. 2012 Jun 29;39(7):4467–4482. doi: 10.1118/1.4730287

Characterization of imaging performance in differential phase contrast CT compared with the conventional CT: Spectrum of noise equivalent quanta NEQ(k)

Xiangyang Tang 1,a), Yi Yang 1, Shaojie Tang 1
PMCID: PMC3396709  PMID: 22830779

Abstract

Purpose: Differential phase contrast CT (DPC-CT) is emerging as a new technology to improve the contrast sensitivity of conventional attenuation-based CT. The noise equivalent quanta as a function over spatial frequency, i.e., the spectrum of noise equivalent quanta NEQ(k), is a decisive indicator of the signal and noise transfer properties of an imaging system. In this work, we derive the functional form of NEQ(k) in DPC-CT. Via system modeling, analysis, and computer simulation, we evaluate and verify the derived NEQ(k) and compare it with that of the conventional attenuation-based CT.

Methods: The DPC-CT is implemented with x-ray tube and gratings. The x-ray propagation and data acquisition are modeled and simulated through Fresnel and Fourier analysis. A monochromatic x-ray source (30 keV) is assumed to exclude any system imperfection and interference caused by scatter and beam hardening, while a 360° full scan is carried out in data acquisition to avoid any weighting scheme that may disrupt noise randomness. Adequate upsampling is implemented to simulate the x-ray beam's propagation through the gratings G1 and G2 with periods 8 and 4 μm, respectively, while the intergrating distance is 193.6 mm (1/16 of the Talbot distance). The dimensions of the detector cell for data acquisition are 32 × 32, 64 × 64, 96 × 96, and 128 × 128 μm2, respectively, corresponding to a 40.96 × 40.96 mm2 field of view in data acquisition. An air phantom is employed to obtain the noise power spectrum NPS(k), spectrum of noise equivalent quanta NEQ(k), and detective quantum efficiency DQE(k). A cylindrical water phantom at 5.1 mm diameter and complex refraction coefficient n = 1 − δ + = 1 −2.5604 × 10−7 + i1.2353 × 10−10 is placed in air to measure the edge transfer function, line spread function and then modulation transfer function MTF(k), of both DPC-CT and the conventional attenuation-based CT. The x-ray flux is set at 5 × 106 photon/cm2 per projection and observes the Poisson distribution, which is consistent with that of a micro-CT for preclinical applications. Approximately 360 regions, each at 128 × 128 matrix, are used to calculate the NPS(k) via 2D Fourier transform, in which adequate zero padding is carried out to avoid aliasing in noise.

Results: The preliminary data show that the DPC-CT possesses a signal transfer property [MTF(k)] comparable to that of the conventional attenuation-based CT. Meanwhile, though there exists a radical difference in their noise power spectrum NPS(k) (trait 1/|k| in DPC-CT but |k| in the conventional attenuation-based CT) the NEQ(k) and DQE(k) of DPC-CT and the conventional attenuation-based CT are in principle identical.

Conclusions: Under the framework of ideal observer study, the joint signal and noise transfer property NEQ(k) and detective quantum efficiency DQE(k) of DPC-CT are essentially the same as those of the conventional attenuation-based CT. The findings reported in this paper may provide insightful guidelines on the research, development, and performance optimization of DPC-CT for extensive preclinical and clinical applications in the future.

Keywords: CT, x-ray CT, x-ray phase CT, x-ray differential phase contrast CT, x-ray tube and grating-based phase CT, imaging performance, modulation transfer function, MTF, noise, noise power spectrum, NPS, noise equivalent quanta, NEQ, detective quantum efficiency, DQE.

INTRODUCTION

Motivated to reach higher subject contrast over soft tissues in x-ray CT imaging, the scientific community has devoted increasing effort to the investigation of x-ray tube and grating-based differential phase contrast CT (DPC-CT) for early detection of cancer and other diseases.1, 2, 3, 4, 5 The initial exploration was relatively qualitative and demonstrated significant improvement in the contrast of soft tissues in human specimens or very small animals.1, 2, 4, 5 Recently, the investigation has become quantitative, and preliminary results on the noise property of DPC-CT,6, 7, 8, 9, 10, 11, 12, 13 such as pixel-based gross variance,6, 7, 8, 9 contrast-to-noise ratio,8, 10 and spatial autocovariance function,7 have been reported. It is a fundamental understanding that the subject contrast of soft tissues in an imaging system is intrinsically determined by their interaction with the x-ray beam,14, 15, 16 while the system's performance is determined by its signal and noise transfer properties.11, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35 In principle, the signal transfer property of an imaging system is dependent on its modulation transfer function MTF(k),17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30 while the noise transfer property can only be thoroughly characterized by its noise power spectrum NPS(k), i.e., the variation of noise intensity as a function over spatial frequency k.17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35 It has been reported that there exists a radical difference in the noise power spectrum NPS(k) between DPC-CT and the conventional attenuation-based CT (referred to as conventional CT hereafter): the NPS(k) of the former manifests itself with the trait 1/|k|, and the latter with |k|.11, 12

In general, an imaging system can be cascaded into two stages—image formation (or record or detection as termed in the literature) and image presentation or display.20, 27, 28 Such a separation is straightforward in digital imaging modalities, e.g., the DPC-CT and conventional CT to be investigated in this work, but quite intricate in the early days when only analog imaging modalities, e.g., x-ray screen-film radiography, were available. Quite a few factors may degrade the performance of an imaging system, which include but are not limited to (i) the ubiquitous random fluctuation—noise—in data acquisition, (ii) the anatomic and physiologic variation in patient population, and (iii) the intra- and interobserver variability when a diagnosis is made based on the images presented.27, 28 These factors work in concert with one another and thus make the performance assessment of an imaging system extremely challenging. With recourse to Bayesian statistical decision and information theories, a framework for the performance assessment of an imaging system has been established through the tremendous and successful effort of leading scientists in this field,17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28 in which the detection of pathological lesions with a medical imaging system is mathematically treated as a decision making process. An observer makes the decision between two hypotheses—H1: lesion present (abnormal or positive); H2: lesion absent (normal or negative)—according to whether the value of a decision function exceeds a threshold or not. If the decision function is the Bayesian likelihood, the decision maker becomes an ideal observer, who minimizes the risks while making the decision, i.e., maximizing the area under the receiver operating characteristics (ROC) curve.20, 27, 28

Suppose the imaging system under study is linear and shift-invariant, and the noise corrupting the system is Gaussian and stationary. Given a signal specified in the spatial frequency domain as ΔS(k), i.e., the signal and background known exactly (SKE/BKE),27, 28, 30 the squared signal-to-noise ratio or the detectability index defined by

SNR ideal 2=|ΔS(k)|2 NEQ (k)dk (1)

is an ideal observer figure of merit (FOM) to assess an imaging system's performance,27, 30 wherein the spectrum of noise equivalent quanta NEQ(k) may be in different functional forms over imaging modalities.20, 21, 26, 28 An ideal observer is assumed to have prior knowledge of the task and the statistical properties of the noise. Equation 1 is an integration of factorization and implies that, given a specific task |ΔS(k)|2, the ideal observer performance of an imaging system can be optimized by maximizing its NEQ(k).27, 30 Usually, the task is to differentiate a lesion from the surrounding tissues or organs or simply an object from its background. Therefore, the signal ΔS(k) in Eq. 1 should be perceived as a difference or contrast between the object to be imaged and background.20, 21, 27, 30 In practice, Eq. 1 can be extended to deal with more complicated situations, wherein the noise observes the Poisson distribution and is not stationary, the imaging system is nonlinear and shift-variant, or the signal is superimposed on a random background.20, 27, 28, 30

Recognizing the important role played by NEQ(k) as a FOM to assess the ideal observer performance of an imaging system,20, 27, 28, 30, 32, 33, 34, 35 we investigate the NEQ(k) and resultant spectrum of detective quantum efficiency DQE(k) of DPC-CT in this work. In a way analogous to the investigation of conventional CT's NEQ(k) and DQE(k), we treat the DPC-CT as a linear and shift-invariant system. Owing to the fact that Gaussian noise approaches Poisson noise if the detected number of photons is large, we assume the noise in the DPC-CT is Gaussian.29 To avoid any interference caused by scatter and beam hardening, a monochromatic x-ray source is assumed in both the DPC-CT and conventional CT in this investigation, though they are actually configured with a polychromatic one in practice. Under the framework of an ideal observer, we derive, analyze, evaluate, and verify the NEQ(k) and DQE(k) of DPC-CT through computer simulation studies and compare the result with that of conventional CT. For clarity in expression henceforth, ΔS(k), SNR2ideal, NPS(k), MTF(k), NEQ(k), and DQE(k) denote the quantities corresponding to an imaging system in general, which can be either DPC-CT or conventional CT. Specifically, the quantities corresponding to the DPC-CT are denoted with subscript “p,” while those corresponding to the conventional CT are denoted with subscript “a.”

MATERIALS AND METHODS

To characterize the spectrum of noise equivalent quanta NEQp(k) and spectrum of detective quantum efficiency DQEp(k) of DPC-CT, we start out by deriving its functional form and comparing it with that of the conventional CT. Subsequently, we introduce the methods and procedures to evaluate and verify the derived NEQp(k) and DQEp(k) and compare them with their counterparts in conventional CT. We constrain ourselves in this work to conducting computer simulation study only; thereby the systematic and random errors, such as the fabrication accuracy and alignment of gratings, and the effects due to beam hardening and scattering that may exist in a physical DPC-CT and compromise or bias the accuracy and precision of evaluation and verification can be excluded.

Imaging mechanism of DPC-CT implemented with x-ray tube and gratings

The architecture of DPC-CT implemented with a microfocus x-ray tube and gratings2 is shown in Fig. 1a. G1 is a phase grating and G2 an absorption grating, which can be fabricated with photolithography, deep chemical etching, and electroplating.36G1 and G2 work together as a shearing interferometer37, 38, 39 to detect the wavefront alteration caused by the object in x-ray beam. The x-ray tube irradiates the specimen, while the specimen stage rotates by a range satisfying the data sufficiency condition, and the CCD x-ray detector is employed for data acquisition. The key component of the imaging chain is grating G1, a diffraction interferometer based on the Talbot effect.37, 38, 39 Figure 1b shows how G1 works by virtually decomposing it into gratings A and B. The extra optical path corresponding to grating B relative to that through grating A is half wavelength, which is equivalent to a 180° phase shift. The beams corresponding to gratings A and B recombine after they pass through the gratings, and interference fringe appears if the object to be imaged is in the x-ray beam.

Figure 1.

Figure 1

The diagrams showing (a) the schematic of an x-ray tube and grating-based DPC-CT and (b) the schematic of virtual grating decomposition, in which the upper and lower Talbot patterns are associated with gratings A and B, respectively.

The imaging mechanism of x-ray tube and grating-based DPC-CT is elaborated in Refs. 11, 12, 13 and 37, 38, 39 and only a concise review is given below. According to Fresnel analysis,40, 41 the irradiance I(x, z) at the CCD detector is

IA+B(x,z)ϕx+Δx2,yϕxΔx2,yϕ(x,y)xΔx, (2)

where Δx is a displacement in the x-direction, and ϕ(x, y) is the phase variation over Δx, which is the projection of refractive coeffiecient δ^(x,y,z) along the x-ray path

ϕ(x,y)=2πλZδ^(x,y,z)dz. (3)

Note that the refractive coefficient has been denoted by δ(x, y, z) in the literature. However, to avoid its confusion with the Dirac function and Kronecker delta that are used to derive the functional form of NEQp(k) later in this paper, δ^(x,y,z) is adopted to represent the refractive coeficient.

Equation 2 shows that the irradiance depends on the derivative of the phase variation along the x axis. After the x-ray passes grating G2, the irradiance at detector D is41

Iu,v(x)=a0(u,v)+m=1am(u,v) cos 2πmxg2+φm(u,v), (4)

where (u, v) is the coordinate of a detector cell and g2 the period of grating G2. By linearly shifting grating G2 along the x axis and carrying out a Fourier analysis, one can determine a0(u, v), a1(u, v), and φ1(u, v) from Eq. 4.41 In fact, one has

ϕ(x,y)x=φ1(u,v)g2λzT, (5)

where zT is the fractional Talbot distance.2, 37, 38, 39, 40, 41 Substituting the ϕ(x, y) defined in Eq. 3 into Eq. 5, one gets

φ1(u,v)=λzTg2ϕ(x,y)x=2πzTg2xZδ^(x,y,z)dz=2πzTg2Zxδ^(x,y,z)dz. (6)

This means that the phase retrieved through a Fourier analysis of Eq. 4 is the projection of the refractive coefficient's derivative, and this is the underlying reason that the phase CT implemented with x-ray tube and gratings is called differential phase contrast CT. Once data φ1(u, v) are acquired, tomographic images of refraction are reconstructed using the filtered backprojection (FBP) algorithms.42, 43, 44, 45 Since the reconstruction is carried out directly from δ^(x,y,z)/x, the ramp kernel is replaced with the Hilbert kernel. It should be pointed out that the modeling of data acquisition in the x-ray tube and grating-based DPC-CT through the schematic of Fig. 1b and Eqs. 2, 3, 4, 5, 6 has been evaluated and verified in Ref. 11, which is in principle the same as the five steps detailed in Ref. 13. The well-known contrast-detail (C-D) phantom29 (see Sec. 3) is used to evaluate and verify the correctness and accuracy of the data acquisition modeling and image reconstruction, and the result is presented in Fig. 2.

Figure 2.

Figure 2

Transverse images of the C-D phantom generated by the x-ray tube and grating-based DPC-CT CT (a) and the conventional CT (b) (x-ray exposure 107 photon/cm2 per projection, detector cell 48 × 48 μm2).

Spectrum of noise equivalent quanta in DPC-CT and conventional CT

In the early days of conventional CT, an observation of the morphologic difference in the noise of CT images against the white noise suggested that there existed an interpixel correlation. By analyzing the distribution of noise power over spatial frequency k, i.e., the noise power spectrum NPSa(k), the interpixel correlation is confirmed, even though there is no intercell correlation in the noise of the detector used for data acquisition. Since then, the groundwork of using noise power spectrum NPSa(k) and spectrum of noise equivalent quanta NEQa(k) to analyze the signal and noise behavior of a conventional CT system or CT imaging method has been laid out.17, 18, 19, 20, 21, 22, 23, 24, 25 In this section, we follow the mathematic treatment to derive the functional form of NPSp(k) and NEQp(k) in the DPC-CT implemented with x-ray tube and gratings and compare it with that of the conventional CT.

Spectrum of noise equivalent quanta NEQa(k) of the conventional CT

A number of strategies, e.g., the method proposed by Barret,18 the central slice theorem,19 information theory,20 and statistical detection theory,21 have been exercised to obtain the noise power spectrum NPSa(k) of the conventional CT, and all lead to the same functional form

NPS a(k)=aπNθN¯k MTF a(k)2=πbNθI0k MTF a(k)2, (7)

where a is the detector pitch, b is the detector height, and k is the radial frequency defined as

k=kx2+ky2. (8)

Nθ is the number of angular locations at which the projection data are acquired, I0 is the measured photon flux, and N¯ is the mean number of x-ray photons detected at each detector cell. Since no object should be placed in the x-ray beam in the investigation of noise property, N¯ is assumed equal across all the detector cells. MTFa(k) is the overall algorithmic contribution, including the windowing and/or boosting in the frequency domain for an optimization between noise and spatial resolution, to the modulation transfer function.

The multiplication of Nθ and I0 has been defined as the noise equivalent quanta.20, 21, 23 Accordingly, Eq. 7 can be expressed as

NPS a(k)=πb NEQ ak MTF a(k)2. (9)

Consequently, the spectrum of noise equivalent quanta NEQa(k) and detective quantum efficiency DQEa(k) of the conventional CT are18, 19, 20, 21

NEQ a(k)=πb NPS a(k)k MTF a(k)2, (10)
DQE a(k)= NEQ a(k)NθI0. (11)

Spectrum of noise equivalent quanta NEQp(k) of the DPC-CT

As indicated above, the phase information φ1(u, v) can be retrieved by stepping grating G2 linearly along the x axis (see Fig. 1). At each step

xg=kMg2,k=1,2,...,M, (12)

the measured x-ray irradiance at (u, v) in a detector represented in Eq. 4 can be rewritten as

N(k)u,v=l=M2M21Nl1+δl02 exp iφlu,v+2πilkM, (13)

where Nl is non-negative and real and δl0 is the Kronecker's delta δlm by setting the 2nd subscript m equal to zero, i.e.,

δl0=1l=00l0. (14)

Note that φl(u, v) (l = 1) is the phase we want to retrieve for reconstruction of the 3D distribution of refractive coefficient [see Eq. 6].

The discrete Fourier transform of both sides of Eq. 13 gives

12MN1 exp iφ1u,v=k=1MN(k)u,vexp2πikM. (15)

Since N1 is non-negative and real, Eq. 15 further gives

exp iφ1u,v=k=1MN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikM, (16)

or equivalently

exp iφ1u,v=k=1MN(k)u,v exp 2πikMk=1MN(k)u,vexp2πikMk=1MN(k)u,v exp 2πikM1/2. (17)

Differentiating both sides of Eq. 17, one gets

i exp iφ1u,vΔφ1u,v=k=1MΔN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikMk=1MNku,vexp2πikM1/212k=1MN(k)u,v exp 2πikM2k=1MΔN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikM3/212k=1MN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikMk=1MΔN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikMk=1MN(k)u,v exp 2πikM3/2. (18)

Substituting Eq. 15 into Eq. 18 yields

i exp iφ1u,vΔφ1u,v=k=1MΔN(k)u,v exp 2πikM12MN112 exp 2iφ1u,v22M2N12k=1MΔN(k)u,v exp 2πikM23M3N131222M2N12k=1MΔN(k)u,v exp 2πikM23M3N13, (19)

which can be concisely rewritten as

Δφ1(u,v)=ik=1M exp 2πikMMN1 exp iφ1(u,v) exp 2πikMMN1 exp iφ1(u,v)ΔN(k)(u,v). (20)

Consequently, one has

Δφ1u,vΔφ1u1,v1=k,k1=1M exp 2πikMMN1 exp iφ1u,v exp 2πikMMN1 exp iφ1u,v× exp 2πik1MMN1 exp iφ1u1,v1 exp 2πik1MMN1 exp iφ1u1,v1ΔNku,vΔNk1u1,v1, (21)

where (u1, v1) represents a location that is different from (u, v) in the detector. Since the measured x-ray flux observes the Poisson distribution and is spatially uncorrelated, one has

ΔNk(u,v)ΔN(k1)(u1,v1)=N(k)(u,v)δkk1δuu1aδvv1b, (22)

where subscripts k and k1 correspond to different shifting steps of grating G2. Inserting Eq. 22 into Eq. 21 and substituting N(k)(u, v) with Eq. 13, one obtains

Δφ1u,vΔφ1u1,v1=k=1M exp 2πikMMN1 exp iφ1u,v exp 2πikMMN1 exp iφ1u,v2N(k)u,vδuu1aδvv1b=2M2N12k=1MN(k)u,v12k=1M exp 4πikM exp 2iφ1u,v+ exp 4πikM exp 2iφ1u,vN(k)u,vδuu1aδvv1b=2M2N12MN0M4N2 exp iφ2u,v exp 2iφ1u,v+N2 exp iφ2u,v exp 2iφ1u,vδuu1aδvv1b. (23)

It is interesting to note that only N−2, N−1, N0, N1, and N2 survive the last step in getting Eq. 23.

Subsequently, defining

ɛ1=N¯1N¯0, (24)
ɛ2=N¯24N¯0 exp iφ¯2 exp 2iφ¯1+ exp iφ¯2 exp 2iφ¯1, (25)

Eq. 23 can be concisely rewritten as

Δφ1u,vΔφ1u1,v1=21ɛ2MN¯0ɛ12δuu1aδvv1b. (26)

Note that the random variables N0, N1, and N2 have been replaced with their corresponding mean values N¯0, N¯1 and N¯2, respectively, in a way analogous to that in Ref. 10. Consequently, one gets

σφ2=21ɛ2MN¯0ɛ12. (27)

Meanwhile, according to Eq. 6, it is not hard for one to get

NθσD2=λzTg22Nθσφ2=λzTg22ɛ1221ɛ2NθMN¯0=λzTg22ɛ1221ɛ2NθMI0abq0, (28)

where Nθ is the total number of projections in data acquisition, σD2 is the noise at each detector cell, and q0 = N0/abI0. In a way analogous to that of the conventional CT,20, 21, 23 we define the noise equivalent quanta as

NEQ p=NθMI0, (29)

and then Eq. 28 can be rewritten as

NθσD2=λzTg22abq0ɛ1221ɛ2 NEQ p. (30)

On the other hand, according to Eq. A25 in the Appendix, one has

NθσD2=a4πk NPS pk MTF p2k. (31)

By equating the right sides of Eqs. 30, 31, one attains

NEQ p(k)=g2λzT21ɛ22πbq0ɛ12k NPS pk MTF p2k, (32)
DQE p(k)= NEQ p(k)NθMI0. (33)

Furthermore, as derived in the Appendix, the noise power spectrum NPSp(k) is

NPS p(k)=g2λzT2a1ɛ22πɛ12kNθMN¯0 MTF p2k=g2λzT21ɛ22πbq0ɛ12k NEQ p MTF p2k. (34)

It should be pointed out that, if N2 and N−2 are assumed to be zero, one gets ɛ2 = 0 and Eq. 27 becomes exactly the same as that derived in Ref. 10, in which only N0, N1, and N−1 are considered. In general, however, N2 and N−2 are small but not equal to zero. Hence, the derivation given here is a more general treatment in physics and mathematics. In addition, the factor |k| in the numerator of Eqs. 9, 10 moves to the denominators of Eqs. 32, 34. Readers are referred to the Appendix of this paper for the details related to the derivation of Eq. 34.

Measurement of the modulation transfer function MTFp(k) of DPC-CT

It has been an established practice in conventional CT to measure the MTFa(k) with a thin wire phantom.29, 46, 47 The wire is usually made of metal, such as tungsten, and placed in either air or water, as long as the attenuation of the thin wire does not exceed the dynamic range of the CT detector. Moreover, it has been well evaluated and verified that, as long as its diameter is substantially smaller than the detector cell dimension, the influence of the wire's thickness on the MTFa(k) measurement can be ignored.47 Unfortunately, however, this approach cannot be directly adopted for measurement of MTFp(k) in the x-ray tube and grating-based DPC-CT, because the detection of the signal generated by such a thin wire may be substantially compromised.13 The major reasons underlying this difficulty are: (i) the signal generated in the DPC-CT is the projection of the refractive coefficient's derivative, which is an odd function in the case of a cylindrical object; (ii) the detection of such a projection can be severely compromised because the integration of an odd function over the finite dimension of a detector cell may be null. As indicated in Ref. 13, a better approach is to utilize a cylindrical object with a diameter substantially larger than the detector cell dimension. The cylinder is placed parallel to the y axis of the DPC-CT and thus it becomes a circle in a reconstructed transverse image. The distribution along a line starting at the center of the circle (namely, a semiradial line) is actually an edge spread function (ESF). In order to reduce random interference, a total of 360 semiradial lines at 1° intervals are engaged, and the intensity profiles reconstructed along these semiradial lines are averaged to get the ESF. The line spread function (LSF) is attained from the ESF via a numerical method, in which an adequate upsampling is carried out to assure that the distance between adjacent pixels along each semiradial line be 4.0 μm for numerical accuracy. Subsequently, the DPC-CT's MTFp(k) is attained through a 1D discrete Fourier transform of the LSF. To assure a fair evaluation and comparison between the DPC-CT and conventional CT, this ESF → LSF → MTF(k) approach, rather than the conventional method using a thin wire, is also employed to obtain the MTFa(k) of the conventional CT.

It should be noted that, in theory, the so-called phase wrapping phenomenon, i.e., the detected phase exceeds the unambiguous 2π phase range, can occur in the measurement of DPC-CT's MTFp(k) with the method specified above. As indicated in the literature [Eq. 19 in Ref. 13], given a cylindrical object, phase wrapping occurs at

xw=R1+(mg2α/π)2, (35)

where m is the index of fractional Talbot distance, R is the radius of the cylindrical object, and α is the phase shift induced by the object per unit length along the x axis. Since g2 is very small in comparison to α, [e.g., α ≈ 46 rad/mm in soft tissues at 25 keV (Ref. 13)], xw is almost equal to R, i.e., phase wrapping occurs at the locations immediately adjacent to the cylinder's edge. It is argued in Ref. 13 that, because of the finite size of a detector cell, the phase wrapping in the case of a cylindrical object made of soft tissue may be reduced to an extent that is not detectable in the DPC-CT. This may be perceived as a form of the partial volume effect as we experienced in the conventional CT, and interested readers are referred to Ref. 13 for more details.

Quantitative evaluation of the spectrum of noise equivalent quanta NEQ(k) and detective quantum efficiency DQE(k)

According to Eqs. 9, 34, the noise power spectrum NPS(k) and modulation transfer function MTF(k) are the two prerequisites to obtain the spectrum of noise equivalent quanta NEQ(k). The noise power spectrum can be calculated by taking the Fourier transform of the autocorrelation function that is obtained using a large number of noise images. An alternative approach that is more efficient in computation is to take the average of the squared Fourier transform of a large number of noise images or regions within the images containing noise only.21 The obtained 2D NPS(k) is circularly symmetric about its origin, as predicted in Eqs. 9, 34. The modulation transfer function MTF(k) is acquired using the method depicted in Subsection 2C and is also in circular symmetry. Thereby, the spectra of noise equivalent quanta NEQ(k) of the conventional CT and DPC-CT can be obtained using Eqs. 10, 32, which are also in circular symmetry, followed by the spectra of detective quantum efficiency DQE(k) using Eqs. 11, 33, respectively.

EVALUATION VIA SIMULATION STUDIES

The spectrum of noise equivalent quanta NEQa(k) of conventional CT specified in Eq. 10 has been evaluated and verified through studies based on computer simulation and physical CT systems.20, 21, 23, 33, 34, 35 It is essential for us to evaluate and verify the DPC-CT's spectrum of noise equivalent quanta NEQp(k) derived in this paper. We carry out a computer simulation study as the initial effort. For the following reasons, we constrain ourselves to conducting the computer simulation study in the parallel beam geometry: (i) the x-ray beam in differential phase contrast CT satisfies the paraxial condition and thus the beam is almost parallel, (ii) the image reconstruction algorithms in the parallel beam geometry outperform those in the fan beam geometry from the perspective of noise uniformity and thus almost all the clinical CT scanners based on the third generation geometry (fan beam or cone beam) adopt the parallel beam reconstruction algorithms via fan-to-parallel rebinning,45, 46 and (iii) most of the simulation studies to investigate the noise power spectrum of CT imaging thus far have been carried out in the parallel beam geometry to exclude the influence of rebinning and weighting schemes. The x-ray source is assumed monochromatic at 30 keV, which irradiates an object by 360° at 1° steps so that no weighting effect can be induced to degrade the noise uniformity. At 30 keV, a 20-fold upsampling is assumed to simulate the x-ray beam's propagation through the gratings G1 and G2 with periods 8 and 4 μm, respectively, while the distance between these two gratings is 193.6 mm, i.e., 1/16 of the Talbot distance. The field of view (FOV) in data acquisition and image reconstruction is 40.96 × 40.96 mm2. The size of each detector cell is set at 32 × 32 μm2, 64 × 64 μm2, 96 × 96 μm2, and 128 × 128 μm2, resulting in the detector consisting of 1280, 640, 426 and 320 cells, respectively. Grating G2 shifts ten times at step 0.4 μm along the x axis to retrieve the phase information corresponding to the refractive property of the object to be imaged.

Since the purpose of this work is to investigate the potential imaging performance of the DPC-CT and compare it with that of the conventional CT, no windowing or boosting techniques42 are adopted in image reconstruction for both the DPC-CT and conventional CT. The data φ1(u, v) specified in Eq. 6 is used to reconstruct the DPC-CT images, in which the classical FBP (Ref. 42) with the Hilbert transform43, 44 is used. The projection data corresponding to the conventional CT are acquired without the G1 and G2 gratings in place, and the ramp filter is used to reconstruct the images with the classical FBP algorithm.42 It has been claimed in the literature that the data corresponding to the a0(u, v) in Eq. 4, which are acquired with the gratings G1 and G2 in place, are equal to the attenuation projection data acquired in the conventional CT. However, we agree with the discussion presented in Ref. 13 that such a claim only holds approximately, i.e., the a0(u, v) in Eq. 4 is not exactly the same as the attenuation projection acquired in the conventional CT. Hence, we do not include the results corresponding to the a0(u, v) in Eq. 4 in this paper.

Prior to analyzing the spectrum of noise equivalent quanta NEQ(k), the C-D phantom24 with outer dimension 37.68 × 28.26 mm2 is employed to evaluate and verify the modeling and simulation accuracy of x-ray propagation, data acquisition and image reconstruction of the DPC-CT. The bulk of the C-D phantom is made of water with its complex refractive coefficient equal to n = 1 − δ + = 1 −2.5604 × 10−7 + i1.2353 × 10−10, which is consistent with that specified in Ref. 2. Other parameters used to simulate the C-D phantom are: rod size (left to right): 16, 32, 64, 96, 128, 192, 256, 384, 512, and 1024 μm; contrast of the rods against phantom body (bottom to top): 5%–50% at step 5%.

To have an accurate and precise measurement of the modulation transfer functions MTFp(k) and MTFa(k), a cylindrical water phantom at diameter 5.1 mm is placed in air and thus the edge is sharp and at high contrast. To avoid any interference caused by noise nonuniformity, other than a water phantom, an air phantom, i.e., nothing placed in the x-ray beam, is employed to study the noise power spectra NPSp(k) and NPSa(k), and the spectra of noise equivalent quanta NEQp(k) and NEQa(k). The x-ray flux observing the Poisson distribution is set at 5 × 106 photon/cm2 per projection in the simulation study, which is consistent with that of an x-ray micro-CT in preclinical applications. At the detector cell size corresponding to 32 × 32 μm2, 64 × 64 μm2, 96 × 96 μm2, and 128 × 128 μm2, ∼360 regions of interest (ROI) at 128 × 128 matrix dimension are used to obtain the noise power spectrum via 2D Fourier transform,16 in which the technique of zero padding is implemented to convert the data matrix from the dimension 128 × 128 to 256 × 256 so that the noise aliasing effect can be avoided.

By definition, the noise equivalent quanta is the total effective number of x-ray quanta detected per unit of detector area in the data acquisition.20, 21, 22, 23, 27, 28, 30, 32 Hence, the specification of the total number of x-ray quanta, i.e., x-ray exposure, is crucial in the investigation of noise equivalent quanta, particularly in the scenario wherein two imaging methods are to be compared. As indicated in Sec. 2A, at each angular position in the data acquisition of DPC-CT implemented with x-ray tube and gratings, grating G2 needs to linearly shift M steps (M = 10 in this work). To take this fact into account, we set the x-ray exposure corresponding to the conventional CT as M times as that of the DPC-CT at each angular position. Thus, the x-ray exposure in the data acquisition of DPC-CT and the conventional CT are kept identical.

RESULTS

MTFp(k) of DPC-CT compared with MTFa(k) of the conventional CT

The modulation transfer functions of the DPC-CT and conventional CT have been thoroughly evaluated in our quantitative investigation. Plotted in Figs. 3456a are the MTFp(k) and MTFa(k) at detector cell size 32 × 32 μm2, 64 × 64 μm2, 96 × 96 μm2, and 128 × 128 μm2, respectively. A close inspection of these plots shows that their spatial resolution measured by the modulation transfer functions MTFp(k) and MTFa(k) is almost identical. We have experienced in the conventional CT that a difference in the noise texture/granularity in general means a difference in the spatial resolution. Fortunately, however, this is not the case with regards to the MTFp(k) and MTFa(k) between DPC-CT and conventional CT. It is important to note that the equality between MTFp(k) and MTFa(k) is of theoretical and practical relevance, and a detailed discussion on this respect is deferred to Sec. 5.

Figure 3.

Figure 3

The imaging performance of DPC-CT compared with that of the conventional CT: (a) MTFp(k) and MTFa(k), (b) NEQp(k) and NEQa(k), and (c) DQEp(k) and DQEa(k) (detector cell dimension: 32 × 32 μm2; x-ray exposure 5 × 106 photon/cm2 per projection).

Figure 4.

Figure 4

The imaging performance of DPC-CT compared with that of the conventional CT: (a) MTFp(k) and MTFa(k), (b) NEQp(k) and NEQa(k), and (c) DQEp(k) and DQEa(k) (detector cell dimension: 64 × 64 μm2; x-ray exposure 5 × 106 photon/cm2 per projection).

Figure 5.

Figure 5

The imaging performance of DPC-CT compared with that of the conventional CT: (a) MTFp(k) and MTFa(k), (b) NEQp(k) and NEQa(k), and (c) DQEp(k) and DQEa(k) (detector cell dimension: 96 × 96 μm2; x-ray exposure 5 × 106 photon/cm2 per projection).

Figure 6.

Figure 6

The imaging performance of DPC-CT compared with that of the conventional CT: (a) MTFp(k) and MTFa(k), (b) NEQp(k) and NEQa(k), and (c) DQEp(k) and DQEa(k) (detector cell dimension: 128 × 128 μm2; x-ray exposure 5 × 106 photon/cm2 per projection).

NEQp(k) of DPC-CT compared with NEQa(k) of the conventional CT

The profiles along the radial line that crosses the 2D spectra of noise equivalent quanta NEQ(k) of both DPC-CT and conventional CT at 45° are plotted in Figs. 3456b, corresponding to detector cell dimension 32 × 32 μm2, 64 × 64 μm2, 96 × 96 μm2, and 128 × 128 μm2, respectively. As such, the variation of NEQp(k) of the DPC-CT as a function over the detector cell dimension can be evaluated. Note that the fluctuation in the profiles of the spectra of noise equivalent quanta NEQp(k) and NEQa(k) are quite severe, because only 360 ROIs at matrix dimension 128 × 128 in the air phantom are engaged in the measurement. With an increasing number, e.g., more than 1000 (see Ref. 21), of ensemble samples (images or ROIs within the images of the air phantom), smoother profiles corresponding to the noise power spectra can be obtained. Note that the order of the polynomial fitting is determined empirically via trial-and-error to assure that the fitted profile adequately follows the trend in NEQp(k) and NEQa(k).

DQEp(k) of DPC-CT compared with DQEa(k) of the conventional CT

The profiles along the radial line that crosses the 2D spectra of detective quantum efficiency DQE(k) of both DPC-CT and conventional CT at 45° are plotted in Figs. 3456c, corresponding to detector cell dimension 32 × 32 μm2, 64 × 64 μm2, 96 × 96 μm2, and 128 × 128 μm2, respectively. It should be noted that the only difference between Figs. 3456b, 6c is the scaling by the x-ray exposure in the DPC-CT and conventional CT, respectively.

DISCUSSIONS

This work is a continuation of our previous investigation focused on the noise power spectrum NPSp(k) of DPC-CT.11 The primary contribution of this work is derivation of the functional forms of noise power spectrum NPSp(k), spectrum of noise equivalent quanta NEQp(k), and DQEp(k) of the DPC-CT. By conducting a computer simulation study, we evaluate and verify the derived functional forms and compare them with their counterparts in the conventional CT. To summarize the results, a number of observations and clarifications are given below.

In general, the spectrum of noise equivalent quanta of an imaging system is in the functional form20, 21, 26, 27, 28, 29, 30

NEQ (k)=G2 MTF 2(k) NPS (k). (36)

By definition, G, MTF(k) and NPS(k) denote the imaging system's transfer characteristics of large area gain, signal, and noise, respectively. In various imaging modalities, Eq. 36 can be in different expressions. For instance, the spectrum of noise equivalent quanta of x-ray radiography and fluoroscopy is expressed in exactly the same way as Eq. 36, whereas that of the conventional CT is expressed in Eq. 10. The spectrum of noise equivalent quanta NEQp(k) of the DPC-CT implemented with x-ray tube and grating is specified in Eq. 32. It is interesting and important to note that the term |k| is in the denominator of NEQp(k), whereas it is in the numerator of NEQa(k).

The computer simulation study shows that, given detector cell dimension, the modulation transfer function MTFp(k) of the DPC-CT is virtually the same as the MTFa(k) of the conventional CT [see Figs. 3456a]. It should be indicated that no widowing or boosting technique is utilized to obtain the modulation transfer functions investigated in this work. This means that, despite the difference in the imaging mechanisms between DPC-CT and conventional CT, their modulation transfer functions are essentially determined by dimension of the detector cells used for data acquisition, warranting that the signal transfer characteristics and spatial resolution are in principle identical in these two imaging methods. This result is consistent with what has been published in the literature.12, 13 Note that such a fact is of not only theoretical but also practical importance, especially in the scenarios wherein the complementary information drawn from the DPC-CT and the conventional CT are jointly of relevance for applications.

The primary finding of this work is that, except at the spatial frequency close to zero, the spectrum of noise equivalent quanta NEQp(k) and detective quantum efficiency DQEp(k) of DPC-CT are in principle identical to their counterparts NEQa(k) and DQEa(k) in the conventional CT, though there exists a radical difference in their noise power spectrum NPSp(k) and NPSa(k). This means that the DPC-CT makes use of x-ray flux as efficiently as the conventional CT if the object to be imaged is of finite size. This fact initially seems surprising since there is a significant difference between NPSp(k) and NPSa(k). However, it should become readily understandable if we inspect Eqs. 10, 32 carefully. The term |k| is at the numerator and denominator of NEQa(k) and NEQp(k), respectively, offsetting the substantial difference in their corresponding NPSa(k) and NPSp(k). Hence, NEQp(k), DQEp(k), NEQa(k), and DQEa(k) are essentially determined by MTFa(k) and MTFp(k), as analytically specified by Eqs. 10, 11, 32, 33 and experimentally verified by the simulation study as demonstrated in Figs. 3456b, 6c.

By referring to Eq. 1, one may reason that the advantage of DPC-CT in squared signal-to-noise ratio or detectability index over the conventional CT is solely dependent on the extent to which the ΔSp(f) in DPC-CT is larger than its counterpart ΔSa(f) in the conventional CT, since the NEQp(k) and NEQa(k) are essentially identical. Again, this seems inconsistent with our anticipation based on the observation that there exists a radical difference in their noise power spectrum NPSp(k) and NPSa(k). However, it should be understood that the SNR2ideal specified in Eq. 1 is defined under the framework of ideal observer, wherein an ideal observer is assumed to be capable of undoing or removing any correlation in noise, i.e., “pre-whitening” colored noise.26, 28 It is interesting and important to note that, the “pre-whitening” is accomplished by the term |k| at the numerator and denominator of NEQa(k) and NEQp(k), respectively, which offsets the substantial difference in their corresponding NPSa(k) and NPSp(k).

We would like to point out and emphasize that a human observer is not an ideal observer, i.e., a human observer is not able to “pre-whiten” colored noise while making a decision. The colored noise plays a significant role in disturbing a human observer's realistic decision-making capacity.26, 28 As we have observed, the noise in DPC-CT is abundant at low frequencies (see Fig. 2), while that in the conventional CT occurs at high frequencies. Hence, it is reasonable to anticipate that, if the object to be imaged is of relatively higher frequency, i.e., smaller in size, a human observer may make a better decision based on DPC-CT images than that made based on conventional CT images. This means that, even if ΔSp(f) = ΔSa(f), the SNR2nonideal of DPC-CT can still be higher than SNR2nonideal of the conventional CT, while the object to be imaged is small in size. In practice, a tomographic imaging method that is less susceptible to high frequency noise has been desired for a long time. The DPC-CT implemented with x-ray tube and grating is just such an imaging modality, which can be of profound significance in clinical and preclinical applications, e.g., the early detection of tumor or atherosclerosis, because a pathophysiological lesion usually starts at a small size.

We have indicated that the root cause for the radical difference between NPSp(k) and NPSa(k) is the adoption of the Hilbert filter kernel, rather than the ramp filter kernel, for image reconstruction in the DPC-CT using a filtered backprojection algorithm.11 However, it is interesting to note that such a radical difference between the NPSp(k) and NPSa(k) has also been reported in Ref. 12, wherein an integration is carried out on the projection data, followed by the well-known ramp filtering, i.e., no Hilbert filtering is carried out explicitly. In fact, however, a cascading of integration and ramp filtering is equivalent to a Hilbert filtering. Thus, we still believe that the root cause underlying the significant difference between NPSp(k) and NPSa(k) and the resultant difference in NEQp(k) and NEQa(k) is that the data acquired in DPC-CT is the projection of the derivative of the refractive coefficient and in principle only a Hilbert filtering is needed for image reconstruction. In addition, we would like to indicate that, though the image reconstruction scheme used in Ref. 12 may be beneficial in terms of inheriting the filtering strategies established in the conventional CT, the degradation in accuracy and spatial resolution due to the cumulative sum cannot be ignored in practice, especially in the cases wherein the detector cell dimension is not small.

The imaging chain of both DPC-CT and conventional CT is assumed ideal, which should not be confused with the concept of an ideal observer mentioned above, in its system modeling, analysis, and evaluation. For example, the x-ray source is assumed monochromatic in this work and thus the conventional CT investigated in this work is actually a “monochromatic” one. However, the conventional CT in reality usually utilizes a polychromatic x-ray source. Hence, the readers are advised to understand that the imaging performance, including the spectrum of noise equivalent quanta and detective quantum efficiency, of both DPC-CT and the conventional CT may be degraded by the imperfection in its imaging chain in practical situations.

The implementation of a DPC-CT is more challenging, because of the stringent requirements on the optoelectronic accuracy of grating fabrication, mechanical alignment, and stability. The time for data acquisition in the grating-based DPC-CT can be longer if grating G1 or G2 needs to shift multiple times. Owing to the phase wrapping phenomenon, the unambiguous phase detection range of DPC-CT is 2π, which may impose limitations on the dynamic range over preclinical and clinical applications. In addition, the noise morphology in DPC-CT images is similar to the texture of some soft tissues. On the other hand, it has been tabulated in Ref. 15 that the refractive coefficients of low atomic number materials are substantially larger than their attenuation counterpart. The preliminary data reported in the literature1, 4, 5 have demonstrated that the ΔSp(f) in soft tissues is significantly larger than its counterpart ΔSa(f). Moreover, as indicated above, there may exist an extra gain in the DPC-CT's SNR2nonideal over that of the conventional CT. Hence, it is hoped that the potential of significantly increased contrast sensitivity over soft tissues may outweigh the DPC-CT's implemental shortcomings and enable it to outperform the conventional CT as an imaging modality for extensive preclinical and clinical applications.

CONCLUSIONS

The spectrum of noise equivalent quanta NEQp(k) and detective quantum efficiency DQEp(k) of the DPC-CT is analytically derived in this paper, followed by an evaluation and verification via computer simulation study, as well as its comparison with that of the conventional CT. The signal transfer property of the DPC-CT characterized by the modulation transfer function MTFp(k) is virtually identical to the MTFa(k) of conventional CT. Moreover, though there exists a radical difference in their noise property characterized by the noise power spectrum NPSp(k) and NPSp(k), the spectrum of noise equivalent quanta NEQp(k) and detective quantum efficiency DQEp(k) of DPC-CT are essentially identical to their counterparts NEQa(k) and DQEa(k) in the conventional CT. It is believed that the NEQp(k) and DQEp(k) characteristics of the DPC-CT unveiled in this study can be of theoretical and practical relevance in the design and optimization of DPC-CT for extensive preclinical and ultimately clinical applications.

ACKNOWLEDGMENTS

This work is partially supported by the U.S. National Institute of Health through Grant No. 5P50CA128301, as well as by the Telemedicine and Advanced Technology Research Center (TATRC) at the U.S. Army Medical Research and Material Command (USAMRMC) Fort Detrick, MD via the award under Contract No. W81XWH-12-1-0138. In addition, the authors would like to express their appreciation to Ms. Jessica Paulishen for her proof reading of this paper.

APPENDIX: DERIVATION OF THE NOISE POWER SPECTRUM NPSP(K) IN DPC-CT

Letting the projection of an object function f(x, y) be

Pθ(t)=dxdyf(x,y)δ(x cos θ+y sin θt), (A1)

the object function f(x, y) can be reconstructed from its projection by

f^(x,y)=0πdθ|k|Sθ(k)× exp [2πik(x cos θ+y sin θ)]dk, (A2)

where the character is used to differentiate the reconstructed object function from the original object function. The Fourier transform of projection Pθ(t) is defined as

Sθk=Pθt exp 2πiktdt, (A3)

and its inverse Fourier transform can be written as

Pθt=Sθk exp 2πiktdk. (A4)

According to the property of Fourier transform, if Dθ(t) denotes the derivative of Pθ(t), one has the Fourier transform pair

Dθ(t)=Pθtt=2πikSθk exp 2πiktdk, (A5a)
Dθ(k)=Dθ(t)exp(2πikt)dt=2πikSθ(k). (A5b)

Substituting Eq. A5b into Eq. A2, one obtains

f^(x,y)=12π0πdθDθkkik exp [2πikx cos θ+y sin θ]dk=12π0πdθDθki sgn (k)exp[2πikx cos θ+y sin θ]dk=12π0πdθDθkH(k) exp [2πikx cos θ+y sin θ]dk, (A6)

where H(k) represents the Fourier transform of Hilbert filter kernel h(t), i.e.,

Hk=h(t) exp 2πiktdt, (A7a)
ht=i sgn (k) exp 2πiktdk. (A7b)

In the way similar to Ref. 20, without losing generality, we assume the object function f(x, y) is circularly symmetric, i.e., f(x, y) = f(r, θ) = f(r, 0) = f(r). In the polar coordinate system, one has

f^r=12π0πdθDθkHk exp 2πikr cos θdk. (A8)

Note that the subscript θ is kept in Dθ(t) for consistence in expression, though Dθ(t) is actually not a function of θ due to the circular symmetry in f(x, y). It should be easy to show that the Fourier transform F(k) of a circularly symmetric function f(r) is an even function and thus

f^(r)=02πdθ0dkF^kkexp2πirk cos θ=0πdθ0dkF^kk exp 2πirk cos θ+π2πdθ0dkF^kk exp 2πirk cos θ=0πdθ0dkF^kk exp 2πirk cos θ+0πdθ0dkF^kk exp 2πirkcosθ=0πdθ0dkF^kk exp 2πirk cos θ0πdθ0dkF^kk exp 2πirkcosθ=0πdθ0dkF^kk exp 2πirk cos θ+0πdθ0dkF^kk exp 2πirkcosθ=0πdθdkF^kk exp 2πirk cos θ. (A9)

Comparing Eqs. A9, A8, one gets

F^k=DθkHk2πk. (A10)

Letting the object function f(x, y) be a delta function, i.e., f(x, y) = δ(x)δ(y), the magnitude of the Fourier Transform of the reconstructed object function is the modulation transfer function, i.e.,

MTF pk=|F^(k)| while f(x,y)=δ(x)δ(y). (A11)

Noting that

Dθk=dt exp 2πiktPθtt=2πikdtPθt exp 2πikt=2πikdt exp 2πiktdxdyδxδyδx cos θ+y sin θt=2πikdt exp 2πiktδt=2πik, (A12)

and substituting Eqs. A11, A12 into Eq. A10, one obtains

MTF pk=ikHkk=i sgn (k)H(k)=H(k). (A13)

On the other hand, the autocovariance function of the reconstructed object function is

C(x,y)=Δf^(x,y)Δf^0,0, (A14)

where Δf^(x,y) represents the deviation of f^(x,y) from its mean value.20 Substituting Eqs. A5b, A7a into Eq. A6, one gets

f^x,y=0πdθ2πdk exp [2πikx cos θ+y sin θ]dtDθt exp 2πikt×dt1ht1 exp 2πikt1=0πdθ2πdtDθtdt1ht1dk exp [2πikx cos θ+y sin θtt1]=0πdθ2πdtDθtdt1ht1δx cos θ+y sin θtt1=0πdθ2πdtDθthx cos θ+y sin θt, (A15)
Δf^x,y=0πdθ2πdtΔDθthxcosθ+ysinθt. (A16)

Subsequently, the autocovariance function becomes

C(x,y)=0πdθ2πdtΔDθthxcosθ+ysinθt0πdθ12πdt1ΔDθ1t1ht1=0πdθ2πdthxcosθ+ysinθt0πdθ12πdt1ht1ΔDθtΔDθ1t1. (A17)

Assuming the recorded data Dθ(t) are uncorrelated, i.e.,

0πdθ1dt1h(x cos θ+y sin θt)h(t1)ΔDθ(t)ΔDθ1(t1) (A18)
=aπNθσD2h(x cos θ+y sin θt)h(t) (A19)

where σD2 is the noise at each detector cell, a is the detector pitch, and Nθ is the total number of projection Dθ(t). Then, Eq. A17 becomes

Cx,y=aσD24πNθ0πdθdthx cos θ+y sin θtht. (A20)

Again, without losing generality, we assume a circular symmetry in C(x, y), i.e.,

Cx,y=Cr,0=Cr=aσD24πNθ0πdθdthr cos θtht. (A21)

Substituting Eq. A7b into Eq. (A2a), one gets

Cr=aσD24πNθ0πdθdtdkHk exp 2πikr cos θtdk1Hk1 exp 2πik1t=aσD24πNθ0πdθdkHk exp 2πikr cos θdk1Hk1dt exp 2πik1t2πikt=aσD24πNθ0πdθdkHk exp 2πikr cos θdk1Hk1δk1+k=aσD24πNθ0πdθdkHk2 exp 2πikr cos θ. (A22)

The noise power spectrum NPSp(k), defined as the Fourier transform of autocovariance function, is an even function. Thus, the autocovariance function becomes20, 21

Cr=02πdθ0dk NPS pkk exp 2πikr cos θ=0πdθ0dk NPS pkk exp 2πikr cos θ+π2πdθ0dk NPS pkk exp 2πikr cos θ=0πdθ0dk NPS pkk exp 2πikr cos θ+0πdθ0dk NPS pkkexp2πikr cos θ=0πdθ0dk NPS pkk exp 2πikr cos θ0πdθ0dk NPS pkkexp2πikr cos θ=0πdθdk NPS pkk exp 2πikr cos θ. (A23)

Comparing Eqs. A23, A22, one obtains

NPS pk=aσD2Hk24πNθk. (A24)

Substituting Eq. A13 into Eq. A24 and reformatting, one finally arrives at

NθσD2=a4πk NPS pk MTF p2k. (A25)

References

  1. Momose A. and Fukuda J., “Phase-contrast radiographs of nonstained rat cerebellar specimen,” Med. Phys. 22, 375–379 (1995). 10.1118/1.597472 [DOI] [PubMed] [Google Scholar]
  2. Pfeiffer F., Weitkamp T., Bunk O., and David C., “Phase retrieval and differential phase-contrast imaging with low-brilliance x-ray sources,” Nat. Phys. 2, 258–261 (2006). 10.1038/nphys265 [DOI] [Google Scholar]
  3. Wu X. and Liu H., “Clinical implementation of x-ray phase-contrast imaging: Theoretical foundations and design considerations,” Med. Phys. 30, 2169–2179 (2003). 10.1118/1.1593836 [DOI] [PubMed] [Google Scholar]
  4. Donath T., Pfeiffer F., Bunk O., Grünzweig C., Hempel E., Popescu S., Vock P., and David C., “Toward clinical x-ray phase-contrast CT demonstration of enhanced soft-tissue contrast in human specimen,” Invest. Radiol. 45, 445–452 (2010). 10.1097/RLI.0b013e3181e21866 [DOI] [PubMed] [Google Scholar]
  5. Bech M., Jensen T. H., Feidenhan'l R., Bunk O., David C., and Pfeiffer F., “Soft-tissue phase-contrast tomography with x-ray tube source,” Phys. Med. Biol. 54, 2747–2753 (2009). 10.1088/0031-9155/54/9/010 [DOI] [PubMed] [Google Scholar]
  6. Cai W. and Ning R., “Dose efficiency consideration for volume-of-interest breast imaging using x-ray differential phase-contrast CT,” Proc. SPIE 7258, 72584D-72584D-9 (2009). 10.1117/12.813837 [DOI] [Google Scholar]
  7. Chou C. and Anastasio M. A., “Noise texture and signal detectability in propagation-based x-ray phase-contrast tomography,” Med. Phys. 37, 270–281 (2010). 10.1118/1.3267548 [DOI] [PubMed] [Google Scholar]
  8. Zambelli J., Bevins N., Qi Z., and Chen G.-H., “Radiation dose efficiency comparison between differential phase contrast CT and conventional absorption CT,” Med. Phys. 37, 2473–2479 (2010). 10.1118/1.3425785 [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Tang X., Yang Y., and Tang S., “Noise characteristics of x-ray tube and grating based phase CT over spatial resolution,” 1068th American Mathematics Society Meeting, Statesboro, GA, 12–13 March, 2011.
  10. Chen G.-H., Zambelli J., Li K., Bevins N., and Qi Z., “Scaling law for noise variance and spatial resolution in differential phase contrast computed tomography,” Med. Phys. 38, 584–588 (2011). 10.1118/1.3533718 [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Tang X., Yang Y., and Tang S., “Characterization of imaging performance in differential phase contrast CT compared with the conventional CT: Noise power spectrum NPS(k),”Med. Phys. 38, 4386–4395 (2011). 10.1118/1.3602071 [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Raupach R. and Flohr Thomas G., “Analytic evaluation of the signal and noise propagation in x-ray differential phase-contrast computed tomography,” Phys. Med. Biol. 56, 2219–2244 (2011). 10.1088/0031-9155/56/7/020 [DOI] [PubMed] [Google Scholar]
  13. Köhler T. and Roessl E., “Noise properties of grating-based x-ray phase contrast computed tomography,” Med. Phys. 38, S106–S116 (2011). 10.1118/1.3532396 [DOI] [PubMed] [Google Scholar]
  14. Born M. and Wolf E., Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light, 7th ed. (Cambridge University Press, Cambridge, England, 1999). [Google Scholar]
  15. Henke B. L., Gullikson E. M., and Davis J. C., “X-ray interaction: Photoabsorption, scattering, transmission, and deflection,” At. Data Nucl. Data Tables 54, 181–342 (1993). 10.1006/adnd.1993.1013 [DOI] [Google Scholar]
  16. Fitzgerald R., “Phase-sensitive x-ray imaging,” Phys. Today 53(7), 23–26 (2000). 10.1063/1.1292471 [DOI] [Google Scholar]
  17. Dainty C. and Shaw R., Image Science (Academic, London, 1974). [Google Scholar]
  18. Barrett H. H., Gordon S. K., and Hershel R. S., “Statistical limitations in transaxial tomography,” Comput. Biol. Med. 6, 307–323 (1976). 10.1016/0010-4825(76)90068-8 [DOI] [PubMed] [Google Scholar]
  19. Riederer S. J., Norbert N. J., and Chesler D. A., “The noise power spectrum in computed x-ray tomography,” Phys. Med. Biol. 23, 446–454 (1978). 10.1088/0031-9155/23/3/008 [DOI] [PubMed] [Google Scholar]
  20. Wagner R. F., Brown D. G., and Pastel M. S., “Application of information theory to the assessment of computed tomography,” Med. Phys. 6, 83–84 (1979). 10.1118/1.594559 [DOI] [PubMed] [Google Scholar]
  21. Hanson K. M., “Detectability in computed tomography images,” Med. Phys. 6, 441–451 (1979). 10.1118/1.594534 [DOI] [PubMed] [Google Scholar]
  22. Barrett H. H. and Swindell W., Radiological Imaging: The Theory of Image Formation, Detection, and Process (Academic, San Diego, CA, 1981). [Google Scholar]
  23. Hanson K. M., “Noise and contrast discrimination in computed tomography,” in Ridiology of the Skull and Brain, Vol. V: Technical Aspects of Computed Tomography, edited by Newton T. H. and Potts D. G. (Mosby, St. Louis, MO, 1981), pp. 3941–3955.
  24. Sandrik J. M. and Wagner R. F., “Absolute measures of physical image quality: Measurement and application to radiographic magnification,”Med. Phys. 9, 540–549 (1982). 10.1118/1.595099 [DOI] [PubMed] [Google Scholar]
  25. Faulkner K. and Moores B. M., “Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function,” Phys. Med. Biol. 29, 1343–1352 (1984). 10.1088/0031-9155/29/11/003 [DOI] [PubMed] [Google Scholar]
  26. Wagner R. F. and Brown D. G., “Unified SNR analysis of medical imaging systems,” Phys. Med. Biol. 30, 489–518 (1985). 10.1088/0031-9155/30/6/001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Metz C. E., Wagner R. F., Doi K., Brown D. A., Nishikawa R. M., and Myers K. J., “Toward consensus on quantitative assessment of medical imaging systems,” Med. Phys. 22, 1057–1061 (1995). 10.1118/1.597511 [DOI] [PubMed] [Google Scholar]
  28. ICRP (International Commission on Radiation Units and Measurements), “Medical Imaging: The assessment of image quality,” ICRP Report No. 54 (ICRP, Bethesda, MD, 1996), pp. 11–67.
  29. Bushberg J. T., Seibert J. A., E. M.LeidholdtJr., and Boone J. M., The Essential Physics of Medical Imaging, 2nd ed. (Lippincott, Philadelphia, PA, 2002). [Google Scholar]
  30. Barrett H. H., “NEQ: Its progenitors and progeny,” Proc. SPIE 7263, 72630F-72630F-7 (2009). 10.1117/12.817795 [DOI] [Google Scholar]
  31. Siewerdsen J. H., Cunningham I. A., and Jeffray D. A., “A framework for noise-power spectrum analysis of multidimensional images,” Med. Phys. 29, 2655–2671 (2002). 10.1118/1.1513158 [DOI] [PubMed] [Google Scholar]
  32. Siewerdsen J. H. and Jeffray D. A., “Three-dimensional NEQ transfer characteristics of volume CT using direct and indirect-detection flat-panel imagers,” Proc. SPIE 5030, 92–102 (2003). 10.1117/12.479970 [DOI] [Google Scholar]
  33. Boedeker K. L., Cooper V. N., and McNitt-Gray M. F., “Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta,” Phys. Med. Biol. 52, 4027–4046 (2007). 10.1088/0031-9155/52/14/002 [DOI] [PubMed] [Google Scholar]
  34. Tward D. J. and Siewerdsen J. H., “Cascaded systems analysis of the 3D noise transfer characteristics of flat-panel cone-beam CT,” Med. Phys. 35, 5510–5529 (2008). 10.1118/1.3002414 [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Baek J. and Pelc N. J., “Local and global 3D noise power spectrum in cone beam CT system with FDK reconstruction,”Med. Phys. 38, 2122–2131 (2011). 10.1118/1.3556590 [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. David C., Bruder J., Rohbeck T., Grunzweig C., Kottler C., Diaz A., Bunk O., and Pfeiffer F., “Fabrication of diffraction gratings for hard x-ray phase contrast imaging,” Microelectron. Eng. 84(5–8), 1172–1177 (2007). 10.1016/j.mee.2007.01.151 [DOI] [Google Scholar]
  37. Keren E. and Kafri O., “Diffraction effects in moiré deflectometry,” J. Opt. Soc. Am. A 2, 111–120 (1985). 10.1364/JOSAA.2.000111 [DOI] [Google Scholar]
  38. Siegel C., Loewenthal F., and Balmer J. E., “A wavefront sensor based on the fractional Talbot effect,” Opt. Commun. 194, 265–275 (2001). 10.1016/S0030-4018(01)01288-3 [DOI] [Google Scholar]
  39. Lohmann A. W., Knuppertz H., and Jahns J., “Fractional montgomery effect: A self-imaging phenomenon,” J. Opt. Soc. Am. A 22, 1500–1508 (2005). 10.1364/JOSAA.22.001500 [DOI] [PubMed] [Google Scholar]
  40. Grella R., “Fresnel propagation and diffraction and paraxial wave equation,” J. Opt. (Paris) 13, 367–374 (1982). 10.1088/0150-536X/13/6/006 [DOI] [Google Scholar]
  41. Bech M., “X-ray imaging with a grating interferometer,” Ph.D. thesis, University of Copenhagen, Denmark, 2009. [Google Scholar]
  42. Kak A. C. and Slaney M., Principles of Computerized Tomographic Imaging (IEEE, New York, 1988). [Google Scholar]
  43. Qi Z., Zambelli J., Bevins N., and Chen G.-H., “A novel quantitative imaging technique for material differentiation based on differential phase contrast CT,” Proc. SPIE 7622, 76220O-76220O-6 (2010). 10.1117/12.844536 [DOI] [Google Scholar]
  44. Noo F., Clackdoyle R., and Pack J. D., “A two-step Hilbert transform method for 2D image reconstruction,” Phys. Med. Biol. 49, 3903–3923 (2004). 10.1088/0031-9155/49/17/006 [DOI] [PubMed] [Google Scholar]
  45. Tang X., Hsieh J., Nilsen R. A., Dutta S., Samsonov D., Hagiwara A., Shaughnessy C., and Drapkin E., “A three-dimensional weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT: Helical scanning,” Phys. Med. Biol. 51, 855–874 (2006). 10.1088/0031-9155/51/4/007 [DOI] [PubMed] [Google Scholar]
  46. Tang X., Narayanan S., Hsieh J., Pack J. D., Mcolash S. M., Sainath P., Nilsen R. A., and Taha B., “Enhancement of in-plane spatial resolution in volumetric computed tomography with focal spot wobbling: Overcoming the constraint on number of projection views per gantry rotation,” J. X-ray Sci. Technol. 18, 251–265 (2010). 10.3233/XST-2010-0258 [DOI] [PubMed] [Google Scholar]
  47. Nickoloff E. L., “Measurement of the PSF for a CT scanner: Appropriate wire diameter and pixel size,” Phys. Med. Biol. 33, 149–155 (1988). 10.1088/0031-9155/33/1/014 [DOI] [PubMed] [Google Scholar]

Articles from Medical Physics are provided here courtesy of American Association of Physicists in Medicine