Skip to main content
Journal of Diabetes Science and Technology logoLink to Journal of Diabetes Science and Technology
letter
. 2012 Mar 1;6(2):479–480. doi: 10.1177/193229681200600239

Minimally Invasive Enzyme Microprobes: An Alternative Approach for Continuous Glucose Monitoring

Anna Radomska Bothelo Moniz 1, Kostis Michelakis 1, Jakub Trzebinski 1,2, Sanjiv Sharma 1, Desmond G Johnston 3, Nick Oliver 3, Anthony Cass 1
PMCID: PMC3380796  PMID: 22538163

Continuous glucose monitoring has been shown to improve glycemia in diabetes.1,2 To minimize the pain, inconvenience, and discomfort associated with long-term implanted sensors, minimally invasive technologies are being developed, including microneedles that remove small volumes of subcutaneous interstitial fluid (ISF).3 We describe here an alternative approach for continuous monitoring of glucose in ISF that involves using solid microprobes that access the interstitium in a minimally invasive and less painful manner than using a lancet to draw capillary blood. The glucose sensor is a small (∼1 cm2) wearable patch covered with arrays of solid microprobes, which penetrate the stratum corneum and monitor glucose in ISF within the epidermis. Unlike microneedles, solid microprobes are not susceptible to clogging and their minimally invasive nature may be less traumatic than existing subcutaneous sensors, reducing the risk of sensor drift and protein adsorption that degrades performance.4,5

The microprobes were made by photolithography using backside illumination6 adapted to create 500–750 mm long structures with tip diameters of 10 mm. Sensitivity to glucose was achieved by assembling glucose oxidase (GOx; EC 1.1.3.4) on the gold-coated microprobes using 1-Ethyl-3-(3-dimethylaminopropyl) carbodiimide/N-hydroxysuccinimide coupling followed by a Nafion membrane incorporating tetrathiafulvalene (TTF) as an electron transfer mediator. In vitro studies with these devices have demonstrated that they perform well in the clinically relevant range (2–20 mM; Figure 1), with a linear response to increasing glucose and measured currents in the microampere domain.

Figure 1.

Figure 1

(A) Scanning electron microscope image showing 600 μm high microprobe array. (B) Calibration of TTF/GOx-modified gold microprobes performed in 0.1 M phosphate-buffered saline pH 7.4 background solution; each step represents an increase of 2 mM in glucose concentration.

Data are presented for glucose measurement, but the platform may be employed with other enzymes such as lactate oxidase or beta-hydroxybutyrate dehydrogenase. It is also possible to partition the microprobe array and immobilize different enzymes across multiple subarrays for continuous multianalyte sensing. Partitioning may also allow multiple sensors to be used in a “voting” system for the same analyte, thus improving sensor reliability.

Glossary

Abbreviations

(GOx)

glucose oxidase

(ISF)

interstitial fluid

(TTF)

tetrathiafulvalene

Funding

This work was funded by the Imperial College Biomedical Research Committee Fund and Imperial Innovations.

References

  • 1.Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Tamborlane WV, Beck RW, Bode BW, Buckingham B, Chase HP, Clemons R, Fiallo-Scharer R, Fox LA, Gilliam LK, Hirsch IB, Huang ES, Kollman C, Kowalski AJ, Laffel L, Lawrence JM, Lee J, Mauras N, O’Grady M, Ruedy KJ, Tansey M, Tsalikian E, Weinzimer S, Wilson DM, Wolpert H, Wysocki T, Xing D. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008;359(14):1464–1476. doi: 10.1056/NEJMoa0805017. [DOI] [PubMed] [Google Scholar]
  • 2.Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Beck RW, Hirsch IB, Laffel L, Tamborlane WV, Bode BW, Buckingham B, Chase P, Clemons R, Fiallo-Scharer R, Fox LA, Gilliam LK, Huang ES, Kollman C, Kowalski AJ, Lawrence JM, Lee J, Mauras N, O’Grady M, Ruedy KJ, Tansey M, Tsalikian E, Weinzimer SA, Wilson DM, Wolpert H, Wysocki T, Xing D. The effect of continuous glucose monitoring in well-controlled type 1 diabetes. Diabetes Care. 2009;32(8):1378–1383. doi: 10.2337/dc09-0108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Mukerjee EV, Collins SD, Isseroff RR, Smith RL. Microneedle array for transdermal biological fluid extraction and in situ analysis. Sens Actuat A Phys. 2004;114(2-3):267–275. [Google Scholar]
  • 4.Noh YW, Kim TH, Baek JS, Park HH, Lee SS, Han M, Shin SC, Cho CW. In vitro characterization of the invasiveness of polymer microneedle against skin. Int J Pharm. 2010;397(1-2):201–205. doi: 10.1016/j.ijpharm.2010.06.050. [DOI] [PubMed] [Google Scholar]
  • 5.Oliver NS, Toumazou C, Cass AE, Johnston DG. Glucose sensors: a review of current and emerging technology. Diabet Med. 2009;26(3):197–210. doi: 10.1111/j.1464-5491.2008.02642.x. [DOI] [PubMed] [Google Scholar]
  • 6.Kim K, Park DS, Lu HM, Che W, Kim K, Lee JB, Ahn CH. A tapered hollow metallic microneedle array using backside exposure of su-8. J Micromech Microeng. 2004;14(4):597–603. [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES