63(2):201–223. doi: 10.1016/j.talanta.2003.10.047. [DOI] [PubMed] [Google Scholar]
- 2.Hu Q, Yang G, Zhao Y, Yin J. Determination of copper, nickel, cobalt, silver, lead, cadmium, and mercury ions in water by solid-phase extraction and the RP-HPLC with UV-Vis detection. Analytical and Bioanalytical Chemistry. 2003;375(6):831–835. doi: 10.1007/s00216-003-1828-y. [DOI] [PubMed] [Google Scholar]
- 3.Zhang M-H, Liang Y-Z. Rapid and simultaneous determination of copper, cadmium, nickel, and cobalt in zinc electrolyte solutions by complex adsorption wave polarography. Journal of Trace and Microprobe Techniques. 2002;20(1):1–14. [Google Scholar]
- 4.Benzo Z, Marcano E, Gómez C, et al. Simultaneous determination of nickel, iron, and copper in margarine by inductively coupled plasma-atomic emission spectroscopy after sample emulsification. Journal of AOAC International. 2002;85(4):967–970. [PubMed] [Google Scholar]
- 5.Sreenivasa Rao K, Balaji T, Prasada Rao T, Babu Y, Naidu GRK. Determination of iron, cobalt, nickel, manganese, zinc, copper, cadmium and lead in human hair by inductively coupled plasma-atomic emission spectrometry. Spectrochimica Acta B. 2002;57(8):1333–1338. [Google Scholar]
- 6.Zendelovska D, Pavlovska G, Cundeva K, Stafilov T. Electrothermal atomic absorption spectrometric determination of cobalt, copper, lead and nickel traces in aragonite following flotation and extraction separation. Talanta. 2001;54(1):139–146. doi: 10.1016/s0039-9140(00)00645-7. [DOI] [PubMed] [Google Scholar]
- 7.Ershova NI, Ivanov VM. Improvement of direct determination of trace nickel in environmental samples by diffuse reflection spectroscopy using chromaticity characteristics. Fresenius’ Journal of Analytical Chemistry. 2000;367(2):210–211. doi: 10.1007/s002160051627. [DOI] [PubMed] [Google Scholar]
- 8.Füchtjohann L, Jakubowski N, Gladtke D, Barnowski G, Klockow D, Broekaert JAC. Determination of soluble and insoluble nickel compounds in airborne participate matter by graphite furnace atomic absorption spectrometry and inductively coupled plasma mass spectrometry. Fresenius’ Journal of Analytical Chemistry. 2000;366(2):142–145. doi: 10.1007/s002160050026. [DOI] [PubMed] [Google Scholar]
- 9.Chimpalee N, Chimpalee D, Keawpasert P, Thorburn Burns D. Flow injection extraction spectrophotometric determination of nickel using bis(acetylacetone)ethylenediimine. Analytica Chimica Acta. 2000;408(1-2):123–127. [Google Scholar]
- 10.Tubino M, Queiroz CAR. Flow injection visible diffuse reflectance quantitative analysis of nickel. Analytica Chimica Acta. 2007;600(1-2):199–204. doi: 10.1016/j.aca.2006.11.034. [DOI] [PubMed] [Google Scholar]
- 11.Hu Q, Yang G, Huang Z, Yin J. Determination of nickel with 2-(2-quinolylazo)-5-diethylaminoaniline as a chromogenic reagent. Analytical Sciences. 2003;19(10):1449–1452. doi: 10.2116/analsci.19.1449. [DOI] [PubMed] [Google Scholar]
- 12.Fries J, Getrost H. Organic Reagents for Trace Analysis. Vol. 99. Darmstadt, Germany: E. Merck; 1977. [Google Scholar]
- 13.Anil KD, Khopkar SM, Chalmer RA. Solvent Extraction of Metal. London, UK: Van Nostrand Reinhold Company Ltd.; 1970. [Google Scholar]
- 14.Memon N, Bhanger MI. Micellar liquid chromatographic determination of aluminum as its complex with 8-hydroxyquinoline-5-sulfonic acid. Acta Chromatographica. 2004;(14):172–179. [Google Scholar]
- 15.de Sousa CS, Korn M. Effects of ultrasonic irradiation on the spectrophotometric determination of nickel with dimethylglyoxime. Analytica Chimica Acta. 2001;444(2):309–315. [Google Scholar]
- 16.Ostlund SG, Pharr DY. Determination of cobalt(II) by flow injection analysis using a surfactant solvent system. Journal of Undergradute Chemistry Research. 2004;2(1) [Google Scholar]
- 17.Memon N, Ahmed MJ, Bhanger MI. Micellar flow injection spectrophotometric determination of indium and its application to the environmental samples. Pakistan Journal of Analytical Chemistry. 2005;6(1):16–21. [Google Scholar]
- 18.Memon N, Memon MA, Bhanger MI, Memon MH. Flow injection spectrophotometric determination of ascorbic acid using iron(III)-bathophenathroline disulfonic sodium salt. The Nucleus. 2003;40(1–4):115–118. [Google Scholar]
- 19.Balouch A, Memon N, Bhanger MI, Khuhawar MY. Utilization of picolinaldehyde-4-phenyl-3-thiosemicarbazone in sodium dodecylsulfate micelles for the spectrophotometric determination of iron, vanadium, and cobalt following partial least-squares regression analysis. Journal of AOAC International. 2009;92(1):248–256. [PubMed] [Google Scholar]
- 20.Gahler A, Mitchell A, Mellon M. Colorimetric determination of nickel with alpha-furildioxime. Analytical Chemistry. 1951;23(3):500–503. [Google Scholar]
- 21.Ojeda CB, de Torres AG, Rojas FS, Pavon JMC. Fluorimetric determination of trace amounts of gallium in biological tissues. Analyst. 1987;112(11):1499–1501. doi: 10.1039/an9871201499. [DOI] [PubMed] [Google Scholar]