Abstract
Objective
Organophosphorus (OP) pesticides are considered hazardous substances because of their high toxicity to nontarget species and their persistence in the environment and agricultural products. Therefore, it is important to develop a rapid, sensitive, and economical method for detecting OP pesticides and their residues in food and the environment.
Methods
A broad, selective monoclonal antibody (MAb) for organophosphorus pesticides was produced. Based on the MAb, an enzyme linked immunosorbent assay (ELISA) and an immunochromatography assay (ICA) for detecting OP pesticides in different agricultural products were developed using a binding inhibition format on microtiter plates and a membrane strip, respectively.
Results
Under the optimized conditions, the IC50 values of the ELISA ranged from 3.7 to 162.2 ng mL–1 for the 8 OP pesticides. The matrix interferences of Apple, Chinese cabbage, and greengrocery were removed by 40-fold dilution, the recoveries from spiked samples ranged from 79.1% to 118.1%. The IC50 values of ICA for the 8 OP pesticides ranged from 11.8 to 470.4 ng mL−1. The matrix interference was removed from the Chinese cabbage and Apple samples with 5-fold dilution, and the interference was removed from the greengrocery samples with 20-fold dilution. The recoveries from the spiked samples ranged between 70.6 and 131.9%. The established ELISA and ICA were specific selectivity for the 8 OP pesticides.
Conclusions
The established ELISA is a sensitive screening method for the detection of OP pesticides, but the ELISA detection method depends on a laboratory platform and requires a relative long assay time and several steps operation. The established ICA is very useful as a screening method for the quantitative, semi-quantitative or qualitative detection of OP pesticides in agricultural products, and it has advantages over ELISA methods with regard to factors such as the testing procedure, testing time, and matrix interferences, among others.
Introduction
Organophosphorus (OP) pesticides are widely used in agriculture for sucking and biting insect pest control, including fruit flies, stem borers, mosquitoes, and Eurygaster cereal bugs. However, OP pesticides are considered hazardous substances because of their high toxicity to nontarget species and their persistence in the environment [1]. Additionally, there is increasing concern over food and environmental contamination resulting from the overuse of pesticides [2]. Therefore, it is important to develop a rapid, sensitive, and economical method for detecting OP pesticides and their residues in food and the environment.
Currently, instrument-based methods, such as gas chromatography and high-performance liquid chromatography, are the most commonly used techniques for detecting OP pesticides in different samples [3]–[6]. However, these commonly used methods require expensive equipment and are only applicable in laboratory settings. In comparison, immunoassays have received considerable attention as a simple, sensitive, cost-effective tool for high-throughput screening analyses in pesticide monitoring programs [7].
Immunoassays are often developed to recognize a single analyte with high specificity [8], [9], and they can also be applied for detecting various related compounds in a single test [10]–[12]. These latter immunoassays have been termed broad-selective, broad-specific, class-specific, or multi-analyte assays. Unlike a single-analyte assay, a multi-analyte assay can be used to detect the total quantity of pesticides [13]. Multi-analyte assays can also be used for pesticide monitoring before chromatographic analyses; if the total quantity of pesticides in a particular sample is less than the maximum limit, then that sample does not require further analysis [14]. To date, several investigators have developed multi-analyte enzyme-linked immunosorbent assays (ELISAs) for the detection of OP pesticides [14]–[16]. However, the ELISA detection method depends on a laboratory platform and requires a relative long assay time.
Colloidal gold has been introduced into immunochemistry and their particles could replace the enzyme to label antibody, and deposited into the conjugate pad. The capture line (test line) is hapten conjugated to a carrier protein immobilized on the membrane. Analytes in samples will compete with antigen immobilized on the membrane to bind to the colloidal gold labeled antibody. The more analytes present in the sample, the more effectively it will be able to block the capture of colloidal gold labeled antibodies. An increase in the amount of analytes in samples will result in a decrease in signal in the test line zone. Therefore, colloidal gold-conjugated immunochromatographic assays (ICAs) have been developed as a rapid and simple test for detecting chemicals in non-laboratory sites, and they have been increasingly applied in various research fields [17]–[22], including for the detection of pesticides [23]–[25]. However, few multi-analyte ICAs have been developed for the detection of OP pesticides in agricultural products.
In this study, a sensitive monoclonal antibody (C8/D3) that can recognize eight OPs was produced and used to develop an ELISA and a colloidal gold ICA for the determination of OP pesticides in agricultural products. The ELSIA showed higher sensitivity than ICA, but the ICA has advantages over ELISA methods with regard to factors such as the testing procedure, testing time, and matrix interferences, among others.
Materials and Methods
Reagents and Equipment
Parathion-methyl, chlorpyrifos-methyl, azinphos-methyl, dimethoate, fenitrooxon, EPN, paraoxon-ethyl, paraoxon-methyl, dicapthon, cyanophos, and famphur were purchased from Dr. Ehrensorfer (Germany). Other pesticide standards were provided by the Jiangsu Pesticide Research Institute (China). Trisodium citrate and methanol were of analytical grade. Tetramethylbenzidine (TMB), Triton x-100 and PEG 20000 were purchased from Sigma-Aldrich (USA). Horseradish peroxidase (HRP)-labeled goat anti-mouse IgG was obtained from Sino-American Biotechnology Co. (Dalian, China). Rabbit anti-mouse IgG was obtained from the Henan Provincial Key Laboratory for Animal Immunology (Henan Academy of Agriculture Sciences, China). Nitrocellulose (NC) membranes, glass-fiber conjugate pads, sample pads, and absorbent pads were produced by Millipore (USA). The BALB/c mice were obtained from the Center of Comparative Medicine of Yangzhou University (Yangzhou, China). All animals used in this study, and animal experiments, were approved by Department of Science and Technology of Jiangsu Province. The license number was SYXK (SU) 2010-0005.
The XYZ-3000 dispensing platform, guillotine cutter (Model: CM 4000) and TSR3000 membrane strip reader were supplied by BioDot (Irvine, CA, USA). Test lines were scanned using a BioDot TSR3000 Membrane Strip Reader (BioDot, CA, USA). The centrifuge (3K30) was purchased from the Sigma-Aldrich Corporation (St. Louis, USA). The protein A/G Spin Kit was purchased from the Thermo Electron Corporation (USA). The CELLine CL 350 was supplied by the Integar Biosciences Corporation (Switzerland). The transmission electron microscope (H-7650) was purchased from the Hitachi Corporation (Japan). The ultraviolet spectrophotometer (ND-1000) was purchased from the NanoDrop Corporation (USA).
Preparation of Antigens
Figure 1 presents the structures of the 12 haptens used in the current study. The preparation of the haptens (haptens 1 to 12) and the antigens (haptens 1- to 12-OVA) has been described previously [26]–[28]. According to the active ester methodology described previously [29], hapten 1 was conjugated with BSA to generate the immunogen (hapten 1-BSA).
Preparation of the Monoclonal Antibody (MAb)
Five 6-week-old BALB/c female mice were immunized with the H1-BSA conjugate; the immunizing strategy and the cell fusion procedures were performed as described previously [30]. When hybridomas were grown to approximately 30% confluence in culture plate wells, the culture supernatants were screened using the coating antigen hapten 1-OVA for the presence of antibodies that could recognize parathion-methyl, parathion, and fenitrothion. These three OP pesticides were chosen because they have structures more similar to the immunizing hapten than other OP pesticides and because previous studies [2], [10], [31]–[33] have demonstrated that cross-reactivity (CR) often occurs among these three pesticides. The selected hybridomas were subcloned using limiting dilution. Stable antibody-producing clones were efficiently cultivated by CELLine CL 350 according to the manufacturer’s protocol. The MAb was purified using the protein A/G Spin Kit according to the manufacturer’s protocol.
Indirect Competitive ELISA Protocol
The concentrations of antibody and coating antigen were optimized by checkerboard titration. The ic-ELISA was performed as follow. Microtiter plates were coated with optimized concentrations of hapten-OVA in carbonate–bicarbonate buffer (50 mM, pH 9.6) by incubating for 2 h. Plates were then blocked by incubating with 1% gelatin in PBS (120 µL well–1) for 1.5 h. Aliquots of 25 µL well–1 of analyte dissolved in working solution and 25 µL well–1 of the supernatant (cultivated by CELLine CL 350) diluted with working solution were added to the blocked plate. After incubating for 1 h, 50 µL well–1 of diluted (1/20,000) goat anti-mouse IgG-HRP was added to the plate. The mixture was incubated for 1 h, followed by the addition of 50 µL well–1 of a TMB solution (100 µL of 1% hydrogen peroxide and 400 µL of 0.6% TMB in dimethyl sulfoxide added to 25 mL of citrate acetate buffer, pH 5.5). After incubating for 10 min, the reaction was stopped by adding 25 µL of 2 M H2SO4 and the solution absorbance was measured at 450 nm. All incubations were performed at 37°C and the plates were washed five times with PBST (10 mM PBS containing 0.05% Tween 20, pH 7.4) after each incubation, unless otherwise specified.
Optimization of the ELSIA
The selection of coating antigens was performed by indirect ELISA and Indirect competitive ELISA. First, Twelve antigens (hapten 1 to 12-OVA) that were able to conjugate with the MAb were screened out by indirect ELISA; the indirect ELISA was run at the same time as the indirect competitive ELISA without addition of analyte. Second, the coating antigens that were screened out in the first step were selected by indirect competitive ELISA, and then the best coating antigen was confirmed based on the sensitivity of the ELISA.
The working solutions, which contained a series of pH values (5.0 ∼ 9.6) and ionic strengths (0 ∼ 3.2 M) were used to dilute the parathion-methyl, parathion, and fenitrothion standards and were tested using the established ELISA. The tested working solution that resulted in the best sensitivity for the ELISA was selected as the optimized working solution.
Preparation of Colloidal Gold
Several investigators have reported that the average particle diameter of 40 nm colloidal gold was the optimal particle size for most diagnostic applications because of its trade-off between the required visibility and steric hindrance [34]–[36]. Therefore, in this study, nanometer colloidal gold with an average particle diameter of 40 nm was prepared according to the method described by Frens [37]. Briefly, under reflux conditions, a 100 mL solution of 0.01% gold chloride was heated to the boiling point, and 1.1 mL of 1% trisodium citrate solution was rapidly added with constant stirring. The solution was boiled for another 5 min after the color of the mixture changed to a brilliant wine red. After cooling at room temperature, the average particle diameter was evaluating using a transmission electron microscope and stored at 4°C for further study.
Conjugation of MAb with Colloidal Gold
The MAb-gold complex was prepared according to the previously described method [38]. The pH value of the prepared colloidal gold solution was adjusted to 8.2 using 0.2 M K2CO3. 125 µL of pH adjusted colloidal gold solution was quickly added into 30 µL of serial concentrations (200, 150, 100, 50, 25, and 12.5 µg mL−1) of MAb. After 5 min, 125 µL of a 10% NaCl solution was added to the mixture. The OD528 (optical density at 528 nm) of each solution was measured using a UV-spectrophotometer in two states (stage I: before the NaCl solution was added; stage II: 60 min after the NaCl solution was added), and the experiment was performed in triplicate. In the next step, the mean decreased value between the two stages was used as the y axis, and the corresponding MAb concentration was used as the x axis to generate a curve. The point where the curve first appeared to be relatively parallel to the x axis was selected as the minimum quantity of MAb required to protect the gold solution from flocculation.
To conjugate MAb with colloidal gold, an amount that corresponded to 10% more than the minimum amount of MAb determined to protect the gold solution from flocculation was used. Then, the MAb solution was rapidly added to 20 mL of the colloidal gold solution (pH 8.2) with rapid stirring and incubated for 1 h. Subsequently, the mixture solution was stabilized using a 10% BSA solution (final BSA concentration was 1%) and was stirred for another 5 min. After incubation for another 1 h, the solution was centrifuged (10,000 rpm) at 4°C for 25 min, and the colorless supernatant was carefully aspirated and discarded. Subsequently, the loose sediment of the gold-labeled MAb was re-suspended using 20 mL of a 2% BSA solution (containing 0.01 M sodium borate), followed by another centrifugation (10,000 rpm at 4°C for 25 min). This procedure was repeated twice to ensure cleanup of the free (unlabeled) MAb. Finally, the loose sediment of the gold-labeled MAb was re-suspended using 4 mL of a TB solution (containing 3% BSA, 3% sucrose, 0.01 M sodium borate and 0.05% sodium azide) and stored at 4°C until needed.
Immobilization of the Reagents
Four percent BSA was dispensed onto the glass-fiber conjugate pad using the XYZ-3000 dispensing platform, and the pad was dried at 42°C for 50 min. Subsequently, the gold-labeled MAb was dispensed onto the aforementioned glass-fiber conjugate pad using the XYZ-3000 dispensing platform, and then the pad containing gold-labeled MAb was dried again at 42°C for another 50 min. After this step, the MAb-gold conjugate pad was stored in a vacuum drying oven until needed.
The antigens and rabbit anti-mouse IgG were dispensed onto the NC membrane using the XYZ-3000 dispensing platform. The antigens were dispensed around the bottom of the NC membrane, which were used as the test line (T line), and the rabbit anti-mouse IgG was dispensed at the upper position of the NC membrane, which is used as the control line (C line). The distance between the T and the C lines was approximately 5 mm. Subsequently, the NC membrane was dried for 2 days and stored at room temperature.
Development of the ICA Test Strip
The ICA test strip was composed of the NC membrane, MAb-gold-conjugated pad, sample pad and absorbent pad, which were pasted onto an adhesive plastic backing (Figure 2). The NC membrane was pasted at the center of the backing plate. The absorbent pad was pasted by over-crossing 1 mm with the upper of the NC membrane. The mAb-gold conjugate pad was pasted by over-crossing 1 mm with the bottom of the NC membrane. The sample pad was pasted by over-crossing 2 mm with the bottom of the mAb-gold conjugate pad. Finally, the entire assembled ICA test strip was cut lengthways into 4.08 mm wide strips using a guillotine cutter.
Test Procedure of ICA
One hundred and fifty microliters of the sample was added to the sample pad of the ICA test strip. After 5 min, the result could be evaluated by the naked eye or scanned using a membrane strip reader. When the sample solution contained the analyte, the analyte and the gold-labeled MAb formed an antigen-gold-labeled MAb complex, and the T line exhibited no color or a weaker color than the C line, which indicated that the test result was positive or weakly positive, respectively. If the sample did not contain the analyte, the T line will exhibited a similar color to the C line, which indicated that the test result was negative. The test was ineffective if no line was observed in the control line region (Figure 2).
Optimization of the ICA Test Strip
Twelve antigens (hapten 1 to 12-OVA) were dispensed around the bottom of the NC membrane (nominal pore size: 8 µm), which was used as the T line. The antigens that conjugated with the gold-labeled MAb were first screened out, and then the best antigen was confirmed based on the sensitivity of the ICA. Four types of NC membranes (Nominal pore size: 6, 8, 10, and 12 µm) were used in this study, and the best type of NC membrane was selected based on the sensitivity and test time.
Optimization of the Assay Conditions
The working solutions, which contained a series of pH values (5.0 ∼ 9.6), ionic strengths (0 ∼ 3.2 M), Triton x-100 (0 ∼ 1%) and polyethylene glycol (PEG) 20000 (0 ∼ 2%), were used to dilute the parathion-methyl standard and were tested using the established ICA test strip. The tested working solution that resulted in the best sensitivity for the ICA test strip was selected as the optimized working solution. The optimized working solution was allowed to migrate on the test strip, and the T line was scanned using the strip reader at different times (3, 5, 10, 20, 30, 40 and 50 min after the working solution was added). The best reading time was determined based on the stability of the G/Peak – ROD value.
Sensitivity of the ELISA and ICA Test Strip
A series of concentrations (ELISA: 0.1∼10, 000 ng mL−1; ICA: 0.5 ∼ 16, 384 ng mL−1) of 23 analyte standards were prepared using the optimized working solution and tested using the ELISA and ICA test strip. In ICA study, the results were scanned by using a membrane strip reader, repeated three times. The average of the scanned values (ELISA: OD450; ICA: G/Peak - ROD) and the concentration of the analytes were used to establish the inhibition curves. Subsequently, the inhibition curves were converted into calibration curves. In the calibration curves, the binding (B/B0) was described as the vertical coordinate axis (y), and the logarithms of the analyte concentrations were designated as the abscissa axis (x). From these curves, the regression equations were obtained, and the IC50 (concentration at which binding of the antibody to the coating antigen on test line is inhibited by 50%) and limit of detection (LOD, IC10) were calculated.
Cross-reactivity (CR) Studies
To evaluate the selectivity of the ELISA and ICA, 23 OP pesticides were evaluated using the established ELISA and ICA. The cross-reactivity, CR, was calculated as follows: CR = [IC50 (parathion-methyl)/IC50 (compound)] × 100%. Here, the CR of parathion-methyl was defined as 100%.
Accuracy (Analysis of Spiked Agricultural Samples)
Three different agricultural samples (Chinese cabbage, Apple, and Greengrocery) were chosen to evaluate the performance of the multi-analyte ELISA and ICA. Chinese cabbage, Apple, and Greengrocery were bought from local markets, before the spiking and recovery studies, each test sample was verified to not contain parathion-methyl, parathion, and fenitrothion by gas chromatography. All of the samples were spiked with known concentrations of parathion-methyl, parathion, and fenitrothion in methanol.
All samples were cut into pieces and homogenized. The spiked samples were thoroughly mixed and allowed to stand at room temperature for 1 h. The sample pretreatment procedure was as follows. Acetonitrile extraction solvent (10 mL) was added to 10 g of homogenized sample, and the mixture was oscillated for 1 min with a vortex mixer. Then, 1.5 g of sodium chloride and 4 g of sodium sulfate were added, followed by oscillation for 3 min. After centrifugation (5 min, 4000 rpm), 1 mL of the supernatant was transferred and evaporated at 40°C under a nitrogen stream. The dry residue was redissolved and diluted using the optimized working solution before analysis (ELISA and ICA).
Results and Discussion
Production of MAb
A hybridoma cell line (C8/D3) that stably produced a monoclonal antibody (MAb) against parathion-methyl, parathion, and fenitrothion was selected from the indirect competitive ELISA and stored in China Center for Type Culture Collection, the certificate number is C201279.
The Optimized ELISA
Four coating antigens (haptens 1, 2, 8, 9), each containing a nitro-aromatic group, were selected by indirect ELISA for their ability to conjugate with the MAb (C8/D3). These results demonstrated that the structure of the nitro-aromatic group, and its position relative to the hapten, has a very important role in the hapten’s reaction with the MAb. The sensitivity of the four antigens decreased in the following order: hapten 9> hapten 8> hapten 2> hapten 1 (Figure S1). Hapten 9-OVA was superior to the other antigens in terms of sensitivity when used as a coating antigen. Based on the IC50 values of parathion-methyl, parathion, and fenitrothion, the optimal ironic strength and pH of working solution was 0.4 M and 9.0, respectively (Figure S2 and S3). And 5% (V/V) methanol was used as cosolvent to improve analyte solubility. In conclusion, the optimized working solution was 0.4 M, pH 9.0 PBS buffer (containing 5% MeOH).
Optimal Amount of mAb for Colloidal Gold
The minimal amount of MAb was determined using a salt-induced precipitation method, and the result is presented in Figure 3. It is clear that as the MAb concentration increased, the corresponding average OD-value of the absorbance in two stages exhibited an evident decrease. Furthermore, when the MAb concentration exceeded 100 µg mL−1, the curve became approximately flat. Therefore, the optimal MAb concentration conjugated to colloidal gold (40 nm) solution was determined to be 110 µg mL−1.
The Optimized ICA Test Strip
Four antigens (hapten 1, 2, 8, 9-OVA) were selected based on their ability to conjugate with the gold-labeled MAb (C8/D3). The sensitivity of the four antigens decreased in the following order: hapten 1-OVA>hapten 2-OVA>hapten 8-OVA>hapten 9-OVA (Table S1). Hapten 9-OVA was superior to the other antigens in terms of sensitivity when dispensed onto the NC membrane as a T line. These results suggest that using an appropriate antigen can enhance the sensitivity of the multi-analyte ICA, and the results are consistent with the ELISA measurements.
The NC membrane type is one of the most significant factors that influence the protein binding levels, sensitivity and testing time of an ICA test strip. In this study, four types of NC membranes (nominal pore sizes: 6, 8, 10, and 12 µm) were used to produce the ICA test strips. As the pore size increased, the testing time decreased and the sensitivity increased, which are good factors for a rapid test. However, the binding level decreased with increasing pore size, and there was an uneven line in the T line region for the 12 µm pore size NC membrane (Table S2). Therefore, the 10 µm pore size NC membrane was selected as the best membrane and used in the future study.
As a rapid detection tool, the result of the negative test should show visible and similar two lines (T and C line) on NC membrane. In this study, the H9-OVA was prepared in normal saline by serial dilutions ranging from 0.075 mg mL−1 to 1.2 mg mL−1. When the concentration of H9-OVA was equal or greater than 0.3 mg mL−1, the T line would show visible signal, but the high concentration (greater than 0.3 mg mL−1) of H9-OVA was used, the more parathion-methyl was needed in the sample to complete with H9-OVA. So, in this study, 0.3 mg mL−1 of H9-OVA solution was used to dispense on the NC membrane, and used as the test line. And the Rabbit anti-mouse IgG was prepared in normal saline solution by serial dilutions ranging from 0.2 mg mL−1 to 1.6 mg mL−1. When the concentration of Rabbit anti-mouse IgG was 0.8 mg mL−1, the C line would show similarly strength signal with T line. Accordingly, 0.8 mg mL−1 of Rabbit anti-mouse IgG solution was used to dispense on the NC membrane, and described as the control line. In order to make the two lines had the similar and easy observation width displayed on ICA strip, the dosage of H9-OVA solution and Rabbit anti-mouse IgG solution were 1.0 µL cm−1 in this study. Finally, 5 µL cm−1 gold-labeled MAb was dispensed onto the glass-fiber conjugate pad to ensure that the two lines presented a visible signal.
The Optimized Assay Conditions
ICAs are based on antigen-antibody interactions, and the pH and ionic strength are additional factors that influence on the equilibrium constant [39]. Therefore, the pH and ionic strength of the working solution were optimized (Table S3 and S4). With the ionic strength of working solution increased, the sensitivity of ICA was improved. But when the ionic strength of the working solution is greater than 0.4 M, the gold-labeled mAb was difficultly redissolved and moved through the NC membrane with the sample solution. Therefore, the 0.4 M NaCl, pH 9.0 borate buffer was selected for the ICA because it exhibited superior sensitivity. Triton x-100 and PEG 20000 mainly affected the sensitivity, reactant flow, background color level, sharpness, and intensity of the test line. Therefore, these compounds were examined to improve the test performance; the optimized concentrations of Triton x-100 and PEG 20000 were 0.02% and 0.2%, respectively (Table S5 and S6). In conclusion, the optimized working solution was a 0.4 M NaCl, pH 9.0 borate buffer (containing 5% methanol, 0.02% Triton x-100 and 0.2% PEG 20000).
As shown in Figure 4, the G/Peak – ROD values increased from 0 to 5 min following the addition of the buffer. After 5 min, the G/Peak – ROD value stabilized, but after 30 min, the test strip began to desiccate and the G/Peak – ROD value began to increase. Therefore, the optimal reading time is 5 to 30 min after adding the samples.
Sensitivity of the ELISA and ICA
In total, 23 OP pesticides were evaluated using the ICA and ELISA methods, and the results are presented in Table 1. The IC50 values ranged from 11.8 to 470.4 ng mL−1 for the eight OP pesticides (parathion-methyl, parathion, fenitrothion, EPN, cyanophos, paraoxon-methyl, paraoxon-ethyl, and fenitrooxon), and the LOD (IC10) ranged from 2.7 to 100.2 ng mL−1 for the eight OP pesticides in the ICA test. The IC50 values of the ELISA ranged from 3.7 to 162.2 ng mL–1 for the eight OP pesticides, and the LOD (IC10) ranged from 0.6 to 14.2 ng mL−1. Furthermore, the average IC50 values from the ICA and ELISA tests were 183.9 and 57.1 ng mL–1, respectively. Figure 5 shows the immunochromatographic detection of parathion-methyl. These results indicate that the sensitivity of the newly developed ICA is lower than that of the ELISA, but this phenomenon was usually observed during the strip test [36].
Table 1. IC50 values and LOD (IC10) of a set of analogs related to parathion-methyl by multi-analyte ICA and ELISA.
ICA | ELISA | ||||
Compound | IC50 (ng mL−1) | IC10 (ng mL−1) | IC50 (ng mL−1) | IC10 (ng mL−1) | |
1 | Parathion-methyl | 11.8 | 2.7 | 3.7 | 0.6 |
2 | Parathion | 20.8 | 4.2 | 11.3 | 1.9 |
3 | Fenitrothion | 47.4 | 10.6 | 8.9 | 0.9 |
4 | Cyanophos | 71.1 | 15.7 | 20.3 | 2.2 |
5 | EPN | 58.6 | 13.3 | 18.3 | 2.8 |
6 | Paraoxon-methyl | 439.2 | 92.3 | 118.7 | 6.9 |
7 | Paraoxon-ethyl | 470.4 | 100.2 | 162.2 | 14.2 |
8 | Fenitrooxon | 352.2 | 71.5 | 113.6 | 10.3 |
9 | Dicapthon | 2018.1 | 479.4 | 576.6 | 83.2 |
10 | Famphur | >100000 | ni | 58215.7 | 9991.8 |
11 | Isocarbophos | >100000 | ni | >100000 | ni |
12 | Fenthion | >100000 | ni | >100000 | ni |
13 | Triazophos | >100000 | ni | >100000 | ni |
14 | Chlorpyrifos | >100000 | ni | >100000 | ni |
15 | Chlorpyrifos-methyl | >100000 | ni | >100000 | ni |
16 | Phoxim | >100000 | ni | >100000 | ni |
17 | Malathion | >100000 | ni | >100000 | ni |
18 | Phorate | >100000 | ni | >100000 | ni |
19 | Dimethoate | >100000 | ni | >100000 | ni |
20 | Acephate | >100000 | ni | >100000 | ni |
21 | Dichlorvos | >100000 | ni | >100000 | ni |
22 | Tolclofos-methyl | >100000 | ni | >100000 | ni |
23 | Azinphose-methyl | >100000 | ni | >100000 | ni |
Mean for 1–8 | 183.9 | 38.8 | 57.1 | 5.0 | |
CV (%) for 1–8 | 108.5 | 107.5 | 110.0 | 100.4 |
ni means no inhibition by a pesticide.
Cross-reactivity
The selectivity of the ICA and ELISA were evaluated by measuring the cross-reactivity with the 23 OP pesticides; the CRs for each compound are summarized in Table 2. These results indicated that the OP pesticides with CR are mainly the eight OP pesticides, and the CRs for the ICA were similar to those from the ELISA. Therefore, the developed ELISA and ICA had specific selectivity for the eight OP pesticides.
Table 2. Cross-reactivity (CR) of a set of analogs related to parathion-methyl by multi-analyte ICA and ELISA.
CR (%) | ||
Compound | ICA | ELISA |
Parathion-methyl | 100 | 100 |
Parathion | 56.7 | 32.9 |
Fenitrothion | 24.9 | 41.4 |
Cyanophos | 16.6 | 18.2 |
EPN | 20.1 | 20.2 |
Paraoxon-methyl | 2.7 | 3.1 |
Paraoxon-ethyl | 2.5 | 2.3 |
Fenitrooxon | 3.4 | 3.3 |
Dicapthon | 0.6 | 0.6 |
Famphur | <0.01 | 0.006 |
Isocarbophos | <0.01 | <0.003 |
Fenthion | <0.01 | <0.003 |
Triazophos | <0.01 | <0.003 |
Chlorpyrifos | <0.01 | <0.003 |
Chlorpyrifos-methyl | <0.01 | <0.003 |
Phoxim | <0.01 | <0.003 |
Malathion | <0.01 | <0.003 |
Phorate | <0.01 | <0.003 |
Dimethoate | <0.01 | <0.003 |
Acephate | <0.01 | <0.003 |
Dichlorvos | <0.01 | <0.003 |
Tolclofos-methyl | <0.01 | <0.003 |
Azinphose-methyl | <0.01 | <0.003 |
Analysis of Spiked Samples
One of the most common challenges in developing an immunoassay is matrix interference. Chemical compounds present in samples or sample extracts, such as humic acids, solvents, pigment, and heavy metal ions, may affect the binding between the antibody and analyte, thereby reducing the sensitivity and reliability of the immunoassay and possibly leading to false positive results [40]. Matrix interferences can be reduced in a number of ways, and dilution with buffers is a commonly used procedure [41], [42]. In this study, the effects of different dilution ratios using the optimized working buffer from 1∶1 to 1∶80 (v/v) on reducing the degree of matrix interferences were tested. These results indicated that the G/Peak–ROD values of the standard curves prepared during the matrix dilution were somewhat similar to those in the matrix free buffer when the dilution ratio was increased. During the ICA, the matrix interference was adequately removed from the Chinese cabbage and Apple samples with at least a 5-fold dilution, and the interference was removed from the greengrocery samples with at least a 20-fold dilution. However, during the ELISA, when the matrixes were diluted equal or greater than 40-fold, the standard inhibition curves prepared during the matrix dilution were somewhat similar to those in the matrix free buffer (Figure 6). Therefore, at least a 40-fold dilution was required to adequately remove the matrix interference from the Chinese cabbage, Apple, and greengrocery samples. These results indicated that the ICA has an advantage over ELISA for matrix interferences. Meanwhile, comparison of the matrix interference of Chinese cabbage and Apple with greengrocery samples in the ICA indicates that the greengrocery samples exhibited the strongest matrix interference. These observations can be explained by the fact that greengrocery is a green leafy vegetable that contains an abundance of chlorophyll; the effect of chlorophyll interference is presented in Figure 7. Figure 7 also indicated that most of the chlorophyll was centered at the bottom of the NC membrane (red signer). This result indicated that the NC membrane has the ability to separate chlorophyll from the buffer, but the NC membrane cannot completely separate the chlorophyll; therefore, the matrix inference of greengrocery samples is stronger than that of Chinese cabbage and Apple samples. The results also indicated that identifying and using materials that have the ability to separate matrix inferences are important for removing matrix inferences and improving the reliability and stability of the ICA.
The measured recoveries from the spiked samples are summarized in Table 3. The recoveries ranged from 70.6 to 131.9 for parathion-methyl, parathion, and fenitrothion. Furthermore, the CV ranged from 5.1% to 35.5%. Therefore, the recoveries of these samples were satisfactory. To confirm the accuracy and applicability of the multi-analyte ICA, the spiked samples were analyzed using ELISA. The results are shown in Table 4. The average recoveries using ELISA from all samples ranged from 79.1% to 118.1%, which were similar to those of the ICA measurements described here. Figure 6 compares the multi-analyte ICA and ELISA data. The line equation and correlation coefficient obtained from linear regression of the combined ICA and ELISA data for parathion-methyl, parathion, and fenitrothion were y = 0.9345x +15.538 and R2 = 0.9706, respectively. The high degree of linearity indicates that the ICA and ELISA techniques yielded comparable results.
Table 3. Recovery studies of samples spiked with parathion-methyl, parathion, and fenitrothion by multi-analyte ICA (n = 4).
Compound | Sample | Spiked (ng g−1) | Measured ±SD (ng g−1) | Recovery (%) | CV (%) |
Parathion-methyl | Chinese cabbage | 50 | 46.8±10.9 | 93.7 | 23.3 |
200 | 206.6±10.6 | 103.3 | 5.1 | ||
Apple | 50 | 45.4±6.9 | 90.8 | 14.3 | |
200 | 180.1±11.0 | 90.1 | 6.1 | ||
Greengrocery | 200 | 218.8±25.2 | 109.4 | 11.5 | |
1000 | 949.4±51.2 | 94.9 | 5.4 | ||
Parathion | Chinese cabbage | 50 | 44.7±16.4 | 89.3 | 36.8 |
200 | 204.1±18.2 | 102.1 | 6.8 | ||
Apple | 50 | 58.5±11.0 | 116.9 | 18.8 | |
200 | 213.4±16.8 | 106.7 | 7.9 | ||
Greengrocery | 200 | 219.2±55.3 | 109.6 | 25.2 | |
1000 | 835.7±102.7 | 83.6 | 12.3 | ||
Fenitrothion | Chinese cabbage | 100 | 97.4±34.6 | 97.4 | 35.5 |
200 | 231.7±58.4 | 115.9 | 18.2 | ||
Apple | 100 | 110.5±7.5 | 110.5 | 6.8 | |
200 | 263.8±30.2 | 131.9 | 11.4 | ||
Greengrocery | 400 | 282.5±79.3 | 70.6 | 28.1 | |
1000 | 1088.8±161.7 | 108.9 | 14.9 |
Table 4. Recovery studies of samples spiked with parathion-methyl, parathion, and fenitrothion by multi-analyte ELISA (n = 3).
Compound | Sample | Spiked (ng g−1) | Measured ±SD (ng g−1) | Recovery (%) | CV (%) |
Parathion-methyl | Chinese cabbage | 100 | 80.1±6.5 | 80.1 | 8.1 |
200 | 158.2±5.4 | 79.1 | 3.4 | ||
Apple | 100 | 89.0±8.1 | 89.1 | 9.1 | |
200 | 236.3±7.3 | 118.1 | 3.1 | ||
Greengrocery | 100 | 105.2±10.7 | 105.2 | 10.1 | |
200 | 215.2±21.9 | 107.6 | 10.2 | ||
Parathion | Chinese cabbage | 100 | 86.1±13.7 | 86.1 | 15.9 |
200 | 164.7±5.9 | 82.4 | 3.6 | ||
Apple | 100 | 89.0±8.9 | 89.0 | 10.1 | |
200 | 177.4±9.6 | 88.7 | 5.4 | ||
Greengrocery | 100 | 112.2±23.3 | 112.2 | 20.8 | |
200 | 221.2±14.7 | 110.0 | 7.4 | ||
Fenitrothion | Chinese cabbage | 100 | 90.7±9.7 | 90.7 | 10.7 |
200 | 172.5±10.9 | 86.2 | 6.3 | ||
Apple | 100 | 78.9±6.7 | 78.9 | 8.5 | |
200 | 175.4±12.1 | 87.7 | 6.9 | ||
Greengrocery | 100 | 89.6±13.9 | 89.6 | 15.6 | |
200 | 217.6±11.1 | 108.8 | 5.1 |
Conclusions
This work presents an ELISA and a gold-labeled antibody ICA test strip for detecting OP pesticides in agricultural samples. The quantitative data indicated that the specificity and accuracy of ICA were ideal and in good agreement with the ELISA measurements. The ELISA has higher sensitivity than ICA for the detection of OP pesticides in agricultural products. But the ICA technique did not require intensive labor or expensive equipment for the sample analysis; the procedure was very simple, and the detection was completed within 5 min. In summary, the established ICA method might provide an alternative tool for the sensitive, rapid, and convenient detection of OP pesticides in agricultural products, and it has advantages over ELISA methods with regard to factors such as the testing procedure, testing time, and matrix interferences, among others.
Supporting Information
Acknowledgments
We acknowledge Guoliang Qian for reviewing the manuscript and assistance throughout the study.
Funding Statement
This work was supported by the National High Technology Research and Development Program of China (2012AA101401-1), the Key Projects in the National Science & Technology Pillar Program (2009BADB9B03), the earmarked fund for the China Agriculture Research System (nycytx-29) and the Natural Science Foundation of Jiangsu Province (BK2010445). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Roex EWM, Keijzers R, van Gestel CAM (2003) Acetylcholinesterase inhibition and increased food consumption rate in the zebrafish, danio rerio, after chronic exposure to parathion. Aquat Toxicol 64: 451–460. [DOI] [PubMed] [Google Scholar]
- 2. Kim MJ, Lee HS, Chung DH, Lee YT (2003) Synthesis of haptens of organophosphorus pesticides and development of enzyme–linked immunosorbent assays for parathion–methyl. Anal Chim Acta 49: 347–62. [Google Scholar]
- 3. Farran A, Ruiz S, Serra C, Aguilar M (1996) Comparative study of high-performance liquid chromatography and micellar electrokinetic capillary chromatography applied to the analysis of different mixtures of pesticides. J Chromatogr A 737: 109–116. [Google Scholar]
- 4. Yao ZW, Jiang GB, Liu JM, Cheng W (2001) Application of solid-phase microextraction for the determination of organophosphorous pesticides in aqueous samples by gas chromatography with flame photometric detector. Talanta 55: 807–814. [DOI] [PubMed] [Google Scholar]
- 5. Padrón-Sanz C, Halko R, Sosa-Ferrera Z, Santana-Rodríguez JJ (2005) Combination of microwave assisted micellar extraction and liquid chromatography for the determination of organophosphorous pesticides in soil samples. J Chromatogr A 1078: 13–21. [DOI] [PubMed] [Google Scholar]
- 6. Xiao Q, Hu B, Yu CH, Xia LB, Jiang ZC (2006) Optimization of a single-drop microextraction procedure for the determination of organophosphorus pesticides in water and fruit juice with gas chromatography-flame photometric detection. Talanta 69: 848–855. [DOI] [PubMed] [Google Scholar]
- 7. Morozova VS, Levashova AI, Ermin SA (2005) Determination of Pesticides by Enzyme Immunoassay. J Anal Chem 60: 202–217. [Google Scholar]
- 8. Kim YJ, Cho YA, Lee HS, Lee YT (2003) Investigation of the effect of hapten heterology on immunoassay sensitivity and development of an enzyme-linked immunosorbent assay for the organophosphorus insecticide fenthion. Anal Chim Acta 494: 29–40. [Google Scholar]
- 9. Liu YH, Jin MJ, Gui WJ, Cheng JL, Guo YR, et al. (2007) Hapten design and indirect competitive immunoassay for parathion determination: Correlation with molecular modeling and principal component analysis. Anal Chim Acta 591: 173–182. [DOI] [PubMed] [Google Scholar]
- 10. Reder S, Dieterle F, Jansen H, Alcock S, Ganglitz G (2003) Multi-analyte assay for triazines using cross-reactive antibodies and neural networks. Biosens Bioelectron 19: 447–455. [DOI] [PubMed] [Google Scholar]
- 12. Xu ZL, Xie GM, Li YX, Wang BF, Beier RC, et al. (2009) Production and characterization of a broad-specificity polyclonal antibody for O,O-diethyl organophosphorus pesticides and a quantitative structure–activity relationship study of antibody recognition. Anal Chim Acta 647: 90–96. [DOI] [PubMed] [Google Scholar]
- 13. Wang CM, Li XB, Liu YH, Guo YR, Xie R, et al. (2010) Development of a Mab-based heterologous immunoassay for the broad-selective determination of organophosphorus pesticides. J Agric Food Chem 58: 5658–5663. [DOI] [PubMed] [Google Scholar]
- 14. Piao YZ, Kim YJ, Kim YA, Lee HS, Hammock BD, et al. (2009) Development of ELISAs for the class-specific determination of organophosphorus pesticides. J Agric Food Chem 57: 10004–10013. [DOI] [PubMed] [Google Scholar]
- 15. Jang MS, Lee SJ, Xue XP, Kwon HM, Ra CS, et al. (2002) Production and characterization of monoclonal antibodies to a generic hapten for class-specific determination of organophosphorus pesticides. Bull Korean Chem Soc 23: 1116–1120. [Google Scholar]
- 16. Johnson JC, Van Emon JM, Pullman DR, Keeper KR (1998) Development and evaluation of antisera for detection of the O,O-diethyl phosphorothionate and ohosphorothionothiolate organophosphorus pesticides by immunoassay. J Agric Food Chem 46: 3116–3123. [Google Scholar]
- 17. Zhang GP, Wang XN, Yang JF, Yang YY, Xing GX, et al. (2006) Development of an immunochromatographic lateral flow test strip for detection of β-adrenergic agonist Clenbuterol residues. J Immunol Methods 312: 27–33. [DOI] [PubMed] [Google Scholar]
- 18. Gandhi S, Caplashb N, Sharma P, RamanSuri C (2009) Strip-based immunochromatographic assay using specific egg yolk antibodies for rapid detection of morphine in urine samples. Biosens Bioelectron 25: 502–505. [DOI] [PubMed] [Google Scholar]
- 19. Li DW, Wei S, Yang H, Li Y, Deng AP (2009) A sensitive immunochromatographic assay using colloidal gold–antibody probe for rapid detection of pharmaceutical indomethacin in water samples. Biosens Bioelectron 24: 2277–2280. [DOI] [PubMed] [Google Scholar]
- 20. Zhou Y, Pan FG, Li YS, Zhang YY, Zhang JH, et al. (2009) Colloidal gold probe-based immunochromatographic assay for the rapid detection of brevetoxins in fishery product samples. Biosens Bioelectron 24: 2744–2747. [DOI] [PubMed] [Google Scholar]
- 21. Gas F, Baus B, Pint L, Compere C, Tanchou V, et al. (2010) One step immunochromatographic assay for the rapid detection of Alexandrium minutum. Biosens Bioelectron 25: 1235–1239. [DOI] [PubMed] [Google Scholar]
- 22. Xu Y, Huang ZB, He QH, Deng SZ, Li LS, et al. (2010) Development of an immunochromatographic strip test for the rapid detection of deoxynivalenol in wheat and maize. Food Chem 119: 834–839. [Google Scholar]
- 23. Gui WJ, Wang ST, Guo YR, Zhu GN (2008) Development of a one step strip for the detection of triazophos residues in environmental samples. Anal Biochem 377: 202–208. [DOI] [PubMed] [Google Scholar]
- 24. Kim YA, Lee EH, Kim KO, Lee YT, Hammock BD, et al. (2011) Competitive immunochromatographic assay for the detection of the organophosphorus pesticide chlorpyrifos. Anal Chim Acta 693: 106–113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Hua XD, Qian GL, Yang JF, Hu BS, Fan JQ, et al. (2010) Development of an immunochromatographic assay for the rapid detection of chlorpyrifos-methyl in water samples. Biosens Bioelectron 26: 189–194. [DOI] [PubMed] [Google Scholar]
- 26. Zhang Q, Wang LB, Ahn KC, Sun Q, Hu BS, et al. (2007) Hapten heterology for a specific and sensitive indirect enzyme-linked immunosorbent assay for organophosphorus insecticide fenthion. Anal Chim Acta 596: 303–311. [DOI] [PubMed] [Google Scholar]
- 27. Liu Y, Lou Y, Xu D, Qian GL, Zhang Q, et al. (2009) Production and characterization of monoclonal antibody for class-specific determination of O, O-dimethyl organophosphorus pesticides and effect of heterologous coating antigens on immunoassay sensitivity. Microchem J 93: 36–42. [Google Scholar]
- 28. Qian GL, Wang LM, Wu YR, Zhang Q, Sun Q, et al. (2009) A monoclonal antibody-based sensitive enzyme-linked immunosorbent assay (ELISA) for the analysis of the organophosphorous pesticides chlorpyrifos-methyl in real samples. Food Chem 117: 364–370. [Google Scholar]
- 29. Langone JJ, Vunakis H (1982) Radioimmunoassay of nicotine, cotinine, and gamma-(3-pyridyl)-gamma-oxo-N-methylbutyramide. Method Enzymol 84: 628–640. [DOI] [PubMed] [Google Scholar]
- 30. Nowinski RC, Lostrom ME, Tam MR, Stone MR, Burnette WN (1979) The isolation of hybrid cell lines producing monoclonal antibodies against the p15(E) protein of ecotropic murine leukemia viruses. Virology 93: 111–126. [DOI] [PubMed] [Google Scholar]
- 31. Kolosova AY, Park JH, Eremin SA, Park SJ, Kang SJ, et al. (2004) Comparative study of three immunoassays based on monoclonal antibodies for detection of the pesticide parathion-methyl in real samples. Anal Chim Acta 511: 323–331. [Google Scholar]
- 32. Wang CM, Liu YH, Guo YR, Liang CZ, Li XB, et al. (2009) Development of a McAb-based immunoassay for parathion and influence of the competitor structure. Food Chem 115: 365–370. [Google Scholar]
- 33. Kim YJ, Kim YA, Lee YT, Lee HS (2007) Enzyme-linked immunosorbent assays for the insecticide fenitrothion influence of hapten conformation and sample matrix on assay performance. Anal Chim Acta 591: 183–190. [DOI] [PubMed] [Google Scholar]
- 34. Chandler J, Gurmin T, Robinson N (2000) The Place of Gold in Rapid Tests. IVD Technol 6: 37–49. [Google Scholar]
- 35. Zhou P, Lu YT, Zhu J, Hong JB, Li B, et al. (2004) Nanocolloidal gold-based immunoassay for the detection of the N-methylcarbamate pesticide carbofuran. J Agric Food Chem 52: 4355–4359. [DOI] [PubMed] [Google Scholar]
- 36. Zhu J, Chen W, Lu Y, Cheng G (2008) Development of an immunochromatographic assay for therapid detection of bromoxynil in water. Environ Pollut 156: 136–142. [DOI] [PubMed] [Google Scholar]
- 37. Frens G (1973) Preparation of gold dispersions of varying particle size: controlled nucleation for the regulation of the particle size in monodisperse gold suspensions. Nat Phys Sci 241: 20–22. [Google Scholar]
- 38.Roth J (1983) Academic Press, London. in: C.R Bullock, P Petrusz, editors. Techniques in Immunocytochemistry. 217–284.
- 39. Reverberi R, Reverberi L (2007) Factors affecting the antigen-antibody reaction. Blood Transfus 5: 227–240. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Krotzky AJ, Zeeh B (1995) Immunoassays for residue analysis of agrochemicals: proposed guidelines for precision, standardization and quality control. Pure Appl Chem 67: 2065–2088. [Google Scholar]
- 41. Gui WJ, Liu YH, Wang CM, Liang X, Zhu GN (2009) Development of a direct competitive enzyme-linked immunosorbent assay for parathion residue in food samples. Anal Biochem 393: 88–94. [DOI] [PubMed] [Google Scholar]
- 42. Sun JW, Dong TT, Zhang Y, Wang S (2010) Development of enzyme linked immunoassay for the simultaneous detection of carbaryl and metolcarb in different agricultural products. Anal Chim Acta 666: 76–82. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.