Summary
Intracranial aneurysm is a fairly common (often asymptomatic) condition. Subarachnoid hemorrhage associated with aneurysmal rupture is a potentially lethal event with a mortality rate as high as 50 percent and a high rate of disability among those who survive the initial hemorrhage, such that recently published guidelines support treatment of intracerebral aneurysms. The current treatment options include surgical clipping and endovascular treatment, but these are not without significant problems. Despite the trend toward endovascular treatment the rate of recurrence and complications is high. Current published evidence of the use of covered stent is limited to stents covered with polytetrafluoroethylene. It is now recognized that mammalian extracellular matrix represents an excellent scaffold material suitable for many therapeutic applications and glutaraldehyde treated pericardium has been widely used for many years due to its desirable features such as low immunogenicity and durability. This report describes the first published experience with the Aneugraft Pericardium Covered Stent (ITGI Medical, OR Akiva, Israel) in the treatment of internal carotid and vertebral artery aneurysms in three patients. In all three cases, the implantation of this novel device has resulted in successful closure of aneurysms.
Key words: covered stent, pericardium covered stent, aneurysm, vertebral artery, internal carotid artery
Introduction
Intracranial aneurysm is a fairly common condition that is often asymptomatic until the time of rupture. Subarachnoid hemorrhage associated with aneurysmal rupture is a potentially lethal event with a mortality rate as high as 50 percent. Many patients who survive the initial hemorrhage have permanent disability 1. Some studies have estimated the prevalence as high as 6% 2. Recently published guidelines support treatment of intracerebral aneurysms 3.
The current treatment options include surgical clipping and endovascular treatment, but these are not without significant problems 4. For instance, a randomized, multicentre trial compared the safety and efficacy of endovascular coiling with standard neurosurgical clipping for aneurysms found that the outcome in terms of survival free of disability at 1 year is significantly better with endovascular coiling 5. In addition, neurosurgery is associated with significantly longer length of stay and significantly higher total hospital charges 6.
Despite the trend toward endovascular treatment the rate of recurrence and complications is high. Recurrence is seven times more likely with coils than clipping, aneurysm sac perfusion persists in completely occluded aneurysms (even when initial or follow-up angiography shows closure) 7,8. Bleeding from incompletely coiled aneurysms is a well-documented threat and rerupture frequently occurs 9-12.
The principal goal of the flow divertor is placement in the parent artery in order to reconstruct the vessel wall 13. The major complications with flow divertors have been found to be perforator artery stroke, aneurysm re-rupture, and in-stent stenosis and thrombosis 14,15.
Covered stents have also shown better closure and shorter procedure time in clinical investigations 16. Current published evidence of the use of covered stent is limited to stents covered with polytetrafluoroethylene (PTFE) 17. Synthetic materials are used in biological settings with limited success. Studies have shown that polyester covered stent graft with 50% re-stenosis and e-PTFE covered stent with 24% re-stenosis in a sheep iliac model 18. It has also been shown that PTFE retards endothelization and that Dacron is prone to infection, due to adherence to and survival of bacteria on its rough surface 19. Other materials are also used as coverings in the context of medical devices. Glutaraldehyde-treated pericardium has been widely used for many years due to its desirable features such as low immunogenicity and durability 20-25. It has been shown that there is significantly less inflammatory cytokine, significantly less antibody response and inflammatory response compared to un-crosslinked decellularized pericardium 26. It is now recognized that mammalian extracellular matrix represents an excellent scaffold material suitable for many therapeutic applications 27. In Neurosurgery, serous sheets are used as dural substitute. An investigation involving 200 patients undergoing a surgical procedure with the application of horse pericardium as a dural prosthesis found that they are free from antigenic effects and do not produce any toxic catabolites 28. The pericardium proved to be resistant to surgical suture, impermeable to cerebrospinal fluid, transparent and does not cause any clinical evidence or radiological artifacts. Pericardium has also shown decreased intraoperative suture line bleeding compared to Dacron 29.
A CE marked coronary stent covered with pericardium (Aneugraft: ITGI Medical limited, Or Akiva, Israel) is available, which so far shows promising clinical results. The pericardium covered stent (PCS) is a percutaneous implantable device consisting of a 316 L stainless steel stent covered by a 100 μ thick equine pericardium cylinder which makes this device flexible and trackable. It is available in diameters of 2.5 mm, 3.0 mm, 3.5 mm, and 4.0 mm, and lengths of 13, 18, 23 and 27 mm. The stent is mounted on a balloon catheter. The device can be seen in Figure 1. It has proved to be safe and effective in two registries 30, and there is also published evidence attesting to this in other indications 31-43. This report describes the first published experience with the PCS in the treatment of internal carotid and vertebral artery aneurysms in three patients.
Figure 1.
Configuration of a novel pericardium-covered stent (PCS).
Case 1
A 52-year-old woman with a pulsatile palpable mass in the left retromandibular space was referred to our hospital. Computed tomography angiography (CTA) revealed a giant false aneurysm of the left cervical segment of the internal carotid artery (ICA) that was probably due to arterial injury caused by an elongated styloid process. CTA revealed significant elongation and tortuosity of the left and right proximal ICA and a large supra-ophthalmic aneurysm of the right ICA (Figure 2). A four-step, multidisciplinary therapeutic plan combining surgical and endovascular modalities was selected: (i) resection and straightening of proximal tortuosity of the right ICA; (ii) endovascular coiling of intracranial aneurysms (Figure 3); (iii) resection and straightening of the proximal left ICA; and (iv) endovascular treatment of the false aneurysms in the left retromandibular space using PCS. Before implantation of the PCS, the patient was given 100 mg aspirin and 300 mg clopidogrel. Right femoral access was used and digital subtraction angiography (DSA) of the left carotid artery confirmed the findings of the CTA (Figure 4A,B). A 6-F guiding catheter (Guider Softip™ XF, Boston Scientific Corp., Fremont, CA, USA) was advanced in the distal part of the common carotid artery (CCA). A 0.014 guidewire (Synchro, Boston Scientific Corp., Fremont, CA, USA) was then passed distal to the neck of the aneurysm. The length of the aneurysm neck necessitated two PCS, and resulted in complete exclusion of the aneurysm demonstrated on post-procedure angiography (Figure 4C).
Figure 2.
CTA finding of a styloid process causing a false aneurysm of the left ICA, elongation of both proximal parts of the ICA and a large aneurysm of the supra-ophthalmic part of the right ICA (A,B).
Figure 3.
DSA before coiling of the aneurysm of the supra-ophthalmic part of the right ICA (A) and after endovascular treatment (B).
Figure 4.
CTA (A) and DSA (B) of a giant false aneurysm of the cervical segment of the left ICA and post-procedure angiography after placement of a covered stent (C).
Case 2
A 44-year-old woman was referred to our hospital after suffering a subarachnoid hemorrhage. Coiling of a small aneurysm of the communicating segment of the left ICA had been done. A giant (20 ↼ 18 mm), large-neck aneurysm was discovered on CTA at the intradural fourth segment of the left vertebral artery (VA) proximal to the posterior inferior cerebellar artery (PICA). Endovascular treatment was considered to be first-line treatment for this VA aneurysm. The patient received 100 mg aspirin and 75 mg clopidogrel for three days before the procedure. Right femoral access was used and a 6-F guiding catheter (Neuron, Penumbra Inc, San Leandro, California, USA) advanced to the V3 segment of the left VA. DSA showed a giant, large-neck aneurysm of the V4 segment of the VA (Figure 5A). After passage of a 0.014 guidewire (Synchro, Boston Scientific Corp., Fremont, CA, USA) distal to the aneurysm neck, a 4 ↼ 27 mm PCS was deployed and inflated to 12 atm. The aneurysm was completely excluded and this was demonstrated at control angiography (Figure 5B). Follow-up CTA at three months demonstrated complete exclusion and shrinkage of the aneurysm to 18 ↼ 16 mm (Figure 6).
Figure 5.
Preoperative DSA of a giant aneurysm of the V4 segment of the left VA (A) and DSA after placement of a covered stent with no filling of the aneurysm (B).
Figure 6.
Three-month follow-up CTA showing aneurysm exclusion, a patent covered stent with no intimal hyperplasia, and aneurysm shrinkage.
Case 3
An 85-year-old woman was admitted to our hospital for endovascular therapy of a symptomatic large-neck aneurysm of the cervical segment of the ICA subsequent to a stroke in the left middle cerebral artery (MCA). The patient was pre-medicated with 100 mg aspirin and 75 mg clopidogrel for three days before the procedure. Endovascular treatment was undertaken after gaining access via the femoral artery and placement of a 6-F guiding sheath (Guider Softip™ XF, Boston Scientific Corp., Fremont, CA, USA) in the CCA. DSA confirmed the CTA findings of an aneurysm of the cervical segment of the ICA (Figure 7A,B). A 0.014 guidewire (Synchro, Boston Scientific Corp., Fremont, CA, USA) was passed distal to the aneurysm and a PCS (4 ↼ 27 mm) advanced over the wire and placed in the optimal position. The balloon was slowly inflated to 10 atm and the PCS successfully deployed. Control DSA confirmed complete exclusion of the aneurysm with preservation of ICA patency (Figure 7C).
Figure 7.
CTA (A) and DSA (B) before endovascular treatment of an aneurysm of the left cervical segment of the ICA and final angiogram after implantation of a covered stent (C).
Discussion
Different endovascular techniques have been used in the treatment of wide neck intracranial and extracranial carotid and vertebral aneurysms. The use of covered stents represents one of the treatment options. A covered stent, the Jostent Graftmaster (Abbott Laboratories, Abbott Park, ILL, USA), has been used for aneurysms up to the ophthalmic artery or up to the vertebrobasilar junction 44,45. The Graftmaster has also been used for carotido-cavernous fistulas and dissections 46. Other types of covered stent have also been used for the treatment of internal carotid and vertebral aneurysms: Fluency stent (Bard Peripheral Vascular, Tempe, AZ, USA), Gore Viabahn (W.L. Gore & Assoc, Newark, DE, USA), iCast (Atrium Medical, Hudson, NH) and Wallgraft (Boston Scientific, MA, USA) 47,48.
Covered stents have shown better closure and shorter procedure time in clinical investigations: 89 patients with cranial internal carotid artery aneurysms were treated nonrandomly with covered stents (n = 43) or coil embolization (n = 46). The initial angiographic results showed 80.9% complete occlusion in covered stents and 52.2% in coil patients (P = 0.004). The six month angiographic results indicated 95.1% and 48.9% respectively (P < 0.001). In addition, the average procedure time was shorter in the covered stents group 49. Stentgraft has also been used for the treatment of ruptured fusiform aneurysm of intracranial vertebral artery and vertebral artery dissecting aneurysms 50,51. There is published evidence of good clinical results of the use of pericardium covered stent in degenerated saphenous vein grafts 30,36,37,42, coronary artery perforations 38,39, coronary artery fistulae 40, hypertrophic obstructive cardiomyopathy 41, aneurysm 32-34,37,39,42,43, acute myocardial infarction 35,52 and iatrogenic radial arteriovenous fistula 31.
The primary drawback in the use of this technology is the potential for blockage of side branches. This is something that the operator must consider prior to treatment. The other drawback with this technology is the lack of published evidence. As mentioned above, it is reasonable to expect that the biological tissue is superior to synthetic tissue, but the absence of published formal clinical investigations (it has even been performed as well as autologous tissue in some animal models) 53,54, may preclude the widespread use of this device. The potential advantage is that whilst slightly bulkier than a bare metal stent, it is thin and flexible and offers better deliverability than other covered stents 55.
In conclusion, the Aneugraft configuration allows for a single stent, single layer cover making for a trackable and flexibile device. This also allows, complete occlusion of the aneurysm, fistula or dissection in one action. We describe the first cases of use of a PCS in the treatment of wide-neck aneurysms of the ICA and VA. More detailed clinical studies will need to be conducted to confirm the overall performance and long-term effect of the PCS in the treatment of extracranial and intracranial aneurysms of the internal carotid artery or vertebral artery.
References
- 1.Vega C, Kwoon JV, Lavine SD. Intracranial aneurysms: current evidence and clinical practice. Am Fam Physician. 2002;66:601–608. [PubMed] [Google Scholar]
- 2.Rinkel GJ, Djibuti M, Algra A, et al. Prevalence and risk of rupture of intracranial aneurysms: a systematic review. Stroke. 1998;29:251–256. doi: 10.1161/01.str.29.1.251. [DOI] [PubMed] [Google Scholar]
- 3.Komotar RJ, Mocco J, Solomon RA. Guidelines for the surgical treatment of unruptured intracranial aneurysms. Neurosurgery. 2008;62:183–194. doi: 10.1227/01.NEU.0000311076.64109.2E. [DOI] [PubMed] [Google Scholar]
- 4.Rinkel GJ. Natural history, epidemiology and screening of unruptured intracranial aneurysms. J Neuroradiol. 2008;35:99–103. doi: 10.1016/j.neurad.2007.11.004. [DOI] [PubMed] [Google Scholar]
- 5.Molyneux A, Kerr R, Stratton I, et al. International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet. 2002;360:1267–1274. doi: 10.1016/s0140-6736(02)11314-6. [DOI] [PubMed] [Google Scholar]
- 6.Hoh BL, Chi YY, Lawson MF, et al. Length of stay and total hospital charges of clipping versus coiling for ruptured and unruptured adult cerebral aneurysms in the Nationwide Inpatient Sample database 2002 to 2006. Stroke. 2010;41:337–342. doi: 10.1161/STROKEAHA.109.569269. [DOI] [PubMed] [Google Scholar]
- 7.Campi A, Ramzi N, Molyneux AJ, et al. Retreatment of ruptured cerebral aneurysms in patients randomized by coiling or clipping in the International Subarachnoid Aneurysm Trial (ISAT) Stroke. 2007;38:1538–1544. doi: 10.1161/STROKEAHA.106.466987. [DOI] [PubMed] [Google Scholar]
- 8.Zhou B, Li MH, Wang W, et al. Three dimensional volume-rendering technique in the angiographic follow-up of intracranial aneurysms embolized with coils. J Neurosurg. 2010;112:674–680. doi: 10.3171/2009.7.JNS081718. [DOI] [PubMed] [Google Scholar]
- 9.Sluzewski M, van Rooij WJ, Rinkel GJ, et al. Endovascular treatment of ruptured intracranial aneurysms with detachable coils: long-term clinical and serial angiographic results. Radiology. 2003;227:720–724. doi: 10.1148/radiol.2273020656. [DOI] [PubMed] [Google Scholar]
- 10.Byrne JV, Sohn MJ, Molyneux AJ, et al. Five-year experience in using coil embolization for ruptured intracranial aneurysms: Outcomes and incidence of late rebleeding. J Neurosurg. 1999;90:656–663. doi: 10.3171/jns.1999.90.4.0656. [DOI] [PubMed] [Google Scholar]
- 11.Henkes H, Fischer S, Liebig T, et al. Repeated endovascular coil occlusion in 350 of 2759 intracranial aneurysms: safety and effectiveness aspects. Neurosurgery. 2006;58:224–232. doi: 10.1227/01.NEU.0000194831.54183.3F. [DOI] [PubMed] [Google Scholar]
- 12.Johnston SC, Dowd CF, Higashida RT, et al. Investigators CARAT. Predictors of rehemorrhage after treatment of ruptured intracranial aneurysms: the Cerebral Aneurysm Rerupture After Treatment (CARAT) study. Stroke. 2008;39:120–125. doi: 10.1161/STROKEAHA.107.495747. [DOI] [PubMed] [Google Scholar]
- 13.Wong GK, Kwan MC, Ng RY, et al. Flow diverters for treatment of intracranial aneurysms: Current status and ongoing clinical trials. J Clin Neurosci. 2011;18:737–740. doi: 10.1016/j.jocn.2010.10.011. [DOI] [PubMed] [Google Scholar]
- 14.Walcott BP, Pisapia JM, Nahed BV, et al. Early experience with flow diverting endoluminal stents for the treatment of intracranial aneurysms. J Clin Neurosci. 2011;18:891–894. doi: 10.1016/j.jocn.2011.01.002. [DOI] [PubMed] [Google Scholar]
- 15.Byrne JV, Beltechi R, Yarnold JA, et al. Early experience in the treatment of intra-cranial aneurysms by endovascular flow diversion: a multicentre prospective study. PLoS One. 2010;5:e12492. doi: 10.1371/journal.pone.0012492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Li MH, Leng B, Li YD, et al. Comparative study of covered stent with coil embolization in the treatment of cranial internal carotid artery aneurysm: a nonrandomized prospective trial. Eur Radiol. 2010;20:2732–2739. doi: 10.1007/s00330-010-1854-z. [DOI] [PubMed] [Google Scholar]
- 17.Tan HQ, Li MH, Zhang PL, et al. Reconstructive endovascular treatment of intracranial aneurysms with the Willis covered stent: medium-term clinical and angiographic follow-up. J Neurosurg. 2011;114:1014–1020. doi: 10.3171/2010.9.JNS10373. [DOI] [PubMed] [Google Scholar]
- 18.Cejna M, Virmani R, Jones R. Biocompatibility and Performance of the Wallstent and Several Covered Stents in a Sheep Iliac Artery Model. J Vasc Interv Radiol. 2001;12:351–358. doi: 10.1016/s1051-0443(07)61916-2. [DOI] [PubMed] [Google Scholar]
- 19.Shintani H. Modification of medical device surface to attain anti-infection. Trends Biomater Artif Organs. 2004;18:1–8. [Google Scholar]
- 20.Butany J, Luk A, Leong SW, et al. A Carpentier-Edwards porcine-valved dacron conduit: at twenty-five years. Int J Cardiol. 2007;117:13–16. doi: 10.1016/j.ijcard.2006.07.063. [DOI] [PubMed] [Google Scholar]
- 21.Jamieson WR, Burr LH, Munro AI, et al. Carpentier-Edwards standard porcine bioprosthesis: a 21-year experience. Ann Thorac Surg. 1998;66:S40–43. doi: 10.1016/s0003-4975(98)01124-2. [DOI] [PubMed] [Google Scholar]
- 22.Cribier A, Eltchaninoff, Tron Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis. J Am Coll Cardiol. 2004;43:698–703. doi: 10.1016/j.jacc.2003.11.026. [DOI] [PubMed] [Google Scholar]
- 23.von Segesser L, Jornod N, Faidutti B. Repeat sternotomy after reconstruction of the pericardial sac with glutaraldehyde-preserved equine pericardium. J Thorac Cardiovasc Surg. 1987;93:616–619. [PubMed] [Google Scholar]
- 24.Imai Y, Takanashi Y, Hoshino S, et al. The equine pericardial valved conduit and current strategies for pulmonary reconstruction. Semin Thorac Cardiovasc Surg. 1995;7:157–161. [PubMed] [Google Scholar]
- 25.Fujiwara H, Sugano T, Someya T. Repair of Postinfarction Ventricular Septal Ruptura with a tailored 3-Dimensional Match. Tex Heart Inst J. 2004;31:69–71. [PMC free article] [PubMed] [Google Scholar]
- 26.Umashankar PR, Arun T, Kumari TV. Short duration glutaraldehyde cross linking of decellularized bovine pericardium improves biological response. J Biomed Mater Res A. 2011;97:311–320. doi: 10.1002/jbm.a.33061. [DOI] [PubMed] [Google Scholar]
- 27.Badylak SF, Freytes DO, Gilbert TW. Extracellular matrix as a biological scaffold material: Structure and function. Acta Biomater. 2009;5:1–13. doi: 10.1016/j.actbio.2008.09.013. [DOI] [PubMed] [Google Scholar]
- 28.Montinaro A, Gianfreda CD, Proto P. Equine pericardium for dural grafts: clinical results in 200 patients. J Neurosurg Sci. 2007;51:17–19. [PubMed] [Google Scholar]
- 29.Marien BJ, Raffetto JD, Seidman CS, et al. Bovine pericardium vs dacron for patch angioplasty after carotid endarterectomy: a prospective randomized study. Arch Surg. 2002;137:785–788. doi: 10.1001/archsurg.137.7.785. [DOI] [PubMed] [Google Scholar]
- 30.Colombo A, Almagor Y, Gaspar J, et al. The pericardium covered stent (PCS) EuroIntervention. 2009;5:394–399. doi: 10.4244/v5i3a61. [DOI] [PubMed] [Google Scholar]
- 31.Summaria F, Romagnoli E, Preziosi P. Percutaneous antegrade transarterial treatment of iatrogenic radial arteriovenous fistula. J Cardiovasc Med (Hagerstown) 2012;13:50–52. doi: 10.2459/JCM.0b013e3283461126. [DOI] [PubMed] [Google Scholar]
- 32.Adlam D, Hutchings D, Channon KM. Optical coherence tomography-guided stenting of a large coronary aneurysm: images at implantation and at 6 months. J Invasive Cardiol. 2011;23:168–169. [PubMed] [Google Scholar]
- 33.Siregar S, Hoseyni Guyomi S, van Herwerden LA. Covered stents in giant coronary artery aneurysm. Eur Heart J. 2010;31:2823. doi: 10.1093/eurheartj/ehq267. [DOI] [PubMed] [Google Scholar]
- 34.Ferlini M, Russo F, Marinoni B, et al. Percutaneous coronary aneurysm obliteration using a novel pericardium-covered stent. J Am Coll Cardiol. 2010;56:2139. doi: 10.1016/j.jacc.2010.04.071. [DOI] [PubMed] [Google Scholar]
- 35.Malkin C, Gunn J. A big cover-up. Heart. 2009;95:1354. doi: 10.1136/hrt.2009.171843. [DOI] [PubMed] [Google Scholar]
- 36.Wykrzykowska JJ, Gutieerez-chico JL, Van Geuns RJ. “Over and under” pericardial covered stent with paclitaxel balloon in a saphenous vein graft. Catheter Cardiovasc Interv. 2010;75:964–966. doi: 10.1002/ccd.22412. [DOI] [PubMed] [Google Scholar]
- 37.Tyczynski P, Kukreja N, Van Geuns RJ, et al. Optical coherence tomography for assessment of treatment with pericardium covered stents of degenerated saphenous vein grafts. Eurointervention. 2010;6:78–85. [PubMed] [Google Scholar]
- 38.Jokhi P, McKenzie D, O’Kane P. Use of a novel pericardial covered stent to seal an iatrogenic coronary perforation. J Invasive Cardiol. 2009;21:187–190. [PubMed] [Google Scholar]
- 39.Hayat SA, Ghani S, Tang A, et al. Percutaneous treatment of a ruptured coronary aneurysm with a novel ‘over and under’ equine pericardium stent. Catheter Cardiovasc Interv. 2009;1(74):367–370. doi: 10.1002/ccd.21937. [DOI] [PubMed] [Google Scholar]
- 40.Heper G, Barcin C Iyisoy, A , et al. Treatment of an iatrogenic left internal mammary artery to pulmonary artery fistula with a bovine pericardium covered stent. Cardiovasc Intervent Radiol. 2006;29:879–882. doi: 10.1007/s00270-005-0149-2. [DOI] [PubMed] [Google Scholar]
- 41.Gaspar J, Martínez-Ríos MA, Vonderwalde C, et al. Pericardium-covered stent for septal myocardial ablation in hypertrophic obstructive cardiomyopathy. Catheter Cardiovasc Interv. 1999;2:241–246. doi: 10.1002/(SICI)1522-726X(199905)47:1<73::AID-CCD17>3.0.CO;2-Z. [DOI] [PubMed] [Google Scholar]
- 42.Gaspar J, Vonderwalde C, Hong MK, et al. [Stent coated with bovine pericardium: in vitro evaluation, in animals, and initial results in humans] [Article in Spanish] Arch Cardiol Mex. 2001;71:286–294. [PubMed] [Google Scholar]
- 43.Gaspar J, Vonderwalde C, Eid-Lidt G. Treatment of coronary artery aneurysms by percutaneous sealing with bovine-pericardium-covered stent. Int J Cardiovasc Interv. 1999;2:241–246. doi: 10.1080/acc.2.4.241.246. [DOI] [PubMed] [Google Scholar]
- 44.Saatci I, Cekirge SH, Ozturk MH, et al. Treatment of internal carotid artery aneurysms with a covered stent: experience in 24 patients with mid-term follow-up results. Am J Neuroradiol. 2004;25:1742–1749. [PMC free article] [PubMed] [Google Scholar]
- 45.Greenberg E, Katz JM, Janardhan V, et al. Treatment of a giant vertebrobasilar artery aneurysm using stent grafts. J Neurosurg. 2007;107:165–168. doi: 10.3171/JNS-07/07/0165. [DOI] [PubMed] [Google Scholar]
- 46.Gomez F, Escobar W, Gomez AM, et al. Treatment of carotid cavernous fistulas using covered stents: midterm results in seven patients. Am J Neuroradiol. 2007;28:1762–1768. doi: 10.3174/ajnr.A0636. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Fischer B, Palkovic S, Wassmann H, et al. Endovascular management of tandem extracranial internal carotid artery aneurysms with a covered stent. J Endovasc Ther. 2004;11:739–741. doi: 10.1583/04-1293R.1. [DOI] [PubMed] [Google Scholar]
- 48.Powell RJ, Rzucidlo EM, Schermerhorn ML. Stent-graft treatment of a large internal carotid artery vein graft aneurysm. J Vasc Surg. 2003;37:1310–1313. doi: 10.1016/s0741-5214(03)00144-7. [DOI] [PubMed] [Google Scholar]
- 49.Li MH, Leng B, Li YD, et al. Comparative study of covered stent with coil embolization in the treatment of cranial internal carotid artery aneurysm: a nonrandomized prospective trial. Eur Radiol. 2010;20:2732–2739. doi: 10.1007/s00330-010-1854-z. [DOI] [PubMed] [Google Scholar]
- 50.Chiaradio JC, Guzman L, Padilla L, et al. Intravascular graft stent treatment of a ruptured fusiform dissecting aneurysm of the intracranial vertebral artery: technical case report. Neurosurgery. 2002;50:213–217. doi: 10.1097/00006123-200201000-00034. [DOI] [PubMed] [Google Scholar]
- 51.He M, Zhang H, Lei D, et al. Application of covered stent grafts for intracranial vertebral artery dissecting aneurysms. J Neurosurg. 2009;110:418–426. doi: 10.3171/2008.3.17470. [DOI] [PubMed] [Google Scholar]
- 52.Gunn J, Siotia A, Malkin CJ, et al. Novel use of a pericardium-covered stent graft to treat bulky coronary artery thrombus. Catheter Cardiovasc Interv. 2011 doi: 10.1002/ccd.23184. Dec 8 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 53.Romaguera R, Waksman R. Covered stents for coronary perforations: is there enough evidence? Catheter Cardiovasc Interv. 2011;78:246–253. doi: 10.1002/ccd.23017. [DOI] [PubMed] [Google Scholar]
- 54.DeBacker CM, Dutton JJ, Proia AD, et al. A comparative study of bovine pericardium (periguard) and homologous sclera as lower eyelid spacer graft analogs in New Zealand white rabbits. Ophthal Plast Reconstr Surg. 2000;16:156–161. doi: 10.1097/00002341-200003000-00011. [DOI] [PubMed] [Google Scholar]
- 55.DeBacker CM, Dutton JJ, Proia AD, et al. Bovine pericardium versus homologous sclera as wrapping materials for hydroxyapatite ocular implants: an animal study. Ophthal Plast Reconstr Surg. 1999;15:312–316. doi: 10.1097/00002341-199909000-00003. [DOI] [PubMed] [Google Scholar]