Abstract
The aim of this work was to investigate the effect of mammographic acquisition parameter variations on the estimation of percent density (PD) produced by a particular semiautomated algorithm. The PD algorithm requires the user to specify a threshold pixel value segmenting breast tissue of greater and lesser density. A whole breast specimen was imaged using a variety of acquisition techniques, and the image data were processed as prescribed by the PD algorithm. PD estimates for all possible values of the user threshold were calculated for all the images. The image data were normalized so that PD varied between 30% and 80% over a fixed threshold range of 23, and a PD value of 50% was obtained for a threshold value of 195. PD differences between all the images and a baseline standard mammographic acquisition technique were calculated. We also estimated PD differences caused by small (3%) variations in operator selection of the threshold value. We found that the largest differences in PD involved changes in the density control of the mammography unit, and changes in the detector (either film type or computed radiography). The maximum PD differences due to technique were all less than 10%, with root-mean-square (RMS) variations less than 4%. PD differences due to operator variation were 24% (maximum) and 6.1% (RMS). These findings suggest that PD differences due to mammographic technique will be considerably less than those inherent to the technique, due to operator variation. All of these estimates are likely larger than differences seen in practice since optimization of the threshold by the operator was not considered in this analysis.
Full Text
The Full Text of this article is available as a PDF (801.0 KB).
References
- 1.Wolf JN, Saftlas AF, Salane M. Mammographic parenchymal patterns and quantitative evaluation of mammographic densities: A case-controlled study. Am J Roentgenol. 1987;148:1087–1092. doi: 10.2214/ajr.148.6.1087. [DOI] [PubMed] [Google Scholar]
- 2.Brisson J, Verrault R, Morrison A, et al. Diet, mammographic features of breast tissue and breast cancer risk. Am J Epidemiol. 1989;130:14–24. doi: 10.1093/oxfordjournals.aje.a115305. [DOI] [PubMed] [Google Scholar]
- 3.Saftlas AF, Hoover RN, Brinton LA, et al. Mammographic densities and risk of breast cancer. Cancer. 1991;67:2833–2838. doi: 10.1002/1097-0142(19910601)67:11<2833::AID-CNCR2820671121>3.0.CO;2-U. [DOI] [PubMed] [Google Scholar]
- 4.Byng JW, Boyd NF, Fishell E, et al. The quantitative analysis of mammographic densities. Phys Med Biol. 1994;39:1629–1638. doi: 10.1088/0031-9155/39/10/008. [DOI] [PubMed] [Google Scholar]
- 5.Boyd NF, Byng JW, Jong RA, et al. Quantitative classification of mammographic densities and breast cancer risk: Results from the Canadian National Breast Screening Study. J Natl Cancer Inst. 1995;87:670–675. doi: 10.1093/jnci/87.9.670. [DOI] [PubMed] [Google Scholar]