Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
letter
. 2013 Jan;51(1):379. doi: 10.1128/JCM.02509-12

Use of Rapid Diagnostic Tests for Diagnosis of Malaria in the United States

April Bobenchik 1, Robyn Shimizu-Cohen 1, Romney M Humphries 1,
PMCID: PMC3536259  PMID: 23135948

LETTER

In the September issue of this journal, performance of BinaxNOW for the diagnosis of malaria in a major U.S. academic medical center was reported (1). Herein, we report performance of this test in a second large academic medical center in the United States. BinaxNOW was introduced at our institution as a component of routine blood parasite examination along with thick- and thin-blood-smear microscopy in January 2011. From 2007 to 2011, BinaxNOW was also selectively performed on specimens that were positive for malaria by microscopy as part of laboratory verification studies. In total, 251 parasite blood exams from 244 patients were performed. Of these, 239 (95.2%) were negative by microscopy. The remaining 12 positive specimens included 4 Plasmodium falciparum isolates, 5 Plasmodium vivax isolates, and 3 Plasmodium malariae isolates, with corresponding parasitemias of <0.01% to 0.5%. BinaxNOW resulted in 9 positive results. Three false-negative BinaxNOW results were obtained, all in patients with P. malariae infection and parasitemia of <0.1%. One false-positive BinaxNOW result was obtained in a 17-month-old child with 2 weeks of recurrent fevers and travel history to a region where malaria is endemic 1 year prior to symptoms. The clinical index of suspicion for malaria in this case was low. All discrepant results were confirmed by repeat testing using BinaxNOW and reevaluation of blood smears. The overall sensitivity, specificity, positive predictive value, and negative predictive value for BinaxNOW were found to be 72.7%, 96.8%, 88.8%, and 98.8%, respectively (Table 1). When patients with P. malariae infection or <0.1% parasitemia, neither of which are covered by BinaxNOW FDA performance claims, were removed from analysis, sensitivity improved to 100%.

Table 1.

Performance of BinaxNOW for 251 blood specimens evaluated for malaria at a large academic hospital laboratory between 2007 and 2012a

Performance measure Overall performance (%) (95% CI) Performance for FDA-approved claims only (%) (95% CI)b
Sensitivity 72.7 (39.3–92.7) 100 (59.8–100.0)
Specificity 99.6 (97.3–99.9) 99.6 (97.3–99.9)
Positive predictive value 88.9 (50.7–99.4) 88.9 (50.7–99.4)
Negative predictive value 96.4 (93.1–98.2) 100 (98.0–100.0)
a

CI, confidence interval.

b

P. falciparum and P. vivax with >0.1% parasitemia.

Based on the results of our study, our laboratory uses BinaxNOW as an adjunct to thin- and thick-smear microscopy, primarily to aid with species identification when only early trophozoites are observed by microscopy and to provide a rapid preliminary positive result identifying P. falciparum from non-falciparum Plasmodium species. A rapid identification of P. falciparum is critical in cases of severe malaria, which necessitate prompt treatment. In particular, intravenous artesunate has recently been made available by the Centers for Disease Control and Prevention for this indication under an investigational new drug protocol (http://www.cdc.gov/malaria/diagnosis_treatment/artesunate.html).

Performance data for BinaxNOW indicate that this test cannot replace microscopy, as we and others have observed both false-negative and false-positive results (25). Clinical ramifications of reporting false-negative results are of most pressing concern, in particular because many patients seen in the United States with P. falciparum malaria will have low-level parasitemia that is below the limit of detection of BinaxNOW. False-negative results may also be obtained at high-level parasitemia due to the prozone phenomenon or possibly a mutation in histidine-rich protein 2 (HRP-2) antigen detected by the test (1, 5).

It is unquestionable that maintaining expertise in malaria diagnostics at the hospital level in the United States is challenging due to regulatory and budget considerations and the fact that laboratories may encounter only 0 to 2 cases of malaria per year. As such, the promise of a rapid, low-complexity test that can replace labor-intensive microscopy techniques is highly attractive. However, current technologies have not yet achieved a sensitivity and specificity to allow the laboratory to discontinue microscopy. Only 17% of laboratories in the United States use rapid diagnostic tests such as BinaxNOW (2), the majority in conjunction with microscopy. We urge laboratories that are considering implementation of this test to ensure that capacity for microscopy diagnostics is maintained. Nevertheless, the implementation of the rapid diagnostic test can be very useful in providing physicians with quick preliminary results, allowing for appropriate clinical management of the patient.

Footnotes

Published ahead of print 7 November 2012

REFERENCES

  • 1.Dimaio MA, Pereira IT, George TI, Banaei N. 2012. Performance of BinaxNOW for diagnosis of malaria in a U.S. hospital. J. Clin. Microbiol. 50:2877–2880 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Abanyie FA, Arguin PM, Gutman J. 2011. State of malaria diagnostic testing at clinical laboratories in the United States, 2010: a nationwide survey. Malar. J. 10:340. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Iqbal J, Sher A, Rab A. 2000. Plasmodium falciparum histidine-rich protein 2-based immunocapture diagnostic assay for malaria: cross-reactivity with rheumatoid factors. J. Clin. Microbiol. 38:1184–1186 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Jelinek T, Grobusch MP, Nothdurft HD. 2000. Use of dipstick tests for the rapid diagnosis of malaria in nonimmune travelers. J. Travel Med. 7:175–179 [DOI] [PubMed] [Google Scholar]
  • 5.Khairnar K, Martin D, Lau R, Ralevski F, Pillai DR. 2009. Multiplex real-time quantitative PCR, microscopy and rapid diagnostic immuno-chromatographic tests for the detection of Plasmodium spp: performance, limit of detection analysis and quality assurance. Malar. J. 8:284. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES