A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications

Ioannis Sechopoulos 1,a)
PMCID: PMC3548896  PMID: 23298127

Abstract

Many important post-acquisition aspects of breast tomosynthesis imaging can impact its clinical performance. Chief among them is the reconstruction algorithm that generates the representation of the three-dimensional breast volume from the acquired projections. But even after reconstruction, additional processes, such as artifact reduction algorithms, computer aided detection and diagnosis, among others, can also impact the performance of breast tomosynthesis in the clinical realm. In this two part paper, a review of breast tomosynthesis research is performed, with an emphasis on its medical physics aspects. In the companion paper, the first part of this review, the research performed relevant to the image acquisition process is examined. This second part will review the research on the post-acquisition aspects, including reconstruction, image processing, and analysis, as well as the advanced applications being investigated for breast tomosynthesis.

Keywords: breast tomosynthesis, digital mammography, breast cancer, review, image reconstruction, image processing, image analysis, contrast enhanced imaging, radiotherapy, elastography, phase contrast

INTRODUCTION

The companion paper provided a review of the medical physics research performed on all aspects of the digital breast tomosynthesis (DBT) image acquisition process. This paper will review all relevant post-acquisition aspects, including image reconstruction, artifact reduction, and other image processing and analyses, in addition to reviewing the research performed on advanced applications and implementations of DBT, including multimodality imaging, in which DBT is combined with other imaging technologies.

TOMOSYNTHESIS RECONSTRUCTION

An in-depth overview of the history and principles of tomosynthesis reconstruction algorithms has been previously published by Dobbins and Godfrey,1 so these aspects will not be discussed here. This discussion will be focused on the different reconstruction methods currently in use or under development in DBT, and the many improvements for each that have been proposed lately.

Filtered backprojection

The most commonly known analytical algorithm for tomosynthesis reconstruction is filtered backprojection (FBP), normally used in computed tomography (CT). As in CT, the choice of filters applied to the projection data before the backprojection is performed can drastically affect the quality of the reconstruction. Mertelmeier et al. presented a general theory for FBP applied to DBT with linear (or arc) x-ray tube motion, in which, in addition to the usual ramp-type and apodization filters, the authors introduce a slice thickness filter that dampens the impact of the incomplete sampling of the frequency space due to the limited angular range used in tomosynthesis, therefore controlling the impact of out-of-plane artifacts on image quality.2 By deriving, validating and using a three-dimensional (3D) cascaded linear system model for DBT, Zhao et al. determined, among other results, that the use of this slice thickness filter reduces aliasing and improves 3D DQE.3, 4 Using task-based analysis, Wang et al. showed a small improvement in mass detectability in the central slice of the lesion, but qualitatively found a decrease in out-of-plane artifacts.5 Orman et al. introduced the concept of an additional filter to avoid the zeroing out of the low frequencies by the ramp filter, thereby introducing an undesirable “flatness”6 to the image.7 The resulting modification of the ramp-type filter was clearly shown by Zhou et al. (Fig. 1).8

Figure 1.

Figure 1

Modification of ramp filter to maintain the low frequency information in FBP reconstructed DBT images. Reprinted with permission from Zhou et al., “A computer simulation platform for the optimization of a breast tomosynthesis system,” Med. Phys. 34(3), 1098–1109 (2007). Copyright © 2007, American Association of Physicists in Medicine (AAPM).

Iterative reconstruction algorithms

Attempting to combine the benefits of an iterative reconstruction method [the simultaneous iterative reconstruction technique (SIRT); Ref. 9] with the speed of FBP, Ludwig et al. estimated the impulse response of SIRT reconstruction in DBT and used it to develop a filter in the frequency domain that could be combined with the other filters used to preprocess the acquired projections before backprojection in FBP.10 This filter, which was found to be weakly projection angle dependent, could be fit to a fourth-order polynomial, yielding a smooth function. The authors found that the use of this filter with FBP results in an important modification of the characteristics of the reconstructed images. Whereas the FBP images present a reduction of the low frequency contrast, resulting in decreased visualization of large area contrast of dense tissue, the images with this iterative-based filter have an overall feel more similar to standard mammograms. In contrast, however, the excellent skin line visualization and sharpness of FBP is somewhat lost with the application of this filter.

In a study relevant to any DBT reconstruction algorithm involving backprojection, Chen et al. investigated the impact of the x-ray tube moving in an arc during acquisition while using a traditional shift-and-add (SAA) type algorithm for backprojection, in which the pixels are shifted only in the direction parallel to the x-ray tube motion.11 As shown by the authors, the slight but non-negligible shift in the projection of each voxel in the direction perpendicular to the tube motion needs to be taken into account to avoid blurring of fine features, which is especially important in the characterization of morphology of microcalcifications.

In a different implementation of FBP, Claus et al. proposed that the pre-backprojection filtering be performed taking into account the size of the features included in the projections, with dimensionally dependent filtering. This method of filtration, combined with an order statistics based algorithm for out-of-plane artifact reduction, was shown to yield results similar to iterative reconstruction methods but with a substantial decrease in computation time.12

Various algebraic methods have been investigated for the general problem of reconstruction. In general, these methods set up a system of simultaneous linear equations, and differ in how this linear algebra problem is solved using iterative methods, such as the algebraic reconstruction technique (ART),13 the SIRT,14 and the simultaneous algebraic reconstruction technique (SART).15 Zhang et al. implemented the SART method for DBT imaging and compared it to standard backprojection (BP) using both homogeneous and structured phantoms.16 While BP performed similarly or better than SART for some lesions in one of the homogeneous phantoms tested, for the more realistic structured phantom the latter yielded better results, both in terms of in-plane image quality and vertical resolution. Lu et al. have proposed two different methods of noise regularization for SART to improve microcalcification visibility without affecting mass detail visibility: one based on selective diffusion17 and, more recently, another based on wavelet decomposition for multiscale regularization.18

A different algebraic reconstruction method, matrix inversion tomosynthesis (MITS), was introduced for DBT by Chen et al.19 MITS was shown to be good at removing out-of-plane artifacts and handling high frequency information, but was also shown to perform poorly with low frequency information.20 To ameliorate this, Chen et al. proposed a hybrid reconstruction algorithm that incorporates the high frequency response of MITS and the good low frequency behavior of FBP, and denoted it Gaussian frequency blending (GFB).20 With GFB, the authors were able to obtain reconstructed images with improved representation of high frequency signals, e.g., microcalcifications, with enhanced representation of the overall breast tissue characteristics, avoiding the commonly found “flat” breast image in which large area density information is lost.

Other iterative reconstruction methods consisting of two steps per iteration, in which the tomosynthesis acquisition process is modeled in a forward step and the reconstructed object is updated in a backward step, have also been proposed for DBT. The most commonly studied method in DBT is the maximum likelihood expectation maximization method (MLEM) Ref. 21 introduced for DBT by Wu et al.22

With most iterative reconstruction methods, the computer processing time required to perform the reconstruction is a concern. Wu et al. discussed that after two to three iterations detailed features are conspicuous, and that radiologists communicated satisfaction with the results after eight to ten iterations, with no further improvement in image quality after ten iterations.22 In early developments, a method to accelerate the MLEM reconstruction algorithm was proposed by Wu et al. in which the reconstructed volume is broken up into oblique sections that can be reconstructed by separate threads of a multiprocessor computer or multiple nodes of a computer cluster.23 This method was then modified for implementation with graphics processing unit (GPU)-based hardware.24 With the advances in this type of hardware, including an ever-increasing number of computing cores and amount of memory,25 the reconstruction time has continued to drop, rendering the many “tricks” needed to avoid the memory issues encountered in the early implementations unnecessary. Implementations of some iterative algorithms optimized for running on a workstation-level GPU card can now just take a few seconds per iteration for a complete patient breast case.

In studying the order in which the acquired projections are processed in each iteration by iterative methods, Wu et al. found that the use of sequences other than that matching the acquisition order can substantially improve the resulting reconstruction quality.26 In this study, the authors used a modified ML-convex algorithm that uses pairs of projections at a time to modify the reconstructed volume, and they found that all three different ordering schemes investigated result in image quality improvement compared to the conventional ordering.

Novel reconstruction methods

Many additional novel reconstruction algorithms have been proposed for DBT with varying levels of testing, success, and continued development. Given their infancy, most of these algorithms are not yet used in any system that is commercially available or close to this stage. Among the proposed algorithms, there are algorithms based on total variation regularization,27, 28 total p variation regularization,29, 30, 31 joint entropy regularization,32, 33 adaptive diffusion regularization,34 iterative penalized maximum likelihood (especially to improve microcalcification visibility),35 iterative maximum a posteriori statistical reconstruction,36 and Bayesian inference.37 In a very different approach, Wu et al. proposed a reconstruction algorithm that results in spherically symmetric “voxels” as opposed to the traditional cubic voxel.38 Another algorithm proposes a spectral reconstruction, in which the polychromatic nature of the x-ray beam is taken into account by inputting the spectrum used for acquisition and the algorithm accounts for the variation in tissue attenuation with x-ray energy and beam hardening.39

Comparison of reconstruction algorithms

Many studies have focused on comparing the capabilities of the various reconstruction algorithms proposed for DBT. The two earliest studies to perform a comparison of reconstruction algorithms were published by Suryanarayanan et al.40, 41 In the first study, using a custom-made structured phantom, the authors compared the acquired projections to the reconstructions obtained with four reconstruction algorithms: tuned-aperture computed tomography (TACT)-backprojection, iterative TACT, an expectation maximization iterative method and a Bayesian-based iterative method.40 For the comparison, Suryanarayanan et al. performed a contrast-detail (CD) study with five radiologists specializing in mammography. The authors found that all reconstructed images result in better CD characteristics than the projections, and that the reconstructions from projections acquired with higher exposure are better than those acquired with lower exposure. CD characteristic comparison among the tomosynthesis reconstruction methods was not performed. In the follow-up study in which the projections were compared to the TACT-backprojection algorithm and the nonlinear TACT-maximization and TACT-minimization algorithms, the authors again found significantly better CD characteristics for all tomosynthesis reconstructions compared to the projection images.41 Among the reconstructions, the TACT-backprojection was only marginally statistically better than the TACT-maximization, with all other algorithm differences not being statistically significant. In another comparison study, Wu et al. tested the standard BP, filtered backprojection and the iterative MLEM reconstruction methods.42 Using phantoms and patient images, the authors found that the BP algorithm resulted in the best in-plane image quality of low contrast, larger features, but suffered from out-of-plane artifacts. The FBP algorithm, as expected, performed better than BP for high frequency features. Finally, the MLEM algorithm provided a good balance of image quality between the low and high frequency features. A comparison of the noise power spectra (NPS) of the SAA, FBP, and MITS algorithms by Chen et al. showed MITS performed best at middle frequencies, while FBP did better at high frequencies, due to the Hamming and Gaussian low pass filters used.43 Zhou et al. compared the detectability of a low contrast lesion in a simulated breast reconstructed with the EM, SART, and BP algorithms with five human observers and found that the EM algorithm had the superior detectability, with SART being somewhat inferior and BP being very inferior.44 Rakowski et al. compared the BP algorithm, with and without an iterative subtraction (IS) algorithm, and an algebraic reconstruction (AR) algorithm.45 Using objective metrics for image quality, the authors found that the BP with the IS algorithm resulted in the highest SNR, but the AR algorithm resulted in the highest normalized signal difference and the best ASF. In another comparison study, Zhang et al. compared the BP, SART, and MLEM algorithms using breast phantoms.46 For both iterative methods, the authors used the result of the BP reconstruction as the initial guess of the reconstructed volume, to improve computational efficiency. When using homogeneous phantoms, the BP algorithm resulted in reconstructions with lower noise and higher CNR, while the iterative methods were superior in all other image quality metrics studied. In particular, with clinically more relevant structured phantoms, BP was again found to lack the ability to reduce out-of-plane artifacts, while both SART and MLEM did reduce these artifacts. In general, SART and MLEM resulted in very comparable image quality, but SART was found to require fewer iterations than MLEM to achieve these results. Zhou et al., as part of a larger study, compared FBP and MLEM using computer simulations and concluded that although MLEM consistently outperformed FBP, the latter can achieve comparable image quality to the former with an optimized acquisition geometry and reconstruction filter.8 Bliznakova et al. used monochromatic synchrotron radiation to image a custom-made structured breast phantom and compared four reconstruction methods: the multiple projection algorithm (MPA, equivalent to BP),47 FBP with a ramp filter and with a ramp and hamming window, and the MPA with noise mask subtraction algorithm48 (MPA-NM).49 Overall, the MPA-NM method was found to be superior due to it being either better than or equivalent to the other algorithms in all tests conducted. By comparing the impulse response of SAA, BP, and FBP, Balla et al. found that BP and FBP performed similarly and better than SAA.50 Using a dedicated breast CT image of an uncompressed patient breast, Van de Sompel et al. simulated a tomosynthesis acquisition and compared the reconstruction obtained with the FBP, SART, and MLEM.51 As metrics, Van de Sompel et al. used the CNR and a numerical observer to estimate the area under the receiver operating characteristic (ROC) curve for detection of a lesion. Among other results mentioned in other sections of this review paper, Van de Sompel et al. performed a comprehensive analysis of the impact of the different reconstruction parameters relevant to each algorithm on the final image quality. The authors again confirmed that SART converges more quickly than MLEM, while the noise amplification of MLEM is lower than that of SART. Overall, SART outperforms MLEM and FBP when tested with noiseless simulations, while MLEM obtains higher metrics in the simulations with noise present.

Reconstruction artifacts

Given the very limited sampling of the frequency domain involved in DBT acquisition due to the narrow angular range of the x-ray source motion, this imaging modality is very susceptible to artifacts of various kinds. Most common among these are the presence of shifted and repeated versions of high-contrast features located in adjacent planes. Various new methods and modifications to existing reconstruction algorithms have been proposed to reduce or eliminate this and other types of artifacts that appear commonly in DBT.

Wu et al. proposed an addition to the MLEM algorithm to reduce the presence of artifacts due to high contrast signals present outside the in-focus plane (Fig. 2).52 In this algorithm, four methods of detecting which projections contain the appropriate information to contribute to the formation of each voxel were tested. The projections that are determined to contain “abnormal” information from out-of-plane artifacts are not used to update the reconstructed voxel. Of the four different “voting” methods investigated to achieve this identification, it was determined that their performance varies depending on the type of artifact present. The combination of two of these methods, the “projection segmentation” and “one-step classification” methods, results in an algorithm that can successfully remove artifacts due to both large and small calcifications. Ge et al. also proposed an algorithm to reduce the artifacts introduced by high contrast objects (large or dense microcalcifications, metal inserts) by estimating the shift variant impulse response of the system, applying it to the isolated object, and subtracting the resulting estimated artifact from the reconstruction.53, 54 Another method to remove artifacts due to calcifications was proposed by Erhard et al., in which a hysteresis thresholding method is applied to the initial reconstruction to isolate the microcalcifications, which are then removed from the projections by interpolation and the results reconstructed.55 Finally, the identified microcalcifications are added back into the reconstruction at their estimated locations. Sun et al. developed an algorithm to reduce the visibility of out-of-plane signals using a 3D nonlinear anisotropic diffusion filter that showed promising results.56

Figure 2.

Figure 2

Typical reconstruction artifacts found in DBT from out-of-plane high contrast objects (left) and the result of the correction proposed by Wu et al. (right). These examples show artifacts from a localization needle (top) and large microcalcifications (bottom). Reprinted with permission from Wu et al., “Voting strategy for artifact reduction in digital breast tomosynthesis,” Med. Phys. 33(7), 2461–2471 (2006). Copyright © 2006, American Association of Physicists in Medicine.

Zhang et al. proposed the use of a breast boundary detection method not only to remove the artifacts commonly seen in the periphery of the reconstructed breast, but also to increase the reconstruction speed (Fig. 3).57 By performing breast boundary detection, the SART reconstruction can be performed only within the volume occupied by the breast, considerably decreasing the reconstruction time. In addition, the inclusion of only the volume occupied by the breast in the reconstruction allows for the removal of the artifact in the breast periphery, which appears due to the breast projection covering one area of the detector in some projection angles and not covering it for other projection angles.

Figure 3.

Figure 3

Result of the correction proposed by Zhang et al. to remove the breast boundary artifacts (arrows) from DBT reconstructions. Reprinted with permission from Zhang et al., “Application of boundary detection information in breast tomosynthesis reconstruction,” Med. Phys. 34(9), 3603–3613 (2007). Copyright © 2007, American Association of Physicists in Medicine (AAPM).

Other common artifacts in DBT are truncation artifacts close to the detector edges due to the finite size of the detector and x-ray beam. The finite size of these items causes two different phenomena, described in Fig. 4, that result in bright horizontal lines and a bright area, as shown in Fig. 5 (left). Different algorithms have been proposed to correct these artifacts, showing promising results. To correct the horizontal line artifact due to wider projections not including portions of tissue (right gray area in Fig. 4), Li et al. proposed an algorithm for FBP,58 actually tested on chest tomosynthesis images but relevant to DBT. In this algorithm, the number of rays that pass through each voxel in the volume being reconstructed is counted, resulting in a 3D “count map.” This map is then used to weigh the corresponding intensities of the backprojected image. Zhang et al. corrected these artifacts when applied to SART reconstructions (Fig. 5).59 To correct the same artifact as that addressed by Li et al., the algorithm proposed by Zhang et al. updates the voxels in the reconstructed volume that are not covered by the projection being processed with the mean value of the neighboring region that is covered by the projection. To correct the artifact that appears as a bright area due to the tissue present outside the reconstructed field of view (left gray area in Fig. 4), Zhang et al. proposed to first detect which rays have traveled through tissue, rather than air, outside the reconstructed volume using a predetermined threshold. Upon identification of these rays, their contribution to the reconstruction is modified by assuming that the tissue outside the reconstructed volume consists of breast tissue with “average” attenuation. As can be seen in Fig. 5, the algorithms result of Zhang et al. in a correction of the artifacts, although in some cases the second algorithm over-compensates for the attenuation outside the reconstructed volume, resulting in a dark area in the final result [white arrow in Fig. 5 (right)]. In a recent follow up study, Lu et al. of the same group improved these correction methods and have been able to remove the dark area artifact in their processed images by refining the method used to estimate the path length traveled by the identified problematic rays within the breast tissue included in the reconstructed volume.60

Figure 4.

Figure 4

Diagram of a typical DBT acquisition showing the source of two reconstruction artifacts. The gray areas show portions of the breast tissue that cause artifacts near the borders of the reconstructed volume (dashed box) in the x-ray tube motion direction. Examples of both of these artifacts can be seen at the top of the left image in Fig. 5; the bright area is due to the gray portion of breast tissue on the left of the diagram, while the horizontal stripes are due to the gray portion of breast tissue on the right.

Figure 5.

Figure 5

Result of the algorithm proposed by Zhang et al. to reduce the detector boundary artifact commonly seen in DBT reconstructions. Reprinted with permission from Zhang et al., “Artifact reduction methods for truncated projections in iterative breast tomosynthesis reconstruction,” J. Comput. Assist. Tomogr. 33(3), 426–435 (2009). Copyright © 2009, Lippincott Williams & Wilkins.

Reconstructed slice thickness

Tomosynthesis reconstruction is normally performed to yield in-focus slices 1 mm apart or 1 mm thick slices. Which of these two theoretical descriptions is applicable depends on the reconstruction algorithm used. Diekmann et al. studied three different methods to combine ten 1 mm slices together to improve image quality mainly by reducing noise.61 The algorithms tested in their work were: maximum intensity projection (MIP), simple averaging, and a novel algorithm called softMIP, which provides a weighted average of the pixels being combined. Using objective metrics, the authors found that MIP works best for microcalcifications, while averaging results in the best image quality for masses, with softMIP performing in between for both types of lesions. In addition to image quality improvement, merely reducing the number of slices that require interpretation while maintaining image quality would be a great contribution to DBT, given the concern for increased reading time of screening DBT images compared to mammography. Van de Sompel et al. also compared reconstructing thicker slices with SART and ML with post-reconstruction binning of thin slices, although the method for this binning was not specified.51 The authors found that post-reconstruction binning results in improved image quality compared to reconstructing thicker slices, with a maximum improvement being achieved at different thicknesses, depending on the reconstruction algorithm used. In this study, however, only masses were investigated, so the impact of binning on microcalcification visibility was not addressed.

Geometric calibration

Except for TACT-based reconstruction, all reconstruction algorithms rely on an accurate and complete understanding of the geometry used to acquire the projections to generate an optimal reconstructed volume. As determined by Mainprize et al., of all the possible errors in acquisition geometry input, the only fixed error (same offset present in all projections) that affects image quality is detector yaw (relative angle between detector edge and x-ray source motion).62 However, random variations of all kinds during the acquisition process result in both a loss of contrast and error in lesion localization.

To determine the geometrical parameters with ease and accuracy, Li et al. proposed to perform geometrical calibration using a phantom with a number of high contrast spherical markers with known locations.63 From acquisition of projections of this phantom, matrix decomposition can yield the projection matrix that relates the volumetric locations to the projection locations. From this projection matrix, the x-ray source and central ray positions can be determined. In this work, Li et al. also showed the sensitivity of small detail to an accurate knowledge of the geometry, by introducing simulated errors in the geometric information and reconstructing a phantom with microcalcifications. In a follow-up study, Li et al. determined that, in practice, reconstruction errors due to sensitivity to the accuracy in the geometric calibration could be avoided by using many more markers than the minimum theoretically required (6), and by distributing them so that a large portion of the imaging volume is covered.64

IMAGE ANALYSIS AND PROCESSING

Synthetic mammograms

Currently, the single DBT system approved for clinical use in the U.S. by the Food and Drug Administration (FDA) is labeled for acquisition of a tomosynthesis image only combined with a mammographic image. Of course, this results in a radiation dose penalty of approximately double the dose of a single mammogram or tomosynthesis acquisition.65 The need, correct or not, to always acquire a mammogram along with a DBT image results mainly from the suspicion that microcalcification detectability in DBT is inferior to that in mammography. Of the many studies comparing microcalcification detectability, some have found this inferiority,66, 67, 68 while others have not.69, 70, 71, 72, 73 To avoid this penalty, research is ongoing in creating a synthetic mammographic image from a tomosynthesis acquisition. In the only study on this matter, Gur et al. reported on an observer study to compare the performance of DBT combined with either an actual mammogram or a synthetic mammogram (Fig. 6).74 With 114 cases and 10 readers, the authors found that combining the DBT image with the synthetic mammogram resulted in a loss of sensitivity with equivalent specificity. Therefore, the synthetic mammograms from tomosynthesis data used in that study were not capable of replacing actual mammographic acquisitions. However, the algorithms to synthesize these images are still being developed, and preliminary (unpublished) evaluations of more recent synthetic mammograms have shown substantial improvement.75

Figure 6.

Figure 6

(a) Digital mammogram (MLO view) of a patient showing two malignant masses, (b) synthetic mammogram constructed from the DBT data, (c) DBT slice. Reprinted with permission from Gur et al., “Dose reduction in digital breast tomosynthesis (DBT) screening using synthetically reconstructed projection images: An observer performance study,” Acad. Radiol. 19(2), 166–171 (2012). Copyright © 2012, Association of University Radiologists (AUR).

Texture and glandular tissue analysis

Although not fully three-dimensional as dedicated breast CT images, DBT images do provide more information on breast tissue localization than mammography. The utility of this additional information for various tissue analysis aspects has been investigated. In two separate studies, Kontos et al. reported on the use of DBT images to determine any correlations between parenchymal texture features and breast glandular areal density or risk estimates.76, 77 In the initial study, Kontos et al. first analyzed the correlation between several texture feature metrics (skewness, coarseness, contrast, energy, homogeneity, and fractal dimension) from DBT and mammography images to the Gail and Claus risk models, finding overall low correlation.76 However, a significant correlation was found between some of the texture feature metrics and glandular density, both with mammography and DBT, with the latter resulting in stronger correlations. This finding was repeated in the follow-up study with a larger number of patient cases.77 It seems that being able to perform the analysis of the parenchymal texture without the effect of overlapping tissue present in mammograms could provide more reliable information on characteristics that may be related to breast cancer development risk. It should be noted, however, that in another study Kontos et al. found that the values of different feature metrics may vary with acquisition parameters (angular range, number of projections). So although this allows for optimization of acquisition parameters for this type of texture analysis, it also means that care should be taken to note the source of images used for feature analysis and interimage comparisons.78

Engstrom et al. studied tomosynthesis projections and reconstructions of 55 patient breasts to determine the power law coefficient (β in the equation PS(f) = k/fβ) associated with the power spectrum of normal breast tissue background.79 In this study, the authors found that β for the tomosynthesis projections had a mean of 3.06, while for the reconstructions the value of β was 2.87, with the difference being statistically significant. In a more recent study, Chen et al. found that the mean values of β of 23 patients in dedicated breast CT, DBT, and mammography images were 1.75–1.83 (3 orthogonal views in breast CT), 3.06–3.10 (CC and MLO in DBT), and 3.17–3.30 (CC and MLO in mammography).80 Although small, the difference in β between tomosynthesis and mammography was statistically significant for all comparisons but one (MLO from tomosynthesis and CC from mammography). Hu et al., however, found that the values of β for the tomosynthesis reconstructions could vary substantially depending on the filters used during FBP reconstruction.81 In previous studies, a smaller value of β has been related to improved mass detectability.82

Bakic et al. compared the breast glandular areal density estimation from mammograms and the central (0°) tomosynthesis projections using previously validated semiautomated density estimation software.83 Although the acquisition geometry of these two images is essentially equal, two major differences negatively impact the image quality of the central tomosynthesis projection compared to the corresponding mammogram. In the first place, the tomosynthesis projection will be considerably noisier, given that the exposure used to acquire it is only a fraction (in this case an average of 22%) of that used for mammogram acquisition. In addition, the overall contrast can be expected to be lower, since currently DBT is acquired without an x-ray scatter grid, and therefore the full impact of the scatter signal on the image contrast is included in the tomosynthesis projection. To perform this comparison, three experienced independent readers used a density estimation software (CUMULUS, version 4.0; University of Toronto, Toronto, Canada)84 involving several manual steps. Each reader obtained a glandular density estimate from all 39 cases twice, with each repeated analysis separated by 2 months. Although the density estimation results obtained by Bakic et al. between the mammography images and tomosynthesis projection images were highly correlated (ρ = 0.91), the latter-based estimations were slightly but significantly higher.83 In addition, the intrareader correlations were slightly, but not significantly, higher for the mammogram-based analysis.

Tagliafico et al. compared the density estimation from mammograms to the average of the estimation of all 15 tomosynthesis projection images acquired in 50 patients.85 Using two different methods, the authors found that the density estimation from mammograms was significantly higher than from tomosynthesis projections. Given that no independent truth is available, it is unclear if mammography overestimated or tomosynthesis underestimated the density. In another study, Bakic et al. compared the mammographic density estimation to the 3D density estimation from DBT using a semiautomated thresholding technique and found moderate agreement.86 They also found that, on average, DBT results in lower density estimates than mammography.

Although several classification algorithms to identify the different breast tissues present in dedicated breast CT images have been developed,87, 88, 89, 90, 91 classification based on DBT images was first proposed by Kontos et al.92 and a method completely developed by Vedantham et al.93 Kontos et al. investigated the use of different texture features (e.g., skewness, coarseness, etc.) to identify regions in DBT reconstructions as dense or fatty. The authors hypothesized that image classification could be achieved using this type of region-based analysis. To test the feasibility of their approach, Kontos et al. performed ROC analysis on the ability of these different metrics to differentiate between the two tissue types, finding that for DBT reconstructions the fractal dimension results in the highest area under the ROC curve. Vedantham et al. on the other hand used a fuzzy c-means based segmentation method, among other processing steps, to perform the tissue classification.93 The method proposed by the authors is capable of classifying breast skin, adipose, and glandular tissues, in addition to muscle and, with some input from the user, suspicious lesions. Such a classification method could be useful for reconstruction, computer aided detection (CADe), breast glandular volumetric density, and registration, which may be useful for comparison of current acquisitions with priors.

Image evaluation with numerical observers

The use of numerical observers in medical imaging research is advantageous since it allows for comparison of image quality for specific tasks which include a very high number of cases and/or imaging parameters without undergoing costly or unfeasible human observer studies. However, for tomosynthesis research, the use of numerical observers is currently challenging due to a lack of models that incorporate the additional pseudo-3D information provided in this imaging modality. As an example of this limitation, some optimization studies, some of which have been mentioned above, have used 2D numerical observers to analyze only the slice that crosses the center of the signal of interest. Although this may be the case, it has not been shown that optimization based on this single tomosynthesis slice perfectly correlates with one based on the entire or portion of the reconstructed volume. It could be expected that for lesions that span more than a single slice, the quality of the representation of that lesion throughout the entire lesion volume does not necessarily correlate perfectly with the quality of the lesion presentation of the center slice. Young et al. proposed to use a numerical observer model to analyze the tomosynthesis projections, not the reconstructions, incorporating into the model the spatial correlations present due to acquisition of the same volume from different projection angles.94 The authors showed that the results of the observer model analysis of the projections are different if these spatial correlations are taken into account. By performing the analysis on the projection space rather than the reconstruction space, the authors suggest that the results obtained provide an upper-bound on the performance results, without introducing additional variability from the reconstruction algorithm methods and parameters. Park et al. proposed a 3D ideal observer model that can be applied to tomosynthesis reconstructed images to analyze lesion detectability.95 A study comparing the 3D observer model to the 2D model, applied both to the slice including the lesion and in projection space, would be of great interest.

Gang et al. derived a 2D detectability index from a 3D cascaded systems analysis and compared its predictions to human observer performance for lesion detection with varying acquisition parameters.96 The authors found reasonable correspondence between model predictions and observer performance, allowing for the use of their developed model for optimization in 3D imaging such as DBT.

Computer aided detection of masses

The use of CADe in digital mammography has become common practice in the clinical realm. It could be expected that similar systems, if available, would be adopted for DBT, especially considering the larger amount of data presented to the reader in a DBT image compared to a mammogram. In an early study, Reiser et al. proposed a CADe method for mass detection based on detection and segmentation of suspicious lesions using a radial gradient index method.97 Inclusion of the vertical direction information did not improve classification performance based on gradient analysis, but it did improve performance based on shape analysis. Overall, the sensitivity for the proposed algorithm was 76% with 11 false positives per exam. In another early study, Chan et al. also proposed a CADe method for mass detection, which they tested on 26 patient DBT cases.98 The method, which involved interpolation of the reconstructed image to obtain cubic voxels with 0.1 mm sides, gradient field analysis to identify candidate lesions, segmentation via 3D region growing, and feature analysis, resulted in an 85% sensitivity with 2.2 false positives per exam and 80% sensitivity with 2.0 false positives per exam, depending on the operating point setting. In a follow-up study, Chan et al. investigated how the performance of this algorithm was affected by the number of projection views and the dose level used to acquire the DBT image.99 For this, they analyzed a set of patient images reconstructed by using either all 21 projections acquired or only 11 of the 21 projections (which resulted in about half the radiation dose). Although the performance of the classifier was unaffected by the lower number of projections and dose, the overall area under the ROC curve was lower for the 11 projection images, due to an increase in the false positive rate.

In a different approach, Reiser et al. proposed to perform automated mass detection by analyzing the acquired tomosynthesis projection images as opposed to the reconstructed images.100 By using the projections, the authors argue that any artifacts introduced by the reconstruction algorithm are avoided. In their method, only an intermediate step involves analysis in reconstruction space, in which the algorithm attempts to identify and eliminate the suspicious lesions due to tissue superposition. The initial lesion identification and final lesion classification is performed in projection space. With their limited data set, the authors found a sensitivity of 90% with 1.5 false positives per exam. Chan et al. later compared the performance of CADe for mass detection under three conditions: when the algorithm performs all analysis on the reconstructed images only, when the algorithm performs the lesion identification and classification on the projection images and then applies a decision threshold on the distribution of mass likelihood scores obtained from the projections and backprojected to reconstruction space, or, finally, when the algorithm combines the likelihood scores of both methods described above to arrive at a final decision.101 The authors found that taking advantage of the most information available, i.e., using the approach that combines the 2D and 3D-based results, yields the highest performance, with a false positive rate of 1.23 and 2.04 per exam at sensitivities of 80% and 90%, respectively. Singh et al. also proposed to perform the initial lesion identification in projection space with the rest of the algorithm performed in reconstruction space.102 As opposed to the previous algorithms proposed, however, the lesion classification performed by Singh et al. was not feature-based, but used information theory-based analysis with mutual information as the metric of choice and three different ways of constructing a knowledge base for training of the system. The authors found that although there was no statistically significant difference in the performance of the system when trained by either of the three knowledge bases, the inclusion in the knowledge base of true and false positive regions of interest (ROI), identified by the first stage of the CADe process, yielded the highest area under the ROC curve. However, the authors argue that a knowledge base consisting of only true positive and normal ROI may be more robust and generalizable, since it does not depend on the performance and characteristics of the first stage of the system. In a subsequent study, Mazurowski et al. demonstrated that a mutual information based algorithm, trained solely with mammographic images, can perform comparably in detecting masses in DBT images, with a simple preprocessing of the DBT reconstructions in which an area close to the breast edge is masked out.103

Computer aided diagnosis of masses

A computer aided diagnosis (CADx) system for characterization of suspicious masses was developed and tested by Chan et al. on both the tomosynthesis reconstructions and the projections.104 After a ROI was marked by the reader in a reconstructed slice, the location of the ROI in the acquired projections was identified, and the same CADx algorithm with few modifications processed both image spaces independently. The authors found that reconstruction-based performance is significantly higher than projection-based performance, with the area under the ROC curve as the figure of merit.

Computer aided detection and diagnosis of microcalcifications

For automated detection of microcalcification clusters, in an early preliminary study, Peters et al. proposed to use wavelet filtering and fuzzy processing of the acquired projections to avoid performance variations due to different reconstruction algorithms and parameters.105 Reiser et al. also proposed a method that performs the initial suspicious lesion identification and final feature-based classification in projection space, while using reconstruction space in an intermediate step to eliminate suspicious areas that appear in few projections.106 In this preliminary study with a limited data set, the authors obtained a sensitivity of 86% with 1.3 false positives per exam, yielding a poorer performance than current digital mammography CADe algorithms for microcalcifications. In another study, Park et al. studied the use of a modified version of an existing CADe algorithm for microcalcification detection in mammography in DBT.107 For this, the authors added a clustering method to the existing algorithm to determine if the calcifications identified in the acquired projections (when the algorithm was applied to projection space) or reconstructed slices (when applied to reconstruction space) belong to the same cluster. Upon identification of the potential clusters, four different methods to arrive at the final identification were compared. The authors found that the best method for projection based analysis yielded a sensitivity of 70% with 4.0 false positives per exam, while the reconstruction based analysis resulted in 88% sensitivity with a 15.9 false positive rate. Using a reconstruction based CADe method for microcalcifications derived from an algorithm for mammography, Bernard et al. obtained promising results, with a sensitivity of 85% with 1.4 false positives per exam.108 The authors note that this performance level is lower than that obtained with the similar algorithm used for mammography (86% sensitivity with 0.95 false positives per exam), again demonstrating that further work is required for tomosynthesis-based CADe to match the performance in mammography in identifying microcalcifications. Sahiner et al., using an algorithm that processes only the reconstructed images, obtained a sensitivity for microcalcification detection of 85% with 3.8 false positives per exam in an abnormal case set and with 3.4 false positives per exam in a normal case set.109 Finally, in early work, Ho et al. have proposed CADe and CADx of microcalcifications for DBT using epipolar curves.110, 111, 112

Image registration

Several clinical applications for registration of DBT images have been identified and methods for each are being investigated. In the first place, comparison of DBT acquisitions with mammographic priors will be required if DBT replaces mammography as a screening technology. For this, as initial first steps, Bakic et al. performed registration of mammograms with the central (0°) tomosynthesis projection113 and with a set of tomosynthesis projections,114 using a previously described non-rigid registration algorithm, resulting in a correction of ≥90% of the per pixel intensity differences originally existing between the image pairs.

Van Schie et al. have proposed that when interpreting two views of the same breast during DBT screening, either by a human reader or a CAD system, there will be a need to determine corresponding locations in the two views.115 For this, the authors proposed a model in which the breast, compressed for acquisition of the first view, is decompressed, rotated, and re-compressed for acquisition of the second view. By characterizing the translation of locations with this model, the authors developed an analytical solution for the transformation, yielding a fast method to perform this location matching. Using two-view patient cases in which different features were located by hand in each view, the method developed was tested and found to yield a median 3D distance between the actual and predicted location of 14.6 mm, with about half of the predictions being one or fewer slices away from the truth. Zhang et al., in an initial study, developed an automated method to detect corresponding feature locations in two tomosynthesis projection images, which allowed for subsequent non-rigid registration.116 Although the authors tested only registration of projection pairs from acquisition of contra-lateral breasts or from different projection angles of the same breast, the application of the same method for temporal registration would be possible. Since Sihna et al. successfully tested the registration of two DBT images acquired at separate time points by manually identifying matching feature locations,117 it seems that incorporating the feature selection process using an algorithm like that proposed by Zhang et al. could result in a successful fully automated temporal registration method.

In a different approach to the registration of two DBT acquisitions separated temporally, Yang et al. proposed to perform the registration of the two volumes during their reconstruction from the acquired projections, using an iterative approach in which the optimization of the registration transformation between the two volumes is also included in each iteration step.118 Using simulated data from a phantom and from a patient breast MRI acquisition, the authors found improved reconstruction and registration performance when the two tasks are performed combined rather than sequentially.

Quantitative breast tomosynthesis

Due to its very limited angular sampling, it is not apparent that DBT can provide quantitative information on the tissues present in the imaged volume. To investigate the feasibility of quantitative analysis in DBT, Shafer et al. used simulated and physical phantoms with varying layers of tissues and signals with a range of fractions of adipose and glandular tissue mixtures.119 Their investigation involved the application of a background uniformity correction to the acquired images and analyzed how the presence of scatter, variation in background tissue composition, lesion vertical location, and x-ray spectrum used for acquisition impacted the estimated glandular density of the inserts. The authors found promising results for the quantitative potential of tomosynthesis, but the use of overall homogeneous phantoms limits the generalizability of the results, calling for further investigation with more realistic breast phantoms and clinical cases. Richard and Samei developed a model to predict estimation task performance based on frequency based analysis, similar to the detectability analysis commonly used to predict signal visibility, but for tasks involving quantitative analysis, e.g., area and volume estimation.120, 121 This model was used to determine the optimal geometrical acquisition parameters that would maximize the precision of area and volume estimation in DBT reconstructions.

ADVANCED APPLICATIONS

Contrast enhanced breast tomosynthesis

As with contrast enhanced digital mammography, DBT imaging with iodinated contrast enhancement aims to interrogate the vascular characteristics of the imaged breast to improve malignant lesion detectability. Contrast-enhanced image characteristics that could help detect and diagnose malignant breast lesions include degree of enhancement, location of enhancement, and, with multi-acquisition dynamic protocols, enhancement kinetics.122 Given the high in-plane spatial resolution and short acquisition time of DBT compared to dynamic contrast-enhanced breast magnetic resonance imaging (DCE-MRI), it seems that this imaging modality could prove useful in determining enhancement details in terms of both location and time. However, acquisition of enhancement kinetic curves with a high number of time points is a potential limitation for contrast-enhanced DBT (CE-DBT) given the use of ionizing radiation during acquisition. This results in an upper limit of probably three or four time points that can be used for constructing a kinetic curve, even with the most advanced CE-DBT technologies discussed below. In contrast, with DCE-MRI, acquisition of more than ten time points for kinetic curve analysis has been recently reported.123, 124 However, the need for more than a few time points to construct a kinetic curve is still being investigated.125 Therefore, given the high spatial resolution and the high temporal resolution (although with few time points) achievable, CE-DBT has been studied extensively. If the limited number of acquisition time points possible with this imaging modality is a true limitation is a clinical question that seems yet to be answered.

Contrast enhanced breast tomosynthesis: Temporal subtraction

Carton et al. were the first to propose performing CE-DBT.126 In their approach, they used temporal subtraction, in which acquisition is performed before and after injection of the contrast agent, and the resulting reconstructions subtracted from each other. Using empirical data and theoretical modeling, the authors investigated the quantification of iodine concentration and the impact of the presence of x-ray scatter, patient motion, and detector stability. For this, the x-ray spectrum had to be modified to increase the fluence of x-rays above the K-shell of iodine (33.2 keV). Therefore, the authors used a tube voltage of 49 kVp and an added copper filter 0.27 mm thick, resulting in a first HVL of 3.06 mm Al. The authors found that the concentration of iodine could be underestimated due to scatter by 28%–54% for breast thicknesses ranging from 20 to 80 mm, respectively, while detector instability could introduce up to an 8% error in quantification. Patient motion between image acquisitions could result in an additional error depending on the amount of displacement and the iodine concentration in the location of interest. Finally, as expected, patient motion also results in subtraction artifacts visible in the resultant iodine only image. In another study on temporal CE-DBT, Chen et al. imaged 13 patients, of which 11 had malignancies proven by pathology, using light breast compression so as to not affect the flow of blood (and contrast agent) in the breast.127 The authors found that CE-DBT resulted in enhancement of suspicious lesions in 10 of the 11 malignancies and no enhancement in the remaining two benign cases.

Contrast enhanced breast tomosynthesis: Dual energy

To minimize patient motion issues, Carton et al. suggested the use of dual energy subtraction, in which both acquisitions are performed post-contrast injection, with one acquisition using an x-ray spectrum with most or all x-rays below the iodine K-shell energy and the other acquisition with an x-ray spectrum with most or all x-rays above this shell energy, rather than temporal subtraction.126 Subsequent investigations on CE-DBT used this approach. Glick and Didier simulated the acquisition of the high and low energy images of a structured breast phantom using the x-ray spectra produced by a tungsten target with a tube voltage of 60 kVp, filtered with 0.4 mm of cerium, and 0.6 mm of tin, respectively.128 The computer simulations showed that the lowest iodine concentration tested, 1.0 mg I/cm3, could be seen in the subtracted image. In another study Puong et al. also proposed dual energy CE-DBT, but, as opposed to Glick and Didier who first reconstructed the low and high energy images and then performed weighted subtraction, this study proposed to first combine the acquired projections to obtain an iodine only projection set using a novel algorithm developed for contrast enhanced mammography,129 and then reconstruct using the SART algorithm.130 In addition to showing feasibility, Puong et al. performed an optimization of the x-ray spectra to be used for acquisition, and studied the effect of x-ray scatter on the quantification of iodine. In a subsequent optimization study, Puong et al. compared the image quality obtained using different target/filter combinations, and found that a Rh/Rh target/filter combination spectrum with a tube voltage of 27 kVp and a Mo/Cu spectrum with 49 kVp and a 0.48:0.52 dose ratio, respectively, results in the lowest image noise.131 In addition, the authors proposed corrections for the quantification of iodine in the presence of scatter, nonconstant breast thickness at the breast borders, and limited vertical resolution. Later on, this same research group introduced a regularization to the SART reconstruction specific to iodine only reconstructions, showing improved image quality.132

In the first dual energy studies involving patients, Carton et al. imaged patients with suspicious findings using both temporal subtraction and dual energy subtraction.133, 134 For this, after one pre-contrast injection acquisition to perform temporal subtraction with a high energy spectrum (49 kVp, 25 μm Rh + 0.25 mm Cu, 3.36 mm Al HVL), the authors acquired two sets of two images post-injection, each with a low energy spectrum (30 kVp, 25 μm Rh, 0.44 mm Al HVL) and the high energy spectrum. The dual energy approach resulted in less motion artifacts than the temporal subtraction (Fig. 7), which is an expected benefit of the former, but the authors proposed that with adequate motion correction, temporal subtraction may ultimately result in better image quality due to better background subtraction and lower noise. The same research group developed and validated a theoretical model to optimize the acquisition parameters to be used for dual energy CE-DBT with a photon counting, scanning multi-slit imaging system.135, 136 In this study, the authors aimed to perform acquisition of the low and high energy images simultaneously by selectively filtering the x-ray beam incident on adjacent detector slits. This necessitates the use of the same tube target, voltage, and current-exposure time product for both images, allowing only for a change in the beam filtration. Therefore, the authors investigated the optimization of the thicknesses of the two filter materials (tin for the low energy beam and copper for the high energy beam), the number of slits covered by each of the filter materials, and the weight factor wt that is used in the equation to obtain the enhancement-only subtracted image:

IDE=lnICuwtlnISn from Ref .134, (1)

where IDE,ICu,ISn are the subtraction, high energy copper-filtered and low energy tin-filtered images, respectively. As the figure of merit, Carton et al. used the signal difference to noise ratio (SDNR) normalized by the square root of the mean glandular dose, a common metric for this type of optimization study. The investigators found that a 2:1 copper and tin slit assignment is optimal, which can be translated in other system designs to assigning 2/3 of the available total exposure to the high energy image, combined with 0.16 mm tin and 0.23 mm copper filters. For this setup, the optimal weight factor varies with breast thickness and the breast tissue glandular fraction aimed to be suppressed. Of course, using a different weight factor for each breast thickness is not challenging, but it is for varying glandular fraction pairs. Overall, the authors found that a weight fraction of between 0.52 and 0.60, depending on the breast thickness, is optimal. Although the system this optimization model was based on is not being further developed, the methods and results of this investigation are still relevant for dual energy CE-DBT performed with other system designs. In a recent optimization study, Samei and Saunders also investigated the optimal weight factor, exposure distribution between the two images, and the x-ray spectra shape, using computer simulated acquisitions of a breast phantom.137 One important distinction between this work and the work of Carton et al. was that the latter performed the logarithmic subtraction of the reconstructed images, while Samei and Saunders found that for their setup this method introduces artifacts, which are avoided if the subtraction is performed on the projections and the result used for reconstruction. Although Samei and Saunders investigated the use of different tube voltages for the two images, they found that a tube voltage of 49 kVp is optimal for both acquisitions, a very important simplification of the required hardware. In terms of filtration, the optimal thicknesses were found to be 92.5 μm of copper and 95 μm of tin for the high and low energy acquisitions, respectively. For this combination of filters, the optimal weight factor and dose distribution is 0.61 and 0.54 (high energy image dose to total dose fraction), respectively.

Figure 7.

Figure 7

ROI of a contrast enhanced DBT reconstruction of a patient with a malignant mass. (Left) Dual energy subtraction; (right) temporal subtraction. The presence of motion in the temporal subtraction is apparent. Reprinted with permission from Carton et al., “Dual-energy contrast-enhanced digital breast tomosynthesis: A feasibility study,” Br. J. Radiol. 83(988), 344–350 (2010). Copyright © 2010, The British Institute of Radiology.

In a different approach to dual energy acquisition taking advantage of new technology, Schmitzberger et al. reported on a study that used an energy discriminating photon counting multi-slit tomosynthesis system.138 The detector used in this system can discriminate between detected x-rays with energy below and above a certain threshold.139 Of course, for iodinated contrast-enhanced imaging, the threshold is set at the K-shell energy of iodine, and therefore a single DBT acquisition can be performed that still results in a low energy and a high energy image. This guarantees perfect tissue registration, so no subtraction artifacts due to patient motion are possible, and it can result in lower total glandular dose per exam. In this study, the glandular dose delivered per acquisition was 0.42 mGy, and three acquisitions at three different time points were performed for each patient to obtain dynamic data. In phantom imaging with a contrast agent concentration of 3 mg I/cm3, the smallest simulated lesion that could be detected was 5 mm diameter (the next smallest lesions present were 1 and 2 mm diameter). The preliminary patient study included in this work showed promising results.

In a study on how contrast dynamics and acquisition timing affect iodine quantification, Hill et al. found that the tomosynthesis-based results will reflect the average concentration during acquisition, but that a correction factor for different lesion sizes is needed to obtain accurate quantification results.140 In addition, the same research group found that the most accurate quantitative assessment is obtained when both the low and high energy projections are acquired in a single x-ray source sweep with fast tube voltage switching.141

Multimodality imaging

The combination of several imaging technologies with DBT has been investigated. Given that DBT as stand-alone technology has only recently been introduced to the clinical realm, it is not surprising that multimodality approaches are still at a very infant stage. One research group has investigated the combination of DBT with electrical impedance tomography for characterization of suspicious lesions, showing promising results with a small number of patient cases.142, 143, 144 Combining DBT with ultrasound has also been investigated, with the development of a prototype system that acquires the tomosynthesis image and immediately afterwards positions and scans an ultrasound probe across the compressed breast, obtaining a co-registered 3D ultrasound image.145, 146 The combination of morphological information with functional information is also being developed by a group that is investigating combining DBT with limited angle Tc-99m sestamibi-based single photon emission computed tomography (SPECT).147, 148, 149 The preliminary patient study performed showed that adding the functional information can improve the performance of DBT, especially in improving specificity.149 Finally, DBT combined with diffuse optical tomography for interrogation of the hemoglobin and oxygen saturation state has been also investigated, with the optical characteristics of malignant lesions having been found to be different than benign lesions and normal tissues.150, 151, 152 Therefore, it seems that diffuse optical tomography could provide useful additional information to help characterize lesions with DBT.

Phase contrast tomosynthesis

Phase contrast imaging has been investigated for various modalities due to its potential to improve contrast and enhance feature edges.153, 154, 155 Given the promising results for applications such as radiography, mammography and CT, the possibility of performing phase contrast DBT is intriguing. Using a benchtop system, Hammonds et al. developed and tested the combination of phase contrast radiography with tomosynthesis imaging on phantoms.156 In their study, the authors determined that this edge enhancement is retained during tomosynthesis reconstruction, therefore making phase contrast DBT feasible and promising.

Tomosynthesis elastography

From ultrasound imaging, it is known that benign and malignant breast lesions exhibit differences in their elasticity. To assess elasticity or differences in elasticity, one theoretically simple method especially applicable to DBT could be to measure strain, in this case defined as the change in lesion size in one direction under varying levels of compression force. To investigate this potential new application for DBT, Kim et al.157 and Engelken et al.158 proposed to perform elastography using a DBT system by acquiring subsequent images of the breast undergoing different amounts of mechanical breast compression. In the preliminary phantom study using DBT performed by Engelken et al., promising results were obtained.158

Therapy

In image-guided radiation therapy, patient positioning pre-irradiation is of utmost importance to ensure accurate dose delivery to the tumor and sparing of the healthy tissue. The introduction of on-board cone-beam CT (CBCT) allowed for 3D verification of patient positioning, but raised concerns of the dose involved during imaging, in addition to extended image scan time and gantry rotation issues. To address these limitations, on-board tomosynthesis imaging using the kV source and planar detector used for CBCT has been proposed. Of course, compared to on-board radiography, tomosynthesis provides improved soft tissue contrast and partial information in the third dimension, while reducing the dose, acquisition time, and gantry travel involved in on-board CBCT. As with the diagnostic imaging application, the research performed on DBT for therapy has focused on optimizing the acquisition geometry159, 160, 161, 162, 163 and characterizing the radiation dose involved,160, 164 in addition to therapy-specific issues, such as comparison to reference images165 and localization of surgical clips.

CONCLUSIONS

The breast tomosynthesis process does not end with acquisition of the projection images. In addition to the obvious importance of the reconstruction process, many algorithms may be used to further improve the resulting images and to extract additional information from them. Due to the infancy of DBT in the clinical realm, some of these additional processes and analyses have had a limited chance of being tested in large clinical patient data sets, hampering our ability to gauge their true performance. Future research will surely involve the testing of these algorithms on larger clinical data sets for further characterization and optimization of their performance. In parallel, the introduction of DBT to the clinic will continue to motivate the development of advanced versions and applications of this imaging technology in addition to its combined use with other modalities that can provide functional and/or physiological information of the breast.

ACKNOWLEDGMENTS

The work by our group discussed here has been supported by the Atlanta Clinical and Translational Science Institute, the University Research Committee of Emory University, the National Science Foundation (DMS-1115627), the National Cancer Institute (R01CA163746, P50CA128301), and the National Center for Research Resources (UL1RR025008). The content is solely the responsibility of the author and does not necessarily represent the official views of these institutions. The author would like to thank Dr. Wei Zhao, Dr. Tao Wu, Dr. Heang-Ping Chan, Dr. David Gur, Dr. Ann-Katherine Carton, and Dr. Andrew Maidment for giving permission to reprint their figures, Dr. James Nagy and Steve Feng for scientific input and Jessica Paulishen for assistance with editing.

References

  1. J. T.DobbinsIII and Godfrey D. J., “Digital x-ray tomosynthesis: Current state of the art and clinical potential,” Phys. Med. Biol. 48, R65–R106 (2003). 10.1088/0031-9155/48/19/R01 [DOI] [PubMed] [Google Scholar]
  2. Mertelmeier T., Orman J., Haerer W., and Dudam M. K., “Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device,” Proc. SPIE 6142, 61420F (2006). 10.1117/12.651380 [DOI] [Google Scholar]
  3. Zhao B. and Zhao W., “Three-dimensional linear system analysis for breast tomosynthesis,” Med. Phys. 35, 5219–5232 (2008). 10.1118/1.2996014 [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Zhao B., Zhou J., Hu Y.-H., Mertelmeier T., Ludwig J., and Zhao W., “Experimental validation of a three-dimensional linear system model for breast tomosynthesis,” Med. Phys. 36, 240–251 (2009). 10.1118/1.3040178 [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Wang X., Mainprize J., Wu G., and Yaffe M., “Task-based evaluation of image quality of filtered back projection for breast tomosynthesis,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 106–113.
  6. Mertelmeier T., Ludwig J., Zhao B., and Zhao W., “Optimization of tomosynthesis acquisition parameters: Angular range and number of projections,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), p. 220–227.
  7. Orman J., Mertelmeier T., and Haerer W., “Adaptation of image quality using various filter setups in the filtered backprojection approach for digital breast tomosynthesis,” in Proceedings of the 8th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Manchester, 2006), pp. 175–182.
  8. Zhou J., Zhao B., and Zhao W., “A computer simulation platform for the optimization of a breast tomosynthesis system,” Med. Phys. 34, 1098–1109 (2007). 10.1118/1.2558160 [DOI] [PubMed] [Google Scholar]
  9. Herman G. T. and Lent A., “Iterative reconstruction algorithms,” Comput. Biol. Med. 6, 273–294 (1976). 10.1016/0010-4825(76)90066-4 [DOI] [PubMed] [Google Scholar]
  10. Ludwig J., Mertelmeier T., Kunze H., and Härer W., “A novel approach for filtered backprojection in tomosynthesis based on filter kernels determined by iterative reconstruction techniques,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 612–620.
  11. Chen Y., Lo J. Y., and J. T.DobbinsIII, “Importance of point-by-point back projection correction for isocentric motion in digital breast tomosynthesis: Relevance to morphology of structures such as microcalcifications,” Med. Phys. 34, 3885–3892 (2007). 10.1118/1.2776256 [DOI] [PubMed] [Google Scholar]
  12. Claus B., Eberhard J., Schmitz A., Carson P., Goodsitt M. and Chan H.-P., “Generalized filtered back-projection reconstruction in breast tomosynthesis,” in Proceedings of the 8th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Manchester,2006), pp. 167–174.
  13. Gordon R., Bender R., and Herman G. T., “Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography,” J. Theor. Biol. 29, 471–481 (1970). 10.1016/0022-5193(70)90109-8 [DOI] [PubMed] [Google Scholar]
  14. Gilbert P., “Iterative methods for the three-dimensional reconstruction of an object from projections,” J. Theor. Biol. 36, 105–117 (1972). 10.1016/0022-5193(72)90180-4 [DOI] [PubMed] [Google Scholar]
  15. Andersen A. H. and Kak A. C., “Simultaneous algebraic reconstruction technique (SART): A superior implementation of the ART algorithm,” Ultrason. Imaging 6, 81–94 (1984). 10.1016/0161-7346(84)90008-7 [DOI] [PubMed] [Google Scholar]
  16. Zhang Y., Chan H.-P., Sahiner B., Wei J., Goodsitt M. M., Hadjiiski L. M., Ge J., and Zhou C., “Tomosynthesis reconstruction using the simultaneous algebraic reconstruction technique (SART) on breast phantom data,” Proc. SPIE 6142, 614249 (2006). 10.1117/12.654891 [DOI] [Google Scholar]
  17. Lu Y., Chan H.-P., Wei J., and Hadjiiski L. M., “Selective-diffusion regularization for enhancement of microcalcifications in digital breast tomosynthesis reconstruction,” Med. Phys. 37, 6003–6014 (2010). 10.1118/1.3505851 [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Lu Y., Chan H.-P., Wei J., Hadjiiski L., and Zhou C., “Multiscale regularized reconstruction for enhancing microcalcification in digital breast tomosynthesis,” Proc. SPIE 8313, 831322–831329 (2012). 10.1117/12.912573 [DOI] [Google Scholar]
  19. Chen Y., Lo J. Y., and J. T.DobbinsIII, “Impulse response analysis for several digital tomosynthesis mammography reconstruction algorithms,” Proc. SPIE 5745, 541–549 (2005). 10.1117/12.595684 [DOI] [Google Scholar]
  20. Chen Y., Lo J. Y., Baker J. A., and J. T.DobbinsIII, “Gaussian frequency blending algorithm with matrix inversion tomosynthesis (MITS) and filtered back projection (FBP) for better digital breast tomosynthesis reconstruction,” Proc. SPIE 6142, 61420E–61429E (2006). 10.1117/12.652264 [DOI] [Google Scholar]
  21. Lange K. and Fessler J. A., “Globally convergent algorithms for maximum a posteriori transmission tomography,” IEEE Trans. Image Process. 4, 1430–1438 (1995). 10.1109/83.465107 [DOI] [PubMed] [Google Scholar]
  22. Wu T., Stewart A., Stanton M., McCauley T., Phillips W., Kopans D. B., Moore R. H., Eberhard J. W., Opsahl-Ong B., Niklason L., and Williams M. B., “Tomographic mammography using a limited number of low-dose cone-beam projection images,” Med. Phys. 30, 365–380 (2003). 10.1118/1.1543934 [DOI] [PubMed] [Google Scholar]
  23. Wu T., Zhang J., Moore R., Rafferty E., Kopans D., Meleis W., and Kaeli D., “Digital tomosynthesis mammography using a parallel maximum-likelihood reconstruction method,” Proc. SPIE 5368, 1–11 (2004). 10.1117/12.534446 [DOI] [Google Scholar]
  24. Goddard I., Wu T., Thieret S., Berman A., and Bartsch H., “Implementing an iterative reconstruction algorithm for digital breast tomosynthesis on graphics processing hardware,” Proc. SPIE 6142, 61424V–61427V (2006). 10.1117/12.652605 [DOI] [Google Scholar]
  25. Pratx G. and Xing L., “GPU computing in medical physics: A review,” Med. Phys. 38, 2685–2697 (2011). 10.1118/1.3578605 [DOI] [PubMed] [Google Scholar]
  26. Wu G., Mainprize J. and Yaffe M., “Characterization of projection ordering in iterative reconstruction methods for breast tomosynthesis,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 601–605.
  27. Kastanis I., Arridge S., Stewart A., Gunn S., Ullberg C. and Francke T., “3D digital breast tomosynthesis using total variation regularization,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 621–627.
  28. Sidky E. Y., Reiser I. S., Nishikawa R. M. and Pan X., “Preliminary study on the impact of digital breast tomosynthesis scanning angle on micro-calcification imaging,” in Proceedings of the IEEE Nuclear Science Symposium Conference Record (IEEE, Dresden, Germany, 2008), pp. 4201–4204.
  29. Sidky E. Y., Pan X., Reiser I. S., Nishikawa R. M., Moore R. H., and Kopans D. B., “Enhanced imaging of microcalcifications in digital breast tomosynthesis through improved image-reconstruction algorithms,” Med. Phys. 36, 4920–4932 (2009). 10.1118/1.3232211 [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Sidky E. Y., Reiser I., Nishikawa R. M., Pan X., Chartrand R., Kopans D. B., and Moore R. H., “Practical iterative image reconstruction in digital breast tomosynthesis by non-convex TpV optimization,” Proc. SPIE 6913, 691328 (2008). 10.1117/12.772796 [DOI] [Google Scholar]
  31. Sidky E. Y., Duchin Y., Reiser I., Ullberg C., and Xiaochuan P., “Optimizing algorithm parameters based on a model observer detection task for image reconstruction in digital breast tomosynthesis,” in Proceedings of the IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) (IEEE, Valencia, Spain, 2011), pp. 4230–4232.
  32. Van de Sompel D. and Brady S. M., “Robust joint entropy regularization of limited view transmission tomography using Gaussian approximations to the joint histogram,” in Proceedings of the International Conference on Information Processing in Medical Imaging (Springer Berlin Heidelberg, Williamsburg, VA, 2009), pp. 638–650. [DOI] [PubMed]
  33. Van de Sompel D. and Brady M., “Regularising limited view tomography using anatomical reference images and information theoretic similarity metrics,” Med. Image Anal. 16, 278–300 (2012). 10.1016/j.media.2011.08.005 [DOI] [PubMed] [Google Scholar]
  34. Lu Y., Chan H.-P., Fessler J. A., Hadjiiski L., Wei J., and Goodsitt M. M., “Adaptive diffusion regularization for enhancement of microcalcifications in digital breast tomosynthesis (DBT) reconstruction,” Proc. SPIE 7961, 796117 (2011). 10.1117/12.878096 [DOI] [Google Scholar]
  35. Das M., Gifford H. C., O’Connor J. M., and Glick S. J., “Penalized maximum likelihood reconstruction for improved microcalcification detection in breast tomosynthesis,” IEEE Trans. Med. Imaging 30, 904–914 (2011). 10.1109/TMI.2010.2089694 [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Jerebko A., Kowarschik M., and Mertelmeier T., “Regularization parameter selection in maximum a posteriori iterative reconstruction for digital breast tomosynthesis,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 548–555.
  37. Xu S. and Chen Y., “A Compton scattering suppression based image reconstruction method for digital breast tomosynthesis,” in Proceedings of the IEEE International Workshop on Genomic Signal Processing and Statistics (GENSIPS) (IEEE, San Antonio, TX, 2011), pp. 190–193.
  38. Wu G., Mainprize J., and Yaffe M., “Breast tomosynthesis reconstruction using a grid of blobs with projection matrices,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 243–250.
  39. Chung J., Nagy J. G., and Sechopoulos I., “Numerical algorithms for polyenergetic digital breast tomosynthesis reconstruction,” SIAM J. Imaging Sci. 3, 133–152 (2010). 10.1137/090749633 [DOI] [Google Scholar]
  40. Suryanarayanan S., Karellas A., Vedantham S., Glick S. J., D’Orsi C. J., Baker S. P., and Webber R. L., “Comparison of tomosynthesis methods used with digital mammography,” Acad. Radiol. 7, 1085–1097 (2000). 10.1016/S1076-6332(00)80061-6 [DOI] [PubMed] [Google Scholar]
  41. Suryanarayanan S., Karellas A., Vedantham S., Baker S. P., Glick S. J., D’Orsi C. J., and Webber R. L., “Evaluation of linear and nonlinear tomosynthetic reconstruction methods in digital mammography,” Acad. Radiol. 8, 219–224 (2001). 10.1016/S1076-6332(03)80530-5 [DOI] [PubMed] [Google Scholar]
  42. Wu T., Moore R. H., Rafferty E. A., and Kopans D. B., “A comparison of reconstruction algorithms for breast tomosynthesis,” Med. Phys. 31, 2636–2647 (2004). 10.1118/1.1786692 [DOI] [PubMed] [Google Scholar]
  43. Chen Y., Lo J. Y., and J. T.DobbinsIII, “Noise power spectrum analysis for several digital breast tomosynthesis reconstruction algorithms,” Proc. SPIE 6142, 614259 (2006). 10.1117/12.652282 [DOI] [Google Scholar]
  44. Zhou L., Oldan J., Fisher P., and Gindi G., “Low-contrast lesion detection in tomosynthetic breast imaging using a realistic breast phantom,” Proc. SPIE 6142, 61425A (2006). 10.1117/12.652343 [DOI] [Google Scholar]
  45. Rakowski J. T. and Dennis M. J., “A comparison of reconstruction algorithms for C-arm mammography tomosynthesis,” Med. Phys. 33, 3018–3032 (2006). 10.1118/1.2219090 [DOI] [PubMed] [Google Scholar]
  46. Zhang Y., Chan H.-P., Sahiner B., Wei J., Goodsitt M. M., Hadjiiski L. M., Ge J. and Zhou C., “A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis,” Med. Phys. 33, 3781–3795 (2006). 10.1118/1.2237543 [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Kolitsi Z., Panayiotakis G., Anastassopoulos V., Scodras A., and Pallikarakis N., “A multiple projection method for digital tomosynthesis,” Med. Phys. 19, 1045–1050 (1992). 10.1118/1.596822 [DOI] [PubMed] [Google Scholar]
  48. Kolitsi Z., Panayiotakis G., and Pallikarakis N., “A method for selective removal of out-of-plane structures in digital tomosynthesis,” Med. Phys. 20, 47–50 (1993). 10.1118/1.597060 [DOI] [PubMed] [Google Scholar]
  49. Bliznakova K., Kolitsi Z., Speller R. D., Horrocks J. A., Tromba G., and Pallikarakis N., “Evaluation of digital breast tomosynthesis reconstruction algorithms using synchrotron radiation in standard geometry,” Med. Phys. 37, 1893–1903 (2010). 10.1118/1.3371693 [DOI] [PubMed] [Google Scholar]
  50. Balla A., Zhou W., and Chen Y., “Impulse response characterization of breast tomosynthesis reconstruction with parallel imaging configurations,” Proc. SPIE 7622, 76225K (2010). 10.1117/12.844320 [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Van de Sompel D., Brady S. M., and Boone J., “Task-based performance analysis of FBP, SART and ML for digital breast tomosynthesis using signal CNR and channelised hotelling observers,” Med. Image Anal. 15, 53–70 (2011). 10.1016/j.media.2010.07.004 [DOI] [PubMed] [Google Scholar]
  52. Wu T., Moore R. H., and Kopans D. B., “Voting strategy for artifact reduction in digital breast tomosynthesis,” Med. Phys. 33, 2461–2471 (2006). 10.1118/1.2207127 [DOI] [PubMed] [Google Scholar]
  53. Ge J., Chan H.-P., Sahiner B., Zhang Y., Wei J., Hadjiiski L. M., and Zhou C., “Digital tomosynthesis mammography: Intra- and interplane artifact reduction for high-contrast objects on reconstructed slices using a priori 3D geometrical information,” Proc. SPIE 6512, 65124Q–65128Q (2007). 10.1117/12.708914 [DOI] [Google Scholar]
  54. Ge J., Chan H.-P., Sahiner B., Zhang Y., Wei J., Hadjiiski L. M., Zhou C., Wu Y.-T., and Shi J., “Digital tomosynthesis mammography: improvement of artifact reduction method for high-attenuation objects on reconstructed slices,” Proc. SPIE 6913, 69134O–69136O (2008). 10.1117/12.771539 [DOI] [Google Scholar]
  55. Erhard K., Grass M., and Nielsen T., “A second pass correction method for calcification artifacts in digital breast tomosynthesis,” Proc. SPIE 7961, 796119 (2011). 10.1117/12.877949 [DOI] [Google Scholar]
  56. Sun X., Land W., and Samala R., “Deblurring of tomosynthesis images using 3D anisotropic diffusion filtering,” Proc. SPIE 6512, 65124P (2007). 10.1117/12.697191 [DOI] [Google Scholar]
  57. Zhang Y., Chan H.-P., Sahiner B., Wu Y.-T., Zhou C., Ge J., Wei J., and Hadjiiski L. M., “Application of boundary detection information in breast tomosynthesis reconstruction,” Med. Phys. 34, 3603–3613 (2007). 10.1118/1.2761968 [DOI] [PMC free article] [PubMed] [Google Scholar]
  58. Li B., Avinash G., Claus B., and Metz S., “3D view weighted cone-beam backprojection reconstruction for digital tomosynthesis,” Proc. SPIE 6510, 65104X–65108X (2007). 10.1117/12.708084 [DOI] [Google Scholar]
  59. Zhang Y., Chan H. P., Sahiner B., Wei J., Zhou C., and Hadjiiski L. M., “Artifact reduction methods for truncated projections in iterative breast tomosynthesis reconstruction,” J. Comput. Assist. Tomogr. 33, 426–435 (2009). 10.1097/RCT.0b013e3181838000 [DOI] [PMC free article] [PubMed] [Google Scholar]
  60. Lu Y., Chan H.-P., Wei J., Hadjiiski L., and Zhou C., “Improving image quality of digital breast tomosynthesis by artifact reduction,” in Proceedings of the 11th International Workshop on Breast Imaging (Springer Berlin Heidelberg, Philadelphia, PA, 2012), pp. 745–752.
  61. Diekmann F., Meyer H., Diekmann S., Puong S., Muller S., Bick U., and Rogalla P., “Thick slices from tomosynthesis data sets: Phantom study for the evaluation of different algorithms,” J. Digit Imaging 22, 519–526 (2009). 10.1007/s10278-007-9075-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  62. Mainprize J. G., Bloomquist A., Wang X., and Yaffe M. J., “Dependence of image quality on geometric factors in breast tomosynthesis,” Med. Phys. 38, 3090–3103 (2011). 10.1118/1.3591990 [DOI] [PubMed] [Google Scholar]
  63. Li X., Da Z., and Liu B., “A generic geometric calibration method for tomographic imaging systems with flat-panel detectors: A detailed implementation guide,” Med. Phys. 37, 3844–3854 (2010). 10.1118/1.3431996 [DOI] [PubMed] [Google Scholar]
  64. Li X., Da Z. and Liu B., “Sensitivity analysis of a geometric calibration method using projection matrices for digital tomosynthesis systems,” Med. Phys. 38, 202–209 (2011). 10.1118/1.3524221 [DOI] [PubMed] [Google Scholar]
  65. Feng S. S. J. and Sechopoulos I., “Clinical digital breast tomosynthesis system: Dosimetric characterization,” Radiology 263, 35–42 (2012). 10.1148/radiol.11111789 [DOI] [PMC free article] [PubMed] [Google Scholar]
  66. Spangler M. L., Zuley M. L., Sumkin J. H., Abrams G., Ganott M. A., Hakim C., Perrin R., Chough D. M., Shah R., and Gur D., “Detection and classification of calcifications on digital breast tomosynthesis and 2D digital mammography: A comparison,” Am. J. Roentgenol. 196, 320–324 (2011). 10.2214/AJR.10.4656 [DOI] [PubMed] [Google Scholar]
  67. Rafferty E. A., “Digital mammography: Novel applications,” Radiol. Clin. North Am. 45, 831–843 (2007). 10.1016/j.rcl.2007.06.005 [DOI] [PubMed] [Google Scholar]
  68. Poplack S. P., Tosteson T. D., Kogel C. A., and Nagy H. M., “Digital breast tomosynthesis: Initial experience in 98 women with abnormal digital screening mammography,” Am. J. Roentgenol. 189, 616–623 (2007). 10.2214/AJR.07.2231 [DOI] [PubMed] [Google Scholar]
  69. Teertstra H., Loo C., van den Bosch M., van Tinteren H., Rutgers E., Muller S., and Gilhuijs K., “Breast tomosynthesis in clinical practice: Initial results,” Eur. Radiol. 20, 16–24 (2010). 10.1007/s00330-009-1523-2 [DOI] [PubMed] [Google Scholar]
  70. Kopans D., Gavenonis S., Halpern E., and Moore R., “Calcifications in the breast and digital breast tomosynthesis,” Breast J. 17, 638–644 (2011). 10.1111/j.1524-4741.2011.01152.x [DOI] [PubMed] [Google Scholar]
  71. Wallis M. G., Moa E., Zanca F., Leifland K., and Danielsson M., “Two-view and single-view tomosynthesis versus full-field digital mammography: High-resolution x-ray imaging observer study,” Radiology 262, 788–796 (2012). 10.1148/radiol.11103514 [DOI] [PubMed] [Google Scholar]
  72. Svane G., Azavedo E., Lindman K., Urech M., Nilsson J., Weber N., Lindqvist L., and Ullberg C., “Clinical experience of photon counting breast tomosynthesis: Comparison with traditional mammography,” Acta Radiol. 52, 134–142 (2011). 10.1258/ar.2010.100262 [DOI] [PubMed] [Google Scholar]
  73. Andersson I., Ikeda D., Zackrisson S., Ruschin M., Svahn T., Timberg P., and Tingberg A., “Breast tomosynthesis and digital mammography: A comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings,” Eur. Radiol. 18, 2817–2825 (2008). 10.1007/s00330-008-1076-9 [DOI] [PubMed] [Google Scholar]
  74. Gur D., Zuley M. L., Anello M. I., Rathfon G. Y., Chough D. M., Ganott M. A., Hakim C. M., Wallace L., Lu A., and Bandos A. I., “Dose reduction in digital breast tomosynthesis (DBT) screening using synthetically reconstructed projection images: An observer performance study,” Acad. Radiol. 19, 166–171 (2012). 10.1016/j.acra.2011.10.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  75. Gur D., University of Pittsburgh (personal communication).
  76. Kontos D., Bakic P. R., Carton A.-K., Troxel A. B., Conant E. F., and Maidment A. D. A., “Parenchymal texture analysis in digital breast tomosynthesis for breast cancer risk estimation: A preliminary study,” Acad. Radiol. 16, 283–298 (2009). 10.1016/j.acra.2008.08.014 [DOI] [PMC free article] [PubMed] [Google Scholar]
  77. Kontos D., Ikejimba L. C., Bakic P. R., Troxel A. B., Conant E. F., and Maidment A. D., “Analysis of parenchymal texture with digital breast tomosynthesis: Comparison with digital mammography and implications for cancer risk assessment,” Radiology 261, 80–91 (2011). 10.1148/radiol.11100966 [DOI] [PMC free article] [PubMed] [Google Scholar]
  78. Kontos D., Zhang C., Ruiter N., Bakic P., and Maidment A., “Evaluating the effect of tomosynthesis acquisition parameters on image texture: A study based on an anthropomorphic breast tissue software model,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 491–498.
  79. Engstrom E., Reiser I., and Nishikawa R., “Comparison of power spectra for tomosynthesis projections and reconstructed images,” Med. Phys. 36, 1753–1758 (2009). 10.1118/1.3116774 [DOI] [PMC free article] [PubMed] [Google Scholar]
  80. Chen L., Abbey C. K., Nosrateih A., Lindfors K. K., and Boone J. M., “Anatomical complexity in breast parenchyma and its implications for optimal breast imaging strategies,” Med. Phys. 39, 1435–1441 (2012). 10.1118/1.3685462 [DOI] [PMC free article] [PubMed] [Google Scholar]
  81. Hu Y.-H., Masiar M., and Zhao W., “Breast structural noise in digital breast tomosynthesis and its dependence on reconstruction methods,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 598–605.
  82. Burgess A. E. and Judy P. F., “Signal detection in power-law noise: Effect of spectrum exponents,” J. Opt. Soc. Am. A 24, B52–B60 (2007). 10.1364/JOSAA.24.000B52 [DOI] [PubMed] [Google Scholar]
  83. Bakic P. R., Carton A.-K., Kontos D., Zhang C., Troxel A. B., and Maidment A. D. A., “Breast percent density: Estimation on digital mammograms and central tomosynthesis projections,” Radiology 252, 40–49 (2009). 10.1148/radiol.2521081621 [DOI] [PMC free article] [PubMed] [Google Scholar]
  84. Byng J. W., Boyd N. F., Fishell E., Jong R. A., and Yaffe M. J., “The quantitative analysis of mammographic densities,” Phys. Med. Biol. 39, 1629–1638 (1994). 10.1088/0031-9155/39/10/008 [DOI] [PubMed] [Google Scholar]
  85. Tagliafico A., Tagliafico G., Astengo D., Cavagnetto F., Rosasco R., Rescinito G., Monetti F., and Calabrese M., “Mammographic density estimation: One-to-one comparison of digital mammography and digital breast tomosynthesis using fully automated software,” Eur. Radiol. 22, 1265–1270 (2012). 10.1007/s00330-012-2380-y [DOI] [PubMed] [Google Scholar]
  86. Bakic P. R., Kontos D., Carton A.-K., and Maidment A. D. A., “Breast percent density estimation from 3D reconstructed digital breast tomosynthesis images,” Proc. SPIE 6913, 691318 (2008). 10.1117/12.773890 [DOI] [Google Scholar]
  87. Packard N. and Boone J. M., “Glandular segmentation of cone beam breast CT volume images,” Proc. SPIE 6510, 651038 (2007). 10.1117/12.713911 [DOI] [Google Scholar]
  88. Chen Z., “Histogram partition and interval thresholding for volumetric breast tissue segmentation,” Comput. Med. Imaging Graph. 32, 1–10 (2008). 10.1016/j.compmedimag.2007.07.007 [DOI] [PubMed] [Google Scholar]
  89. Nelson T. R., Cervino L. I., Boone J. M., and Lindfors K. K., “Classification of breast computed tomography data,” Med. Phys. 35, 1078–1086 (2008). 10.1118/1.2839439 [DOI] [PMC free article] [PubMed] [Google Scholar]
  90. Yang X., Sechopoulos I., and Fei B., “Automatic tissue classification for high-resolution breast CT images based on bilateral filtering,” Proc. SPIE 7962, 79623H (2011). 10.1117/12.877881 [DOI] [PMC free article] [PubMed] [Google Scholar]
  91. Yang X., Wu S., Sechopoulos I., and Fei B., “Bias correction and automated classification for high-resolution dedicated breast CT images,” Med. Phys. 39, 6397–6406 (2012). 10.1118/1.4754654 [DOI] [PMC free article] [PubMed] [Google Scholar]
  92. Kontos D., Berger R., Bakic P. R., and Maidment A. D. A., “Breast tissue classification in digital breast tomosynthesis images using texture features: A feasibility study,” Proc. SPIE 7260, 726024 (2009). 10.1117/12.813812 [DOI] [Google Scholar]
  93. Vedantham S., Shi L., Karellas A., Michaelsen K. E., Krishnaswamy V., Pogue B. W., and Paulsen K. D., “Semi-automated segmentation and classification of digital breast tomosynthesis reconstructed images,” in Proceedings of the 33rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE, Boston, 2011), pp. 6188–6191. [DOI] [PMC free article] [PubMed]
  94. Young S., Park S., Anderson S. K., Badano A., Myers K. J., and Bakic P., “Estimating breast tomosynthesis performance in detection tasks with variable-background phantoms,” Proc. SPIE 7258, 72580O–72589O (2009). 10.1117/12.813900 [DOI] [Google Scholar]
  95. Park S., Jennings R., Liu H., Badano A., and Myers K., “A statistical, task-based evaluation method for three-dimensional x-ray breast imaging systems using variable-background phantoms,” Med. Phys. 37, 6253–6270 (2010). 10.1118/1.3488910 [DOI] [PMC free article] [PubMed] [Google Scholar]
  96. Gang G. J., Lee J., Stayman J. W., Tward D. J., Zbijewski W., Prince J. L., and Siewerdsen J. H., “Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance,” Med. Phys. 38, 1754–1768 (2011). 10.1118/1.3560428 [DOI] [PMC free article] [PubMed] [Google Scholar]
  97. Reiser I., Nishikawa R. M., Giger M. L., Wu T., Rafferty E., Moore R. H. and Kopans D. B., “Computerized detection of mass lesions in digital breast tomosynthesis images using two- and three dimensional radial gradient index segmentation,” Technol. Cancer Res. Treat. 3, 437–441 (2004). [DOI] [PubMed] [Google Scholar]
  98. Chan H.-P., Wei J., Sahiner B., Rafferty E. A., Wu T., Roubidoux M. A., Moore R. H., Kopans D. B., Hadjiiski L. M., and Helvie M. A., “Computer-aided detection system for breast masses on digital tomosynthesis mammograms: Preliminary experience,” Radiology 237, 1075–1080 (2005). 10.1148/radiol.2373041657 [DOI] [PMC free article] [PubMed] [Google Scholar]
  99. Chan H.-P., Wei J., Zhang Y., Sahiner B., Hadjiiski L., and Helvie M., “Detection of masses in digital breast tomosynthesis mammography: Effects of the number of projection views and dose,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 279–285.
  100. Reiser I., Nishikawa R. M., Giger M. L., Wu T., Rafferty E. A., Moore R., and Kopans D. B., “Computerized mass detection for digital breast tomosynthesis directly from the projection images,” Med. Phys. 33, 482–491 (2006). 10.1118/1.2163390 [DOI] [PubMed] [Google Scholar]
  101. Chan H. P., Wei J., Zhang Y., Helvie M. A., Moore R. H., Sahiner B., Hadjiiski L., and Kopans D. B., “Computer-aided detection of masses in digital tomosynthesis mammography: Comparison of three approaches,” Med. Phys. 35, 4087–4095 (2008). 10.1118/1.2968098 [DOI] [PMC free article] [PubMed] [Google Scholar]
  102. Singh S., Tourassi G. D., Baker J. A., Samei E., and Lo J. Y., “Automated breast mass detection in 3D reconstructed tomosynthesis volumes: A featureless approach,” Med. Phys. 35, 3626–3636 (2008). 10.1118/1.2953562 [DOI] [PMC free article] [PubMed] [Google Scholar]
  103. Mazurowski M. A., Lo J. Y., Harrawood B. P., and Tourassi G. D., “Mutual information-based template matching scheme for detection of breast masses: From mammography to digital breast tomosynthesis,” J. Biomed. Inf. 44, 815–823 (2011). 10.1016/j.jbi.2011.04.008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  104. Chan H. P., Wu Y. T., Sahiner B., Wei J., Helvie M. A., Zhang Y., Moore R. H., Kopans D. B., Hadjiiski L., and Way T., “Characterization of masses in digital breast tomosynthesis: Comparison of machine learning in projection views and reconstructed slices,” Med. Phys. 37, 3576–3586 (2010). 10.1118/1.3432570 [DOI] [PMC free article] [PubMed] [Google Scholar]
  105. Peters G., Muller S., Bernard S., Iordache R., Wheeler F., and Bloch I., “Reconstruction-independent 3D CAD for calcification detection in digital breast tomosynthesis using fuzzy particles,” in Proceedings of the 10th Iberoamerican Congress on Pattern Recognition (Springer Berlin Heidelberg, Havana, Cuba, 2005), pp. 400–408.
  106. Reiser I., Nishikawa R. M., Edwards A. V., Kopans D. B., Schmidt R. A., Papaioannou J., and Moore R. H., “Automated detection of microcalcification clusters for digital breast tomosynthesis using projection data only: A preliminary study,” Med. Phys. 35, 1486–1493 (2008). 10.1118/1.2885366 [DOI] [PMC free article] [PubMed] [Google Scholar]
  107. Park S. C., Zheng B., Wang X.-H., and Gur D., “Applying a 2D based CAD scheme for detecting micro-calcification clusters using digital breast tomosynthesis images: An assessment,” Proc. SPIE 6915, 691507–691508 (2008). 10.1117/12.768534 [DOI] [Google Scholar]
  108. Bernard S., Muller S., and Onativia J., “Computer-aided microcalcification detection on digital breast tomosynthesis data: A preliminary evaluation,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 151–157.
  109. Sahiner B., Chan H. P., Hadjiiski L. M., Helvie M. A., Wei J., Zhou C., and Lu Y., “Computer-aided detection of clustered microcalcifications in digital breast tomosynthesis: A 3D approach,” Med. Phys. 39, 28–39 (2012). 10.1118/1.3662072 [DOI] [PMC free article] [PubMed] [Google Scholar]
  110. Ho C., Tromans C., Schnabel J., and Brady S., “A clustering method for the extraction of microcalcifications using epipolar curves in digital breast tomosynthesis,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 682–688.
  111. Ho C., Tromans C., Schnabel J., and Brady S., “Microcalcification detection in digital breast tomosynthesis using an epipolar curve approach,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 704–711.
  112. Ho C. P., Tromans C., Schnabel J. A., and Brady M., “Classification of clusters of microcalcifications in digital breast tomosynthesis,” in Proceedings of the 32nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE, Buenos Aires, Argentina, 2010), pp. 3166–3169. [DOI] [PubMed]
  113. Bakic P., Richard F., and Maidment A., “Registration of mammograms and breast tomosynthesis images,” in Proceedings of the 8th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Manchester, 2006), pp. 498–503.
  114. Bakic P., Richard F., and Maidment A., “2D-to-3D x-ray breast image registration biomedical image registration,” in Proceedings of the 8th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Manchester, 2006), pp. 84–91.
  115. van Schie G., Tanner C., Snoeren P., Samulski M., Leifland K., Wallis M. G., and Karssemeijer N., “Correlating locations in ipsilateral breast tomosynthesis views using an analytical hemispherical compression model,” Phys. Med. Biol. 56, 4715–4730 (2011). 10.1088/0031-9155/56/15/006 [DOI] [PubMed] [Google Scholar]
  116. Zhang W. and Brady S., “Feature point detection for non-rigid registration of digital breast tomosynthesis images,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 296–303.
  117. Sinha S. P., Narayanan R., Ma B., Roubidoux M. A., Liu H., and Carson P. L., “Image registration for detection and quantification of change on digital tomosynthesis mammographic volumes,” Am. J. Roentgenol. 192, 384–387 (2009). 10.2214/AJR.08.1388 [DOI] [PMC free article] [PubMed] [Google Scholar]
  118. Yang G., Hipwell J., Clarkson M., Tanner C., Mertzanidou T., Gunn S., Ourselin S., Hawkes D., and Arridge S., “Combined reconstruction and registration of digital breast tomosynthesis,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 760–768.
  119. Shafer C. M., Samei E., and Lo J. Y., “The quantitative potential for breast tomosynthesis imaging,” Med. Phys. 37, 1004–1016 (2010). 10.1118/1.3285038 [DOI] [PMC free article] [PubMed] [Google Scholar]
  120. Richard S. and Samei E., “Quantitative imaging in breast tomosynthesis and CT: Comparison of detection and estimation task performance,” Med. Phys. 37, 2627–2637 (2010). 10.1118/1.3429025 [DOI] [PubMed] [Google Scholar]
  121. Richard S. and Samei E., “Quantitative breast tomosynthesis: From detectability to estimability,” Med. Phys. 37, 6157–6165 (2010). 10.1118/1.3501883 [DOI] [PubMed] [Google Scholar]
  122. Schnall M. D. et al. , “Diagnostic architectural and dynamic features at breast MR imaging: Multicenter study,” Radiology 238, 42–53 (2006). 10.1148/radiol.2381042117 [DOI] [PubMed] [Google Scholar]
  123. El Khouli R. H., Macura K. J., Jacobs M. A., Khalil T. H., Kamel I. R., Dwyer A., and Bluemke D. A., “Dynamic contrast-enhanced MRI of the breast: Quantitative method for kinetic curve type assessment,” Am. J. Roentgenol. 193, W295–W300 (2009). 10.2214/AJR.09.2483 [DOI] [PMC free article] [PubMed] [Google Scholar]
  124. El Khouli R. H., Macura K. J., Kamel I. R., Jacobs M. A., and Bluemke D. A., “3-T dynamic contrast-enhanced MRI of the breast: Pharmacokinetic parameters versus conventional kinetic curve analysis,” Am. J. Roentgenol. 197, 1498–1505 (2011). 10.2214/AJR.10.4665 [DOI] [PMC free article] [PubMed] [Google Scholar]
  125. El Khouli R. H., Macura K. J., Barker P. B., Habba M. R., Jacobs M. A., and Bluemke D. A., “Relationship of temporal resolution to diagnostic performance for dynamic contrast enhanced MRI of the breast,” J. Magn. Reson Imaging 30, 999–1004 (2009). 10.1002/jmri.21947 [DOI] [PMC free article] [PubMed] [Google Scholar]
  126. Carton A.-K., Li J., Albert M., Chen S., and Maidment A. D. A., “Quantification for contrast-enhanced digital breast tomosynthesis,” Proc. SPIE 6142, 61420D (2006). 10.1117/12.654682 [DOI] [Google Scholar]
  127. Chen S. C., Carton A. K., Albert M., Conant E. F., Schnall M. D., and Maidment A. D., “Initial clinical experience with contrast-enhanced digital breast tomosynthesis,” Acad. Radiol. 14, 229–238 (2007). 10.1016/j.acra.2006.10.022 [DOI] [PMC free article] [PubMed] [Google Scholar]
  128. Glick S. J. and Didier C., “A computer simulation for evaluating dual-energy, contrast-enhanced breast tomosynthesis,” Proc. SPIE 6510, 65102V–65107V (2007). 10.1117/12.713817 [DOI] [Google Scholar]
  129. Puong S., Bouchevreau X., Patoureaux F., Iordache R., and Muller S., “Dual-energy contrast enhanced digital mammography using a new approach for breast tissue canceling,” Proc. SPIE 6510, 65102H–65112H (2007). 10.1117/12.710133 [DOI] [Google Scholar]
  130. Puong S., Patoureaux F., Iordache R., Bouchevreau X., and Muller S., “Dual-energy contrast enhanced digital breast tomosynthesis: concept, method, and evaluation on phantoms,” Proc. SPIE 6510, 65100U–65112U (2007). 10.1117/12.710106 [DOI] [Google Scholar]
  131. Puong S., Bouchevreau X., Duchateau N., Iordache R., and Muller S., “Optimization of beam parameters and iodine quantification in dual-energy contrast enhanced digital breast tomosynthesis,” Proc. SPIE 6913, 69130Z (2008). 10.1117/12.770148 [DOI] [Google Scholar]
  132. Iordache R., Lohezic M., Palma G., Puong S., and Muller S., “Noise reduction in dual-energy contrast enhanced digital breast tomosynthesis using regularization,” in Proceedings of the 10th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Girona, Spain, 2010), pp. 92–99.
  133. Carton A.-K., Currivan J. A., Conant E., and Maidment A., “Temporal subtraction versus dual-energy contrast-enhanced digital breast tomosynthesis: A pilot study,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 166–173.
  134. Carton A. K., Gavenonis S. C., Currivan J. A., Conant E. F., Schnall M. D., and Maidment A. D., “Dual-energy contrast-enhanced digital breast tomosynthesis: A feasibility study,” Br. J. Radiol. 83, 344–350 (2010). 10.1259/bjr/80279516 [DOI] [PMC free article] [PubMed] [Google Scholar]
  135. Carton A.-K., Ullberg C., Lindman K., Acciavatti R., Francke T., and Maidment A. D. A., “Optimization of a dual-energy contrast-enhanced technique for a photon-counting digital breast tomosynthesis system: I. A theoretical model,” Med. Phys. 37, 5896–5907 (2010). 10.1118/1.3490556 [DOI] [PMC free article] [PubMed] [Google Scholar]
  136. Carton A.-K., Ullberg C., and Maidment A. D. A., “Optimization of a dual-energy contrast-enhanced technique for a photon-counting digital breast tomosynthesis system: II. An experimental validation,” Med. Phys. 37, 5908–5913 (2010). 10.1118/1.3488889 [DOI] [PMC free article] [PubMed] [Google Scholar]
  137. Samei E. and Saunders R. S., “Dual-energy contrast-enhanced breast tomosynthesis: Optimization of beam quality for dose and image quality,” Phys. Med. Biol. 56, 6359–6378 (2011). 10.1088/0031-9155/56/19/013 [DOI] [PMC free article] [PubMed] [Google Scholar]
  138. Schmitzberger F. F., Fallenberg E. M., Lawaczeck R., Hemmendorff M., Moa E., Danielsson M., Bick U., Diekmann S., Pollinger A., Engelken F. J., and Diekmann F., “Development of low-dose photon-counting contrast-enhanced tomosynthesis with spectral imaging,” Radiology 259, 558–564 (2011). 10.1148/radiol.11101682 [DOI] [PubMed] [Google Scholar]
  139. Fredenberg E., Lundqvist M., Aslund M., Hemmendorff M., Cederstrom B., and Danielsson M., “A photon-counting detector for dual-energy breast tomosynthesis,” Proc. SPIE 7258, 72581J (2009). 10.1117/12.813037 [DOI] [Google Scholar]
  140. Hill M., Mainprize J., and Yaffe M., “Sensitivity of contrast-enhanced digital breast tomosynthesis to changes in iodine concentration during acquisition,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 643–650.
  141. Hill M. L., Mainprize J. G., Puong S., Carton A.-K., Iordache R., Muller S., and Yaffe M. J., “Impact of image acquisition timing on image quality for dual energy contrast-enhanced breast tomosynthesis,” Proc. SPIE 8313, 831308–831309 (2012). 10.1117/12.911627 [DOI] [Google Scholar]
  142. Boverman G., Kao T. J., Kulkarni R., Kim B. S., Isaacson D., Saulnier G. J., and Newell J. C., “Robust linearized image reconstruction for multifrequency EIT of the breast,” IEEE Trans. Med. Imaging 27, 1439–1448 (2008). 10.1109/TMI.2008.922187 [DOI] [PMC free article] [PubMed] [Google Scholar]
  143. Kao T. J., Boverman G., Isaacson D., Saulnier G. J., Newell J. C., Moore R. H., and Kopans D. B., “Regional admittivity spectra with tomosynthesis images for breast cancer detection,” in Proceedings of the 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE, Shanghai, China, 2007), pp. 4142–4145. [DOI] [PubMed]
  144. Kao T. J., Boverman G., Kim B. S., Isaacson D., Saulnier G. J., Newell J. C., Choi M. H., Moore R. H., and Kopans D. B., “Regional admittivity spectra with tomosynthesis images for breast cancer detection: Preliminary patient study,” IEEE Trans. Med. Imaging 27, 1762–1768 (2008). 10.1109/TMI.2008.926049 [DOI] [PMC free article] [PubMed] [Google Scholar]
  145. Sinha S. P., Roubidoux M. A., Helvie M. A., Nees A. V., Goodsitt M. M., LeCarpentier G. L., Fowlkes J. B., Chalek C. L., and Carson P. L., “Multi-modality 3D breast imaging with x-ray tomosynthesis and automated ultrasound,” in Proceedings of the 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE, Shanghai, China, 2007), pp. 1335–1338. [DOI] [PubMed]
  146. Kapur A. et al. , “Combination of digital mammography with semi-automated 3D breast ultrasound,” Technol. Cancer Res. Treat. 3, 325–334 (2004). [DOI] [PMC free article] [PubMed] [Google Scholar]
  147. More M. J., Heng L., Goodale P. J., Yibin Z., Majewski S., Popov V., Welch B., and Williams M. B., “Limited angle dual modality breast imaging,” IEEE Trans. Nucl. Sci. 54, 504–513 (2007). 10.1109/TNS.2007.897828 [DOI] [Google Scholar]
  148. Williams M.et al. , “Tomographic dual modality breast scanner,” in Proceedings of the 9th International Workshop on Digital Mammography (Springer Berlin Heidelberg, Tucson, AZ, 2008), pp. 99–107.
  149. Williams M. B., Judy P. G., Gunn S., and Majewski S., “Dual-modality breast tomosynthesis,” Radiology 255, 191–198 (2010). 10.1148/radiol.09091160 [DOI] [PMC free article] [PubMed] [Google Scholar]
  150. Boverman G., Fang Q., Carp S. A., Miller E. L., Brooks D. H., Selb J., Moore R. H., Kopans D. B., and Boas D. A., “Spatio-temporal imaging of the hemoglobin in the compressed breast with diffuse optical tomography,” Phys. Med. Biol. 52, 3619–3641 (2007). 10.1088/0031-9155/52/12/018 [DOI] [PubMed] [Google Scholar]
  151. Fang Q., Moore R. H., Kopans D. B., and Boas D. A., “Compositional-prior-guided image reconstruction algorithm for multi-modality imaging,” Biomed. Opt. Express 1, 223–235 (2010). 10.1364/BOE.1.000223 [DOI] [PMC free article] [PubMed] [Google Scholar]
  152. Fang Q., Selb J., Carp S. A., Boverman G., Miller E. L., Brooks D. H., Moore R. H., Kopans D. B., and Boas D. A., “Combined optical and x-ray tomosynthesis breast imaging,” Radiology 258, 89–97 (2011). 10.1148/radiol.10082176 [DOI] [PMC free article] [PubMed] [Google Scholar]
  153. Wilkins S. W., Gureyev T. E., Gao D., Pogany A., and Stevenson A. W., “Phase-contrast imaging using polychromatic hard x-rays,” Nature (London) 384, 335–338 (1996). 10.1038/384335a0 [DOI] [Google Scholar]
  154. Honda C. and Ohara H., “Advantages of magnification in digital phase-contrast mammography using a practical x-ray tube,” Eur. J. Radiol. 68, S69–S72 (2008). 10.1016/j.ejrad.2008.04.041 [DOI] [PubMed] [Google Scholar]
  155. Tang X., Yang Y., and Tang S., “Characterization of imaging performance in differential phase contrast CT compared with the conventional CT: Noise power spectrum NPS(k),” Med. Phys. 38, 4386–4395 (2011). 10.1118/1.3602071 [DOI] [PMC free article] [PubMed] [Google Scholar]
  156. Hammonds J. C., Price R. R., Donnelly E. F., and Pickens D. R., “Phase-contrast digital tomosynthesis,” Med. Phys. 38, 2353–2358 (2011). 10.1118/1.3574871 [DOI] [PubMed] [Google Scholar]
  157. Kim G. W., Han B. H., Cho M. H., and Lee S. Y., “X-ray elastography: A feasibility study,” in Proceedings of the 31st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE, Minneapolis, 2009), pp. 3513–3516. [DOI] [PubMed]
  158. Engelken F. J., Sack I., Klatt D., Fischer T., Fallenberg E. M., Bick U., and Diekmann F., “Evaluation of tomosynthesis elastography in a breast-mimicking phantom,” Eur. J. Radiol. 81, 2169–2173 (2012). 10.1016/j.ejrad.2011.06.033 [DOI] [PubMed] [Google Scholar]
  159. Lyatskaya Y., Buehler A., Ng S. K., Wong J., Bellon J. R., and Zygmanski P., “Optimal gantry angles and field sizes in kilovoltage cone-beam tomosynthesis for set-up of women with breast cancer undergoing radiotherapy treatment,” Radiother. Oncol. 93, 633–638 (2009). 10.1016/j.radonc.2009.10.016 [DOI] [PubMed] [Google Scholar]
  160. Winey B. A., Zygmanski P., Cormack R. A., and Lyatskaya Y., “Balancing dose and image registration accuracy for cone beam tomosynthesis (CBTS) for breast patient setup,” Med. Phys. 37, 4414–4423 (2010). 10.1118/1.3464716 [DOI] [PubMed] [Google Scholar]
  161. Zhang J., Wu Q. J., Godfrey D. J., Fatunase T., Marks L. B., and Yin F. F., “Comparing digital tomosynthesis to cone-beam CT for position verification in patients undergoing partial breast irradiation,” Int. J. Radiat. Oncol., Biol., Phys. 73, 952–957 (2009). 10.1016/j.ijrobp.2008.10.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
  162. Ng S. K., Zygmanski P., Jeung A., Mostafavi H., Hesser J., Bellon J. R., Wong J. S., and Lyatskaya Y., “Optimal parameters for clinical implementation of breast cancer patient setup using Varian DTS software,” J. Appl. Clin. Med. Phys. 13, 60–73 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  163. Sarkar V., Shi C., Rassiah-Szegedi P., Diaz A., Eng T., and Papanikolaou N., “The effect of a limited number of projections and reconstruction algorithms on the image quality of megavoltage digital tomosynthesis,” J. Appl. Clin. Med. Phys. 10, 155–172 (2009). 10.1120/jacmp.v10i3.2970 [DOI] [PMC free article] [PubMed] [Google Scholar]
  164. Winey B., Zygmanski P., and Lyatskaya Y., “Evaluation of radiation dose delivered by cone beam CT and tomosynthesis employed for setup of external breast irradiation,” Med. Phys. 36, 164–173 (2009). 10.1118/1.3036113 [DOI] [PubMed] [Google Scholar]
  165. Godfrey D. J., Ren L., Yan H., Wu Q., Yoo S., Oldham M., and Yin F. F., “Evaluation of three types of reference image data for external beam radiotherapy target localization using digital tomosynthesis (DTS),” Med. Phys. 34, 3374–3384 (2007). 10.1118/1.2756941 [DOI] [PubMed] [Google Scholar]

Articles from Medical Physics are provided here courtesy of American Association of Physicists in Medicine

RESOURCES