Summary
Low-dose images obtained by electron cryo-microscopy (cryo-EM) are often affected by blurring caused by sample motion during electron beam exposure, degrading signal especially at high resolution. We show here that we can align frames of movies, recorded with a direct electron detector during beam exposure of rotavirus double-layered particles, thereby greatly reducing image blurring caused by beam-induced motion and sample stage instabilities. This procedure increases the efficiency of cryo-EM imaging and enhances the resolution obtained in three-dimensional reconstructions of the particle. Employing movies in this way is generally applicable to all cryo-EM samples and should also improve the performance of mid-range electron microscopes that may have limited mechanical stability and beam coherence.
Introduction
Electron cryo-microscopy (cryo-EM) has developed over the last four decades as a technique for visualizing the three-dimensional (3D) structures of macromolecules and their assemblies. Two-dimensional (2D) crystals provided resolutions of 4 Å or better in the 1990s (e.g. Henderson et al., 1990; Kuhlbrandt et al., 1994; Nogales et al., 1998); helical assemblies yielded comparable resolutions about a decade later (e.g. Sachse et al., 2007; Unwin, 2005; Yonekura et al., 2003); most recently, large icosahedral viruses have done the same (see Grigorieff and Harrison, 2011 for a review). The steady advances in the technique have included better sample preparation methods (e.g. microfabricated grids, Ermantraut et al., 1998; Quispe et al., 2007), better electron microscopes (e.g. field emission guns, Henderson et al., 1990); spot-scan, (Bullough and Henderson, 1987), improved image processing algorithms (e.g., MRC image processing package, Crowther et al., 1996; EMAN, Tang et al., 2007; FREALIGN, Grigorieff, 2007; CTFFIND3 & CTFTILT, Mindell and Grigorieff, 2003), and more powerful computing systems. Further technological improvements are currently underway, including phase plates to enhance low-resolution contrast (Cambie et al., 2007; Danev and Nagayama, 2001; Majorovits et al., 2007; Muller et al., 2010), direct electron detectors that increase the signal-to-noise ratio (SNR) in the recorded images (McMullan et al., 2009; Milazzo et al., 2005), and new algorithms to accommodate conformational and compositional variability in the imaged sample (Penczek et al., 2011; Scheres et al., 2007).
Among the remaining problems currently limiting the attainable resolution is the motion of the sample that occurs while it is exposed to the high-energy electron beam. In a recent study, we showed that the beam causes particles suspended in a thin layer of vitreous ice inside holes in a perforated carbon support film to shift by up to a few nm and rotate by up to a few degrees, causing significant blurring in many of the images (Brilot et al., 2012). As the motion occurs in the ice layer, neighboring particles move in a similar fashion. The blurring due to beam-induced motion (BIM) is likely one of the main reasons why the number of images required to achieve 4 Å resolution or better is significantly larger than predicted by calculations based on the SNR of idealized low-dose images (Glaeser, 1999; Henderson, 1995). We earlier characterized BIM using exposure series and movies obtained using a direct electron detector (Brilot et al., 2012). As we noted, the movies can be used to reduce image blurring caused by BIM and stage instabilities by averaging the movie frames after aligning the individual frames to bring them into register with each other. In the present work, we demonstrate this possibility by comparing reconstructions calculated from movies without frame alignment and reconstructions calculated from movies with frames that have been aligned to correct for sample motion. We show that the movie-frame alignment substantially increases the overall resolution that can be obtained from a relatively small cryo-EM data set.
Results
Movie processing
We recorded 501 movies of rotavirus double-layered particles (DLPs), the same sample that was used in our previous work (Brilot et al., 2012). These particles previously yielded a ~4 Å reconstruction using data recorded on photographic film (Zhang et al., 2008) (referred to here as “the 2008 data set”). To track particle motion during beam exposure, we recorded movies consisting of 16 frames during continuous electron exposure (see Experimental Procedures). To track sample motion, we followed our previous procedure (Brilot et al., 2012) and generated averages of four movie frames at a time, corresponding to an electron dose of about 8 electrons/Å2 in each average, providing sufficient signal to align images of individual particles against a reference reconstruction of the DLP (Zhang et al., 2008). The tracking showed temporal dependence of the motion similar to what we saw previously, with faster initial motion that slowed towards the end of the exposure (Brilot et al., 2012). We used the particle coordinates and Euler angles measured in the four-frame averages as time points to describe particle trajectories and derived particle parameters for individual movie frames from fits to a quadratic function for each particle trajectory.
Correcting beam-induced motion and sample drift
Fig. 1 shows an area of sample that experienced a shift of about 10 Å during the exposure. The average calculated without alignment from all 16 frames is shown in Fig. 1A. There is evidence of significant blurring and loss of detail in the Thon ring pattern (Thon, 1966) calculated for this frame average (Fig. 1C). After shift alignment of the frames to compensate for sample drift and BIM (Fig. 1B), the Thon ring pattern (Fig. 1D) appears more uniform, indicating near-isotropic resolution.
We obtained a total of 1,915 particles from the movies and selected the best ~40% (807 particles) for 3D reconstruction, based on their correlation coefficients with a reference reconstruction. We chose the percentage of included particles to correspond to the percentage used in the final reconstructions from the 2008 data set, which we used as a reference to evaluate the quality of the movie data (see below). Fig. 2 shows a histogram of the measured rotations and shifts that occurred between averages of the first and second group of four movie frames (frames 1 – 4 and 5 – 8, respectively), and between averages of the first and last group of four movie frames (frames 1 – 4 and 13 – 16, respectively). The motions occurring between the first and second group of movie frames is directly comparable with our earlier experiments in which we analyzed the motions between single 8 electrons/Å2 exposures (Brilot et al., 2012). Each histogram is fitted with a Rayleigh distribution to obtain the median value λ (Brilot et al., 2012). Separate histograms are shown for the full data set and the 807 particles selected for 3D reconstruction, showing that there is no significant difference in the distributions of observed rotations and shifts.
Between averages of the first and second group of four movie frames, measured rotations range from a fraction of a degree to a few degrees with a distribution maximum at about 0.2° (Fig. 2A). This is substantially smaller than the previously observed 0.44° (Brilot et al., 2012) for a similar dose rate (80 electrons/Å2/second). In our previous analysis, we estimated the angular alignment accuracy to be about 0.2°, close to the majority of rotations plotted in Fig. 2A. The distribution of rotations is therefore broadened by an alignment error and actual rotations are likely smaller than suggested by the histogram in Fig. 2A. Shifts range from a fraction of an Ångstrom to over 20 Å with a distribution maximum at about 2 Å (Fig. 2B). Larger motions were measured between averages of the first and last group of four frames, with distribution maxima at about 0.3° (Fig. 2C) and about 5 Å for rotations and shifts, respectively. The broadening of the histogram of rotations may in part be the result of somewhat increased alignment errors due to degraded signal in the final four-frame average, which shows increased radiation damage compared with the earlier frame averages. Fig. 1A shows an example of measured particle rotations within a movie. The measured angles are relatively small and their directions are uncorrelated, consistent with these measurements being at the limit of our alignment accuracy.
The observed shifts are likely a combination of beam-induced motion and stage drift, which are difficult to distinguish in our experiments. However, as part of the Leginon recording cycle, drift rates were measured each time the stage was moved to a different area of the grid (translations larger than 10 μm). Following 96% of these translational movements, an average drift rate of 1.6 Å/second (standard deviation = 0.7 Å/second) was achieved after 20s. A much smaller translation (less than 2 μm) was subsequently used to move to the area where movies were recorded. Although no further monitoring of the stage drift was performed after these smaller translations it is reasonable to assume that the average drift rate did not change significantly. Assuming a drift rate of 1.6 Å/second, the sample moved only by about 1 Å during the course of the movies (0.6 seconds). The measured shifts plotted in Fig. 2D can therefore mainly be attributed to BIM. Nevertheless, it should be noted that our motion tracking procedure does not depend on the source of the measured shifts, and blurring due to both stage drift and BIM will be corrected.
Improved 3D reconstructions
A particle moving by about 5 Å will have a standard deviation from its average position of about 1.5 Å, producing a temperature factor (B-factor) of about 200 Å2 (Jensen, 2001). Without motion tracking, the majority of particles experience shifts larger than 5 Å during the entire length of the movies (Fig. 2D) and will contribute with an even higher effective temperature factor. Rotations of 0.3° will add an average B-factor of 160 Å2 (Jensen, 2001), giving a total B-factor of about 360 Å2 due to motion. Using our motion tracking procedure to reduce blurring in the particle images, we calculated icosahedrally symmetrized reconstructions of DLP. As expected, the B-factors of these reconstructions were significantly reduced, by about 160 Å2, relative to the reconstructions that were not motion-compensated (Fig. 3A). Fig. 3B shows FSC curves for the DLP reconstruction, and for the rotavirus VP6 coat protein after additional 13-fold non-icosahedral averaging (Zhang et al., 2008), calculated without motion tracking, with motion tracking that accounts only for particle shifts, and with motion tracking that accounts for shifts and rotations. The curves indicate significant improvements in the resolution in both DLP and VP6 densities when compensating for shifts, and further minor improvements when including rotations. The resolution of the DLP reconstruction improved from 7.5 Å to 6.4 Å and 6.2 Å, respectively, while that of the averaged VP6 reconstruction improved from 4.9 Å to 4.4 Å and 4.4 Å, respectively (at FSC = 0.143, Rosenthal and Henderson, 2003). The improvement agrees with the observed B-factor reduction of 160 Å2, which is equivalent to an 8-fold signal amplification at 4.4 Å. The 8-fold signal amplification translates into a 60-fold reduction of the number of particles required to achieve the same signal without amplification.
In Fig. 4, we compare details that can be seen in the 13-fold averaged VP6 density calculated with and without motion tracking, to similar details seen in the 2008 reconstruction (Zhang et al., 2008). We fit densities with an atomic model of VP6 obtained by X-ray crystallography (PDB ID: 3KZ4, McClain et al., 2010) to help the interpretation. All maps show density corresponding to amino-acid side chains. Compared to the density calculated without motion tracking (Fig. 4A), the reconstruction that takes sample shifts into account improves substantially, and the density for many of the larger side chains is significantly stronger (Fig. 4B). Taking into account both shifts and rotations leads to a map with further minor improvements (Fig. 4C), as expected (see above). The improved detail in the densities shown in Fig. 4B and 4C are also indicated by the increase in the Fourier shell correlation (FSC) calculated for the reconstructions (Fig. 3B), showing a resolution increase from 4.9 Å (without motion tracking) to 4.4 Å (with motion tracking). Comparison of the motion-compensated map with the map obtained in 2008 (Fig. 4D), which was calculated using 10 times as many particle images to yield a resolution of 4.1 Å (FSC = 0.143), suggests that the two maps are of similar quality.
Discussion
One of the main problems limiting the resolution attainable in cryo-EM reconstructions of ordered arrays, such as 2D crystals and helical assemblies, and of single particles, such as viruses and ribosomes, is beam-induced motion (BIM), which causes blurring in the recorded images. We have shown here that blurring due to beam-induced motion and stage drift can be reduced by recording movies of the sample as it is exposed to the electron beam, followed by alignment of individual movie frames (Fig. 1). The reduced blurring leads to a substantial improvement in resolution (Fig. 3) and map quality. To achieve the resolution of the final 13-fold averaged density in Fig. 4B (4.4 Å) required a total of 807 × 60 × 13 = 630,000 VP6 subunits, while it took 6.6 million subunits to obtain a resolution of 4.1 Å with the 2008 data set, and both maps are of similar quality. The 10-fold reduction of required data arises partly due to the compensation for sample motion, but the observed decrease in B-factor by 160 Å2 would indicate a 60-fold reduction in required data. The remaining differences may arise due to several other factors that differ between the data sets, including the accelerating voltage (300 vs. 200 kV), detector (film vs. electronic detector), pixel size (1.23 Å vs. 1.42 Å), grids (a combination of lacey carbon and C-flat™ vs. solely C-flat™), beam diameter and stage drift rates; all of these factors preclude a fully quantitative comparison.
While motion-tracking significantly enhances the data, the need for over half a million subunits to obtain a ~4 Å map still exceeds by at least an order of magnitude the number of particles that would be required based on calculations of the SNR of idealized low-dose images (Glaeser, 1999; Henderson, 1995). Several factors may contribute to this difference. The data in our study were collected on a non-backthinned CMOS direct electron detector that has a detective quantum efficiency (DQE) inferior to that of photographic film at the resolution evaluated here (about 70% Nyquist frequency, McMullan et al., 2009). We thus expect the map quality to improve further when data is collected at 300 kV using backthinned CMOS chips that will be available with new direct electron detectors (McMullan et al., 2009). Furthermore, the motion tracking we have implemented may not completely compensate for the particle shifts, due to positional inaccuracy. The tracking of particle motion in our current work was done for individual particles, which provided sufficiently strong contrast due to the high molecular mass (70 MDa) of the particle. We showed in our earlier work (Brilot et al., 2012) that neighboring particles move together because the motion appears to be caused by the embedding ice layer moving as a whole. The positional tracking can therefore be made more accurate by imposing correlations between particles. This will be especially important when tracking particles with smaller mass that do not produce strong contrast. An interesting further improvement may be possible when imaging particles with lower molecular mass. Due to their weak low-resolution contrast, these particles are more prone to misalignment than particles with a high molecular mass (Henderson et al., 2011). The low-resolution contrast can be enhanced by using a higher dose, at the expense of the high-resolution signal, which fades more rapidly with dose than the low-resolution signal (Baker et al., 2010; Hayward and Glaeser, 1979). Using movies, the effective dose that contributes to a reconstruction can be adjusted by limiting the number of included frames. Thus a dose-dependent frame filtering could be used to amplify both the low- and high-resolution contrast. The traditional dose limit in low-dose cryo-EM must therefore be re-evaluated to find the optimum when using this new approach (Baker et al., 2010; Hayward and Glaeser, 1979).
An important result from our data is the reduction in the amount of observed particle rotations compared with our previous study (Brilot et al., 2012). There are two differences in the way the sample was illuminated by the beam. In the previous work, the beam was adjusted to a diameter of 2.0 μm to cover a variety of hole diameters, including the 1.6 μm holes found on Quantifoil grids. In the present study, the beam diameter was kept at 1.3 μm, covering less than half the area compared with the previous study. The reduction in BIM may thus be similar to that observed with the spot-scan technique (Bullough and Henderson, 1987), arising primarily from a smaller beam size. Secondly, while in the earlier work the area exposed to the electron beam included entire holes in the carbon support film, we limited the exposed area in our present work to include only one side of a hole. It is not yet clear how much of each of these differences is responsible for the reduced particle rotations. However, as our analysis shows, taking into account the particle rotations in the movie frame alignment has a fairly insignificant effect on the reconstructions. This indicates that particle rotations can be ignored in a simplified processing scheme, especially when dealing with smaller particles, for which a given rotational blur will have less of an impact as compared to larger particles.
The recording of movies also allows for compensation of stage and microscope instabilities. While exposure times have been traditionally limited to 1 – 2 seconds when working with side-entry stages to avoid blurring due to stage drift, the ability to record movies can reduce or eliminate this constraint. As a result, exposures can be longer, enabling the use of smaller condenser spot sizes and apertures so as to increase beam coherence, which in turn will enhance the high-resolution signal in the recorded images. The ability to record movies therefore represents a way to overcome limited mechanical stability and enhance beam coherence of midrange electron microscopes.
Experimental Procedures
Electron microscopy
Rotavirus DLPs were prepared as described (Street et al., 1982). Three microliters of sample with a concentration of 2.5 mg/mL were applied to a 1.2–1.3 C-flat™ grid, which had been glow discharged for 45s at 20 mA. The grid was immediately plunge-frozen using an FEI Vitrobot Mark 2. The relative humidity was set at 85% with 7 seconds blot time. Movies were collected at 25 frames/second on a DE-12 4k × 3k pixel direct electron detector (Direct Electron, San Diego, USA) mounted on an FEI TF20 electron microscope operating at 200 kV, using a side-entry Gatan 626 cryo holder. Data collection was automated using Leginon (Suloway et al., 2005), which had been adapted to handle movies. The movies contained 18 frames, the first two of which were recorded with a blanked beam to ensure that the entire exposures were captured by the movies. These first two frames were excluded from subsequent processing, leaving 16 usable frames. The beam was adjusted to deliver a dose of about 2 electrons/Å2/frame, or a total dose of about 32 electrons/Å2 over the duration of the movies. The relatively high dose rate of 50 electrons/Å2/second was used to keep the total exposure time short, thereby minimizing stage instabilities as a source of the observed particle motions (see Results). However, as shown here, movie frame alignment compensates for both stage drift and beam-induced motion. In future experiments, it may therefore be more appropriate to use a lower dose rate (e.g. 25 electrons/Å2/second) to increase beam coherence. Magnification adjustment was performed by maximizing the correlation coefficient between density derived from the DLP crystal structure (PDB ID: 3KZ4, McClain et al., 2010) and 3D reconstructions of DLP, giving a calibrated magnification of 42,135x. This corresponded to a pixel size on the specimen of 1.424 Å. Images were underfocused between 0.8 and 2.7 μm.
Movie processing
We collected 561 movies of DLP samples and selected 1,915 particles from 501 movies after visual inspection, excluding micrographs that showed clear evidence of thick ice, contamination or large sample drift. Particles were selected manually from images that were calculated by simple averaging of the frames. Simple frame averages were also used to determine image defocus using the computer program CTFFIND3 (Mindell and Grigorieff, 2003), followed by further contrast transfer function correction with the computer program FREALIGN (Grigorieff, 2007) during refinement. The particle coordinates were then used to excise stacks of 16 frames with 800 × 800 pixel dimensions from the gain and dark-current corrected movies using computer program label from the MRC image processing suite (Crowther et al., 1996). These “particle” movies were used for further processing to track sample motion and calculate 3D reconstructions.
Euler angles and x,y coordinates of the excised DLPs were initially determined from the simple averages of the 16 movie frames and used for 3D reconstruction using FREALIGN (Grigorieff, 2007). Refinement of the reconstruction was done as described (Chen et al., 2009) until there was no further improvement in the resolution. During the refinement, only data up to a resolution of 7.5 Å were used to avoid possible over-fitting and bias in the FSC curves at higher resolution (Grigorieff, 2000; Stewart and Grigorieff, 2004). Based on the correlation coefficients calculated by FREALIGN, the best 42% (807) of the particles were selected for further processing. With respect to reconstructions using the full data set, the selection of this subset of the particles did not change the resolution of the reconstructions as measured by different FSC criteria (FSC = 0.5, Bottcher et al., 1997; FSC = 0.143, Rosenthal and Henderson, 2003). This result indicates that about 60% of the particles did not contribute significantly to the high-resolution signal in the reconstruction. A similar percentage of particle images (46%) was also selected from the 2008 data set of DLP (Zhang et al., 2008), thus ensuring similar conditions used in both cases for a more meaningful comparison of the final density maps in Fig. 4. 13-fold non-icosahedral averaging was performed as described (Zhang et al., 2008), using the same symmetry matrices that were derived for the 2008 data set, again to maintain comparable conditions.
For the alignment of movie frames to compensate for sample motion, we developed two new computer programs: INTERPOLATE, which calculated shifts for individual frames based on FREALIGN alignment parameters determined from frame averages, and UNBLUR, which applies the interpolated shifts to the movie frames. These programs are freely available from the Grigorieff lab web page (http://emlab.rose2.brandeis.edu/).
Highlights.
Cryo-EM images are blurred by sample motion during electron beam exposure.
The sample motion degrades signal especially at high resolution.
Aligning frames of movies recorded with a direct electron detector reduces blurring.
This procedure enhances the resolution obtained in 3D reconstructions of the sample.
Acknowledgments
The authors thank John Crum, Jim Pulokas and Sargis Dallakyan (National Resource for Automated Molecular Microscopy at the Scripps Research Institute) for technical support, Tilak Jain for supervising data collection during the night, and Joel Quispe for help with grid preparation. The work was supported by National Institutes of Health Grant P01 GM62580 (awarded to S.C.H. and N.G.) and RR017573 (B.C., C.P., A.C.). The work presented here was conducted at the National Resource for Automated Molecular Microscopy which is supported by the National Institutes of Health though the National Center for Research Resources (2P41RR017573-11) and the National Institute of General Medical Sciences (9 P41 GM103310-11). DV was supported by a FP7 Marie-Curie IOF fellowship (273427).
Uncorrected and motion-compensated 3D reconstructions of DLP and the 13-fold locally averaged VP6 trimer have been deposited in the Electron Microscopy Data Bank (accession numbers: DLP (uncorrected), EMD-5485; DLP (compensated for particle rotations and shifts), EMD-5486; VP6 (uncorrected), EMD-5487; VP6 (compensated for particle rotations and shifts), EMD-5488).
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- Baker LA, Smith EA, Bueler SA, Rubinstein JL. The resolution dependence of optimal exposures in liquid nitrogen temperature electron cryomicroscopy of catalase crystals. J Struct Biol. 2010;169:431–437. doi: 10.1016/j.jsb.2009.11.014. [DOI] [PubMed] [Google Scholar]
- Bottcher B, Wynne SA, Crowther RA. Determination of the fold of the core protein of hepatitis B virus by electron cryomicroscopy. Nature. 1997;386:88–91. doi: 10.1038/386088a0. [DOI] [PubMed] [Google Scholar]
- Brilot AF, Chen JZ, Cheng A, Pan J, Harrison SC, Potter CS, Carragher B, Henderson R, Grigorieff N. Beam-induced motion of vitrified specimen on holey carbon film. J Struct Biol. 2012;177:630–637. doi: 10.1016/j.jsb.2012.02.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bullough P, Henderson R. Use of spot-scan procedure for recording low-dose micrographs of beam-sensitive specimens. Ultramicroscopy. 1987;21:223–229. [Google Scholar]
- Cambie R, Downing KH, Typke D, Glaeser RM, Jin J. Design of a microfabricated, two-electrode phase-contrast element suitable for electron microscopy. Ultramicroscopy. 2007;107:329–339. doi: 10.1016/j.ultramic.2006.09.001. [DOI] [PubMed] [Google Scholar]
- Chen JZ, Settembre EC, Aoki ST, Zhang X, Bellamy AR, Dormitzer PR, Harrison SC, Grigorieff N. Molecular interactions in rotavirus assembly and uncoating seen by high-resolution cryo-EM. Proc Natl Acad Sci U S A. 2009;106:10644–10648. doi: 10.1073/pnas.0904024106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Crowther RA, Henderson R, Smith JM. MRC image processing programs. J Struct Biol. 1996;116:9–16. doi: 10.1006/jsbi.1996.0003. [DOI] [PubMed] [Google Scholar]
- Danev R, Nagayama K. Transmission electron microscopy with Zernike phase plate. Ultramicroscopy. 2001;88:243–252. doi: 10.1016/s0304-3991(01)00088-2. [DOI] [PubMed] [Google Scholar]
- Ermantraut E, Wohlfart K, Tichelaar W. Perforated support foils with pre-defined hole size, shape and arrangement. Ultramicroscopy. 1998;74:75–81. [Google Scholar]
- Glaeser RM. Review: electron crystallography: present excitement, a nod to the past, anticipating the future. J Struct Biol. 1999;128:3–14. doi: 10.1006/jsbi.1999.4172. [DOI] [PubMed] [Google Scholar]
- Goddard TD, Huang CC, Ferrin TE. Visualizing density maps with UCSF Chimera. J Struct Biol. 2007;157:281–287. doi: 10.1016/j.jsb.2006.06.010. [DOI] [PubMed] [Google Scholar]
- Grigorieff N. Resolution measurement in structures derived from single particles. Acta Crystallogr D Biol Crystallogr. 2000;56:1270–1277. doi: 10.1107/s0907444900009549. [DOI] [PubMed] [Google Scholar]
- Grigorieff N. FREALIGN: high-resolution refinement of single particle structures. J Struct Biol. 2007;157:117–125. doi: 10.1016/j.jsb.2006.05.004. [DOI] [PubMed] [Google Scholar]
- Grigorieff N, Harrison SC. Near-atomic resolution reconstructions of icosahedral viruses from electron cryo-microscopy. Curr Opin Struct Biol. 2011;21:265–273. doi: 10.1016/j.sbi.2011.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hayward SB, Glaeser RM. Radiation damage of purple membrane at low temperature. Ultramicroscopy. 1979;04:201–210. doi: 10.1016/s0304-3991(79)90211-0. [DOI] [PubMed] [Google Scholar]
- Henderson R. The potential and limitations of neutrons, electrons and X-rays for atomic resolution microscopy of unstained biological molecules. Q Rev Biophys. 1995;28:171–193. doi: 10.1017/s003358350000305x. [DOI] [PubMed] [Google Scholar]
- Henderson R, Baldwin JM, Ceska TA, Zemlin F, Beckmann E, Downing KH. Model for the structure of bacteriorhodopsin based on high-resolution electron cryo-microscopy. J Mol Biol. 1990;213:899–929. doi: 10.1016/S0022-2836(05)80271-2. [DOI] [PubMed] [Google Scholar]
- Henderson R, Chen S, Chen JZ, Grigorieff N, Passmore LA, Ciccarelli L, Rubinstein JL, Crowther RA, Stewart PL, Rosenthal PB. Tilt-pair analysis of images from a range of different specimens in single-particle electron cryomicroscopy. J Mol Biol. 2011;413:1028–1046. doi: 10.1016/j.jmb.2011.09.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jensen GJ. Alignment error envelopes for single particle analysis. J Struct Biol. 2001;133:143–155. doi: 10.1006/jsbi.2001.4334. [DOI] [PubMed] [Google Scholar]
- Kuhlbrandt W, Wang DN, Fujiyoshi Y. Atomic model of plant light-harvesting complex by electron crystallography. Nature. 1994;367:614–621. doi: 10.1038/367614a0. [DOI] [PubMed] [Google Scholar]
- Majorovits E, Barton B, Schultheiss K, Perez-Willard F, Gerthsen D, Schroder RR. Optimizing phase contrast in transmission electron microscopy with an electrostatic (Boersch) phase plate. Ultramicroscopy. 2007;107:213–226. doi: 10.1016/j.ultramic.2006.07.006. [DOI] [PubMed] [Google Scholar]
- McClain B, Settembre E, Temple BR, Bellamy AR, Harrison SC. X-ray crystal structure of the rotavirus inner capsid particle at 3.8 A resolution. J Mol Biol. 2010;397:587–599. doi: 10.1016/j.jmb.2010.01.055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McMullan G, Chen S, Henderson R, Faruqi AR. Detective quantum efficiency of electron area detectors in electron microscopy. Ultramicroscopy. 2009;109:1126–1143. doi: 10.1016/j.ultramic.2009.04.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Milazzo AC, Leblanc P, Duttweiler F, Jin L, Bouwer JC, Peltier S, Ellisman M, Bieser F, Matis HS, Wieman H, et al. Active pixel sensor array as a detector for electron microscopy. Ultramicroscopy. 2005;104:152–159. doi: 10.1016/j.ultramic.2005.03.006. [DOI] [PubMed] [Google Scholar]
- Mindell JA, Grigorieff N. Accurate determination of local defocus and specimen tilt in electron microscopy. J Struct Biol. 2003;142:334–347. doi: 10.1016/s1047-8477(03)00069-8. [DOI] [PubMed] [Google Scholar]
- Muller H, Jin J, Danev R, Spence J, Padmore H, Glaeser RM. Design of an electron microscope phase plate using a focused continuous-wave laser. New J Phys. 2010;12 doi: 10.1088/1367-2630/12/7/073011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nogales E, Wolf SG, Downing KH. Structure of the alpha beta tubulin dimer by electron crystallography. Nature. 1998;391:199–203. doi: 10.1038/34465. [DOI] [PubMed] [Google Scholar]
- Penczek PA, Kimmel M, Spahn CM. Identifying conformational states of macromolecules by eigen-analysis of resampled cryo-EM images. Structure. 2011;19:1582–1590. doi: 10.1016/j.str.2011.10.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Quispe J, Damiano J, Mick SE, Nackashi DP, Fellmann D, Ajero TG, Carragher B, Potter CS. An improved holey carbon film for cryo-electron microscopy. Microsc Microanal. 2007;13:365–371. doi: 10.1017/S1431927607070791. [DOI] [PubMed] [Google Scholar]
- Rosenthal PB, Henderson R. Optimal determination of particle orientation, absolute hand, and contrast loss in single-particle electron cryomicroscopy. J Mol Biol. 2003;333:721–745. doi: 10.1016/j.jmb.2003.07.013. [DOI] [PubMed] [Google Scholar]
- Sachse C, Chen JZ, Coureux PD, Stroupe ME, Fandrich M, Grigorieff N. High-resolution electron microscopy of helical specimens: a fresh look at tobacco mosaic virus. J Mol Biol. 2007;371:812–835. doi: 10.1016/j.jmb.2007.05.088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Scheres SH, Gao H, Valle M, Herman GT, Eggermont PP, Frank J, Carazo JM. Disentangling conformational states of macromolecules in 3D-EM through likelihood optimization. Nat Methods. 2007;4:27–29. doi: 10.1038/nmeth992. [DOI] [PubMed] [Google Scholar]
- Stewart A, Grigorieff N. Noise bias in the refinement of structures derived from single particles. Ultramicroscopy. 2004;102:67–84. doi: 10.1016/j.ultramic.2004.08.008. [DOI] [PubMed] [Google Scholar]
- Street JE, Croxson MC, Chadderton WF, Bellamy AR. Sequence diversity of human rotavirus strains investigated by northern blot hybridization analysis. J Virol. 1982;43:369–378. doi: 10.1128/jvi.43.2.369-378.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Suloway C, Pulokas J, Fellmann D, Cheng A, Guerra F, Quispe J, Stagg S, Potter CS, Carragher B. Automated molecular microscopy: the new Leginon system. J Struct Biol. 2005;151:41–60. doi: 10.1016/j.jsb.2005.03.010. [DOI] [PubMed] [Google Scholar]
- Tang G, Peng L, Baldwin PR, Mann DS, Jiang W, Rees I, Ludtke SJ. EMAN2: an extensible image processing suite for electron microscopy. J Struct Biol. 2007;157:38–46. doi: 10.1016/j.jsb.2006.05.009. [DOI] [PubMed] [Google Scholar]
- Thon F. Zur Defokussierungsabhängigkeit des Phasenkontrastes bei der elektronenmikroskopischen Abbildung. Zeitschrift für Naturforschung. 1966;21a:476–478. [Google Scholar]
- Unwin N. Refined structure of the nicotinic acetylcholine receptor at 4A resolution. J Mol Biol. 2005;346:967–989. doi: 10.1016/j.jmb.2004.12.031. [DOI] [PubMed] [Google Scholar]
- Yonekura K, Maki-Yonekura S, Namba K. Complete atomic model of the bacterial flagellar filament by electron cryomicroscopy. Nature. 2003;424:643–650. doi: 10.1038/nature01830. [DOI] [PubMed] [Google Scholar]
- Zhang X, Settembre E, Xu C, Dormitzer PR, Bellamy R, Harrison SC, Grigorieff N. Near-atomic resolution using electron cryomicroscopy and single-particle reconstruction. Proc Natl Acad Sci U S A. 2008;105:1867–1872. doi: 10.1073/pnas.0711623105. [DOI] [PMC free article] [PubMed] [Google Scholar]