Abstract
For MR applications such as contrast-enhanced MR angiography (CE-MRA), it is desirable to achieve simultaneously high spatial and temporal resolution. The current clinical standard uses view sharing methods combined with parallel imaging; however this approach still provides limited spatial and temporal resolution. To improve on the clinical standard, we present an Interleaved Variable Density (IVD) sampling method that pseudorandomly undersamples each individual frame of a 3D Cartesian ky-kz plane combined with parallel imaging acceleration. From this data set, time-resolved images are reconstructed with a method that combines parallel imaging with a multiplicative constraint. Total acceleration factors on the order of 20 are achieved for CE-MRA of the lower extremities, and improvements in temporal fidelity of the depiction of the contrast bolus passage are demonstrated relative to the clinical standard.
Keywords: Interleaved variable density sampling, constrained reconstruction, contrast-enhanced MR angiography, parallel imaging, fast MRI, HYPR
INTRODUCTION
3D time-resolved MR angiography (MRA) is a widely used technique in clinic for the diagnosis of a variety of vascular disease (1). This technique often requires high spatial and temporal resolution to achieve better visualization of blood vessels as well as the ability to separate of arterial from venous phases. However, for applications that employ large sampling matrices, such as imaging the peripheral vasculature, the Nyquist criterion precludes the possibility of obtaining both goals simultaneously using conventional Fourier encoding techniques. Acquisition strategies that undersample k-space (with respect to the Nyquist criterion) and reconstruction algorithms that minimize artifacts are needed to advance the capabilities of MRI as a dynamic imaging modality.
To meet this need, several Cartesian acquisition methods, such as CAPR (2,3), TRIPPS (4), and random sampling (5), have been reported. CAPR and TRIPPS are similar approaches in that for every “time frame” that is reconstructed, the center of the phase-encoding plane (ky-kz plane) is fully sampled on the accelerated (CAPR) or original (TRIPPS) Cartesian grid, and the peripheral region is undersampled by collecting views either within radial sectors (CAPR) or along radial lines (TRIPPS). Over several time frames, the undersampled views are interleaved, producing an average k-t density in the periphery that is constant (CAPR) or falls off radially as approximately 1/kr (TRIPPS). Random sampling is an alternative approach that can be customized to achieve more scattered views in the ky-kz plane with the corresponding artifacts less coherent in image space.
It is well understood that reconstructing undersampled k-space data leads to artifacts in image space. Reconstructing data from undersampled Cartesian acquisition methods in which views are grouped along radial sectors or lines, i.e. CAPR or TRIPPS, results in streak-like artifacts appearing in the images, similar to those observed when an undersampled radial stack of stars acquisition is reconstructed. However, reconstructing undersampled Cartesian data in which the views are randomly arranged results in less coherent, speckle-like image artifacts, rather than streaks (6). Several techniques have been developed to ameliorate the effects of undersampling. Temporal view-sharing methods (2,4,7) are techniques that are used to approximate high frequency data for a given time frame by using data from neighboring time frames prior to the image reconstruction of the selected time frame. However, these techniques may lead to compromises in the temporal resolution of the reconstructed images (8,9). Parallel imaging methods allow for acceleration of the data acquisition by regular undersampling, and use the coil sensitivity information to unfold the images in the reconstruction. Among these methods, the coil sensitivity information is used either explicitly (i.e. model-driven) such as SMASH(10), SENSE (11) and GEM (12), or implicitly (i.e. data-driven) such as AUTO-SMASH (13), VD-AUTO-SMASH (14) and GRAPPA (15). The combination of view-sharing and parallel imaging is able to improve the temporal resolution of dynamic MR imaging (2,3). Other reconstruction approaches to improve time-resolved imaging include model-based reconstruction methods (16,17) and constrained reconstruction methods including compressed sensing (5,18–20) and algorithms in the HYPR-like family (21–25). Compressed sensing requires that the reconstructed image be sparse in a known transformed domain, and HYPR methods have been demonstrated to be capable of suppressing radial undersampling artifacts while improving temporal fidelity during in-vivo studies (26).
We have recently proposed an Interleaved Variable Density (IVD) sampling method in which the 3D Cartesian ky-kz plane matrix is pseudorandomly sampled such that data from individual time frames are undersampled but when data from multiple time frames are combined a fully-sampled data set may be achieved (9). We also have proposed that incoherent artifacts caused by the pseudorandom undersampling may be suppressed by applying a multiplicative constraint similar to that used by the HYPR techniques (9,27). In this work, we present a novel technique that integrates parallel imaging (PI) with an IVD sampling method and the multiplicative constrained reconstruction, referred to here as IVD HYCR (HighlY constrained Cartesian Reconstruction). IVD HYCR uses the combined undersampling resulting from PI and IVD to allow total acceleration factors on the order of 20 relative to fully-sampled Cartesian imaging when applied to time-resolved contrast-enhanced peripheral vascular imaging.
THEORY
Sampling
Let S(ky,kz) represent all view locations in the ky-kz (phase-encoding) plane that are sampled during the course of a 3D acquisition. Anticipating that a parallel-imaging reconstruction will be used, S(ky,kz) may have view spacing that violates the Nyquist criterion over the FOV to be imaged (in either or both phase encode directions), yielding a regularly-undersampled pattern as shown in Fig. 1(a). For the work shown here, a typical acceleration factor of 4 (2 × 2) was used. To further accelerate the acquisition, let st(ky,kz) be a single subset of these k-space locations sampled within the “time frame” t. This subset of views may be subsampled from S(ky,kz) in a pseudorandom fashion given by a pre-determined k-t density function, such as 1/kr as shown in Fig. 1(b). Acceleration factors of 3–6 from the pseudorandom undersampling were empirically found to cause minimal incoherent undersampling artifacts producing a total acceleration factor in the range of 12–24 shown in Fig. 1(c). The pseudorandom subsampling patterns can be interleaved such that over N time frames, all sampling locations in S(ky,kz) are sampled at least once. The sampling pattern is calculated for each individual scan based on patient-specific FOV (i.e. ky-kz matrix size), parallel imaging factor along ky and kz directions, and the desired temporal resolution while maintaining the 1/kr fall-off. This acquisition strategy is referred to as Interleaved Variable Density (IVD) sampling.
Reconstruction
In this section, we first assume that S(ky,kz) meets the Nyquist criterion (e.g. no parallel-imaging acceleration) and demonstrate the principles of the multiplicative constrained reconstruction. After that, the integration of parallel imaging is presented.
Let Kt(ky,kz) be the k-space data acquired at locations st(ky,kz) during a time frame t. If S(ky,kz) met the Nyquist criterion, then a Fourier transform of the Kt(ky,kz) would yield an image, It, albeit one with spatially blurring due to the variable sampling density and incoherent artifacts due to the unacquired “holes” left in k-space.
Let K̅t(ky,kz) be the k-space data integrated over a sliding window of N frames. This integrated data set (one per time frame) also could be reconstructed using a Fourier transform to yield an image Īt, free of spatial blurring and incoherent artifacts, but temporally blurred due to integrating data over several time frames. It should be noted that the integration uses a density correction factor to account for the variable k-t sampling density.
Let K̃t(ky,kz) be the subset of k-space points from the integrated data set corresponding to the same sampled locations given by st(ky,kz):
[1] |
In this formulation, st(ky,kz) can be treated as the sampling function used to subsample K̅t(ky,kz). A Fourier transform of these data yields an image Ĩt. It should be noted that both It and Ĩt have the same data sampling pattern st(ky,kz) and therefore have the same blurring and similar incoherent artifact due to the point spread function of st(ky,kz).
Finally, these three images (per time frame t) are combined in a multiplicative constrained reconstruction described by the following function:
[2] |
where c is a small regularization constant to avoid zero division. This operation has the effect of imposing the dynamic characteristics of It upon the spatially sharp and fully-sampled constraining image Īt. The undersampling artifacts that appear in It and Ĩt tend to cancel because they are derived from the same undersampling function, st(ky,kz). This multiplicative constraint is similar to the image-space variation of HYPR reconstruction described in (23), and similar to the “local reconstruction” variation of HYPR described in (22).
The integration of parallel imaging is achieved by applying a parallel imaging reconstruction method prior to the multiplicative constrained reconstruction. More specifically, to unfold the aliased images, a parallel imaging reconstruction method is applied to generate the images It, Īt and Ĩt prior to combining these into the final image Ît. Unaliasing coefficients may be derived from a separate calibration scan or from data acquired during the acquisition. A diagram summarizing the proposed method with parallel imaging is shown in Fig. 2.
In the experiments presented here, all parallel imaging (PI) reconstructions were performed using ARC (GE Healthcare, Waukesha, WI, USA): proprietary software that employs an indirect calibration procedure (28) to generate unaliasing coefficients in k-space together with hybrid-space data synthesis, described in Ref. (29) as “Method 5”. The software was instructed to generate unaliasing coefficients for the regularly undersampled pattern S(ky,kz) (not st(ky,kz)), using data from a pre-contrast acquisition in which the center of k-space was fully sampled. In the synthesis phase, the software was instructed to apply these coefficients to data sets Kt, K̃t and K̅t. As demonstrated in the experiments below, applying parallel imaging in such a fashion suppresses coherent (parallel imaging) artifacts; the missing data propagates into the intermediate reconstructed images as incoherent artifacts, but such artifacts are largely suppressed in the final image by the constrained reconstruction (application of Eq. [2]).
MATERIALS AND METHODS
Contrast-enhanced MR angiography exams of the calf were performed on ten normal volunteers using 3T clinical MRI scanners (Signa HDx and Discovery MR750, GE Healthcare, Waukesha, WI, USA) with two different commercially available coils (8-channel and 32-channel Torso Arrays, GE Healthcare, Waukesha, WI, USA) to validate the proposed methods. The protocol was IRB approved and informed consent was obtained from all volunteers. A range of parallel imaging and IVD undersampling factors were investigated to determine parameters that yielded maximum spatial and temporal resolution while maintaining clinically acceptable image quality. Following this parameter optimization, two volunteers (referred to below as volunteer 1 and volunteer 2) were imaged to demonstrate the technique optimized for each of the two coils. The first experiment, using an 8-channel torso coil, demonstrates the process of integrating parallel imaging into the multiplicative constrained reconstruction, as well as the feasibility of the proposed method. The second experiment, using a 32-channel torso coil, was used to demonstrate the capability of improving the spatial resolution and SNR using the proposed method and a coil array with higher channel count.
Volunteer 1 was imaged using an 8-channel torso coil. A 15× acceleration factor was achieved: 4× parallel imaging acceleration (R = 2 along L/R and A/P directions) and 3.8× IVD acceleration. Acquisition parameters included: matrix size of 512 (freq) × 298 (phase) × 70 (slice) over an FOV of 480 (S/I) × 278 (L/R) × 126 (A/P) mm3 for a spatial resolution of 0.94 × 0.94 × 1.8 mm3 with 75% fractional readout (zero-filled in the reconstruction), BW= ±62.5kHz, and TR/TE= 5.6/2.0 ms. Twenty-four time frames, resolved at 7.7 seconds, were acquired over 3 minutes. In the reconstruction, a kernel of size Wx × Wy × Wz = 3 × 7 × 7 was used to determine the calibration coefficients for the parallel imaging reconstruction. The data were reconstructed in two ways for comparison. First, the data were processed using the proposed method with sliding window length of 9 frames for Īt. Second, the data were also reconstructed using view-sharing (nearest temporal-neighbor interpolation within a window length of 9 frames). In both methods, this temporal window was cropped for the first and last four frames.
Volunteer 2 was imaged using a 32-channel torso coil. A 23× acceleration factor was achieved: 4× parallel imaging (R = 2 along L/R and A/P directions) and 5.8× IVD. Acquisition parameters included: matrix size of 480 (freq) × 290 (phase) × 120 (slice) over an FOV of 480 (S/I) × 290 (L/R) × 120 (A/P) mm3 for an isotropic resolution of 1.0 mm3, 75% fractional readout (zero-filled in the reconstruction), BW= ±62.5kHz, TR/TE= 5.2/1.9 ms. Twenty-four time frames were resolved at 8.0 seconds. A kernel size of Wx × Wy × Wz = 3 × 5 × 5 was used for estimating the calibration coefficients for the parallel imaging reconstruction. As above, the data were reconstructed with the proposed method (11-frame sliding window for Īt), and also with a view-sharing (11-frame window) reconstruction. An ROI was placed on the popliteal trifurcation and the signal was measured within the ROI as a function of time and compared for the proposed reconstruction method vs. the view sharing approach. In addition, a zero-filled parallel imaging reconstruction (zero-filling followed by parallel imaging synthesis) was performed on the same data set to generate images for comparison. The zero-filled parallel imaging reconstruction used the weighting coefficients that were determined by a calibration for a regularly-undersampled pattern, i.e. S(ky,kz).
RESULTS
In order to demonstrate that the proposed reconstruction effectively removes coherent aliasing in intermediate images and incoherent aliasing in the final image, data from the first normal volunteer exam was used. Fig. 3, first row, shows the data sets Kt, K̃t and K̅t reconstructed with zero-filling FFT. The images exhibit coherent aliasing in an axial slice and coronal MIP. The second row shows the same data reconstructed with parallel imaging. These images are unaliased, although the images reconstructed from the IVD-undersampled data (columns 1 and 2) exhibit the expected blurring and incoherent artifacts. It is worth pointing out here that the image It is temporally resolved, i.e. only arteries in the superior half of the FOV were enhanced; the image Īt is sharp and free of spatial artifacts, but temporally blurred, i.e. arteries over the entire FOV were enhanced; the image Ĩt is derived from Īt using the same sampling function st(ky,kz), yielding similar spatial blurring and incoherent artifacts as in It. Finally, the application of the constrained reconstruction (Eq. [2]) is able to impose the accurate temporal dynamics of It onto the constraining image Īt, generating a final image Ît of both high spatial and temporal resolution, shown in Fig. 4(a).
Fig. 4 shows example time-frame images from the time-resolved exam of volunteer 1 acquired with the 8-channel coil. Coronal MIP images of three consecutive time frames demonstrate improved temporal depiction of the contrast bolus (Fig. 6(a–c) arrows). Sagittal reformats of the data for a single leg show improved visualization of small vessels (Fig. 6(d)). No aliasing artifacts due to parallel imaging undersampling are observed in the images. Images reconstructed with view-sharing are also provided for comparison (Fig. 6(e–g)). As shown by the arrows, the movement of the bolus from superior (top) to inferior (bottom) is better depicted in the constrained reconstruction. The temporal blurring effect of the view-sharing technique is typically manifested as early enhancement of the vessels, which can be best observed by comparing Fig. 4(a) and (e).
Higher resolution and image quality were achieved using the proposed method with a 32-channel phased array coil. Images acquired with decreased voxel size (isotropic spatial resolution) are shown in Fig. 5 for volunteer 2. Fig. 5(a–c) shows three consecutive arterial time frames produced using the proposed constrained reconstruction method. The high spatial resolution can be appreciated by observing an enlarged image of the gray dashed box (Fig. 5(d)), where small arteries are well visualized. A sagittal reformat (Fig. 5(e)) demonstrates the improved resolution in the slice-encoding dimension compared to Fig. 4(d). The view-sharing method (Fig. 5(f–h)) results in significant temporal blurring, as indicated by the arrows showing the leading edge of the contrast bolus. The images produced by zero-filling (Fig. 5(i–k)) suffer from spatial artifacts and blurring; however, since the zero-filling method does not employ any temporal interpolation, these images can be considered as temporally closest to the truth. As shown by the arrows, the constrained reconstruction provides more accurate temporal depiction of the filling arteries than the view-sharing method. This effect can be best appreciated by comparing Fig. 5(a), (f) and (i). In both the constrained reconstruction image (Fig. 5(a)) and the zero-filling image (Fig. 5(i)), only a small portion of the popliteal artery are enhanced at t = 32s; however, in the view-sharing (Fig. 5(f)), the arteries are enhanced over the entire FOV. The improved temporal fidelity is also demonstrated by plotting the signal within the selected ROI (white box in Fig. 5(c) and (h)) as a function of time as shown in Fig. 5(l). Fig. 5(l) compares the waveforms obtained from both the constrained reconstruction and view-sharing methods. As indicated by the arrow in Fig. 5(l), view-sharing causes a temporal blurring at the sharp uptake. It is worth noting that no apparent coherent aliasing (parallel imaging) artifacts are seen in the zero-filled images (Fig. 5(i–k)). This supports the previous results in Fig. 3, suggesting that coherent aliasing artifacts can be effectively removed by parallel imaging when it is calibrated for a 4-fold regularly-undersampled pattern and applied to the 23-fold undersampled data set.
DISCUSSION
In this work, a method for high acceleration that utilizes IVD Cartesian sampling, parallel imaging, and a constrained reconstruction was described and feasibility demonstrated in-vivo. The increased acceleration provided using this method can be utilized to improve spatial resolution and/or temporal resolution, depending on the specific clinical need. While all the studies presented in this work used moderate IVD undersampling factors of 4 to 5 to maintain minimal level of incoherent undersampling artifacts, it can be expected that increased temporal resolution may be achievable in other applications where the clinical need for very high temporal resolution may drive a different trade-off with spatial resolution (30,31).
When performing parameter optimization, a range of parallel imaging and IVD undersampmling factors was studied to understand the trade-offs between these two forms of acceleration. It was found that if parallel imaging is not used (i.e. R = 1), requiring a higher IVD undersampling factor, then residual undersampling artifacts in the final image was observed, manifesting when visualized as a time-series movie as “flashing” effect. With the 8-channel torso coil, a parallel imaging factor of 4 (2 along each direction) can be achieved with no apparent coherent aliasing artifacts. While keeping the total undersampling factor the same, the integration of parallel imaging can decrease the burden for the IVD undersampling, which results in a significant reduction of the flashing effect. Using a 32-channel coil, 1.0 mm isotropic resolution can be achieved because of the increased SNR from the higher channel count. More recent work (32) has demonstrated the feasibility of using parallel imaging R = 6 with this coil to achieve 1.0 mm isotropic resolution and temporal resolution of about 5 sec/frame, with total acceleration factor of about 36. Generally speaking, the maximum acceleration factors (parallel imaging factor and IVD factor) depend on the spatial-temporal sparsity of the specific application, coil used, and particular clinical requirements.
As demonstrated experimentally, parallel imaging can reconstruct highly undersampled data with coherent aliasing artifacts in intermediate images It and Ĩt significantly reduced. If one were to attempt to produce the highest quality image reconstructed from only the highly undersampled time frame data Kt, one might consider using a much larger kernel size and calibrating for the actual sampling pattern st(ky,kz), rather than the pattern of the integrated data S(ky,kz). We found that this results in much (100×) longer reconstruction times, and may suffer from noise amplification. Such an observation does not imply that the method proposed here is the optimal reconstruction of the data, but it does appear to be a practical solution for rapid reconstruction with good image quality.
Conceivably, Eq. [2] could be applied either before or after coil-combing processing. In this work, for each time frame t, intermediate images (It, Īt and Ĩt) were generated for each coil, and Eq. [2] was applied on a coil-by-coil basis followed by a sum-of-squares reconstruction (33). The reason for applying Eq. [2] prior to coil combination is that each coil image is sparser than the coil-combined image, and thus is better conditioned for the multiplicative constrained reconstruction.
It is also worth noting that the coil images It, Īt and Ĩt, are complex-valued. In applications such as CE-MRA, phase information is typically not needed. Therefore, one way to enforce real-positive constraint is to use the magnitude of these complex coil images. Alternatively, a low-resolution phase map can be obtained from Īt, and subsequently used to rephase all the images and enforce the real-positive constraint. The SNR of the final images reconstructed using the rephasing method was found to be better than the SNR of the final images reconstructed using the magnitude method, because in the former method, incoherent background noise from the imaginary channel is excluded. This rephasing method is essentially the noise reduction application of homodyne detection (34).
Due to the division operation, two types of regularization methods were included in the proposed method. First, in Eq. [2], a small constant c (usually chosen as 2% of the maximum value of Īt) may be added as a base-line shift in the image intensity to avoid zero division. Second, an empirically-chosen maximum value may be used to threshold the ratio image to further avoid noise amplification, prior to the multiplication with Īt. For all the calf studies in this work, the threshold value was chosen between 1.5 and 2. Other regularization methods to further reduce noise amplification will be investigated in future work.
The proposed constrained reconstruction (HYCR) can be considered as a Cartesian implementation of the HYPR technique (21–23), in which a time-averaged, high spatial resolution, high SNR image is multiplied by a lower spatial resolution, high SNR, temporally-resolved weighting image. The application of the HYPR principle to Cartesian-sampling was first proposed in Ref. (35); however that earlier work did not describe interleaved sampling or parallel imaging (it was targeted at combining a number of fully sampled data sets to improve SNR) and it applied the multiplicative constraint in hybrid space, not image space.
Besides Cartesian approaches, several non-Cartesian techniques have also been developed for time-resolved CE-MRA that aim to improve image quality by reducing coherent artifacts compared to undersampled Cartesian methods. These techniques include radial acquisition approaches, such as Stack-of-Stars (SOS) (36) and VIPR (7,37), and spiral acquisitions (38,39). However, for some applications such as peripheral run-off CE-MRA, Cartesian sampling trajectories offer some important advantages over non-Cartesian approaches. First, spin-warp phase-encoding (acquiring parallel lines of k-space along the readout direction) is very robust to k-space trajectory errors caused by gradient timing errors and eddy-currents. Second, parallel readout acquisitions permit establishment of a single frequency-encoding direction, which enables application of filters to limit aliasing of signal from outside the field of view (FOV), typically along the longest axis of the acquisition volume. Third, separable view spacing in the two phase-encoding directions enables scan times to be reduced by tailoring the FOV to the anatomy. Fourth, parallel imaging techniques for Cartesian acquisitions are more computationally efficient than similar parallel imaging reconstruction techniques for non-Cartesian acquisitions.
A number of previous studies that compared a constrained reconstruction to standard view-sharing techniques have been performed and reported (8,9,40). Keith et al. (8) and Busse et al. (9) used the concept of temporal point-spread-function (PSF) in numerical simulations, where the temporal input function was a step function. They both demonstrated that the constrained reconstructions provided superior temporal fidelity by better resolving rapidly occurring dynamic processes. These results are consistent with the in-vivo results reported here, which suggest a reduction in temporal blurring compared to the view-sharing reconstruction. It should be recognized that the relative performance of the proposed method and the view-sharing approach may also depend on spatial and temporal characteristics including vessel structures and contrast kinetics, which are application-dependent. Generally, the depiction of smaller vessels requires more high frequency data for accurate depiction and view-sharing methods are expected to produce more temporal blurring than the constrained method when applied to smaller vessels. It should also be noted that the optimal sampling pattern for constrained reconstruction and view-sharing reconstruction may be different. For example, we found that the 1/kr density is optimal for the HYCR algorithm, while other sampling density functions may be optimal for view-sharing reconstruction in terms of minimizing the temporal footprint.
There are other reconstruction methods that can be used for dynamic MR imaging, such as k-t BLAST/SENSE (17) and compressed sensing (5). k-t BLAST/SENSE are model-based techniques that utilize a training data set to guide the reconstruction of the high resolution time frames, and have shown improved spatial and temporal resolution primarily in cardiac cine MRI, because the cardiac motion is approximately periodic and can be approximated with a few Fourier coefficients. However, such feature does not exist for CE-MRA. In contrast, the proposed method in this work generates a time-averaged composite image from the undersampled time frame, and thus does not require the acquisition of the training data set. Conventional compressed sensing methods explore the sparsity in the spatial domain, or transformed-spatial domain, as demonstrated in Fig. 9 in Ref. (5), but it may be time-consuming for time-resolved MRI with large matrix sizes. Differently, the proposed method in this work aims to explore the sparsity in the spatial-temporal domain, and preliminary results from other applications which involve spatially non-sparse images have been reported (25,41). In the mean time, there are some recent works on using compressed sensing framework to explore the sparsity in the spatial-temporal domain, such as k-t SPARSE (42) and k-t FOCUSS (43).
Iterative HYPR-like approaches can be employed to enforce data consistency, further improve temporal waveform fidelity and increase robustness to motion (24,44). The iterative formulation of Eq. [2] can be written as:
[3] |
with the first image estimation being the composite image, i.e. Ît(0) = Īt, and Ĩt(n) being the re-sampled Ît(n) at location st(ky,kz). This iterative formulation has similarities to some expectation maximization algorithms, such as the Multiplicative Algebraic Reconstruction Technique (MART) (45) that was first developed for x-ray imaging.
CONCLUSION
In this work, a novel interleaved variable density Cartesian sampling method and an image reconstruction technique that combines a multiplicative constrained reconstruction with parallel imaging (IVD HYCR), was described. It was demonstrated in volunteers that the technique is capable of producing high spatial resolution and high SNR images while maintaining high temporal resolution in small, enhancing vessels. The proposed data acquisition and image reconstruction methods can also be applied to other areas where higher temporal resolution is needed.
ACKNOWLEDGEMENT
The authors would like to thank Lauren Keith, Yijing Wu and Patrick Turski for their help in this project.
Grant sponsors: NIH; American Heart Association (AHA);
Grant numbers: NIH-R01 EB006882; AHA-0865146F (to Du J.)
REFERENCES
- 1.Zhang H, Maki JH, Prince MR. 3D contrast-enhanced MR angiography. J Magn Reson Imaging. 2007;25(1):13–25. doi: 10.1002/jmri.20767. [DOI] [PubMed] [Google Scholar]
- 2.Haider CR, Hu HH, Campeau NG, Huston J, 3rd, Riederer SJ. 3D high temporal and spatial resolution contrast-enhanced MR angiography of the whole brain. Magn Reson Med. 2008;60(3):749–760. doi: 10.1002/mrm.21675. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Haider CR, Glockner JF, Vrtiska TJ, Macedo TA, Borisch EA, Riederer SJ. A Comparison of Time-Resolved 3D CE-MRA with Peripheral Run-off CTA in the Calves. East Lansing, MI: MR Angio Club; 2009. [Google Scholar]
- 4.Du J. Contrast-enhanced MR angiography using time resolved interleaved projection sampling with three-dimensional cartesian phase and slice encoding (TRIPPS) Magn Reson Med. 2009;61(4):918–924. doi: 10.1002/mrm.21805. [DOI] [PubMed] [Google Scholar]
- 5.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58(6):1182–1195. doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
- 6.Busse RF, Wang K, Holmes JH, Brittain JH, Korosec FR. Optimization of Variable-Density Cartesian Sampling for Time-Resolved Imaging. Proc 17th Annual Meeting ISMRM; Hawai'i, USA. 2009. p. 4534. [Google Scholar]
- 7.Du J, Carroll TJ, Brodsky E, Lu A, Grist TM, Mistretta CA, Block WF. Contrast-enhanced peripheral magnetic resonance angiography using time-resolved vastly undersampled isotropic projection reconstruction. J Magn Reson Imaging. 2004;20(5):894–900. doi: 10.1002/jmri.20189. [DOI] [PubMed] [Google Scholar]
- 8.Keith L, Kecskemeti S, Velikina J, Mistretta C. Simulation of relative temporal resolution of time-resolved MRA sequences. Magn Reson Med. 2008;60(2):398–404. doi: 10.1002/mrm.21658. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Busse RF, Pineda AR, Wang K, Holmes JH, Brittain JH, Korosec FR. Time-Resolved Imaging with Multiplicative Algebraic Reconstruction Technique (MART): An Application of HYPR Principles for Variable Density Cartesian Acquisitions. Proc 17th Annual Meeting ISMRM; Hawai'i, USA. 2009. p. 2834. [Google Scholar]
- 10.Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38(4):591–603. doi: 10.1002/mrm.1910380414. [DOI] [PubMed] [Google Scholar]
- 11.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42(5):952–962. [PubMed] [Google Scholar]
- 12.Sodickson DK, McKenzie CA. A generalized approach to parallel magnetic resonance imaging. Med Phys. 2001;28(8):1629–1643. doi: 10.1118/1.1386778. [DOI] [PubMed] [Google Scholar]
- 13.Jakob PM, Griswold MA, Edelman RR, Sodickson DK. AUTO-SMASH: a self-calibrating technique for SMASH imaging. SiMultaneous Acquisition of Spatial Harmonics. Magma. 1998;7(1):42–54. doi: 10.1007/BF02592256. [DOI] [PubMed] [Google Scholar]
- 14.Heidemann RM, Griswold MA, Haase A, Jakob PM. VD-AUTO-SMASH imaging. Magn Reson Med. 2001;45(6):1066–1074. doi: 10.1002/mrm.1141. [DOI] [PubMed] [Google Scholar]
- 15.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA) Magn Reson Med. 2002;47(6):1202–1210. doi: 10.1002/mrm.10171. [DOI] [PubMed] [Google Scholar]
- 16.Chandra S, Liang ZP, Webb A, Lee H, Morris HD, Lauterbur PC. Application of reduced-encoding imaging with generalized-series reconstruction (RIGR) in dynamic MR imaging. J Magn Reson Imaging. 1996;6(5):783–797. doi: 10.1002/jmri.1880060512. [DOI] [PubMed] [Google Scholar]
- 17.Tsao J, Boesiger P, Pruessmann KP. k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003;50(5):1031–1042. doi: 10.1002/mrm.10611. [DOI] [PubMed] [Google Scholar]
- 18.Candes EJ, Romberg J, Tao T. Robust uncertainty principles: Exact signal reconstruction from highly incomplete frequency information. IEEE Transactions on Information Theory. 2006;52(2):489–509. [Google Scholar]
- 19.Donoho DL. Compressed sensing. IEEE Transactions on Information Theory. 2006;52(4):1289–1306. [Google Scholar]
- 20.Lustig M, Alley M, Vasanawala S, Donoho DL, Pauly JM. L1 SPIR-iT: Autocalibrating Parallel Imaging Compressed Sensing. Proc 17th Annual Meeting ISMRM; Hawai'i, USA. 2009. p. 379. [Google Scholar]
- 21.Mistretta CA, Wieben O, Velikina J, Block W, Perry J, Wu Y, Johnson K, Wu Y. Highly constrained backprojection for time-resolved MRI. Magn Reson Med. 2006;55(1):30–40. doi: 10.1002/mrm.20772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Johnson KM, Velikina J, Wu Y, Kecskemeti S, Wieben O, Mistretta CA. Improved waveform fidelity using local HYPR reconstruction (HYPR LR) Magn Reson Med. 2008;59(3):456–462. doi: 10.1002/mrm.21505. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Huang Y, Wright GA. Time-resolved MR angiography with limited projections. Magn Reson Med. 2007;58(2):316–325. doi: 10.1002/mrm.21312. [DOI] [PubMed] [Google Scholar]
- 24.O'Halloran RL, Wen Z, Holmes JH, Fain SB. Iterative projection reconstruction of time-resolved images using highly-constrained back-projection (HYPR) Magn Reson Med. 2008;59(1):132–139. doi: 10.1002/mrm.21439. [DOI] [PubMed] [Google Scholar]
- 25.Wang K, Du J, O'Halloran R, Fain S, Kecskemeti S, Wieben O, Johnson KM, Mistretta C. Ultrashort TE spectroscopic imaging (UTESI) using complex highly-constrained backprojection with local reconstruction (HYPR LR) Magn Reson Med. 2009;62(1):127–134. doi: 10.1002/mrm.21986. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Wu Y, Johnson K, Wieben O, Velikina J, Carrillo A, Kecskemeti S, Mistretta C, Korosec F. Comparison of HYPR Stack-Of-Stars and HYPR VIPR to TRICKS in Peripheral CE MRA. Proc 16th Annual Meeting ISMRM; Toronto, Canada. 2008. p. 108. [Google Scholar]
- 27.Wang K, Du J, Wu Y, Busse RF, Johnson K, Korosec FR. Accelerated 3D Time-Resolved MR Angiography using Cartesian HYPR LR Reconstruction. Proc 17th Annual Meeting ISMRM; Hawai'i, USA. 2009. p. 3884. [Google Scholar]
- 28.Beatty PJ, Brau AC, Chang S, Joshi SM, Michelich CR, Bayram E, Nelson TE, Herfkens RJ, Brittain JH. A Method for Autocalibrating 2-D Accelerated Volumetric Parallel Imaging with Clinically Practical Reconstruction Times. Berlin, Germany: 2007. p. 1749. [Google Scholar]
- 29.Brau AC, Beatty PJ, Skare S, Bammer R. Comparison of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods. Magn Reson Med. 2008;59(2):382–395. doi: 10.1002/mrm.21481. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Velikina JV, Johnson KM, Wu Y, Samsonov AA, Turski P, Mistretta CA. PC HYPR flow: a technique for rapid imaging of contrast dynamics. J Magn Reson Imaging. 2010;31(2):447–456. doi: 10.1002/jmri.22035. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Holmes JH, O'Halloran RL, Brodsky EK, Bley TA, Francois CJ, Velikina JV, Sorkness RL, Busse WW, Fain SB. Three-dimensional imaging of ventilation dynamics in asthmatics using multiecho projection acquisition with constrained reconstruction. Magn Reson Med. 2009;62(6):1543–1556. doi: 10.1002/mrm.22150. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Holmes JH, Wang K, Busse RF, Beatty PJ, Brittain JH, Francois CJ, Reeder SB, Keith L, Wu Y, Korosec FR. 3D Time-Resolved MRA of Lower Extremities using Interleaved Variable Density Sampling, Parallel Imaging and Cartesian HYPR Reconstruction. Seoul, Korea: MR Angio Club; 2010. Abstract 8.5. [Google Scholar]
- 33.Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990;16(2):192–225. doi: 10.1002/mrm.1910160203. [DOI] [PubMed] [Google Scholar]
- 34.Noll DC, Nishimura DG, Macovski A. Homodyne detection in magnetic resonance imaging. IEEE Trans Med Imaging. 1991;10(2):154–163. doi: 10.1109/42.79473. [DOI] [PubMed] [Google Scholar]
- 35.Yutzy S, Griswold MA, Duerk JL. Is HYPR Compatible with a Cartesian Acquisition?; Proc 15th Annual Meeting ISMRM; Berlin, Germany. 2007. p. 832. [Google Scholar]
- 36.Peters DC, Korosec FR, Grist TM, Block WF, Holden JE, Vigen KK, Mistretta CA. Undersampled projection reconstruction applied to MR angiography. Magn Reson Med. 2000;43(1):91–101. doi: 10.1002/(sici)1522-2594(200001)43:1<91::aid-mrm11>3.0.co;2-4. [DOI] [PubMed] [Google Scholar]
- 37.Barger AV, Block WF, Toropov Y, Grist TM, Mistretta CA. Time-resolved contrast-enhanced imaging with isotropic resolution and broad coverage using an undersampled 3D projection trajectory. Magn Reson Med. 2002;48(2):297–305. doi: 10.1002/mrm.10212. [DOI] [PubMed] [Google Scholar]
- 38.Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn Reson Med. 1992;28(2):202–213. doi: 10.1002/mrm.1910280204. [DOI] [PubMed] [Google Scholar]
- 39.Du J, Bydder M. High-resolution time-resolved contrast-enhanced MR abdominal and pulmonary angiography using a spiral-TRICKS sequence. Magn Reson Med. 2007;58(3):631–635. doi: 10.1002/mrm.21298. [DOI] [PubMed] [Google Scholar]
- 40.Jeong HJ, Cashen TA, Hurley MC, Eddleman C, Getch C, Batjer HH, Carroll TJ. Radial sliding-window magnetic resonance angiography (MRA) with highly-constrained projection reconstruction (HYPR) Magn Reson Med. 2009;61(5):1103–1113. doi: 10.1002/mrm.21888. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Wang K, Korosec FR, Artz NS, Schiebler ML, Francois CJ, Reeder SB, Grist TM, Busse RF, Holmes JH, Fain S, Nagle SK. Dynamic Pulmonary Perfusion Imaging using Interleaved Variable Density Sampling, Parallel Imaging and Cartesian HYPR Reconstruction. Seoul, Korea: MR Angio Club; 2010. Abstract 2.3. [Google Scholar]
- 42.Lustig M, Santos J, Donoho D, Pauly J. k-t SPARSE: High frame rate dynamic MRI exploiting spatio-temporal sparsity. Proc 14th Annual Meeting ISMRM; Seattle, USA. 2006. p. 2420. [Google Scholar]
- 43.Jung H, Sung K, Nayak KS, Kim EY, Ye JC. k-t FOCUSS: a general compressed sensing framework for high resolution dynamic MRI. Magn Reson Med. 2009;61(1):103–116. doi: 10.1002/mrm.21757. [DOI] [PubMed] [Google Scholar]
- 44.Griswold M, Barkauskas K, Blaimer M, Moriguchi H, Sunshine J, Duerk J. More optimal HYPR reconstructions using a combination of HYPR and conjugate-gradient minimization. Basel, Switzerland: Proceedings of MRA Club; 2006. p. 29. [Google Scholar]
- 45.Gordon R, Bender R, Herman GT. Algebraic reconstruction techniques (ART) for three-dimensional electron microscopy and x-ray photography. J Theor Biol. 1970;29(3):471–481. doi: 10.1016/0022-5193(70)90109-8. [DOI] [PubMed] [Google Scholar]