Abstract
Objective
A dosimetric study was performed to evaluate the performance of volumetric modulated arc radiotherapy with RapidArc on locally advanced nasopharyngeal carcinoma (NPC).
Methods
The CT scan data sets of 20 patients of locally advanced NPC were selected randomly. The plans were managed using volumetric modulated arc with RapidArc and fixed nine-field coplanar dynamic intensity-modulated radiotherapy (IMRT) for these patients. The dosimetry of the planning target volumes (PTV), the organs at risk (OARs) and the healthy tissue were evaluated. The dose prescription was set to 70 Gy to the primary tumor and 60 Gy to the clinical target volumes (CTV) in 33 fractions. Each fraction applied daily, five fractions per week. The monitor unit (MU) values and the delivery time were scored to evaluate the expected treatment efficiency.
Results
Both techniques had reached clinical treatment’s requirement. The mean dose (Dmean), maximum dose (Dmax) and minimum dose (Dmin) in RapidArc and fixed field IMRT for PTV were 68.4±0.6 Gy, 74.8±0.9 Gy and 56.8±1.1 Gy; and 67.6±0.6 Gy, 73.8±0.4 Gy and 57.5±0.6 Gy (P<0.05), respectively. Homogeneity index was 78.85±1.29 in RapidArc and 80.34±0.54 (P<0.05) in IMRT. The conformity index (CI: 95%) was 0.78±0.01 for both techniques (P>0.05). Compared to IMRT, RapidArc allowed a reduction of Dmean to the brain stem, mandible and optic nerves of 14.1% (P<0.05), 5.6% (P<0.05) and 12.2% (P<0.05), respectively. For the healthy tissue and the whole absorbed dose, Dmean of RapidArc was reduced by 3.6% (P<0.05), and 3.7% (P<0.05), respectively. The Dmean to the parotids, the spinal cord and the lens had no statistical difference among them. The mean MU values of RapidArc and IMRT were 550 and 1,379. The mean treatment time of RapidArc and IMRT was 165 s and 447 s. Compared to IMRT, the delivery time and the MU values of RapidArc were reduced by 63% and 60%, respectively.
Conclusion
For locally advanced NPC, both RapidArc and IMRT reached the clinic requirement. The target volume coverage was similar for the different techniques. The RapidArc technique showed some improvements in OARs and other tissue sparing while using reduced MUs and delivery time.
Key words: Volumetric modulated arc therapy, Intensity-modulated radiotherapy, Dosimetry, Target volume, Nasopharyngeal carcinoma
INTRODUCTION
Recently, nasopharyngeal carcinoma (NPC) patients more likely have received intensity-modulated radio- therapy (IMRT) than three dimensionals’ conformal radiotherapy (3D-CRT). The IMRT has got better dose distribution in target volume and lower dose for organs at risk (OARs), especially for the parotids[1-3]. But IMRT needs complex plan management, more fixed fields and monitor units (MUs) especially for NPC which has large target volume, more OARsand more overlapping of target volume. All of the above may bring the prolonging of the treatment time (for example, 7-9 fixed-field dynamic IMRT needs about 7-10 minutes of treatment delivery) which may cause the increasing of the movement of the swallowing and the position shift during the treatment, so that the clinical efficacy may be reduced consequently.
Generally, the number of the fixed fields can increase the freedom of the plan management, and the development of the volumetric modulated arc therapy technique brings the design of the treatment plan to a new stage which can produce various optimization methods based on the differences of the peak value optimization[4]. Nowadays, RapidArc was developed mainly by the optimization of the multi-leaf collimators shape, the change of the dose rate delivery and the rotation of the gantry[5-8]. It was based on the volumetric modulated arc therapy technique and can obtain the similar distribution of the fixed IMRT. The analytical anisotropic algorithm (AAA) system was used for dose calculation[9], and GLAaS[10]and PTW-729[11] methods were applied for quality control to ensure the accuracy and security in the clinical application[12].
Currently, multiple centers compared the dose distribution of the both techniques and generally suggested that volumetric modulated arc therapy with RapidArc was a rapid, safe and accurate radiotherapy technique for many tumors like gliomas, brain metastases and some lung tumors according to the preliminary results[13-17].
NPC needs large and complex target volume and has many OARs around, so that complicated fields’ management was necessary. Therefore, in this study, RapidArc was compared at the reference of the fixed 9-field IMRT in dosimetry for locally advanced NPC.
MATERIALS AND METHODS
Patients’ Characteristics
Twenty cases of location CT scan data (layer 3 mm) was randomly selected from the locally advanced NPC patients who had received radiotherapy (RT) continuously in Radiotherapy Department, Beijing Cancer Hospital. The patients’ characteristics are shown in Table 1, and the clinical stage was according to the Stage of NPC (AJCC 2002) followed as below. Among these 20 patients, three patients had been diagnosed as T4 and six patients as N3. The gross tumor volume (GTV) was ranged from 51.4 cm3 to 421.8 cm3 and the median volume was 130.9±83.2 cm3.
Table 1. Patients’ General and Clinical Characteristics.
| Parameters (20 cases) | Values | |
|---|---|---|
| Age (y) | 48 (23−70) | |
| Sex (male:female) | 7:3 | |
| Stage III | 13 | |
| Stage IV | 7 | |
Treatment Plan Management
Two treatment plans were performed for each patient. RapidArc was compared to the fixed 9-field coplanar dynamic IMRT in dosimetry. The Eclipse system from the Varian Company (Denver, USA) was used for the two RT plans, with 6MV-X ray and 120 multi-leaf collimator in it. Considering the large target volume of the locally advanced NPC and complex OARs around, the double-arc plan was adopted for RapidArc and the coplanar fixed 9-field plan was selected for IMRT. Meantime, the AAA 8.6 edition system was applied for calculation.
The simultaneous boost plan was used and the dose prescribed as: GTV 70 Gy/33f, 95% planning target volume (PTV) 60 Gy/33f was generated by 5-mm outer margin of clinical target volume (CTV) and 5 mm apart from the skin at least.
The quality control of the plan was in accordance with the standard dose-volume histogram (DVH) of D98% and D2% which represent the doses of 98% and 2% PTV and they indicated the minimum and maximum doses of the plan respectively. The conformity index (CI) of the target volume is expressed as CI95%=(PTV60Gy/ VPTV)×(PTV60Gy/V60Gy)[18]. PTV60Gy represents the volume receiving the prescription dose 60 Gy in the target volume, VPTV stands for the volume of the PTV, V60Gy is in the name of the volume which has received the prescribed dose. The homogeneity index (HI) of the target volume is defined as HI=100×[1−(D5%−D95%)/ Dmean][15]. OARs, D33%, Dmean, D50%, and D66%were adopted to evaluate the dose distribution of both sides of the parotids. Dmeanand Dmax(the maximum dose which was defined as the dose received by less than 2 ml volume of the following OARs) were applied to evaluate the dose of the lens, spinal cord, mandible and optic nerves. For the healthy tissue, integrity absorption dose (DoseInt) was used as the evaluation standard accompaniedby Dmeanand V10Gy at the same time. The time interval of the treatment delivery and the MU values of the techniques were also compared.
Statistical Analyses
The SPSS 13.0 (SPSS Inc., IL, USA) was applied for statistical data management and analysis, and double- side t-test was employed to compare the difference between two treatment plans at statistically significant level of P<0.05. The null hypothesis of no difference in dosimetry between study groups was tested with the use of the log-rank test at a two-sided level of significance of 0.05. Confidence intervals were calculated and study groups were compared by means of the log-rank test. All other hypothesis tests were two-sided at a significance level of 0.05.
RESULTS
Dosimetry Analysis of Target Volume
Table 2 shows the comparison of the dose distribution of the PTVs and the CI and HI of the target volumes. According to the clinical requirement, Dmean and Dmax of the PTV were lower while Dmin was slightly higher and statistically significant for IMRT. The CI95% of RapidArc and IMRT were both 0.78±0.07 (P>0.05). However, the HI for RapidArc and IMRT were 78.9±1.3 and 80.4±0.5, respectively (P<0.05).
Table 2. PTV dose distribution comparison of two RT techniques.
| Indices | RapidArc | IMRT | P |
|---|---|---|---|
| PTV Dmean (Gy) | 68.4±0.6 | 67.6±0.6 | <0.05 |
| PTV Dmax (Gy) | 74.8±0.9 | 73.8±0.5 | <0.05 |
| PTV Dmin (Gy) | 56.8±1.1 | 57.5±0.6 | <0.05 |
| CI95% | 0.78±0.07 | 0.78±0.07 | >0.05 |
| HI | 78.9±1.3 | 80.4±0.5 | <0.05 |
CI, conformity index; HI, homogeneity index.In table 2,both of PTV Dmean and PTV Dmax in RapidArc were slightly higher and statistically significant than those in IMRT. The CI95% of RapidArc and IMRT were the same and the HI for RapidArc was lower thanthat of IMRT at statistically significant level (P<0.05).
Dosimetry Comparison of OARs and Healthy Tissues
Table 3 manifests the dose comparison of the OARs for the two plans. As we can see, the plans were similar in the mean dose of the spinal cord and had no differences at statistically significant levels (P<0.05). With the standard of the clinical treatment, the mean dose of RapidArc for the brain stem was really lower than that of IMRT; however, for the maximum dose, IMRT was indeed very low. The mean and maximum doses of the lens were similar in two plans and have no differences at statistically significant levels (P>0.05). For both sides of the optic nerves, the doses (except for the maximum dose of the right optic nerve) in RapidArc were less than that in IMRT, the differences were statistically significant at P<0.05 levels. The maximum doses for the mandible of the RapidArc and IMRT plans were similar in statistically significant level (P<0.05) with the mean doses less than 50 Gy for the both, while the former was lower at statistically significant levels (P<0.05) in the mean dose.
Table 3. Dose comparison of OARs for two plans.
| Indices | RapidArc | IMRT | P |
|---|---|---|---|
| Spinal cord Dmean (Gy) | 27.7±1.4 | 27.8±1.7 | >0.05 |
| Spinal cord Dmax (Gy) | 33.7±1.7 | 33.5±1.2 | >0.05 |
| Brain stem Dmean (Gy) | 23.9±2.8 | 27.9±2.2 | <0.05 |
| Brain stem Dmax (Gy) | 45.2±2.4 | 43.6±1.7 | <0.05 |
| Mandible Dmean (Gy) | 46.6±2.5 | 50.5±2.1 | <0.05 |
| Mandible Dmax (Gy) | 73.8±1.8 | 74.0±2.1 | >0.05 |
| Left lens Dmean (Gy) | 5.3±0.5 | 5.4±0.7 | >0.05 |
| Left lens Dmax (Gy) | 6.2±0.5 | 6.2±0.7 | >0.05 |
| Right lens Dmean (Gy) | 5.3±0.4 | 5.2±0.6 | >0.05 |
| Right lens Dmax (Gy) | 6.2±0.4 | 6.4±0.7 | >0.05 |
| Left optic nerve Dmean (Gy) | 32.8±5.2 | 37.6±4.9 | <0.05 |
| Left optic nerve Dmax (Gy) | 53.5±4.2 | 55.6±3.0 | <0.05 |
| Right optic nerve Dmean (Gy) | 32.9±5.6 | 37.6±4.6 | <0.05 |
| Right optic nerve Dmax (Gy) | 51.6±8.9 | 55.1±2.6 | >0.05 |
The Dmean of RapidArc for the brain stem was really lower than that of IMRT. As was shown above, the Dmean and the Dmax for the left optic nerve of RapidArc were really lower than that of IMRT. For the Dmean of the mandible, RapidArc was also lower than IMRT (P<0.05).
For both sides of the parotids, we performed detailed analysis in different volumes. The Dmean, D33%, D50% and D66% were all lower for IMRT. However, all of them above had no statistical differences between both sides of the parotids (Table 4). Figure 1 shows a demographic of the two different plans.
Table 4. Comparison of dose distribution for parotids.
| Indices | RapidArc | IMRT | P |
|---|---|---|---|
| Left parotids Dmean (Gy) | 32.9±7.5 | 32.0±3.6 | >0.05 |
| Left parotids D33% (Gy) | 34.2±7.5 | 32.2±5.3 | >0.05 |
| Left parotids D50% (Gy) | 30.8±12.7 | 27.0±3.2 | >0.05 |
| Left parotids D66% (Gy) | 25.3±8.5 | 24.2±2.3 | >0.05 |
| Right parotids Dmean (Gy) | 33.4±9.8 | 33.4±7.9 | >0.05 |
| Right parotids D33% (Gy) | 35.1±10.2 | 33.7±9.1 | >0.05 |
| Right parotids D50% (Gy) | 29.1±11.3 | 28.8±9.8 | >0.05 |
| Right parotids D66% (Gy) | 26.0±11.7 | 26.0±9.5 | >0.05 |
The Dmean, D33%, D50% and D66% for both sides of the parotids of RapidArc were likely higher than that of IMRT but not statisticallysignificant.
Figure 1.

DVH for two treatment plans.The solid lines represent the doses of the different OARs and target volume in different volumes for RapidArc, and the dotted lines stand for those of IMRT. RapidArc appeared some priority in the mandible and the brain stem visually.
There was no clear definition of the radiation dose to the healthy tissue. But in principle, lower integrity absorption dose (DoseInt) and smaller radiation volume are better. The Dmean and DoseInt were significantly lowered (P<0.05) for RapidArc. The radiation volume of 10 Gy was smaller for IMRT, but has no statistical significance (Table 5).
Table 5. Dose comparison of the healthy tissue.
| Indices | RapidArc | IMRT | P |
|---|---|---|---|
| Healthy tissue Dmean (Gy) | 29.0±2.2 | 30.1±2.2 | <0.05 |
| Volume of ≤10Gy (cm3) | 5,745±1,150 | 5,708±1,142 | >0.05 |
| DoseInt (Gy×cm3) | 216,190±41,308 | 224,491±43,318 | <0.05 |
DoseInt, integrity absorption dose. The Dmean and DoseInt of the healthy tissue were lowered significantly (P<0.05) for RapidArc than IMRT.
Comparison of MUValue and Treatment Time Interval
Figure 2 shows the MU values and time interval for the treatment plans. The mean MU value of RapidArc was 550, reduced by 60% compared to that of IMRT with the mean MU value of 1379. The mean treatment time of RapidArc was 165 s, reduced by 63% compared to that of IMRT with the mean time of 447 s.
Figure 2.

The MU value and treatment time for the two plans.
DISCUSSION
As mentioned above, the treatment time interval is much shorter for RapidArc which may bring the promotion of the biologic effect. Although accuracy and repeatability of the RapidArc had been evaluated in some studies, there is no dose comparison for the two treatment plans in NPC patients currently.
In this research, the enrolled patients were satisfied with clinical treatment. The data of RapidArc implied that the dose distribution may be improved for some OARs and we can get more accurate plans, but this is not the point of the study. Moreover, in the program of the optimum, the weight for the target volume and the OARs were not clearly defined. Therefore, the dosimetry comparison for the two plans has its limitation.
The double-arc RapidArc was selected to guarantee the dose distribution due to a large targeting volume and lots of OARs for NPC patients. The reduction of the healthy tissue dose may decrease the risk of the secondary primary carcinoma, which needs follow up to clarify. The MU value was reduced by 60% meanly for RapidArc.
The result showed the 63% reduction of the mean treatment time can decrease the effect of the organ movement, and affect treatment effectiveness. In American Society for Therapeutic Radiology and Oncology (ASTRO) meeting in 2009, it was reported by Wang that the swallowing frequency was 3-19 times in one treatment fraction (11 minutes meanly in IMRT), the time interval was 4.8 secends for swallowing, and the shift of the target volume was 3.13-12.32 mm during the course. The data demonstrated that the position shift and the swallowing movement would inevitably lower the accuracy of the target volume because of the fairly long time interval for the treatment delivery of the head and neck cancer, thereby decreasing the clinical treatment effectiveness. Therefore, RapidArc can increase the accuracy of the radiotherapy and improve the effectiveness of the treatment by reducing the time interval of the delivery.
Furthermore, the reduction of the fraction treatment time may increase the biologic effect of radiotherapy. Shibamoto, et al.[19]showed that between two fractions of radiotherapy, sublethal repair occurred in 2-3 minutes or longer time, while using breast cancer cell strain EMT 6 and head and neck squamous cell strain SCCVn of mice in vitro.
The previous data about RapidArc and fixed field IMRT showed that the treatment time interval was less than 2.5 minutes for RapidArc and about 7-9 minutes for the fixed field IMRT (7-9 fields generally). The superiority of RapidArc in this work was the obvious reduction of the MU value and the treatment time[20,21]. Therefore, we believe that RapidArc may promote the clinical result since it can reduce the shift of the patient location[20] and the influence of the swallowing movement on the OARs during the delivery.
To achieve satisfactory dose distribution, it will take more than 17 hours for the plan optimization to RapidArc for one locally advanced NPC while only 0.5-1 hour is needed for IMRT. Consequently, the calculation system for the RapidArc plan needs to be improved in future studies.
In conclusion, for the local advanced NPC, RapidArc can reach similar target conformity comparing to IMRT for target dose distribution and homogeneity, and IMRT is a little better than RapidArc. Compare to IMRT, though the parotids received higher dose for RapidArc, it would not damage parotids statistical significantly; brain stem, mandible and both sides of the optic nerves had better dose distribution in RapidArc plans; for the lens and spinal cord, the dose distribution was similar in both plans. Considering the mean dose of the healthy tissue and the integrity absorbed dose, RapidArc was superior to IMRT. The reduction of the treatment time and the MU value were notable for RapidArc.
ACKNOWLEDGEMENTS
Sincerely appreciate to Dr. Zheng Junxiu for the great help in the arrangement of the data and the modification of the writing for this article. Thanks again for her unselfish help.
REFERENCES
- 1.Kam MK, Chau RM, Suen J, et al. Intensity-modulated radiotherapy in nasopharyngeal carcinoma: dosimetric advantage over conventional plans and feasibility of dose escalation. Int J Radia Oncol BiolPhys 2003;56:145-57 [DOI] [PubMed] [Google Scholar]
- 2.Chau RM, Teo PM, Kam MK, et al. Dosimetric comparison between 2-dimensional radiation therapy and intensity modulated radiation therapy in treatment of advanced T-stage nasopharyngeal carcinoma: to treat less or more in the planning organ-at-risk volume of the brainstem and spinal cord. Med Dosim 2007;32:263-70 [DOI] [PubMed] [Google Scholar]
- 3.Wu Q, Manning M, Schmidt-Ullrich R, et al. The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head-and-neck cancers: a treatment design study. Int J Rad Oncol Biol Phys 2000;46:195-205 [DOI] [PubMed] [Google Scholar]
- 4.Thieke C, Küfer KH, Monz M, et al. A new concept for interactive radiotherapy planning with multi-criteria optimization: First clinical evaluation. RadiotherOncol 2007;85:292-8 [DOI] [PubMed] [Google Scholar]
- 5.Ling CC, Zhang P, Archambault Y, et al. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Rad Oncol Biol Phys 2008;72:575-81 [DOI] [PubMed] [Google Scholar]
- 6.Berg M, Bangsgaard JP, Vogelius IS. Absorption measurements on a new cone beam CT and IMRT compatible tabletop for use in external radiotherapy. Phys Med Biol 2009;54:N319-28 [DOI] [PubMed] [Google Scholar]
- 7.Otto K, Letter to the editor on ‘Single-Arc IMRT’? Phys Med Biol 2009;54:L37-41 [DOI] [PubMed] [Google Scholar]
- 8.Falk M, af Rosenschöld PM, Keall P, et al. Real-time dynamic MLC tracking for inversely optimized arc radiotherapy. Radiother Oncol 2010;94:218-23 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Gagne IM, Ansbacher W, Zavgorodni S, et al. A Monte Carlo evaluation of RapidArc dose calculations for oropharynx radiotherapy. Phys Med Biol 2008;53:7167-85 [DOI] [PubMed] [Google Scholar]
- 10.Nicolini G, Vanetti E, Clivio A, et al. The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008;3:24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Vanetti E, Nicolini G, Clivio A, et al. The impact of treatment couch modeling on RapidArc. Phys Med Biol 2009;54:157-66 [DOI] [PubMed] [Google Scholar]
- 12.Korreman S, Medin J, Kjer Kristofferson F.Dosimetric verification of RapidArc treatment delivery. Acta Oncol 2009;48:185-91 [DOI] [PubMed] [Google Scholar]
- 13.Otto K.Volumetric Modulated Arc therapy: IMRT in a single arc. Med Phys 2008;35:310-7 [DOI] [PubMed] [Google Scholar]
- 14.Shaffer R, Nichol AM, Vollans E, et al. A comparison of volumetric modulated arc therapy and conventional intensity- modulated radiotherapy for frontal and temporal high-grade gliomas. Int J Radiat Oncol Biol Phys 2010;76:1177-84 [DOI] [PubMed] [Google Scholar]
- 15.Lagerwaard FJ, Van Der Hoorn EA, Verbakel WF, et al. Whole- brain radiotherapy with simultaneous integrated boost to multiple brain metastases using volumetric modulated arc therapy. Int J Radiat Oncol Biol Phys 2009;75:253-9 [DOI] [PubMed] [Google Scholar]
- 16.Lagerwaard FJ, Meijer OW, Van Der Hoorn EA, et al. Volumetric modulated arc radiotherapy for vestibular schwannomas. Int J Rad Oncol Biol Phys 2009;74:610-5 [DOI] [PubMed] [Google Scholar]
- 17.Verbakel WF, Senan S, Cuijpers JP, et al. Rapid delivery of stereotactic radiotherapy for peripheral lung tumors using volumetric intensity-modulated arcs. Radiother Oncol 2009;93:122-4 [DOI] [PubMed] [Google Scholar]
- 18.Crijns W, Budiharto T, Defraene G, et al. IMRT-based optimization approaches for volumetric modulated single arc radiotherapy planning. Radiother Oncol 2010;95:149-52 [DOI] [PubMed] [Google Scholar]
- 19.Shibamoto Y, Ito M, Sugie C, et al. Reeovery from sublethal damage during iniermittent exposures in cultured tumor cells: imPlieations for dose modification in radiosurgery and imrt. Int J RadiatOncolBiolPhys 2004;59:1484-90 [DOI] [PubMed] [Google Scholar]
- 20.Popescu CC, Olivotto IA, Beckham WA, et al. Volumertric modulated arc therapy improves dosimetry and reduces treatment time compared to conventional intensity- modulated radiotherapy of left-sided breast cancer and internal mammary nodes. Int J RadiaOncolBiolPhys 2010;76:287-95 [DOI] [PubMed] [Google Scholar]
- 21.Yoo S, Wu QJ, Lee WR, et al. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vesicles and lymph node. Int J RadiatOncolBiolPhys 2010;76:935-42 [DOI] [PubMed] [Google Scholar]