Skip to main content
Journal of Digital Imaging logoLink to Journal of Digital Imaging
. 1997 Aug;10(Suppl 1):128–132. doi: 10.1007/BF03168675

The laser film digitizer: Density, contrast, and resolution

Terrance N Teslow 1,2,
PMCID: PMC3452839  PMID: 9268857

Abstract

This work evaluates differences in density linearity, contrast, and resolution of digital images acquired from a laser film scanner. A calibrated photographic step tablet tested linearity of diffuse optical density (OD). It was digitized within center openings of a dark and light film. The 14- × 17-inch films were scanned into 1024 × 1280 picture elements of 8 bits. A photographic tapered resolution guide evaluated the scanner’s detail resolution. A single characteristic curve adequately represents the rate of change of OD per pixel value, but two piecewise linear functions are better. One ranges in light OD and a second in dark OD. Contrast shows a steeper slope in a light film rather than a dark film, yet a greater difference exists between two identical digitizers. The resolution test diminishes at 1.3 line pairs per mm (LP/mm), which compares to a Nyquist limit of 1.44 LP/mm. Such density, contrast, and resolution tests yield a benchmark to assure quality operations.

Key Words: quality assurance, image quality, laser film digitizer, film densitometry, PACS, laser scanner, digital radiography

Full Text

The Full Text of this article is available as a PDF (1.1 MB).

References

  • 1.Fang-Fang Y, Giger ML, Doi K, et al. Evaluation of imaging properties of a laser film digitizer. Phys Med Biol. 1992;37:273–280. doi: 10.1088/0031-9155/37/1/020. [DOI] [PubMed] [Google Scholar]
  • 2.Halpern EJ. A test pattern for quality of laser scanner and charge-coupled device film digitizers. J Digit Imaging. 1995;8:3–9. doi: 10.1007/BF03168050. [DOI] [PubMed] [Google Scholar]
  • 3.Halpern EJ, Esser PD. An improved phantom for quality control of laser scanner digitizers in picture archiving and communication systems. J Digit Imaging. 1991;4:241–247. doi: 10.1007/BF03173906. [DOI] [PubMed] [Google Scholar]
  • 4.Meeder RJJ, Jaffray DA, Munro P. Tests for evaluating laser film digitizers. Med Phys. 1995;22:635–642. doi: 10.1118/1.597550. [DOI] [PubMed] [Google Scholar]
  • 5.Siegel EL, Templeton AW, Cook LT, et al. Image calibration of laser digitizers, printers, and gray-scale displays. Radiographics. 1992;12:329–335. doi: 10.1148/radiographics.12.2.1561421. [DOI] [PubMed] [Google Scholar]
  • 6.Gray JE, Lisk KG, Haddick DH, et al. Test pattern for video displays and hard-copy cameras. Radiology. 1985;154:519–527. doi: 10.1148/radiology.154.2.3966138. [DOI] [PubMed] [Google Scholar]
  • 7.Dwyer SJ, Stewart BK, Sayre JW, et al. Performance characteristics and image fidelity of gray-scale monitors. Radiographics. 1992;12:765–772. doi: 10.1148/radiographics.12.4.1636038. [DOI] [PubMed] [Google Scholar]
  • 8.Lim AJ. Image quality in film digitizers: testing and quality assurance. In: Gould RG, Boone JM, editors. Categorical Course in Physics: Technology Update and Quality Improvement of Diagnostic X-ray Imaging Equipment-Laser Cameras. Film Digitizers, and Soft-copy Displays. Oak Brook, IL: Radiological Society of North America; 1996. pp. 183–193. [Google Scholar]
  • 9.Teslow TN, Gilbert RA, Grant WH, et al. A teleradiology case conference system. Journal of Telemedicine and Telecare. 1995;1:95–99. doi: 10.1177/1357633X9500100205. [DOI] [PubMed] [Google Scholar]