ISway: a sensitive, valid and reliable measure of postural control

Martina Mancini 1,2,, Arash Salarian 1, Patricia Carlson-Kuhta 1, Cris Zampieri 3, Laurie King 1, Lorenzo Chiari 2, Fay B Horak 1
PMCID: PMC3481400  PMID: 22913719

Abstract

Background

Clinicians need a practical, objective test of postural control that is sensitive to mild neurological disease, shows experimental and clinical validity, and has good test-retest reliability. We developed an instrumented test of postural sway (ISway) using a body-worn accelerometer to offer an objective and practical measure of postural control.

Methods

We conducted two separate studies with two groups of subjects. Study I: sensitivity and experimental concurrent validity. Thirteen subjects with early, untreated Parkinson’s disease (PD) and 12 age-matched control subjects (CTR) were tested in the laboratory, to compare sway from force-plate COP and inertial sensors. Study II: test-retest reliability and clinical concurrent validity. A different set of 17 early-to-moderate, treated PD (tested ON medication), and 17 age-matched CTR subjects were tested in the clinic to compare clinical balance tests with sway from inertial sensors. For reliability, the sensor was removed, subjects rested for 30 min, and the protocol was repeated. Thirteen sway measures (7 time-domain, 5 frequency-domain measures, and JERK) were computed from the 2D time series acceleration (ACC) data to determine the best metrics for a clinical balance test.

Results

Both center of pressure (COP) and ACC measures differentiated sway between CTR and untreated PD. JERK and time-domain measures showed the best test-retest reliability (JERK ICC was 0.86 in PD and 0.87 in CTR; time-domain measures ICC ranged from 0.55 to 0.84 in PD and from 0.60 to 0.89 in CTR). JERK, all but one time-domain measure, and one frequency measure were significantly correlated with the clinical postural stability score (r ranged from 0.50 to 0.63, 0.01 < p < 0.05).

Conclusions

Based on these results, we recommend a subset of the most sensitive, reliable, and valid ISway measures to characterize posture control in PD: 1) JERK, 2) RMS amplitude and mean velocity from the time-domain measures, and 3) centroidal frequency as the best frequency measure, as valid and reliable measures of balance control from ISway.

Keywords: Postural control, Accelerometers, Inertial sensors, Parkinson’s disease

Background

Postural control is the foundation of our ability to stand and to walk independently. Deterioration in postural control due to normal ageing or neurodegenerative disease such as Parkinson’s disease (PD) is associated with an increase in risk of falls incurred during activities of daily life [1,2]. Deterioration in balance control predisposes 68% of people with PD to fall at least once each year [3]. Although it is well-known that subjects with PD show postural instability in advanced stages of the disease [4-7], very few studies have investigated whether it is possible to identify abnormal balance in early stages of the disease, prior to starting antiparkinsonian medication [8-10]. Early identification of abnormal balance in patients with PD is important because new neuroprotective medications are currently being tested to slow the progression of PD. Neuroprotection needs to begin early in the disease, prior to significant loss of neurons. Also, early in the disease, it is often difficult to distinguish idiopathic PD from other, Parkinson-plus syndromes that have very different Prognoses and it is possible that sway characteristics will differ among these different basal ganglia diseases.

Currently, the most common way to evaluate postural control in the clinic is to use clinical rating scales that are limited by clinicians bias, insensitivity to mild impairments (ceiling effects), and poor reliability [11-13]. These limitations are serious concerns for clinicians and researchers who want to monitor disease progression, determine intervention efficacy or treat people with mild balance deficits [14]. Technology currently available for clinicians and researchers to measure postural control generally uses force plate analysis of center of pressure (COP) displacement during quiet stance [6,15]. Experimental studies have demonstrated the sensitivity of COP measures to postural disorders such as Parkinson’s disease [8,9] and to fall risk in the elderly [16,17]. However, force-plate-based posturography (or stabilometry) is quite large and expensive and requires a proper installation that may not be practical for clinical use.

Body-worn accelerometers (ACC) have been proposed as a portable, low-cost alternative to a force plate for measurements of postural sway [2,18-21]. Recently, we showed that ACC-based measures of postural sway are sensitive to balance disorders in patients with untreated PD [10]. Moreover, Najafi et al., demonstrated that ACC-based derived measures of center of mass during postural sway detect deterioration in patients with diabetic neuropathy [2]. These results suggest that such measures may provide a sensitive means of measuring subtle balance deficits in clinical settings.

Despite the potential advantages of accelerometric systems in clinical practice, they still have several drawbacks, such as the need to pre-process data and the question of how to translate sway measures into clinically understandable outcomes. However, the major limitation is that there is no consensus as to which sway-related measures should be considered. Studies have shown that root mean square (RMS) of the acceleration signal can be sensitive to test conditions (eye closure, standing on one foot), to ageing, and to history of falls [22-24]. We recently presented a relatively new measure of sway, “JERK”, as the most discriminative measure to differentiate sway in patients with untreated PD compared to age-matched control subjects [10].

In order to make ACC-based measures useful for clinical applications, it is important to assess their validity, sensitivity, and reliability, compared to gold standard lab and clinical assessments [25,26]. The relationship between the same postural sway measures calculated from the force plate COP and from ACC that would support the experimental validity of ACC measures have only been reported in three studies, the first by Adlerton et al. [20] in one-legged stance trials in healthy adults, the second by Najafi et al. [2] in diabetic patients, and the third by Whitney et al., [18] in healthy control subjects during different sensory conditions. No studies have shown experimental validity of ACC-based measures and force plate-based measures of postural sway for subjects with very mild balance deficits, such as those with untreated PD. Our previous study on ACC measures of postural control only showed which are the best measures that discriminate between untreated PD and age-matched control [10]. Moreover, to our knowledge, only one study has presented test-retest reliability of balance measures calculated from an ACC placed on the belt of young, healthy adults [27]. There are no studies that have evaluated test-retest reliability of postural sway in patients with PD and elderly subjects, or studies that have systematically examined the relative reliability of amplitude, velocity, and frequency components of sway. Finally, it is important that a new objective test of postural control is validated with clinical scales. For patients with PD, the most common clinical measure of balance impairment is the postural instability and gait disability (PIGD) subscore of the Unified Parkinson’s Disease Rating Scale (UPDRS) [28].

The goal of this study was to develop and validate a practical tool that allows clinicians to measure postural sway in a clinical setting with body-worn accelerometers. We call our tool the instrumented sway system (ISway). Our vision is that this tool will provide reliable, automatic analysis of sway that is sensitive, accurate, robust, and consistent, without the need for clinical experts to deal with the raw data. To achieve this objective, we carried out two studies in order to determine: i) the sensitivity and experimental concurrent validity of ACC compared to force-plate measures of postural sway; and ii) test-retest reliability and clinical concurrent validity of ACC-based measures compared to the PIGD. From this information, we recommend a subset of the most sensitive, reliable, and valid ISway measures to characterize postural control in PD.

Methods

Study design

We carried out two studies. The first study was performed in the motion analysis laboratory, to determine the sensitivity and experimental concurrent validity of ACCs compared to force-plate measures (Study I). The second study took place in the hospital neurology clinic to determine the test-retest reliability and clinical concurrent validity of the proposed automatic clinical system, ISway (Study II).

Subjects

Study I: sensitivity and experimental concurrent validity

To assess the sensitivity of sway measures to detect untreated PD and the relationship between ACC and COP measures, 13 subjects with idiopathic PD and 13 age-matched healthy control subjects participated in the study (Table 1). One control subject was excluded because he was diagnosed with a neurological disorder after he had been selected to participate, leaving the control group with 12 subjects. Only subjects with PD who were early- to mid-stage in the disease course and had never been treated with dopaminergic or other anti-parkinsonian medication were invited to participate. The PD and control groups showed no significant difference in age or body mass index (BMI) but the PD group had more male (7) subjects than control group (5).

Table 1.

Summary of Subject Characteristics for Study I and II

  n UPDRS Age Medications
Study I
13 PD
28.1(±11.2)
60.4 ± 8.5 years
None
 
12 CTRL
 
60.2 ± 8.2 years
 
Study II
17 PD
28.3 (±10.4)
67.1 ± 7.3 years
PD ON medication
  17 CTRL   67.9 ± 6.1 years  

Study II: test-retest reliability and clinical concurrent validity

To assess test-retest reliability of the measures and the relationship with clinical scales, a different group of 17 PD and 17 age-matched control subjects were tested in the neurology clinic by a research assistant. The patients with PD were tested in their ON medication condition (Table 1). The PD and control groups showed no significant difference in age or BMI, but the PD group had more males (12) than the control group (6).

In both studies, subjects were excluded if they presented with any neurological disorders other than PD or if they had any other condition that could affect their balance. Patients were clinically rated by a trained examiner on the Motor Section (III) of the UPDRS and the Hoehn and Yahr Scale immediately before the experimental sessions. The UPDRS Part III consists of 23 items related to bradykinesia, rigidity, tremor and posture and gait signs of PD, rated on a 4-point scale [28]. The PIGD consists of the sum of 4 UPDRS sub-items, posture, gait, sit-to-stand and pull test, with score from 0 (normal) to 16 (severe) [7].

Measurement protocol and data acquisition

Study I: sensitivity and experimental concurrent validity

All participants were instructed to maintain an upright standing position on a force-plate (AMTI OR6-6, Watertown, MA), with arms crossed and heel-to-heel distance fixed at 10 cm. Feet were allowed to be externally rotated at a comfortable amount for each subject [29], and all the participants wore similar athletic shoes. Initial stance position was consistent from trial to trial by tracing foot outlines on the force-plate.

Subjects wore a MTX Xsens sensor (49A33G15, Xsens, Enschede, NL) with 3-D accelerometers (±1.7 g range) mounted on the posterior trunk at the level of L5, near the body center of mass. The sensing axes were oriented along the anatomical antero-posterior (AP), medio-lateral (ML), and vertical directions. A total of three, 2-minute trials were performed with eyes open gazing straight ahead at an art poster.

The COP displacement was calculated from the ground reaction forces recorded by the force-plate at a 100-Hz sampling frequency and after applying a 10-Hz cut-off, zero-phase, low-pass Butterworth filter. Acceleration signals were collected with a 50-Hz sampling frequency, transformed to a true horizontal-vertical Cartesian coordinate system [30] and filtered with a 3.5 Hz cut-off, zero-phase, low-pass Butterworth filter. This filter was applied also to the COP in order to eliminate possible contributions of tremor at rest, a well-known PD symptom, which may be present in the range from 4-to-7 Hz [31].

Study II: test-retest reliability and clinical concurrent validity

Subjects wore the same MTX Xsens sensors on the posterior trunk at the level of L5. To ensure a consistent foot-width position, we constructed a styrofoam wedge that was placed between the feet before each trial. Three 30 s trials of quiet standing were collected.

The sensor was removed after finishing the 3 ISway trials. After 30 min resting in a chair the sensor was put back on and the protocol was repeated. We assumed that the subjects’ performances remained the same within this time period. The same examiner used the same device and the same protocol to test the subject for the second time.

Automatic instrumented Sway in the clinic: ISway

For the experiments in the clinic, subjects wore a portable data-receiver (X-Bus) wired to the MTX XSens sensors. The sensor recorded 3-D linear accelerations and angular velocity while the controller continuously, wirelessly streamed data to a laptop via Bluetooth.

A custom MATLAB (MathWorks, Nantick, MA) graphical interface was built to acquire, store and analyze different components of balance. The software also automatically compared each subject’s balance-related measures compared to normative ranges (based on metrics from healthy subjects) and uploaded the data to a server for additional analysis.

In addition to the previously described pre-processing of the acceleration signals, the algorithm includes an automatic inspection of acceleration signals to make sure that trials in which subjects may have lost their balance were excluded from the analysis. Specifically, signals were divided into three 10s windows and the standard deviation (SD) was computed on each window. If the SD of at least one of the three windows exceeded 5 times the SD of one of the other windows, the trial was discarded. After this automatic check, a total of 13 ACC measures were computed from the 2D acceleration time series, similar to COP analysis (details in Table 2). In the time-domain, we computed six measures that characterized the ACC trajectory and one measure that estimated the area covered by the 2D ACC trace. In the frequency-domain, spectral properties were assessed by five measures: one measure that quantifies the total power of the ACC signal, one measure that estimates the variability of the frequency content of the power spectral density of the ACC, and three measures of characteristic frequencies in the power spectral density of the ACC signal. We included jerkiness of sway, as described in a previous paper [10].

Table 2.

Summary of the extracted measures

Measure abbreviation Description
JERK
Sway jerkiness, time derivative of acceleration [m2/s5] JERK=120tdACCLAPdt2+dACCLMLdt2
Time-domain measures
 
DIST
Mean distance from center of COP (ACC) trajectory [mm] ([m/s2])
RMS
Root mean square of COP (ACC) time series [mm] ([m/s2])
PATH
Sway path, total length of COP (ACC) trajectory [mm] ([m/s2])
RANGE
Range of COP displacement (acceleration) [mm] ([m/s2])
MV
Mean velocity COP: PATH/trial duration [m/s]; ACC: ACCLAP2+ACCLML2 [mm/s]
MF
Mean frequency, the number, per second, of loops that have to be run by the COP (ACC), to cover a total trajectory equal to PATH (MF = PATH/ (2*π*DIST*trial duration) (Hz)
AREA
Sway area, computed as the area spanned from the COP (ACC) per unit of time [mm2/s] ([m2/s5])
Frequency-domain measures
 
PWR
Total power [mm2] ([m2/s4])
F50
Median frequency, frequency below which the 50% of PWR is present (Hz)
F95
95% power frequency, frequency below which the 95% of PWR is present (Hz)
CF
Centroidal frequency (Hz)
FD Frequency dispersion (−)

Data analysis

Algorithms for signal analysis and statistical evaluation of the outcomes were written in MATLAB. Differences between untreated PD and control groups were determined by using a t-test. Differences were assumed significant when P < 0.05. A Pearson Product moment correlation was used to assess both the relationship between COP and ACC metrics in Study I and the relationship between ACC metrics and clinical scores (UPDRS III and sub-items) in Study II. To evaluate test-retest reliability of the ISway, Intra-Class Correlation (ICC) was used [32]. Since the same subjects and same device was used for reliability, an ICC(1,1) was used. The ρ and 95% confidence intervals were reported.

Results

Study I: sensitivity and experimental concurrent validity

ACC-based measures of sway were just as sensitive as COP measures in differentiating between the untreated PD and control groups. The COP and ACC traces of a representative control subject, a mild, untreated PD subject (UPDRS = 17), and a moderate, untreated PD subject (UPDRS = 35) during a quiet stance trial are illustrated in Figure 1. The size and jerkiness of both the COP and ACC traces are larger in the mild untreated PD subject and even larger in the moderate, untreated PD subject compared to control subject.

Figure 1.

Figure 1

Center of pressure (left panel) and acceleration (right panel) traces in the horizontal plane for three representative subjects.

Consistent with the representative trials in Figure 1, most time- and frequency-domain measures of both ACC and COP were sensitive to untreated PD. Table 3 summarizes p-values of the time- and frequency-domain outcome measures comparing untreated PD and control subjects, as well as the correlations between ACC and COP measures.

Table 3.

Sensitivity of COP and ACC-based measures

  COP
ACC
Correlation
Control
PD
p values Control
PD
p values r p
Mean SEM Mean SEM Mean SEM Mean SEM
Jerk
 
 
 
 
NA
0.241
0.023
0.394
0.053
0.0003
 
 
Trajectory Measures
DIST
4.370
0.329
6.261
0.423
0.002
0.062
0.005
0.088
0.010
0.02
0.73
0.0001
RMS
5.141
0.371
7.329
0.552
0.003
0.073
0.006
0.106
0.012
0.02
0.74
0.0000
PATH
944.015
89.277
817.707
62.661
0.25
17.183
1.028
22.035
2.128
0.05
0.46
0.200
RANGE
32.247
2.750
40.824
3.385
0.06
0.421
0.031
0.610
0.065
0.01
0.64
0.0007
MV
7.867
0.744
6.814
0.522
0.26
1.328
0.229
1.789
0.241
0.18
0.12
0.56
MF
0.290
0.018
0.181
0.016
0.0001
0.393
0.029
0.378
0.045
0.78
0.33
0.11
Area Measures
AREA
10.699
1.581
13.138
2.121
0.36
0.002
0.0003
0.005
0.001
0.004
0.68
0.002
Frequency Measures
PWR
9870.8
1761.7
12966.1
1585.1
0.2
2.009
0.269
3.178
0.479
0.04
0.41
0.05
F50
0.410
0.015
0.298
0.012
0.00009
0.385
0.014
0.305
0.012
0.0003
0.86
0.000001
F95
1.518
0.069
0.990
0.057
0.0007
1.849
0.121
1.523
0.128
0.07
0.49
0.01
CF
0.666
0.032
0.451
0.023
0.0002
0.722
0.043
0.591
0.038
0.03
0.47
0.02
FD 0.767 0.009 0.827 0.009 0.00005 0.777 0.009 0.839 0.008 0.0003 0.89 0.000001

Those measures in boldface were significantly different between untreated PD and control subjects. Correlations between COP and ACC measures are also reported.

Most ACC measures of sway were significantly correlated with COP measures of sway, attesting to their concurrent validity (Table 3). Only two measures, MV and MF were not significantly correlated with COP.

Study II: test-retest reliability and clinical concurrent validity

Table 4 summarizes the test-retest reliability of ACC measures for the control and PD groups. Overall, the PD group showed higher test-retest reliability than the control group. The JERK and time-domain measures showed the best reliability (highlighted in Table 4), whereas the frequency-domain measures had poorer test-retest reliability.

Table 4.

Test-Retest Reliability of ACC-based measures

  CTRL
PD
Test I
Test II
ICC(1-1)
95%Cl bounds
Test I
Test II
ICC(1-1)
95%Cl bounds
Mean SEM Mean SEM ρ lower upper Mean SEM Mean SEM ρ lower upper
Jerk
0.065
0.007
0.067
0.007
0.87
0.67
0.95
0.235
0.110
0.188
0.066
0.86
0.66
0.95
Trajectory Measures
DIST
0.043
0.003
0.048
0.004
0.70
0.34
0.88
0.089
0.020
0.074
0.014
0.84
0.61
0.94
RMS
0.052
0.003
0.057
0.005
0.71
0.35
0.89
0.108
0.026
0.089
0.017
0.83
0.59
0.93
PATH
4.428
0.220
4.484
0.234
0.89
0.72
0.96
8.232
1.285
7.286
0.909
0.81
0.56
0.93
RANGE
0.275
0.016
0.283
0.020
0.74
0.41
0.90
0.567
0.150
0.494
0.102
0.82
0.58
0.93
MV
0.104
0.012
0.122
0.013
0.68
0.31
0.87
0.211
0.049
0.176
0.029
0.75
0.44
0.90
MF
0.58
0.03
0.55
0.05
0.60
0.17
0.84
0.54
0.04
0.54
0.04
0.55
0.12
0.81
Area Measures
AREA
0.0019
0.0002
0.0021
0.0002
0.76
0.45
0.91
0.0127
0.0059
0.0078
0.0031
0.73
0.40
0.89
Frequency Measures
PWR
0.41
0.04
0.47
0.08
0.61
0.19
0.84
3.54
2.24
2.04
1.33
0.85
0.64
0.94
F50
0.43
0.03
0.41
0.05
0.30
-0.20
0.68
0.36
0.01
0.37
0.02
0.35
-0.13
0.70
F95
1.95
0.10
1.92
0.10
0.59
0.17
0.83
1.78
0.16
2.00
0.14
0.67
0.31
0.87
CF
0.78
0.05
0.76
0.05
0.61
0.19
0.84
0.69
0.05
0.76
0.06
0.69
0.34
0.87
FD 0.76 0.01 0.79 0.02 0.25 -0.25 0.65 0.78 0.01 0.80 0.01 0.61 0.20 0.84

Table 5 summarizes the clinical concurrent validity. Several ACC measures showed significant correlation with the PIGD sub-score of the UPDRS III. All except one time-domain measure (MF), one frequency-domain measure (PWR), and JERK were significantly and positively correlated with the PIGD sub-score related to clinical postural instability. No significant correlations were found between ACC measures and the total Motor UPDRS.

Table 5.

Correlation of ACC measures with UPDRS III and PIGD

  UPDRS III PIGD
JERK
0.29 (p = 0.29)
0.55 (p = 0.03)
Time-domain measures
 
 
DIST
0.29 (p = 0.29)
0.57 (p = 0.02)
RMS
0.29 (p = 0.28)
0.57 (p = 0.02)
PATH
0.22 (p = 0.42)
0.50 (p = 0.05)
RANGE
0.29 (p = 0.30)
0.56 (p = 0.03)
MV
0.29 (p = 0.28)
0.63 (p = 0.01)
MF
−0.09 (p = 0.72)
−0.37 (p = 0.17)
AREA
0.29 (p = 0.29)
0.55 (p = 0.03)
Frequency-domain measures
 
 
PWR
0.30 (p = 0.27)
0.54 (p = 0.04)
F50
−0.17 (p = 0.54)
−0.18 (p = 0.49)
F95
−0.24 (p = 0.38)
−0.42 (p = 0.12)
CF
−0.18 (p = 0.50)
0.34 (p = 0.20)
FD 0.23 (p = 0.41) 0.31 (p = 0.26)

Discussion

ISway is an innovative tool that can allow clinicians to objectively measure posture control during stance easily and quickly in a clinical setting. The approach of using accelerometers on the belt to quantify postural sway was shown to be sensitive, valid, and reliable for patients with PD.

The JERK of sway acceleration was found to be the most sensitive measure to discriminate untreated PD and control subjects [10]. The ISway measures also showed significant differences between the untreated PD and age-matched control groups for 5 out of 7 time-domain and 4 out of 5 frequency-domain ACC measures. Similarly, significant differences between groups were observed using COP measures: 3 out of 7 time-domain and 4 out of 5 frequency-domain COP measures. Subjects with untreated PD showed larger sway amplitude and area, larger sway jerkiness but lower frequencies of sway than control subjects. A more detailed discussion of ACC differences in postural sway between mild PD and control groups is presented in a separate paper [10].

ISway measures of postural sway were validated by force-plate measures of COP displacement. Many, but not all, ISway measures were correlated with the gold-standard laboratory measures of sway from a force-plate. If the body was moving like an inverted pendulum, a correlation close to 1 would be expected between trunk acceleration and COP displacement [33]. In fact, highly correlated measures of COP and center of mass amplitudes have also been reported during quiet stance [34,35]. Moderate to good correlations were found between COP and ACC parameters, except for MV and MF. One explanation can be that we used a different formula to calculate MV for COP and ACC; for ACC, we integrated the acceleration signal to get the velocity and for COP, we differentiated the COP displacement. Another explanation for the difference may be that subjects do not sway strictly as inverted pendulums [36]. In fact, even quiet stance in young, healthy subjects includes some hip strategy and the amount of hip strategy used to control stance posture have been shown to increase with age.

To our knowledge, this is the first paper reporting reliability of postural sway measures, in control or PD subjects, using an accelerometer-based approach. As expected, ICCs were larger in PD subjects because there was larger intra-subject variability in sway among PD subjects than control subjects, without any increase in variability of performance across sessions. In the PD group, 6 ACC parameters out of 13 showed excellent reliability (ICC > .80), 5 good (.60 < ICC < .80), 1 moderate (.40 < ICC < .60), and 1 poor reliability (ICC < .40). In control subjects, 2 out of 13 ACC parameters showed excellent reliability, 8 good, 1 moderate, and 2 parameters showed poor reliability. In general, JERK and time-domain parameters showed better reliability than frequency-domain parameters, and this has been previously reported in COP based measure of postural sway [37,38]. Although it is possible that the poor reliability of sway frequency metrics in people with PD, compared to other spatial-temporal metrics, could be due to fluctuating tremor, we think this unlikely because we filtered the acceleration signals below 3.5 Hz. In addition, healthy control subjects showed poorer reliability of the frequency metrics. A possible explanation for the worse test-retest reliability of frequency-domain parameters might be the variation in a subject’s balance strategy across the testing sessions, or might be attributed to the shorter trial duration of trials in study II.

The ISway measures showed good clinical validity as the measures are related to clinical scores. JERK, 6 out of 7 time-domain measures, and total power significantly correlated significantly with PIGD, a UPDRS III subscore that is used in clinical trials to evaluate postural instability and gait disability based on clinical rating of postural alignment, the pull test of postural stepping response, sit to stand and gait [39]. However, ISway measures were not significantly related with the overall Motor UPDRS III. Since UPDRS motor scale is composed of scores mostly related to tremor, bradykinesia and rigidity, its poor correlation with ISway might suggest that the balance deficiency in early PD is not explained by those symptoms. In fact, our recent meta-analysis of long-term effects of deep brain stimulation surgery in patients with PD show independent decline in PIGD, even when the rest of the UPDRS cardinal signs slowed their decline after surgery [40].

The potential application of the ISway is not to be limited to testing subjects with PD. The ISway provides a large number of measures that automatically, fully characterize body sway in amplitude, smoothness, and frequency; measures that are relevant for testing any individual with balance deficits. In fact, it is likely that a different subset of measures in the ISway might be sensitive to different neurological or musculoskeletal constraints. For example, JERK has been shown to be lower than normal in patients with mild multiple sclerosis [41]. Further studies are needed to determine the best subset of postural sway parameters that can predict future falls or disability during daily activities.

Conclusion

Previous results [23,24,42-44] suggest that postural sway could be characterized by three relatively independent characteristics: amplitude, velocity and frequency. Our results showed the most sensitive and reliable ISway measures to characterize posture control in PD are: 1) JERK, for its excellent reliability and sensitivity; 2) RMS for its best sensitivity and reliability among the time-domain measures and MV, for the best correlation with clinical score (even if it is not sensitive to untreated PD); and 3) CF as the best compromise from the frequency-domain measures (sensitive, reliable, and correlated with COP CF).

In summary, the present accelerometric-based measurement of sway offers an inexpensive and efficient alternative for quantifying posture control and provides excellent opportunities for on-line sway analyses. The method is reliable and can be used together with clinical balance and mobility tests in various circumstances, particularly in outcome and screening studies.

Competing interests

ISway is used in Mobility Lab by APDM. Drs. Horak and Salarian have significant financial interests in APDM, a company that has a commercial interest in the results of this research and technology. This potential conflict of interest has been reviewed and managed by OHSU and the Integrity Oversight Council.

Authors´ contribution

MM participated to study design and data collection, was responsible for data analysis and interpretation, drafting the manuscript. AS participated to study design and conception, participated in data analysis and interpretation, revising manuscript. PCK, CZ, and LK were responsible for data collection and interpretation, revising critically the manuscript. LC and FBH were responsible for study design and conception, drafting and revising critically the manuscript. All authors read and approved the final manuscript.

Contributor Information

Martina Mancini, Email: mancinim@ohsu.edu.

Arash Salarian, Email: salarian@ohsu.edu.

Patricia Carlson-Kuhta, Email: carlsonp@ohsu.edu.

Cris Zampieri, Email: Cristiane.Zampieri-gallagher@nih.gov.

Laurie King, Email: kingla@ohsu.edu.

Lorenzo Chiari, Email: lorenzo.chiari@unibo.it.

Fay B Horak, Email: horakf@ohsu.edu.

Acknowledgements

We thank Kelsey Priest and Triana Nagel for scheduling and testing subjects.

This publication was made possible with support from Kinetics Foundation, NIH grants RC1 NS068678 and R37 AG006457, and in part by the Intramural Research Program of the NIH, Clinical Center, Rehabilitation Medicine Department.

References

  1. Melzer I, Benjuya N, Kaplanski J. Postural stability in the elderly: a comparison between fallers and non-fallers. Age Ageing. 2004;33:602–607. doi: 10.1093/ageing/afh218. [DOI] [PubMed] [Google Scholar]
  2. Najafi B, Horn D, Marclay S, Crews RT, Wu S, Wrobel JS. Assessing postural control and postural control strategy in diabetes patients using innovative and wearable technology. J Diabetes Sci Technol. 2010;4:780–791. doi: 10.1177/193229681000400403. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Menant JC, Latt MD, Menz HB, Fung VS, Lord SR. Postural sway approaches center of mass stability limits in Parkinson's disease. Mov Disord. 2011;26:637–643. doi: 10.1002/mds.23547. [DOI] [PubMed] [Google Scholar]
  4. Bronte-Stewart HM, Minn AY, Rodrigues K, Buckley EL, Nashner LM. Postural instability in idiopathic Parkinson's disease: the role of medication and unilateral pallidotomy. Brain. 2002;125:2100–2114. doi: 10.1093/brain/awf207. [DOI] [PubMed] [Google Scholar]
  5. Maurer C, Mergner T, Xie J, Faist M, Pollak P, Lucking CH. Effect of chronic bilateral subthalamic nucleus (STN) stimulation on postural control in Parkinson's disease. Brain. 2003;126:1146–1163. doi: 10.1093/brain/awg100. [DOI] [PubMed] [Google Scholar]
  6. Rocchi L, Chiari L, Horak FB. Effects of deep brain stimulation and levodopa on postural sway in Parkinson's disease. J Neurol Neurosurg Psychiatry. 2002;73:267–274. doi: 10.1136/jnnp.73.3.267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Bloem BR, Beckley DJ, van Hilten BJ, Roos RA. Clinimetrics of postural instability in Parkinson's disease. J Neurol. 1998;245:669–673. doi: 10.1007/s004150050265. [DOI] [PubMed] [Google Scholar]
  8. Beuter A, Hernandez R, Rigal R, Modolo J, Blanchet PJ. Postural sway and effect of levodopa in early Parkinson's disease. Can J Neurol Sci. 2008;35:65–68. doi: 10.1017/s0317167100007575. [DOI] [PubMed] [Google Scholar]
  9. Chastan N, Debono B, Maltete D, Weber J. Discordance between measured postural instability and absence of clinical symptoms in Parkinson's disease patients in the early stages of the disease. Mov Disord. 2008;23:366–372. doi: 10.1002/mds.21840. [DOI] [PubMed] [Google Scholar]
  10. Mancini M, Horak FB, Zampieri C, Carlson-Kuhta P, Nutt JG, Chiari L. Trunk accelerometry reveals postural instability in untreated Parkinson's disease. Parkinsonism Relat Disord. 2011;17:557–562. doi: 10.1016/j.parkreldis.2011.05.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Dibble LE, Lange M. Predicting falls in individuals with Parkinson disease: a reconsideration of clinical balance measures. J Neurol Phys Ther. 2006;30:60–67. doi: 10.1097/01.npt.0000282569.70920.dc. [DOI] [PubMed] [Google Scholar]
  12. Ebersbach G, Baas H, Csoti I, Mungersdorf M, Deuschl G. Scales in Parkinson's disease. J Neurol. 2006;253(Suppl 4):IV32–IV35. doi: 10.1007/s00415-006-4008-0. [DOI] [PubMed] [Google Scholar]
  13. Haaxma CA, Bloem BR, Borm GF, Horstink MW. Comparison of a timed motor test battery to the Unified Parkinson's Disease Rating Scale-III in Parkinson's disease. Mov Disord. 2008;23:1707–1717. doi: 10.1002/mds.22197. [DOI] [PubMed] [Google Scholar]
  14. Horak FB, Wrisley DM, Frank J. The Balance Evaluation Systems Test (BESTest) to differentiate balance deficits. Phys Ther. 2009;89:484–498. doi: 10.2522/ptj.20080071. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Frenklach A, Louie S, Koop MM, Bronte-Stewart H. Excessive postural sway and the risk of falls at different stages of Parkinson's disease. Mov Disord. 2009;24:377–385. doi: 10.1002/mds.22358. [DOI] [PubMed] [Google Scholar]
  16. Swanenburg J, de Bruin ED, Uebelhart D, Mulder T. Falls prediction in elderly people: a 1-year prospective study. Gait Posture. 2010;31:317–321. doi: 10.1016/j.gaitpost.2009.11.013. [DOI] [PubMed] [Google Scholar]
  17. Keus SH, Nieuwboer A, Bloem BR, Borm GF, Munneke M. Clinimetric analyses of the Modified Parkinson Activity Scale. Parkinsonism Relat Disord. 2009;15:263–269. doi: 10.1016/j.parkreldis.2008.06.003. [DOI] [PubMed] [Google Scholar]
  18. Whitney SL, Roche JL, Marchetti GF, Lin CC, Steed DP, Furman GR, Musolino MC, Redfern MS. A comparison of accelerometry and center of pressure measures during computerized dynamic posturography: a measure of balance. Gait Posture. 2011;33:594–599. doi: 10.1016/j.gaitpost.2011.01.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Mayagoitia RE, Lotters JC, Veltink PH, Hermens H. Standing balance evaluation using a triaxial accelerometer. Gait Posture. 2002;16:55–59. doi: 10.1016/S0966-6362(01)00199-0. [DOI] [PubMed] [Google Scholar]
  20. Adlerton AK, Moritz U, Moe-Nilssen R. Forceplate and accelerometer measures for evaluating the effect of muscle fatigue on postural control during one-legged stance. Physiother Res Int. 2003;8:187–199. doi: 10.1002/pri.289. [DOI] [PubMed] [Google Scholar]
  21. Moe-Nilssen R, Helbostad JL. Trunk accelerometry as a measure of balance control during quiet standing. Gait Posture. 2002;16:60–68. doi: 10.1016/S0966-6362(01)00200-4. [DOI] [PubMed] [Google Scholar]
  22. Amiridis IG, Hatzitaki V, Arabatzi F. Age-induced modifications of static postural control in humans. Neurosci Lett. 2003;350:137–140. doi: 10.1016/S0304-3940(03)00878-4. [DOI] [PubMed] [Google Scholar]
  23. Maki BE, Holliday PJ, Fernie GR. Aging and postural control. A comparison of spontaneous- and induced-sway balance tests. J Am Geriatr Soc. 1990;38:1–9. doi: 10.1111/j.1532-5415.1990.tb01588.x. [DOI] [PubMed] [Google Scholar]
  24. Prieto TE, Myklebust JB, Hoffmann RG, Lovett EG, Myklebust BM. Measures of postural steadiness: differences between healthy young and elderly adults. IEEE Trans Biomed Eng. 1996;43:956–966. doi: 10.1109/10.532130. [DOI] [PubMed] [Google Scholar]
  25. Mancini M, Horak FB. The relevance of clinical balance assessment tools to differentiate balance deficits. Eur J Phys Rehabil Med. 2010;46:239–248. [PMC free article] [PubMed] [Google Scholar]
  26. Moe-Nilssen R, Nordin E, Lundin-Olsson L. Criteria for evaluation of measurement properties of clinical balance measures for use in fall prevention studies. J Eval Clin Pract. 2008;14:236–240. doi: 10.1111/j.1365-2753.2007.00839.x. [DOI] [PubMed] [Google Scholar]
  27. Moe-Nilssen R. Test-retest reliability of trunk accelerometry during standing and walking. Arch Phys Med Rehabil. 1998;79:1377–1385. doi: 10.1016/S0003-9993(98)90231-3. [DOI] [PubMed] [Google Scholar]
  28. Fahn S, Elton R. In: Recent Developments in Parkinson's Disease. Fahn S, Marsden CD, Calne D, Goldstein M, editor. Florham Park: Macmillan Healthcare Information; 1987. The UPDRS Development Committee. Unified Parkinson's disease rating scale; pp. 153–163. [Google Scholar]
  29. McIlroy WE, Maki BE. Preferred placement of the feet during quiet stance: development of a standardized foot placement for balance testing. Clin Biomech (Bristol, Avon) 1997;12:66–70. doi: 10.1016/S0268-0033(96)00040-X. [DOI] [PubMed] [Google Scholar]
  30. Moe-Nilssen R. A new method for evaluating motor control in gait under real-life environmental conditions. Part 1: The instrument. Clin Biomech (Bristol, Avon) 1998;13:320–327. doi: 10.1016/S0268-0033(98)00089-8. [DOI] [PubMed] [Google Scholar]
  31. Kerr G, Morrison S, Silburn P. Coupling between limb tremor and postural sway in Parkinson's disease. Mov Disord. 2008;23:386–394. doi: 10.1002/mds.21851. [DOI] [PubMed] [Google Scholar]
  32. McGraw KO, Wong SP. Forming Inferences About Some Intraclass Correlation Coefficients. Psychol Methods. 1996;1:30–46. [Google Scholar]
  33. Winter DA. Human balance and posture control during standing and walking. Gait Posture. 1995;3:193–214. doi: 10.1016/0966-6362(96)82849-9. [DOI] [Google Scholar]
  34. Hasan SS, Robin DW, Szurkus DC, Ashmead DH, Peterson SW, Schiavi R. Simultaneous measurement of body center of pressure and center of gravity during upright stance. Part II: Amplitude and frequency data. Gait Posture. 1996;4:11–20. doi: 10.1016/0966-6362(95)01031-9. [DOI] [Google Scholar]
  35. Peterka RJ. Sensorimotor integration in human postural control. J Neurophysiol. 2002;88:1097–1118. doi: 10.1152/jn.2002.88.3.1097. [DOI] [PubMed] [Google Scholar]
  36. Creath R, Kiemel T, Horak F, Peterka R, Jeka J. A unified view of quiet and perturbed stance: simultaneous co-existing excitable modes. Neurosci Lett. 2005;377:75–80. doi: 10.1016/j.neulet.2004.11.071. [DOI] [PubMed] [Google Scholar]
  37. Benvenuti F, Mecacci R, Gineprari I, Bandinelli S, Benvenuti E, Ferrucci L, Baroni A, Rabuffetti M, Hallett M, Dambrosia JM, Stanhope SJ. Kinematic characteristics of standing disequilibrium: reliability and validity of a posturographic protocol. Arch Phys Med Rehabil. 1999;80:278–287. doi: 10.1016/S0003-9993(99)90138-7. [DOI] [PubMed] [Google Scholar]
  38. Lafond D, Corriveau H, Hebert R, Prince F. Intrasession reliability of center of pressure measures of postural steadiness in healthy elderly people. Arch Phys Med Rehabil. 2004;85:896–901. doi: 10.1016/j.apmr.2003.08.089. [DOI] [PubMed] [Google Scholar]
  39. Bakker M, Esselink RA, Munneke M, Limousin-Dowsey P, Speelman HD, Bloem BR. Effects of stereotactic neurosurgery on postural instability and gait in Parkinson's disease. Mov Disord. 2004;19:1092–1099. doi: 10.1002/mds.20116. [DOI] [PubMed] [Google Scholar]
  40. St George RJ, Nutt JG, Burchiel KJ, Horak FB. A meta-regression of the long-term effects of deep brain stimulation on balance and gait in PD. Neurology. 2010;75:1292–1299. doi: 10.1212/WNL.0b013e3181f61329. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Spain RI, St George RJ, Salarian A, Mancini M, Wagner JM, Horak FB, Bourdette D. Body-worn motion sensors detect balance and gait deficits in people with multiple sclerosis who have normal walking speed. Gait Posture. 2012;35:573–578. doi: 10.1016/j.gaitpost.2011.11.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Maurer C, Peterka RJ. A new interpretation of spontaneous sway measures based on a simple model of human postural control. J Neurophysiol. 2005;93:189–200. doi: 10.1152/jn.00221.2004. [DOI] [PubMed] [Google Scholar]
  43. Rocchi L, Chiari L, Cappello A. Feature selection of stabilometric parameters based on principal component analysis. Med Biol Eng Comput. 2004;42:71–79. doi: 10.1007/BF02351013. [DOI] [PubMed] [Google Scholar]
  44. Rocchi L, Chiari L, Cappello A, Horak FB. Identification of distinct characteristics of postural sway in Parkinson's disease: a feature selection procedure based on principal component analysis. Neurosci Lett. 2006;394:140–145. doi: 10.1016/j.neulet.2005.10.020. [DOI] [PubMed] [Google Scholar]

Articles from Journal of NeuroEngineering and Rehabilitation are provided here courtesy of BMC

RESOURCES