Skip to main content
Journal of Digital Imaging logoLink to Journal of Digital Imaging
. 2000 May;13(Suppl 1):117–120. doi: 10.1007/BF03167640

A new needle-crystalline computed radiography detector

Paul Leblans 1,, Luc Struye 1, Peter Willems 1
PMCID: PMC3453297  PMID: 10847378

Abstract

The most successful digital radiography detectors to date have been storage phosphor plates used in computed radiography (CR). The detector is cheap, has good producibility, and is robust. Direct radiography (DR) systems are being developed based on flat-panel technology. Better image quality is claimed for some DR systems. On the other hand, DR detectors have low producibility and robustness, and a high price. A new CR detector is being developed at Agfa that combines the advantages of CR and DR. It is a storage phosphor plate made up of needle-shaped crystals. The phosphor efficiently converts absorbed x-ray quanta into photostimulable centers for efficient read out. It has a large dynamic range and its emision is efficiently detected with both photomultiplier tube (PMT) and charge coupled device (CCD). It is shown that CR systems based on the new detector offer image quality that matches that of the best DR systems.

Full Text

The Full Text of this article is available as a PDF (880.4 KB).

References

  • 1.Hillen W, Eckenbach W, Quadflieg P, et al. Signal-to-noise performance in cesium iodide x-ray fluorescent screens. SPIE Med Imaging. 1991;1443:120–131. [Google Scholar]
  • 2.Thoms M. Photostimulated luminescence: A tool for the determination of optical properties of defect centers. J Luminescence. 1994;60/61:585–587. doi: 10.1016/0022-2313(94)90223-2. [DOI] [Google Scholar]
  • 3.Nakazawa M, Morikawa O, Nitta M, et al. Effect of protective layer on resolution properties of photostimulable phosphor detector for digital radiographic system. SPIE Med Imaging. 1990;1231:350–350. [Google Scholar]
  • 4.Lubinsky AR, Whiting BR, Owen JF. Storage phosphor system for computed radiography: Optical effects and detective quantum efficiency (DQE) SPIE Med Imaging. 1987;767:167–167. [Google Scholar]
  • 5.Shaw R, Metter RL. An analysis of the fundamental limitations of screen-film systems for x-ray detection II, model calculations. SPIE Med Imaging. 1984;454:128–128. [Google Scholar]
  • 6.Dobbins JT, Ergun DL, Rutz L, et al. DQE(f) of four generations of computed radiography acquisition devices. Med Phys. 1995;22:1581–1581. doi: 10.1118/1.597627. [DOI] [PubMed] [Google Scholar]
  • 7.Siewerdsen JH. Signal, noise, and detective quantum efficiency of a-Si:H flat-panel imagers. Ann Arbor, MI: University of Michigan; 1998. [DOI] [PubMed] [Google Scholar]
  • 8.Lee D, Cheung LK, Rodricks BG, et al. Improved imaging performance of a 14×17-in direct radiography system using a Se/TFT detector. Proc SPIE. 1998;3336:14–14. doi: 10.1117/12.317017. [DOI] [Google Scholar]
  • 9.Granfors PR. Performance characteristics of an amorphous silicon flat panel x-ray imaging detector. SPIE Med Imaging. 1999;3659:480–480. [Google Scholar]

Articles from Journal of Digital Imaging are provided here courtesy of Springer

RESOURCES