. 2012 Jun;13(6):503–510. doi: 10.1631/jzus.B1100337

Electrocatalytic response of poly(cobalt tetraaminophthalocyanine)/multi-walled carbon nanotubes-Nafion modified electrode toward sulfadiazine in urine*

Xiao-ping Hong 1,2, Yan Zhu 1, Yan-zhen Zhang 3,†,
PMCID: PMC3370295  PMID: 22661213

Abstract

A highly sensitive amperometric sulfadiazine sensor fabricated by electrochemical deposition of poly(cobalt tetraaminophthalocyanine) (poly(CoIITAPc)) on the surface of a multi-walled carbon nanotubes-Nafion (MWCNTs-Nafion) modified electrode is described. This electrode showed a very attractive performance by combining the advantages of CoIITAPc, MWCNTs, and Nafion. Compared with the bare glassy carbon electrode (GCE) and the MWCNTs-Nafion modified electrode, the electrocatalytic activity of poly(CoIITAPc)-coated MWCNTs-Nafion GCE generated greatly improved electrochemical detections toward sulfadiazine including low oxidation potential, high current responses, and good anti-fouling performance. The oxidation peak currents of sulfadiazine obtained on the new modified electrode increased linearly while increasing the concentration of sulfadiazine from 0.5 to 43.5 μmol/L with the detection limit of 0.17 μmol/L.

Keywords: Sulfadiazine, Electrochemical detection, Cobalt tetraaminophthalocyanine, Multi-walled carbon nanotubes

1. Introduction

Sulfonamides are broad-spectrum and low-cost synthetic antibiotics used in treatment of diseases such as urinary tract infections, pneumocystis pneumonia, chronic bronchitis, meningococcal meningitis, and toxoplasmosis (Catalano-Pons et al., 2004; Joseph and Kumar, 2010). Over-dosage of sulfonamides can cause allergic reactions, antibiotic resistance, and carcinogenic effects (Hasebe and Osteryoung, 1975). Because of poor water solubility, sulfonamides can also lead to urinary tract lithiasis such as renal calculus and vesical calculus (Catalano-Pons et al., 2004). Therefore, quick systematic monitoring of sulfonamides in body fluids is an important analytical task.

Several methods have been developed for measurement of sulfonamide concentration including high performance liquid chromatography coupled with fluorimetric detection (HPLC-Fluo) (Pang et al., 2003; Maudens et al., 2004) or ultra-violet detection (HPLC-UV) (Malintan and Mohd, 2006), gas chromatography (GC) (Chiavarino et al., 1998), mass spectrometry (MS) (Hartig et al., 1999; Göbel et al., 2004; Conley et al., 2008), enzyme-linked immunosorbent assay (ELISA) method (Assil et al., 1992; Díaz-Cruz et al., 2003; Shelver et al., 2008), and time-resolved fluoroimmunoassay (TRFIA) (Zhang et al., 2010). Although these methods have advantages of high sensitivity and accuracy, they have drawbacks such as the high cost of the instruments, tedious sample preparation, and the necessity of skilled laboratory personnel. Consequently, it is of great importance to develop a sensitive, selective, cheap, and convenient analytical method for monitoring the level of sulfonamides and their metabolites in body fluids such as blood and urine.

Electrochemical methods have several advantages over other methods such as simplicity, low cost, and facility of use (Tsai and Chiu, 2007). Mediators such as rhodium particles deposited polyvinyl pyridine film (Shaidarova et al., 2004), palladium particles deposited polyaniline film (Shaidarova et al., 2006), cobalt phthalocyanine complex (Wang et al., 1988), and ruthenium oxide-ruthenium cyanide composites film (Shaidarova et al., 2008; 2011) were successfully used as the electrode modifier for enhancing electron-transfer kinetics and lowering the overvoltage. Modified electrodes such as carbon nanotubes (CNTs)-Nafion modified electrodes (Issac and Kumar, 2009), boron-doped electrodes (Preechaworapun et al., 2006; Souza et al., 2008), metalloporphyrin modified electrodes (Joseph and Kumar, 2010), bismuth film electrodes (Campestrini et al., 2010), and amperometric DNA-peroxidase sensors (Evtugyn et al., 2005) were designed for sulfonamide antibiotics determination. The results demonstrated that electrochemical methods have advantages of high sensitivity, good stability, and promising selectivity. Hence, electrochemical determination methods will be a new avenue for sulfonamide antibiotics detection.

Electrochemical detection based on CNTs-modified electrodes is an attractive research area. As electrode materials, CNTs exhibit fast electron transfer kinetics, high surface area, low over-potential, and promising surface anti-fouling capability (Kachoosangi et al., 2009; Zhou et al., 2011). Surface-bound mediators, such as gold nanoparticles (Li and Lin, 2007), metalloporphyrins (Chen et al., 2001), and metallophthalocyanines (Wang et al., 1988), were covalently or non-covalently stabilized onto CNTs fibers to acquire fantastic synergistic effect in electrocatalytic reactions. As versatile electrocatalysts, cobalt phthalocyanine (CoIIPc) complexes were shown to decrease the electro-oxidation potential by several hundred millivolts toward a wide variety of oxidizable species (Wang et al., 1988). However, non-covalently adsorbed CoIIPc complexes are not stable at higher pH conditions due to the decomposition of CoIIPc (Tse et al., 1995). It was noticed that potential scanning around the cobalt tetraaminophthalocyanine (CoIITAPc) oxidation potential in dimethylformamide or dimethylsulfoxide solution could cause electrochemical oxidative polymerization of CoIITAPc and deposition of an electroactive poly(CoIITAPc) onto the surface of the electrode (Li and Guarr, 1989).

In the present study, we aimed at developing a voltammetric sensor fabricated by using poly(CoIITAPc) coated multi-walled carbon nanotubes (MWCNTs) modified glassy carbon electrode (GCE) for trace sulfadiazine determination in urinary samples.

2. Materials and methods

2.1. Chemicals and apparatus

Sulfadiazine (analytical grade) and Nafion (5% (w/w) in low aliphatic alcohols and water) were purchased from Sigma-Aldrich (Shanghai, China). MWCNTs crude material (diameter 10–30 nm, length 5–15 μm, purity >95%) was obtained from Shenzhen Nanotech Port, China. Other pure analytical quality chemicals such as N,N′-dimethylformamide (DMF), sulfuric acid (98% (w/w)), nitric acid (70% (w/w)), and hydrogen peroxide (30% (w/w)) were obtained from standard suppliers and used as received. All solutions were prepared with Milli-Q water from an Academic Millipore system (Millipore Inc., MA, USA).

Electrochemical measurements were performed by using a CHI 660A electrochemical workstation (CH Instruments Inc., Austin, USA). The electrochemical cell was assembled with a conventional three-electrode system: GCE (3.0 mm diameter), MWCNTs-Nafion or poly(CoIITAPc)/MWCNTs-Nafion modified GCE as working electrodes, an AgCl/Ag electrode (saturated potassium chloride) as the reference electrode, and a platinum wire as the auxiliary electrode. Indium-tin oxide (ITO) glass was used as the substrate for morphologic and elementary characterization. All experiments were carried out under ambient conditions.

2.2. Fabrication of modified electrode

Crude MWCNTs were pretreated according to the method of Liu et al. (1998) prior to use. In brief, 100 mg of MWCNTs crude material was refluxed in 2.6 mol/L of nitric acid for 48 h at 80 °C. Then, the purified MWCNTs were rinsed to neutral with double-distilled water. After drying in a vacuum drying oven at 60 °C, the nitric acid purified MWCNTs were dispersed in a mixture of concentrated sulfuric and nitric acids (3:1 in volume) and then sonicated for 6 h. Then, the pretreated MWCNTs were rinsed to neutral with double-distilled water and dried at 60 °C in a vacuum drying oven. A suspension of MWCNTs was prepared by dispersing 2 mg of pretreated MWCNTs into 2 ml of DMF solution containing 0.1% (w/w) of Nafion through ultrasonication and a black suspension was obtained.

Prior to electrode modification, a GCE was polished to a mirror-like surface with 0.05 μm alumina slurry, thoroughly rinsed with double-distilled water, and then sonicated in anhydrous ethanol and double-distilled water (each for 3 min). Then, 2 μl of above-mentioned black suspension was cast onto the surface of the cleaned GCE and dried in ambient temperature to fabricate MWCNTs-Nafion modified GCE. The poly(CoIITAPc)/MWCNTs-Nafion modified GCE was prepared by successive voltammetric scanning in 0.2 mol/L phosphate buffer solution (pH 7.8) containing 110 mg/L of CoIITAPc with MWCNTs-Nafion modified GCE as the working electrode. MWCNTs-Nafion modified ITO electrode and poly(CoIITAPc)/MWCNTs-Nafion modified ITO electrode were prepared for morphological and elementary characterization by using cleaned ITO glass as electrode substrate. The morphologies of electrode surfaces and energy-dispersive X-ray spectra were studied by field-emission scanning electron microscope (FESEM, Hitachi S-4800, Japan) equipped with an energy-dispersive X-ray spectroscopy (EDS, Hitachi, Japan).

2.3. Sample preparation

Drug-free human urine samples used in this study were obtained from a healthy donor and stored in a refrigerator immediately after collection. After centrifugation for 5 min at 1 500 r/min, the urine supernatant was diluted five times with sulfuric acid to reach pH 0.7. Then, the solution was applied for analysis without any further pretreatment. The standard addition method was used to calculate the recovery of this method.

3. Results and discussion

3.1. Electrochemical deposition of CoIITAPc

CoIITAPc was electrochemically deposited onto MWCNTs-Nafion modified electrode by cyclic voltammetric scanning in a phosphate buffer solution (pH 7.8) containing 110 mg/L of CoIITAPc. As depicted in Fig. 1, the currents of redox couple around 0.45 V steadily increased with voltammetric cycling. After scanning for 32 cycles, a stable redox response was obtained. This signifies the electro-deposition of CoIITAPc onto the surface of MWCNTs-Nafion modified electrode. The redox peaks occurring between 0.4 and 0.6 V are assigned to CoIITAPc film oxidation and reduction, respectively.

Fig. 1.

Fig. 1

Electrochemical deposition of CoIITAPc onto MWCNTs-Nafion modified electrode

The scanning electron microscopic (SEM) characterization of the electrode modified with different complexes was carried out. Fig. 2 shows the morphologies of MWCNTs-Nafion (Fig. 2a) and poly(CoIITAPc)/MWCNTs-Nafion (Fig. 2b) electrode surfaces. It is discernible that poly(CoIITAPc) film deposition took place as the morphology of these images was different. In Fig. 2a, MWCNTs (about 20–30 μm in diameter) can be seen clearly and distribute homogeneously on the surface of the electrode. However, in Fig. 2b, it can be observed that small particles are adsorbed on the tubes and the edges of tubes become indistinct. This phenomenon could be attributed to the aggregation of poly(CoIITAPc) molecules onto the surface of MWCNTs-Nafion.

Fig. 2.

Fig. 2

SEM images of MWCNTs-Nafion film (a) and poly(CoIITAPc)/MWCNTs-Nafion film (b) on ITO substrates

The EDS surface analysis was also performed to provide information on materials present on the electrodes’ surfaces. In Fig. 3, the dominating carbon (C) peaks were expected, as the CNTs immobilized on the electrode surface. Strong peaks due to silicon (Si), indium (In), sodium (Na), and calcium (Ca) are observed and these attribute to the ITO substrate. Weak peaks (Co) at about 0.8 and 6.9 keV are also observed in the EDS spectrum. The presence of the Co (L and K) orbital peaks from the modified electrode shows the presence and the successful deposition of poly(CoIITAPc) onto MWCNTs-Nafion surface to form poly(CoIITAPc)/MWCNTs-Nafion modified electrode.

Fig. 3.

Fig. 3

EDS spectrum for a poly(CoIITAPc)/MWCNTs-Nafion modified electrode

KCnt: 1 000 counts

3.2. Electrocatalytic oxidation of sulfadiazine at poly(CoIITAPc)/MWCNTs-Nafion modified electrode

In order to test the potential electrocatalytic oxidation of the modified electrodes, cyclic voltammetric responses of poly(CoIITAPc)/MWCNTs-Nafion modified GCE, MWCNTs-Nafion modified GCE, and GCE were obtained in 0.1 mol/L sulfuric acid solution with the presence of 0.09 mmol/L of sulfadiazine. As shown in Fig. 4, a weak anodic peak at about 1.1 V is observed (solid line) which is a typical electrochemical response of sulfadiazine corresponding to the oxidation of a phenylamino group (Msagati and Ngila, 2002). As expected, when MWCNTs-Nafion modified GCE was used as the working electrode, the anodic peak of sulfadiazine was greatly enhanced (dashed line) with the anodic potential shifting negatively to 1.0 V. Surprisingly, when poly(CoIITAPc)/MWCNTs-Nafion modified GCE was used as the working electrode, the anodic peak was enhanced much greater (dot line) than that at MWCNTs-Nafion modified GCE. The anodic potential shifted to 0.9 V (dotted line), which was lower than that obtained on GCE or MWCNTs-Nafion modified GCE. The enhanced current responses and low anodic potential observed on poly(CoIITAPc)/MWCNTs-Nafion modified GCE probably attribute to the synergistic effect of MWCNTs and poly(CoIITAPc) coated on GCE, as depicted in Fig. 5. The enhanced sensibility of poly(CoIITAPc)/MWCNTs-Nafion modified GCE suggests the possibility of sensitive determination of sulfadiazine at trace levels.

Fig. 4.

Fig. 4

Cyclic voltammograms of 0.09 mmol/L of sulfadiazine in sulfuric acid solution (0.1 mol/L) at GCE, MWCNTs-Nafion modified GCE, and poly(CoIITAPc)/MWCNTs-Nafion modified GCE

Fig. 5.

Fig. 5

Mechanism of sulfadiazine oxidation on poly(CoIITAPc)/MWCNTs-Nafion surface

3.3. Effect of pH

The pH value of the support electrolyte solution is regarded as a most important impact factor in sensor performance and its detection of the analyte. The influence of pH on electrocatalytic oxidation of sulfadiazine was studied in the presence of 0.09 mmol/L sulfadiazine. Several kinds of solutions, such as 0.1 mol/L sulfuric acid, 0.1 mol/L hydrochloric acid, 0.2 mol/L sodium acetate buffer, and 0.2 mol/L phosphate buffer, at pH values ranging from 0.7 to 7.0 were used as the supporting electrolytes, respectively. The results showed that the pH value of the buffer medium has an important effect on the electrochemical oxidation of sulfadiazine. The anodic peak currents decreased as pH increased from 0.7 to 5.8. When the pH value of the electrolyte was higher than 7.0, the anodic peak disappeared. Hence, sulfuric acid (0.1 mol/L, pH 0.7) is the optimum electrolyte for the sensor.

3.4. Analytical properties

Since amperometry under stirred conditions has a much higher current sensitivity than cyclic voltammetry, it was chosen for sensitive determination of trace sulfadiazine. Under the optimal detection conditions (electrolyte of 0.1 mol/L sulfuric acid, pH 0.7; amperometry with applied potential of 0.9 V vs. AgCl/Ag (saturated KCl)), the response current change of sensor was plotted against sulfadiazine concentration by amperometric detection (Fig. 6, inset). When an aliquot of sulfadiazine solution was added into 10.0 ml of electrolytes, the oxidation currents increased steeply to reach a stable value in less than 3 s (Fig. 6). The current response increased linearly with sulfadiazine concentration in the range of 0.5 to 43.5 μmol/L. The corresponding regression equation was I=0.091 44+0.040 42C, where I was the current change (μA), C was the sulfadiazine concentration (μmol/L), and the coefficient R was 0.998. The detection limit of 0.17 μmol/L was calculated according to 3S b/m (Hasebe and Osteryoung, 1975), where m was the slope of the linear range of the respective calibration plot, and S b was estimated as the standard deviation of the signals obtained in a blank solution. In Table 1, references on determination of sulfadiazine were listed and compared with this work. Although the detection limit of this work is inferior to HPLC methods, it is the lowest in all electrochemical methods listed in Table 1.

Fig. 6.

Fig. 6

Amperometric current-time curve response obtained for successive addition of 5 (a), 10 (b), 25 (c) and 50 (d) μl of 1.0 mmol/L sulfadiazine in each step

Inset shows the plot of current change versus sulfadiazine concentration

Table 1.

Comparison of different methods for determination of sulfadiazine

Study Method Linear range (μmol/L) Detection limit (μmol/L)
Rao et al., 2000 HPLC (AD) 0.05–50 0.05
Preechaworapun et al., 2006 HPLC (AD) 0.2–400 0.044
Pingarron Carrazon et al., 1987 DPV (GCE) 15–60 5.4
Souza et al., 2008 SWV (BDDE) 8.01–119 2.19
Diaz et al., 1996 DPP (HgE) 2.0–32 4.9
Msagati and Ngila, 2002 SWV (P3MT/GCE) 20–3 200 4.0
Braga et al., 2010 SWV (GCE) 62.7–340 10.9
Campestrini et al., 2010 DPV (BiFE) 3.2–20 2.1
This work Amperometry 0.5–43.5 0.17

HPLC: high performance liquid chromatography; AD: amperometric detector; DPV: differential pulse voltammetry; GCE: glassy carbon electrode; SWV: square wave voltammetry; BDDE: boron doped diamond electrode; DPP: differential pulse polarography; HgE: mercury electrode; P3MT: poly(3-methylthiophene); BiFE: bismuth film electrode

The repeatability of five measurements for solutions of 10 μmol/L sulfadiazine was excellent with a relative standard deviation (RSD) of 3.6%. The proposed poly(CoIITAPc)/MWCNTs-Nafion modified GCE was stored in air under ambient conditions and its sensitivity was checked daily. The current response of poly(CoIITAPc)/MWCNTs-Nafion modified GCE maintained 93% of its initial response after one week, indicating good stability of the new electrode.

According to the interference in real samples, ascorbic acid (AA) widely exists in biological samples. The oxidation potential of AA is about 0.35 V in 0.1 mol/L sulfuric acid, which means AA will be oxidized simultaneously with sulfadiazine at applied potential of 0.9 V during amperometry monitoring. Hence, determination of sulfadiazine with AA coexisting was carried out by background subtraction. The obtained current is 0.27 μA for a solution containing 10 μmol/L AA and 10 μmol/L sulfadiazine when 0.6 V was applied. The observed current was due to the oxidation of AA. However, when 0.9 V was applied to the same solution, the current was 0.78 μA due to the oxidation of AA and sulfadiazine. Subtracting the current obtained for AA (0.27 μA) from the total current gave the current of 0.51 μA for sulfadiazine alone. This current was consistent with the current obtained for the individual addition of 10 μmol/L sulfadiazine in Fig. 6. Potential interference from dopamine and uric acid can also be calibrated by background subtraction due to their low anodic potential (Xiao et al., 2007).

3.5. Analytical application

The practical application of the modified electrode was demonstrated by measuring the concentration of sulfadiazine in real urine samples. The standard addition method was used for the determination of sulfadiazine in urine samples supplied by a healthy donor. The anodic currents of sulfadiazine in real samples were calculated by subtracting background currents (obtained at 0.6 V) from total currents (obtained at 0.9 V). Based on the anodic currents of amperometric measurements and using the calibration plots which are shown in Fig. 6 (inset), the sulfadiazine concentration in spiked urine samples was obtained. The results are given in Table 2. The proposed method showed a better recovery of spiked sulfadiazine in urine samples suggesting that this method could be used for practical applications.

Table 2.

Determination and recovery results of sulfadiazine in urine samples at a modified electrode

Sample cs (μmol/L)
RSD (%) n=5 Recovery (%)
Added Found
1 <LOD
2 5.0 4.4 5.8 88.0
3 20.0 18.6 3.4 92.8

c s: sulfadiazine concentration; RSD: relative standard deviation; LOD: limit of detection

4. Conclusions

In this work, we have demonstrated a highly sensitive electrochemical determination of sulfadiazine using a new poly(CoIITAPc)/MWCNTs-Nafion modified GCE. The attachment of poly(CoIITAPc) on MWCNTs was confirmed by SEM and EDS analyses. The synergistic effects of poly(CoIITAPc) and MWCNTs efficiently catalyzed the electrochemical oxidation of sulfadiazine on electrode surface. The modified electrode showed a fast response, high precision and sensitivity, and could detect sulfadiazine concentrations as low as 0.17 μmol/L in acidic buffer solution. The practical application of the modified electrode was demonstrated by measuring the concentration of sulfadiazine in urine samples.

Footnotes

*

Project supported by the National Natural Science Foundation of China (Nos. 20803067 and 20911140271), the Qianjiang Talents Project of the Technology Office of Zhejiang Province (No. 2009R10029), and the Natural Science Foundation of Zhejiang Province (No. Y5110054), China

References

  • 1.Assil HI, Sheth H, Sporns P. An ELISA for sulfonamide detection using affinity-purified polyclonal antibodies. Food Res Int. 1992;25(5):343–353. doi: 10.1016/0963-9969(92)90109-I. [DOI] [Google Scholar]
  • 2.Braga OC, Campestrini I, Vieira IC, Spinelli A. Sulfadiazine determination in pharmaceuticals by electrochemical reduction on a glassy carbon electrode. J Braz Chem Soc. 2010;21(5):813–820. doi: 10.1590/S0103-50532010000500008. [DOI] [Google Scholar]
  • 3.Campestrini I, Braga OC, Vieira IC, Spinelli A. Application of bismuth-film electrode for cathodic electroanalytical determination of sulfadiazine. Electrochim Acta. 2010;55(17):4970–4975. doi: 10.1016/j.electacta.2010.03.105. [DOI] [Google Scholar]
  • 4.Catalano-Pons C, Bargy S, Schlecht D, Tabone MD, Deschênes G, Bensman A, Ulinski T. Sulfadiazine-induced nephrolithiasis in children. Pediatr Nephrol. 2004;19(8):928–932. doi: 10.1007/s00467-004-1519-8. [DOI] [PubMed] [Google Scholar]
  • 5.Chen RJ, Zhang Y, Wang D, Dai H. Noncovalent sidewall functionalization of single-walled carbon nanotubes for protein immobilization. J Am Chem Soc Vol. 2001;123(16):3838–3839. doi: 10.1021/ja010172b. [DOI] [PubMed] [Google Scholar]
  • 6.Chiavarino B, Crestoni ME, Dimarzio A, Fornarini S. Determination of sulfonamide antibiotics by gas chromatography coupled with atomic emission detection. J Chromatogr B. 1998;706(2):269–277. doi: 10.1016/S0378-4347(97)00568-9. [DOI] [PubMed] [Google Scholar]
  • 7.Conley JM, Symes SJ, Kindelberger SA, Richards S. Rapid liquid chromatography-tandem mass spectrometry method for the determination of a broad mixture of pharmaceuticals in surface water. J Chromatogr A. 2008;1185(2):206–215. doi: 10.1016/j.chroma.2008.01.064. [DOI] [PubMed] [Google Scholar]
  • 8.Diaz TG, Cabanillas AG, Valenzuela MIA, Sallinas F. Polarographic behaviour of sulfadiazine, sulfamerazine, sulfamethazine and their mixtures. Use of partial least squares in the resolution of the non-additive signals of these compounds. Analyst. 1996;121(4):547–552. doi: 10.1039/an9962100547. [DOI] [Google Scholar]
  • 9.Díaz-Cruz MS, de Alda MJL, Barceló D. Environmental behavior and analysis of veterinary and human drugs in soils, sediments and sludge. Trends Anal Chem. 2003;22(6):340–351. doi: 10.1016/S0165-9936(03)00603-4. [DOI] [Google Scholar]
  • 10.Evtugyn GA, Goldfarb OE, Budnikov HC, Ivanov AN, Vinter VG. Amperometric DNA-peroxidase sensor for the detection of pharmaceutical preparations. Sensors. 2005;5(6):364–376. doi: 10.3390/s5060364. [DOI] [Google Scholar]
  • 11.Göbel A, McArdell CS, Suter MJF, Giger W. Trace determination of macrolide and sulfonamide antimicrobials, a human sulfonamide metabolite, and trimethoprim in wastewater using liquid chromatography coupled to electrospray tandem mass spectrometry. Anal Chem. 2004;76(16):4756–4764. doi: 10.1021/ac0496603. [DOI] [PubMed] [Google Scholar]
  • 12.Hartig C, Storm T, Jekel M. Detection and identification of sulfonamide drugs in municipal waste water by liquid chromatography coupled with electrospray ionization tandem mass spectrometry. J Chromatogr A. 1999;854(1-2):163–173. doi: 10.1016/S0021-9673(99)00378-7. [DOI] [PubMed] [Google Scholar]
  • 13.Hasebe K, Osteryoung J. Differential pulse polarographic determination of some carcinogenic nitrosamines. Anal Chem. 1975;47(14):2412–2418. doi: 10.1021/ac60364a002. [DOI] [PubMed] [Google Scholar]
  • 14.Issac S, Kumar KG. Voltammetric determination of sulfamethoxazole at a multiwalled carbon nanotube modified glassy carbon sensor and its application studies. Drug Test Anal. 2009;1(7):350–354. doi: 10.1002/dta.69. [DOI] [PubMed] [Google Scholar]
  • 15.Joseph R, Kumar KG. Differential pulse voltammetric determination and catalytic oxidation of sulfamethoxazole using [5,10,15,20-tetrakis(3-methoxy-4-hydroxy phenyl)porphyrinato] Cu(III) modified carbon paste sensor. Drug Test Anal. 2010;2(6):278–283. doi: 10.1002/dta.129. [DOI] [PubMed] [Google Scholar]
  • 16.Kachoosangi RT, Musameh MM, Abu-Yousef I, Yousef JM, Kanan SM, Xiao L, Davies SG, Russell A, Compton RG. Carbon nanotube-ionic liquid composite sensors and biosensors. Anal Chem. 2009;81(1):435–442. doi: 10.1021/ac801853r. [DOI] [PubMed] [Google Scholar]
  • 17.Li H, Guarr TF. Formation of electrohically conductive thin films of metal phthalocyanines via electropolymerization. J Chem Soc Chem Commun. 1989;(13):832–834. doi: 10.1039/c39890000832. [DOI] [Google Scholar]
  • 18.Li J, Lin X. Electrocatalytic oxidation of hydrazine and hydroxylamine at gold nanoparticle-polypyrrole nanowire modified glassy carbon electrode. Sensors Actuat B-Chem. 2007;126(2):527–535. doi: 10.1016/j.snb.2007.03.044. [DOI] [Google Scholar]
  • 19.Liu J, Rinzler AG, Dai H, Hafner JH, Bradley RK, Boul PJ, Lu A, Iverson T, Shelimov K, Huffman CB, et al. Fullerene pipes. Science. 1998;280(5367):1253–1256. doi: 10.1126/science.280.5367.1253. [DOI] [PubMed] [Google Scholar]
  • 20.Malintan NT, Mohd MA. Determination of sulfonamides in selected Malaysian swine wastewater by high-performance liquid chromatography. J Chromatogr A. 2006;1127(1-2):154–160. doi: 10.1016/j.chroma.2006.06.005. [DOI] [PubMed] [Google Scholar]
  • 21.Maudens KE, Zhang GF, Lambert WE. Quantitative analysis of twelve sulfonamides in honey after acidic hydrolysis by high-performance liquid chromatography with post-column derivatization and fluorescence detection. J Chromatogr A. 2004;1047(1):85–92. doi: 10.1016/j.chroma.2004.07.007. [DOI] [PubMed] [Google Scholar]
  • 22.Msagati TAM, Ngila JC. Voltammetric detection of sulfonamides at a poly(3-mehtylthiophene) electrode. Talanta. 2002;58(3):605–610. doi: 10.1016/S0039-9140(02)00327-2. [DOI] [PubMed] [Google Scholar]
  • 23.Pang GF, Cao YZ, Fan CL, Zhang JJ, Li XM, Li ZY, Jia GQ. Liquid chromatography-fluorescence detection for simultaneous analysis of sulfonamide residues in honey. Anal Bioanal Chem. 2003;376(4):534–541. doi: 10.1007/s00216-003-1883-4. [DOI] [PubMed] [Google Scholar]
  • 24.Pingarron Carrazon JM, Corona Corona P, Polo Diez LM. Electroanalytical study of sulphadiazine at solid electrodes. Determination in pharmaceutical preparations. Electrochim Acta. 1987;32(11):1573–1575. doi: 10.1016/0013-4686(87)90006-5. [DOI] [Google Scholar]
  • 25.Preechaworapun A, Chuanuwatanakul S, Einaga Y, Grudpan K, Motomizu S, Chailapakul O. Electroanalysis of sulfonamides by flow injection system/high-performance liquid chromatography coupled with amperometric detection using boron-doped diamond electrode. Talanta. 2006;68(5):1726–1731. doi: 10.1016/j.talanta.2005.08.040. [DOI] [PubMed] [Google Scholar]
  • 26.Rao TN, Sarada BV, Tryk DA, Fujishima A. Electroanalytical study of sulfa drugs at diamond electrodes and their determination by HPLC with amperometric detection. J Electroanal Chem. 2000;491(1-2):175–181. doi: 10.1016/S0022-0728(00)00208-4. [DOI] [Google Scholar]
  • 27.Shaidarova LG, Gedmina AV, Chelnokova IA, Budnikov GK. Electrocatalytic oxidation of hydroquinone and pyrocatechol at an electrode modified with a polyvinyl pyridine film with electrodeposited rhodium and its use in the analysis of pharmaceuticals. J Anal Chem. 2004;59(11):1025–1031. doi: 10.1023/B:JANC.0000047003.02044.6b. [DOI] [Google Scholar]
  • 28.Shaidarova LG, Gedmina AV, Chelnokova IA, Budnikov GK. Electrocatalytic oxidation and flow-injection determination of ascorbic acid at a graphite electrode modified with a polyaniline film containing electrodeposited palladium. J Anal Chem. 2006;61(6):601–608. doi: 10.1134/S1061934806060165. [DOI] [Google Scholar]
  • 29.Shaidarova LG, Gedmina AV, Chelnokova IA, Budnikov GK. Determination of tetracycline antibiotics using the electrocatalytic response of an electrode modified by a mixed-valence ruthenium oxide-ruthenium cyanide film. Pharm Chem J. 2008;42(9):545–549. doi: 10.1007/s11094-009-0165-7. [DOI] [Google Scholar]
  • 30.Shaidarova LG, Gedmina AV, Chelnokova IA, Budnikov GK. Selective determination of paracetamol and acetylsalicylic acid on electrode modified with a mixed-valent film of ruthenium oxide-ruthenium cyanide. Russ J Appl Chem. 2011;84(4):620–627. doi: 10.1134/S1070427211040112. [DOI] [Google Scholar]
  • 31.Shelver WL, Shappell NW, Franek M, Rubio FR. ELISA for sulfonamides and its application for screening in water contamination. J Agric Food Chem. 2008;56(15):6609–6615. doi: 10.1021/jf800657u. [DOI] [PubMed] [Google Scholar]
  • 32.Souza CD, Braga OC, Vieira IC, Spinelli A. Electroanalytical determination of sulfadiazine and sulfamethoxazole in pharmaceuticals using a boron-doped diamond electrode. Sensors Actuat B. 2008;135(1):66–73. doi: 10.1016/j.snb.2008.07.020. [DOI] [Google Scholar]
  • 33.Tsai YC, Chiu CC. Amperometric biosensors based on multiwalled carbon nanotube-Nafion-tyrosinase nanobiocomposites for the determination of phenolic compounds. Sensor Actuat B. 2007;125(1):10–16. doi: 10.1016/j.snb.2007.01.032. [DOI] [Google Scholar]
  • 34.Tse YH, Janda P, Lam H. Electrode with electropolymerized tetraaminophthalocyanatocobalt(II) for detection of sulfide ion. Anal Chem. 1995;67(5):981–985. doi: 10.1021/ac00101a028. [DOI] [Google Scholar]
  • 35.Wang J, Golden T, Li R. Cobalt phthalocyanine/cellulose acetate chemically modified electrodes for electrochemical detection in flowing streams. Multifunctional operation based upon the coupling of electrocatalysis and permselectivity. Anal Chem. 1988;60(15):1642–1645. doi: 10.1021/ac00166a038. [DOI] [PubMed] [Google Scholar]
  • 36.Xiao Y, Guo C, Li CM, Li Y, Zhang J, Xue R, Zhang S. Highly sensitive and selective method to detect dopamine in the presence of ascorbic acid by a new polymeric composite film. Anal Biochem. 2007;371(2):229–237. doi: 10.1016/j.ab.2007.07.025. [DOI] [PubMed] [Google Scholar]
  • 37.Zhang Z, Liu JF, Shao B, Jiang GB. Time-resolved fluoroimmunoassay as an advantageous approach for highly efficient determination of sulfonamides in environmental waters. Environ Sci Technol. 2010;44(3):1030–1035. doi: 10.1021/es903328b. [DOI] [PubMed] [Google Scholar]
  • 38.Zhou XH, Xi FN, Zhang YM, Lin XF. Reagentless biosensor based on layer-by-layer assembly of functional multiwall carbon nanotubes and enzyme-mediator biocomposite. J Zhejiang Univ-Sci B (Biomed & Biotechnol) 2011;12(6):468–476. doi: 10.1631/jzus.B1000280. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Zhejiang University. Science. B are provided here courtesy of Zhejiang University Press