Abstract
Purpose
To improve tumor dose conformity and homogeneity for COMS plaque brachytherapy by investigating the dosimetric effects of varying component source ring radionuclides and source strengths.
Material and methods
The MCNP5 Monte Carlo (MC) radiation transport code was used to simulate plaque heterogeneity-corrected dose distributions for individually-activated source rings of 14, 16 and 18 mm diameter COMS plaques, populated with 103Pd, 125I and 131Cs sources. Ellipsoidal tumors were contoured for each plaque size and MATLAB programming was developed to generate tumor dose distributions for all possible ring weighting and radionuclide permutations for a given plaque size and source strength resolution, assuming a 75 Gy apical prescription dose. These dose distributions were analyzed for conformity and homogeneity and compared to reference dose distributions from uniformly-loaded 125I plaques. The most conformal and homogeneous dose distributions were reproduced within a reference eye environment to assess organ-at-risk (OAR) doses in the Pinnacle3 treatment planning system (TPS). The gamma-index analysis method was used to quantitatively compare MC and TPS-generated dose distributions.
Results
Concentrating > 97% of the total source strength in a single or pair of central 103Pd seeds produced the most conformal dose distributions, with tumor basal doses a factor of 2-3 higher and OAR doses a factor of 2-3 lower than those of corresponding uniformly-loaded 125I plaques. Concentrating 82-86% of the total source strength in peripherally-loaded 131Cs seeds produced the most homogeneous dose distributions, with tumor basal doses 17-25% lower and OAR doses typically 20% higher than those of corresponding uniformly-loaded 125I plaques. Gamma-index analysis found > 99% agreement between MC and TPS dose distributions.
Conclusions
A method was developed to select intra-plaque ring radionuclide compositions and source strengths to deliver more conformal and homogeneous tumor dose distributions than uniformly-loaded 125I plaques. This method may support coordinated investigations of an appropriate clinical target for eye plaque brachytherapy.
Keywords: eye plaque, TG-43 hybrid technique, dosimetry, conformity index, homogeneity index
Purpose
Investigations into the dosimetric and radiobiological effects of plaque size selection, radionuclide type, prescription dose and plaque positioning for eye plaque brachytherapy have been performed [1–5]. Recommendations exist regarding pre-operative quality assurance (QA) procedures such as source strength verification using a well chamber with a NIST-traceable calibration [6, 7]. Recent efforts have also been made to incorporate material and anatomical heterogeneities into treatment planning for eye plaque brachytherapy [7–11]. Yet despite these advancements, it remains primitive compared to other conventional brachytherapy modalities such as intracavitary accelerated partial breast irradiation and vaginal cylinder high dose-rate 192Ir brachytherapy [12], particularly with regards to conformity of the delivered dose to the clinical tumor volume. Additionally, though professional society recommendations are only available for COMS-plaques [1, 6], non-COMS eye plaque brachytherapy is practiced in many institutions worldwide [13–16]. Standardized and accurate dose calculation methods are required to facilitate evaluation of these dissimilar treatment devices.
The current prescriptive paradigm for eye plaque brachytherapy simply recommends that the prescription isodose line pass through the tumor apex and cover the entire tumor [1]. As no guidelines exist regarding tumor dose uniformity, typical COMS plaque tumor basal doses are a factor of 2.5-8 times higher than those at the apex [9]. For non-COMS plaques, tumor basal doses may be as much as 70 times higher than the prescription dose [17]. Such he-terogeneous dose distributions risk toxicities in the proximal retina and sclera and inhibit identification of an angiogenic choroidal tumor base or a neoplastic advancing apical nevus as the true clinical target for eye plaque brachytherapy. Without knowledge of such a target, physical prescription dose de-escalation studies for eye plaque brachytherapy are unguided, and thus the same prescription dose has persisted for over the last 25 years, in contrast to other ocular melanoma treatment modalities such as stereotactic radiosurgery [18]. A dosimetric advantage of the COMS-plaques is the Silastic seed-carrier insert, which provides a reproducible source arrangement and dosimetric attenuation through tangential angles. To our knowledge, no research exists on the dosimetric effects of optimization methods such as varied intra-plaque source strengths and radionuclide choices or the clinical feasibilities therein for COMS-style eye plaque brachytherapy. The absence of such data restricts further coordinated refinement or dosimetric optimization of COMS-plaque eye brachytherapy towards achieving more conformal and uniform tumor dose distributions with less dose to critical structures. This study endeavors to explore these aspects.
Material and methods
Rings
The TG-43 hybrid technique [19] currently requires a cylindrically-symmetric dose distribution; this is due to treatment planning system (TPS) limitations with respect to TG-43 parameter [20] dataset entry. As such, it is outside the scope of this current investigation to explore the dosimetric effects of varying individual sources within a particular COMS-plaque. It is, however, possible to parse each COMS-plaque into constituent rings of sources, as the Silastic seed-carrier arranges sources about concentric circles of different radii (Fig. 1). In this manner, each COMS-plaque dose distribution may be considered as the superposition of these rings. The dosimetric influence of altering the source strength and radionuclide of component rings can then be explored. In this manuscript, the rings are enumerated in ascending order from the center of the plaque outward for reference: the 16 mm plaque in Fig. 1 thus consists of ring #1 (innermost, 2 sources), ring #2 (middle, 4 sources) and ring #3 (outermost, 7 sources).
Fig. 1.
Example of a 16 mm standard COMS-style eye plaque as viewed from the concave side of the Silastic insert (light grey). The plaque central-axis Z points out of the page (i.e., into the globe). The thirteen sources are designated by rectangles (dark grey), and the Modulay backing with six suture lugs (yellow) are also shown. Three constituent concentric rings of brachytherapy sources are shown (green). Figure adopted from Chiu-Tsao et al. [7]
Following the methodology proposed above, the MCNP5 Monte Carlo (MC) radiation transport code [21] was used to simulate dose distributions for 14, 16 and 18 mm diameter COMS eye plaques with material compositions and reference geometry identical to those employed previously [5, 9]. These plaque sizes were selected to investigate a range of customization capabilities afforded by their distinct numbers of rings (three in the 16 mm plaque, four in the 14 mm plaque and five in the 18 mm plaque). The plaques were populated with each of three distinct low-energy brachytherapy sources: 103Pd (model 200, Theragenics, Corp.®), 125I (model 6711, GE HealthCare, Inc.®) and 131Cs (model CS-1 Rev2, IsoRay Medical, Inc.®). Source design specifications, absorbed dose determination methods and tally geometry were identical to those developed in Gagne et al. [5]. For the three radionuclides, each constituent ring of the three fully-loaded COMS plaques was activated to determine its contribution to the total dose distribution while accounting for the heterogeneous dosimetric influences of inter-seed attenuation and scatter. The dosimetric influence of replacing inactive seeds with air or water was also investigated for the range of plaques and radionuclides in this study to anticipate the possibility that the optimal COMS ring weighting for a particular tumor could involve the removal of one or more rings of seeds.
In a manner similar to Rivard et al. [19], a unique virtual source position and virtual set of brachytherapy dosimetry parameters was then ascribed to each ring. These brachytherapy dosimetry parameters were based on the AAPM TG-43 report formalism [20], and were formatted for Pinnacle3 TPS (Philips Medical Systems, Cleveland, OH) entry with a grid size of 0.2 mm. Such virtual ring sources were used to investigate the dosimetric advantages of using distinct ring weightings with a single radionuclide or with intra-plaque radionuclide rings versus uniformly-loaded single-radionuclide COMS plaques.
Ellipsoidal tumors
Three clinically representative tumors (dome-shaped ellipsoids) with apical heights of 4, 5, and 6 mm, basal diameters of 9, 11, and 13 mm, and volumes of 144, 253, and 399 mm3 were contoured for the 14, 16, and 18 mm COMS plaques, respectively. A MATLAB program (MathWorks, Natick, MA®) was developed to generate tumor dose distributions for all possible ring weighting and radionuclide permutations for a given plaque size and source strength resolution. For example, with three radionuclides (103Pd, 125I, and 131Cs) for each of the three rings of a 16 mm plaque, there are a total of 33 = 27 possible combinations of radionuclide rings; this number scales exponentially with increased source strength resolution. Tumor dose distributions were generated in this study from > 108 permutations of ring weightings and radionuclides for a given apical prescription dose. The total plaque heterogeneity-corrected physical prescription dose from all ring combinations for the three tumors was set to 75 Gy in agreement with the uniformly-loaded 125I plaque results of Melhus and Rivard [9] for equivalent prescription depths as considered in this study. This prescription dose is lower than the nominal 85 Gy – which assumes a homogeneous water calculation environment – to account primarily for the Modulay plaque, Silastic seed-guide insert, and interseed attenuation effects. 125I was selected as the reference radionuclide for comparison due to its use in the COMS protocol.
Unified dosimetry index analysis
These TUMOR DVHs were evaluated using a brachytherapy-specific adaptation of the unified dosimetry index (UDI) of Akpati et al. [22]. The UDI is originally defined as:
1 |
The DI K are indices analyzing tumor coverage (DI 1), dose conformity (DI 2), dose homogeneity (DI 3), and dose gradient (DI 4). The UDI thus incorporates the weighted (W k) contributions from all four components into a single metric that scores the overall deviation of a treatment plan from an ‘ideal’ plan with UDI equal to unity.
Akpati et al. define DI 3, the homogeneity index, as the ratio of the maximum dose at any point 2 mm beyond the PTV to the prescription dose. However, in COMS brachytherapy with the eye plaque sutured directly over the tumor, a more relevant investigation of tumor dose homogeneity would consider points most proximal to the plaque surface, such as at the tumor base. Though alternate homogeneity indices which incorporate volumetric considerations of the target have been investigated [23, 24], our definition of DI 3 in this study preserved the point-dose-ratio nature of the original from Akpati et al. Moreover, since a fixed apical prescription point and appropriate tumor margin ensure excellent tumor coverage, the base-to-apex homogeneity of an eye plaque brachytherapy dose distribution is appropriately examined by investigating TUMOR D MAX.
In an effort to improve clarity of the results and subsequent discussion, the symbolism employed in this study is slightly different than that of Akpati et al. As such, the individual DI k investigated were defined as:
2 |
3 |
4 |
and
5 |
In Eqs. (2)-(5), TUMOR V 100% is the tumor volume receiving 100% of the prescription dose (i.e., 75 Gy), TUMOR V is the entire tumor volume, V 100% is the entire eye volume receiving 75 Gy and, V 50% is the entire eye volume receiving half of the prescription dose (i.e., 37.5 Gy). Volumes attributable to the plaque itself – the Modulay backing, Silastic insert, and radionuclide sources – were excluded from analysis entirely.
By their construction in the above equations, each DI K has an ‘ideal’ score of unity; one thus seeks to minimize these metrics in the treatment planning process. Equal weighting factors W k = 1 in Eq.(1) were assigned to the component metrics in this study. This prevented inflation/deflation of the absolute score, and as such the UDI was confined to a parameter space ≥ 1 (see Fig. 1 from Akpati et al.), with an ‘ideal’ score of unity.
For the three tumor sizes, the component metrics DI K and overall UDI were calculated for each of the TUMOR DVHs, and compared to those of a uniformly-loaded (subscript “REF”) 125I plaque. As will be discussed, the dose gradient of Akpati et al. (Eq. (5)) was found to be an inappropriate metric and was excluded from the overall UDI calculation. Thus, the overall UDI calculated for each candidate and reference plaque loading was determined by:
6 |
TPS dosimetry
To assess dosimetric effects to ocular organs-at-risk (OARs) versus uniformly-loaded 125I plaques, the three plaque loadings responsible for the most conformal (DI 2,MIN), most homogeneous (DI 3,MIN), and best overall (UDI MIN) tumor dose distributions for each plaque size were also simulated in a Pinnacle3 TPS reference eye environment. This reference eye environment contained OAR contours for the ciliary body, cornea, eyelid, lacrimal gland, lens, and optic nerve, as well as dose points for the foveola and optic disc. Dose-volume histograms (DVHs) and maximum, minimum, and average doses from each plaque loading were calculated for these structures for comparison.
Gamma-index analysis
The method of Yang et al. [25] was employed to quantitatively compare the raw MC dose distribution data from a uniformly-loaded 16 mm plaque with the TPS dose distribution data from a summation of the plaque's three constituent rings using the OmniPro I'mRT software suite (IBA Dosimetry, Bartlett, TN). Both dose distributions were calculated with (0.2 mm)2 dose grid resolution, were normalized to 75 Gy at a depth of 5 mm, and comprised a (50 mm)2 field-of-view.
Results
Inactive seed locations
The most dramatic dosimetric effects are observed when all but the active central 131Cs seed of an 18 mm plaque are replaced with water (i.e., the assumption that all empty seed slots are filled with fluid upon implant). In this scenario, central axis doses are nonetheless only < 1.5% higher and off-axis doses only < 2.0% higher within the globe, with the maximum dose increase occurring near the scleral surface beneath the plaque edge. Assuming a 6 mm apical prescription depth of 75 Gy for this plaque, such a removal of radioactive sources leads to a total expansion of the V 50%, V 100% and V 150% isodose volumes of less than 24.0, 8.4, and 1.2 mm3, respectively. These effects were less pronounced for all other plaque sizes and radionuclide combinations considered. For instance, minimum central- and off-axis dose increases of < 0.6% within the globe resulted from an active central 103Pd seed in a 14 mm plaque having all remaining slots filled with water.
Ellipsoidal tumors
DI1, coverage
Since the 2003 ABS report [1] recommends that dose be prescribed to the tumor apex from an approximately cylindrically-symmetric arrangement of typically uniformly-loaded sources beneath and encompassing the base of the tumor, one might assume it is rare that the prescription isodose volume covers less than 95% of the tumor volume. However, a recent study has shown that following these recommendations, COMS eye plaque tumor coverage is a sensitive function of both prescription depth and assumed plaque placement uncertainty, and that D 95% may be as low as 68% of D Rx for certain lesions [26]. Nonetheless, the present study assumes that the plaque is placed perfectly coaxial above a cylindrically-symmetric tumor; likewise the tumor dimension and plaque size combinations in the present study also ensure adequate dose volumetric coverage. Thus, it is unlikely in the present study that any combination of radionuclide rings and source strengths would provide significantly better (or worse) coverage than a uniformly-loaded 125I plaque with DI 1,REF = 1.010, 1.004, and 1.000 for the 14 mm, 16 mm, and 18 mm plaques, respectively.
Also, any combination of rings whose 75 Gy isodose volume at least encompasses the tumor scores a DI 1 of unity, and thus is indistinguishably ‘better’ than any other such combination with respect solely to this index. As such, further evaluation of candidate rings specifically with respect to coverage was abandoned in this study. The component DI 1 scores were nonetheless calculated and incorporated into the analysis of overall UDI for each ring combination as shown in Eq. (6).
DI2, conformity
For the 14 mm plaque, the most conformal tumor dose distribution (Fig. 2) had a DI 2,MIN = 6.2 versus a DI 2,REF = 22.1 for the reference plaque, and was achieved by concentrating > 99% of the total source strength in the center 103Pd source. This improved conformity came at the expense of tumor dose homogeneity, resulting in a DI 3 = 50.4 vs. DI 3,REF = 18.6; TUMOR D MAX was also more than twice that of the reference 14 mm plaque.
Fig. 2.
Ellipsoidal tumor cross-sectional views of the most conformal (DI 2,MIN ), most homogeneous (DI 3,MIN ) and best overall (UDI MIN ) dose distributions as compared to uniformly-loaded 125I (UDI REF ) for the 14 mm, 16 mm and 18 mm plaques. The radionuclide composition (“Pd” = 103Pd, “I” = 125I, “Cs” = 131Cs) and total source strength in U (parenthesized) of each constituent ring assuming a 7 day implant duration are provided at the top of the resultant dose distribution, with ring number increasing from left to right. The component metrics (DI 1 , DI 2 , and DI 3 ) are provided at the bottom of each respective dose distribution, with the overall UDI score in bold at the upper-right corner of each figure. The ellipsoidal tumors are cylindrically-symmetric about X/Y = 0.0. Tumor edges and the inner scleral edge are outlined in black. All volumes outside the inner sclera are in white at the upper-right corner of each figure
For the 16 mm plaque, the most conformal tumor dose distribution (Fig. 2) had a DI 2,MIN = 5.8 versus a DI 2,REF = 18.5 for the reference plaque, and was achieved by again concentrating > 99% of the total source strength in the innermost pair of 103Pd sources. This improved conformity again compromised tumor dose homogeneity, and resulted in a DI 3 = 53.2 vs. DI 3,REF = 19.9. TUMOR D MAX was more than twice that of the reference 16 mm plaque.
For the 18 mm plaque, a minimum DI 2,MIN = 11.7 (Fig. 2) was achieved versus a DI 2,REF = 31.6 by concentrating > 97% of the total source strength in one central 103Pd source, with < 1% of the total source strength in the 131Cs sources filling each of rings #2-#5. This most conformal dose distribution resulted in a DI 3 = 100.2 vs. DI 3,REF = 23.8, and a TUMOR D MAX > 800 Gy, more than three times higher than that for a reference 18 mm plaque.
DI3, homogeneity
The most homogeneous dose distribution for the 14 mm plaque (Fig. 2) had a DI 3,MIN = 11.8 vs. DI 3,REF = 18.6. Rings #1 and #2 were filled with 103Pd sources comprising 5.4% and 8.2% of the total source strength, respectively and rings #3 and #4 were filled with 131Cs sources accounting for the remaining 0.1% and 86.2%, respectively. This resulted in a 23% reduction of TUMOR D MAX compared to a reference 125I plaque, albeit with compromised conformity: DI 2 = 28.1 as compared to DI 2,REF = 22.1.
For the 16 mm plaque, the most homogeneous dose distribution (Fig. 2) gave a DI 3,MIN = 15.1 vs. DI 3,REF = 19.9. Ring #1 was comprised of 103Pd sources having 15.2% of the total source strength and rings #2 and #3 filled with 131Cs sources accounting for the remaining 0.1% and 84.7% of the total source strength, respectively. This resulted in a 17% reduction of TUMOR D MAX compared to a reference 125I plaque, at the expense again of reduced conformity: DI 2 = 21.0 compared to DI 2,REF = 18.5.
For the 18 mm plaque, DI 3,MIN = 15.4 vs. DI 3,REF = 23.8 was achieved (Fig. 2) with rings #1, #4, and #5 filled with 131Cs sources of 0.7%, 0.9% and 82.2% of the total source strength, respectively. Rings #2 and #3 were filled with 103Pd sources of 12.9% and 3.3% of the total source strength, respectively. This loading resulted in a 25% reduction of TUMOR D MAX compared to a reference 125I plaque, though conformity was compromised with DI 2 = 37.7 compared to DI 2,REF = 31.6.
DI4, dose gradient
For the 14 mm plaque, 8 × 106 permutations of radionuclide and source strength could only achieve DI 4,MIN = 6.22 vs. DI 4,REF = 6.82. The two other plaque sizes in this study demonstrated similar insensitivity of this metric to varied ring compositions and weightings. As discussed below, this index was found to be dosimetrically irrelevant for the purposes of the current study, and was excluded from calculation of the overall UDI results below.
UDI
Based on the DI 4 results above, the UDI in this study was a weighted product of DI 1, DI 2, and DI 3 only as expressed in Eq. (6). For the 14 mm 125I plaque, the lowest UDI (Fig. 2) was achieved using 103Pd in all four rings, with rings #1-#4 contributing 5.5%, 0.2%, 94.2% and 0.1% of the total source strength, respectively. This loading resulted in components DI 1 = 1.01, DI 2 = 15.8, and DI 3 = 18.3 for an overall UDI MIN = 292 vs. UDI REF = 414.
For the 16 mm plaque, the lowest UDI (Fig. 2) was achieved using 103Pd in all three rings with rings #1-#3 contributing 12.7%, 86.7% and 0.6% of the total source strength, respectively. This loading resulted in components DI 1 = 1.01, DI 2 = 11.8, and DI 3 = 24.1 for an overall UDI MIN = 285 vs. UDI REF = 371.
For the 18 mm plaque, the lowest UDI (Fig. 2) was achieved using 125I in ring #1, 103Pd in rings #2 and #3, and 131Cs in rings #4 and #5; the rings contributed 1.7%, 5.7%, 19.1%, 2.4% and 71.1% of the total source strength, respectively. This loading resulted in components DI 1 = 1.00, DI 2 = 35.7, and DI 3 = 15.6 for an overall UDI MIN = 558 vs. UDI REF = 750.
TPS dosimetry
Dose distributions in the reference eye environment for the 14 mm plaque loadings described above are shown in Figure 3 with resultant OAR DVHs shown in Figure 4. Table 1 provides OAR dose information for all three plaque sizes.
Fig. 3.
Pinnacle3 TPS dose distributions for the most conformal (DI 2,MIN ), most homogeneous (DI 3,MIN ), best overall (UDI MIN ), and uniformly-loaded 125I (UDI REF ) 14 mm plaque loadings visualized on an axial (inferior/superior) cross section of a right (OD) eye at the nasal equator of the globe. OAR contours include the ciliary body (aqua), cornea (yellow), eyelid (beige), lens (red), and optic nerve (orange). Concentric circles establish the inner- and outer scleral boundaries (white), and a dose point simulates the optic disc (orange crosshair). The lacrimal gland and foveola are not pictured. Isodose lines are provided for 1, 5, 10, 30, 50, 65, 75, 85, 100, 150, 300, and 500 Gy, with the thick 75 Gy prescription isodose line in green
Fig. 4.
OAR DVHs for the corresponding 14 mm plaque dose distributions from Figure 3
Table 1.
OAR dosimetric information for the most conformal (DI 2), most homogeneous (DI 3), best overall (UDI MIN), and uniformly-loaded 125I (UDI REF) 14 mm, 16 mm, and 18 mm plaque solutions positioned at the nasal equator of the globe, as shown in Figure 3. Minimum, mean, and maximum OAR doses are provided where appropriate and the reference plaque results are emboldened and italicized. Only point doses are provided for the foveola and optic disc, as these were simply dose calculation points rather than volumetric OARs. The ‘–‘ symbol for the eyelid and optic nerve indicate that these OARs had volumes outside the dose calculation grid which therefore received no physical dose. All values presented are in units Gy
14 mm | 16 mm | 18 mm | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DI 2 | DI 3 | UDI MIN | UDI REF | DI 2 | DI 3 | UDI MIN | UDI REF | DI 2 | DI 3 | UDI MIN | UDI REF | ||
Ciliary body | min | 1.80 | 5.44 | 2.33 | 4.37 | 2.61 | 6.83 | 3.09 | 5.89 | 3.52 | 9.22 | 8.72 | 7.73 |
mean | 8.05 | 19.11 | 11.40 | 16.15 | 11.87 | 25.49 | 15.44 | 22.82 | 15.59 | 37.02 | 35.35 | 31.42 | |
max | 59.23 | 145.11 | 106.13 | 123.10 | 92.05 | 175.99 | 146.18 | 167.14 | 113.22 | 191.93 | 188.67 | 195.91 | |
Cornea | min | 2.12 | 6.22 | 2.78 | 5.05 | 3.08 | 7.86 | 3.70 | 6.85 | 4.15 | 10.68 | 10.10 | 9.02 |
mean | 3.96 | 9.22 | 5.27 | 8.14 | 5.76 | 13.05 | 7.04 | 11.68 | 7.66 | 17.67 | 16.85 | 15.45 | |
max | 9.25 | 19.66 | 12.71 | 17.57 | 13.54 | 28.48 | 17.11 | 25.57 | 17.85 | 41.56 | 39.64 | 35.17 | |
Eyelid | min | 0.45 | 1.90 | 0.74 | 1.54 | 0.54 | 1.82 | 0.67 | 1.53 | – | – | – | – |
mean | 3.02 | 6.52 | 3.61 | 5.96 | 4.13 | 8.64 | 4.69 | 7.99 | 5.14 | 10.54 | 10.10 | 9.70 | |
max | 10.49 | 14.67 | 9.93 | 14.87 | 13.64 | 19.13 | 14.08 | 19.65 | 15.15 | 22.72 | 22.43 | 23.54 | |
Foveola | 3.74 | 9.55 | 4.94 | 8.09 | 5.44 | 12.47 | 6.60 | 11.06 | 7.25 | 17.16 | 16.32 | 14.68 | |
Lacrimal gland | min | 0.94 | 3.42 | 1.22 | 2.66 | 1.36 | 4.30 | 1.62 | 3.59 | 1.87 | 5.79 | 5.43 | 4.71 |
mean | 1.50 | 4.75 | 1.95 | 3.81 | 2.17 | 6.01 | 2.59 | 5.16 | 2.94 | 8.14 | 7.67 | 6.78 | |
max | 2.99 | 7.72 | 3.94 | 6.62 | 4.34 | 10.45 | 5.26 | 9.21 | 5.81 | 14.26 | 13.53 | 12.18 | |
Lens | min | 3.42 | 8.85 | 4.47 | 7.32 | 4.96 | 11.14 | 5.94 | 9.90 | 6.64 | 15.12 | 14.40 | 13.02 |
mean | 6.82 | 15.34 | 8.97 | 12.98 | 9.90 | 19.35 | 11.95 | 17.59 | 13.11 | 26.48 | 25.42 | 23.25 | |
max | 14.33 | 29.13 | 19.12 | 25.03 | 20.85 | 36.77 | 25.54 | 34.03 | 27.38 | 51.13 | 49.30 | 45.19 | |
Optic disc | 7.29 | 13.27 | 9.90 | 13.00 | 10.66 | 21.59 | 13.31 | 20.12 | 14.05 | 29.39 | 28.31 | 26.94 | |
Optic nerve | min | – | – | – | – | – | – | – | – | – | – | – | – |
mean | 1.42 | 2.91 | 1.61 | 2.69 | 1.87 | 3.72 | 2.06 | 3.47 | 2.16 | 4.33 | 4.15 | 3.97 | |
max | 9.71 | 18.24 | 13.45 | 17.25 | 14.25 | 29.87 | 18.17 | 27.24 | 18.80 | 44.71 | 42.63 | 38.11 |
Gamma-index analysis
Employing the most stringent comparative criteria (i.e. a dose difference of 1%, a distance difference of 1 mm, a search distance of 1 mm, and a 5% minimum relative signal) of the OmniPro I'mRT software, the gamma-index analysis of the full plaque MC and TPS rings dose distributions resulted in a 99.03% pass rate. These gamma-index results are shown in Figure 5.
Fig. 5.
Gamma-index analysis of the MC full 125I 16 mm plaque dose distribution versus the TPS dose distribution from the sum of the three constituent 16 mm 125I plaque rings. Both dose distributions were normalized to 75 Gy at a depth of 5 mm. Gamma-index results, parameters, and passing rate information are provided in the right-most image
Discussion
Inactive seed locations
Of the permutations of source ring removal investigated in this study, the most clinically relevant – based on the DI 2 results (Fig. 2 and Table 1) above – is perhaps the case of an activated center seed(s) with all other sources removed. The case of a single central 131Cs source in an 18 mm plaque somewhat unsurprisingly produces the largest of the observed dosimetric effects in this study. Firstly, this source position possesses the largest number of proximal neighbors, and is thus subject to the largest attenuation and scatter influences from the replacement of inactive surrounding high-Z seeds with lower-Z media such as air or water. Secondly, the higher energy (∼30 keV) photon emissions from 131Cs likewise demonstrate the most pronounced lateral dosimetric effects as compared to the lower energy emissions from 125I and 103Pd. Lastly, this scenario involves replacement of > 95% (20 out of 21) of the plaque's seeds, whereas retention of any other radionuclide ring involves necessarily less substantial changes to the dosimetric environment (i.e., it intrinsically possesses more sources). Nonetheless, the results of this “worst case scenario” above indicate a negligible dosimetric effect due to the removal of inactive rings of seeds for the purpose of eye plaque brachytherapy dosimetric optimization. Similar dosimetric effects have been investigated previously and also deemed to constitute less than a 2% variation from a fully loaded plaque [27]. This is due in part to the source configurations of the COMS plaques having been designed explicitly to prevent interseed attenuation and shadowing and also attributed to the attenuating Silastic seed carrier insert.
Ellipsoidal tumors
UDI calculations
The UDI and component metrics of Akpati et al. were evaluated as applied to eye plaque brachytherapy in this study. As mentioned in the results above, tumor coverage (Eq. (1)) was not a significant concern given the assumptions of the current study, as well as the tumor dimensions and plaque sizes employed herein. As such, the optimization of this specific metric using the MATLAB code returned arbitrarily many candidate solutions whose dose distributions each resulted in the tumor volume being entirely encompassed by the 75 Gy isodose volume. For instance, all plaque loadings in Figure 2 provide nearly perfect coverage. These solutions are thus essentially equivalent with respect to coverage and are therefore of little interest to the reader or clinician in themselves. This metric was retained in the calculation of overall UDI, however, to emphasize the importance of tumor dose coverage while striving for better conformity or homogeneity.
Defined in Eq. (3) as the ratio of the entire eye volume (plaque excluded) receiving 75 Gy to the tumor volume, dose conformity for eye plaque brachytherapy is necessarily reduced by the presence of 1 mm thick sclera between the plaque and the tumor. Additionally, since plaque selection incorporates a 2-3 mm margin around the tumor base for conservative coverage [1], peripheral conformity to the tumor volume is largely absent in traditional uniformly-loaded plaques. Thus, the dosimetric flexibility provided by varied ring weightings and radionuclides resulted in the greatest improvement of this metric for all three ellipsoidal tumors. In this manner, DI 2 also became the most important component index in UDI calculations for an ‘optimal’ loading of the 14 mm and 16 mm plaques, assuming each W k = 1.
As shown in Figure 3 and Figure 4 for the 14 mm plaque, the most conformal plaque loadings resulted in the greatest OAR sparing of those considered. Minimum, average, and maximum OAR doses were also typically 2 times lower for all plaque sizes than corresponding uniformly-loaded 125I plaques (Table 1). This critical structure sparing, however, is offset by the 2x-3x increase in the tumor base dose versus that of the reference plaques, which likewise implies higher doses to the proximal sclera. These most conformal plaque loadings each employed a single central 103Pd seed (or pair of seeds for the 16 mm plaque) containing > 97% of the total source strength to deliver the requisite dose distributions. This corresponded to source strengths of 23-57 U depending on plaque size. Such high-strength sources are beyond the current production capabilities of commercial brachytherapy source manufacturers, and also pose significant challenges to clinical implementation regarding health physics and quality assurance concerns [28].
Modification of the original DI 3 from Akpati et al. in favor of that considered in this study – the ratio of the maximum and apical prescription tumor doses – nonetheless resulted in improved tumor dose homogeneity versus the reference plaques as shown in the second column of Figure 2. It is unlikely that more homogeneous tumor dose distributions could be produced using current COMS-style brachytherapy applicators given the limited range of photon energies for the three commercially-available radionuclides considered in this study, as well as the limited number of discrete source rings per plaque. This homogeneity came largely at the expense of conformity (Fig. 2) and OAR sparing (Figs. 3 and 4), however, due to the high strength 131Cs seeds at the plaque peripheries. Though TUMOR D MAX was reduced by 17-25% depending on plaque size, the OAR doses in Table 1 were typically 1.2 times higher than those from the reference plaques.
As defined in Eq. (5), the dose gradient is a measure of the ratio of the prescription isodose volume, V 100%, to the half-prescription isodose volume, V 50%. For eye plaque brachytherapy, an ideal optimization of this index thus seeks to minimize the difference of these isodose volumes by shrinking the V 50% more proximal to the tumor volume (sharper dose falloff), but the same mathematical effect is achieved by instead expanding the V 100% to encompass more volume beyond the tumor while maintaining or decreasing the V 50%. The former is dosimetrically advantageous for eye plaque treatments and results in less dose to normal ocular structures adjacent to or opposite of the plaque, whereas the latter increases the volume of healthy eye receiving 75 Gy and risks increasing ocular toxicities. The three radionuclides in this study represent a relatively narrow range of average photon energies (20–30 keV) whose physical properties unfortunately prohibit significant reductions in V 50% while maintaining a fixed apical prescription dose. The dose gradient of COMS eye plaque brachytherapy implants is thus somewhat limited in its improvement. The DI 4 solutions in this study with expanded V 100%, however, point to the hazards of simple parameter optimization methodologies and underscore the importance of guided and thoughtful algorithmic programming. Applying a significantly lower weighting W 4 to this metric while increasing the DI 2 weighting W 2 may have constrained the MATLAB program to retain solutions with both improved conformity and steeper dose gradients, thus permitting its inclusion in the overall UDI calculations. The decision to maintain equal weightings W K in this study, however, was deliberate to emphasize the ambiguities surrounding selection of a clinically ‘optimal’ tumor dose distribution.
The best overall (i.e., UDI MIN) plaque loadings for the 14 mm and 16 mm plaques produced more conformal dose distributions than the corresponding uniformly-loaded 125I plaques, with typically 1.6 times lower average OAR doses (Table 1). For the 18 mm plaque, however, the plaque loading resulting in the UDI MIN produced instead a more homogeneous tumor dose distribution than the reference loading, despite being less conformal. For this plaque size, the DI 3 was the dominant component metric in the overall UDI calculations: with a 6 mm apical prescription depth, the conformity resulting from focusing source strength solely on inner plaque rings results in higher tumor base doses which quickly sabotage the DI 3 score as seen in the first column of Figure 2. Thus, the optimal plaque loading positioned 71% of the total source strength in 131Cs seeds in the outermost ring. Resultant OAR doses were typically 10% higher than those for the reference 18 mm plaque.
Gamma-index analysis
The gamma-index method for dosimetry quality analysis has been developed and widely used for IMRT patient QA to compare measured dose distributions with those generated from the treatment planning system. Application of this method directly to brachytherapy dosimetry is difficult since the clinical medical physicist does not measure brachytherapy dose distributions. Thus, an alternate application of the gamma-index method for brachytherapy dose distributions is to compare reference and planned results as investigated in Yang et al. [25]. The uniformly-loaded 125I plaque MC results represent the reference dose distributions in this study. The methods of this study parsed these full MC plaques into their individual rings for simulation, formatted virtual source parameter datasets from these MC ring results, entered these datasets into the TPS via the TG-43 hybrid technique, and created individual TPS ring dose distributions. By comparing these resultant TPS ring dose distributions against the standard fully-loaded plaque MC results, a quantitative dosimetric analysis of the entire process is permitted. The gamma-index results for the two 16 mm 125I dose distributions in Figure 5 demonstrate agreement within 1%, and provide the dosimetric QA necessary for clinical adaptability of this technique.
Implementation
The candidate plaque loadings examined in this study offering improved conformity or homogeneity nonetheless present impediments to clinical implementation due to their employment of mixed radionuclides and non-uniform source strength weightings. Melhus and Rivard [9] provide plaque heterogeneity-corrected physical prescription doses for the range of COMS plaques yet assume the use of a single radionuclide per plaque. It is thus unclear what corrections should be applied to the ABS recommended homogeneous 85 Gy apical prescription dose for plaques containing a variety of radionuclide rings. Varied intra-plaque radionuclide rings and source strength weightings also presents challenges to traditional clinical brachytherapy QA procedures. The importance of source distinction throughout the QA process from source strength verification to loading the Silastic insert could not be overstated in these instances.
As mentioned previously, the conformity requirements of a particular tumor dose distribution may lead to high concentrations of source strength at the upper tolerances of or even outside of current brachytherapy source manufacturer capabilities. Such high-strength sources may prove prohibitively expensive if not impossible to manufacture, and would present unique clinical concerns regarding personnel shielding and brachytherapy QA as well. Additionally, source manufacturers cannot always make seeds at the desired strength due to batching practices. This will quantize the available source strengths, especially for 125I, and limit the potential UDI min.
Further, minimization of UDI min will be limited based on practical concerns such as potential for clinical medical physicists to measure low-energy source strengths to an accuracy (k = 2) of about 3%.
Limitations and future directions
Though the dosimetric effects of inactive seed removal have been investigated and found negligible for the range of radionuclides and COMS plaques in this study, the plaque heterogeneity-corrected MC ring results presented herein assume that all Silastic seed slots in the plaque are filled with brachytherapy sources, active or otherwise. Additionally, these results are specific to the model 200 (103Pd, Theragenics Corp.®), model 6711 (125I, Oncura/GE HealthCare Inc.®) and model CS-1 Rev2 (131Cs, IsoRay Medical Inc.®) sources. The relatively narrow range of lower photon energies offered by these commercially-available radionuclide sources also necessarily restricts the tumor dose homogeneity afforded by their utilization in an episcleral treatment methodology such COMS plaque brachytherapy.
The dosimetric customization demonstrated by the various plaque loadings in this study is necessarily limited by the fixed number of rings in the COMS plaque. Other plaque types with significantly greater numbers of sources [29] or surface patch-source modeling capabilities [30] likely provide improved customization. A further constraint in this investigation is the requirement that all brachytherapy sources within a given ring contain the same radionuclide. This limitation is due to the dosimetry parameter input requirements of current commercial TPS which confine the TG-43 hybrid technique [19] to cylindrically-symmetric calculation geometries. Further dosimetric customization is allowed by varying individual brachytherapy source strengths and radionuclides throughout the plaque, though these aspects are outside the scope of the current study.
The UDI of Akpati et al. was modified in this study to evaluate eye plaque brachytherapy dose distributions for coverage, conformity, and homogeneity. These dosimetric characteristics were considered with equal importance in the optimization process, though this method permits unequal weighting and thus custom target prioritization and OAR sparing on a per-patient basis. The DI K employed in these investigations are merely a sampling of the various quantitative metrics and evaluation criteria available to the clinical medical physicist, yet no such consensus objective measures currently exist towards identifying an ‘optimal’ treatment specifically for eye plaque brachytherapy. Future research should investigate both the suitability and discriminatory ability of other evaluative metrics for this treatment modality.
Conclusions
A method has been developed for COMS plaque brachytherapy to select intra-plaque ring radionuclides and source strengths on an individual, per-patient basis to deliver more conformal and more homogeneous tumor dose distributions than a uniformly-loaded 125I plaque. The accurate, plaque heterogeneity-corrected dosimetry employed herein supports development of a prospective clinical protocol to determine the appropriate target (e.g., tumor base, tumor apex, tumor periphery, etc.) for eye plaque brachytherapy.
References
- 1.Nag S, Quivey JM, Earle JD, et al. The American Brachytherapy Society recommendations for brachytherapy of uveal melanomas. Int J Radiat Oncol Biol Phys. 2003;56:544–555. doi: 10.1016/s0360-3016(03)00006-3. [DOI] [PubMed] [Google Scholar]
- 2.Rivard MJ, Melhus CS, Sioshansi S, et al. The impact of prescription depth, dose rate, plaque size, and source loading on the central axis using 103Pd, 125I, and 131Cs. Brachytherapy. 2008;7:327–335. doi: 10.1016/j.brachy.2008.05.002. [DOI] [PubMed] [Google Scholar]
- 3.Leonard KL, Gagne NL, Mignano JE, et al. A 17-year retrospective study of institutional results for eye plaque brachytherapy of uveal melanoma using 125I, 103Pd, and 131Cs and review of the literature. Brachytherapy. 2011;10:331–339. doi: 10.1016/j.brachy.2011.01.006. [DOI] [PubMed] [Google Scholar]
- 4.Gagne NL, Leonard KL, Huber KE, et al. BEDVH – a method for evaluating biologically effective dose volume histograms: Application to eye plaque brachytherapy implants. Med Phys. 2012;39:976–983. doi: 10.1118/1.3679010. [DOI] [PubMed] [Google Scholar]
- 5.Gagne NL, Leonard KL, Rivard MJ. Radiobiology for eye plaque brachytherapy and evaluation of radionuclide choice and implant duration using an objective function. Med Phys. 2012;39:3332–3342. doi: 10.1118/1.4718683. [DOI] [PubMed] [Google Scholar]
- 6.Butler WM, Bice WS, Jr, DeWerd LA, et al. Third-party brachytherapy source calibrations and physicist responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group. Med Phys. 2008;35:3860–3865. doi: 10.1118/1.2959723. [DOI] [PubMed] [Google Scholar]
- 7.Chiu-Tsao S-T, Astrahan MA, Finger PT, et al. Dosimetry of 125I and 103Pd COMS eye plaques for intraocular tumors: Report of Task Group 129 by the AAPM and ABS. Med Phys. 2012;39:6161–6184. doi: 10.1118/1.4749933. [DOI] [PubMed] [Google Scholar]
- 8.Astrahan MA. Improved treatment planning for COMS eye plaques. Int J Radiat Oncol Biol Phys. 2005;61:1227–1242. doi: 10.1016/j.ijrobp.2004.09.062. [DOI] [PubMed] [Google Scholar]
- 9.Melhus CS, Rivard MJ. COMS eye plaque brachytherapy dosimetry simulations for 103Pd, 125I, and 131Cs. Med Phys. 2008;35:3364–3371. doi: 10.1118/1.2940604. [DOI] [PubMed] [Google Scholar]
- 10.Thomson RM, Rogers DWO. Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy with various seed models. Med Phys. 2010;37:368–376. doi: 10.1118/1.3271104. [DOI] [PubMed] [Google Scholar]
- 11.Rivard MJ, Chiu-Tsao S-T, Finger PT, et al. Comparison of dose calculation methods for brachytherapy of intraocular tumors. Med Phys. 2011;38:306–316. doi: 10.1118/1.3523614. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Rivard MJ, Venselaar JLM, Beaulieu L. The evolution of brachytherapy treatment planning. Med Phys. 2009;36:2136–2153. doi: 10.1118/1.3125136. [DOI] [PubMed] [Google Scholar]
- 13.Lommatzsch PK. Results after beta-irradiation (106Ru/106Rh) of choroidal melanomas: 20 years’ experience. Br J Ophthalmol. 1986;70:844–885. doi: 10.1136/bjo.70.11.844. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Granero D, Perez-Calatayud J, Ballester F, et al. Dosimetric study of the 15 mm ROPES eye plaque. Med Phys. 2004;31:3330–3336. doi: 10.1118/1.1812605. [DOI] [PubMed] [Google Scholar]
- 15.Finger PT. Finger's “slotted” eye plaque for radiation therapy: treatment of juxtapapillary and circumpapillary intraocular tumours. Br J Ophthalmol. 2007;91:891–894. doi: 10.1136/bjo.2007.114082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Thomson RM, Furutani KM, Pulido JS, et al. Modified COMS plaques for 125I and 103Pd iris melanoma brachytherapy. Int J Radiat Oncol Biol Phys. 2010;78:1261–1269. doi: 10.1016/j.ijrobp.2009.12.002. [DOI] [PubMed] [Google Scholar]
- 17.Flühs D, Anastassiou G, Wening J, et al. The design and the dosimetry of bi-nuclide radioactive ophthalmic applicators. Med Phys. 2004;31:1481–1488. doi: 10.1118/1.1755471. [DOI] [PubMed] [Google Scholar]
- 18.Schirmer CM, Chan M, Mignano JE, et al. Dose de-escalation with Gamma Knife radiosurgery in the treatment of choroidal melanoma. Int J Radiat Oncol Biol Phys. 2009;75:170–176. doi: 10.1016/j.ijrobp.2008.10.077. [DOI] [PubMed] [Google Scholar]
- 19.Rivard MJ, Melhus CS, Granero D, et al. An approach to using conventional brachytherapy software for clinical treatment planning of complex, Monte Carlo-based brachytherapy dose distributions. Med Phys. 2009;36:1968–1975. doi: 10.1118/1.3121510. [DOI] [PubMed] [Google Scholar]
- 20.Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31:633–674. doi: 10.1118/1.1646040. [DOI] [PubMed] [Google Scholar]
- 21.X-5 Monte Carlo Team. Los Alamos: Los Alamos National Laboratory; 2003. MCNP: a general Monte Carlo N-particle transport code, version 5. [Google Scholar]
- 22.Akpati H, Kim C, Kim B, et al. Unified dosimetry index (UDI): a figure of merit for ranking treatment plans. J Appl Clin Med Phys. 2008;9:99–108. doi: 10.1120/jacmp.v9i3.2803. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Chaswal V, Yoo S, Thomadsen BR, et al. Multi-species prostate implant treatment plans incorporating 192Ir and 125I using a Greedy Heuristic based 3D optimization algorithm. Med Phys. 2007;34:436–444. doi: 10.1118/1.2400827. [DOI] [PubMed] [Google Scholar]
- 24.Yoon M, Park S, Shin D, et al. A new homogeneity index based on statistical analysis of the dose volume histogram. J Appl Clin Med Phys. 2007;8:9–17. doi: 10.1120/jacmp.v8i2.2390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Yang Y, Melhus CS, Sioshansi S, et al. Treatment planning of a skin-sparing conical breast brachytherapy applicator using conventional brachytherapy software. Med Phys. 2011;38:1519–1525. doi: 10.1118/1.3552921. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Gagne NL, Rivard MJ. Quantifying the dosimetric influences of COMS eye plaque radiation coverage and brachytherapy implant placement uncertainty on plaque size selection; Brachytherapy; 2012-Ph16, accepted. [DOI] [PubMed] [Google Scholar]
- 27.Thomson RM, Taylor REP, Rogers DWO. Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy. Med Phys. 2008;35:5530–5543. doi: 10.1118/1.3002412. [DOI] [PubMed] [Google Scholar]
- 28.Williamson JF, Dunscombe PB, Sharpe MB, et al. Quality assurance needs for modern image-based radiotherapy: recommendations from 2007 interorganizational symposium on quality assurance of radiation therapy: challenges of advanced technology. Int J Radiat Oncol Biol Phys. 2008;71:S2–S12. doi: 10.1016/j.ijrobp.2007.08.080. [DOI] [PubMed] [Google Scholar]
- 29.Astrahan MA, Luxton G, Pu Q, et al. Conformal episcleral plaque therapy. Int J Radiat Oncol Biol Phys. 1997;39:505–519. doi: 10.1016/s0360-3016(97)00118-1. [DOI] [PubMed] [Google Scholar]
- 30.Astrahan MA. A patch source model for treatment planning of ruthenium ophthalmic applicators. Med Phys. 2004;30:1219–1228. doi: 10.1118/1.1573971. [DOI] [PubMed] [Google Scholar]