Abstract
Self-monitoring of blood glucose (SMBG) implicates a number of handling steps with the meter and the lancing device. Numerous user errors can occur during SMBG, and each step adds to the complexity of use. This report compares the required steps to perform SMBG of one fully integrated (the second generation of the Accu-Chek® Mobile), three partly integrated (Accu-Chek Compact Plus, Ascensia® Breeze®2, and Accu-Chek Aviva), and six conventional (Bayer Contour®, Bayer Contour USB, BGStar™, FreeStyle Lite®, OneTouch® Ultra® 2, and OneTouch Verio™Pro) systems. The results show that the fully integrated system reduces the number of steps to perform SMBG. The mean decrease is approximately 70% compared with the other systems. We assume that a reduction of handling steps also reduces the risk of potential user errors and improves the user-friendliness of the system.
Keywords: handling steps, integrated blood glucose measurement system, potential user errors, self-monitoring of blood glucose, user-friendliness
Introduction
Self-monitoring of blood glucose (SMBG) is an integral part of diabetes self-management. The reliability of the SMBG value depends heavily on if and how well the user prepares and performs the measurement.1,2
Innovative technologies have led to high-quality SMBG systems that provide a wide range of user-friendly features such as short measurement times, small blood samples, or less painful lancing devices.3–6
However, nonadherence to recommended test frequencies is still a common issue.7 User-friendliness of a system can help overcome barriers to testing that are closely linked to the SMBG adherence of patients.8
A number of user-errors can occur during SMBG, including miscoding of blood glucose (BG) meters, use of expired test strips, or inadequate storage conditions of test strips.2,9,10 Current systems are often characterized by more integrative features. Some of the more integrated systems incorporate the tests (test fields or test strips) in cassettes, drums, or discs instead of using single test strips. Other integrative features concern the lancing technology, such as lancing devices that contain a drum with six sterile lancets instead of inserting single lancets for every measurement.
The aim of this study was to evaluate the required handling steps for performing a BG measurement with fully integrated, partly integrated, and conventional SMBG systems.
Methods
In this study, 10 SMBG systems (fully integrated, partly integrated, and conventional) were used (Table 1 and Figure 1). Accu-Chek® Mobile, Accu-Chek Compact Plus, and Ascensia® Breeze®2 incorporate the tests in cassettes, drums, or discs (Figure 1, Table 1). The other systems are single-strip systems. Regarding the functional characteristics of the corresponding lancing devices, the Accu-Chek FastClix (Accu-Chek Mobile) and the Accu-Chek Multiclix (Accu-Chek Aviva) use a drum containing six sterile lancets. The lancing devices of the other systems require the insertion of a new lancet for each measurement (Table 1).
Table 1.
Features and Characteristics of the Investigated Self-Monitoring of Blood Glucose Measurement Systems
| BG meter | Lancing device | Test strip system | Codinga | Weight (g)b | Required blood volume (µl) | Manufacturer | ||
|---|---|---|---|---|---|---|---|---|
| Fully Integrated system | Accu-Chek Mobile | Accu-Chek FastClix | Drum (6 lancets) | Cassette (50 test elements) | Aut | 149 | 0.3 | Roche Diagnostics GmbH, Mannheim, Germany |
| Accu-Chek Compact Plus | Accu-Chek Softclix Plus | Single lancets | Drum (17 test elements) | Aut | 267.5 | 1.5 | Roche Diagnostics GmbH, Mannheim, Germany | |
| Partly integrated systems | Ascensia Breeze2 | Ascensia Microlet® | Single lancets | Disc (10 test elements) | Aut | 170 | 1 | Bayer Consumer Care AG, Basel, Switzerland |
| Accu-Chek Aviva | Accu-Chek Multiclix | Drum (6 lancets) | Single test strips | Man | 160 | 0.6 | Roche Diagnostics GmbH, Mannheim, Germany | |
| Conventional systems | Bayer Contour® | Bayer Microlet 2 | Single lancets | Single test strips | Aut | 119 | 0.6 | Bayer Consumer Care AG, Basel, Switzerland |
| Bayer Contour USB | Bayer Microlet 2 | Single lancets | Single test strips | Aut | 104.7 | 0.6 | Bayer Consumer Care AG, Basel, Switzerland | |
| BGStar™ | BGStar | Single lancets | Single test strips | No | 102.4 | 0.5 | AgaMatrix Inc. Salem, NH | |
| FreeStyle Lite® | FreeStyle™ | Single lancets | Single test strips | No | 100.5 | 0.3 | Abbott Diabetes Care Inc., Alameda, CA | |
| OneTouch Ultra2 | OneTouch Ultra Soft® | Single lancets | Single test strips | Man | 145.5 | 1 | LifeScan Inc., Milpitas, CA | |
| OneTouch Verio™Pro | OneTouch Comfort™ | Single lancets | Single test strips | No | 185.3 | 0.4 | LifeScan Europe, Zug, Switzerland | |
Coding technologies of the SMBG systems used are differentiated into manual coding, automatic coding, and no coding according to the manual’s description. Man, manual coding; Aut, automatic coding; No, no coding.
The calculated weight (g) includes meter + test strips/test strip drum(s)/test strip disc(s)/tape cassette(s) for 50 (51) measurements; lancing device (including lancet) + 5 lancets or lancing device, including drum with 6 lancets; and carrying case (if provided by manufacturer).
Figure 1.
Self-monitoring of blood glucose measurement systems. Fully integrated system: the BG meter incorporates the tests in cassettes and the corresponding lancing device uses a drum containing six sterile lancets. Partly integrated systems: the BG meter incorporates the tests in discs or drums or the corresponding lancing device uses a drum containing six sterile lancets. Conventional systems: the systems are single-strip systems and the corresponding lancing devices require the insertion of a new lancet for each measurement.
An important criterion is the coding technology of the systems (Table 1). Coding is a process for addressing batch-specific variations in the production of tests to ensure reliable SMBG values. Some systems require manual input of a batch-specific code number into the BG meter for every test strip batch. Other systems have an automatic coding feature by putting the test fields or test strips into the BG meter or do not require coding (Table 1).
Handling steps for each system were analyzed and counted for two different categories: routine measurement and routine measurement with additional tasks (Tables 2 and 3). Additional tasks are not required for each measurement but repeat regularly in a series of 100 routine measurements.
Table 2.
Analyzed Handling Step Categories and Corresponding Handling Steps within a Routine Blood Glucose Measurement
| Routine BG measurement | Handling steps |
|---|---|
| Unpacking the carrying case |
|
| Preparation of the lancing device |
|
| Preparation of the BG meter |
|
| Measurement |
|
| Cleaning up BG meter |
|
| Cleaning up lancing device |
|
| Packing in |
|
Table 3.
Analyzed Additional Tasks and the Corresponding Handling Steps
| Additional tasks | Handling steps |
|---|---|
| Change code chip/code number |
|
| Change tape cassette/drum/strip disc |
|
| Change test strip vial |
|
| Perform control measurement |
|
| Change lancet drum |
|
Table 2 gives an overview on handling step categories to perform a routine BG measurement. Table 3 shows possible additional tasks with the used systems. System-independent steps that are required to perform SMBG correctly (e.g., washing hands before the measurement) were not counted.
Each handling step with each system was performed and documented several times by two investigators. Only steps mentioned in the manual were counted, including associated steps, e.g., if the manual mentioned “put the strip into BG meter,” we also counted the steps “open the test strip vial” and “take the test strip out of the vial.” Handling steps required for 100 measurements were calculated by combining the number of steps performed for 100 measurements as well as the number of steps of additional tasks.
Additionally, the relative decreases of handling steps required for the fully integrated system (Accu-Chek Mobile) compared with each partly integrated and conventional system and the respective mean values were calculated.
Results
The smallest number of handling steps was found for the Accu-Chek Mobile system (9 steps) to perform a single routine BG measurement. The other systems needed between 25 and 38 steps (Table 4).
Table 4.
Number of Required Handling Steps to Perform a Single Routine Measurement and Number of Required Handling Steps to Perform Additional Tasks for 100 Routine Blood Glucose Measurements
| BG meter | Accu-Chek Mobile | Accu-Chek Compact Plus | Ascensia Breeze2 | Accu-Chek Aviva | Bayer Contour | Bayer Contour USB | BGStar | FreeStyle Lite | OneTouch Ultra2 | OneTouch VerioPro |
|---|---|---|---|---|---|---|---|---|---|---|
| Lancing device | Accu-Chek FastClix | Accu-Chek Softclix Plus | Ascensia Microlet | Accu-Chek Multiclix | Bayer Microlet 2 | Bayer Microlet 2 | BGStar | FreeStyle | OneTouch Ultra Soft | OneTouch Comfort |
| Routine Measurement | ||||||||||
| Unpacking | 0 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| Preparation of lancing device | 3 | 11 | 13 | 5 | 12 | 12 | 12 | 12 | 12 | 12 |
| Preparation of BG meter | 1 | 1 | 3 | 9 | 8 | 8 | 8 | 8 | 9 | 8 |
| Measurement | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 |
| Cleaning up BG meter | 1 | 2 | 3 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| Cleaning up lancing device | 0 | 6 | 7 | 1 | 7 | 7 | 7 | 7 | 7 | 7 |
| Packing in | 0 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| Steps for a single routine measurement | 9 | 28 | 34 | 25 | 37 | 37 | 37 | 37 | 38 | 37 |
| Additional tasks | ||||||||||
| Change code chip/code number | 0 | 0 | 0 | 12a | 0 | 0 | 0 | 0 | 12b | 0 |
| Change tape cassette/drum/strip disc | 20a | 48c | 80d | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Change test strip vial | 0 | 0 | 0 | 8a | 8a | 8a | 10a | 8a | 20b | 20b |
| Perform control measurement | 18a | 22e | 0 | 34a | 0 | 0 | 0 | 0 | 72b | 68b |
| Change lancet drum | 102f | 0 | 0 | 153f | 0 | 0 | 0 | 0 | 0 | 0 |
| Additional steps/100 routine measurements | 140 | 70 | 80 | 207 | 8 | 8 | 10 | 8 | 104 | 88 |
The required steps for the additional tasks are counted for 100 routine measurements, and the additional tasks are repeated every
50 measurements,
25 measurements,
17 measurements,
10 measurements,
51 measurements,
6 measurements.
The calculated numbers of handling steps to perform 100 routine BG measurements are shown in Figure 2. Figure 3 shows the calculated numbers of lancing-device-specific and BG-meter/measurement-specific steps to perform 100 measurements (routine measurements, including additional tasks). Considering the required steps for 100 routine measurements and taking into account the additional tasks, the minimum number of required steps is found in the Accu-Chek Mobile system (1040; 402 lancing device specific, 638 BG meter/measurement-specific steps; Figures 2 and 3). The other systems require between 2707 (Accu-Chek Aviva meter with 753 lancing-device-specific and 1954 BG-meter/measurement-specific steps) and 3904 (OneTouch® Ultra®2 with 1900 lancing-device-specific and 2004 BG-meter/measurement-specific steps) steps (Figures 2 and 3).
Figure 2.
The calculated numbers of handling steps to perform 100 routine BG measurements and additionally the numbers of handling steps to perform 100 routine BG measurements and the required additional tasks within 100 measurements are shown. The minimum number of required handling steps is found for the fully integrated Accu-Chek Mobile system (100 routine measurements, 900; 100 routine measurements, including additional tasks, 1040) The partly integrated and conventional systems require between 2500 (Accu-Chek Aviva) and 3800 (OneTouch Ultra2) handling steps for 100 routine BG measurements and between 2707 (Accu-Chek Aviva) and 3904 (OneTouch Ultra2) when additional tasks are taking into account.
Figure 3.
Calculated numbers of lancing-device-specific and BG-meter/measurement-specific handling steps to perform 100 measurements (routine measurements, including the additional tasks). Lancing device specific: preparation/cleaning up lancing device (routine measurement); change lancet drum (additional tasks). BG-meter/measurement specific: unpacking/packing in the carrying case, preparation/cleaning up BG meter, measurement (routine measurement); change code chip/code number, change tape/cassette/drum/strip disc, change test strip vial, perform control measurement (additional tasks).
The reduction of the required handling steps with the Accu-Chek Mobile system compared with partly integrated and conventional systems leads to a decrease in mean by 73% (ranges between 64% and 76%) for one single routine measurement and to a decrease in mean by 70% (ranges between 62% and 73%) for 100 measurements (routine measurements, including additional tasks; Table 5).
Table 5.
Relative Difference in the Number of Required Handling Stepsa
| Steps decrease with the fully integrated Accu-Chek Mobile compared with partly integrated and conventional systems (%) | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| BG meter Lancing device | Accu-Chek Compact Plus Accu-Chek Softclix Plus |
Ascensia Breeze2 Ascensia Microlet |
Accu-Chek Aviva Accu-Chek Multiclix |
Bayer Contour Bayer Microlet 2 |
Bayer Contour USB Bayer Microlet 2 |
BGStar BGStar |
FreeStyle Lite FreeStyle |
OneTouch Ultra 2 OneTouch Comfort |
OneTouch VerioPro OneTouch Verio |
Mean | |
| Accu-Chek Mobile versus partly integrated and conventional systems | Steps decrease/ 1 testb | 68 | 74 | 64 | 76 | 76 | 76 | 76 | 76 | 76 | 73 |
| Accu-Chek Mobile versus partly integrated and conventional systems | Steps decrease/ 100 testsc | 64 | 70 | 62 | 72 | 72 | 72 | 72 | 73 | 73 | 70 |
Minimum and maximum values are given in bold font.
1 test = 1 routine BG measurement.
100 tests = 100 routine BG measurements + additional tasks necessary within this series.
Discussion
Today, SMBG systems have to meet the challenge of providing improved features to maximize user-friendliness on one hand and to minimize errors on the other hand. User-friendliness may be improved by a reduction of handling steps.8
The results of this study show that the fully integrated system (Accu-Chek Mobile) requires the least number of handling steps to perform SMBG compared with partly integrated and conventional systems.
One main reason for the reduction of handling steps with fully integrated systems is the elimination of single-test-strip insertion, which was achieved by integrating a test cassette with 50 tests. Handling with single test strips and test strip vials (opening, closing) and the expiration date of test strips is a commonly underestimated but frequent source of error that potentially impacts the reliability of SMBG values.2,10 Integration of a test strip drum (Accu-Chek Compact Plus) or disc (Ascensia Breeze2) into the BG meter also eliminates the single insertion of test strips for every measurement.
Another important reason for the reduction of steps with fully integrated systems is the insertion of a lancet drum containing six sterile lancets instead of handling with single lancets. The insertion of lancets represents a frequent user-error source.2 Insertion of single lancets for every measurement requires good handling skills, as the sharp lancets represent a potential risk of accidental sticks for the user. The development of lancing drums (Accu-Chek FastClix, Accu-Chek Multiclix) improves the user-friendliness because the insertion of the lancet drum is much easier and safer than the insertion of single lancets, as the user does not come into contact with the sharp lancets when inserting the drum into the BG meter. The reuse of lancets, which is done by many patients, has not been taken into consideration, as steps were counted strictly according to the manual. Another limitation of this study is the counting of handling steps under laboratory conditions by well-trained personnel, which does not represent the patient’s measurement procedure. Additionally, for some patients, the complexity of steps to perform a measurement might be more important than the number.
Compared with partly integrated and fully integrated systems, most of the conventional systems are smaller and lighter. On the other hand, the conventional systems require separate carrying of all measurement components (lancing device, lancets, test strip vial), not to forget the disposal of waste, which can be reduced with more integrated systems.
Most of the systems used in this study provide a clear simplification for users—they do not require manual coding. Previous studies demonstrated that some miscoded meters resulted in readings of up to 30% median difference when compared with the results obtained with a laboratory glucose analyzer.11–13 Miscoded meters can result in significant insulin dose errors.11
In summary, we demonstrated that fully integrated systems require the least number of handling steps to perform SMBG. It can be assumed that the reduction of handling steps is associated with an improved user-friendliness and a reduction of potential errors. It seems likely that, in the future, the trend will keep moving toward more integrated SMBG systems, as they have the potential to improve user-friendliness and SMBG adherence and concurrently avoid user errors.
Glossary
- (BG)
blood glucose
- (SMBG)
self-monitoring of blood glucose
Funding
This work was funded by Roche Diagnostics GmbH, Mannheim, Germany.
Disclosure
Guido Freckmann is general manager of the Institut fuer Diabetes-Technologie GmbH, Ulm, Germany. This institute performs clinical trials in cooperation with different medical device companies. Guido Freckmann has received speaker’s honoraria and travel support from such companies.
References
- 1.Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician. Clin Chem. 2002;48(7):994–1003. [PubMed] [Google Scholar]
- 2.Müller U, Hämmerlein A, Casper A, Schulz M. Evaluation der Durchführung von Glukoseselbstkontrollen in Apotheken (EDGAr) Diabetes Stoffwechsel Herz. 2006;15(4):9–17. [Google Scholar]
- 3.Fruhstorfer H, Schmelzeisen-Redeker G, Weiss T. Capillary blood sampling: relation between lancet diameter, lancing pain and blood volume. Eur J Pain. 1999;3(3):283–286. doi: 10.1053/eujp.1999.0132. [DOI] [PubMed] [Google Scholar]
- 4.Fruhstorfer H, Schmelzeisen-Redeker G, Weiss T. Capillary blood volume and pain intensity depend on lancet penetration. Diabetes Care. 2000;23(4):562–563. doi: 10.2337/diacare.23.4.562. [DOI] [PubMed] [Google Scholar]
- 5.Hönes J, Müller P, Surridge N. The technology behind glucose meters: test strips. Diabetes Technol Ther. 2008;10(Suppl 1):S10–S26. [Google Scholar]
- 6.Westhoff A, Jendricke N, Haug C, Freckmann G. Schmerzempfinden an Finger und Handballen bei Verwendung verschiedener Blutzuckermesssysteme. Diabetologie Stoffwechsel. 2010;5:41. [Google Scholar]
- 7.Hansen MV, Pedersen-Bjergaard U, Heller SR, Wallace TM, Rasmussen AK, Jorgensen HV, Pramming S, Thorsteinsson B. Frequency and motives of blood glucose self-monitoring in type 1 diabetes. Diabetes Res Clin Pract. 2009;85(2):183–188. doi: 10.1016/j.diabres.2009.04.022. [DOI] [PubMed] [Google Scholar]
- 8.Bergenstal R, Pearson J, Cembrowski GS, Bina D, Davidson J, List S. Identifying variables associated with inaccurate self-monitoring of blood glucose: proposed guidelines to improve accuracy. Diabetes Educ. 2000;26(6):981–989. doi: 10.1177/014572170002600610. [DOI] [PubMed] [Google Scholar]
- 9.Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol. 2009;3(4):903–913. doi: 10.1177/193229680900300438. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Bamberg R, Schulman K, MacKenzie M, Moore J, Olchesky S. Effect of adverse storage conditions on performance of glucometer test strips. Clin Lab Sci. 2005;18(4):203–219. [PubMed] [Google Scholar]
- 11.Raine CH, 3rd, Schrock LE, Edelman SV, Mudaliar SR, Zhong W, Proud LJ, Parkes JL. Significant insulin dose errors may occur if blood glucose results are obtained from miscoded meters. J Diabetes Sci Technol. 2007;1(2):205–210. doi: 10.1177/193229680700100211. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Raine CH, 3rd, Pardo S, Parkes JL. Predicted blood glucose from insulin administration based on values from miscoded glucose meters. J Diabetes Sci Technol. 2008;2(4):557–562. doi: 10.1177/193229680800200404. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Baum JM, Monhaut NM, Parker DR, Price CP. Improving the quality of self-monitoring blood glucose measurement: a study in reducing calibration errors. Diabetes Technol Ther. 2006;8(3):347–357. doi: 10.1089/dia.2006.8.347. [DOI] [PubMed] [Google Scholar]