Abstract
Background
Aerogels are a versatile group of nanostructured/nanoporous materials with physical and chemical properties that can be adjusted to suit the application of interest. In terms of biomedical applications, aerogels are particularly suitable for implants such as membranes, tissue growth scaffolds, and nerve regeneration and guidance inserts. The mesoporous nature of aerogels can also be used for diffusion based release of drugs that are loaded during the drying stage of the material. From the variety of aerogels polyurea crosslinked silica aerogels have the most potential for future biomedical applications and are explored here.
Methodology
This study assessed the short and long term biocompatibility of polyurea crosslinked silica aerogel implants in a Sprague-Dawley rat model. Implants were inserted at two different locations a) subcutaneously (SC), at the dorsum and b) intramuscularly (IM), between the gluteus maximus and biceps femoris of the left hind extremity. Nearby muscle and other internal organs were evaluated histologically for inflammation, tissue damage, fibrosis and movement (travel) of implant.
Conclusion/Significance
In general polyurea crosslinked silica aerogel (PCSA) was well tolerated as a subcutaneous and an intramuscular implant in the Sprague-Dawley rat with a maximum incubation time of twenty months. In some cases a thin fibrous capsule surrounded the aerogel implant and was interpreted as a normal response to foreign material. No noticeable toxicity was found in the tissues surrounding the implants nor in distant organs. Comparison was made with control rats without any implants inserted, and animals with suture material present. No obvious or noticeable changes were sustained by the implants at either location. Careful necropsy and tissue histology showed age-related changes only. An effective sterilization technique for PCSA implants as well as staining and sectioning protocol has been established. These studies further support the notion that silica-based aerogels could be useful as biomaterials.
Introduction
Porous biocompatible materials have received particular attention in recent years for a broad range of applications. From filters and prostheses to scaffolds for tissue engineering, porous biomaterials have been under constant development and improvement for biological and biomedical applications [1], [2], [3], [4], [5], [6], [7]. Complications such as poor mechanical performances, batch-to-batch purity variations, and large pore sizes (µm –range) have limited the extent of use of naturally occurring biomaterials [8] even though they most closely simulate the native cellular environment.
Large pore sizes restrict the use of this class of porous materials to whole cell penetration and tissue infiltration applications only, making them unsuitable for drug delivery and protein loading applications where the physical size of the entities of interest are on the order of 10 s of nanometers rather than microns [9], [10]. Studies have also shown that cellular uptake of drugs is more efficient at the nm scale level [11] emphasizing the need for nm-scale porous materials and membranes as ideal tools for drug delivery. The cell size of porous biomaterials also plays an important role in the formation of scar tissue and fibrosis where major effort is invested to minimize these formations. There is evidence to suggest that closely spaced nanometer-sized pores prevent formation of extensive fibrous connective and scar tissue and in fact promotes superior tissue integration [12], [13]. Additionally, from a mechanical behavior point of view stress concentration greatly depends on pore size and is proven to be significantly less for materials with smaller pore diameters such as mesoporous materials (pore diameter<50 nm) [14], [15].
Aerogels are nanostructured, open-mesoporous (pore diameter<50 nm) ultra low-density lightweight materials with a high surface-to-volume ratio [16] and tunable surface and bulk properties allowing for control over potentially key parameters such as surface wettability, density, opacity, pore size and shape to name a few [17]. Of particular interest to biological applications are the polyurea crosslinked silica aerogels (PCSA) where significant mechanical strength has been accomplished by covalent crosslinking of the skeletal nanoparticles with polyurea without significant compromise of the porosity and the low bulk density of the native material [18], [19]. The biocompatibility of silica particles has been widely explored and has shown a great deal of promise and compatibility with living matter [20], [21], [22], [23]. Drug delivery by porous silica has also been explored and offers great promise [24], [25]. At the cellular level, recent in vitro studies on PCSA has shown good biocompatibility of this material [26], [27], [28] and the ability to manipulate the growth of neurons on the PCSA surface with the aid of a laminin layer [26].
The versatility of sol-gel method for material preparation allows for doping and pigmenting techniques to be incorporated at the synthesis stage of aerogels with relative ease [29], [30]. Previous investigations of pigmenting PCSA demonstrated a stable and non-leaching chemistry while also strengthening the mechanically properties and retaining the porosity of the aerogels [31]. This can be of significant interest for biomedical applications since the majority of polymeric implants used today are translucent or transparent. Post-surgical identification of a colored implant may be easier.
In this work, we report the first in vivo assessment of the biocompatibility of both clear and pigmented PCSA implants in a rat model sterilized following an ethylene oxide (EtO) sterilization protocol. The presence of the implants did not result in any morbidity. In addition, histological evaluation of the tissues surrounding the implants as well as distant organ sites did not demonstrate inflammation or substantial tissue damage. The results of the PCSA implant study were compared to three control groups: 1) Suture material present (peripheral nerve transection followed by suture repair), 2) sham surgery, and 3) no surgery. Results presented in this study demonstrate the biocompatibility of PCSA thereby supporting the notion that these materials could be useful for biomedical applications.
Materials and Methods
Preparation of clear and pigment-doped aerogels
Pigmented and clear polyurea crosslinked silica aerogels were synthesized according to our previously described formulation [30]. The aerogel implants were cut and roughly shaped from the bulk material using a diamond tipped abrasive disk mounted in a rotary tool. The samples were then polished by hand to a final size of 5×2×2 mm3 for the back muscle study and 5×2×2 mm3 for the deep muscle study. The final stage of implant preparation was exposure to a small vibratory tumbler containing aluminum oxide grinding media (approximately 200 mesh), and processed for about 2 days. The surface contact angle of these aerogels was measured to be around 45° and a density of 0.4 g/cm3 for clear and 0.5 g/cm3 for pigmented samples was measured. All implants were sterilized for 24 hrs by a standard ethylene oxide sterilization process in an Amprolene system (Anderson Products) prior to surgery.
Surgical procedure(s)
All surgical procedures were performed on male or female Sprague-Dawley rats weighing 200–300 gr. Rats were anesthetized 15 min prior to surgery via an initial intramuscular injection of Telazol (0.03–0.05 ml at 0.3–0.5 mg/kg), followed by isoflurane inhalation at 1–2%. During the post-surgery recovery period animals were not restrained and were allowed to continue with their normal grooming routine. They were fed a routine diet and kept under close observation for signs of infection or abnormal behavior. Rats were euthanized post implant recovery by an overdose of carbon dioxide. This study was approved by the Animal Care and Use Committee at the University of Memphis.
Surgical group 1: Subcutaneous Implant insertion (n = 4)
An area on the back, approximately 3 cm2, was shaved and a 1 cm incision was made with a scalpel. One pigmented and one clear implant was inserted subcutaneously in the back of rats (total of 8 implants). Samples were inserted and immobilized without the use of any sutures, adhesives, or staples. Figure 1 shows clear (Figure 1a) and pigmented (Figure 1b) SC PCSA implants after incubation period, prior to extraction.
Surgical group 2: Intramuscular implant insertion (n = 2)
The left hind limb of rats were abducted and shaved. Under a dissecting microscope, a transverse incision was made through the skin of the limb halfway between the iliac crest and the femur's articulation with the tibia. A self retracting retainer was replaced, and dissection was carried down between the biceps femoris and gluteus muscle until the sciatic nerve was identified. The implant was positioned between the muscle and the sciatic nerve such that the nerve was in direct contact with the surface of the implant. No adhesive or sutures were used for immobilizing the aerogel implant. The skin was then closed with staples and the animal was then allowed to emerge from anesthesia.
Surgical group 3: Suture repair (n = 2)
The sciatic nerve branches of rats were exposed and transected sharply as described above. The nerve segments were then coapted using standard epineural suture technique with interrupted 9-0 nylon.
Surgical group 4: Sham surgery (n = 1)
The sciatic nerve of rats was exposed and severed again sharply, similar to the method used for suture repair but after severance, nerve endings were abandoned and no coaptation was attempted.
Implant and tissue retrieval
At time of sacrifice each animal was weighed and its weight recorded. All animals under investigation in this study continued to gain weight steadily throughout the implant incubation period at the same rate as the control groups. At appropriate recovery time points previous incisions were reopened and the aerogel implant was identified and removed with a cuff of the surrounding tissue intact for further analysis. At two weeks, four of the subcutaneously implanted clear and pigmented PCSA samples (surgical group 1) were removed for analysis. At twenty months four more subcutaneously inserted clear and pigmented PCSA samples (surgical group 1) were removed along with vital organs. Finally, at seven months time point PCSA implants and vital organs were removed from surgical group 2. For the sake of comparison tissue samples from the surgery sites as well as vital organs were removed from surgical groups 2, 3, and 4 at the same time point. Post surgery, animals were anesthetized with inhaled agent and sacrificed at each time point with an overdose of carbon dioxide. All tissue samples were transferred to jars containing 10% formaldehyde.
Histological staining and examination
Immediately after dissection, tissues containing implants as well as tissues from distant organs were fixed for at least 18 hrs in 10% formaldehyde. Tissue segments obtained from organs without implant were embedded in paraffin and the sections obtained from these paraffin blocs were stained by hematoxylin eosin (H&E) following routine procedures.
Initial experiments revealed that adhesion of sectioned PCSA aerogel implants to glass slides was challenging. Two protocols were tested to overcome this difficulty. First, implant containing tissue samples were embedded in polymethylmethacrylate (PMMA) and the sections obtained from these blocs were mounted onto plastic slides for staining. This resulted in satisfactory adhesion of tissue/aerogel samples onto the slides but the refractive index of the plastic hindered optimal light microscopy observations. In the second protocol, the implant containing tissue sections were embedded in paraffin following standard procedures. Sections of 5 µm thickness were obtained from the paraffin blocks and various functionalized glass slide surfaces were tested with respect to their ability to bind the implant present in the tissue sections. Slides coated with silane (Electron Microscopy Sciences), poly-L-lysine (Electron Microscopy Sciences), or super-frost coating (Thermo Fisher Scientific) were unsatisfactory as the aerogel implants detached readily early during the staining process. In contrast, the PCSA implant material adhered through the whole staining and mounting procedure on glass slides coated with egg albumin (Newcomers Supply). Sections containing the implant were stained with H&E. Double blinded tissue section evaluations were performed by a pathologist who, in the case of sections without implants, was unaware of whether or not the tissue examined was obtained from an animal with or without implants. Figure 2 shows the effect of the staining protocol on the clear (Figure 2a) and pigmented (Figure 2b) control PCSA sections. It can be seen that the PCSA alone does uptake the stain to a certain extent. Striations seen in the image are considered to be staining artifacts and do not reflect PCSA surface morphology. At times, sectioning the paraffin-embedded tissue-implant samples was difficult due to the hardness of the PCSA implant and several attempts had to be made in order to create a uniform section.
Results and Discussion
Implant incubation times were chosen based on the three contact duration categories that biomaterials and medical devices are recommended for testing namely 1) Limited (<24 hrs), 2) Prolonged (>24 hrs and <30 days) and 3) Permanent (>30 days) [32]. Aerogel-based biomaterials are expected to serve under classifications 2 and 3 and incubation periods were chosen accordingly
Both clear and pigmented PCSA samples were implanted a) subcutaneously (SC) at the dorsum and b) intramuscularly (IM) between the gluteus maximus and biceps femoris of the left hind extremity in Sprague–Dawley rats. Subcutaneous implants were removed at two weeks and twenty month time points while IM implants were extracted after seven months of incubation.
Implant Condition
The physical appearance and size of the recovered implants was not altered as a result of short or long term in vivo incubation. Macroscopic and microscopic (optical) evaluation of implants showed no signs of erosion, surface deterioration, or fragmenting. Additionally, the implants inserted at the different locations did not appear to have traveled from their original location even though no immobilization (sutures, adhesives, etc) was used. This is particularly surprising in the case of the implants inserted in the hind leg, since the animal had continuous motion and grooming ability and its motion was not restricted during the recovery period. We attribute this to the low density and light weight nature of the aerogel implant. It is likely that the nanoporous/mesoporous and three dimensional nature of the surface of the aerogel implants created anchoring sites between the nearby tissue and the surface of the PCSA implant.
Inflammation
In general, irritation and inflammation of the tissue that is in direct contact with implants is a concern [33]. In this case however, no significant inflammation was observed for any of the time points, at the interface of the aerogel implants and the nearby tissue for both SC and IM surgery types. The amount of inflammation (according to pathologists report) was categorized as mild in all cases such that the amount of inflammation observed for PCSA implants was comparable to the amount of inflammation observed for biocompatible materials particularly surgical steel and nylon sutures (Tables 1 and 2). This is likely due to the nanoporous/mesoporous nature of the aerogel implant surface that while it is highly non-uniform at the nanometer scale, to the large muscle cells nearby it presents itself smooth and as a result causes minimum irritation and inflammation. A mild inflammation is expected for all foreign materials including biomaterials and is recognized as part of the body's foreign body response [34]. The response seen in this study to the PCSA implants is identified as a normal reaction.
Table 1. Summary of short and long term effect of subcutaneous aerogel implant on local tissue.
Incubation PeriodResponse | Two weeks | Twenty months | ||
Clear | Pigmented | Clear | Pigmented | |
Inflammation | Mild | Mild | Mild | Mild |
Fibrosis | Mild | Mild | Mild | Mild |
Infection | None | None | None | None |
Implant travel | None | None | None | None |
Age of rat at time of extraction | ∼3 months | ∼2 years |
Slides from both control and implant-containing rats were reviewed blindly by pathologist, specifically looking for signs of inflammation, infection, fibrosis, and implant travel.
Table 2. Summary of the effect of intramuscular aerogel implant on local tissue.
Seven month period | ||||
ProcedureResponse | Sham | Suture | Implant-clear | Implant-pigmented |
Inflammation | Mild | Mild | Mild | Mild |
Fibrosis | Mild | Mild | Mild | Mild |
Infection | None | None | None | None |
Age of rat at time of extraction | 9–10 months old |
Slides from both control and implant-containing rats were reviewed blindly by pathologist, specifically looking for signs of inflammation, infection, and fibrosis.
Fibrosis
In general, the biocompatibility of novel biomaterials with tissue is evaluated based on the in vivo inflammatory responses as well as the fibrosis formed around the implant [35]. Fibrous capsule formation is a well-established reaction to implanted biomaterials and is recognized as the end stage of the foreign body reaction [34]. In this study a mild fibrosis and capsule formation was observed for almost all of aerogel implant-tissue interfaces. Figure 3 shows the SC PCSA-tissue interface behavior for a pigmented sample. In Figure 3a the whole implant is imaged while Figures 3b, 3c, and 3d show high magnification images of interface and the fibrous layer formed around the implant for spots 1, 2, and 3 respectively as identified on the image (Figure 3a).
In Figure 4 example cross sectional images of the SC implant insertion after twenty months of incubation are shown for both clear (Figure 4a) and pigmented (Figure 4b) PCSA samples. The dotted line in each case outlines the boundary between the aerogel and the nearby tissue. Some staining of the PCSA section can be seen in each case. The thickness of the fibrous layer for the SC surgery, around the clear and pigmented aerogel implants between two weeks (Figure 3) and twenty months (sample images shown in Figure 4) time points did not show a significant difference. In all cases native cellular tissue had grown up to the implant surfaces. It is not clear from the tests performed here if any tissue had grown into the aerogel pores and interstices. From the histological evaluations it was concluded that the fibrosis around the pigmented implants wasn't significantly greater than fibrosis around clear implants. Table 1 summarizes these results.
Next, the fibrous layer formed around biocompatible materials such as surgical sutures and surgical tools (images not shown) were compared with the fibrous layer formed around PCSA implants inserted intramuscularly. High magnification images of the interface of clear (Figure 5a) and pigmented (Figure 5b) PCSA with nearby tissue extracted after seven months of IM incubation was studied and results are summarized in Table 2. Fibrosis observed in all cases was classified as “mild” and was in agreement with the amount of fibrosis observed at the surgery site for suture and sham surgery procedures. Again, cellular tissue had grown at least up to the implant and perhaps growth has also occurred into the structure of the aerogels although, no specific tests were performed to confirm this.
Infection
No infection was observed in any of the implant or control conditions, at any of the time points, suggesting that ethylene oxide sterilisation is an effective method for sterilising PCSA implants.
Toxicity
Toxicity of novel biomaterials is often evaluated by monitoring the response of vital organs distant to the implant. In particular organs such as spleen, lung, heart, kidney, and intestine must be evaluated histopathologicaly [36] for any signs of abnormal responses. In Table 3 the long term effect of aerogels' presence in the system of Sprague-Dawley rats has been summarized. The evaluation of organs removed from animals with 1) suture surgery, 2) no surgery (control), and 3) sham surgery was classified as unremarkable. The organs removed from animals carrying an aerogel implant of some kind were indistinguishable from those previously mentioned. This indicates that in the initial study, no noticeable systemic reaction or immune response has been triggered by the presence of the aerogel implants.
Table 3. Summary of long term effect of intramuscular and subcutaneous aerogel implants on organs.
ProcedureOrgan | Sham surgery | Suture surgery | Control (no surgery) | Pigmented/clear aerogel implant-SC | Pigmented/clear aerogel implant-IM |
Spleen | Unremarkable | Unremarkable | Unremarkable | Undistinguishable | Undistinguishable |
Lung | Unremarkable | Unremarkable | Unremarkable | Undistinguishable | Undistinguishable |
Heart | Unremarkable | Unremarkable | Unremarkable | Undistinguishable | Undistinguishable |
Kidney | Unremarkable | Unremarkable | Unremarkable | Undistinguishable | Undistinguishable |
Intestine | Unremarkable | Unremarkable | Unremarkable | NA | NA |
Summary and Conclusion
This study involves the utilization of in vivo tests to determine the general biocompatibility of PCSA as a biomaterial. Double blinded reviews by pathologists showed no statistical difference between tissue samples collected from different surgical groups suggesting tolerance and biocompatibility of PCSA. Early in vivo assessment of tissue compatibility presented here can be used to influence the design criteria of future PCSA-based medical devices.
Ethylene oxide sterilization has proven to be an effective method for sterilizing aerogel implants. The in vivo studies performed here on a small group of Sprague-Dawley rats have demonstrated biocompatibility of polyurea crosslinked silica aerogels over a maximum of twenty month incubation period. Fibrosis observed was at the normal level that would be observed with any foreign object including sutures. Although this work was performed on a small group of animals, it provides the basis for continuation of the study on a larger group of animals and shows promise for PCSA as a biomaterial. The nanometer scale roughness of the aerogel surface seems to play a significant role in the limited range of motion and travel of the implant from the surgery site therefore, eliminating the need for sutures. The results of our in vivo experiments suggest the need for further careful study to better understand the fundamental processes involved in the interaction between this specific type of material and the living body.
Acknowledgments
The authors would like to acknowledge Ms. Renada Scott at the Integrated Microscopy Center (IMC), the University of Memphis for histology and sectioning work as well as University of Alabama at Birmingham, Center for Metabolic Bone Disease–Histomorphometry and Molecular Analysis Core Laboratory. Authors would also like to thank Dr Anand Kulkarni and his team at UTHSC- Department of Pathology for pathology evaluations. Finally we are grateful to Dr Karyl Buddington for all surgical procedures conducted at UoM.
Funding Statement
The first author would like to thank the FedEx Institute of Technology, University of Memphis, for finanical support. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Power M, Hosticka B, Black E, Daitch C, Norris P (2001) Aerogels as biosensors: viral particle detection by bacteria immobilized on large pore aerogel. Journal of Non-Crystalline Solids 285: 303–308. [Google Scholar]
- 2. Ren L, Tsuru K, Hayakawa S, Osaka A (2002) Novel approach to fabricate porous gelatin-siloxane hybrids for bone tissue engineering. Biomaterials 23: 4765–4773. [DOI] [PubMed] [Google Scholar]
- 3. Chang MC, Ko CC, Douglas WH (2003) Preparation of hydroxyapatite-gelatin nanocomposite. Biomaterials 24: 2853–2862. [DOI] [PubMed] [Google Scholar]
- 4. Lyons L, Cauble M, Cole JA, Sabri F (1999) Investigation of biological adhesives and polyurea-crosslinked silica-based aerogels. 76th Annual Meeting of the Southeastern Section of APS Vol. 54, No. 16. [Google Scholar]
- 5. Karageorgiou V, Kaplan D (2005) Porosity of 3D biomaterial scaffolds and osteogenesis. Biomaterials 26: 5474–5491. [DOI] [PubMed] [Google Scholar]
- 6. Thomas V, Dean DR, Vohra YK (2006) Nanostructured Biomaterials for Regenerative Medicine. Current Nanoscience 2: 155–177. [Google Scholar]
- 7. Yang S, Leong KF, Du Z, Chua CK (2001) The Design of Scaffolds for Use in Tissue Engineering. Part I. Traditional Factors. Tissue Engineering 7: 679–689. [DOI] [PubMed] [Google Scholar]
- 8. Kim BS, Mooney DJ (1998) Development of biocompatible synthetic extracellular matrices for tissue engineering. TIBTECH 16: 224–229. [DOI] [PubMed] [Google Scholar]
- 9. Jahanshahi M, Babaei Z (2008) Protein nanoparticle: A unique system as drug delivery. Vehicles African Journal of Biotechnology 7: 4926–4934. [Google Scholar]
- 10.Hughes GA (2005) Nanostructure-mediated drug delivery Nanomedicine: Nanotechnology, Biology, and Medicine 1: :22–30. [DOI] [PubMed] [Google Scholar]
- 11. Liu H, Liu T, Li L, Hao N, Tan L, et al. (2012) Size dependent cellular uptake, in vivo fate and light–heat conversion efficiency of gold nanoshells on silica nanorattles. Nanoscale 4: 3523–3529. [DOI] [PubMed] [Google Scholar]
- 12. Slack M, Sandhu JS, Staskin DR, Grant RC (2006) In Vivo Comparison of Suburethral Sling Materials. International Urogynecology Journal 17: 106–110. [DOI] [PubMed] [Google Scholar]
- 13. Klinge U, Klosterhalfen B, Birkenhauer V, Junge K, Conze J, et al. (2002) Impact of Polymer Pore Size on the Interface Scar Formation In a Rat Model. Journal of Surgical Research 103: 208–214. [DOI] [PubMed] [Google Scholar]
- 14. Boccaccini AR (1998) Influence of stress concentrations on the mechanical property-porosity correlation in porous materials. Journal of Materials Science Letters 17: 1273–1275. [Google Scholar]
- 15. Liu DM (1997) Influence of porosity and pore size on the compressive strength of porous hydroxyapatite ceramic. Ceramics International 23: 135–139. [Google Scholar]
- 16.Leventis N, Lu H (2011) Polymer Crosslinked Aerogels. In: Aegerter M, Leventis N, Koebel M, editors. Aerogels Handbook - Advances in Sol-Gel Derived Materials and Technologies. Springer: New York. pp 251–285.
- 17. Hrubesh LW (1998) Aerogel Applications. Journal of non-Crystalline Solids 225: 335–342. [Google Scholar]
- 18. Leventis N, Sotiriou-Leventis C, Zhang G, Rawashdeh AMM (2002) Nanoengineering strong silica aerogels. Nano Letters 2: 957–960. [Google Scholar]
- 19. Leventis N (2007) Three dimensional core-shell superstructures: Mechanically strong aerogels. Acc Chem Res 40: 874–884. [DOI] [PubMed] [Google Scholar]
- 20. Roy I, Ohulchanskyy TY, Bharali DJ, Pudavar HE, Mistretta RA, et al. (2005) Optical tracking of organically modified silica nanoparticles as DNA carriers: A nonviral, nanomedicine approach for gene delivery. Proceedings of the National Academy of Sciences of the United States of America 102: 279–284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Giri S, Trewyn BG, Lin VSY (2007) Mesoporous silica nanomaterial-based biotechnological and biomedical delivery systems. Nanomedicine 2: 99–111. [DOI] [PubMed] [Google Scholar]
- 22. Giri S, Trewyn BG, Stellmaker MP, Lin VSY (2005) Stimuli-Responsive Controlled-Release Delivery System Based on Mesoporous Silica Nanorods Capped with Magnetic Nanoparticles. Angewandte Chemie International Edition 44: 5038–5044. [DOI] [PubMed] [Google Scholar]
- 23. Smirnova I, Suttiruengwong S, Arlt W (2004) Feasibility study of hydrophilic and hydrophobic silica aerogels as drug delivery systems. Journal of Non-Crystalline Solids 350: 54–60. [Google Scholar]
- 24. Lai CY, Trewyn BG, Jeftinija DM, Jeftinija K, Xu S, et al. (2003) Mesoporous Silica Nanosphere-Based Carrier System with Chemically Removable CdS Nanoparticle Caps for Stimuli-Responsive Controlled Release of Neurotransmitters and Drug Molecules. Journal of the American Chemical Society 125: 4451–4459. [DOI] [PubMed] [Google Scholar]
- 25. Torney F, Trewyn BG, Lin VSY, Wang K (2007) Mesoporous silica nanoparticles deliver DNA and chemicals into plants. Nature Nanotechnology 2: 295–300. [DOI] [PubMed] [Google Scholar]
- 26. Sabri F, Cole JA, Scarbrough MC, Leventis N (2012) Investigation of Polyurea-Crosslinked Silica Aerogels as a Neuronal Scaffold: A Pilot Study. PLoS ONE 7 3: e33242 doi:10.1371/journal.pone.0033242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Sabri F, Cole JA, Scarbrough MC, Leventis N (2011) Investigation of crosslinked silica Aerogels for implant applications. In Proceedings of the Third Annual ORNL Biomedical Science and Engineering Conference: 15–17 March 2011; Knoxville. Catalog number CFP1147G-ART [Google Scholar]
- 28. Yin W, Venkitachalam SM, Jarrett E, Staggs S, Leventis N, et al. (2010) Biocompatibility of surfactant-templated polyurea–nanoencapsulated macroporous silica aerogels with plasma platelets and endothelial cells. Journal of Biomedical Materials Research Part A 92: 1431–1439. [DOI] [PubMed] [Google Scholar]
- 29. Dunn B, Miller JM, Dave BC, Valentine JS, Zink JI (1998) Strategies for encapsulating biomolecules in sol gel matrices. Acta Materialia 46: 737–741. [Google Scholar]
- 30. Schneider M, Baiker A (1997) Titania-based aerogels. Catalysis Today 35: 339–365. [Google Scholar]
- 31. Sabri F, Leventis N, Hoskins J, Schuerger AC, Sinden-Redding M, et al. (2011) Spectroscopic evaluation of polyurea crosslinked aerogels, as a substitute for RTV-based chromatic calibration targets for spacecraft. Advances in Space Research 47: 419–427. [Google Scholar]
- 32. Schuh JCL (2008) Medical Device Regulations and Testing for Toxicologic Pathologists. Toxicologic Pathology 36: 63–69. [DOI] [PubMed] [Google Scholar]
- 33. Anderson JA (2001) Biological Responses to Materials. Annu. Rev. Mater. Res 31: 81–110. [Google Scholar]
- 34.Doherty P (2009) Inflammation, Carcinogenicity and Hypersensitivity. In: Roger Narayan editor. Biomedical Materials. Springer: New York. pp.201–214.
- 35. Spector M, Cease C, Tong-Li X (1989) The local tissue response to Biomaterials. Crit. Rev. Biocompatibility 5: 269–295. [Google Scholar]
- 36. Park JC, Lee DH, Suh H (1999) Preclinical Evaluation of Prototype Products. Yonsei Medical Journal 40: 530–535. [DOI] [PubMed] [Google Scholar]