Abstract
Laminar flow patterning (LFP) is a characteristic method of microfluidic systems that allows two (or more) different solutions to flow side-by-side in a channel without convective mixing. This fluid behavior can be used to pattern cell suspensions, particles, and treatments as well as to create chemical gradients. LFP is typically implemented using syringe pumps and, for this reason, is most effective in constant flow scenarios such as long-term gradient generation. However, the complexity of using syringe pumps for patterning cell suspensions typically makes it a less attractive option than other standard patterning methods. We present a passive microfluidic method that enables short-term LFP of multiple fluids using a single pipette and allows each sample to be loaded in any sequence, at any point in time relative to one another. The proposed method reduces the complexity of LFP to be on a similar level as other cell patterning methods and is advantageous for cell-based applications such as co-culture and wound healing assays that require maximal exchange of soluble factors.
Laminar flow patterning (LFP) is a prominent and established method used in microfluidics to pattern cells, particles, and treatments within a single channel1. The method can be used to precisely control the nature and location of an interface between two miscible or non-miscible fluids, without requiring a physical divider2. In addition, the laminar behavior of fluids at the microscale allows diffusion to be leveraged as a highly controllable mixing force enabling the creation of precise gradients within a microchannel3. The method has also been leveraged in areas such as the dynamics of chemical and enzymatic reactions4 and the response of cell cultures to patterned treatments5. LFP has proven to be an enabling method for microfluidic cell-based assays where it is predominantly used for generating gradients of soluble factors in chemotaxis assays6, but is also used in the study of soluble factor signaling between different cell compartments7,8, and for patterning a gap between two cell culture populations in wound-healing/invasion assays9.
Despite the multitude of applications demonstrated using LFP, use of the method has not spread further than a restricted number of microfluidic labs, especially for applications where only a brief period of laminar flow is needed (i.e., in short-term LFP applications such as transient gradients or cell patterning). In short-term LFP applications, the complexity and expertise required to connect and operate syringe pumps for a brief period of laminar flow is prohibitive compared to alternative approaches. Further, tight control over flow rates is required to maintain synchronization of the flow through each branch. Using syringe pumps and tubes, it is difficult to ensure that flow through each branch ceases at exactly the same time due to different flow relaxation times caused by upstream capacitance and pumping mechanics (Fig 1A). Other difficulties of working with tubing include dead-volumes and air bubbles10. Thus, ease-of-use and accessibility are the biggest barriers to wide-spread use of short-term LFP for the purpose of cell patterning and transient gradients. Passive methods to perform LFP using capillary action have been demonstrated11, but require the use of surface patterning and surfactants which increase the fabrication complexity and prevent its use in cell-based applications.
We describe here a passive microfluidic device that requires only a single pipette to operate and addresses the challenge of manually synchronizing flows for LFP. One half of the device is a typical branched channel for LFP. The other half, placed upstream, consists of synchronization components that provides an intermediate storage location for sample fluids and unified pressure source. We achieve this by using reservoir channels and a surface tension-based fluid capacitor. These elements maintain constant fluid interfaces within the patterning area throughout sample loading. Specifically, the ability to cease flow down each branch at exactly the same time to preserve patterning is a major advantage of this method. The synchronization components can be used to stabilize LFP when using syringe pumps but the components also enable the use of other pumping strategies such as surface-tension base pumping12. In the case of surface-tension-based pumping, sample droplets can be placed using a pipette in any order and at any time relative to one another using a standard micropipette without disturbing the patterned interfaces.
The robust, pipette-based LFP method is particularly wellsuited for patterning cell populations in close proximity and would find immediate utility in co-culture and wound-healing assays. In order to characterize the functional range of the proposed method, numerical simulations were performed to understand the impact of different design considerations. Experimental demonstrations are used to illustrate the operation and potential applications of this method. Finally, a three-flow device was used to pattern two different cell types separated by a tunable gap as a means of modulating cell-cell communication.
Methods
Numerical Simulations
To better understand device operation, numerical modeling was performed using an electrical analogy of the microfluidic device that is valid for low Reynolds number laminar flows (Figure 3A). Simulations were performed using matlab with a time step of 0.1 ms and recorded for two different durations to characterize the beginning and ending of the flow (Fig 3). The resistance of each channel is calculated using the Washburn law while pressure drops at channel turns and junctions are neglected (ESI). Ports are modeled as non-ideal capacitors with a volume-pressure relationship given in Eq 1. The modeled device consists of two input ports of radius 500 μm, a main capacitor port of radius 1.25 mm and an output port of radius 1.5 mm. The reservoir ports are 250 μm tall, 4mm long and 2 mm wide. Each small branch of the Y channel is 100 μm tall, 750 μm wide, and 3 mm long and join into a channel that is 100 μm tall, 1 mm wide, and 5 mm long.
Device Fabrication
The device is made by bonding a PDMS (polydimethylsiloxane) mold of the channel network fabricated using standard softlithography techniques to a glass slide13. Plasma treatment of the PDMS and the glass was performed to render the surfaces hydrophilic and enable capillary filling of the device using PBS applied to each of the large ports. Initial filling can also be achieved using vacuum filling14 or by filling initially with ethanol and immediately replacing with PBS or media.
Device characterization
The device is filled with PBS such that the fluid is roughly flush with the surface of the device. A volume of 2.5 μ L was used to load samples. Samples were loaded with a minimum of a 3 second interval between each addition. PBS with red and green food colorant was used to illustrate device operation (Fig 2). Imaging of the device was performed on an Olympus SZX12 stereoscope (Olympus, Center Valley, PA). Similarly fluorescent analysis was performed using Alexa488 dye (Molecular Probes, Carlsbad, CA).
Cell culture
Cell lines (MCF7-GFP, MCF7-RFP, BEAS-2B, LNCaP and MC3T3-E1) were cultured according to the ATCC guidelines. At 80% confluency, the cell cultures were detached using trypsin, centrifuged and resuspended in PBS at 3.106 cells / mL. For 3-D applications, a cell suspension at a final concentration of 3.106 cells / mL was achieved by mixing 1:4 parts of collagen type I stock solution at 8 mg/mL, 1:4 parts of HEPES buffer and 1:2 parts of cell suspension at 6.106 cells / mL.
Cell patterning
A device was fabricated with three sample inputs to allow a cell-free solution to be patterned between two outer streams containing cells. Two MCF7 cell lines (ATCC), one expressing GFP (green fluorescent protein) and the other RFP (red fluorescent protein) were flown in the two outmost channels, green on one side and red on the other. Burst image acquisition was performed to record a movie of the actual loading process (Fig. 4B). Bone marrow MC3T3-E1 cells and LNCaP prostate cancer cells were flown in a similar manner in 3-D collagen suspension (Fig. 4D). In this case, the device was pre-filled with collagen to provide uniform viscosity for loading of the collagen-cell suspension.The migration of the bone marrow cells was observed in 2-D at 24 and 48 hours in the presence or absence of the LNCaP cells (Fig. 5B). Finally a wound healing-like assay was performed by patterning two compartments of lung epithelial cells, BEAS-2B, after coating with bovine collagen, and images were recorded at 24 and 48 hours.
Results and Discussion
Demonstration of Laminar Flow Patterning
The device used in this demonstration consists of a basic Y-channel with additional components to synchronize flows (Fig 1B). Each branch of the Y-channel contains a port to load the fluid of interest and a reservoir channel to hold that fluid temporarily. All reservoirs are connected to a central capacitor port that charges with fluid when samples are inserted into the input ports (Fig 2). Rapid charging of the capacitor occurs due to the low resistance between the inputs and capacitor and ensures that an equal pressure is applied on each branch of the Y channel. By applying an equal pressure to each branch, the proportion of fluid that travels down each branch is determined by their fluidic resistance. Thus, a loading event at any input, at any point in time, causes the contents of each reservoir to be pushed downstream in accordance with the ratio of each path resistance and results in LFP with a steady fluid interface.
Fig 2B demonstrates the steady flow-interface produced throughout pumping using colored fluids. This example also shows that the samples do not need to be loaded simultaneously for robust patterning. This is particularly beneficial in the context of a biology lab where the user often handles a multitude of reagents/solutions/dilutions. Fig 2C shows that branch resistances can be altered to adjust the location of the flow interface (2:1 flow ratio between branches). Similar experiments using fluorescent dye illustrate a well-defined flow interface and are summarized in the images of Fig 2D.
Analysis of Each Device Component
Analytical and numerical modeling have been performed to better understand the contribution of the different components of the microfluidic device. In order to adapt the use of this method to other applications, each component and the design considerations required for its proper function are discussed.
Surface tension and port capacitors
The fluid-air interface formed by a port can be used as a capacitor in a fluidic circuit, having a pressure-volume relationship. The surface tension-generated pressure, Pport, in a droplet of curvature Rcurv, at each port acts as a capacitor of fluid according to the Laplace relationship shown in Eq 1 where γ is the surface tension of the fluid. The volume of liquid, V, contained in a port, can also be linked to the radius of curvature (Eq 2), which allows to deduce, using Eq 1, a pressure-volume relationship defining the capacitor. For instance, a large port will be able to contain a large volume of fluid and will act as a weak capacitor, whereas a small port will act a stiffer capacitor generating large pressures when fluid is added.
(1) |
(2) |
As seen in Fig 1B, small ports (low-volume, stiff capacitors) are used for the loading ports of the device while large ports (high-volume, weak capacitors) are used to link the inputs and act as low pressure points for the fluidic circuit. Each large port is made big enough to absorb the whole amount of liquid that will be inserted for patterning. The loading ports are made small enough to quickly drive fluid to the capacitor and reduce synchronization times (Fig 3).
Synchronization of the flow
The experimental results of Fig 2 show that laminar flow patterning is maintained throughout pumping. However, a short-lived disturbance of the flow pattern occurs upon each addition of fluid to an input port, which is difficult to identify in the images. The disturbance results from the short amount of time it takes for the capacitor to be charged. Simulations performed to characterize the transient flow profile in each branch of the Y-channel after adding a single droplet of 1.0 μ L in one of the input ports show that syfirnchronization of the flows occurs in 15 ms (Fig 3B). During the rest of the flow (4 s) the flows in both branches of the Y channel match. An additional drop can be applied to any port at anytime (before or after the flow stops), causing the process of synchronization described here to be repeated. Shorter synchronization times minimize the time that flows are imbalanced and can be achieved by reducing the radius of the sample loading port, decreasing the volume being loaded, and reducing the resistance of the reservoir channels.
Reservoir channel design
The reservoir channels keep samples from mixing in the capacitor port. If the samples are allowed to reach the capacitor, samples might mix or possibly travel down an alternate branch of the device. Thus, the reservoir channels must be able to accommodate the entire sample volume. Based on previous characterizations of flow in rectangular channels, the amount of fluid needed to go beyond the end of the reservoir, Voverfill, is at most 0.48 times the volume of the reservoir due to parabolic flow profiles (Eq 3)15. Using this piece of information, the length of the reservoir channel can be designed appropriately.
(3) |
Y channel design
The ratio of flow down each branch of the Y-channel after synchronization, Q1/Q2, is proportional to the ratio of the resistances of each branch, R2/R1. Although the primary function of the branches is to determine where the interface between the flows will be patterned, they must also be designed to allow enough time for synchronization (Fig 3B). As mentioned previously, the Y-channel has a significantly higher resistance than the reservoirs such that when fluid is added to a loading port, most of the fluid travels directly to the capacitor port, charging it with fluid. Channel dimensions can be reduced to pattern narrower regions of interest (50-100 μm) and would result in longer pumping times. The pumping mechanism can be scaled to increase pumping velocities to avoid cell settling during patterning.
Sample loading
The timing of the loading no longer influences patterning but can affect how much sample is left in each reservoir channel after patterning. In the case of symmetric patterning, after the first sample is loaded and comes to rest, half of the sample will remain in the reservoir channel. After loading the second sample in the other port, the remainder of the first sample will be pushed out of its reservoir channel and half of the second sample will remain in its reservoir. Alternatively, if the loading of the first and second sample had been perfectly synchronized, none of either sample would remain in the reservoir channels. Both cases result in symmetric patterning but have different amounts of sample left over in the reservoir channels. Furthermore, this illustrates that a limited amount of time may still be required between sample loading if sedimentation of particles in the sample fluid is an issue of importance.
Patterning of More Than Two Streams
Fig 4 illustrates the ability to scale the technique to incorporate more than two streams for 2-D or 3-D culture as well as for increased throughput and screening applications. Cell suspensions of MCF7 cells transfected with either GFP or RFP were patterned a set distance apart using a third stream of PBS directed down the center of the channel. In each case, only the width of the center branch was varied to produce a narrow (150 μm), medium (250 μm), and wide (600 μm) spacing (Fig 4A). Fig 4B is the superposition of multiple movie frames taken using phase contrast in order to illustrate the path of each cell in suspension as it travels. Fig 4C shows the final result of patterning using a false-colored overlay of the fluorescent images with an enhanced phase contrast image. Fig 4D demonstrates the ability to pattern cells in 3-D matrices such as collagen. As shown in Fig 4E, the individually addressable devices can be arrayed to enable inclusion of multiple experimental conditions on a single chip.
Patterning for Cell-Cased Assays
The pipette-based strategy makes the method significantly more accessible to biologists by eliminating the need for syringes, tubes, and fluidic connections. The method is also more appropriate and robust for cell-based applications than other pipette-based methods which require the use of surface treatments and surfactants in the fluid to make patterning predictable11. As with other pipette-based methods, flow is transient and best-suited for applications where only a short period of flow is needed to establish the desired pattern, such as transient gradient generation, co-culture assays, wound-healing assays, and other particle patterning applications in 2-D or 3-D. The following figures illustrate the ability to use this method with cells to perform such assays.
Wound Healing and Migration
The patterning technique was used with a lung epithelial cell line (BEAS-2B) to illustrate the potential for performing wound healing assays. Dotted lines show the initial patterning location and the advance of the cells onto the blank substrate in the center. Groups of cells form protrusions via collective migration instead of each cell migrating independently to advance the front uniformly (Fig 5A, Day 2, white arrow). Fig 5B illustrates the same patterning method with two different cell types, MC3T3-E1 and LNCaP, to demonstrate a directed cell migration or invasion assay. Given the channels are individually addressable without the use of valves or other microfluidic components, appropriate controls can be easily performed on the same array (i.e. blank+A, A+A, B+B, and blank+B). Similarly, different treatments can be applied to each channel at different time-points. A subtle, but important, capability of this method is the ability to change the cell ratio (i.e. 2 x A + B, or A + 3 x B) using different flow ratios (see Fig 2) while maintaining a constant surface cell density – something that cannot be done in standard transwell assays16. This ability can be crucial for controlling the confounding effects of density dependent cell behavior seen in many in vitro models of cell signaling.
This pipette-based approach offers multiple specific advantages for cell-based assays in addition to increased accessibility. Each channel is individually addressable allowing any number of fluids and cells suspensions to be loaded into a each channel of an array at any time without valving or changing fluidic connections. Arrays of the pipette-based devices can be addressed using liquid handling automation to increase throughput. Soft-lithography can be used to tailor the device geometry to change the ratio of one cell-type to another without changing the surface cell density. The distance between the groups of cells can be tuned to modulate soluble factor signaling to study soluble factor sensitivity. Additional branches can easily be added to create more complex co-culture assays containing more than two cell types. The device also supports the culture of cells in small volumes to enable more rapid accumulation of factors, a characteristic which has shown to improve co-culture sensitivity16. Taken, together, the method offers significant advantages that increase the flexibility of the method and can broaden the use of LFP in cell-based assays.
Conclusions
While LFP is an attractive method for studying cell migration, soluble factor signaling, and other biological phenomena, it has not been readily available for robust every-day-use in biology laboratories. The passive LFP method described here allows samples to be loaded at any time and in any order using only a pipette without depending on the use of syringes, tubes, fluidic connections, surface treatments, or surfactants. This advance enables the design and creation of practical assays that can leverage advantages of LFP to study soluble factor signaling and are amendable to use in biology labs. Furthermore, the synchronization method described here is compatible with patterning and in-situ polymerization of cell-laden hydrogels such as collagen or matrigel. These are increasingly gaining momentum as being central to the study of cell migration/invasion in three-dimensional matrices as they offer more physiologically relevant environments8.
Acknowledgements
David J. Beebe has an ownership interest in Bellbrook Labs LLC, which has licensed technology reported in this publication. The developments reported were accomplished through funding from the following grants: NIH-NCI R33 CA137673, DOD PCRP Idea (W81XWH-09-1-0192), Korea Research Foundation Grant (KRF-2008-220-D00133) and a doctoral fellowship from the Morgridge Institute for Research (E. Berthier). We wish to thank Sameer Mathur for providing the BEAS-2B cells.
Footnotes
Electronic Supplementary Information (ESI) available. See DOI: 10.1039/b000000x/
References
- 1.Takayama S, McDonald JC, Ostuni E, Liang MN, Kenis PJ, Ismagilov RF, Whitesides GM. Proc Natl Acad Sci U S A. 1999;96:5545–8. doi: 10.1073/pnas.96.10.5545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lucchetta EM, Lee JH, Fu LA, Patel NH, Ismagilov RF. Nature. 2005;434:1134–8. doi: 10.1038/nature03509. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sia S, Whitesides G. Electrophoresis. 2003;24:3563–3576. doi: 10.1002/elps.200305584. [DOI] [PubMed] [Google Scholar]
- 4.Regenberg B, Krühne U, Beyer M, Pedersen LH, Simon M, Thomas ORT, Nielsen J, Ahl T. Lab Chip. 2004;4:654–7. doi: 10.1039/b409141h. [DOI] [PubMed] [Google Scholar]
- 5.Tourovskaia A, Figueroa-Masot X, Folch A. Lab On a Chip. 2005;5:14–19. doi: 10.1039/b405719h. [DOI] [PubMed] [Google Scholar]
- 6.Li Jeon N, Baskaran H, Dertinger SKW, Whitesides GM, Van de Water L, Toner M. Nat Biotechnol. 2002;20:826–830. doi: 10.1038/nbt712. [DOI] [PubMed] [Google Scholar]
- 7.Torisawa Y.-s., Mosadegh B, Luker GD, Morell M, O’Shea KS, Takayama S. Integr Biol (Camb) 2009;1:649–54. doi: 10.1039/b915965g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Sung KE, Yang N, Eliceiri KW, Keely PJ, Friedl A, Beebe D. Integrative Biology. 2010 doi: 10.1039/c0ib00063a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wong AP, Perez-Castillejos R, Christopher Love J, Whitesides GM. Biomaterials. 2008;29:1853–61. doi: 10.1016/j.biomaterials.2007.12.044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Paguirigan AL, Beebe DJ. Bioessays. 2008;30:811–821. doi: 10.1002/bies.20804. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kim S, Lim Y, Yang H, Shin Y, Kim K, Lee D, Park S, Kim Y. Analytical Chemistry. 2005;77:6494–6499. doi: 10.1021/ac0504417. [DOI] [PubMed] [Google Scholar]
- 12.Walker G, Beebe D. Lab On a Chip. 2002;2:131–134. doi: 10.1039/b204381e. [DOI] [PubMed] [Google Scholar]
- 13.Jackman R, Duffy D, Ostuni E, Willmore N, Whitesides G. Analytical Chemistry. 1998;70:2280–2287. doi: 10.1021/ac971295a. [DOI] [PubMed] [Google Scholar]
- 14.Zhao Y, Zeng H. Sensors and Actuators B-Chemical. 2009;139:673–681. [Google Scholar]
- 15.Warrick J, Meyvantsson I, Ju J, Beebe DJ. Lab On a Chip. 2007;7:316–321. doi: 10.1039/b613350a. [DOI] [PubMed] [Google Scholar]
- 16.Domenech M, Yu H, Warrick JW, Badders NM, Meyvantsson I, Alexander CM, Beebe DJ. Integrative Biology. 2009;1:267–274. doi: 10.1039/b823059e. [DOI] [PMC free article] [PubMed] [Google Scholar]