Medical device vigilance systems: India, US, UK, and Australia

Pooja Gupta 1, Manthan D Janodia 1,, Puralea C Jagadish 1, Nayanabhirama Udupa 1
PMCID: PMC3417867  PMID: 22915923

Abstract

The term medical device includes a wide category of products ranging from therapeutic medical devices exerting their effects locally such as tissue cutting, wound covering or propping open clogged arteries, to highly sophisticated computerized medical equipment and diagnostic medical devices. To achieve uniformity among the national medical device regulatory systems and increase the access to safe, effective, and clinically beneficial medical technologies, the Global Harmonization Task Force (GHTF) was conceived in 1992 by five members: European Union, United States, Australia, Japan, and Canada. All regulated countries have clearly defined medical devices, as has the GHTF. Although GHTF has tried to achieve harmonization with respect to medical devices, some differences still exist in the national laws of the countries of GHTF. Further, regulated countries have classified medical devices on the basis of their associated risk. In the Indian regulatory system, medical devices are still considered as drugs. In 2006, the Medical Device Regulation Bill was recommended to consolidate laws for medical devices and to establish the Medical Device Regulatory Authority of India. In addition, medical devices are not classified by any Indian regulatory authority. Although India has moved towards harmonizing its medical device regulations with those of regulated countries, this study aims to identify whether India should have a vigilance system in harmony with those of GHTF or develop its own system for medical devices.

Keywords: medical device, vigilance, regulatory systems, GHTF, India

Introduction

The term “medical device” includes a broad category of products ranging from therapeutic medical devices with local applications, such as tissue cutting, wound covering, or propping open clogged arteries, to highly sophisticated computerized medical equipment and diagnostic medical devices. Because these devices vary widely in type and are highly essential for patients’ care, their manufacture, distribution, and sale must be regulated to ensure their quality, safety, and efficacy.1

The postmarket surveillance of medical devices was initiated in the United States with the enactment of the Food and Drug Administration Modernization Act 1970 under Section 522 for class II and class III devices.2 In 1989, the Therapeutic Goods Act provided the legislative basis for uniform national controls over goods used in the prevention, diagnosis, curing, or alleviation of a disease, ailment, defect, or injury in Australia.3 In June 1993, the vigilance requirement for medical devices for member states and manufacturers was published as Council Directive 90/385/EEC and 93/42/ EEC, followed by incorporation of amendments of revision 5 of MEDDEV guidance 2.12–1 [1] in 2007 by the European Union.47 To achieve uniformity between the national medical device regulatory systems and to increase the access to safe, effective, and clinically beneficial medical technologies, the Global Harmonization Task Force (GHTF) was conceived in 1992 by five members: European Union, United States, Australia, Japan, and Canada wherein the vigilance of devices was among the study groups.8

All regulated countries have distinctly defined medical devices, but GHTF defined a medical device as any instrument, apparatus, implement, machine, appliance, implant, in vitro reagent or calibrator, software, material, or other similar or related article, which is thereby intended to be used by the manufacturer for human beings for one or more of the specific purposes of:1

  1. Diagnosis, prevention, monitoring, treatment, or alleviation of disease or compensation for an injury

  2. Investigation, replacement, modification, or support of the anatomy or of a physiological process

  3. Supporting or sustaining life

  4. Control of conception

  5. Disinfection of medical devices

  6. Providing information for medical purposes by means of in vitro examination (such as reagents, calibrators, sample collection kits, control materials, and related instruments) of specimens derived from the human body and which does not achieve its primary intended action in or on the human body by pharmacological, immunological, or metabolic means, but which may be assisted in its function by such means.

The Medicines and Healthcare products Regulatory Agency (MHRA)9 has excluded materials used for disinfecting medical devices from the definition, whereas the Therapeutic Goods and Administration (TGA)3 has excluded tampons and hospital, household, and commercial grade disinfectants. However, to date, India has considered medical devices as drugs.10

Scope and objective

This study was the result of the examination and comparison of regulations to monitor medical devices in national regulatory authorities: United States, United Kingdom, and Australia (There were 5 founding members of GHTF). The issue of whether India should establish a vigilance system in harmony with those of the GHTF members or start with a clean slate was also taken into account. Hence, the study was initiated to examine the proposed framework, and its implication, for the Indian medical device vigilance system with respect to global harmonization, ie, GHTF.

Plan of study

During this study, the information was collected through secondary media, ie, the official Websites of the respective regulatory authorities and the knowledge gained by interaction with various industrial professionals in the field of regulatory affairs.

Indian regulatory system

In the Indian regulatory system, medical devices are considered as drugs by the Ministry of Health and Family Welfare.10 In October 2005, the manufacturing of devices came under the control of the Central Licensing Approving Authority. It declared 10 devices, such as cardiac stents, drug-eluting stents, catheters, intraocular lenses, bone cements, heart valves, scalp vein sets, and so on, to be considered drugs and included another 19 sterile medical devices (on March 20, 2009 [on hold]) such as extension tubes, arterial venous fistulas and spinal needles, volume measuring sets, heart lung packs, and so on, under the provisions as such.11

In 2006, the Medical Device Regulation Bill12 was recommended to consolidate laws for medical devices and to establish the Medical Device Regulatory Authority of India for establishing and maintaining a national system of controls relating to quality, safety, efficacy, and availability of medical devices that are used in India, whether produced in India or elsewhere and exported from India. It was recommended that the provisions of this Act should come into force by December 31, 2009, but the bill was not passed by the Rajya Sabha and and the bill was abolished. It was expected that a definition of medical devices and other likely changes shall be incorporated in Schedule M(III) by 2010. With the initiation of such amendments, it was addressed to include medical device with the Drugs and Cosmetics Act and Rules in a recent press release from the Medical Device Regulatory Authority of India.13 In India, the Federation of Indian Chambers of Commerce and Industry (FICCI) is being recognized as the focal point for regulations of medical device. It has been working closely with the Central Drug Standard Control Organization (CDSCO) and Indian medical regulators (both importer and native manufacturers) to increase access to medical device, promoting its manufacturing and streamlining the regulatory process toward global harmonization.14

If the product is being imported to India, the importer has to submit postmarket surveillance data including protocol and report not exceeding 5 years for the following:10

  1. Procedures for distribution of records

  2. Complaint handling

  3. Adverse incident reporting

  4. Procedure for product recall

However, the requirements specified do not require the manufacturer to provide the data for a specified period, but the manufacturers have been required to submit postmarket surveillance data for at least a period of 3–5 years. Recently, the CDSCO has issued the guidelines for Adverse Event Reporting (ADR).15

Medical device classification

Each regulatory authority has classified medical devices in its own way. In general, the basis for medical device classification is:16

  1. The risk associated with the medical device

  2. Manufacturers’ intended purpose for the device

  3. The device’s indications for use

The US Food and Drug Administration (FDA)17 has classified medical devices into three classes on the basis of level of control necessary to assure the safety and effectiveness of the device and on the basis of information about marketing requirements and has grouped them into 16 medical specialties:

  1. Class I (general controls): eg, elastic bandage and examination gloves

  2. Class II (general and special controls): eg, infusion pumps, surgical drapes, and ultrasound imaging systems

  3. Class III (general controls and premarket approval): eg, heart valves, and silicone gel-filled breast implants

The TGA18 has classified medical devices into five classes on the basis of set of four classification rules: noninvasive, invasive, active, and special type of devices, which are:

  1. Class I (low risk): eg, examination lights and surgical microscopes

  2. Class IIa (low–medium risk): eg, warming blankets and electrical acupuncture

  3. Class IIb (medium–high risk): eg, infant incubators and external defibrillators

  4. Class III (high risk): eg, heparin-coated catheters and biological heart valves

  5. Class active implantable medical device (AIMD): eg, contraceptive intrauterine devices

Likewise, the MHRA4,5,19 has classified medical devices into three groups:

  1. General medical devices

  2. AIMD

  3. In vitro diagnostic medical devices

The general medical devices were then further classified into four classes as described in Table 1.

Table 1.

MHRA classification of general medical devices

Class of device Risk level Requirements Example
Class I Low risk Premarket notification Dressings
Class IIa Low–medium risk Certification by notified body X-ray film
Class IIb Medium–high risk Certification by notified body Blood bags, contact lens care products
Class II High risk Certification by notified body Bone cement, cardiac stents

Also, the MHRA classified medical devices within a series of 18 rules depending on functions, parts of body treated, and properties of medical devices. If more than one rule applies to a medical device, the higher would be its classification.

In comparison to the regulated countries, medical devices in India are not classified on the basis of risk. Rather, the 10-device category of medical devices has been notified to be regulated as drugs.10

Medical device tracking

Once the medical devices are marketed, they must be tracked up to the end-user for the useful life of the device to facilitate patient notification or recall at the time of any defect or problem with the device. The FDA2,2023 has included medical device tracking as one of its postmarket surveillance activities, but tracking is still in the development stage for inclusion in the TGA vigilance system.2426 For tracking of patients with implantable medical devices, an Implantable Medical Device Tracking Subcommittee (IMDTS) has been developed by TGA. However, the European Medicines Agency has developed the Adverse Incident Tracking System (AITS).

The devices require tracking if they meet statutory requirements according to the FDA act and if the FDA has issued a tracking order. It has also listed some implantable devices that are subjected to tracking. Furthermore, it requires submission of tracking information by the manufacturer, distributor, and user facility. The manufacturer must provide critical information about the devices within 3 or 10 days if a device has not or has been distributed to a patient, respectively. The written records for the useful life of device must also be maintained by the manufacturer and the distributor.

However, in revised MHRA9 directives, incidents reported with the device are entered into the AITS database via a user-reporting system. Each incident is given an AITS reference number, and the risk associated with the device is assessed. The adverse incident is then accordingly allocated into one of the five investigation categories, ie, urgent in depth, in depth, standard, information, and others. It is the responsibility of manufacturer or authorized representative to track incident reports, but the information to be tracked is not being specified. In-depth investigations lead to the issue of a Medical Device Alert, whereas standard investigations are initially conducted via a series of standardized letters issued directly by the adverse incident center.

In India, the labeling provisions for devices include identification of lot number or batch number for easy traceability of the device, but maintenance of the tracking records by the manufacturer is not a mandatory requirement. However, in a recent press release by the FICCI,13 the Health Minister Dinesh Trivedi had promised to set up a National Health Portal for sharing information in the public domain on standardization and protocols and to ensure that the medical records of all citizens are electronically stored for ease of access by pathologists and doctors for diagnosis and treatment of patients. The statement seemed to represent promising progress toward the establishment of a tracking portal for medical devices.

Adverse event reporting

The adverse event reporting system has been considered as a tool to improve and protect the health and safety of patients and users, thereby reducing the likelihood of adverse events, to prevent repetition of adverse events, or to alleviate consequences of such repetition. Despite the steps taken by the GHTF to achieve uniformity among adverse event reporting systems, the system still differs in founding member countries.

The GHTF27 provided the guidance on mandatory reporting of adverse events for device manufacturers and voluntary reporting for users. Also, the manufacturer must decide the reportability of adverse events on the basis of information available. The FDA’s medical device reporting (MDR) regulations2831 require manufacturers and importers to report serious injuries, deaths, and malfunctions, and user facilities to report serious injuries and deaths. The regulations do not require foreign manufacturer to meet FDA regulations, but if employed, the agent could forward reports and will be viewed as an employee of a foreign firm. The TGA24,32 has encouraged the reporting of adverse events not only by the manufacturers and sponsors but also by users. The EEC9,33,34 directives include requirements not only for manufacturers or authorized representatives of medical device to report certain types of incidents to a competent authority such as the MHRA, but also for the MHRA to disseminate the information to other competent authorities and the EEC. Also, in India, CDSCO13 considers that an adverse event should be reported by the manufacturer.

Reporting criteria

The reporting criteria include the following:

  • If an event has occurred, and the manufacturer becomes aware of the information

  • If it is assessed that the manufacturer’s device is associated with the event based on the opinion from the available information

  • If the event has led to or might have led to death or serious injury of a patient, user, or other person.

However, the FDA2831 also requires the manufacturer to report malfunctions and events occurring due to user errors. The TGA24,32 also requires the sponsor or manufacturer to report malfunctions. In addition to the above reporting criteria, a manufacturer in India15 must also report events that do not require to be reported under regulations, so that trends or patterns of their occurrence can be monitored.

Not-reportable incidents or events

The regulated countries, along with India,15 defined not-reportable events similarly with few exceptions. The following events2745 are exempted from reporting in all countries:

  • If the deficiency of a device is found by the user before its use and no serious injury has occurred

  • If the root cause of the adverse event is due to a patient’s pre-existing condition

  • If the shelf life or service life of the device was exceeded before its use by a patient

  • If the design feature for protection against malfunction complied with the relevant standards and operated correctly

  • If the deficiency had a negligible likelihood of causing death or serious injury and had been established and documented as acceptable after risk assessment

  • If the side effects are expected and foreseeable from the manufacturer’s labeling, are clinically well known, and are documented in the device master record, with an appropriate risk assessment

  • If the adverse event was caused by errors of use and abnormal use

As mentioned above, the manufacturer is required to report adverse events occurring in Australia24,32 and India if they occur after the issuance of an advisory notice, and their occurrence and frequency are submitted with the annual report, whereas in the United States,35 the manufacturer can request remedial action exemption (RAE) with or after the submission of one or more 5-day or 30-day initial reports on events. The RAE can be requested if the information received is erroneous; the malfunction does not result in death, serious injury, or other adverse events; death or serious injury did not occur; and another manufacturer has made the device.

Reporting time frame

Although the reportability of the adverse event is uncertain, the GHTF27,45 requires immediate reporting by the manufacturer of unanticipated death, serious injury, or public health threat and all other reportable events as soon as possible, but no later than 30 elapsed calendar days following the date of awareness of the event. Also, if the adverse events are reported within the first 2 weeks based on an incomplete investigation, it may require a subsequent follow-up report.

The FDA2831 requires reporting of events not only by the manufacturer, whether domestic or foreign, but also by the user facility and distributor. The manufacturer must submit four reports depending on the event reported: first, 30-day reports for death, serious injury, or malfunctions; second, 5-day reports for events requiring immediate remedial action (FDA form 3500A); third, baseline report (FDA form 3417) to provide basic data on the device, subject to MDR report (30 or 5 days); and finally, annual certification (FDA form 3381). The user facility and distributor need to report death and serious injury within 10 working days on FDA form 3500A. The user facility is also required to submit semi-annual reports to FDA (form 3419) from January 1 to July 1. Eventually, form 3500A should report events for each device, and only one report must be submitted for the same patient or same event irrespective of the multiple sources of information for an event. Voluntary reports may be submitted at any time for adverse events, product problems, and product use errors through completion of online MedWatch FDA form 3500.

In Australia,24,32 although it is the manufacturer who must assess an incident, the sponsor is held accountable for forwarding information about events to the manufacturer and then for forwarding the results of any analysis to the TGA. Hence, the reporting time frame for reporting adverse events is as follows:

  1. Sponsor or manufacturer: they must report on form MDIR01 the following events.

    1. Death or serious deterioration in the state of health within 10 calendar days

    2. Near adverse event or event that did not result in death or serious injury within 30 calendar days

    3. Serious public health threat or concern and an event that requires prompt remedial action within 48 hours

    4. Complaints, malfunctions, and adverse events for class III and AIMD devices that have been reported for the year July 1 and October 1 following the entry of device into the Australian Register of Therapeutic Goods (ARTG) and for the subsequent 3 years

  2. Voluntary reporting by users of a suspected health hazard created by a medical device, on form UDIR01.

In certain cases, if all information is not available, the sponsor or manufacturer must provide it later when it is, as additional information along with a statement to the effect that the report is made by the manufacturer and sponsor without prejudice and does not imply any admission of liability for the incident or its consequences. All reports should be submitted to the Coordinator of the Medical Device

Incident Report Investigation Scheme of the TGA. In United Kingdom,33,34 the manufacturer is required to report within the time frame relating to the type of incident upon becoming aware that an event has occurred and one of its devices has caused or contributed to the incident, ie,

  1. Serious public threat within two calendar days after the date of awareness

  2. Death or serious deterioration in state of health within 10 elapsed calendar days after the date of awareness

  3. Other incidents, immediately after assessing the link between the device and the event within 30 elapsed calendar days

  4. Manufacturer’s written acknowledgment of user reports from MHRA to manufacturer within three working days of receiving user report

  5. Voluntary reports may be submitted at any time, and may be on the events other than death, serious injury, or malfunction as defined

The manufacturer’s incident report form should be used for initial, follow-up, and final incident reports. If the report is made orally, it should be followed by a written report as soon as possible and a statement to the effect that the report is made by the manufacturer without prejudice and does not imply any admission of liability for the incident or its applicability. The manufacturer or authorized representative must submit an initial incident report to MHRA for record and evaluation, followed by final reports, which should not be delayed by incomplete information. Incidents should preferably reported on form MORE, ie, MHRA’s Manufacturers’ On-line Reporting Environment.

According to the recent amendments directed by the CDSCO,15 the manufacturer is liable to report unanticipated death or serious injury or a serious public health threat within 10 days of becoming aware of the event, and all other reportable events not later than 30 elapsed calendar days. Because of the uncertainty of the reportable events, the manufacturer must report within the expected time frame.

Types of reports

The reports being submitted are determined on the grounds of stipulated reporting time frames. Thus, GHTF27,45 has not recommended any report types for manufacturers. The type of report to be submitted differs among the regulated countries, as the reporting time frame differs. The FDA2831 has determined the manufacturer is required to submit five MDR reports, ie,

  1. 30-day report: must contain information on occurrence of death, serious injury, malfunction, and reporting of events via complaint information.

  2. 5-day report: is generally submitted for severe, unusual, or unexpected events and those events that require immediate remedial action or for which FDA has made a written request.

  3. Baseline report: must accompany the corresponding form 3500A when an event involving the device model or device family is reported for the first time, providing basic device identification information. The report can be submitted by model type (one baseline report for each model) or by device family (one baseline report for all models in that family).

  4. Supplemental report: must be submitted using form 3500A within 1 month (30 calendar days) after receipt of the additional information. The number of follow-ups would depend on the nature and severity of the event reported.

  5. Annual certification: the manufacturer must designate more than one certifying official, each of whom should sign a certification statement for his or her identified organizational component or site. The certification is required to minimize the unintentional reporting errors that have been submitted during the 12-month period. The annual certification has to be submitted during the firm’s annual registration date.

However in United Kingdom,33,34 the manufacturer or authorized representative can report incidents by submission of following reports:

  1. Initial reporting of adverse event to MHRA for record and evaluation

  2. Periodic summary reporting: these reports are submitted in an agreed format and frequency for the device and incident between the manufacturer and MHRA after submission of one or more initial reports.

  3. Trend reports: these reports must be submitted when there is a significant increase in the rate of already reportable events, incidents that are usually exempted from reporting, and events that are usually not reportable.

  4. Final reports

All device complaints, potential use error events, and abnormal use events must be reported and investigated by the manufacturer. The manufacturer must perform the investigation after the initial report in consultation with the user while informing the MHRA of the progress through periodic reports or trend reports as appropriate. If initial assessment involves an alteration of the device that might affect subsequent analysis, the MHRA must be informed before proceeding with alteration.

The TGA24,32 also has not determined the distinct report types for reporting adverse events by the manufacturer, whereas CDSCO15 in India has determined notification reports as initial, final, and/or trend report. The choice of report type depends on the availability of applicable data within the appropriate report time.

Vigilance reporting

The dissemination of information among the National Competent Authorities (NCAs) can help to improve the health and safety of patients, users, and others by reducing the likelihood of repeated similar adverse events. The GHTF3638 defines a vigilance exchange program as a National Competent Authority Report (NCAR) exchange program, which aims at exchanging two types of information: information that may be considered highly sensitive or confidential, and selected public (nonconfidential) information. The countries participating in GHTF receive guidance on how to exchange NCARs. The necessity for NCAR exchange would depend on the seriousness of the incident associated with the device and its extent of global distribution. The criteria for NCAR exchange are as follows:

  1. Seriousness

  2. Unexpectedness of the incident or event

  3. Population that is vulnerable (infants or elderly)

  4. Preventability (can useful recommendations be made?)

  5. Public concern or outrage, eg, lead aprons containing radioactive material

  6. Benefit/risk ratio

  7. Lack of scientific data, especially long-term effects

  8. Repeated device problems that resurface, eg, heating pads or fires

  9. Written notifications by the NCA to the public (eg, hospitals, physicians)

  10. Will active exchange help protect public health or have the manufacturer’s actions been sufficient?

The GHTF has posted NCAR forms and reports that could be used for exchanging information directly to the NCAR Secretariat for appropriate global distribution. However, the FDA does not specify a vigilance exchange program as one of its postmarket activities.

In Australia,24,32 the NCAR program is obliged to exchange vigilance information with overseas regulatory agencies through its participation in the GHTF and various mutual recognition agreements. Consequently, the information will be exchanged on incidents and events for which corrective action, including recalls, is to be taken and there is a serious risk to the safety of patients or other users, but where no corrective action has yet been established, although measures are under consideration, or where there is not yet a final report from the sponsor.

The dissemination of information to other NCAs is termed NCAR in MHRA.9,33,34 The information should be copied to the Commission and disseminated to other NCAs to help prevent a recurrence of incidents, depending on the outcome of investigation carried out by the manufacturer and the limitation of their consequences. The information should be disseminated when field safety corrective actions (FSCA) have been performed by the manufacturer or requested by MHRA; changes already initiated in FSCA are requested by MHRA, but corrective action has not yet been taken yet and is under consideration, although there is serious risk to the health of patient/user; and the manufacturer has not provided the final report in a timely manner.

The NCAR should be distributed without delay, but no later than 14 calendar days after the manufacturer has reported, and the manufacturer’s report must also be circulated along with the NCAR. In spite of communicating the information to the public, it is recommended to communicate notification directly to the medical practitioner or health professional, unless required.

Records

Records provide a history of each batch of product, including its distribution, and also of all other relevant circumstances pertinent to the quality of the final product.9,2224 The records should be made or completed at the time each action is taken in such a way that all significant activities concerned are traceable.

Manufacturers are required to establish and maintain written procedures for implementation of the MDR regulation, including the following:

  1. Evaluated information that determines the reportability of an event

  2. All reports and information of the medical device (MDR) submitted to FDA in either written or electronic form

  3. Any information that was evaluated during the preparation of annual certification report(s)

  4. Systems that ensure access to information that facilitates timely follow-up and inspection by FDA

  5. Investigation protocol that would be followed.

The records related to an event (whether reportable or not) must be maintained for 2 years from the date of the event or a period equivalent to the expected life of the device, whichever is longer.

The sponsors marketing medical devices in Australia are required to keep distribution records and retain them for 5 years after the last product has been made and to provide the records, or copies of the records, when requested by the TGA. The sponsors should maintain records for the products manufactured by them in accordance with the following:

  1. A system should be in operation whereby the complete and up-to-date histories of all batches of components from the starting materials to the finished products can be progressively recorded.

  2. The system should allow the determination of utilization and disposal of device components.

All sponsors should maintain records of problem reports received about each device and problem reports evaluated by competent personnel, and the appropriate action taken should also be shown in the records.

For MHRA, the manufacturers must establish a documented procedure for a feedback system and for operating in compliance with the standards of a “Medical device Quality Management System”. The documented procedure must aim at providing the following:

  1. Early warnings of quality problems and implementation of timely corrective and preventive actions

  2. Records of adverse incident reports received from any source for the device associated with the incident. The procedure must also detail the evaluation of each report and the consequent action taken.

  3. Review of customer or user complaints, such as customer or user surveys, literature reviews, postmarket clinical follow-up, and so on, depending on the nature of device

  4. Records of manufacturing by date and batch or serial number and supply of medical devices directly to users and distributors

  5. For active implantable devices, the code should be recorded.

These records should be retained throughout the specified lifetime of the device for which an FSCA may be necessary.

Enforcement actions

To ensure the supply of safe, efficacious, and quality medical devices in the Australian market and establishment and maintenance of national systems of controls, the vigilance program in TGA39,40 has incorporated the following enforcement actions:

  1. Offences and penalties, eg, failure to notify adverse events (section 41MP), with maximum penalties of A$44,000 for individuals and A$220,000 for corporations

  2. Suspension or cancellation of the medical device from the ARTG (parts 4–6 of the act)

  3. Recall of medical devices supplied to either batch level or all medical devices

However, the device can be suspended for a maximum period of 6 months, which can be extended for a further 6 months (maximum) if the TGA is satisfied that the manufacturer has demonstrated they are correcting the issue that led to the suspension. If this extension period expires, then the entry on the ARTG is automatically cancelled. The TGA will cancel the products from the register only in those cases where there has been a severe breach of the law or, more often, where there is a safety concern associated with the use of the product. The notice of the cancellation will be published in the Commonwealth Gazette.

For the compliance with UK medical device regulations, MHRA33,34 investigates any complaints about the products marked or not marked with the CE mark, inspects the manufacturer’s facilities in case of breach of regulations, and investigates the results of vigilance reports. MHRA has defined a range of enforcement powers under the Consumer Protection Act 1987, Medical Device Regulations 2002, and General Product Safety Regulations 2005.

Consumer Protection Act 1987

  1. Offence against safety regulations

  2. Prohibition notices

  3. Notices to warn

  4. Suspension notices

  5. Test purchase

  6. Forfeiture orders

Medical Device Regulations 2002

  1. Compliance notice

  2. Restriction notice

Under the General Product Safety Regulations 2005 a recall notice is issued on the grounds that the device is a dangerous product and has already been supplied or made available to the consumers.

Recall

While marketing a device, the manufacturer distributor, or consumer might report complaints as some quality defects. If a complaint about a defect is not justified, then it is considered a failure of the quality system and immediate corrective action is undertaken by a product recal.

The guidelines of MHRA42 and GHTF43,44 have termed recall a FSCA to reduce the risk of harm to patients, operators, or others or to minimize the recurrence of the event. The FSCA would include the following actions:

  1. Return of a medical device to the manufacturer or its representative (which is termed recall)

  2. Device modification

  3. Device exchange

  4. Device destruction

  5. Advice given by manufacturer regarding the use of the device

The MHRA42 has recommended timescales to issue a FSCA (Table 2).

Table 2.

Timescales to issue a FSCA

Draft field safety notice Minimum of 48 h for MHRA to comment
Response to MHRA on queries concerning FSCA 21 working days or as specifically requested in writing by MHRA, eg, when a serious public threat

Abbreviation: FSCA, field safety corrective actions.

The FSCA could include actions such as return of a medical device to supplier; device modification, exchange or destruction; retrofit by the purchaser of manufacturer’s modification or design change; and any advice being given by the manufacturer on the use of the device. However, the manufacturer must distribute a field safety notice (FSN) by appropriate means such as by conformation of receipt. The FSN should be sent in the official language of the recipient, and a common layout technique should be used by the manufacturer to highlight the most important parts of the letter and to have a clearly arranged notice. The FSN itself should include the following items:

  1. A clear title like “Urgent Safety Notice” on the notice itself, on the envelope if sent by mail, and as the subject line if sent by email or fax

  2. The intended audience: clear statement about the intended recipient of the notice

  3. Concise description of subject device (model, batch, or serial number)

  4. A factual statement explaining the reasons for the FSCA

  5. A clear description of the hazards associated with the specific failure of the device and, where appropriate, the likelihood of occurrence, being mindful of the intended audience

  6. The recommended action to be taken by the recipient of the FSN

  7. Time frames by which the action should be taken by the manufacturer and user, where appropriate

  8. Designated contact point for the recipient of the FSN to use to obtain further information.

The notice must also include a request to inform customers or patients who received the product. If relevant, the notice must also include a request for the details of any affected devices that have been transferred to other organizations or been destroyed to be given to the manufacturer so that follow-up can take place and a request for a copy of the FSN to be passed on to the organization to which the device has been transferred. Manufacturers using a unique reference number to identify the FSCA should include this in the FSN, or include a date.

In the United States9 and Australia,41 medical device recalls are usually conducted voluntarily by the manufacturer, but in rare instances when the manufacturer or importer (foreign manufacturer) fails to voluntarily recall a device that is a risk to health or the device is removed from ARTG, the respective authority could issue a recall order to the manufacturer. However, the recall does not include market withdrawal or a stock recovery. Also in Australia, the manufacturers, sponsors, and distributors are required to report the necessity for a recall on a Medical Device Incident Report form notifying not only serious problems for devices but also any kind of tampering with the medical device. But if the sponsor fails to carry out the mandatory recall and fails to notify a safety-related recall, then they are subjected to substantial fines.

The recalls have been classified on the same principles (associated relative health hazard) by both FDA and TGA as follows:

  1. Class I: where severe adverse health consequences or death are likely

  2. Class II: where temporary or medically reversible health consequences are likely

  3. Class III: where use or exposure to the offending product will not likely cause adverse health consequences.

Class I or class II recalls are considered to be urgent safety-related recalls, whereas class III recalls are considered to be routine nonsafety-related recalls. In addition, the TGA has classified recalls on the basis of the following:

  1. The incidence of complaints

  2. Distribution networks

  3. Recovery procedures

  4. Resources for corrective action

  5. Availability of alternative products

The TGA and FDA have also defined the elements of recall strategy, including classification of recall; necessity of initiating a recall; level (or depth) of a recall, ie, wholesale, retail, hospital, or consumer level; significance of the hazard (if any); the channels by which the goods have been distributed; and the level to which distribution has taken place. The manufacturer must also specify the method used and the level of consignees’ communication for recall. These levels have been specified by the FDA as below:

  1. Level A: 100% of the total number of consignees to be contacted

  2. Level B: some percentage, ie, 10%–100% of the total number of consignees to be contacted on a case-by-case basis

  3. Level C: 10% of the total number of consignees to be contacted

  4. Level D: 2% of the total number of consignees to be contacted

  5. Level E: no effectiveness checks

The manufacturer marketing devices in the United States should submit regular a recall status report, whose frequency is decided by the FDA, but generally the reporting interval is between 2 and 4 weeks. According to TGA, the sponsor should provide the Australian Recalls Coordinator with an interim and a final report on the recall at 2 and 6 weeks after the implementation of the recall or at other agreed times, to establish the effectiveness of the recall. The MHRA has recommended inclusion of the following in the follow-up progress of FSCA report:

  1. Monitoring of the extent of notification and reconciliation of FSCA

  2. In-house corrective action to prevent the recurrence of problem

The final FSCA report should also incorporate validation of corrective measures.

Result

In 1992, although the GHTF was established with a view to achieve uniformity in medical device regulations globally among founding members, the regulations of the United States, United Kingdom, and Australia still differ. Also taking into account the guideline issued by CDSCO, it was observed that the regulators in India have moved towards harmonizing the vigilance system with those of regulated countries. Major differences are still apparent, however (see Table 3).

Table 3.

Differences in vigilance systems of US, UK, Australia, and India

Parameters of countries FDA TGA (Australia) MHRA CDSCO (India)
Definition of medical device Includes all instruments, appliances, materials, machines, in vitro diagnostic agents, implants, software, accessories, and disinfectants Excludes tampons and hospital, household, and commercial-grade disinfectants Excludes materials used for disinfection of medical devices 10-device category regulated as drug
Medical device classification 3 classes: class I, class II, and class III 5 classes: class I, classes II a and II b, class III, and class AIMD 4 classes: class I, class IIa, class IIb, and class III No defined classes for devices
Basis of classification Level of control
Medical specialties
Classification rules Classification rules NA
Postmarketing surveillance activities Medical device tracking
MDR
MDR event files, records, and written procedures
Complaint handling
Recall procedure and seizures
Adverse event reporting
Vigilance exchange program
Enforcement activities
Distribution records
Audits
Adverse event reporting
FSCA and field safety notices
Investigations
Enforcement
Postmarket clinical follow-up
Records
Adverse event reporting
For importers
 complaint handling
 adverse event reporting
 procedure for distribution of records
 procedure for recall
Medical device tracking Have established tracking system since 1993 IMDTS developed recently for tracking of patients with implantable medical devices AITS developed to investigate the failure modes of the device by assessment of user reports In labeling provisions, the lot number/batch number for device is mandatory for easy traceability
Who need to report AE Manufacturers, importers, user facilities, users, distributors, and health professionals Manufacturers, sponsors, users, health professionals, and TGA Manufacturers, users, health professionals, authorized representatives, and MHRA Manufacturers only
Criteria for reporting Death or serious injury
Device malfunctions
User error
Injury/illness requiring medical intervention
Event has occurred
Medical device’s association with the event
Event led/might lead to death/serious injury
Event has occurred
Medical device’s association with the event
Event led/might lead to death/serious injury
Event has occurred
Medical device’s association with the event
Event led/might lead to death/serious injury
Not-reportable incidents/events Manufacturers can apply for RAE, eg,
 Erroneous information
 When other manufacturer makes the device
User-detected deficiencies
Root cause of the adverse event is due to the patients’ preexisting condition
Exceeded service life of device
Likelihood of adverse event is acceptable after risk assessment Side effects clearly identified in the manufacturer’s labeling and documented in device master record
User-detected deficiencies
Root cause of the adverse event is due to the patients’ pre-existing condition
Exceeded service life of device
Likelihood of adverse event is acceptable after risk assessment
Side effects clearly identified in the manufacturer’s labeling and documented in device master record
User-detected deficiencies
Root cause of the adverse event is due to the patients’ preexisting condition
Exceeded service life of device
Likelihood of adverse event is acceptable after risk assessment
Side effects clearly identified in the manufacturer’s labeling and documented in device master record
Reporting time frame Manufacture: death, serious injury, and malfunctions – 30 calendar days, and events requiring immediate remedial action – 5 working days
User facility: death and serious injury – 10 working days
Distributors and importers: death, serious injury, and malfunction to manufacturer – 10 working days
Death/serious deterioration – 10 calendar days
Reportable near adverse event – 30 calendar days
Serious public health threat requiring remedial action – 48 h
Serious public threat – 2 calendar days
Death/serious deterioration – 10 elapsed calendar days
Other incidents – 30 elapsed calendar days
After receiving user reports from MHRA, reporting 3 working days
Unanticipated death or serious injury within 10 days
All other reportable events not later than 30 elapsed calendar days
Types of report 30-day reports
5-day reports
Baseline reporting
Supplemental reporting
Annual reports
Adverse event report for each incident or medical device
Annual report
Initial reporting of adverse events
Final reports
Periodic summary reporting
Trend reporting
Initial reporting
Trend reporting
Final reporting
Applicable forms Form 3500 – online
Form 3500A for manufacturers, importers, and distributors
Form 3419
Form 3417
Form 3381
Form MDIR01
Form UDIR01 – online
Manufacturer’s incident report form
Online reporting for manufacturers by MORE
Adverse event reporting form
Vigilance exchange NA With overseas regulatory agencies Exchange information for similar incidents and for FSCA within and outside Not defined
Vigilance exchange form NA No Yes NA
Records AE records
Evaluation records
Records for follow-up and inspection
Investigation protocol
Copies of test, laboratory reports, and service records
Distribution records
Records for products manufactured
Records of problem report, its evaluation, and appropriate action taken
AE records
Evaluation records
Customer/user complaint record
Records for products manufactured
Records of distributors
CAPA records
A mandatory specification for importers only
Recall/FSCA Manufacturers need to initiate recall Sponsors need to initiate recall Manufacturers need to initiate recall A mandatory specification for importers only
Recall communication Telephone calls, telegrams, and mailgrams
First class letters approved by FDA
General public warning
Public warning through specialized news media
Recall letters approved within 48 hours of recall agreement
Paid advertisements to consumer/retail level approved by TGA
FSN approved by MHRA as per specified format within 48 hours of FSCA agreement
In case of urgency, through telephone, fax, or by a visit

Abbreviations: AE, adverse event; AIMD, active implantable medical device; AITS, Adverse Incident Tracking System; CAPA, corrective and preventive actions; CDSCO, Central Drug Standard Control Organization; FDA, Food and Drug Administration; FSCA, field safety corrective actions; FSN, field safety notice; IMDTS, Implantable Medical Device Tracking Subcommittee; MDR, medical device reporting; MHRA, Medicines and Healthcare products Regulatory Agency; RAE, remedial action exemption; TGA, Therapeutic Goods and Administration.

Conclusion

Despite efforts by the founding members of GHTF to achieve uniformity for medical device regulations globally since 1992, these regulations still differ. Like ICH, these major differences need to be eliminated to facilitate the introduction by manufacturers of emerging technologies for medical devices in both developed and developing countries.

Taking into account these differences, the question remains as to what is the aim of harmonization. The consensus should be that it should aim at establishing regulations that not only alleviate and improve patients’ health but also promote international trade. To this effect, India has begun to introduce an adverse event reporting system that is uniform with that of regulated countries. Hence, in order to maintain a national system of controls for medical devices, CDSCO should promote the adoption of existing vigilance systems of regulated countries and incorporate timely amendments. For example,

  1. The regulators should define clearly the medical device and classify the devices on the basis of risk involved.

  2. Like the United States, the reporting time frame must include not only manufacturers but also the user facility and distributors.

  3. The regulations should include distinct tracking provisions for tracking of devices through to end-users, such as inclusion of General Medical Device Nomenclature codes.

  4. CDSCO should also incorporate a vigilance exchange program.

  5. Unlike the FDA, CDSCO should define distinctly the enforcement actions including a recall system in case of breach of regulations.

  6. CDSCO should also indicate a specific format for FSCA and FSN.

  7. CDSCO should also establish an online adverse event reporting system.