Vibro-acoustography Beam Formation with Reconfigurable Arrays

Hermes A S Kamimura 1, Matthew W Urban 2, Antonio A O Carneiro 3, Mostafa Fatemi 4, Azra Alizad 5
PMCID: PMC3408637  NIHMSID: NIHMS350990  PMID: 22828838

Abstract

In this work, we present a numerical study of the use of reconfigurable arrays (RCA) for vibro-acoustography (VA) beam formation. A parametric study of the aperture selection, number of channels, number of elements, focal distance, and steering parameters is presented in order to show the feasibility and evaluate the performance of VA imaging based on RCA. The transducer aperture was based on two concentric arrays driven by two continuous-wave or toneburst signals at slightly different frequencies. The mathematical model considers a homogeneous, isotropic, inviscid medium. The point-spread function of the system is calculated based on angular spectrum methods using the Fresnel approximation for rectangular sources. Simulations considering arrays with 50 × 50 to 200 × 200 elements with number of channels varying in the range of 32 to 128 are evaluated to identify the best configuration for VA. Advantages of two dimensional and RCA arrays and aspects related to clinical importance of the RCA implementation in VA such as spatial resolution, image frame-rate, and commercial machine implementation are discussed. It is concluded that RCA transducers can produce spatial resolution similar to confocal transducers, steering is possible in elevation and azimuthal planes, and optimal settings for number of elements, number of channels, maximum steering, and focal distance are suggested for VA clinical application. Furthermore, an optimization for beam steering based on the channel assignment is proposed for balancing the contribution of the two waves in the steered focus.

Index Terms: reconfigurable array, transducer, ultrasound, vibro-acoustography

I. Introduction

Developments in transducer technology including mechanical design, material properties, and signal processing are important for providing improvements in ultrasound imaging, such as enhanced resolution, contrast, signal-to-noise ratios, frame rate (real-time concept), etc. Array transducers enable ultrasound imaging without the need of mechanical scanning, which decreases the elapsed time in acquisition and avoids motion artifacts arising from low frame rate in mechanical scanning [1]. In addition, arrays enable the amplitude weighting (apodization) of the applied signal on the elements which controls the grating and side lobes. The point spread function (PSF) describes the spatial distribution of the excitation energy on an observation plane and facilitates the analysis of the scanning process and apodization effects. Two-dimensional arrays present more flexibility in phase, apodization, and aperture control. These features allow improvements in the PSF and enable the electronic steering of the focus over the target object in both azimuthal and elevational directions as opposed to linear and phased array, where mechanical displacement of the transducer is required for elevational scanning.

Reconfigurable arrays (RCA) are two-dimensional ultrasound transducers in which elements are dynamically connected in groups using microelectronic processes [2]. The dynamic combining of elements allows different settings of apertures, focusing, and electronic steering, with the advantage of requiring fewer number of channels [3]. The aperture control provides an accurate beamforming for electronic volume scanning including elevational, azimuthal, and axial directions. The electronic steering avoids the mechanical displacement of the transducer which prevents degradation in the image caused by mechanical errors and motion artifacts as well as noise inherent in the mechanical scanning.

Vibro-acoustography (VA) is an imaging technique which acquires the map of the acoustic response of an object under a low frequency excitation. In this method, the low frequency (kHz) response is generated by the interference of two co-focused ultrasound waves (MHz) [4] [5] [6]. The best results for VA, including resolution and intensity analysis, are provided by concave transducers due to the axisymmetric pattern of the PSF. However, mechanical scanning is necessary for this type of transducer which leads to problems related to coupling between the concave transducer and the object. These problems can be partially solved by using a water tank to host the transducer. Multiple laboratory VA studies using the confocal transducer in a water tank have been performed in breast [7] [8], thyroid [9], and prostrate tissues [10] [11] [12] [13]. The same confocal transducer has been also used for in vivo measurements in human breast [14] [15] [16]. More clinically oriented studies need to be performed using linear and phased arrays, which is facilitated by implementation of VA on a clinical imaging system [17].

A numerical investigation of the RCA beam formation for VA is presented in this paper. RCA transducers present thousands of rectangular elements which demands a long computational simulation time. An algorithm using Fresnel approximation is used for evaluating simulations with large number of elements (thousands of elements). Optimal values for the RCA parameters, such as number of elements, number of channels, steering, focal distance, for annular aperture, are assessed in order to evaluate the RCA performance and to define the most feasible configuration for VA imaging. The RCA makes possible both volume steering without mechanical scanning, as well as the flexibility to define the aperture. It enables the design of an annular pattern which improves the image quality and decreases the acquisition time in VA.

II. Theory

A. Dynamic acoustic radiation force

The frequency interference phenomenon described in VA is simulated by calculating pressures p1(r : ω1) and p2(r : ω2) at the same plane at distance r for two sources at slightly different frequencies ω1 and ω2, where r = [x0, y0, z0]. The resultant force F in the focus is given by

F=drSρc2{p12(r)2+p22(r)2+p1(r)p2(r)2cos[(ω2ω1)t+(ϕ2(r)ϕ1(r))]} (1)

where dr is the drag coefficient, S is the surface area, ρ is the density, c is the speed of wave, and ϕ1 and ϕ2 are the phases of the two waves 1 and 2.

The dynamic part of the force in (1) is given by

Fd=drSρc2{p1(r)p2(r)2cos[Δωt+(ϕ2(r)ϕ1(r))]} (2)

where Δω is the difference of frequencies ω1 and ω2.

The drag coefficient accounts for the effects of both absorption and scattering of incident waves. It is defined per unit incident energy density and unit projected area. The contribution of the total absorbed power Πa and the total scattered power Πs in the drag coefficient is given as

dr=aS(Πa+Πsγcos(θs)dS)+bSγsin(θs)dS (3)

where a and b are respectively the unit vectors parallel and perpendicular to the beam direction, γ is the scattered intensity, and θs is the angle between the incident and scattered intensity.

A brief summary of the implications of (3) as it relates to radiation force application in soft tissue is dicussed by Nightingale, et al. [18]. The contribution of absorption in soft tissue is much greater than that from scattering. The drag coefficient can be simplified to dr = Πa.

B. Analytic solution for the acoustic field

Clinical applications for VA such as thyroid and breast imaging involve interrogation of tissue from depths of 10–50 mm. In order to properly evaluate the use of RCA transducers for VA, we need to examine effects of beam formation over these ranges of depths. Numerical evaluation of the acoustic field emitted by rectangular sources can be evaluated by the impulse response method or the angular spectrum method. In the impulse response method, elements on the surface are considered as independent sources and their individual contributions for specific time and location are summed and weighed by the apodization function [19]. In the angular spectrum method, the two-dimensional (2D) Fast Fourier Transform (FFT) of the velocity potential Φ on the source plane is multiplied by the transfer function (from the source plane to the plane of interest) and finally the inverse Fourier transform of the result is calculated [20]. The pressure phasor p is given by

p=jωρ0Φ (4)

where ω is the angular frequency and ρ0 is the equilibrium density and the calculated pressure plane (plane of interest) is expressed in terms of an angular distribution of plane waves.

The Fresnel approximation method is an angular spectrum method that allows simple analytic solutions for rectangular sources in the near-field [21]. In our application, we are assuming a 2D array of rectangular elements as shown in Fig. 1 that are labeled with the subscripts m and n in the x and y direction, respectively. The analytic solution for the potential velocity calculated in time t at a distance Rmn = |rrs| (Fig. 1) is derived from Rayleigh integral

ϕN(r:t)=Sξ0υ0ejω(tRmnc0)αRmn2πRmndS (5)

where ξ0 is the spatial variation related to the apodization, υ0 is the velocity normal to the surface S, c0 is the wave propagation speed, Rmn=(x0xm)2+(y0yn)2+z02, and α is the attenuation coefficient.

Fig. 1.

Fig. 1

Reconfigurable array with rectangular sources (xm, yn) that lies on the source plane and an observation point (x0, y0, z0) in a observation plane at a Rmn distance.

In the Fresnel approximation, binomial-series expansion of the distance Rmn that is given by

Rmn=z0{1+1/(2z02)[(x02+y02)2(x0xm+y0yn)+(xm2+yn2)]+} (6)

is truncated at second order. This method enables a fast, straightforward, and relatively accurate evaluation of the field in the desired observation plane in comparison to the impulse response method [1].

C. Numeric implementation for the acoustic field analysis

The Fresnel approximation was implemented in the numerical simulations of the acoustic field emitted by the RCA in a homogeneous, isotropic, inviscid fluid taking into account focusing and apodization effects. The reasons for choosing the Fresnel approach involved the compromise between the computational cost and the reliability of results in the evaluation of focusing of VA.

The numerical implementation for rectangular transducers of the Fresnel approach in the frequency domain based on the product between the aperture and apodization functions is given by [1]

pmn(r:ω)ωρ0υ0z0λ8πRmnej(kz0ϕ0)αRmn{erf[σ(W2x0)]erf[σ(W+2x0)]}×{erf[σ(H2y0)]erf[σ(H+2y0)]} (7)

where ϕ0 = ωRmn/c, α is the attenuation coefficient, and the term σ is defined as

σ=jπ/(4λz0) (8)

and er f is the error function given by

erf(z0)=2π0z0eς2dς (9)

The error function is a simple mathematical representation of the Gaussian function convolved with a rectangular distribution of the elements [21].

Equation (7) is used to calculate the acoustic pressure emitted independently by each element. Each pixel on the observation plane corresponds to the contribution (sum) of the acoustic response of all elements on the source plane. The acoustic responses for frequencies f1 and f2 are given by

p(r:ωk)=mnAmn(ωk)pmn(r:ωk),k=1,2 (10)

where Amn is an apodization term to define the different apertures for ω1 and ω2 which represents the assignment of channel considering time delay and frequency of excitation (k = 1,2), and pmn is the contribution of pressure of each element in the source plane at mn-coordinate. Amn is 1 when an element is assigned ω and 0 otherwise.

Early simulations with large apertures caused crashes in MATLAB with Field II [22] [23] so the Fresnel method was adopted. Recent simulations with newer versions of Field II and MATLAB allowed for simulations of larger apertures, but at an added computational time. The numerical implementation of the Fresnel theory was checked against simulations using Field II. The comparison that was performed used a 50 × 50 element array that was set to be focused at 25 mm. We compared the fields by normalizing both of them and taking the difference. The fields were each independently normalized. We then computed the mean of the absolute values of the differences (Equation (11)) and found the MAD was 0.0023. The mean was taken of all the calculated points in the volume. This small difference is quantitative evidence that the Field II and Fresnel methods are within good agreement.

MAD=mean[|pN,FieldIIpN,Fresnel|] (11)

where the normalized pressure fields are given as pN,FieldII = pFieldII/max(pFieldII ) and pN,Fresnel = pFresnel/max(pFresnel) provided by Field II and the Fresnel-based simulations, respectively.

III. Method

A. General description

In this simulation, we considered a RCA with a maximum of 200 by 200 elements for a total of 40,000 elements. The distances between each pair of elements in x and y directions (kerf) were 0.001 mm with elements size of 0.2 mm × 0.2 mm. The central frequency f1 was set to 5 MHz and the frequency generated by the interference of the two incident waves Δf was set to 50 kHz.

The influence of the RCA parameters (Table I) including aperture size, number of channels, and focus position (steering in azimuth, elevation, and z-axis) were evaluated in order to define the optimal configuration for vibro-acoustography imaging. PSF profiles and acoustic fields were generated in x and y directions for quantitative comparisons of the parameters. The element size, the kerf (in x and y), and the radiating power density (power per element) were fixed. The numbers of channels assessed were 32, 64, 96, and 128. The aperture size was controlled by varying the number of active elements per axis (available for transmission). The active numbers of elements per axis assessed were 50, 100, 150, and 200. The total output power was directly proportional to the number of active elements.

TABLE I.

General setting parameters for RCA simulation.

Parameters Settings
Aperture Concentric circles
Elements size 0.2 mm × 0.2 mm
Kerf (x and y) 0.001 mm
Central frequency f1 5 MHz
Difference frequency Δf 50 kHz
Number of elements per axis Ne 50, 100, 150, 200
Number of channels Nc 32, 64, 96, 128
focal distance Df 20, 30, 40, 50

The steering performance of the RCA was analyzed in the azimuthal plane at x0-positions 0, 3, 6, 9, and 12 mm. The focus evaluation at different distances 20, 30, 40, and 50 mm was given on the elevation and azimuthal planes (Fig. 1).

B. Aperture definition for numeric study

A confocal-like implementation using concentric rings was adopted due to the good lateral resolution, angle independence, and long depth of field provided by its cylindrical symmetry [1]. The beamforming evaluation describes 2 concentric apertures used as ultrasound sources for generating the interference phenomenon in a common focus region. The different annular rings are depicted as 8 different gray levels in Figs. 2(a) and 2(d) for the cases where the beams are not steered and for when the beams are steered, respectively. The concentric apertures for f1 and f2 are defined in Figs. 2(b) and 2(e) by the white and gray elements, respectively. The delay profiles for the center of the images in Figs. 2(a) and 2(d) are shown in Figs. 2(c) and 2(f), respectively.

Fig. 2.

Fig. 2

Aspects for aperture definition: (a) channels, represented by equal areas showing activated elements for each channel; (b) frequency regions, areas with frequency excitation f1 in white (green, in the online version of the paper) and f2 in gray (red, in the online version of the paper); and (c) delays profile, where different levels represent the activation sequence of the channels. Pictures (d), (e), and (f) show the respective aspects for aperture definition for a 6 mm steering in x direction. The range of channels used in simulations was from 32 to 128. For a better visualization we represent only 8 channels in this scheme.

First, for a given focal position at (x0, y0, and z0) the time delays are calculated for each element in the array. For the purposes of this study, we will only use an active circular aperture inside the full square aperture. For each of the channels in the active part of the array, we want to have an equal number of elements. We divide the available active elements into Nc concentric rings. In the RCA transducer, each channel can be given a different signal with a particular time delay. All of the elements connected to this channel, will transmit the same signal. In this study, the signals are assumed to be continuous wave at a single frequency. For each channel, we determine a single time delay for all the elements assigned to that channel. We compute a mean value from the time delay values associated with the elements assigned to a channel. The next step is to assign the transmit signals at f1 and f2 to different sets of channels. Unless otherwise stated, we use equal numbers of channels for the beams for f1 and f2. When the focal point is off the central axis of the transducer, the delay will shift such that the maximum delay is provided for elements near to focal position (x0, y0). Concentric rings are used as in [24]. As seen in Fig. 2(d) the rings are shifted when steering to x0 = 6 mm. The frequency assignment is mathematically represented by Amn1) and Amn2) in equation (10). In this study, we want to mimic our confocal transducer which consists of a central disc and annular elements. With the RCA, we define a similar central disc and annular element. We numerically evaluate the pressure fields for the beams at f1 and f2 using equation (10). To find the magnitude of the low-frequency dynamic term at Δf, we calculate the product of the spatial distributions of the two beams as in equation (2).

C. PSF analysis

The spatial PSF analysis includes focal distance Df , focal zone Fz, and lateral resolution Rl. The spatial parameters are depicted in the illustration in Fig. 3. The focal distance Df is defined as the distance between the transducer and the maximum intensity point along the beam axis. The effective focal distance is the distance at which peak of the field occurs in the simulation. The effective focal distance was compared to the specified focal distance in the simulation (expected value). The focal zone Fz refers to the length of focus delimited by the full-width at half-maximum (FWHM) along the beam axis. The lateral resolution Rl is evaluated in this analysis by measuring the length of the region delimited by the FWHM in the focus region along azimuthal direction [25].

Fig. 3.

Fig. 3

Spatial parameters used in the focal evaluation. FWHM was used as reference to define the focal zone and lateral resolution. The focal distance is defined as the distance between the transducer and the point of maximum acoustic force.

The PSF analysis also includes the normalized force. According to the aim of the study, an independent normalization and an overall normalization of force were adopted resulting in the normalized forces Fd and Fn, respectively. The force was independently normalized for a most effective comparison of resolution cell for different number of elements and steering. The normalizing factor was the maximum value of all PSF’s that allowed the overall comparison of data for different aperture parameters such as aperture size and number of channels.

D. Steering optimization

In the simulations described above, we used an equal number of channels for the beams produced at f1 and f2. When the beam is steered, the steering angle for some of the elements far away from the focal position is so large that they contribute little or no pressure for beam formation. Essentially, the product p1(r).p2(r) in equation (2) is reduced, and the magnitude of the force at Δf is reduced. The product p1(r).p2(r) is maximized when the pressure in the two beams are equal. Steering the beam will create an imbalance between the amplitude of the two beams. One way to correct this imbalance is to devote more channels to one beam versus the other. We can define Nc as

Nc=Nc,1+Nc,2 (12)

where Nc,1 and Nc,2 are the number of channels devoted to the beams for f1 and f2, respectively. We can vary Nc,1 from 1 to Nc−1 and evaluate the magnitude of Δf component. We performed simulations with Ne = 200, Nc = 128 and varied the focal position of x0 = 1–28 mm and z0 = 10–50 mm. We found the value of Nc,1 that optimizes the magnitude of Δf component for each focal position.

IV. Results

A. Spatial focusing analysis

Fig. 2 exemplifies the confocal-like implementation adopted in this simulation. Fig. 2(a) shows the channels distribution for focus formation. The disk and annular regions in Fig. 2(b) represent the excited regions by frequencies f1 and f2. The interference of frequencies f1 and f2 generated a difference frequency Δf in the focus region. For providing a good visualization, the set of figures in the Fig. 2 shows delays for 8 channels. However, the simulated range of channels was from 32 to 128 channels.

First, we analyzed the effect of the focal distance and number of channels in the force magnitude. Fig. 4 shows the normalized force Fn provided by the RCA in the beam axis with Ne = 200 for different focal distances and number of channels. Varying Nc did not significantly affect the spatial parameters of PSF for a fixed focal distance, including effective focal distance Df , focus zone Fz, and lateral resolution Rl. However, the number of channels highly affected the peak force of PSF mainly for shallow focus distances. The set of PSF plots in Fig. 4 shows the decay of peak forces and the decreasing of difference of peaks for different channels with the increasing of distance of focus. The focal zone Fz was modified only for distance variation and it was increased with the distance.

Fig. 4.

Fig. 4

Amplitude pressure variation in relation to focal distances at 20, 30, 40, and 50 mm represented respectively in (a), (b), (c), and (d). The numbers of channels assessed were 32, 64, 96, and 128. The number of elements was fixed in 200 × 200. An overall normalization of force Fn was adopted.

Second, we observed the variation of the lateral resolution with the focal distance for fixed number of channels and elements. The lateral resolution Rl increased with focal distance and Fig. 5 shows the differences in azimuthal profiles for different focus distances and 128 channels.

Fig. 5.

Fig. 5

Lateral resolution for focal depths at 20, 30, 40, and 50 mm for 128 channels and 200 × 200 elements. An independent normalization of force Fd was adopted.

Third, we assessed the effect of the number of elements in the resolution cell for fixed focal distance and number of channels. Fig. 6 shows the maps of the Fd fields for different numbers of elements for a desired focused field at z0 = 20 mm. The effective focal distances were highly affected by the number of elements. The field was not properly focused at 20 mm unless the aperture was increased to at least Ne = 100. Also, we observe in Fig. 6 that the lateral resolution is degraded by decreasing the number of elements. The fnumber for each combination of Ne and focal distance Df is given in Table II.

Fig. 6.

Fig. 6

Maps of acoustic force with 128 channels and focal distance set to 20 mm for number of elements in the range of 50 × 50 to 200 × 200 elements. An independent normalization of force Fd was adopted.

TABLE II.

fnumber for apertures and focal distances used in the simulations.

Ne Expected focal distance
20 mm 30 mm 40 mm 50 mm
50 × 50
100 × 100
150 × 150
200 × 200
2.00
1.00
0.67
0.50
3.03
1.49
1.00
0.75
4.00
2.00
1.33
1.00
5.00
2.50
1.67
1.25

Fig. 7 shows the dispersion and the differences in the effective focal distance of focus, with specified focusing value at 20 mm, by using fewer elements.

Fig. 7.

Fig. 7

Axial profile of PSF showing the effective focal distance for number of elements varying between 50 × 50 to 200 × 200. An independent normalization of force Fd was adopted.

We evaluated the effect of steering in the focus formation at different focal distances. The Fd PSF maps shown in Fig. 8 present the steering in the azimuthal-axial plane using Nc = 128 and Ne = 200 for different values of focal distance. Due to the symmetry of the problem, the steering was evaluated only for half of the plane. At shallow depths, the PSF has a shorter focal zone and improved spatial resolution. As the focal distance increases, the focal zone and beam width increases.

Fig. 8.

Fig. 8

Maps of acoustic force for focal distances at (a) 20, (b) 30, (c) 40, and (d) 50 mm. The number of channels was set to 128 and equally assigned to each frequency, number of elements to 200 × 200, and the beam was steered in azimuthal direction for values at x0 = 0, 3, 6, 9, and 12 mm. An independent normalization of force Fd was adopted for each panel.

Fig. 9 was obtained by taking the maximum values for magnitudes in Fig. 8 at the specified x-distances (or steering positions). It shows that the normalized force is decreased by steering especially for focal distances near to the transducer. At 20 mm of focal distance, the steering at x = 12 mm decreased the magnitude of the acoustic force to 28% of the value without steering. However, for focus distances greater than 30 mm the magnitude is decreased less than 50% at x = 12 mm.

Fig. 9.

Fig. 9

Normalized maximum forces extracted from maps of Fig. 8 for expected focal distances at 20, 30, 40, and 50 mm. An independent normalization of force Fd was adopted for each focal depth.

Then, a summary of overall performance of the RCA in the focus formation for different number of channels and elements at different focal distance is reported. Table III shows quantitative results for effective focal distance, focus zone, lateral resolution, and normalized force for different values of Nc, Ne, and focal distances. Different tissues have different attenuation values and so we show the results with no attenuation in Table III to provide a baseline evaluation of the focusing and then more specialized optimization can be performed for different applications. The number of channels assigned to each frequency is equal for all cases in Table III. The 50 × 50 elements array degraded the effective focal distance by more than 20% of the desired value for every depth of focus assessed. Table III shows a tendency of the necessity to increase the number of elements for deeper depths not only due to the acoustic force decreasing but also due to the focus zone increasing and lateral resolution degradation. For a given focal distance, the number of elements highly affected the acoustic force magnitude and the difference between arrays with 150 and 200 elements was about 60%. The number of channels did not significantly change the spatial parameter results for fixed focal distance and number of elements. However for a focus distance at 20 mm using 100-element array, the acoustic force was reduced by 20% in comparison to using 32 and 64 channels. This difference was even more for larger values of Ne at the same distance of focus, but decreased for deeper distances.

TABLE III.

Summary of overall performance of the RCA for different number of channels and elements at different focal distance in terms of effective focal distance Df, focal zone Fz, lateral resolution Rl, and Normalized force Fn.

Expected focal distance
20 mm 30 mm 40 mm 50 mm
Ne Nc Df Fz Rl Fn Df Fz Rl Fn Df Fz Rl Fn Df Fz Rl Fn
50 × 50 32 16.4 11.9 0.55 0.008 21.2 18.9 0.65 0.005 24.5 24.5 0.75 0.003 27.0 28.7 0.85 0.002
64 16.4 11.9 0.55 0.009 21.2 18.8 0.65 0.005 24.5 24.4 0.75 0.003 27.0 28.7 0.85 0.002
96 16.4 11.9 0.55 0.009 21.2 18.8 0.65 0.005 24.5 24.4 0.75 0.003 27.0 28.7 0.85 0.002
128 16.3 11.9 0.55 0.009 21.2 18.9 0.65 0.005 24.5 24.4 0.75 0.003 27.0 28.7 0.85 0.002

100 × 100 32 19.8 4.9 0.35 0.081 28.9 9.8 0.45 0.044 37.5 16.0 0.55 0.027 45.5 23.0 0.75 0.018
64 19.7 4.9 0.35 0.094 29.9 9.7 0.45 0.047 37.4 15.9 0.55 0.028 45.4 23.0 0.75 0.019
96 19.7 4.9 0.35 0.098 28.8 9.7 0.45 0.048 37.4 15.8 0.55 0.029 45.4 23.0 0.75 0.019
128 19.7 4.9 0.35 0.099 28.8 9.7 0.45 0.048 37.4 15.8 0.55 0.029 45.4 23.0 0.75 0.019

150 × 150 32 20.1 2.7 0.25 0.164 30.0 5.0 0.35 0.138 39.5 8.3 0.45 0.100 49.0 12.3 0.55 0.072
64 19.9 2.7 0.25 0.345 29.7 5.0 0.35 0.197 39.4 8.2 0.45 0.123 48.9 12.3 0.55 0.083
96 19.9 2.7 0.25 0.392 29.7 5.0 0.35 0.209 39.4 8.2 0.45 0.127 48.9 12.3 0.55 0.085
128 19.9 2.7 0.25 0.409 29.7 5.0 0.35 0.214 39.4 8.2 0.45 0.129 48.8 12.3 0.55 0.086

200 × 200 32 20.5 2.1 0.15 0.031 30.2 3.2 0.25 0.142 40.1 5.1 0.35 0.166 49.9 7.5 0.45 0.150
64 20.1 1.8 0.15 0.613 30.0 3.2 0.25 0.454 39.9 5.1 0.35 0.317 49.7 7.5 0.45 0.227
96 20.0 1.8 0.15 0.887 30.0 3.2 0.25 0.548 39.8 5.1 0.35 0.355 49.6 7.4 0.45 0.244
128 20.0 1.8 0.15 1.000 30.0 3.2 0.25 0.583 39.8 5.1 0.35 0.369 49.6 7.4 0.45 0.251
1

Units for Df, Fz, Rl in mm. The number of channels assigned to each frequency is equal for all cases.

Absorption is the dominant source of attenuation in soft tissues. In Table IV, we recompiled some of the results of Table III to assess the effect of acoustic attenuation of 0.5 dB/cm/MHz in the focusing. As expected, the effect of the attenuation on the normalized force was significant. For a focal distance set to 50 mm, 200 × 200 of number of elements, and 128 channels, the force was reduced to 18% of the simulated value in Table III (without attenuation). For shallow distances, the focal zone and lateral resolution were less affected by the attenuation. However, the attenuation degraded the focus for deeper distances. The attenuation most affected the results for less number of elements and bigger focal distances, since the delivery of energy to the region of interest is highly reduced.

TABLE IV.

Recompiled results of Table III considering an attenuation of 0.5 dB/cm/MHz of the ultrasound waves in the medium where Df is the effective focal distance, Fz is the focal zone, Rl is the lateral resolution, and Fn is the normalized force.

Expected focal distance
20 mm 30 mm 40 mm 50 mm
Ne Nc Df Fz Rl Fn Df Fz Rl Fn Df Fz Rl Fn Df Fz Rl Fn
50 × 50 128 15.4 9.5 0.45 0.011 19.1 12.8 0.55 0.005 21.6 15.9 0.65 0.003 23.3 18.5 0.75 0.002

100 × 100 128 19.5 4.7 0.35 0.102 28.0 8.8 0.45 0.030 35.4 12.9 0.55 0.011 41.7 16.7 0.65 0.005

150 × 150 128 19.9 2.6 0.25 0.412 29.5 4.9 0.35 0.123 38.7 7.7 0.45 0.043 47.4 11.0 0.55 0.017

200 × 200 128 20.1 1.8 0.15 1.000 29.9 3.1 0.25 0.330 39.5 5.0 0.35 0.119 49.1 7.1 0.35 0.046
2

Units for Df, Fz, Rl in mm. The number of channels assigned to each frequency is equal for all cases.

B. Steering optimization

We varied the number of channels assigned to each ultrasound frequency and found that there was an optimal setting depending on the steered location. Fig. 10 shows the normalized magnitude of the Δf component for varying x0 positions. The total number of channels was Nc = 128 and the normalization for Fig. 10 was based on using Nc,1 = Nc,2 = 64. The increase in the magnitude of Δf component varies with the x0 position. We varied the positions of x0 and z0 and evaluated the optimal value of Nc,1 and the normalized magnitude of the Δf component at that optimal setting. Those results are summarized in Fig. 11. In Fig. 11(a) the optimal value of Nc,1 is shown for different combinations of focal positions x0 and z0. The highest values of Nc,1 were observed with shallow focal distances. Fig. 11(b) shows the gains in the magnitude of the Δf component relative to Nc,1 = Nc,2. For example at z0 = 20 mm and x0 = 15 mm, the optimal value of Nc,1 = 85 provided a gain of 15.8% over an equal distribution of the channels. Using this steering optimization may serve to change the curves in Fig. 9 such that the losses with steering are less pronounced. The largest gains again come at shallow focal distances and steering of ≈ 15 mm. Plots of the normalized magnitude of Δf component at different focal distances are shown in Fig. 12. Increases up to 4–16% in the magnitude of Δf component are evident compared to an equal split of channels, depending on the focal distance.

Fig. 10.

Fig. 10

Variation of normalized magnitude of Δf component with Nc,1 varying focal location for x0 = 0–25 mm and z0 = 20 mm. The total number of channels was Nc = 128. The values were normalized for the case of Nc,1 = Nc,2 = 64.

Fig. 11.

Fig. 11

Steering optimization results for different focal positions. (a) Optimal Nc,1 values, (b) Normalized magnitude of the Δf component with respect to Nc,1 = Nc,2 = 64.

Fig. 12.

Fig. 12

Optimized normalized magnitude of Δf component for different focal positions.

V. Discussion

When Nc was varied, the spatial parameters of PSF did not change appreciably, but the magnitude did vary. Using 32 channels for the 200 × 200 element array for focal distance at 20 mm, the field presented proper focusing and high resolution of focusing. However, for the same parameters, the normalized force was just 3.1% of maximum value. The number of quantization steps of the time delays was reduced when smaller numbers of channels were used. This resulted in much lower magnitudes of the Df component that was realized. This relationship is difficult to explicitly examine because of the number of channels involved and the involvement of two separate beams of different frequencies, but it is a topic that needs further investigation beyond the scope of this feasibility study. This would degrade the VA imaging since the radiation force would not be enough for exciting the target object. Although this variation reduced as the focal distance was increased, those results encourage the use of higher number of channels for apertures bigger than 100 × 100 elements since the number of channels is not a parameter that greatly affects the time of image acquisition.

Increasing Ne has a dramatic effect on the ability to focus the beam at the desired depth. This is a direct function of the aperture size used for focusing the beam. To focus the beam at z0 = 20 mm, the minimum number of elements necessary is 100 × 100. Table II shows the fnumber for the different apertures and focal distances simulated. By comparing Table II and Table III, it can be observed that to achieve proper focal distance, the fnumber values should be less than 1.67. Using a lower number of elements, the fnumber will increase which will lead to improper focusing. Increasing of the focus zone for low number of elements does not necessarily mean that this behavior can not be useful. Depending on the clinical problem such as bone imaging by VA [26] a larger extent of the focus zone can excite larger volumes. Thus, an optimization study of the coverage of focus zone and lateral resolution is necessary for defining the best setting for the number of elements for different applications.

To perform VA imaging entirely with electronic focusing, the beams will need to be steered. We evaluated the effects of steering on the magnitude and shape of PSF. The PSF resolution cell does not drastically change, but becomes tilted with respect to the face of the transducer and slightly asymmetric in line with the steering angle. The magnitude of the Δf component decreases as the beam is steered away from the center of the array.

One way to improve the loss in the magnitude of Δf component is to use unequal numbers of channels for the two ultrasound beams. It was shown that by changing the number of channels assigned to each frequency we can find an optimal setting for different focal positions. The gain in the magnitude of Δf component varies from 0–26 % for Nc = 128 and Ne = 200. The highest gains were found close to the transducer and steered to about x0 = 15 mm. The gains decrease with focal distance. This type of optimization could be readily performed with an RCA probe because of the ability to dynamically change the channels for each beam.

With the confocal transducer, the whole transducer is mechanically translated in a raster pattern to cover the area of interest. Steering with the RCA transducer shows that the magnitude of the force will decrease as the beam is steered farther. One approach to minimize these effects, is to use an active aperture smaller than the full aperture, for example Ne = 100. Then the beams could be steered, but the aperture could be translated within the full aperture so that the PSF is uniform for a portion of the image, very similar to scanning using a confocal transducer, and steered at the edges. This same type of process is used with linear array transducers as reported by Urban, et al [17]. To use this approach it is important to find the minimum value of Ne that provides proper focusing and enough force for adequate signal-to-noise ratio in the resultant image.

The simulation model presented in this paper provides a very useful platform for optimizing the implementation of VA using a RCA transducer. Based on the application and focal distance of interest, we can use the simulation model to find the best combination of Ne, Nc,1, and Nc,2.

Vibro-acoustography implemented using a RCA transducer offers some very important advantages. A very prominent advantage of using a 2D RCA transducer is that there would be no need for mechanical translation of the transducer as is needed with a confocal transducer [5] or a linear array transducer [17]. The time needed for mechanical translation of the confocal transducer leads to long scan times of up to 8 minutes for a 50 × 50 mm image with 200 × 200 pixels. The RCA electronic scanning is expected to take about 2 ms per pixel which is six-fold improvement over the confocal transducer (480 s/4000 pixels = 12 ms/pixel). The 2 ms interval is selected based on the time needed to record the acoustic emission at low frequencies for up to 300 µs and wait for the acoustic emission to attenuate so that it does not interfere with subsequent measurements as well as considerations for tissue and transducer heating.

Using the RCA transducer would provide another route for clinical implementation of VA imaging. The spatial resolution varies with focal distance, but is comparable to that of a confocal transducer with equivalent aperture size [27]. An advantage of RCA over using a linear array transducer is that the field is well focused in the elevation direction as well as the azimuthal and axial directions. B-mode ultrasound images could be acquired in register with the VA images for comparison if desired. The resolution of the system can be optimized based on multiple parameters discussed in this study, including the number of elements, and the number of channels assigned to each ultrasound frequency.

Examining the effects of attenuation more thoroughly will be a subject of further research. Implementation of VA requires an evaluation of possible transducer heating and bioeffects such as heating of tissue and production of cavitation. When an experimental transducer is available, the pressures and intensities would be evaluated and scaled to lie within FDA regulatory limits [28]. The tissue heating could be evaluated with numerical methods based on the methods detailed by Chen, et al [29].

The RCA transducer also requires specialized hardware for assigning the signals produced by a clinical scanner such as the General Electric Vivid 7 that we have used for VA imaging [17], to the defined channels. Channel and delay definitions will also need to be changed for each beam.

VI. Conclusion

We have demonstrated a numerical model for implementation of vibro-acoustography using a two-dimensional reconfigurable array transducer. We have studied the effects of number of elements, number of channels, and focal position and their effects on the PSF for the low-frequency force at Δf. We found that the PSF dimensions change with the number of elements and the focal position. Using fewer numbers of channels reduced the overall force magnitude. We also explored an optimization of the field produced based on assigning different numbers of channels for each ultrasound beam and found that we could achieve increased force by optimizing the number of channels assigned to each frequency. These simulations provide a strong framework for evaluation and optimization of VA imaging with RCA transducer for different applications in medical imaging.

Acknowledgment

This work was supported in part by the National Institute of Health under Grant CA 121579 and in part by the Brazilian Counsel of Technological and Scientific Development under grant CNPq 571801/2008-0.

The authors would like to thank Dr. James F. Greenleaf and Dr. Kai E. Thomenius (GE Global Research) for their valuable input to this project.

Biographies

graphic file with name nihms350990b1.gif

Hermes A. S. Kamimura obtained the B.Sc. in medical physics and M.Sc. in Physics Applied to Medicine and Biology from Universidade de São Paulo, Ribeirão Preto, São Paulo, Brazil. He is currently PhD student in Physics Applied to Medicine and Biology at the same institution. He was a visiting graduate student at the Mayo Clinic College of Medicine in 2010/2011. His research interests include image processing and biomedical instrumentation.

graphic file with name nihms350990b2.gif

Matthew W. Urban (IEEE/S’02, M’07) the B.S. degree in electrical engineering at South Dakota State University, Brookings, SD in 2002 and the Ph.D. in biomedical engineering at the Mayo Clinic College of Medicine in Rochester, MN in 2007. He is currently an Assistant Professor in the Department of Biomedical Engineering, Mayo Clinic College of Medicine. He has been a Summer Undergraduate Fellow at the Mayo Clinic Biomechanics laboratory. His current research interests are ultrasonic signal and image processing, vibro-acoustography, and vibrometry applications. Dr. Urban is a member of Eta Kappa Nu, Tau Beta Pi, IEEE, and the Acoustical Society of America.

graphic file with name nihms350990b3.gif

Antonio A. O. Carneiro received the B.Sc. degree in physics from Universidade Federal da Bahia, Salvador-Ba, Brazil, in 1995, and the M.Sc. and Ph.D. degrees in physics applied to medicine and biology from Universidade de São Paulo, Ribeirão Preto-SP, Brazil, in 1997 and 2001, respectively. From 2002 to 2003, he was a postdoctoral research associate at the department of Medical clinic at the Universidade de São Paulo, Ribeirão Preto-SP, Brazil. He is currently an associate professor in medical physics at the Universidade de São Paulo, Ribeirão Preto-SP, Brazil. His current research interests include bioinstrumentation and tissue characterization using ultrasound. He is a member of the Brazilian Physical society, Brazilian association of Biomedical Engineering. Currently, he is also an associate researcher to the Brazilian national research council (CNPq) level 1D.

graphic file with name nihms350990b4.gif

Mostafa Fatemi (M77SM07, F’12) received his Ph.D. degree in electrical engineering from Purdue University, West Lafayette, In. Currently, he is a professor of biomedical engineering at Mayo Clinic College of Medicine in Rochester, MN, and a member of the Mayo clinic Cancer Center, Cancer Imaging Program, and Prostate Cancer Program. His current research interests include ultrasound imaging and evaluation of the elastic properties of biological tissues by means of ultrasound radiation force. He leads research programs in the areas of breast and prostate imaging and tissue characterization. Dr. Fatemis pioneering work includes a novel imaging method known as vibro-acoustography, which was presented in two papers in Science (1998) and the Proceedings of the National Academy of Sciences (1999). He has been awarded eight patents on various aspects of vibro-acoustography and auditory effects of ultrasound on fetuses. He has published more than 150 peer-reviewed and proceedings papers and 10 book chapters. Also, he is the editor of a book entitled Biomedical Applications of Vibration and Acoustics in Imaging and Characterization. Dr. Fatemi is a fellow of IEEE, the American Institute for Medical and Biological Engineering, the Acoustical Society of America, and the American Institute of Ultrasound in Medicine.

graphic file with name nihms350990b5.gif

Azra Alizad received her M.D. degree from Tehran University of Medical Sciences, Tehran, Iran. Her past academic appointments include assistant professor of pediatrics at Tehran University and visiting assistant professor of pediatrics at the University of Florida, Gainesville, FL. Currently, she is an associate professor at the Mayo Clinic College of Medicine in Rochester, MN, and also a member of Mayo Cancer Center and Mayo Cancer Imaging Program. Her current research interests include medical ultrasound and applications of ultrasound radiation force for imaging and tissue characterization. Dr. Alizad is a senior member of American Institute of Ultrasound in Medicine, and a member of American Cancer Society, American Society of Breast Disease, Sigma-Xi Society, and IEEE.

Footnotes

Disclosure of financial interest: Mayo Clinic and one of the authors (MF) have a financial interest associated with the technology used in this research; the technology has been licensed in part for some specific application areas to industry.

Contributor Information

Hermes A. S. Kamimura, Departmento de Física e Matemática, Faculdade de Filosofia Ciências e Letras de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SP, Brazil

Matthew W. Urban, Email: urban.matthew@mayo.edu, Department of Physiology and Biomedical Engineering, Mayo Clinic, College of Medicine, Rochester, MN, 55905, USA.

Antonio A. O. Carneiro, Departmento de Física e Matemática, Faculdade de Filosofia Ciências e Letras de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SP, Brazil

Mostafa Fatemi, Department of Physiology and Biomedical Engineering, Mayo Clinic, College of Medicine, Rochester, MN, 55905, USA.

Azra Alizad, Department of Physiology and Biomedical Engineering, Mayo Clinic, College of Medicine, Rochester, MN, 55905, USA.

References

  • 1.Cobbold R. Foundations of Biomedical Ultrasound. USA: Oxford University Press; 2007. [Google Scholar]
  • 2.Fisher R, Thomenius K, Wodnicki R, Thomas R, Cogan S, Hazard C, Lee W, Mills D, Khuri-Yakub B, Ergun A, et al. Reconfigurable arrays for portable ultrasound. Proc. IEEE Ultrason. Symposium. 2005;vol. 1(no. 4):495–499. [Google Scholar]
  • 3.Hooi FM, Thomenius KE, Fisher R, Carson PL. Hybrid beamforming and steering with reconfigurable arrays. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2010 Jun.vol. 57(no. 6):1311–1319. doi: 10.1109/TUFFC.2010.1551. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Fatemi M, Greenleaf J. Ultrasound-stimulated vibro-acoustic spectrography. Science. 1998 Apr;vol. 280(no. 5360):82–85. doi: 10.1126/science.280.5360.82. [DOI] [PubMed] [Google Scholar]
  • 5.Fatemi M. Vibro-acoustography: An imaging modality based on ultrasound-stimulated acoustic emission. Proc. Nat. Acad. of Sci. USA. 1999 Jun.vol. 96(no. 12):6603–6608. doi: 10.1073/pnas.96.12.6603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Fatemi M, Manduca A, Greenleaf J. Imaging elastic properties of biological tissues by low-frequency harmonic vibration. Proc. IEEE. 2003 Oct.vol. 91(no. 10):1503–1519. [Google Scholar]
  • 7.Fatemi M, Wold L, Alizad A, Greenleaf J. Vibro-acoustic tissue mammography. IEEE Trans. Med. Imag. 2002 Jan.vol. 21(no. 1):1–8. doi: 10.1109/42.981229. [DOI] [PubMed] [Google Scholar]
  • 8.Alizad A, Fatemi M, Wold L, Greenleaf J. Performance of vibro-acoustography in detecting microcalcifications in excised human breast tissue: A study of 74 tissue samples. IEEE Trans. Med. Imag. 2004 Mar.vol. 23(no. 3):307–312. doi: 10.1109/TMI.2004.824241. [DOI] [PubMed] [Google Scholar]
  • 9.Alizad A, Mitri F, Kinnick R, Greenleaf J, Fatemi M. Vibro-acoustography of thyroid. J. Acoust. Soc. Am. 2007;vol. 122:3026. [Google Scholar]
  • 10.Mitri F, Trompette P, Chapelon J. Improving the use of vibro-acoustography for brachytherapy metal seed imaging: A feasibility study. IEEE Trans. Med. Imag. 2004 Jan.vol. 23(no. 1):1–6. doi: 10.1109/TMI.2003.819934. [DOI] [PubMed] [Google Scholar]
  • 11.Mitri FG, Davis BJ, Alizad A, Greenleaf JF, Wilson TM, Mynderse LA, Fatemi M. Prostate cryotherapy monitoring using vibro-acoustography: preliminary results of an ex vivo study and technical feasibility. IEEE Trans. Biomed. Eng. 2008 Nov.vol. 55(no. 11):2584–2592. doi: 10.1109/TBME.2008.2001284. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Mitri FG, Davis BJ, Greenleaf JF, Fatemi M. In vitro comparative study of vibro-acoustography versus pulse-echo ultrasound in imaging permanent prostate brachytherapy seeds. Ultrasonics. 2009 Jan.vol. 49(no. 1):31–38. doi: 10.1016/j.ultras.2008.04.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Mitri FG, Davis BJ, Urban MW, Alizad A, Greenleaf JF, Lischer GH, Wilson TM, Fatemi M. Vibro-acoustography imaging of permanent prostate brachytherapy seeds in an excised human prostate - Preliminary results and technical feasibility. Ultrasonics. 2009 Mar.vol. 49(no. 3):389–394. doi: 10.1016/j.ultras.2008.10.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Alizad A, Whaley D, Greenleaf J, Fatemi M. Potential applications of vibro-acoustography in breast imaging. Tech. Canc. Res. Treat. 2005 Apr.vol. 4(no. 2):151–157. doi: 10.1177/153303460500400204. [DOI] [PubMed] [Google Scholar]
  • 15.Alizad A, Whaley DH, Greenleaf JF, Fatemi M. Critical issues in breast imaging by vibro-acoustography. Ultrasonics. 2006 Dec.vol. 44(no. Suppl. 1):E217–E220. doi: 10.1016/j.ultras.2006.06.021. [DOI] [PubMed] [Google Scholar]
  • 16.Hosseini H, Alizad A, Fatemi M. Integration of vibro-acoustography imaging modality with the traditional mammography. Int. J. Biomed. Imag. 2007;vol. 2007 doi: 10.1155/2007/40980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Urban MW, Chalek C, Kinnick RR, Kinter TM, Haider B, Greenleaf JF, Thomenius K, Fatemi M. Implementation of vibro-acoustography on a clinical ultrasound system. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2011;vol. 58:1169–1181. doi: 10.1109/TUFFC.2011.1927. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Nightingale K, Palmeri M, Nightingale R, Trahey G. On the feasibility of remote palpation using acoustic radiation force. J. Acoust. Soc. Am. 2001;vol. 110:625. doi: 10.1121/1.1378344. [DOI] [PubMed] [Google Scholar]
  • 19.Stepanishen P. The Time-Dependent Force and Radiation Impedance on a Piston in a Rigid Infinite Planar Baffle. J. Acoust. Soc. Am. 1971;vol. 49:841. [Google Scholar]
  • 20.Williams E, Maynard J. Numerical evaluation of the Rayleigh integral for planar radiators using the FFT. J. Acoust. Soc. Am. 1982;vol. 72:2020. [Google Scholar]
  • 21.Mast TD. Fresnel approximations for acoustic fields of rectangularly symmetric sources. J. Acoust. Soc. Am. 2007 Jun.vol. 121(no. 6):3311–3322. doi: 10.1121/1.2726252. [DOI] [PubMed] [Google Scholar]
  • 22.Jensen J, Svendsen N. Calculation of pressure fields from arbitrarily shaped, apodized and excited ultrasound transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 1992;vol. 39(no. 2):262–267. doi: 10.1109/58.139123. [DOI] [PubMed] [Google Scholar]
  • 23.Jensen J. Field: A program for simulating ultrasound systems. 10th Nordic-Baltic Conference on Biomedical Imaging; Citeseer. 1996. pp. 351–353. [Google Scholar]
  • 24.Cogan S, Fisher R, Thomenius K, Wodnicki R. Solutions for reconfigurable arrays in ultrasound; Proc. IEEE Ultrason. Symposium; 2006. Oct. pp. 116–119. [Google Scholar]
  • 25.Ziskin M, Lewin P. Ultrasonic Exposimetry. CRC; 1992. [Google Scholar]
  • 26.Calle S, Remenieras J, Matar O, Defontaine M, Patat F. Application of nonlinear phenomena induced by focused ultrasound to bone imaging. Ultrasound Med. Biol. 2003;vol. 29(no. 3):465–472. doi: 10.1016/s0301-5629(02)00729-9. [DOI] [PubMed] [Google Scholar]