Abstract

The MIC-2000 96-channel dispenser was evaluated for accuracy and repeatability of dispensing 50-μl volumes of water. It was found to perform within the manufacturer's specifications for accuracy of ±5%, but not within the limits for repeatability of ±1%. Overall, instrument performance was found to be satisfactory for the antimicrobial susceptibility testing technique which has been implemented in this laboratory. A method is suggested for simple evaluation of the volumes dispensed. A variety of antibiotics may be stored in Trypticase soy broth in Microtiter plates without loss of potency for periods of at least 2 weeks at −20°C and 4 months at −70°C.

Full text

PDF
842

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Barry A. L., Garcia F., Thrupp L. D. An improved single-disk method for testing the antibiotic susceptibility of rapidly-growing pathogens. Am J Clin Pathol. 1970 Feb;53(2):149–158. doi: 10.1093/ajcp/53.2.149. [DOI] [PubMed] [Google Scholar]
  2. Bauer A. W., Kirby W. M., Sherris J. C., Turck M. Antibiotic susceptibility testing by a standardized single disk method. Am J Clin Pathol. 1966 Apr;45(4):493–496. [PubMed] [Google Scholar]
  3. Foster J. W., Woodruff H. B. Microbiological Aspects of Penicillin: I. Methods of Assay. J Bacteriol. 1943 Aug;46(2):187–202. doi: 10.1128/jb.46.2.187-202.1943. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. MacLowry J. D., Jaqua M. J., Selepak S. T. Detailed methodology and implementation of a semiautomated serial dilution microtechnique for antimicrobial susceptibility testing. Appl Microbiol. 1970 Jul;20(1):46–53. doi: 10.1128/am.20.1.46-53.1970. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. MacLowry J. D., Marsh H. H. Semiautomatic microtechnique for serial dilution-antibiotic sensitivity testing in the clinical laboratory. J Lab Clin Med. 1968 Oct;72(4):685–687. [PubMed] [Google Scholar]
  6. Marymont J. H., Jr, Wentz R. M. Serial dilution antibiotic sensitivity testing with the microtitrator system. Am J Clin Pathol. 1966 May;45(5):548–551. doi: 10.1093/ajcp/45.5.548. [DOI] [PubMed] [Google Scholar]