Tissue composition measurement may provide a quantitatively and physically meaningful method to objectively determine the “mammographic density” linked to breast cancer risk. A single energy hybrid (SEH) technique is described for measuring the tissue composition on a pixel-by-pixel basis with a single digital mammogram. Theoretical models were derived and used to compute signal-to-noise ratios (SNRs) in tissue composition measurement using the SEH method. The results were compared with those computed for measurements using the dual kVp and dual screen methods. SNRs were theoretically related to the pixel area, total unattenuated detector exposure and fluence spectra of the incident X-rays. SNRs were computed for measurement of the adipose tissue thickness for a 6 cm thick breast, consisting of 50% of adipose tissue and 50% of glandular tissue. Effects of kVp and prepatient filtration were studied by computing the SNRs for various kVps and filters and optimal kVps and filters are determined. The results showed that the SNRs obtained with the SEH method is an order of magnitude better than the dual kVp method, which, in turn, is an order of magnitude better than the dual screen method. When using the optimal kVp’s and no prepatient filtration, the SNRs were computed to be 84.2, 13.2, and 2.0 for the SEH, dual kVp, and dual screen methods, respectively. Prepatient filtration can improve the SNR by as much as 35% for the dual kVp and dual screen techniques with reasonable tube loading factors (8–10).
Key Words: computed radiography (CR), digital mammography, numerical simulation, optimization, tissue composition measurement, signal-to-noise ratio
Full Text
The Full Text of this article is available as a PDF (2.1 MB).
Footnotes
This work was supported in part by a grant (CA 51248) from the National Cancer Institute, a grant (DAMD 17-93-J-3009) from the U.S. Army Medical Research and Material Command, a grant (DAAH 04-93-G-0299) from the Department of Defense, and a grant, “Breast Fat Measurement,” from the Pittsburgh Cancer Institute.
References
- 1.Shaw de Paredes E. Radiographic breast anatomy: Radiologic signs of breast cancer. In: Haus AG, Yaffe MJ, editors. Syllabus: A Categorical Course in Physics-Technical Aspects of Breast Imaging. ed 3. Oak Brook, IL: RSNA; 1994. pp. 35–46. [Google Scholar]
- 2.Yaffe MJ, Jennings R, Fahrig R, et al. X-ray spectral considerations for mammography. In: Haus AG, Yaffe MJ, et al., editors. Syllabus: A Categorical Course in Physics-Technical Aspects of Breast Imaging. ed 3. Oak Brook, IL: RSNA; 1994. pp. 63–74. [Google Scholar]
- 3.Byng JW, Boyd NF, Fishell E, et al. The quantitative analysis of mammographic densities. Phys Med Biol. 1994;39:1629–1638. doi: 10.1088/0031-9155/39/10/008. [DOI] [PubMed] [Google Scholar]
- 4.Leinster SJ, Walsh PV, Whitehouse GH, et al. Factors associated with mammographic parenchymal patterns. Clin Radiol. 1988;39:252–256. doi: 10.1016/s0009-9260(88)80519-1. [DOI] [PubMed] [Google Scholar]
- 5.Saftlas AF, Hoover RN, Brinton LA, et al. Mammographic parenchymal patterns as indicators of breast cancer risk. Cancer. 1991;67:2833–2838. doi: 10.1002/1097-0142(19910601)67:11<2833::AID-CNCR2820671121>3.0.CO;2-U. [DOI] [PubMed] [Google Scholar]
- 6.Warner E, Lockwood G, Tritchler D. The risk of breast cancer associated with mammographic parenchymal patterns: A meta-analysis of the published literature to examine the effect of method of classification. Cancer Detect Prev. 1992;16:67–72. [PubMed] [Google Scholar]
- 7.Byrne C, Schairer C, Wolfe J, et al. Mammographic features and breast cancer risk: effects with time, age, and menopause status. J Natl Cancer Inst. 1995;87:1622–1629. doi: 10.1093/jnci/87.21.1622. [DOI] [PubMed] [Google Scholar]
- 8.Boyd NF, Byng JW, Jong RA, et al. Quantitative classification of mammographic densities and breast cancer risk: Results from the Canadian National Breast Screening Study. J Natl Cancer Inst. 1995;87:670–675. doi: 10.1093/jnci/87.9.670. [DOI] [PubMed] [Google Scholar]
- 9.Rowlands JA, Hunter DM, Araj N. X-ray imaging using amorphous selenium: A photoinduced discharge readout method for digital mammography. Med Phys. 1991;18:421–431. doi: 10.1118/1.596689. [DOI] [PubMed] [Google Scholar]
- 10.Yaffe MJ. Direct digital mammography using a scanned-slot CCD imaging system. Med Prog Technol. 1993;19:13–21. [PubMed] [Google Scholar]
- 11.Karellas A, Harris LJ, D’Orsi CJ. Small field digital mammography with a 2048×2048 pixel charge coupled device. Radiology. 1990;177:288–288. [Google Scholar]
- 12.Williams MB, Liu H, Fajardo LL, et al. Imaging properties of a prototype digital detector for full-breast mammography Radiology 1995197291–291.7568840 [Google Scholar]
- 13.Shaw CC, Wang TP, Breitenstein DS, et al. A high resolution storage phosphor imaging system for mammographic imaging. Med Phys. 1995;22:997–997. [Google Scholar]
- 14.Workman A, Cowen AR, Brettle DS. Physical evaluation of computed radiography as a mammographic x-ray imaging system. Brit J Radiol. 1994;67:988–996. doi: 10.1259/0007-1285-67-802-988. [DOI] [PubMed] [Google Scholar]
- 15.Lehmann LA, Alvarez RE, Macovski A, et al. Generalized image combinations in dual kVp digital radiography. Med Phys. 1981;8:659–667. doi: 10.1118/1.595025. [DOI] [PubMed] [Google Scholar]
- 16.Gauntt DM, Barnes GT. X-ray tube potential, filtration, and detector considerations in dual-energy chest radiography. Med Phys. 1994;21:203–218. doi: 10.1118/1.597373. [DOI] [PubMed] [Google Scholar]
- 17.Cardinal HN, Fenster A. An accurate method for direct dual-energy calibration and decomposition. Med Phys. 1990;17:327–341. doi: 10.1118/1.596512. [DOI] [PubMed] [Google Scholar]
- 18.Johns PC, Drost DJ, Yaffe MJ, et al. Dual-energy mammography: Initial experimental results. Med Phys. 1985;12:297–304. doi: 10.1118/1.595767. [DOI] [PubMed] [Google Scholar]
- 19.Johns PC, Yaffe MJ. Theoretical optimization of dual-energy x-ray imaging with applications to mammography. Med Phys. 1985;12:289–296. doi: 10.1118/1.595766. [DOI] [PubMed] [Google Scholar]
- 20.Boone JM, Shaber GS, Tecotzky M. Dual-energy mammography: A detector analysis. Med Phys. 1990;17:665–675. doi: 10.1118/1.596548. [DOI] [PubMed] [Google Scholar]
- 21.Brettle DS, Cowen AR. Dual-energy digital mammography utilizing stimulated phosphor computed radiography. Phys Med Biol. 1994;39:1989–2004. doi: 10.1088/0031-9155/39/11/012. [DOI] [PubMed] [Google Scholar]
- 22.Johns PC, Beauregard RM. Incorporation of scattered radiation into dual-energy radiologic theory and application to mammography. Med Phys. 1994;21:1455–1462. doi: 10.1118/1.597404. [DOI] [PubMed] [Google Scholar]
- 23.Gingold EL, Hasegawa BH. Systemic bias in basismaterial decomposition applied to quantitative dual-energy x-ray imaging. Med Phys. 1992;19:25–33. doi: 10.1118/1.596889. [DOI] [PubMed] [Google Scholar]
- 24.Moreau M, Holdsworth DW, Fenster A. Dual-energy x-ray imaging technique for in vitro tissue composition measurement. Med Phys. 1994;21:1807–1815. doi: 10.1118/1.597283. [DOI] [PubMed] [Google Scholar]
- 25.Nishitani H, Umezu Y, Kazuhisa O, et al. Dual-energy projection radiography using condeser x-ray generator and digital radiography aparatus. Radiology. 1986;161:533–535. doi: 10.1148/radiology.161.2.3763927. [DOI] [PubMed] [Google Scholar]
- 26.Ishigaki T, Sadayuki S, Yoshimi H, et al. One-shot dual-energy subtraction imaging. Radiology. 1986;161:271–273. doi: 10.1148/radiology.161.1.3532182. [DOI] [PubMed] [Google Scholar]
- 27.Stewart BK, Huang HK. Single-exposure dual-energy computed radiography. Med Phys. 1990;17:866–875. doi: 10.1118/1.596479. [DOI] [PubMed] [Google Scholar]
- 28.Shaw CC, Gur D. Comparison of three different techniques for dual-energy subtraction imaging in digital radiography: A signal-to-noise analysis. J Digit Imaging. 1992;5:262–270. doi: 10.1007/BF03167808. [DOI] [PubMed] [Google Scholar]
- 29.White DR, Martin RJ, Darlison R. Epoxy resin based tissue substitutes. Brit J Radiol. 1977;50:814–821. doi: 10.1259/0007-1285-50-599-814. [DOI] [PubMed] [Google Scholar]
- 30.Nishikawa RM, Yaffe MJ. Model of the spatial-frequency-dependent DQE of phosphor screens. Med Phys. 1990;17:894–904. doi: 10.1118/1.596583. [DOI] [PubMed] [Google Scholar]
- 31.Fewell TR, Shuping RE. Handbook of Mammographic X-ray Spectra. Rockville, MD: U.S. Department of Health, Education, and Welfare; 1978. [Google Scholar]
- 32.Storm E, Israel HI. Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100,* in Nuclear Data Tables. New York, NY: Academic; 1970. pp. 565–681. [Google Scholar]
- 33.Johns HE, Cunningham JR. The Physics of Radiology. Springfield, IL: Thomas; 1983. [Google Scholar]
- 34.Fuji Imaging Plates ST-III and HR-III and Fuji CR Film Types 633 and 780 for Computed Radiography. Tokyo, Japan, Fuji Photo Film Co, Ltd.