Skip to main content
International Orthopaedics logoLink to International Orthopaedics
. 2003 Jul 4;27(5):282–285. doi: 10.1007/s00264-003-0479-0

Implant wear and osteolysis with a hydroxylapatite-coated screw cup

E H van Haaren 1, I C Heyligers 1,2,
PMCID: PMC3461862  PMID: 12845483

Abstract

Radiographs of 72 uncemented hip implants with a mean follow-up of 6.5 years were analysed to measure wear of the implant. The wear rate was measured using the modified technique of Livermore and was found to be 0.13 mm per year (CI 95% 0.080–0.18). There was no statistical correlation between wear and age, gender and cup angle, but there was a correlation between wear and the number and size of cysts. This has not been previously published. In view of the risk of loosening and of a dramatic increase in the rate of wear when the femoral head articulates with the metal backing, it is important to plan revision of these cups in good time and to ensure regular radiographic surveillance of these patients.

Electronic Supplementary Material

Supplementary material is available in the online version of this article at http://dx.doi.org/10.1007/s00264-003-0479-0

Full Text

The Full Text of this article is available as a PDF (178.6 KB).

References

  • 1.BankstonClin Orthop 199531777671498 [Google Scholar]
  • 2.ChenClin Orthop 1995317447671495 [Google Scholar]
  • 3.D'Antonio Clin Orthop. 1992;285:102. [PubMed] [Google Scholar]
  • 4.Devane PA, Robinson EJ, Bourne RB, Rorabeck CH, Nayak NN, Horne JG (1997) Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J Bone Joint Surg [Am] 79:682–689 [DOI] [PubMed]
  • 5.Geesink RG, Hoefnagels NH (1995) Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg.[Br] 77:534–547 [PubMed]
  • 6.Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg [Br] 84:641–646 [DOI] [PubMed]
  • 7.Livermore J, Ilstrup D, Morrey B (1990) Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg [Am] 72:518–528 [PubMed]
  • 8.Morscher EW, Hefti A, Aebi U (1998) Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating. J Bone Joint Surg [Br] 80:267–272 [DOI] [PubMed]
  • 9.Nashed RS, Becker DA, Gustilo RB (1995) Are cementless acetabular components the cause of excess wear and osteolysis in total hip arthroplasty? Clin Orthop 317:19–28 [PubMed]
  • 10.Oosterbos Int Orthop. 2001;25:17. doi: 10.1007/s002640000211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Rokkum Acta Orthop Scand. 1998;69:253. doi: 10.3109/17453679809000925. [DOI] [PubMed] [Google Scholar]
  • 12.RossiClin Orthop 1995310987641466 [Google Scholar]
  • 13.Sochart Clin Orthop. 1999;363:135. [PubMed] [Google Scholar]
  • 14.Sychterz CJ, Moon KH, Hashimoto Y, Terefenko KM, Engh CA Jr, Bauer TW (1996) Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem. J Bone Joint Surg [Am] 78:1193–1200 [DOI] [PubMed]
  • 15.XenosClin Orthop 1995317297671492 [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table 1 (PDF 36 KB) (35.4KB, pdf)

Articles from International Orthopaedics are provided here courtesy of Springer-Verlag

RESOURCES