PEDOT polymer coatings facilitate smaller neural recording electrodes

Kip A Ludwig 1,*, Nicholas B Langhals 1, Mike D Joseph 1, Sarah M Richardson-Burns 2, Jeffrey L Hendricks 2, Daryl R Kipke 1
PMCID: PMC3415981  NIHMSID: NIHMS269939  PMID: 21245527

Abstract

We investigated using poly(3,4-ethylenedioxythiophene) (PEDOT) to lower the impedance of small, gold recording electrodes with initial impedances outside of the effective recording range. Smaller electrode sites enable more densely packed arrays, increasing the number of input and output channels to and from the brain. Moreover, smaller electrode sizes promote smaller probe designs; decreasing the dimensions of the implanted probe has been demonstrated to decrease the inherent immune response, a known contributor to the failure of long-term implants. As expected, chronically implanted control electrodes were unable to record well-isolated unit activity, primarily as a result of a dramatically increased noise floor. Conversely, electrodes coated with PEDOT consistently recorded high-quality neural activity, and exhibited a much lower noise floor than controls. These results demonstrate that PEDOT coatings enable electrode designs 15 microns in diameter.

Keywords: Microelectrodes, neural recordings, PEDOT, conductive polymers

1. INTRODUCTION

The utility of implantable microelectrode arrays is currently limited by the size of the individual electrode sites for a number of reasons. First, smaller electrode sites facilitate the development of more densely packed microelectrode arrays. At present, researchers are limited to sampling from a few neurons out of the billions of neurons which execute function in the brain1. As a result, the study of how large networks of neurons interact to produce biologically relevant behaviors is severely hampered. Similarly, the efficacy of modern neuroprosthetic devices is greatly dependent upon the observable number of neural inputs/outputs. Second, smaller electrode sites promote the design of smaller arrays, which in turn cause less damage upon implantation. Recent studies indicate that probe dimensions smaller than 12 microns minimize the reactive cell responses that negatively impact long-term neural recordings2-4.

Unfortunately, decreasing the size of an electrode site increases the impedance, which can degrade recordings. Impedance impacts recordings primarily through two mechanisms: noise and shunt loss 5. Noise at the electrode/electrolyte interface can arise from random fluctuations of charged carriers - either electrons in an electrical conductor (known as Johnson or thermal noise), or ions in an electrolytic medium 6-10. These fluctuations can be caused by Brownian motion of electrons, drift and diffusion of charged ions due to concentration gradients, oxidation/reduction reactions occurring at the electrode/electrolyte interface, etc 10. As these random movements occur they create current perturbations, which increase the voltage noise in proportion to impedance10. Noise resulting from the fluctuation of charged particles (which we will refer to as fluctuation noise hereafter) represents only one source of noise out of many possible sources: instrumentation noise, shot noise, flicker noise, biological noise, etc. Sources of noise sum in quadrature (a2 + b2 = c2), meaning that the single largest source of noise tends to dominate the noise floor. Functionally, this means that the fluctuation noise for an electrode will only marginally contribute to the total noise floor until a nominal impedance magnitude is reached. Beyond this impedance magnitude, fluctuation noise will tend to dominate the observed noise floor.

Shunt loss is defined as the loss of signal from the electrode and measurement system to ground. This can be separated into three distinct circuit elements. First, there is a capacitive loss from the metal traces on the microelectrode to the surrounding Cerebral Spinal Fluid (CSF)9. Second, a resistive element models the loss of signal from the inherent resistive nature of the metal wiring from the electrode to the measurement system9. Finally, there is a capacitor that models the capacitive signal loss in the measurement system9. Like fluctuation noise, shunt loss increases in proportion to impedance9. Typically, electrodes with impedances of 5 MΩ or greater have levels of fluctuation noise and shunt loss that make recording from individual neurons problematic 5,9,11.

Conductive polymer coatings have been hypothesized to be an enabling technology for smaller electrode designs12-17. Conductive polymers increase the electrochemical surface area of an electrode (ESA) without changing its geometric surface area (GSA - e.g. diameter, circumference), lowering impedance 18-21. In 2006, the Kipke lab demonstrated that surfactant-templated poly(3,4-ethylenedioxythiophene) (PEDOT) could be used to improve chronic neural recordings from standard-sized electrodes 12.

Building on this earlier work, in this study we demonstrate that PEDOT coatings effectively lower the impedance of small, gold recording electrodes (15 microns in diameter) with initial impedances outside of the effective range. Chronically implanted unmodified electrodes were unable to record well-isolated unit activity, primarily because of an increased noise floor arising from large site impedances. Conversely, sites coated with PEDOT were able to consistently record high-quality neural activity, and exhibited a markedly lower noise floor than controls.

2. METHODS

2.1. Microelectrodes

Three male Sprague-Dawley rats were implanted with three 16-channel chronic silicon ‘Michigan’ microelectrode arrays, using experimental procedures outlined previously 12,22. Arrays consisted of four shanks, each with four evenly spaced gold electrodes. Site and shank separations were sufficient (100 μm or greater) to limit the probability of an individual neuron being recorded from multiple sites 23. All electrodes used in this study were 15 microns in diameter, or a GSA of 177 μm2.

2.2. Electrochemical Deposition & Initial Evaluation

Electrochemical deposition of PEDOT in this study was accomplished using an electrochemical potentiostat/ galvanostat (Autolab PGSTAT12, Eco Chemie, Urtecht, The Netherlands) with associated General Purpose Electrochemical System (GPES) software. PEDOT doped with tetraethylammonium perchlorate and dissolved in 20 wt% surfactant poly(oxythylene)10-oleyl ether was galvanostatically deposited onto the gold sites of the neural probes 24. Based on results from our prior paper, PEDOT films generated with a deposition charge of 260 mC/cm2 were chosen for in vivo testing. Scanning electron microscopy (SEM) demonstrated that the PEDOT film deposited using a charge of 260 mC/cm2 did not increase the diameter of the recording site (Figure 1). Consequently, at 260 mC/cm2 the PEDOT film increased the electrochemical surface area of the electrode (ESA) without increasing the geometric surface area (GSA). Eight sites on each probe were deposited with surfactant-templated ordered PEDOT film. The deposited sites were staggered in relative location to prevent bias due to specific shank location or cortical depth (Figure 3). The remaining eight sites on each probe were left uncoated as controls for comparison.

Figure 1. Scanning Electron Microscopy, PEDOT coatings.

Figure 1

A) Depicts a PEDOT film generated using a deposition charge of approximately 260 mC/cm2, and B) depicts a PEDOT film generated using a deposition charge of approximately 1600 mC/cm2. Note that the PEDOT film generated using 260 mC/cm2 does not increase the geometric diameter of the underlying gold site, whereas the film generated using a deposition charge of 1600 mC/cm2 increases the effective diameter of the gold site by approximately 20 percent. For this study, PEDOT films were generated using a deposition charge of 260 mC/cm2.

Figure 3. Representative High Speed Data Across One 4-shank Array.

Figure 3

Black traces are PEDOT sites, while Blue traces denote Controls. The noise floor on control sites is dramatically larger than the noise floor on PEDOT sites, presumably obscuring unit activity.

2.2. Surgical Techniques

All of the arrays in this study were implanted in motor cortex, targeting cortical layer V, as outlined in previous work12,22,25 . Initial anesthesia was administered via intra-peritoneal injections of a mixture of 50 mg/ml ketamine, 5 mg/ml xylazine, and 1 mg/ml acepromazine at an injection volume of 0.125 ml/100g body weight. Updates of 0.1 ml ketamine (50 mg/ml) were delivered as needed to maintain anesthesia during the surgery. Animals were secured to a standard stereotaxic frame, and three stainless steel bone-screws were inserted into the skull. The electrode connector was grounded to a bone-screw over parietal cortex using a stainless steel wire.

A craniotomy approximately 3 mm by 2 mm was made over the target area (target location 3.0 mm anterior to bregma, 2.5 mm lateral from bregma, and 1.4 mm deep from the surface of the brain). Two incisions were made in the dura mater to create four flaps, which were subsequently folded back over the edge of the craniotomy. The electrodes were then hand inserted using microforceps into the approximate target cortical area. Cortical depth was estimated using the known location of the electrode sites on the individual shanks in conjunction with the known length of the individual shanks. Next, the surface of the brain was covered with GelFoam® (Henry Schein, Inc., Miami, FL) for protection. The silicon cable connector was covered with either remaining Gelfoam or Kwik-Sil silicone polymer (World Precision Instruments, Inc). The entire assembly excluding the connector was then enclosed using dental acrylic (Co-Oral-Ite, Dental Mfg. Co., Santa Monica, Ca). Finally, sutures were used to close the skin around the acrylic and triple-antibiotic ointment was applied. All procedures complied with the United States Department of Agriculture guidelines for the care and use of laboratory animals and were approved by the University of Michigan Animal Care and Use Committee.

2.3. Neural Recordings & Data Analysis

Recorded neural signals were acquired using a Plexon Multi-channel Neural Acquisition Processor (MNAP; Plexon Inc, Dallas, TX). Neural electrophysiological recordings for all sixteen channels were amplified and bandpass filtered; single and multi-unit recordings were sampled at 40 kHz (Plexon), and bandpass filtered from 450-5000 Hz. All recordings were taken in reference to a distant stainless steel (316-SS grade) bone screw inserted through the skull during surgery. During recording sessions, animals were placed in an electrically shielded recording booth and multiple 30-second segments of continuous neural recordings were taken. Animals were lightly anesthetized with ketamine/xylazine throughout the data collection sessions.

Neural recording segments were analyzed offline to determine number of neurons recorded, noise levels, and signal amplitudes using custom automated MATLAB (Mathworks Inc., MA) software, as described in detail12 and utilized elsewhere 12,21,25-29. In summary, an amplitude threshold window was set 3.5 standard deviations above and below the mean of the sample distribution. For each peak exceeding the threshold window, a 2.4 ms candidate waveform snippet centered on the absolute minimum of the waveform was removed from the recorded segment and stored. The amplitude of the noise voltage for every recording site in each recorded segment was calculated after all candidate waveforms had been removed.

After initial principal component analysis and fuzzy C-means clustering12, waveforms with a cluster membership index of greater than 0.8 were used to determine a mean waveform for a cluster. An interspike interval histogram for each cluster was generated and visually inspected for an obvious absolute refractory period as an additional measure of noise rejection. Signal amplitude for a cluster was defined as the peak-to-peak amplitude of the mean waveform for each cluster.

The signal-to-noise ratio (SNR) for a given cluster was defined as follows:

SNR=Signal Amplitude(2CalculatedRMSNoise Voltage for Recording Site)

Clusters were then separated into one of four categories based on calculated SNR. Clusters with an SNR of greater than 4 were categorized as quality units. Clusters with an SNR between 3 and 4 were categorized as moderate units. Clusters with an SNR between 2 and 3 were categorized as poor units, while clusters with an SNR of less than 2 were not considered units. These four categories correspond well with observations of unit quality based on signal-to-noise ratio made in similar recording studies 12,23,25.

Isolating action potentials from an individual neuron using an individual recording site is inherently prone to classification errors 30-31. The methodology employed in this study is intended to minimize these errors, and should accurately parallel the true number of underlying neural sources. The sorting routine produces similar results to manual sorting performed by experienced researchers over the same data sets, but with the advantage of being objective and automated 12.

2.4. Impedance Spectroscopy Measurements

Impedance spectroscopy measurements were made using an Autolab potentiostat PGSTAT12 (Eco Chemie, Utrecht, The Netherlands) with associated frequency response analyzer (Brinkmann, Westbury, NY). Impedance measurements were made by applying a 25 mV RMS sine wave with frequencies varied logarithmically from 10 Hz to 10 kHz. Prior to implantation, measurements were made by immersing the electrode recording sites in 0.1 M phosphate buffer saline (PBS) and a platinum foil was used as the reference electrode. After implantation, a distant stainless steel (316-SS grade) bone screw was used as the reference electrode.

2.5. Statistical Analysis

For this study, comparative statistical significance between groups was determined using standard analysis of variance techniques (ANOVA). There were 24 PEDOT modified sites and 24 control sites on any specific day in the experiment. The factors used in initial comparative ANOVA calculations for any given metric were coated vs. control, animal, and day number. The associated calculated standard deviation has been included in the text with all average measurements.

3. RESULTS/DISCUSSION

3.1. Impedances

The impedances of the electrodes at 1 kHz were used for comparison purposes as action potentials have a characteristic frequency band centered at 1 kHz. Prior to implantation, the 1 kHz impedance of the unmodified gold sites ranged between 6-11 MΩ, with a mean impedance of 9.1±1.4 MΩ (See Table 1). After electrochemical deposition with PEDOT, in vitro site impedances ranged between 0.3 -0.6 MΩ, with a mean impedance of 0.37 MΩ±0.05.

Table 1.

Summary of Results for Groups of Days after Surgery.

Average Impedance (in MΩ) Pk-Pk Noise (in μV) Signal Amplitude (in μV) Units per Site
PEDOT Control PEDOT Control PEDOT Control PEDOT Control
Days 0-2 0.45±0.1 9.2±1.7 35.3±5.3 106.2±8.2 55.2±20.2 N.A. 0.8±0.1 0
Days 3-5 1.31±0.4 10.3±1.9 46.2±8.1 111.3±10.1 84±14.4 N.A. 0.5±0.2 0
Days 6-8 2.21±0.7 11.4±2.2 51.3±9.3 116.8±9.5 95.3±17.1 N.A. 0.4±0.2 0

PEDOT site averages are listed in black, while control site averages are listed in blue.

Over the seven days following implantation, impedance increases were noted in both PEDOT and control sites that were consistent with trends observed in other microelectrode recording studies performed in the rat model12,21-22,25. Figure 2 depicts Bode plots of the average impedance magnitude for PEDOT and control sites at Day 0, Day 4, and Day 7 post surgery. Over the first two days following implantation, the average impedance of the PEDOT sites at 1 kHz slightly increased to 0.45±0.1MΩ, while the average impedance of the control sites also slightly increased to 9.2±1.7 MΩ (p<0.001, See Table 1). Between days three and five post-implantation, a more marked impedance increase at 1kHz was observed on both PEDOT and control sites, to 1.31±0.4 MΩ and 10.3±1.9 MΩ respectively (p<0.001, See Table 1).

Figure 2. Bode plot of average measured impedance versus frequency.

Figure 2

The dotted lines denote standard error of the data set on the given day (n=24). (a) Day 0 post-implantation. (b) Day 4 post-implantation. (c) Day 7 post-implantation. As the immune response to the implant progresses the first seven days from surgery, impedances at the critical 1 kHz frequency increase for both PEDOT and control sites.

During the peak of the early reactive response, spanning from days 6 to 8, the mean impedance of PEDOT sites further increased to 2.21±0.7 MΩ while the mean impedance of the unmodified sites also increased to 11.4±2.2 MΩ (p<0.001, See Table 1).

3.2. Noise

As anticipated, the average peak-to-peak noise levels over the timeframe of this experiment were dramatically larger on control sites (Control Mean: 106.2±8.2 μV, PEDOT Mean: 35.3±5.3 μV). Over the seven days following implantation, the trend in average peak-to-peak noise levels on both PEDOT and control sites paralleled their impedance trends (See Table 1). In the period spanning days three and five post-surgery, the average peak to peak noise level on PEDOT sites increased to 46.2±8.1 μV, while control sites increased to 111.3±10.1 μV. During days six to eight post-surgery, the average peak to peak noise levels on PEDOT and control sites yet again increased, to 51.3±9.3 μV and 116.8±9.5 μV respectively.

Quality unit activity was evident on at least some PEDOT treated sites on all three arrays at every time point in this study (See Table 1). The average number of recorded units for PEDOT sites was highest in the days immediately following surgery (0.8±0.1 average units per site; total number of quality units/total number of sites), and then diminished over the week following surgery (0.4±0.2 average units per site between days 6 and 8 following surgery). A decreasing trend in unit activity in the week following surgery has been noted elsewhere 12,25-26,32 and is hypothesized to be caused by edema/swelling coupled with the initial immune response.

In contrast to the PEDOT sites on the three implanted arrays, no quality unit activity was evident on the control sites at any point during this study (see Table 1). Recordings on control sites were dominated by a large noise floor, which was sufficient to obscure all unit activity on these sites (see Figure 3). The high initial impedance of the untreated sites resulted in a large fluctuation noise contribution, dramatically reducing the likelihood of observing neural activity sufficient in amplitude to be differentiated from the noise floor.

3.4. Contribution of Encapsulation

Inserting a microelectrode into brain tissue elicits a reactive foreign body response, which produces a fibrous encapsulation of the array, effectively creating a high impedance barrier between the microelectrode and the neuron population 33-35. A well-described theoretical circuit model of this phenomenon has been widely accepted 36-39. The tissue encapsulation of the array can be characterized by a sealing resistance, describing protein adsorption and in some cases a layer of connective tissue. In addition, the model incorporates adjacent cellular layers of glia and macrophages given by a membrane capacitance, a membrane resistance, and a membrane area scaling term, m, related to encapsulation thickness and cell-to-cell adhesion within the cellular layer (See Figure 4). The extracellular pathway between cells is defined as a resistance.

Figure 4.

Figure 4

One Individual Element of the Lumped Circuit Model for Encapsulation.

Based on this model, the chronic foreign body response can dramatically increase the effective overall impedance of a chronically implanted electrode. In previous studies describing the changes in electrode impedance of typically-sized cortical microelectrodes following implant, the contribution of the encapsulation response to total electrode impedance became dominant after a few weeks 12,21-22,26. Consequently, increasing the pre-implant surface area using a conductive polymer coating provided limited chronic benefit to neural recordings 12.

In this study of small electrode sites 15 microns in diameter, the obvious benefit of PEDOT coatings to neural recordings was still evident at eight days post implant. This time point is important, as the contribution of encapsulation to electrode impedance typically reaches a maximum one to two weeks post surgery 12,21,28,40, demonstrating that PEDOT coatings enable neural recordings from small electrode sites even when the impedance contribution of encapsulation is at a maximum. These results indicate that as the size of the electrode decreases and total impedance approaches 5 MΩ and beyond, coatings to increase the surface area of the electrode become critical to minimize fluctuation noise and shunt loss.

Although the PEDOT films enabled recordings through 15 micron diameter sites, by day 8 the average number of quality units per PEDOT site had diminished to 0.4 units per site, presumably due to the inherent foreign body response to the electrode. Consequently, future work will focus on utilizing PEDOT coatings to enable smaller multi-electrode arrays that cause less damage upon implantation. Recent studies have indicated that by reducing the size of the implanted electrode, the foreign-body response to the electrode can be minimized, decreasing the contribution of encapsulation to impedance 3-4,41-42 and possibly limiting neuronal death in the vicinity of the electrode array 43. Although PEDOT is one enabling component of manufacturing smaller multi-electrode arrays, a number of technical problems need to be addressed - e.g. insulation requirements, interconnect reliability, crosstalk between adjacent traces, etc. - before these arrays become a practical reality.

4. CONCLUSIONS

The next generation of neural electrodes needs to facilitate more channels of communication to and from the brain over a smaller tissue area. Moreover, these electrodes need to be smaller in order to limit neural damage upon implantation and long-term insult to the surrounding tissue. In this study, PEDOT successfully decreased the impedance of previously unusable small gold electrode sites to within viable neural recordings range. PEDOT coatings help to alleviate a major electrode design constraint, enabling the fabrication of ultra-small, high-channel count arrays to interface chronically with the brain.

ACKNOWLEDGEMENTS

The authors of this paper would like to acknowledge all of the members of the Neural Engineering Laboratory at the University of Michigan for their assistance in this study.

GRANTS

This work was supported by the NIH P41 Center for Neural Communication Technology (EB002030) and the Whitaker Foundation.

This work was supported by the Center for Wireless Integrated Microsystems NSF EEC-9986866, the NIH P41 Center for Neural Communication Technology (EB002030), and the Whitaker Foundation

Footnotes

DISCLOSURES

Daryl R. Kipke has significant financial interest in NeuroNexus Technologies, a leading supplier of microelectrode technology. Nicholas B. Langhals was a consultant for NeuroNexus Technologies. Sarah Richardson-Burns and Jeffrey Hendricks have significant financial interest in Biotectix, a company that specializes in conductive polymer coatings for electrodes.

REFERENCES

  • 1.Kipke DR, et al. Advanced neurotechnologies for chronic neural interfaces: new horizons and clinical opportunities. J Neurosci. 2008;28:11830–11838. doi: 10.1523/JNEUROSCI.3879-08.2008. doi:28/46/11830 [pii] 10.1523/JNEUROSCI.3879-08.2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bernatchez SF, Parks PJ, Gibbons DF. Interaction of macrophages with fibrous materials in vitro. Biomaterials. 1996;17:2077–2086. doi: 10.1016/0142-9612(96)00014-2. doi:0142961296000142 [pii] [DOI] [PubMed] [Google Scholar]
  • 3.Sanders JE, Stiles CE, Hayes CL. Tissue response to single-polymer fibers of varying diameters: evaluation of fibrous encapsulation and macrophage density. J Biomed Mater Res. 2000;52:231–237. doi: 10.1002/1097-4636(200010)52:1<231::aid-jbm29>3.0.co;2-e. doi:10.1002/1097-4636(200010)52:1<231::AID-JBM29>3.0.CO;2-E [pii] [DOI] [PubMed] [Google Scholar]
  • 4.Seymour JP, Kipke DR. Neural probe design for reduced tissue encapsulation in CNS. Biomaterials. 2007;28:3594–3607. doi: 10.1016/j.biomaterials.2007.03.024. [DOI] [PubMed] [Google Scholar]
  • 5.Najafi K, Ji J, Wise KD. Scaling limitations of silicon multichannel recording probes. Biomedical Engineering, IEEE Transactions. 1990;37:1–11. doi: 10.1109/10.43605. [DOI] [PubMed] [Google Scholar]
  • 6.Kovacs GTA. In: Enabling Technologies for Cultured Neural Networks. Stenger DA, McKenna T, editors. Academic; London: 1994. pp. 121–165. [Google Scholar]
  • 7.Schmidt E, Humphrey DR. Neurophysiological techniques. Vol. 2. Humana Press; 1990. pp. 1–64. [Google Scholar]
  • 8.Shoham S, Nagarajan S. In: Neuroprosthetics: Theory and Practice. Horch Kenneth W., Dhillon Gurpreet S., editors. World Scientific Publishing; 2003. pp. 448–465. [Google Scholar]
  • 9.Robinson DA. The electrical properties of metal microelectrodes. Proceedings of the IEEE. 1968;56:1065. [Google Scholar]
  • 10.Hassibi A, Navid R, Dutton RW, Lee TH. Comprehensive study of noise processes in electrode electrolyte interfaces. Journal of Applied Physics. 2004;96:9. [Google Scholar]
  • 11.Hetke JF, Lund JL, Najafi K, Wise KD, Anderson DJ. Silicon ribbon cables for chronically implantable microelectrode arrays. Biomedical Engineering, IEEE Transactions. 1994;41:314–321. doi: 10.1109/10.284959. [DOI] [PubMed] [Google Scholar]
  • 12.Ludwig KA, Uram JD, Yang J, Martin DC, Kipke DR. Chronic neural recordings using silicon microelectrode arrays electrochemically deposited with a poly(3,4-ethylenedioxythiophene) (PEDOT) film. Journal of Neural Engineering. 2006;3:59. doi: 10.1088/1741-2560/3/1/007. [DOI] [PubMed] [Google Scholar]
  • 13.Cogan SF. Neural stimulation and recording electrodes. Annu Rev Biomed Eng. 2008;10:275–309. doi: 10.1146/annurev.bioeng.10.061807.160518. doi:10.1146/annurev.bioeng.10.061807.160518. [DOI] [PubMed] [Google Scholar]
  • 14.Keefer EW, Botterman BR, Romero MI, Rossi AF, Gross GW. Carbon nanotube coating improves neuronal recordings. Nat Nanotechnol. 2008;3:434–439. doi: 10.1038/nnano.2008.174. doi:nnano.2008.174 [pii] 10.1038/nnano.2008.174. [DOI] [PubMed] [Google Scholar]
  • 15.Abidian MR, Corey JM, Kipke DR, Martin DC. Conducting-polymer nanotubes improve electrical properties, mechanical adhesion, neural attachment, and neurite outgrowth of neural electrodes. Small. 2010;6:421–429. doi: 10.1002/smll.200901868. doi:10.1002/smll.200901868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Richardson-Burns SM, et al. Polymerization of the conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) around living neural cells. Biomaterials. 2007;28:1539–1552. doi: 10.1016/j.biomaterials.2006.11.026. doi:S0142-9612(06)00997-5 [pii] 10.1016/j.biomaterials.2006.11.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Wilks SJ, Richardson-Burns SM, Hendricks JL, Martin DC, Otto KJ. Poly(3,4-ethylenedioxythiophene) as a Micro-Neural Interface Material for Electrostimulation. Front Neuroengineering. 2009;2:7. doi: 10.3389/neuro.16.007.2009. doi:10.3389/neuro.16.007.2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Cui X, Martin DC. Electrochemical deposition and characterization of poly(3,4-ethylenedioxythiophene) on neural microelectrode arrays. Sensors and Actuators. 2003;B 89:92–102. [Google Scholar]
  • 19.Richardson-Burns SM, Hendricks JL, Martin DC. Electrochemical polymerization of conducting polymers in living neural tissue. J Neural Eng. 2007;4:L6–L13. doi: 10.1088/1741-2560/4/2/L02. doi:S1741-2560(07)30000-1 [pii] 10.1088/1741-2560/4/2/L02. [DOI] [PubMed] [Google Scholar]
  • 20.Abidian MR, Martin DC. Experimental and theoretical characterization of implantable neural microelectrodes modified with conducting polymer nanotubes. Biomaterials. 2008;29:1273–1283. doi: 10.1016/j.biomaterials.2007.11.022. doi:S0142-9612(07)00961-1 [pii] 10.1016/j.biomaterials.2007.11.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Abidian MR, et al. Interfacing Conducting Polymer Nanotubes with the Central Nervous System: Chronic Neural Recording using Poly(3,4-ethylenedioxythiophene) Nanotubes. Advanced Materials. 2009;21:8. doi: 10.1002/adma.200900887. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Vetter RJ, Williams JC, Hetke JF, Nunamaker EA, Kipke DR. Spike recording performance of implanted chronic silicon-substrate microelectrode arrays in cerebral cortex. IEEE Transactions on Neural Systems and Rehabilitation Engineering. 2004;52 [Google Scholar]
  • 23.Henze DA, et al. Intracellular features predicted by extracellular recordings in the hippocampus in vivo. J Neurophysiol. 2000;84:390–400. doi: 10.1152/jn.2000.84.1.390. [DOI] [PubMed] [Google Scholar]
  • 24.Yang J, et al. Ordered surfactant-templated poly(3,4-ethylenedioxythiophene) (PEDOT) conducting polymer on microfabricated neural probes. Acta Biomaterialia. 2005;1:125–136. doi: 10.1016/j.actbio.2004.09.006. [DOI] [PubMed] [Google Scholar]
  • 25.Ludwig KA, et al. Using a common average reference to improve cortical neuron recordings from microelectrode arrays. J Neurophysiol. 2009;101:1679–1689. doi: 10.1152/jn.90989.2008. doi:90989.2008 [pii] 10.1152/jn.90989.2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Purcell EK, Thompson DE, Ludwig KA, Kipke DR. Flavopiridol reduces the impedance of neural prostheses in vivo without affecting recording quality. J Neurosci Methods. 2009 doi: 10.1016/j.jneumeth.2009.06.026. doi:S0165-0270(09)00344-6 [pii] 10.1016/j.jneumeth.2009.06.026. [DOI] [PubMed] [Google Scholar]
  • 27.Langhals NB, Kipke DR. Validation of a novel three-dimensional electrode array within auditory cortex. Conf Proc IEEE Eng Med Biol Soc. 2009;2009:2066–2069. doi: 10.1109/IEMBS.2009.5333958. doi:10.1109/IEMBS.2009.5333958. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Purcell EK, Thompson DE, Ludwig KA, Kipke DR. Flavopiridol reduces the impedance of neural prostheses in vivo without affecting recording quality. J Neurosci Methods. 2009;183:149–157. doi: 10.1016/j.jneumeth.2009.06.026. doi:S0165-0270(09)00344-6 [pii] 10.1016/j.jneumeth.2009.06.026. [DOI] [PubMed] [Google Scholar]
  • 29.Rohatgi P, Langhals NB, Kipke DR, Patil PG. In vivo performance of a microelectrode neural probe with integrated drug delivery. Neurosurg Focus. 2009;27:E8. doi: 10.3171/2009.4.FOCUS0983. doi:10.3171/2009.4.FOCUS0983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Harris KD, Henze DA, Csicsvari J, Hirase H, Buzsaki G. Accuracy of tetrode spike separation as determined by simultaneous intracellular and extracellular measurements. J Neurophysiol. 2000;84:401–414. doi: 10.1152/jn.2000.84.1.401. [DOI] [PubMed] [Google Scholar]
  • 31.Lewicki MS. A review of methods for spike sorting: the detection and classification of neural action potentials. Network-Computation in Neural Systems. 1998;9:R53–R78. [PubMed] [Google Scholar]
  • 32.Santhanam G, et al. HermesB: a continuous neural recording system for freely behaving primates. IEEE Trans Biomed Eng. 2007;54:2037–2050. doi: 10.1109/TBME.2007.895753. doi:10.1109/TBME.2007.895753. [DOI] [PubMed] [Google Scholar]
  • 33.Szarowski DH, et al. Brain responses to micro-machined silicon devices. Brain Research. 2003;983:23–35. doi: 10.1016/s0006-8993(03)03023-3. [DOI] [PubMed] [Google Scholar]
  • 34.Turner JN, et al. Cerebral astrocyte response to micromachined silicon implants. Exp Neurol. 1999;156:33–49. doi: 10.1006/exnr.1998.6983. [DOI] [PubMed] [Google Scholar]
  • 35.Liu X, et al. Stability of the interface between neural tissue and chronically implanted intracortical microelectrodes. IEEE transactions on rehabilitation engineering. 1999;7:315–326. doi: 10.1109/86.788468. [DOI] [PubMed] [Google Scholar]
  • 36.Buitenweg JR, Rutten WL, Willems WP, van Nieuwkasteele JW. Measurement of sealing resistance of cell-electrode interfaces in neuronal cultures using impedance spectroscopy. Medical & biological engineering & computing. 1998;36:630–637. doi: 10.1007/BF02524436. [DOI] [PubMed] [Google Scholar]
  • 37.Grill WM, Mortimer JT. Electrical properties of implant encapsulation tissue. Annals of biomedical engineering. 1994;22:23–33. doi: 10.1007/BF02368219. [DOI] [PubMed] [Google Scholar]
  • 38.Otto KJ, Johnson MD, Kipke DR. Voltage pulses change neural interface properties and improve unit recordings with chronically implanted microelectrodes. Biomedical Engineering, IEEE Transactions. 2006;53:333–340. doi: 10.1109/TBME.2005.862530. [DOI] [PubMed] [Google Scholar]
  • 39.Williams JC, Hippensteel JA, Dilgen J, Shain W, Kipke DR. Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants. J Neural Eng. 2007;4:410–423. doi: 10.1088/1741-2560/4/4/007. [DOI] [PubMed] [Google Scholar]
  • 40.Vetter RJ, Williams JC, Hetke JF, Nunamaker EA, Kipke DR. Chronic neural recording using silicon-substrate microelectrode arrays implanted in cerebral cortex. IEEE transactions on bio-medical engineering. 2004;51:896–904. doi: 10.1109/TBME.2004.826680. [DOI] [PubMed] [Google Scholar]
  • 41.Chen CS, Mrksich M, Huang S, Whitesides GM, Ingber DE. Geometric control of cell life and death. Science. 1997;276:1425–1428. doi: 10.1126/science.276.5317.1425. [DOI] [PubMed] [Google Scholar]
  • 42.Turner AM, et al. Attachment of astroglial cells to microfabricated pillar arrays of different geometries. J Biomed Mater Res. 2000;51:430–441. doi: 10.1002/1097-4636(20000905)51:3<430::aid-jbm18>3.0.co;2-c. doi:10.1002/1097-4636(20000905)51:3<430::AID-JBM18>3.0.CO;2-C [pii] [DOI] [PubMed] [Google Scholar]
  • 43.Biran R, Martin DC, Tresco PA. Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Exp Neurol. 2005;195:115–126. doi: 10.1016/j.expneurol.2005.04.020. [DOI] [PubMed] [Google Scholar]

RESOURCES