SUMMARY
To correlate different CT system: MSCT (multislice computed tomography) with different acquisition parameters (100KV, 80KV), different reconstruction algorithm (ASIR) and CBCT (cone beam computed tomography) examination in terms of absorbed X-ray dose and diagnostic accuracy.
80 KV protocols compared with 100 KV protocols resulted in reduced total radiation dose without relevant loss of diagnostic image information and quality. CBCT protocols compared with 80 KV MSCT protocols resulted in reduced total radiation dose but loss of diagnostic image information and quality although no so relevant.
In addition the new system applies to equipment ASIR applicable on MSCT allows 50% of the dose without compromising image quality.
Keywords: Multislice computed tomography, cone beam CT, radiation dose, orthodontics
RIASSUNTO
L’obiettivo del presente lavoro è di correlare sistema TC differenti quali TC multistrato con differenti parametri di acquisizione (100KV, 80KV) e differenti algoritmi di ricostruzione (ASIR) e CBCT (tomografia a fascio conico) in termini di dose assorbita in raggi X e di accuratezza diagnostica. Il protocollo con 80 KV rispetto a quello con 100 KV ha portato una riduzione di dose di radiazioni totale senza perdita rilevante di informazioni diagnostiche di qualità. Nella CBCT si ha una riduzione di dose di radiazione totale, con modesta perdita di informazioni e di qualità dell’immagine. Il nuovo sistema di riduzione della dose ASIR applicabile alla TCMS consente di abbassare fino al 50% della dose di radiazioni senza compromettere assolutamente la qualità delle immagini.
Introduction
CT images are not part of routine orthodontics records, but the cranio-facial complex can’t be correctly investigated by traditional X-ray techniques due to superimposition of the bones structures.
The Italian law, Article 2 of Legislative Decree 230/95, regulates all activities carrying a risk of ionizing radiation. Therefore, our study continues this line of research by modifying the acquisition parameters in order to reduce the absorbed X-ray (19,20).
For these reasons the demand for Dentascan exams and the use of this technique on children is increased constantly from 1982 when the use of spiral CT was proposed by Timms in orthodontic treatment. The risk factors associated with being exposed to ionized radiation must be considered and more attention to be paid to this diagnostic tool to optimise the technique and reduce dose to the lowest possible level.
The CT, a radiological three-dimensional technique based on computer assessment of the radiation-absorbing characteristics of the various tissues of the body, offers many advantages (15–18):
it provides true tomography images of excellent contrast resolution;
tissue density differences of 0.5% are resolvable;
all the photons that have penetrated the body are used in the formation of the images;
X-ray beams are highly collimated into a band varying from 2–13 mm, so that very little tissue, other than that being imaged, is irradiated.
The problems in adapting helical-CT scans for orthodontic use include high cost, scanning time and most importantly, high radiation exposure, which is of particular concern given that the orthodontic treatment often involves young or adolescent patients.
Techniques
MSCT
The MDCT scanner sufficient for orthodontic applications is at 16 slice.
The scanner equipped with 16 detector rows has a minimal rotation time of 0.5 sec. given a collimation between 0.75 and 1.5 mm with dose calibrated. The scanning tecnique should be applied consists of: preliminary scan performed in the anteroposterior (AP) and laterolateral (LL) projections, with the following acquisition parameters: 100 KV or 80 KV, with 200 mA (1). The subsequent scans should taken with a 1.25 mm slice thickness, 0.6 mm interval, 11.25 mm table speed/rotation, 100 mA, 13.7 cm FOV, 512x512 matrix and 0° gantry angle (Fig. 1). Data acquisition, planning of examinations and printing of axial images and standardized multiplanar reconstructions should be performed on a dedicate workstation. The radiologists should have immediate access for visualization and further multi-planar and 3D reformatting on the workstation. Additional reconstructions could be necessary to clarify the diagnosis or demanded by the referring clinician, and should be performed and printed by radiologists. The examinations and additional reconstructions should sent via a fast 100-MB/seclocal area network-Ethernet to the PACS. The radiologic evaluation process should be based exclusively on image interpretation at the workstation rather than on hard copies or PACS.
Dose estimated data is possible to obtain from the console of the CT scanner:
CTDI (Computer Tomography Dose Index) = dose measured in central and peripheral region of the subjects as a direct result of a CT section acquisition T millimetres thick;
DLP (Dose Length Product)= total dose deposited over the length of the acquisition.
CBCT
Cone beam computed tomography (CBCT) has recently been introduced as an alternative imaging technology for acquiring 3D data for diagnostic tasks such as implant treatment planning (4, 5). This technology requires less expensive components than conventional CT and is reported to produce lower patient exposures (2).
The technique also allows, in principle, to obtain the data acquisition of a volume quickly and with low dose.
In the cone-beam technique is necessary to produce the X-ray photons only the number of times equal to the number of slices needed to cover the volume of interest. The system has an automatic exposure control (AEC), which allows automatic selection of the intensity of X-rays of departure in accordance with the size of the patient and changes in accordance with the thickness.
Cone-beam CT scanners utilize a two dimensional detector, which allows for a single rotation of the gantry to generate a scan of the entire region of interest, as compared to conventional CT scanners whose multiple “slices”. must be stacked to obtain a complete image (9, 10). In comparison with conventional fan-beam or spiral-scan geometries, cone-beam geometry has higher efficiency in X-ray use, inherent quickness in volumetric data acquisition, and potential for reducing the cost of CT. The cone beam technique, on the other hand, requires only a single scan to capture the entire object with a cone of X-rays. Thus, the time required to acquire a single cone-beam projection is the same as that required by a single fan-beam projection. The quality of the reconstructed CT image depends significantly on the quality of the data that is acquired during the scan. The characteristics of the two-dimensional detector used by a cone beam CT scanner affect therefore directly the quality of the CT scan (Fig. 2).
Its high spatial resolution, smaller size and lower cost has made CBCT a natural fit dentomaxillofacial imaging. The image obtained consists of an image on a 2D axial volume acquisition. Moreover, the acquisition of a volume can not only aimed at reconstruction of the arch and the like but rework similpanoramich and Telecrane. In this way, with only one purchase you can get three types of tests (reconstruction of the arches, and ortopanoramic - Telecrane) saving the administered dose, time and obviously cost savings (11–13).
Technical limits of the cone beam CT is a revised version of the images in real time, which requires at least half minute of acquisition. The processing time of images on the dedicated console appears comparable for the two techniques.
The primary axial reconstructions require a few seconds. The secondary reconstructions are performed in real time by an operator or directly on a dedicated workstation.
ASIR (Adaptive Statistical Iterative Reconstruction)
ASIR is a revolutionary patented image reconstruction algorithm (GE medical systems) that allows to overcome the concept of reconstruction back projecton that, starting from the birth of CT technology, has characterized the principles of operation. At the base of Asir is the concept of dualism in which the trend-particle physics studies the waves elettromagnetic including X-rays, the principle is that not all the phenomena of the waves Electromagnetic you can think of a photon as a point particle, some phenomena can be explained only by modeling the photon as a wave, which is why we say that electromagnetic waves living in wave-particle duality (19).
With ASIR we introduce the concept of a statistical measure of the signal produced by the photon system X detection, exceeding the simple measure of the attenuation in Hounsfield units of the photon after passing through the patient’s anatomy comes on the detector, which is why you move from a mathematical model for solving systems of linear equations a mathematical model heuristic and iterative, the operator can thus determine the value of standard deviation, that may cause the opening of the bell curve that you want. Please use as a basis for statistical model.
The benefit brought by Asir, comparable to that obtained with the generation before subsidy-driven 64-slice CT and demonstrated in numerous scientific studies can be summarized in:
- increase of 40% of low-resolution contrast
- reduce the Xray-dose of approximately 50%
- reduction in the number of artifacts
- percentage applicable to all anatomic areas and all types of examination (Fig. 3).
So reducing the dose of approximately 50% you get images with the same resolution as the higher KV.
Results
80 KV protocols compared with 100 KV protocols resulted in reduced total radiation dose without relevant loss of contrast and resolution of images. 80 KV protocols allow good quality of images for orthodontics diagnosis and treatment.
At the Department of Diagnostic Imaging of the University of Rome “Tor Vergata”, the standard setting of the equipment was modified for use in orthodontic diagnosis. The choice of the lower dose of 80 KV was made as a necessary compromise to reach a good diagnostic accuracy. According to our experiences in the 80 KV MSCT scanning we obtained a mean CTDI of 6.29 mGy and a total scan time of 19 sec. In the 100 KV scanning we obtained a mean of CTDI of 13.22 mGy and a total scan time of 20 sec. In the CBCT we obtained a mean of CTDI of 1.04mGy and a scan total scan time of 29 sec.
The 80 KV scanning has a significantly lower affective radiation dose compared to the 100 KV scanning ( p< 0.05). The CBCT scanning has a significantly lower affective radiation dose compared to the 80 KV MSCT scanning ( p< 0.05) (1).
No significant differents was found for image reconstruction time between 100KV and 80KV protocols. CBCT reconstruction time was least twice than MSCT.
In the retroanalysis, the images of all patients were used for comparing the protocols in terms of image quality. The mean score for 80 KV scanning images was 1.1 +/− 0.5, 1.3 +/− 0.5 for dose obtained by 100 KV scanning and 1 +/− 0.5. The median image quality was 2 (good) for MSCT both protocols and 3 (satisfactory) for CBCT.
Both the protocols of examination introduce the same resolution and the same quality of the image. In the 80 KV scanning (time T0) we obtained a mean DLP of 43.1 mGy per cm, CTDI of 6.29 mGy and a total scan time of 19 sec (Table 1). The 80 KV scanning has a significantly lower effective radiation dose (p< 0.05) compared to 100 KV scanning (time T1) (Tables 2–3) (1).
Table 1.
Time T0 | Time T1 | Change | ||||
---|---|---|---|---|---|---|
| ||||||
Mean | SD | Mean | SD | Mean | P | |
CTDI | 6.231 | 1.945 | 13.282 | 4.104 | −7.051 | .000*** |
DLP | 42.710 | 14.277 | 100.382 | 34.619 | −57.672 | .000*** |
Image quality | 4.18 | .81 | 4.41 | .80 | −0.23 | .417 |
p<0.05 Statistically significant
p<0.01 Statistically significant
p<0.001 Statistically significant ns = Non statistically significant
Table 2.
Patient | CTDI mGy | DLP | Image Quality |
---|---|---|---|
1 | 2.64 | 37.97 | Satisfactory 3 |
2 | 5.44 | 38.08 | Good 4 |
3 | 3.47 | 37.97 | Moderate 2 |
4 | 8.13 | 69.49 | Excellent 5 |
5 | 4.06 | 26.41 | Good 4 |
6 | 8.07 | 38.03 | Excellent 5 |
7 | 7.26 | 35.02 | Excellent 5 |
8 | 8.07 | 42.07 | Excellent 5 |
9 | 7.26 | 39.41 | Good 4 |
10 | 5.41 | 37.91 | Good 4 |
11 | 7.08 | 27.07 | Good 4 |
12 | 3.47 | 37.97 | Satisfactory 3 |
13 | 5.24 | 29.95 | Good 4 |
14 | 6.08 | 33.54 | Good 4 |
15 | 9.08 | 68.08 | Excellent 5 |
16 | 7.16 | 69.49 | Good 4 |
17 | 8.01 | 57.61 | Good 4 |
Media | 6.231 | 42.710 | 4.18 |
Sum | 105.93 | 726.07 | 69 |
Table 3.
Patient | CTDI mGy | DLP | Image quality |
---|---|---|---|
1 | 9.5 | 77,45 | Satisfactory 3 |
2 | 16.14 | 125,84 | Excellent 5 |
3 | 14.52 | 100,6 | Excellent 5 |
4 | 17.01 | 171,24 | Excellent 5 |
5 | 12.10 | 91,34 | Good 4 |
6 | 9.08 | 59,42 | Excellent 5 |
7 | 19.36 | 146,16 | Excellent 5 |
8 | 16.94 | 136,37 | Excellent 5 |
9 | 14.52 | 96,94 | Excellent 5 |
10 | 9.03 | 65,07 | Satisfactory 3 |
11 | 12.10 | 95,91 | Good 4 |
12 | 8.07 | 72,04 | Satisfactory 3 |
13 | 10.09 | 61,03 | Excellent 5 |
14 | 9.08 | 63,54 | Good 4 |
15 | 10.09 | 86,24 | Good 4 |
16 | 21.16 | 105,8 | Excellent 5 |
17 | 17.01 | 151,5 | Excellent 5 |
Media | 13.282 | 100.381 | 4.41 |
Sum | 225.8 | 1706.49 | 75 |
Discussion
The data obtained indicate that the parotid glands, the thyroid and the mouth are the anatomical structures most exposed to radiation during Dentascan CT examinations, both of the upper and low jaw bone, in line with what has been reported in previous studies (7, 8).
The images obtained with the 80 KV protocol were good enough for orthodontic diagnosis, so we decided to compare these images not with those obtained with 100 KV, setting typically used in Dentascan examinations, but with the low dose protocol of 100 KV (1, 6, 15,18, 21).
When milliamperes are kept constant, the radiation dose delivered by an X-ray tube increases, as the X-ray energy, that is the kilovoltage, increases. This effect is largely exploited in diagnostic imaging performed on babies or children, where a low radiation dose is required. As X-ray energy ( kilovoltage) decreases, the contrast between the structures crossed by the X-ray beam increases (9–11). This concept proved to be useful in our experience, as it enabled a marked reduction in the dose administered without affecting image quality.
The dose value obtained are significantly lower than the minimum values reported in the literature that contains numerous studies on dosimetry in odontostomatologic radiology, often with comparisons between conventional CT, CBCT and OPT (16, 17, 20).
Even when compared with the doses reported for CBCT, with the obvious approximations linked to the different dosimetric techniques, our results were comparable, and the absorbed dose measured in our study was even lower in some anatomical sites (12, 14).
The measurements taken for each examination all proved to be in the same order of dose, which confirms the reliability of the values obtained. A 50% to ten-fold reduction of the radiation dose was seen in the 80 KV protocol compared with what is reported in the literature. Image quality with the lower kilovoltage remains acceptable both as regards quantitative measurements and the evaluation of bone quality. CBCT, which are characterized by a reasonably low radiation dose, are only indicated in basic implant and trauma diagnostics. Anatomically complex cases or masses requiring more detailed studies can only be analysed with blade-beam CT scanners, which produce higher-quality images (Fig. 4).
In orthodontics we are just beginning to establish the potential diagnostic and therapeutic applications of three dimensional images. Evolution of the technology and software will allow further studies to demonstrate that valuable diagnostic results may be obtained with even lower kilovoltage and tube current in the near future. Future technical developments, such as further increases in axial resolution combined with intelligent dose modulation protocols, may help strengthen the role of CT in the assessment in the orthodontic use (3, 4).
Three-dimensional investigations are necessary in orthodontic daily practice every time conventional X-ray techniques are not enough, such as impacted maxillary canines or incisors, supernumeraries teeth, trauma and orthopaedic treatment.
MSCT ASIR allows new scanning modes to set new dose standards and clinical cases showing 40% average dose reduction vs MSCT. It improve image IQ on very low dose exams and allows to decrease dose while improving image quality. With specific clinical indications whereby high-contrast lesions are surrounded by low-contrast structures, dedicated low-dose protocols have been routinely used in an effort to minimize radiation dose. For example, several studies have suggested that low-dose CT may be adequate for renal calculus evaluation, CT colonography, and pulmonary CT angiography. It can be satisfactory applied to the dento-maxillo-facial district. Although satisfactory for the dedicated evaluation of the intended anatomy, these protocols were implemented with the realization that image quality would be significantly reduced outside the area of interest.
Although desirable for all patients, dose reduction is of particularly high priority for young patients (5), patients who are pregnant, and patients requiring multiple follow-up or multiphase studies. However, low-dose imaging with ASIR reconstructions appears to have potential in displaying similar or better overall image quality compared with routine-dose. New imaging methods will replace the way we look at a variety of common diagnostic and treatment issues in daily orthodontic practice.
References
- 1.Ballanti F, Lione R, Fiaschetti V, Fanucci E, Cozza P. Low-dose CT protocol for orthodontic diagnosis. Paediatric Dentistry. 2008;9/2 [PubMed] [Google Scholar]
- 2.Bamgbose BO, Adeyemo WL, Ladeinde AL, Ogunlewe MO. Cone Beam computed tomography (CBCT): the new vista in oral and maxillofacial imaging. Nig Q J Hosp Med. 2008 Jan-Mar;18(1):32–5. doi: 10.4314/nqjhm.v18i1.44955. [DOI] [PubMed] [Google Scholar]
- 3.Chaushu S, Chaushu G, Becker A. The role of digital volume tomography in the imaging of impacted teeth. World J Orthod. 2004;5:120–132. [PubMed] [Google Scholar]
- 4.Deguchi T, Nasu M, Murakami K, Yabuuchi T, Kamioka H, Takano-Yakamoto T. Quanitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. Am J Orthod Dentofacial Orthop. 2006;129:721.e7–721-12. doi: 10.1016/j.ajodo.2006.02.026. [DOI] [PubMed] [Google Scholar]
- 5.Espelid I, Mejàre I, Weerheijm K. EAPD guidelines for use of radiographs in children. Eur J Paed Dent. 2003 Mar;4(1):40–8. [PubMed] [Google Scholar]
- 6.Fanucci E, Leporace M, Di Costanzo G, Fiaschetti V, Simonetti G. Multidetector CT and dentascan software: dosimetric evaluation and technique improvement. Radiol Med. 2006;111:130–8. doi: 10.1007/s11547-006-0014-7. [DOI] [PubMed] [Google Scholar]
- 7.Hajeer MY, Millet DT, Ayoub AF, Siebert JP. Current product and practices. Applications of 3D imaging in orthodontics: Part I. Journal of orthodontics. 2004;31:62–70. doi: 10.1179/146531204225011346. [DOI] [PubMed] [Google Scholar]
- 8.Hajeer MY, Millet DT, Ayoub AF, Siebert JP. Current product and practices. Applications of 3D imaging in orthodontics: Part II. Journal of orthodontics. 2004;31:154–162. doi: 10.1179/146531204225020472. [DOI] [PubMed] [Google Scholar]
- 9.Hechler SL. Cone-beam CT: application in orthodontics. Dent Clin North Am. 2008 Oct;52(4):809–23. doi: 10.1016/j.cden.2008.05.001. [DOI] [PubMed] [Google Scholar]
- 10.Katsumata A, Hirukawa A, Noujeim M, Okumura S, Naitoh M, Fujishita M, Ariji E, Langlais RP. Image artifact in dental cone-beam CT. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 May;101(5):652–7. doi: 10.1016/j.tripleo.2005.07.027. Epub 2005 Nov. [DOI] [PubMed] [Google Scholar]
- 11.Kau CH, Boži M, English J, Lee R, Bussa H, Ellis RK. Cone-beam computed tomography of the maxillofacial region-an update. Int J Med Robot. 2009 Sep 23; doi: 10.1002/rcs.279. [DOI] [PubMed] [Google Scholar]
- 12.Lecomber AR, Yoneyama Y, Lovelock DJ, et al. Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and computer tomography. Dentomaxillofacial Radiol. 2001;30:255–259. doi: 10.1038/sj/dmfr/4600627. [DOI] [PubMed] [Google Scholar]
- 13.Ludlow JB, Davies-Ludlow LE, Brooks SL. Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol. 2003;32:229–234. doi: 10.1259/dmfr/26310390. [DOI] [PubMed] [Google Scholar]
- 14.Macchi A, Carrafiello G, Cacciafesta V, Norcini A. Three-dimensional digital modeling and setup. Am J Orthod Dentofacial Orthop. 2006;129(5):605–610. doi: 10.1016/j.ajodo.2006.01.010. [DOI] [PubMed] [Google Scholar]
- 15.Matarese G, Portelli M, Mazza M, Militi A, Nucera R, Gatto E, Cordasco G. Evaluation of skin dose in a low dose spiral CT protocol. Eur J Paed Dent. 2006;2:77–80. [PubMed] [Google Scholar]
- 16.Miller AJ, Maki K, Hatcher DC. New diagnostic tools in orthodontics. Am J Orthod Dentofacial Orthop. 2004;126:395–396. doi: 10.1016/j.ajodo.2004.04.014. [DOI] [PubMed] [Google Scholar]
- 17.Redmond WR, Huang J, Bumann A, Mah J. Three-dimensional radiographic analysis in orthodontics. JCO. 2005;39(2):421–428. [PubMed] [Google Scholar]
- 18.Scheck RJ, Coppenrath EM, Kellner MW, et al. Radiation dose and image quality in spiral CT: multicentre evaluation at 6 institutions. Br J Radiol. 2002;75:140–150. doi: 10.1259/bjr.71.847.9771384. [DOI] [PubMed] [Google Scholar]
- 19.Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol. 2010 Jan;194(1):191–9. doi: 10.2214/AJR.09.2953. [DOI] [PubMed] [Google Scholar]
- 20.Suomalainen A, Kiljunen T, Käser Y, Peltola J, Kortesniemi M. Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac Radiol. 2009 Sep;38(6):367–78. doi: 10.1259/dmfr/15779208. [DOI] [PubMed] [Google Scholar]
- 21.Ziegler CM, Woertche R, Brief J, Hassfeld S. Clinical indications for Digital Volume Tomography in oral and maxillofacial surgery. Dentomaxillofac Radiol. 2002;31:126–130. doi: 10.1038/sj/dmfr/4600680. [DOI] [PubMed] [Google Scholar]