Summary
Embolization using n-butyl-cyanoacrylate (NBCA) for arteriovenous malformation (AVM) is now a daily practice over the world, but there exists no objective data that can be a basis for discussion or decision-making on the best concentration and injection rate of NBCA mixture. The purpose of this study was to obtain objective data on control and behavior of NBCA mixture with an in vitro simulation system of NBCA embolization for AVM.
A nidus model made of a one-ml syringe filled with small beads was connected to a pulsatile flow circuit. A microcatheter was introduced just before the nidus model. Endoluminal pressures proximal and distal to the nidus and flow volume through the nidus were measured. Digital subtraction angiography (DSA) was performed to calculate transit time of the contrast medium (CM) through the nidus. NBCA was injected at various rates with an autoinjector and transit time of NBCA through the nidus was calculated.
27 trials were completed. Transit time of CM through the nidus model is well correlated to flow volume per unit of time through the nidus model. Shorter the transit time, larger was the flow volume per unit of time. The correlation was statistically significant (P < .0001). Though statistical significance was not attained, transit time of NBCA mixture at 50% concentration had a tendency to be correlated to flow volume per unit of time through the nidus, and slower injection of the NBCA mixture led to slower filling of the nidus model.
Though this simulation system was artificial and the results should be interpreted carefully, it was shown with this system that transit time of CM through the nidus could be a good index for flow volume per unit of time through the nidus. Also suggested was a possibility to utilize this in vitro system for research and training on NBCA embolization of AVM.
Key words: AVM, embolization, NBCA
Introduction
Embolization using n-butyl-cyanoacrylate (NBCA) for arteriovenous malformation (AVM) is now a daily practice over the world. But its control or handling is learned only on clinical practice, and we have no objective indices to decide the concentration and injection rate of NBCA mixture in relation to the flow characteristics of the target of our therapy. Though in vitro simulation study cannot fully reproduce the clinical setting of NBCA embolization and its results should be carefully interpreted, it may be helpful in order to findsome objective indices with which we can analyze our clinical experience and improve therapeutic decision-making.
In this study we tried to develop an in vitro system to simulate NBCA embolization of AVM, to evaluate some indices of flow dynamics of AVM, and to clarify relations among haemodynamic factors, operational factors, and behavior of NBCA.
Material and Methods
Pulsatile Flow Circuit:
Isoda’s pulsatile flow circuit1 was modified and applied to generate pulsatile flow (figure 1A). In brief explanation, this circuit consisted of two pumps; a constant flow pump and a pulsatile flow pump connected in parallel. The fluid running in this circuit was composed of; 2L of 10 times concentrated phosphate-buffered saline, 18L of non-ionized water, and 4L of low molecular weight dextran with a pH of 7.327.33 at its temperature of 21-25 degrees C.
Figure 1.
A) Diagram of the pulsatile flow circuit. B) Diagram of the simulation component with an artificial nidus model. C) An artificial nidus model made by putting small beads in a one-ml syringe.
Simulation Component
The simulation component comprised of an artificial nidus model connected via an extension tube and a Y-connector to the pulsatile flow circuit (figure 1B). The artificial nidus model was made by putting small beads (3.4 mm in radius or 2.5 × 4.5 mm in size) in a one-ml syringe (figure 1C). The proximal end of the artificial nidus was connected to an extension tube (3.2 mm in inner diameter and 50 cm long or 1.5 mm in inner diameter and 2 cm long) and then through one limb of a Y-connecter to the pulsatile flow circuit. Through the other limb of the Y-connecter, a microcatheter was inserted to the proximal end of the nidus model. The distal end of the nidus model is connected to a series of three extension tubes (3.2 mm in inner diameter and 50 cm long each) and the end of the last tube was then opened into a bucket.
Procedure
1. Monitor simultaneously endoluminal pressures just proximal to the nidus, through the microcatheter, and just distal to the nidus through a three-way cock.
2. Measure flow volume running through the nidus model by receiving the fluid pouring out from the distal end of the extension tube into the bucket, and calculate flow volume per unit of time (ml/min) through the nidus.
3. Run digital subtraction angiography (DSA) at 10 acquisitions/sec with contrast medium (CM) (300 mgI/100 ml) injected through the microcatheter at 0.3 ml/sec of injection rate with help of an autoinjector.
4. Replay the series of acquired images. Find the images where the top of CM just passed the proximal or distal end of the nidus model. Assume the time that elapsed between the two images as the transit time for the trial.
5. Under DSA, inject 1 ml of mixture of NBCA and Lipiodol at 80,50, or 25% of NBCA concentration and push it with 5% glucose solution with an autoinjector at 2ml/min. Trials were additionally done with NBCA mixture at 50% concentration injected at 16, 4, and 1 ml/min of injection rate, as well as injected manually.
6. Replay the series of acquired images. Find the images where the top of NBCA mixture just passed the proximal or distal end of the nidus model. Assume the time that elapsed between the two images as the transit time for the trial.
Results
The full length of the procedure was accomplished in 27 trials: two trials with NBCA mixture at 25% concentration at 2 ml/min of injection rate, 11 trials with NBCA mixture at 50% concentration at 2 ml/min, 10 trials with NBCA mixture at 80% concentration at 2 ml/min, one trial with NBCA mixture at 50% concentration injected at each 16, 4, and 1 ml/min, one trial with NBCA mixture at 50% concentration injected manually.
Pressure waves monitored at the proximal and the distal to the nidus were pulsatile, but not physiological in the shape (figure 2). Mean pressure difference over the nidus model varied between 1-69 mmHg. Flow volume per unit of time through the nidus measured between 21.2136.0 ml/min. Transit time of CM varied from 0.53 to 1.20 sec. NBCA mixture passed through the nidus model in 23 trials and transit time of the mixture varied from 0.40 to 40.14 sec. NBCA mixture did not pass the nidus model and occluded a proximal part of the nidus in 4 trials; two trials with NBCA mixture at 80% concentration of NBCA injected at 2 ml/min, one trial with NBCA mixture at 50% concentration at 2 ml/min, and one trial with NBCA mixture at 50% concentration injected manually.
Figure 2.
Pressure waveforms: an example;A) pressure monitored just proximal to the nidus model, B) pressure monitored just distal to the nidus model. Pulsatile, biphasic change is observed, though not fully physiological in the form.
Mean pressure difference over the nidus model was not correlated to flow volume per unit of time through the nidus with statistical significance (P = .270278). Transit time of CM was well correlated to flow volume per unit of time through the nidus model (figure 3). Larger the flow volume per unit of time, shorter was the transit time. This correlation between transit time of CM and flow volume per unit of time was statistically significant (P = .000282).
Figure 3.
Transit time of contrast medium (CM) (sec) through the nidus model is plotted as a function of flow volume per unit of time (ml/min) through the nidus model. Transit time of CM is well correlated to flow volume per unit of time. Shorter the transit time, larger the flow volume per unit of time is. The correlation coefficient (r) is - 0.644882, and the coefficient of determination (R2) is 0.415872. The P value is 0.000282, which means that the correlation is statistically significant.
Transit time of NBCA mixture at 50% concentration injected at 1,2,4,16 ml/min of injection rate as well as injected manually is shown in the same figure (figure 4). When NBCA mixture at 50% concentration was injected at 2 ml/min of injection rate, transit time of the mixture showed a tendency to be correlated to flow volume per unit of time through the nidus model. But in one trial, NBCA mixture did not pass the nidus model and occluded a proximal part of it. This precluded statistical significance. The transit time for this trial is plotted as infinity in the figure.
Figure 4.
Transit time through the nidus model of NBCA at 50% concentration injected at 1,2, 4, 16 ml/min and injected manually is plotted as a function of flow volume per unit of time (ml/min) through the nidus model. When NBCA mixture does not pass the nidus model, the transit time is plotted as infinity. When the mixture is injected at 2ml/min of injection rate, transit time of the mixture shows a tendency to be correlated to flow volume per unit of time through the nidus model, but no statistical significance is noted. Injected at 16,4,2,1 ml/min in the flow through the nidus model around 120 ml/min, slower injection leads to slower filling of the nidus model. Injected manually, NBCA mixture slowly occluded proximal 70% of the nidus model and did not pass the nidus.
When NBCA mixture at 50% concentration was injected at 16, 4, 2, 1 ml/min, in one trial for each injection rate, in the flow through the nidus model around 120 ml/min, slower injection led to slower filling of the nidus model (figure 4). When NBCA mixture at the same concentration was injected manually, it slowly occluded proximal 70% of the nidus model and did not pass the nidus.
Transit time of NBCA mixture at 25, 50, and 80% concentration injected at the same rate of 2 ml/min is shown in the same figure (figure 5). Transit time of 80% NBCA and that of 25% NBCA showed a tendency to be shorter than that of 50% NBCA. NBCA mixture at 50% concentration did not pass the nidus model in one trial as well as that at 80% concentration in two trials. In these cases where the mixture did not pass the nidus model, the transit time is plotted as infinity in the figure.
Figure 5.
Transit time of NBCA mixture at 25,50, and 80% concentration injected at the same rate of 2 ml/min is plotted as a function of flow volume per unit of time (ml/min) through the nidus model. Transit time of 80% NBCA and that of 25% NBCA showed a tendency to be shorter than that of 50% NBCA. NBCA mixture at 50% concentration did not pass the nidus model in one trial as well as that at 80% concentration in two trials, and the transit time in these trials is plotted as infinity in the figure.
Discussion
In these years, embolization with NBCA for AVM is a therapy of choice undertaken all over the world. But its control or handling is learned only on clinical practice in days of our apprenticeship, and even later as a main operator, we have no objective indices when we are to decide the concentration and injection rate of NBCA mixture on the basis of the flow characteristics of the target of our therapeutic intervention, or to later discuss with others on the choice of the concentration and injection rate taken at the therapeutic session.
In vitro simulation of AVM embolization have been reported since late 1970’s2-7, but the purposes of these studies were mainly for teaching and training, and not for research, though possibility of application of these models in research domain was mentioned in some of these reports. Another article8 tackled on in vitro simulation of embolization to collect objective data including flow characteristics and results of embolization, but they used polyvinyl acetate as an embolization agent.
In our study, we focused our efforts on deriving some objective indices that can help us decide on the concentration and injection rate of NBCA mixture in a clinical setting. We made clear the relationship of transit time of CM through the artificial nidus and flow volume per unit of time through the nidus, and analyzed the influence of concentration and injection rate of NBCA mixture and flow volume per unit of time through the nidus upon transit time of NBCA mixture through the nidus. To our knowledge, this is the first report dealing with the relation between transit time of CM through an in vitro AVM nidus model and its flow characteristics as well as other relations among haemodynamic factors, operational factors, and results in in vitro simulation of NBCA embolization.
With materials available to us at this period of time, full reproduction of vascular structure of AVM is impossible. For example, endoluminal surface of an artificial tube is different from that of physiological vessels. Even though an interesting report9 showed that materials of vessels, either vinyl or animal vein, made no difference in polymerization of NBCA mixture in vessels filled with still blood, straight extrapolation to the flowing fluid condition should not be made at the moment. Walls of vessels in living creatures are elastic, while tubing material available for in vitro simulation is rigid and non-elastic. Blood is not a Newtonian fluid, while fluid we usually use is Newtonian. These were some of the limitations in our in vitro simulation system. Even if we took elastic materials and non-Newtonian fluid, adjustment of elasticity, resistance of vessels, and characteristics of the non-Newtonian fluid to reproducethe real conditions in living beings would be of tremendous complexity, and make another field of study, the field of flow dynamics and biomechanics. Nevertheless, if we interpret the results of in vitro simulation study carefully with its limitations in our mind, it may be helpful in order to find some objective indices with which we can analyze our clinical experience and to improve therapeutic decision-making.
For the fluid running the circuit, we adjusted its pH to 7.32-7.33. Though it was still a little bit lower than in the normal arterial blood, it roughly reproduced the negatively charged condition of the normal blood, which is one of the prerequisites for polymerization of NBCA10. The temperature of the fluid was at the room temperature of 21-25 degrees C and may have made slower the polymerization time of NBCA than in the human blood 10,11. These are other conditions to consider when we evaluate the results of this study. The range of the flow volume per unit of time through the nidus in this experiment was between 21.2-136.0 ml/min, and can be considered to represent low to at least moderate flow volume running into an AVM nidus from one of the feeding arteries, as the flow in feeding arteries at the level of the circle of Willis was reported to measure from 50 to 550 ml/min12.
Mean pressure difference over the nidus model was not correlated to flow volume per unit of time through the nidus with statistical significance. One of the reasons could lie in the different ways of measuring pressure at the proximal and distal points to the nidus model, through a microcatheter at the former and through a three-way cock at the latter. Diameter of the inner lumen of the three-way cock was smaller at the point just below the third arm of the cock than the other parts proximal or distal to this point, which could also influence the pressure measured at this point in the flowing fluid condition.
We assumed as transit time of CM or NBCA mixture the time that it took for the top of CM or NBCA mixture to pass the full length of the nidus model. This is a simpler way of measuring transit time than computer-assisted calculation based on the time-density curve of CM or NBCA mixture, and can be employed more easily at daily clinical practice.
Transit time of CM was well correlated to the flow volume per unit of time through the nidus model (figure 3) with statistical significance (P = .000282). Shorter the transit time, larger was the flow volume per unit of time. This result suggests that using transit time of CM as an index of flow per unit of time through a nidus in a clinical setting is on a significant scientific basis, even though this is derived from an in vitro simulation with some of the limits described above.
Transit time of NBCA mixture at 50% concentration injected at 2 ml/min of injection rate showed a tendency to be correlated to flow volume per unit of time through the nidus model (figure 4), though this tendency is not statistically significant. In one trial, NBCA mixture did not pass the nidus model and occluded only a proximal part of it, which may suggest existence of uncontrolled elements in the study setting. When NBCA mixture at 50% concentration was injected at 16, 4, 2, 1 ml/min, slower injection led to slower filling of the nidus model, and injection was well controlled manually to achieve a good obliteration of the proximal 70% of the nidus model. Though only one trial was done for each injection rate, this result corresponded well to our experience of daily practice.
Transit time of NBCA mixture at 25, 50, and 80% concentration injected at the same rate of 2 ml/min was not well correlated to the concentration of the mixture (figure 5). This can also suggest uncontrolled elements in the study setting. Or it may reveal another possibility that higher concentration of NBCA mixture does not necessarily lead to shorter polymerization time in some condition of flow rate and injection rate, where anions necessary for polymerization can not be recruited from the surrounding flowing fluid so much as higher concentration of NBCA in the mixture, as the contact surface of the NBCA mixture with the surrounding flowing fluid is rather constant among different concentrations of NBCA.
Conclusions
An in vitro system to simulate NBCA embolization of AVM was developed. Though this simulation system was artificial and the results should be interpreted carefully, it was shown with this system that transit time of CM through the nidus could be a good index for flow volume per unit of time through the nidus. Also suggested was a possibility to utilize this in vitro system for research and training on NBCA embolization of AVM.
Note Added in Proof
A part of this study was financially supported with a national grant for scientific research from Japanese government in the fiscal years 2000 and 2001.
A part or full of this study was presented at the 31st Meeting of Japanese Society of Neuroradiology, February 20-22, 2002, Akita, Japan, and also at the 17th Symposium Neuroradiologicum, August 18-24, 2002, Paris, France.
References
- 1.Isoda H, Ramsay RG, et al. MR angiography of aneurysm models of various shapes and neck sizes. Am J Neuroradiol. 1997;18:1463–72. [PMC free article] [PubMed] [Google Scholar]
- 2.Kerber CW, Bank WO, et al. Calibrated leak balloon microcatheter: a device for arterial exploration and occlusive therapy. AM J Roentgenol. 1979;132:207–212. doi: 10.2214/ajr.132.2.207. [DOI] [PubMed] [Google Scholar]
- 3.Kerber CW, Flaherty LW. A teaching and research simulation for therapeutic embolization. Am J Neuroradiol. 1980;1:167–169. [PMC free article] [PubMed] [Google Scholar]
- 4.Debrun GM, Vinuela FV, et al. Two different calibrated leak balloons: experimental work and applications in humans. Am J Neuroradiol. 1982;3:407–414. [PMC free article] [PubMed] [Google Scholar]
- 5.Bartinsky WS, O’Reilly GV, et al. High flow-rate arteriovenous malformation model for simulated therapeutic embolization. Radiology. 1988;167:419–421. doi: 10.1148/radiology.167.2.3357949. [DOI] [PubMed] [Google Scholar]
- 6.Kerber CW, Hecht ST, Knox K. Arteriovenous malformation model for training and research. Am J Neuroradiol. 1997;18:1229–1932. [PMC free article] [PubMed] [Google Scholar]
- 7.Gailloud P, Muster M, et al. In vitro models of intracranial arteriovenous fistulas for the evaluation of new endovascular treatment materials. AM J Neuroradiol. 1999;20:291–295. [PMC free article] [PubMed] [Google Scholar]
- 8.Park S, Yoon H, Suh DC, Song HY. An arteriovenous malformation model for testing liquid embolic materials. Am J Neuroradiol. 1997;18:1892–1896. [PMC free article] [PubMed] [Google Scholar]
- 9.Suga T, Akamatsu T, et al. Actual behaviour of n-butyl-2-cyanoacrylate (Histoacryl) in a blood vessel: a model of the varix. Endoscopy. 2002;34:73–77. doi: 10.1055/s-2002-19384. [DOI] [PubMed] [Google Scholar]
- 10.Brothers MF, Kaufmann JCE, et al. n-Butyl-2-cyanoacrylate - substitute for IBCA in interventional neuroradiology: histopathologic and polymerisation time studies. Am J Neuroradiol. 1989;10:777–786. [PMC free article] [PubMed] [Google Scholar]
- 11.Cromwell LD, Kerber CW. Modification of cyanoacrylate for therapeutic embolization: preliminary experience. AM J Roentgenol. 1979;132:799–801. doi: 10.2214/ajr.132.5.799. [DOI] [PubMed] [Google Scholar]
- 12.Nornes H, Grip A. Haemodynamic aspects of cerebral arteriovenous malformations. J Neurosurgery. 1980;53:456–464. doi: 10.3171/jns.1980.53.4.0456. [DOI] [PubMed] [Google Scholar]