Computer-aided diagnosis system for bone scintigrams from Japanese patients: importance of training database

Hiroyuki Horikoshi 1, Akihiro Kikuchi 2,, Masahisa Onoguchi 2, Karl Sjöstrand 3, Lars Edenbrandt 4
PMCID: PMC3475966  PMID: 22729550

Abstract

Aim

Computer-aided diagnosis (CAD) software for bone scintigrams have recently been introduced as a clinical quality assurance tool. The purpose of this study was to compare the diagnostic accuracy of two CAD systems, one based on a European and one on a Japanese training database, in a group of bone scans from Japanese patients.

Method

The two CAD software are trained to interpret bone scans using training databases consisting of bone scans with the desired interpretation, metastatic disease or not. One software was trained using 795 bone scans from European patients and the other with 904 bone scans from Japanese patients. The two CAD softwares were evaluated using the same group of 257 Japanese patients, who underwent bone scintigraphy because of suspected metastases of malignant tumors in 2009. The final diagnostic results made by clinicians were used as gold standard.

Results

The Japanese CAD software showed a higher specificity and accuracy compared to the European CAD software [81 vs. 57 % (p < 0.05) and 82 vs. 61 % (p < 0.05), respectively]. The sensitivity was 90 % for the Japanese CAD software and 83 % for the European CAD software (n.s).

Conclusion

The CAD software trained with a Japanese database showed significantly higher performance than the corresponding CAD software trained with a European database for the analysis of bone scans from Japanese patients. These results could at least partly be caused by the physical differences between Japanese and European patients resulting in less influence of attenuation in Japanese patients and possible different judgement of count intensities of hot spots.

Keywords: Computer-aided diagnosis, Bone scintigram, Bone metastases, Artificial neural networks

Introduction

Patients with prostate or breast cancer often have bone metastases, and the bone metastases of the primary carcinomas may greatly affect the management and prognosis of the patient, so the diagnosis of metastatic disease is important [13]. The interpretation of bone scans and other diagnostic images are generally made visually as a pattern recognition task where for example the accumulation of a radiopharmaceutical is assessed. This type of task is subjective and largely dependent on the experience of the interpreter. Computer-aided diagnosis (CAD) software has been introduced in practical use as an objective approach to interpret diagnostic images. The computer interpretations can be used as a second opinion. CAD software has been developed for mammography, colonography and myocardial perfusion scintigrams [47]. In the USA, CAD software has been used for a part of clinical routine for detecting breast cancers by mammography for several years [8]. Lindahl et al. [9] reported that physicians improved their diagnostic capability significantly by the use of CAD software in for the analysis of myocardial perfusion scintigrams.

CAD software for the interpretation of bone scans have also been developed recently [1014]. Sadik et al. [12] reported that even an experienced physician obtained benefits for diagnosis through the CAD software and that differences in interpretations between observers decreased. This bone scan CAD software was trained to interpret bone scans using training databases consisting of bone scans from European patients with the desired image interpretation, metastatic disease or not. It is well known that American and European standard values are not appropriate to use in myocardial scintigrams for Japanese patients [15]. An assumption could be that it is of value to use a database consisting of bone scans from Japanese patients for training of CAD software for Japanese patients. The purpose of this study was to compare the diagnostic accuracy of two CAD systems, one based on a European and one on a Japanese training database, in a group of bone scans from Japanese patients.

Materials and methods

Patients

The evaluation group of this study comprised 257 patients (males 113, females 144; age 63± years), who had underwent bone scintigrams for detecting bone metastatic diseases in 2009. The primary cancers included 125 breast cancers, 64 prostatic cancers, 34 lung cancers, 9 bladder cancers, 4 gastric cancers, 5 renal cancers, 3 esophagus cancers, 2 testicular cancers, 2 ureteral cancers, 2 rectal cancers, 1 hepatoma, 1 endometrial cancer, 1 multiple myeloma, 1 bile duct cancer, 1 oropharyngeal cancer, 1 ovarian cancer and 1 lymphoma.

The final clinical assessment of each patient, made by an experienced physician, was used as the gold standard classification for the presence or absence of bone metastases. These assessments were based on increase and decrease of the high accumulation areas, changes in intensity and sizes by follow-up bone scans and other diagnostic imaging modalities such as MRI, CT, radiographic images or blood data.

CAD software

The CAD software interprets the bone scans automatically regarding metastatic disease. The method for interpretation consists of image analysis and artificial neural networks (ANNs). The CAD software is fed with the anterior and posterior images in digital format. The first steps are image segmentation, detection and characterization of high accumulation areas. The segmentation of the skeleton is performed using a normal atlas comprising 14 anatomical segments. The skeleton is normalized and high accumulation areas are detected in an iterative process. A characterization of high accumulation area is performed and the resulting features include intensity, shape and localization. The resulting image features are used as input to ANNs classifying high accumulation areas. A second set of neural networks is used to provide the ANN value on the basis of the results of the first set of networks. The neural networks are trained using databases of classified examples of bone scans. Two different European and Japanese training databases were used. After the training process, all bone scans of the evaluation group were processed with the two CAD softwares.

European training database

The European database consists of data of the cases which underwent examination at Sahlgrenska University Hospital in Gothenburg, Sweden from January 1999 to June 2002. The final number of cases was 795 including 514 males (70.2 ± 9.7 years old, age range 27–91 years) and 281 females (58.9 ± 13.1 years old, age range 25–92 years). Cases with technical problems during acquisition were excluded.

Japanese training database

The Japanese database is based on cases with diagnoses established at Gunma Prefectural Cancer Center in the years of 2007 and 2008. The final number of cases was 904 including 457 males (68.4 ± 10.4 years old, age range 20–91 years) and 447 females (58.5 ± 12.5 years old, age range 26–91 years). For diagnosis of metastasis, two expert radiologists made final diagnosis based on blood data, MRI or CT, and clinical course of 1 year or more. The cases included lung cancer, breast cancer, prostatic cancer, colorectal cancer and degenerative diseases. Cases of the ileal conduit were excluded.

Aiming at bone whole-body scan images (anterior and posterior images), a result is displayed through roughly two analysis flows comprising image analysis and ANN analysis as shown in Fig. 1. The analysis software BONENAVI (FUJIFILM RI Pharma Co., Ltd.) used in this study performs from analysis after data input to output almost automatically, and is excellent in convenience or repeatability.

Fig. 1.

Fig. 1

A 55-year-old man with lung cancer. Increased radiotracer uptake can mainly be seen in cranial bone, mandible, cervical spine, sternum, and right femur on whole-body scan (a, d). CAD software with European database (b, e) classifies as this patient as having no metastases (ANN 0.29, grade 2). However, CAD software with Japanese database (c, f) correctly classifies as having metastases (ANN 0.87, grade 4). CT imaging shows osteolytic bone metastasis corresponding to increased radiotracer uptake of cranial bone and right femur (not shown)

Image evaluation using CAD software

A cutoff ANN value of 0.5 was chosen to provide the CAD software with the binary classification of ‘‘bone metastases’’ or ‘‘no bone metastases.’’ All cases with ANN values were assigned following grades.

Grade 1 (ANN 0–0.25)

Some cases without findings of suspected bone metastases had high accumulation areas by bone scintigrams. For high accumulation areas, images showed normal accumulation patterns or degenerative change or bone fracture. Any patterns indicating bone metastases were also not observed by other modalities.

Grade 2 (ANN 0.26–0.50)

Cases in which bone metastasis may not be excluded securely by bone scintigrams. Although changes such as increase of intensity or expansion of the range were not observed in high accumulation areas at follow-up, changes due to bone fracture or degenerative diseases were observed by other modalities. It was determined that bone metastases were less suspicious by clinical judgment.

Grade 3 (ANN 0.51–0.75)

Cases with bone metastases suspected by bone scintigrams. Changes such as increase of intensity or expansion of the range were observed in high accumulation areas at follow-up, or findings suspecting bone metastases were obtained by other modalities and bone metastases were clinically suspected.

Grade 4 (ANN: 0.76–1.00)

Cases with bone metastases highly suspected by bone scintigrams. Typical bone metastatic patterns were observed in high accumulation areas. Typical bone metastatic patterns were observed by other modalities as well.

Bone scintigrams

Whole-body bone scan images, anterior and posterior views, were acquired 3–5 h after an intravenous injection of 740 MBq Technetium-99 m methylene diphosphonate (FUJIFILM RI Pharma Co., Ltd., Japan). A dual-detector gamma-camera system equipped with a low energy high resolution parallel hole collimators was used Siemens ECAM. The acquisition conditions of the whole-body scan were set as follows: scan speed 15 cm/min, matrix size 256 × 1024, energy window of energy peak 140 keV and window width 15 %.

Statistical analysis method

Cases classified as Grade 1 or Grade 2 were regarded as having no metastases and Grade 3 or Grade 4 had metastases. Differences in sensitivity, specificity and accuracy between the two CAD were tested using a McNemar test.

Results

The final clinical assessment showed 42 cases with bone metastases and 215 cases without bone metastases. The number of true positive classifications was 38 cases for the Japanese CAD software and 35 cases for the European CAD software and this difference was not statistically significant. The number of true negative classifications was 174 cases for the Japanese CAD software and 122 cases for the European CAD software. The classifications of the CAD software for the Japanese and European databases are presented in Table 1. The sensitivity, specificity, and accuracy for the Japanese and European CAD software are presented in Table 2. The Japanese CAD software showed a higher specificity and accuracy compared to the European CAD software [81 vs. 57 % (p < 0.05) and 82 vs. 61 % (p < 0.05), respectively]. The sensitivity was 90 % for the Japanese CAD software and 83 % for the European CAD software (n.s).

Table 1.

Classifications of the CAD software for the Japanese and European databases

Grade 1 2 3 4 Total
Japanese database CAD
 Bone metastases 4 0 3 35 42
 No bone metastases 163 11 7 34 215
European database CAD
 Bone metastases 3 4 2 33 42
 No bone metastases 67 55 30 63 215

n numbers

Table 2.

Comparison of CAD software according to Japanese and European databases

Japan database CAD (%) European database CAD (%)
Sensitivity 90 83
Specificity 81* 57*
Accuracy 82* 61*

p < 0.05

Discussion

Usefulness of the CAD system for bone scintigrams using the ANN technology used in this study was reported by the Sadik et al. [1012]. A high disease detection rate and sensitivity of almost 90 % were indicated in the report [12]. In the analysis software used in this study, skeletal segmentation, detection of a high accumulation area, characterization and an ANN algorithm were used.

In this study, difference in the analysis results due to the difference in the races of the databases was compared, showing that a better diagnostic result for Japanese patients could be obtained by use of the Japanese database (Fig. 1). As long as seeing the reference analysis result, the dose and collection conditions of the radiopharmaceutical of each database were almost equivalent. The system resolution was 8.1 mm for Maxxus and 7.4 mm for ECAM, which were comparable. In analysis, the ANN value in the high accumulation area has greatly accounted for the final result, and the distribution and intensity may be affected by difference in the absorption of radiation due to difference in the physical constitution.

In judgment using the ANN value, the count intensity of the high accumulation area is a key factor. For the count intensity, influence of internal absorption depending on the physical constitution becomes a major variation factor if the dose and the collection conditions are the same. Especially absorption in the trunk area tends to be different, and this area varies significantly by races. The high accumulation area in the trunk area is critical in characterization by the ANN [13], and since Japanese are small in stature on the average compared with Europeans, influence of internal absorption is less than in Europeans, and the count intensity in the high accumulation area may tend to be higher consequently. Using the European database tends to result in false positive, while sensitivity and specificity may have been improved using the Japanese database including the characteristics of the Japanese physical constitution.

Conclusion

It was suggested that the precision, especially the specificity and the accuracy of the CAD system could be improved significantly by creating a Japanese database when analyzing the Japanese data.

Acknowledgments

We appreciate the advices and the cooperation from Jens Richter and Thomas Jonsson at EXINI Diagnostics AB, Lund Sweden during this study.

Conflict of interest

Kikuchi is employed in FUJIFILM RI Pharma Co., Ltd. Japan, which distributes the software BONENAVI and Sjöstrand and Edenbrandt are employed in EXINI Diagnostics AB, Lund Sweden, which has developed the EXINI bone and BONENAVI software.

Open Access

This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

References

  • 1.Maffioli L, Florimonte L, Pagani L, Butti I, Roca IN. Current role of bone scan with phosphonates in the follow-up of breast cancer. Eur J Nucl Med Mol Imaging. 2004;31(Suppl 1):S143–S148. doi: 10.1007/s00259-004-1537-6. [DOI] [PubMed] [Google Scholar]
  • 2.Coleman R, Heidenreich A, Bell R. Managing metastatic bone disease: three case studies. Semin Oncol. 2004;31:83–86. doi: 10.1053/j.seminoncol.2004.07.028. [DOI] [PubMed] [Google Scholar]
  • 3.Coleman RE. Metastatic bone disease: clinical features, pathophysiology and treatment strategies. Cancer Treat Rev. 2001;27:165–176. doi: 10.1053/ctrv.2000.0210. [DOI] [PubMed] [Google Scholar]
  • 4.Doi K. Computer-aided diagnosis in medical imaging: historical review, current status and future potential. Comput Med Imaging Graph. 2007;31:198–211. doi: 10.1016/j.compmedimag.2007.02.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Gromet M. Comparison of computer-aided detection to double reading of screening mammograms: review of 231,221 mammograms. AJR. 2008;190:854–859. doi: 10.2214/AJR.07.2812. [DOI] [PubMed] [Google Scholar]
  • 6.Goldin JG, Brown MS, Petkovska I. Computer-aided diagnosis in lung nodule assessment. J Thorac Imaging. 2008;23:97–104. doi: 10.1097/RTI.0b013e318173dd1f. [DOI] [PubMed] [Google Scholar]
  • 7.Kawamoto K, Houlihan CA, Balas EA, Lobach DF. Improving clinical practice using clinical decision support systems: a systemic review of trials to identify features critical to success. BMJ. 2005;330:765–768. doi: 10.1136/bmj.38398.500764.8F. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Freer TW, Ulissey MJ. Screening mammography with computer-aided detection: prospective study of 12,860 patients in a community breast center. Radiology. 2001;220:781–786. doi: 10.1148/radiol.2203001282. [DOI] [PubMed] [Google Scholar]
  • 9.Lindahl D, Lanke J, Lundin A, Palmer J, Edenbrandt L. Improved classifications of myocardial bull’s-eye scintigrams with computer-based decision support systems. J Nucl Med. 1999;40:96–101. [PubMed] [Google Scholar]
  • 10.Sadik M, Hamadeh I, Nordblom P, Suurkula M, Hoglund P, Ohlsson M, et al. Computer-assisted interpretation of planar whole-body bone scans. J Nucl Med. 2008;49(12):1958–1965. doi: 10.2967/jnumed.108.055061. [DOI] [PubMed] [Google Scholar]
  • 11.Sadik M, Suurkula M, Hoglund P, Jarund A, Edenbrandt L. Quality of planar whole-body bone scan interpretations: a nationwide survey. Eur J Nucl Med Mol Imaging. 2008;35:1464–1472. doi: 10.1007/s00259-008-0721-5. [DOI] [PubMed] [Google Scholar]
  • 12.Sadik M, Suurkula M, Höglund P, Järund A, Edenbrandt L. Improved classifications of planar whole-body bonescans using a computer-assisted diagnosis system: a multicenter, multiple-reader, multiple-case study. J Nucl Med. 2009;50(3):368–375. doi: 10.2967/jnumed.108.058883. [DOI] [PubMed] [Google Scholar]
  • 13.Sadik M, Jakobsson D, Olofsson F, Ohlsson M, Suurkula M, Edenbrandt L. A new computer-based decision-support system for the interpretation of bone scans. Nucl Med Commun. 2006;27:417–423. doi: 10.1097/00006231-200605000-00002. [DOI] [PubMed] [Google Scholar]
  • 14.Yin T-K, Chiu N-T. A computer-aided diagnosis for locating abnormalities in bone scintigraphy by a fuzzy system with a three-step minimization approach. IEEE Trans Med Imag. 2004;23:639–654. doi: 10.1109/TMI.2004.826355. [DOI] [PubMed] [Google Scholar]
  • 15.Nakajima K, Okuda K, Kawano M, Matsuo S, Slomka P, Germano G, et al. The importance of population-specific normal database for quantification of myocardial ischemia: comparison between Japanese 360 and 180-degree databases and a US database. J Nucl Cardiol. 2009;16:422–430. doi: 10.1007/s12350-009-9049-1. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Annals of Nuclear Medicine are provided here courtesy of Springer

RESOURCES