Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 2012 Nov;50(11):3747–3750. doi: 10.1128/JCM.02037-12

Comparative Evaluation of Tigecycline Susceptibility Testing Methods for Expanded-Spectrum Cephalosporin- and Carbapenem-Resistant Gram-Negative Pathogens

Olympia Zarkotou a, Spyros Pournaras b,, George Altouvas a, Vassiliki Pitiriga c, Maria Tziraki a, Vassiliki Mamali a, Katerina Themeli-Digalaki a, Athanassios Tsakris c
PMCID: PMC3486197  PMID: 22933593

Abstract

We evaluated the Vitek2, Etest, and MIC Test Strip (MTS) methods of tigecycline susceptibility testing with 241 expanded-spectrum cephalosporin-resistant and/or carbapenem-resistant Enterobacteriaceae and Acinetobacter baumannii clinical isolates by using dry-form broth microdilution (BMD) as the reference method. The MIC50/90s were as follows: BMD, 1/4 μg/ml; Vitek2, 4/≥8 μg/ml; Etest, 2/4 μg/ml; MTS, 0.5/2 μg/ml. Vitek2 produced 9.1/21.2% major errors, Etest produced 0.4/0.8% major errors, and MTS produced no major errors but 0.4/3.3% very major errors (FDA/EUCAST breakpoints). Vitek2 tigecycline results require confirmation by BMD or Etest for multidrug-resistant pathogens.

TEXT

Carbapenem resistance has steadily increased in several regions, representing the most significant resistance issue among multidrug-resistant (MDR) Gram-negative pathogens (2, 17, 18). Tigecycline and colistin are among the few antimicrobials active and are commonly used for infections caused by carbapenem-resistant (CR) Enterobacteriaceae and Acinetobacter baumannii (2, 4, 16, 21).

The increasing clinical use of tigecycline necessitates rapid, simple, and accurate susceptibility testing methods. Several methods have been evaluated for routine tigecycline susceptibility testing (1, 12, 14, 19, 22) with broth microdilution (BMD) as the reference method. It has been previously noted that discrepancies may exist when different methods are used (3, 7, 14, 19, 22). It was also reported that disk diffusion and Etest have a poor correlation with BMD, especially for A. baumannii (3, 10, 12, 22). Furthermore, the accuracy of the Vitek2 automated system (bioMérieux, Marcy l' Etoile, France) has not been adequately evaluated for tigecycline (15).

In this study, we evaluated three routine tigecycline susceptibility methods, Vitek2, Etest (bioMérieux), and MIC Test Strip (MTS; Liofilchem SRL, Roseto degli Abruzzi, Italy) in comparison with BMD. These methods were applied to a large collection of CR Enterobacteriaceae and A. baumannii and expanded-spectrum cephalosporin-resistant (ESCR) Enterobacteriaceae isolates.

Bacterial isolates.

This study included 241 clinical isolates recovered during 2008 to 2011 from patients in five tertiary-care hospitals located in different regions of Greece, consisting of CR (K. pneumoniae carbapenemase [KPC]-producing K. pneumoniae, n = 73; VIM-producing K. pneumoniae, n = 39; KPC- and VIM-producing K. pneumoniae, n = 13; OXA-58-producing A. baumannii, n = 56) and ESCR isolates (extended-spectrum β-lactamase [ESBL]-producing Escherichia coli, n = 20; ESBL-producing K. pneumoniae, n = 20; Enterobacter spp., n = 20). Identification was performed with Vitek2. Phenotypic carbapenemase detection was performed as described previously (23). Common broad-spectrum β-lactamase genes (for KPC, OXA-48, OXA-58, metallo-β-lactamases, and ESBLs) were sought by PCR (20).

Susceptibility testing methods.

The FDA-cleared commercial dry-form microtiter panels and cation-adjusted Mueller-Hinton broth with N-Tris(hydroxymethyl)methyl-2-aminoethanesulfonic acid (Sensititre JustOne for tigecycline; TREK Diagnostic Systems, Cleveland, OH) was carried out according to CLSI procedures (6); panels were inoculated manually and read optically, and MICs ranged from 0.06 to 64 μg/ml. The Vitek2 AST-EXN8 susceptibility card containing tigecycline at concentrations of 0.75, 2, and 4 μg/ml was used according to the manufacturer's recommendations. Etest (0.016 to 256 μg/ml) and MTS (0.016 to 256 μg/ml) for tigecycline were performed with Mueller-Hinton II agar (Mast Group Ltd., Bootle, Merseyside, United Kingdom) according to the manufacturer's instructions. Etest and MTS MICs were rounded up to the next 2-fold BMD dilution, as is standard for MIC gradient tests.

All methods were performed simultaneously with a single inoculum of each strain. E. coli ATCC 25922 was used as the quality control strain in susceptibility assays.

Definitions and data analysis.

There is discordance between the interpretative tigecycline MIC susceptibility breakpoints of Enterobacteriaceae issued by the European Committee on Antimicrobial Susceptibility Testing (EUCAST) (susceptible MIC, ≤1 μg/ml; resistant MIC, >2 μg/ml) (8) and the U.S. Food and Drug Administration (FDA) (susceptible MIC, ≤2 μg/ml; resistant MIC, ≥8 μg/ml) (24). For that reason, interpretation of susceptibility results was performed using both the EUCAST and FDA breakpoints. The breakpoints listed for Enterobacteriaceae were also applied to A. baumannii.

Data were analyzed by comparing the results from each method to those produced by the reference BMD method. MIC50s and MIC90s were calculated. Categorical agreement (CA) was defined as the percentage of isolates classified in the same susceptibility category by BMD and the method under evaluation. Category discrepancies were classified as follows: (i) very major errors (VME), cases where BMD indicated resistance and the comparative method indicated susceptibility; (ii) major errors (ME), an isolate categorized as susceptible by BMD and resistant by the comparative method; (iii) minor errors (mE), one interpretation category difference between BMD and the comparative method (5). Essential agreement (EA) was considered the percentage of MICs within ±1 doubling dilution of the MIC determined by BMD (5).

Acceptable performance was evaluated according to the criteria established by the International Organization for Standardization as follows: ≥90% for EA or CA, ≤3% for VME or ME, and ≤7% for ME plus mE (11).

Susceptibility to tigecycline, EA and CA.

The susceptibilities to tigecycline using FDA/EUCAST breakpoints and the MIC50 and MIC90 for the study isolates determined by each method are presented in Table 1.

Table 1.

Tigecycline susceptibilities of the study isolates and MIC50s and MIC90s determined by BMD, Vitek2, Etest, and MTS

Test method and isolate group No. (%) of isolates
MIC (μg/ml)
Susceptible
Intermediate
Resistant
FDA EUCAST FDA EUCAST FDA EUCAST 50% 90%
BMD
    All isolates 201 (83.4) 150 (62.2) 35 (14.5) 51 (21.2) 5 (2.1) 40 (16.6) 1 4
    CR K. pneumoniae 105 (84.0) 80 (64.0) 18 (14.4) 25 (20.0) 2 (1.6) 20 (16.0) 1 4
    CR A. baumannii 42 (75.0) 25 (44.6) 12 (21.4) 17 (30.4) 2 (3.6) 14 (25.0) 2 4
    ESCR Enterobacteriaceae 54 (90.0) 45 (75.0) 5 (8.3) 9 (15.0) 1 (1.7) 6 (10.0) 0.5 2
Vitek2
    All isolates 103 (42.7) 53 (22.0) 84 (34.9) 50 (20.7) 54 (22.4) 138 (57.3) 4 ≥ 8
    CR K. pneumoniae 50 (40.0) 12 (9.6) 50 (40.0) 38 (30.4) 25 (20.0) 75 (60.0) 4 ≥ 8
    CR A. baumannii 10 (17.9) 3 (5.4) 27 (48.2) 7 (12.5) 19 (33.9) 46 (82.1) 4 ≥ 8
    ESCR Enterobacteriaceae 43 (71.7) 38 (63.3) 7 (11.7) 5 (8.3) 10 (16.7) 17 (28.3) 1 ≥ 8
Etest
    All isolates 198 (82.2) 108 (44.8) 33 (13.7) 89 (36.9) 10 (4.1) 44 (18.3) 2 4
    CR K. pneumoniae 105 (84.0) 48 (38.4) 17 (13.6) 56 (44.8) 3 (2.4) 21 (16.8) 2 4
    CR A. baumannii 39 (69.6) 16 (28.6) 11 (19.6) 23 (41.1) 6 (10.7) 17 (30.4) 2 4
    ESCR Enterobacteriaceae 54 (90.0) 44 (73.3) 5 (8.3) 10 (16.7) 1 (1.7) 6 (10.0) 0.5 2
MTS
    All isolates 229 (95.0) 190 (78.8) 9 (3.7) 39 (16.2) 3 (1.2) 12 (5.0) 0.5 2
    CR K. pneumoniae 124 (99.2) 106 (84.8) 1 (0.8) 18 (14.4) 0 (0) 1 (0.8) 1 2
    CR A. baumannii 47 (83.9) 32 (57.1) 6 (10.7) 15 (26.8) 3 (5.4) 9 (16.1) 1 4
    ESCR Enterobacteriaceae 58 (96.7) 52 (86.7) 2 (3.3) 6 (10.0) 0 (0) 2 (3.3) 0.25 2

By BMD, 201 isolates (83.4%) were tigecycline susceptible using FDA breakpoints while 150 (62.2%) were susceptible using EUCAST breakpoints. CR A. baumannii exhibited the lowest susceptibility rates (75/44.6% using FDA/EUCAST breakpoints).

As shown in Table 2, Etest produced susceptibility results similar to those obtained with BMD and CA was high (>80%) for all pathogens using FDA breakpoints. With EUCAST breakpoints, susceptibility rates were lower for most species and CA was 71.8%. Discordant susceptibility rates with serious interpretative errors and rather low CA (48.1/39.4% with FDA/EUCAST breakpoints) were observed for Vitek2. The in vitro activity of tigecycline determined by Vitek2 was limited; 57.3/78% of the isolates were classified as either intermediate or resistant. The most pronounced differences in susceptibility rates were noted for CR A. baumannii and K. pneumoniae. In contrast to Vitek2, a slight shift toward susceptibility was noted for MTS. However, relatively high rates of CA were obtained overall (86.3/68.5%).

Table 2.

EA, CA, and types of errors produced when testing tigecycline susceptibility by Vitek2, Etest, and MTS compared to BMD

Test method and isolate group No. (%) of isolates with:
EA CA
VME
ME
mE
FDA EUCAST FDA EUCAST FDA EUCAST FDA EUCAST
Vitek2
    All isolates 148 (61.4) 116 (48.1) 95 (39.4) 0 (0) 0 (0) 22 (9.1) 51 (21.2) 103 (42.7) 95 (39.4)
    CR K. pneumoniae 65 (52.0) 57 (45.6) 34 (27.2) 0 (0) 0 (0) 10 (8.0) 32 (25.6) 58 (46.4) 59 (47.2)
    CR A. baumannii 31 (55.4) 14 (25.0) 18 (32.1) 0 (0) 0 (0) 7 (12.5) 16 (28.6) 35 (62.5) 22 (39.3)
    ESCR Enterobacteriaceae 52 (86.7) 45 (75.0) 43 (71.7) 0 (0) 0 (0) 5 (8.3) 3 (5.0) 10 (16.7) 14 (23.3)
Etest
    All isolates 229 (95.0) 220 (91.3) 173 (71.8) 0 (0) 0 (0) 1 (0.4) 2 (0.8) 20 (8.3) 66 (27.4)
    CR K. pneumoniae 121 (96.8) 117 (93.6) 81 (64.8) 0 (0) 0 (0) 1 (0.8) 1 (0.8) 7 (5.6) 43 (34.4)
    CR A. baumannii 52 (92.9) 47 (83.9) 39 (69.6) 0 (0) 0 (0) 0 (0) 1 (1.8) 9 (16.1) 16 (28.6)
    ESCR Enterobacteriaceae 56 (93.3) 56 (93.3) 53 (88.3) 0 (0) 0 (0) 0 (0) 0 (0) 4 (6.7) 7 (11.7)
MTS
    All isolates 187 (77.6) 208 (86.3) 165 (68.5) 1 (0.4) 8 (3.3) 0 (0) 0 (0) 32 (13.3) 68 (28.2)
    CR K. pneumoniae 103 (82.4) 105 (84.0) 80 (64.0) 1 (0.8) 6 (4.8) 0 (0) 0 (0) 19 (15.2) 39 (31.2)
    CR A. baumannii 50 (89.3) 48 (85.7) 35 (62.5) 0 (0) 1 (1.8) 0 (0) 0 (0) 8 (14.3) 20 (35.7)
    ESCR Enterobacteriaceae 34 (56.7) 55 (91.7) 50 (83.3) 0 (0) 1 (1.7) 0 (0) 0 (0) 5 (8.3) 9 (15.0)

BMD, Etest, and MTS resulted in MIC50s of ≤2 μg/ml for all of the species tested, while by Vitek2 the MIC50 was 4 μg/ml for CR bacteria. The Etest MIC90 was identical to the BMD MIC90 for all bacterial groups. MIC90s obtained with Vitek2 (≥8 μg/ml) were inconsistent, compared with those obtained with BMD, while MTS produced lower MIC90 values for CR K. pneumoniae (Table 1). MICs obtained by MTS were 1, 2, and 3 log2 dilutions lower for 50.2%, 17.8%, and 2.5% of the isolates, respectively, than those obtained by BMD. In contrast, compared with BMD, Vitek2 resulted in MICs being 1, 2, and 3 log2 dilutions higher for 41.1%, 36.5%, and 1.7% of the isolates, respectively. By Etest, 47.7% of the isolates exhibited MICs identical to those obtained with BMD and 47.4% of the remaining isolates displayed MICs within ±1 log2 dilution (Table 3).

Table 3.

Differences in log2 dilutions of MICs obtained by Vitek2, Etest, and MTS compared to BMD

Test method No. (%) of isolates showing a MIC difference (in log2 dilutions) of:
>−3 −3 −2 −1 0 1 2 3
Vitek2 4 (1.7) 45 (18.7) 99 (41.1) 89 (36.5) 4 (1.7)
Etest 1 (0.4) 4 (1.7) 37 (15.4) 115 (47.7) 77 (32.0) 7 (2.9)
MTS 5 (2.1) 6 (2.5) 43 (17.8) 121 (50.2) 56 (23.2) 10 (4.1)

EA was highest for Etest (95.0% overall), being ≥90% for all subsets of isolates, exceeding the acceptable performance rate for antimicrobial susceptibility test (AST) methods. MTS resulted in a relatively high rate of EA (77.6%), displaying better performance with CR isolates (EA, 82.4/89.3% for K. pneumoniae/A. baumannii. On the contrary, Vitek2 generated an overall EA rate of 61.4%, with lower EA rates for CR isolates (52.0 and 55.4%). The EA and CA of MIC results between testing methods compared to the reference BMD are presented in Table 2.

Error classification.

Error rates were significantly higher by EUCAST. Etest exhibited the lowest overall error rates (ME, 0.4/0.8%; mE, 8.3/27.4%). No MEs were detected for MTS; however, it yielded VMEs (0.4/3.3%) and relatively high mE rates (13.3/28.2%). Vitek2 produced high rates of both MEs (9.1/21.2%) and mEs (42.7/39.4%). With FDA breakpoints, Etest approaches the criteria for acceptable AST performance, while no method displayed acceptable performance with EUCAST breakpoints (Table 2).

The shortage of available treatment options for infections by CR Gram-negative bacteria highlights the importance of accurate tigecycline susceptibility results (9). The interchangeability of tigecycline susceptibility results of CR Enterobacteriaceae and A. baumannii is not well defined (13). Previous studies have reported discrepant Etest and BMD results with tigecycline, mainly for A. baumannii (3, 7, 10, 12, 14, 19, 22). Scarce data are available on the accuracy of VItek2 with MDR pathogens (15). In the present study, we assessed the performance of three routine methods of tigecycline susceptibility testing against ESCR/CR Enterobacteriaceae and A. baumannii clinical isolates. The results were interpreted by using FDA and EUCAST recommendations.

Of the methods evaluated, Etest showed the best correlation with BMD, exhibiting the lowest error rates and the highest EA and CA. Etest susceptibility results obtained by using FDA criteria meet the AST acceptable-performance criteria for CR K. pneumoniae and ESCR Enterobacteriaceae, and this could be considered a reliable method for tigecycline testing, as also shown previously (1, 19). A relatively high mE rate of Etest, with slightly decreased susceptibility rates compared to those of BMD was observed for CR A. baumannii. In these few cases, Etest generated 1 to 2 log2 dilution higher MICs. Similar but more pronounced findings have been published previously including higher MICs by Etest for A. baumannii (3, 14, 19) and Enterobacter spp. (7). Concerns have been expressed regarding the suitability of Etest for tigecycline susceptibility testing against A. baumannii (19). With EUCAST breakpoints, a significant increase in mEs was observed, more pronounced for CR K. pneumoniae. The elevated mE rates could be explained by the MIC distribution of the organisms' population being close to the breakpoint concentration. The analysis of mEs revealed no trend toward susceptibility or resistance, and thus, error rates were considered acceptable.

We also evaluated MTS for tigecycline susceptibility testing. The level of CA, though lower, was comparable to that of Etest. MTS produced ≥1 log2 dilution lower MICs than BMD for 72.6% of our isolates, resulting in a shift toward susceptibility. The generally lower MTS MICs were more pronounced for isolates with low MICs. When the results were evaluated upon the performance criteria for susceptibility testing, MTS produced unacceptable results with marginally elevated VMEs with EUCAST breakpoints and increased mE rates with both FDA and EUCAST breakpoints.

Vitek2 is widely used in clinical laboratories. Our data highlight its important limitations in tigecycline susceptibility testing. It produced results with low EA and CA rates; ME and mE rates were severalfold higher than the acceptable performance with both FDA and EUCAST breakpoints and generally produced higher MICs, resulting in false resistance findings. These findings were more pronounced for CR bacteria (ME plus mE, >50%), against which, however, tigecycline therapy is more commonly necessary. The only published study evaluating the performance of Vitek2 in the determination of the tigecycline susceptibility of KPC-producing K. pneumoniae strains reported a higher EA level and much lower mE rates than our results (15). These results could be explained by the lower MIC90s for the isolates included in that study.

Major discrepancies were noted for CA and error rates between EUCAST and FDA data analyses, reflecting the high number of isolates for which the MICs were within ±1 log2 dilution of the categorical breakpoints set by each organization.

Since tigecycline is commonly used against infections with CR pathogens, reliable susceptibility results are important for therapeutic decisions. Our study underlines the shortcomings of automated and manual susceptibility testing methods, which may falsely restrict the available treatment options or lead to inappropriate antimicrobial therapy. Clinical laboratories should be aware of the interpretive problems. Confirmation of susceptibility results by a reference method is therefore recommended, particularly when tigecycline administration is deemed necessary.

(This work was presented in part in the 22nd European Congress of Clinical Microbiology and Infectious Diseases, London, United Kingdom, April 2012.)

ACKNOWLEDGMENTS

This work was funded by a research grant from Pfizer (WS963566). The funding source was not involved in the study design, execution, result interpretation, or publication policy.

Footnotes

Published ahead of print 29 August 2012

REFERENCES

  • 1.Bolmström A, et al. 2007. Validation and reproducibility assessment of tigecycline MIC determinations by Etest. J. Clin. Microbiol. 45:2474–2479 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Carmeli Y, et al. 2010. Controlling the spread of carbapenemase-producing Gram-negatives: therapeutic approach and infection control. Clin. Microbiol. Infect. 16:102–111 [DOI] [PubMed] [Google Scholar]
  • 3.Casal M, et al. 2009. Influence of testing methodology on the tigecycline activity profile against presumably tigecycline-non-susceptible Acinetobacter spp. J. Antimicrob. Chemother. 64:69–72 [DOI] [PubMed] [Google Scholar]
  • 4.Castanheira MH, Sader S, Deshpande LM, Fritsche TR, Jones RN. 2008. Antimicrobial activities of tigecycline and other broad-spectrum antimicrobials tested against serine carbapenemase- and metallo-beta-lactamase-producing Enterobacteriaceae: report from the SENTRY Antimicrobial Surveillance Program. Antimicrob. Agents Chemother. 52:570–573 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Clinical and Laboratory Standards Institute 2008. Development of in vitro susceptibility testing criteria and quality control parameters, 3d edition M23-A3 Clinical and Laboratory Standards Institute, Wayne, PA [Google Scholar]
  • 6.Clinical and Laboratory Standards Institute 2008. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; approved standard, 8th edition M08-A8 Clinical and Laboratory Standards Institute, Wayne, PA [Google Scholar]
  • 7.Cohen Stuart J, et al. 2010. Evaluation of Etest to determine tigecycline MICs for Enterobacter species. Antimicrob. Agents Chemother. 54:2746–2747 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.European Committee on Antimicrobial Susceptibility Testing (EUCAST) 2012. Breakpoint tables for interpretation of MICs and zone diameters. Version 2.0, valid from 2012-01-01. http://www.eucast.org/clinical_breakpoints/ [Google Scholar]
  • 9.Falagas ME, Karageorgopoulos DE, Nordmann P. 2011. Therapeutic options for infections with Enterobacteriaceae producing carbapenem-hydrolyzing enzymes. Future Microbiol. 6:653–666 [DOI] [PubMed] [Google Scholar]
  • 10.Hope R, et al. 2010. Tigecycline activity: low resistance rates but problematic disc breakpoints revealed by a multicentre sentinel survey in the UK. J. Antimicrob. Chemother. 65:2602–2609 [DOI] [PubMed] [Google Scholar]
  • 11.International Organization for Standardization 2007. ISO 20776-2:2007(E) Clinical laboratory testing and in vitro diagnostic test systems. Susceptibility testing of infectious agents and evaluation of performance of antimicrobial susceptibility test devices. Part 2: evaluation of performance of antimicrobial susceptibility test devices. International Organization for Standardization, Geneva, Switzerland [Google Scholar]
  • 12.Jones RN, et al. 2007. Multicenter studies of tigecycline disk diffusion susceptibility results for Acinetobacter spp. J. Clin. Microbiol. 45:227–230 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Kelesidis T, Karageorgopoulos DE, Kelesidis I, Falagas ME. 2008. Tigecycline for the treatment of multidrug-resistant Enterobacteriaceae: a systematic review of the evidence from microbiological and clinical studies. J. Antimicrob. Chemother. 62:895–904 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kulah C, et al. 2009. Unexpected tigecycline resistance among Acinetobacter baumannii isolates: high minor error rate by Etest. J. Chemother. 21:390–395 [DOI] [PubMed] [Google Scholar]
  • 15.Lat A, et al. 2011. Comparison of polymyxin B, tigecycline, cefepime, and meropenem MICs for KPC-producing Klebsiella pneumoniae by broth microdilution, Vitek2, and Etest. J. Clin. Microbiol. 49:1795–1798 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Mendes RE, Farrell DJ, Sader HS, Jones RN. 2010. Comprehensive assessment of tigecycline activity tested against a worldwide collection of Acinetobacter spp. (2005-2009). Diagn. Microbiol. Infect. Dis. 68:307–311 [DOI] [PubMed] [Google Scholar]
  • 17.Miyakis S, Pefanis A, Tsakris A. 2011. The challenges of antimicrobial drug resistance in Greece. Clin. Infect. Dis. 53:177–184 [DOI] [PubMed] [Google Scholar]
  • 18.Nordmann P, Naas T, Poirel L. 2011. Global spread of carbapenemase-producing Enterobacteriaceae. Emerg. Infect. Dis. 17:1791–1798 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Pillar CM, Draghi DC, Dowzicky MJ, Sahm DF. 2008. In vitro activity of tigecycline against gram-positive and gram-negative pathogens as evaluated by broth microdilution and Etest. J. Clin. Microbiol. 46:2862–2867 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Pournaras S, et al. 2009. Clonal spread of KPC-2 carbapenemase-producing Klebsiella pneumoniae strains in Greece. J. Antimicrob. Chemother. 64:348–352 [DOI] [PubMed] [Google Scholar]
  • 21.Sader HS, Farrell DJ, Jones RN. 2011. Tigecycline activity tested against multidrug-resistant Enterobacteriaceae and Acinetobacter spp. isolated in US medical centers (2005-2009). Diagn. Microbiol. Infect. Dis. 69:223–227 [DOI] [PubMed] [Google Scholar]
  • 22.Torrico M, et al. 2010. Influence of media and testing methodology on susceptibility to tigecycline of Enterobacteriaceae with reported high tigecycline MIC. J. Clin. Microbiol. 48:2243–2246 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Tsakris A, et al. 2010. A simple phenotypic method for the differentiation of metallo-beta-lactamases and class A KPC carbapenemases in Enterobacteriaceae clinical isolates. J. Antimicrob. Chemother. 65:1664–1671 [DOI] [PubMed] [Google Scholar]
  • 24.Wyeth Pharmaceuticals Inc. 2005. Tygacil package insert. Wyeth Pharmaceuticals Inc., Philadelphia, PA [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)