Abstract.
We achieved fast and automatic retinal vessel segmentation by employing the active contours method in Doppler spectral-domain optical coherence tomography (SD-OCT). In a typical OCT B-scan image, we first extracted the phase variations between adjacent A-lines and removed bulk motion. Then we set the initial contour as the boundary of the whole image and iterated until all of the segmented vessel contours became stabilized. Using a typical office computer, the whole segmentation took no more than 50 s, making real-time retinal vessel segmentation possible. We tested the active contours method segmentation in both controlled phantom and in vivo rodent eye images.
Keywords: image processing, retinal imaging, optical coherence tomography
1. Introduction
Alterations in retinal blood flow velocity and volume can be associated with several ocular and systemic diseases, including, for example, ischemic optic neuropathy, glaucoma, diabetes, and HIV.1–3 Accurately measuring blood flow can potentially help to better understand their pathophysiologies and to achieve early diagnosis of these diseases. Spectral-domain optical coherence tomography (SD-OCT) is a primary modality used to image the retinal anatomy with high spatial resolution.4,5 By taking advantage of the Doppler effect impinging on the probing light induced by moving optical scatterers, OCT can measure retinal blood flow velocity, which further enables the OCT angiogram.6–11
Accurate estimation of blood flow volume in a retinal vessel requires OCT to obtain first, the spatial location and boundary of retinal vessels, second, the Doppler angles between the retinal blood flow and the OCT probing light, and third, the phase variance within the vessel caused by flowing blood. Spatial localization and boundary of a vessel could be obtained from OCT amplitude images,12,13 which usually takes a long time, or by human operator,6 which is subjective and lacks consistency among operators. Only after the spatial location and boundary of a vessel are identified, phase variance within the vessel boundary can be averaged for mean flow velocity, and centerline of the vessel can be further obtained to calculate the Doppler angle. From this point of view, retinal vessel segmentation is the key first step to achieve correct blood flow measuring.
Previously, histogram-based filtering14 and support vector machine (SVM)15 were used for retinal vessel segmentation in phase contrast OCT. Histogram-based filtering method worked well in images with high signal to noise ratios (SNRs); but showed poor results in relatively noisy images. Good segmentation results can also be acquired by SVM method; however, SVM method is time-consuming and requires a lengthy prior training. Here, we report a new, fast, and automatic vessel segmentation method for SD-OCT based on active contours method (ACM) and phase contrast.
2. Materials and Methods
2.1. Phantom and Animal Experiment
Both controlled phantom and in vivo rodent experiments were conducted. Phantom experiments were designed to test the accuracy of ACM segmentation. Specifically, a capillary tube was filled with a 1% Intralipid solution. The tube (CTPS125-250-5, Paradigm; inner diameter: 125 μm; outer diameter: 250 μm) was connected to a perfusion pump (A-99, Razel) at a flow rate of . We imaged the phantom using a home-made SD-OCT system, which has been described in detail previously.16 The OCT light source (IPSDD0804, InPhenix) has a center wavelength of 840 nm and a 3-dB bandwidth of 50 nm. The A-line rate was 24 kHz. The spatial resolutions of the SD-OCT in tissue were 6 μm (axial) and 20 μm (lateral). In total, 32 B-scans phantom images were acquired to test ACM segmentation accuracy, where results from all the 32 images were acquired independently and used to obtain mean and standard deviation (STD) of ACM segmentation result of the capillary tube. For comparison, histogram-based filtering segmentation was also performed.
Besides phantom experiment, we imaged rodent eyes in vivo. All animal experimental procedures were approved by the Institutional Animal Care and Use Committee at Northwestern University. Detailed information of the experimental procedure can be found in our previous paper.17
2.2. Preprocessing of Retinal SD-OCT Images
Figure 1 shows the preprocessing steps for in vivo SD-OCT images. Figure 1(a) is an SD-OCT fundus image of a rat eye, where the circle and arrow highlight the trajectory and direction of a circular scan around the optic disk. We first acquired a B-scan amplitude image containing 2048 A-lines [Fig. 1(b)] and further extracted the phase contrast [Fig. 1(c)] by subtracting the phase values between spatially adjacent A-lines. Then we corrected the bulk motion following the method reported in literature.18 After that, we applied a median-filter for noise reduction to get a cleaner background in the phase contrast OCT image [Fig. 1(d)]. From both Fig. 1(c) and 1(d), we can observe that, indeed, all of the retinal blood vessels were located within a small depth range. In order to speed up the segmentation and improve the segmentation accuracy, we reduced the depth range of the OCT image and only applied ACM to the reduced region. The reduced-size image was determined by calculating the STD of the OCT phase variation along the horizontal axis. We observed that STD values were significantly higher within depth range exhibiting the highest vascular density compared to STD values from depth ranges without vessels [Fig. 1(e)]. Based on the calculated depth-resolved STD, we digitally removed the region beyond 100 pixels above and below the STD max [Fig. 1(f)].
Fig. 1.
Preprocessing steps of in vivo rodent eye image. (a) OCT fundus image of rat retina. The circle highlights the scanning trajectory around the optic disk. Bar 200 μm; (b) OCT B-scan amplitude image from the position highlighted in panel a. Bar: 250 μm; (c) OCT B-scan phase image; (d) OCT B-scan phase image after bulk motion correction; (e) Phase STD distribution along the depth axis; and (f) reduced-size image OCT B-scan phase image.
2.3. Active Contours Method
The underlying principle of ACM is to evolve a contour enclosing the targeted objects in an image based on region-based gradient inside and outside the contour. A final contour may consist of boundaries of multiple objects.
According to Chan and Vese,19 for a given image in domain (real number), ACM method is formulated by minimizing the energy defined by:
| (1) |
where is the contour, and and are the average intensities inside and outside the contour, respectively. Equation (1) can be solved iteratively through the following steps.
-
Step 1.Initialize a contour containing all the objects that are of interest. In the present study, the initial contour encompasses the boundary of the whole image. Then apply an initial level set (iteration index ):
(2) - Step 2.
-
Step 3.Update the level set function with iteration number:20
where is a factor to increase the segmentation speed, and is the function estimating the gradient. The reason why we applied this level set function is that it improves the traditional level set methods by avoiding the estimation of the signed distance function (SDF) and re-initialization;20 thus, it’s more efficient.(5) -
Step 4.
Re-estimate and as in step 2 for the iteration; then compare the results between the th and th iterations. If the difference of and between the th and th iterations are very small (such as ), the contour is stabilized, and segmentation is finished; otherwise, iteration continues. Although convergence may affect segmentation results very much, our tests of ACM segmentation typically converged within 100 iterations.
3. Results and Discussion
3.1. Phantom Imaging Result
We first tested the performance of ACM segmentation using a controlled phantom as shown in Fig. 2. Figure 2(a) is the OCT amplitude image of the capillary tube, where we can observe the Intralipid solution, the outer boundary, and the bottom of the container. The corresponding OCT phase image is shown in Fig. 2(b). After applying our algorithm described in Sec. 2.3, we extracted the boundary profile of the tube and further estimated the tube center based on the acquired boundary [Fig. 2(b)]. Based on the segmented result and the calibrated OCT resolutions, we estimated the inner diameter of the tube to be (after averaging 32 B-scan), which agreed with the specification (125 μm). The phantom imaging result confirmed the accuracy of the ACM segmentation method in OCT phase images. Figure 2(c) is the segmentation result using histogram-based filtering method and the estimated the tube’s inner diameter was , which agreed with the specification. However, we can clearly observe segmentation errors in the background as pointed out by the arrows in Fig. 2(c).
Fig. 2.
Imaging and segmentation results of the capillary tube phantom. (a) OCT B-scan amplitude image. Bar: 100 μm; (b) ACM segmentation results overlaid on OCT B-scan phase image. The black circle is detected contour and black cross is the calculated vessel center; and (c) Histogram-based filtering segmentation result.
3.2. In Vivo Rodent Eye Imaging
The segmentation results of the in vivo image are shown in Fig. 3. The boundaries and the center positions of all retinal vessels obtained from ACM are showed in Fig. 3(a). Altogether 12 vessels were identified by the algorithm, which agreed with our visual observation in the OCT fundus and amplitude OCT image [Fig. 1(a) and 1(b)]. By overlaying the detected boundaries onto the OCT phase image; we segmented all the retinal vessels from the background as shown in Fig. 3(b). The color in Fig. 3(b) represents the direction of the blood flow. Based on the size of the vessels and the magnitude of phase variances, it is reasonable to hypothesize that the venous blood flows outwards (red) and arterial blood flows inwards (blue) with respect to the image plane. Moreover, the vessels no. 1 and no. 5 (from the left) in Fig. 3(b) showed phase wrap, presumably caused by high flow velocity of the arterial blood; however, ACM is not affected by such phase wrap. We tested in vivo images from 10 rodent eyes (data not shown); the segmentation results were similar to Fig. 3.
Fig. 3.
Segmentation results of in vivo retinal image. (a) Vessel boundaries and centers obtained from ACM method. Bar: 100 μm; (b) OCT phase image after all vessels are segmented; (c) segmentation results from histogram-based filtering method. The color bars in panels b and c represent OCT phase variation; and (d) comparison of estimated vessel diameters obtained by the two segmentation methods.
Once the boundaries and center positions of all vessels are obtained, accurate calculation of mean vessel phase variation and the Doppler angle for complete retinal blood flow measurements can proceed. During the whole segmentation and center identification processes, we provided no user input. As a comparison, we showed the segmentation result using the histogram-based filtering method14 in Fig. 3(c). Although all vessels were segmented, the boundaries are not as uniform as Fig. 3(b), and it also has a noisy background (as pointed out by arrows), all of which may affect further information extraction based on the segmented results. Vessel diameters obtained from both ACM and histogram-based filtering were compared [Fig. 3(d)] and they, in general, agreed with each other. However, we cannot conclude which method is more precise because there is no gold standard to measure retinal vessel diameters in vivo.
We also tested the speed of our ACM segmentation in ten different in vivo OCT B-scan images with the same image size (2048 points per A-line, 4096 A-lines per B-scan). Using a typical office laptop (Thinkpad™ T400 with RAM 2 GB and Intel™ P8600 Processor at 2.4 GHz), our Matlab™ (MathWorks, version 2011a) code took for OCT raw image reconstruction, for Doppler signal extraction, and for ACM segmentation. The histogram-based filtering was faster than ACM in image segmentation, where was needed in the segmentation step. However, the results from histogram-based method were not as good as results from ACM. Such performance of ACM can allow real-time clinical applications in the future if we further optimize the algorithm and, especially, take advantage of advanced hardware such as graphic processing unit.21,22
4. Conclusion
In this work, we applied a fast and automatic retinal vessel segmentation method, active contours method, in phase contrast OCT images. We first examined the segmentation accuracy of the ACM in controlled phantom experiments. In an Intralipid-filled capillary tube, the measured inner diameter agreed well with the manufacturer specification. Then we tested our method in multiple in vivo rodent retinal images. In every tested image, we obtained all the vessel boundaries and vessel centers with a processing time less than 1 min. The significance of ACM method lies in its automatic capabilities in obtaining vessel centers and boundaries fast, which permits future real-time operations and better estimation of Doppler angles and blood flow velocities in clinical settings.
Acknowledgments
This work is supported in part by NIH Grants 1R01EY019951 and 1RC4E021357, and NSF Grants CBET-1055379 (CAREER) and CBET-1066776.
References
- 1.Padnick-Silver L., et al. , “Retinal oxygenation and oxygen metabolism in Abyssinian cats with a hereditary retinal degeneration,” IOVS 47(8), 3683–3689 (2006). 10.1167/iovs.05-1284 [DOI] [PubMed] [Google Scholar]
- 2.Puro D. G., “Retinovascular physiology and pathophysiology: new experimental approach/new insights,” Prog. Retin. Eye Res. 31(3), 258–70 (2012). 10.1016/j.preteyeres.2012.01.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Polak K., et al. , “Regulation of human retinal blood flow by endothelin-1,” Exp. Eye Res. 76(5), 633–640 (2003). 10.1016/S0014-4835(02)00312-3 [DOI] [PubMed] [Google Scholar]
- 4.Wojtkowski M., et al. , “In vivo human retinal imaging by Fourier domain optical coherence tomography,” J. Biomed. Opt. 7(3), 457–463 (2002). 10.1117/1.1482379 [DOI] [PubMed] [Google Scholar]
- 5.Srinivasan V. J., et al. , “Ultrahigh-speed optical coherence tomography for three-dimensional and en face imaging of the retina and optic nerve head,” IOVS 49(11), 5103–5110 (2008). 10.1167/iovs.08-2127 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Wang Y., et al. , “Retinal blood flow measurement by circumpapillary Fourier domain Doppler optical coherence tomography,” J. Biomed. Opt. 13(6), 064003 (2008). 10.1117/1.2998480 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Wang R. K., An L., “Doppler optical micro-angiography for volumetric imaging of vascular perfusion in vivo,” Opt. Express 17(11), 8926–8940 (2009). 10.1364/OE.17.008926 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Zhi Z., et al. , “Impact of intraocular pressure on changes of blood flow in the retina, choroid, and optic nerve head in rats investigated by optical microangiography,” Biomed. Opt. Express 3(9), 2220–2233 (2012). 10.1364/BOE.3.002220 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Zhang J., Chen Z., “In vivo blood flow imaging by a swept laser source based Fourier domain optical Doppler tomography,” Opt. Express 13(19), 7449–7457 (2005). 10.1364/OPEX.13.007449 [DOI] [PubMed] [Google Scholar]
- 10.White B., et al. , “In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography,” Opt. Express 11(25), 3490–3497 (2003). 10.1364/OE.11.003490 [DOI] [PubMed] [Google Scholar]
- 11.Walther J., et al. , “Optical coherence tomography in biomedical research,” Anal. Bioanal. Chem. 400(9), 2721–2743 (2011). 10.1007/s00216-011-5052-x [DOI] [PubMed] [Google Scholar]
- 12.Pilch M., et al. , “Automated segmentation of retinal blood vessels in spectral domain optical coherence tomography scans,” Biomed. Opt. Express 3(7), 1478–1491 (2012). 10.1364/BOE.3.001478 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wehbe H., et al. , “Automatic retinal blood flow calculation using spectral domain optical coherence tomography,” Opt. Express 15(23), 15193–15206 (2007). 10.1364/OE.15.015193 [DOI] [PubMed] [Google Scholar]
- 14.Kolbitsch C., Schmoll T., Leitgeb R. A., “Histogram-based filtering for quantitative 3D retinal angiography,” J. Biophoton. 2(6–7), 416–425 (2009). 10.1002/jbio.v2:6/7 [DOI] [PubMed] [Google Scholar]
- 15.Singh A. S. G., Schmoll T., Leitgeb R. A., “Segmentation of Doppler optical coherence tomography signatures using a support-vector machine,” Biomed. Opt. Express 2(5), 1328–1339 (2011). 10.1364/BOE.2.001328 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Liu T., et al. , “Combined photoacoustic microscopy and optical coherence tomography can measure metabolic rate of oxygen,” Biomed. Opt. Express 2(5), 1359–1365 (2011). 10.1364/BOE.2.001359 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Song W., et al. , “Integrating photoacoustic ophthalmoscopy with scanning laser ophthalmoscopy, optical coherence tomography, and fluorescein angiography for a multimodal retinal imaging platform,” J. Biomed. Opt. 17(6), 061206 (2012). 10.1117/1.JBO.17.6.061206 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Yang V., et al. , “High speed, wide velocity dynamic range Doppler optical coherence tomography (part I): system design, signal processing, and performance,” Opt. Express 11(7), 794–809 (2003). 10.1364/OE.11.000794 [DOI] [PubMed] [Google Scholar]
- 19.Chan T. F., Vese L. A., “Active contours without edges,” IEEE Trans. Image Process. 10(2), 266–277 (2001). 10.1109/83.902291 [DOI] [PubMed] [Google Scholar]
- 20.Zhang K., et al. , “Active contours with selective local or global segmentation: a new formulation and level set method,” Image Vis. Comput. 28(4), 668–676 (2010). 10.1016/j.imavis.2009.10.009 [DOI] [Google Scholar]
- 21.Zhang K., Kang J. U., “Real-time 4D signal processing and visualization using graphics processing unit on a regular nonlinear-k Fourier-domain OCT system,” Opt. Express 18(11), 11772–11784 (2010). 10.1364/OE.18.011772 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Huang Y., Liu X., Kang J. U., “Real-time 3D and 4D Fourier domain Doppler optical coherence tomography based on dual graphics processing units,” Biomed. Opt. Express 3(9), 2162–2174 (2012). 10.1364/BOE.3.002162 [DOI] [PMC free article] [PubMed] [Google Scholar]