Abstract
The MIC-2000 96-channel dispenser was evaluated for accuracy and repeatability of dispensing 50-μl volumes of water. It was found to perform within the manufacturer's specifications for accuracy of ±5%, but not within the limits for repeatability of ±1%. Overall, instrument performance was found to be satisfactory for the antimicrobial susceptibility testing technique which has been implemented in this laboratory. A method is suggested for simple evaluation of the volumes dispensed. A variety of antibiotics may be stored in Trypticase soy broth in Microtiter plates without loss of potency for periods of at least 2 weeks at −20°C and 4 months at −70°C.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Barry A. L., Garcia F., Thrupp L. D. An improved single-disk method for testing the antibiotic susceptibility of rapidly-growing pathogens. Am J Clin Pathol. 1970 Feb;53(2):149–158. doi: 10.1093/ajcp/53.2.149. [DOI] [PubMed] [Google Scholar]
- Bauer A. W., Kirby W. M., Sherris J. C., Turck M. Antibiotic susceptibility testing by a standardized single disk method. Am J Clin Pathol. 1966 Apr;45(4):493–496. [PubMed] [Google Scholar]
- Foster J. W., Woodruff H. B. Microbiological Aspects of Penicillin: I. Methods of Assay. J Bacteriol. 1943 Aug;46(2):187–202. doi: 10.1128/jb.46.2.187-202.1943. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MacLowry J. D., Jaqua M. J., Selepak S. T. Detailed methodology and implementation of a semiautomated serial dilution microtechnique for antimicrobial susceptibility testing. Appl Microbiol. 1970 Jul;20(1):46–53. doi: 10.1128/am.20.1.46-53.1970. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MacLowry J. D., Marsh H. H. Semiautomatic microtechnique for serial dilution-antibiotic sensitivity testing in the clinical laboratory. J Lab Clin Med. 1968 Oct;72(4):685–687. [PubMed] [Google Scholar]
- Marymont J. H., Jr, Wentz R. M. Serial dilution antibiotic sensitivity testing with the microtitrator system. Am J Clin Pathol. 1966 May;45(5):548–551. doi: 10.1093/ajcp/45.5.548. [DOI] [PubMed] [Google Scholar]