A Standard Approach to Continuous Glucose Monitor Data in Pregnancy for the Study of Fetal Growth and Infant Outcomes

Teri L Hernandez 1,2,3,, Linda A Barbour 1,4
PMCID: PMC3558676  PMID: 23268584

Abstract

The power of continuous glucose monitoring system (CGMS) technology to profile glycemic patterns throughout a 24-h period has benefited the care of individuals with diabetes mellitus for over 10 years. Recently, this technology has been utilized to better understand glucose patterns in pregnancy, especially as they relate to abnormal fetal growth given that adiposity at birth is associated with increased risks for childhood obesity and metabolic syndrome. However, the lack of a standardized approach to defining glucose measures associated with maternal outcomes and fetal growth has greatly limited comparison and pooling of CGMS data among pregnancy trials, hindering our ability to take advantage of the enormous amount of data available to explore these relationships. The purpose of this article is to offer a methodical approach to the identification and extraction of CGMS-derived glucose variables for the characterization of glycemic profiles in pregnant women, particularly focusing on women with gestational diabetes or obesity who are at risk for abnormal fetal growth. A review of the properties of CGMS data and examples of how CGMS data in pregnancy have been reported to date are included. We further define several pregnancy-relevant, CGMS-derived glucose variables and directly apply them to unpublished data to illustrate how these measures might be utilized. This approach offers one possible standardized method to define and analyze these time-sensitive glucose measures to facilitate comparisons among studies and to increase our understanding of how glycemic profiles contribute to excess infant adiposity in pregnant women with and without diabetes.

Introduction

The treatment goal for glycemia in diabetes is to mirror normoglycemia, and this is particularly important and challenging in pregnancy.1 In addition to fasting glucose, higher postprandial excursions as well as nocturnal glycemia clearly contribute to excess fetal growth.2 Yet, historically our understanding of maternal patterns of glycemia in relation to fetal growth, as characterized by glucometers, has been incomplete.3 The Hyperglycemia and Adverse Pregnancy Outcomes trial recently challenged our understanding of normoglycemia in pregnancy: women with much lower glucose levels than previously recognized were at risk of delivering an infant with excess adiposity4 and fetal hyperinsulinemia (measured by cord blood C-peptide).5 Furthermore, there is an increasing body of literature to support that infants who are born large for gestational age (LGA), especially those with increased adiposity, have a higher risk of developing obesity and metabolic syndrome as children.6 A better appreciation of these two extremes, hyper- and hypoglycemia in pregnancy, has been made possible by the advent of continuous glucose monitoring system (CGMS) technology. We and others have previously reported the application of this technology in pregnancy and have demonstrated that other high-risk populations, specifically obese pregnant women who do not meet criteria for gestational diabetes mellitus, demonstrate occult or unrecognized hyperglycemia.710 This finding might partially provide an explanation for the high risk of macrosomia (birth weight >9 pounds or >4,000 g) or LGA infants in the obese pregnant population despite what has been considered normal glucose tolerance. The power of CGMS technology to profile glycemic patterns throughout a 24-h period has revolutionized the care of individuals with diabetes mellitus over the last 10 years. However, the lack of a standardized approach in defining glucose measures associated with maternal outcomes and fetal growth has greatly limited our ability to compare and pool CGMS data in pregnancy trials and take advantage of the enormous amount of data available.

Research investigators have adopted CGMS as a method to characterize populations without type 1 or type 2 diabetes but who are at high risk for metabolic diseases. If the CGMS device is worn for 72 h, over 800 glucose measurements are possible through sampling of interstitial glucose every 5 min. Although increasing numbers of investigators are using this technology in pregnancy, the literature lacks a uniform approach to the use of CGMS in pregnant women who are at risk for delivering offspring with abnormal fetal growth. Preliminary data in our studies of pregnant women demonstrate strong correlations between CGMS-derived glucose variables and infant outcomes, particularly neonatal adiposity,79 indicating that CGMS may be a promising methodology for the study of glucose-driven neonatal outcomes. However, there is a lack of any standardization in defining specific CGMS variables (i.e., fasting, nocturnal, postprandial glucose) clinically important for pregnant women. In addition, data management procedures are poorly described at best in most published reports, which severely limits the ability to compare or pool data between studies or directly apply the findings to clinical practice.

Recommendations have been made to suggest approaches to analysis and interpretation of the enormous data output generated by CGMS,1113 especially those with type 1 diabetes.14,15 However, these approaches and the glycemic measures utilized are not specifically tailored to exploring variables of high clinical relevance in pregnancy associated with abnormal fetal growth (e.g. fasting, 1- and 2-h postprandial glucose). Space constraints within published articles make it impossible for authors to fully describe how CGMS variables were defined and handled, making comparison, replication, and interpretation of data among studies in pregnant women extremely difficult. For example, the seemingly simple but extremely clinically important question of what defines a fasting glucose in pregnant women (utilizing CGMS) in a free-living environment has become subject to interpretation. Should it be defined by the number of hours after a bedtime snack, at a certain time period during the morning, or the value immediately before breakfast independent of the actual time spent fasting? Should a single value be used, or should several values be averaged over a period of time since measures every 5 min are available? Within hundreds of glucose concentrations per monitoring period, per day, and per research participant, CGMS can provide an unwieldy volume of data. From these data, important variables specific to the population of interest must be extracted, while incongruous/missing data are dealt with in a consistent and methodical manner.

The purpose of this article is to offer a methodical approach to the identification and extraction of CGMS-derived glucose variables for the characterization of glycemic profiles in pregnant women, particularly focusing on women at risk for abnormal fetal growth. After review of the properties of CGMS data and examples of how data in pregnancy have been reported to date, we propose and define several pregnancy-relevant, CGMS-derived glucose variables that may be be considered with a systematic approach to their identification. These CGMS-derived glucose variables were used to examine the correlation of fetal growth patterns in our published study.7 For this article, we include some of our unpublished data to specifically exemplify how these CGMS measures might be adapted to illustrate their potential application. This approach offers one possible standard methodology to define time-sensitive glucose measures and analyze variables of interest that other investigators may find clinically useful and may facilitate comparisons among studies.

Properties of Interstitial Glucose Measured by CGMS

Properties of CGMS glucose concentrations are unlike other measures of blood glucose (BG). The gold standard for the measurement of glucose is within plasma using a high-precision enzymatic laboratory method (glucose oxidase, glucose dehydrogenase, or hexokinase).16 Since 1987, however, glucometers have been standardized to report plasma-adjusted values within ±15%17,18 and are recognized as the standard for adjustment of insulin therapy and monitoring treatment adherence in diabetes.19 Because it is calibrated to capillary glucose by glucometer (meter glucose),20 interstitial glucose as measured by CGMS (CGMS glucose) is highly correlated with meter glucose (r=0.91–0.92).21,22 However, an important difference between the two glucose measures is that CGMS glucose is an in vivo, indirect measure of glucose. Meter glucose is an in vitro, direct test of plasma glucose. Thus, in vivo CGMS glucose is calibrated to an in vitro meter glucose measure. Clarke and Kovatchev13 described that CGMS glucose is further different from meter glucose because of the inherent differences in their properties. Fluctuating in vivo CGMS glucose concentrations reflect a continuous process in time. Because the process is continuous, consecutive glucose measures are highly associated. Each CGMS glucose concentration is determined by the one before it, and the chronological time series represents the rate and direction of the change in glucose. The subcutaneous sensor samples interstitial fluid, collects information continuously, and then derives an average glucose value every 5 min. Thus, each CGMS glucose value represents an average of measures and information during the preceding 5 min. Individual CGMS glucose measures, therefore, should not be considered alone. In vitro meter glucose, on the other hand, is a purely isolated measure of an independent concentration of capillary BG, unaffected by previous measures. CGMS glucose measures are further dependent on the physiological diffusion of blood into capillaries and separation to interstitial fluid, which creates a measurement time delay.20,23 The time delay, in combination with the time-dependent properties of CGMS glucose, makes accuracy evaluation challenging. Although software specific to CGMS attempts to correct for the time delay, in vitro meter glucose remains the standard for evaluation of CGMS precision clinically.20 Therefore, the continuous and interdependent structure of CGMS patterns, along with the in vivo nature of the measures, are important considerations when working with these data.

Lack of Uniformity in Reporting of CGMS Data in Pregnancy

The pregnancy literature suffers from the absence of a uniform methodology for an approach to using and reporting CGMS data in pregnancy, even within similar study populations. Although some researchers have attempted to identify their variable selection criteria in limited manuscript space, either the outlined procedures are briefly defined such that replication is impossible, or the chronological properties of CGMS measures are not clearly distinguished. For example, some studies report the “mean glucose” via CGMS but do not identify the time frame from which the mean was derived.10,2428 Thus, it is unclear if the “mean glucose” comprises one 24-h period (approximately 288 CGMS glucose measures) or is a 24-h average across several days. It is further unclear what time frame comprises the “mean glucose” (i.e., Does it begin at midnight? Does it begin when the patient goes to sleep? Is it simply an average of all sensor data across the monitoring period?). Some studies have defined the mean nocturnal glucose as those values between 2300 and 0600 h10 or 2330 and 0630 h,7 in contrast with others27 who have reported the same variable in name but without definition of the nocturnal period. However, if a pregnant woman has a midnight snack (which is common), the values surrounding the snack might be included in the “nocturnal” period, when the woman is neither fasting nor asleep.

Other CGMS variables are often not clearly described in the literature that may have significant clinical relevance. Studies have reported mean pre- and postprandial CGMS glucose variables29 and glucose area under the curve (AUC)27 without explicit explanation as to how the variables were defined or calculated. Another frequently reported CGMS glucose variable with important clinical relevance to pregnancy and fetal growth is the fasting glucose. With the exception of our publication,7 other studies have not clearly defined the fasting glucose in terms of timing, or if it was one isolated value versus the mean of more than one CGMS glucose value.10,25 Ben Haroush et al.26 provided a helpful graphic representation of how they determined the pre- and postprandial peak and nadir glucose measures from CGMS, but it appears that in this study and in one other10 one isolated CGMS glucose value was used for each variable. This practice, as discussed, overlooks the dependence on the rate and direction of change in measures, which is an intrinsic property of CGMS data. Duration of glycemia above and below specific thresholds has further been reported.25,27,30 Space constraints may prevent investigators from being able to report their definition of glycemic measures in sufficient detail. We attempted to identify our data procedures in a recent publication,7 but space constraints precluded the type of detail that would allow for replication, and definitions of CGMS variables were relegated to a figure legend. The lack of a uniform definition and approach to CGMS data in pregnancy significantly compromises our ability to compare data among studies or gain a clearer understanding of glycemic patterns in normal and metabolically high-risk pregnant populations.

Selection of CGMS Glucose Variables in Pregnancy

Glucose variables derived from CGMS data should optimally be specific to the population of interest. For example, in adolescents with type 1 diabetes, a frequent variable of interest is to report the duration of hypo- or hyperglycemia above defined thresholds within a 24-h period.31 In this population, glucose at specific time periods might be less informative because of wide variation in sleep–wake patterns, meal intake patterns, and/or physical activity. In nonpregnant individuals with type 1 diabetes, especially young children, a threshold for hyperglycemia might be >200 mg/dL.13 However, in pregnant women a mean glucose level of >130 mg/dL is highly predictive of fetal macrosomia.32 Thus the threshold for “hyperglycemia” in a pregnant mother is actually lower than outside of pregnancy. Even within pregnant women, glycemic variability is higher in those who are obese versus normal weight7 and in those with preexisting versus gestational diabetes.26

The goal for the management of diabetes in pregnancy is to achieve tight glycemic control in the first trimester in order to prevent major malformations and pregnancy loss; in the second and third trimester tight glycemic control attempts to avoid excess fetal growth,1 prevent neonatal respiratory distress, and minimize metabolic abnormalities at birth. Optimal ranges of glycemia in pregnancy have been identified as premeal, bedtime, or nocturnal glucose 60–99 mg/dL,1 1- and 2-h postprandial glucose of <140 mg/dL and <120 mg/dL, respectively, fasting BG ≤95 mg/dL,33 and 24-h mean BG of 87–104 mg/dL.34 Postprandial as opposed to preprandial glucose is closely monitored and targeted because of its association with excess fetal growth patterns.1,2

A list of suggested CGMS-derived glucose variables that are of particular use for the clinical management of pregnant women is presented in Table 1, and we recognize that others may be highly relevant for specific studies. The properties of CGMS glucose concentrations, dependence on the previous glucose, the lag time inherent in the measures, and the clinically relevant timing of glucose measures were considered in determining these definitions. None of the variables is defined as an isolated CGMS glucose measure, given the properties of CGMS. Time frames were determined based on typical patient life-styles to define daytime (0630–2330 h) and nocturnal (2330–0630 h) periods. However, if a patient or subject does not assume these typical life-style patterns (e.g., eating past 2300 h or staying up through the night), it is critical for the investigator to identify and account for this deviation. To determine pre- and 1-h and 2-h postmeal glucose variables, the use of three consecutive values within a single day was chosen. Three values, 5 min apart, is likely to capture glucose variability or lack thereof during that vicinity of time.13 The use of six consecutive values within a single day for determination of fasting glucose was chosen in an effort to understand fasting conditions for a period of time when glucose is minimally fluctuating.35 The mean values of daytime glucose, which reflect primarily the fed state in pregnant women, nocturnal glucose, which reflects maternal hepatic gluconeogenesis and fetoplacental demands, and mean 24-h glucose may also be highly clinically relevant to pregnancy and fetal growth.

Table 1.

Continuous Glucose Monitoring System Variables and Their Definitions

CGMS variable Definition
FBG Mean of six consecutive values starting at 0600 h and/or after at least 7 h fastinga
Preprandial BG Mean of three consecutive values directly before breakfast, lunch, and dinner meal start time during BG stability
1-h PP BG Mean of three consecutive measures 1 h after the meal start time
2-h PP BG Mean of three consecutive measures 2 h after the meal start time
Mean daytime BG Mean of all measures between 0630 and 2330 h
Mean nocturnal BG Mean of all measures between 2330 and 0630 h
Lowest nocturnal BG Mean of six lowest consecutive measures between 2330 and 0630 h
1-h PP excursion (1-h PP BG) – (preprandial BG) (calculated)
Mean 24-h BG Mean of all measures in 24 h: 2330–2330 h
Peak PP BG Highest PP glucose within 2 h of meal start time
Time to PP peak Time from meal start to peak PP BG
Percentage >120 mg/dL Percentage of time glucose was >120 mg/dL
24-h AUC AUC 2330–2330 h
Daytime AUC AUC 0630–2330 h
Nocturnal AUC AUC 2330–0630 h
2-hr PP AUC AUC 2 h after meal start time for breakfast, lunch, or dinner

Each variable is determined from a single 24-h period of continuous glucose monitoring system (CGMS) data. Identical variables can then be averaged across several days according to the research study design.

a

If the timing of CGMS glucose measures does not correspond to exact 5-min clock times (i.e., 0600, 0605, 0610 h), then the closest value to 0600 h can be used for the first consecutive measure. Moreover, if the 7-h fasting period extends past 0600 h, the first consecutive measure should be 7 hours after the patient reports eating. Times are 24-h clock time. Meal start time was defined as the start of meal consumption as recorded by the research participant.

AUC, glucose area under the curve; BG, blood glucose; FBG, fasting blood glucose; PP, postprandial.

Use of Glucose AUC

The power of CGMS technology data is in part due to its ability to define specific glycemic patterns of variable duration that would otherwise be impossible to discern using self-monitoring of BG. In pregnancy, the profound effect of maternal glucose on the fetus has long been appreciated.36 The placenta allows for facilitated transfer of maternal glucose in a gradient fashion: higher maternal glucose causes more glucose transport to the placenta, and subsequently the fetus secretes insulin to utilize the glucose load, which is a potent growth factor.37 Thus, in pregnant women, the pattern of glycemia visible by CGMS represents total potential fetal glucose exposure. For this reason, we use the calculation of total glucose AUC using the trapezoid method, as opposed to the incremental AUC, for descriptions of maternal glycemia (Table 1). To describe glycemia in pregnancy, use of the incremental AUC (requiring adjustment for the baseline glucose) becomes problematic because it is possible to calculate a negative AUC if post-baseline values are lower than baseline. Thus, if only areas above the baseline value are included in the calculation, important information about glucose availability to the fetus could be lost.38 Potteiger et al.38 described that the method used to calculate AUC can affect data interpretation within intervention trials. Therefore, careful consideration of differences in AUC calculations is advised outside of descriptive study designs. The glucose AUC can be further used to characterize the 2-h postprandial AUC for each meal, which incorporates all of the postprandial values in the first 2 h after ingestion. If a higher fat diet is consumed, it may be more relevant to include a 3- or 4-h postprandial AUC to account for the delayed glucose absorption. The AUC is also useful to characterize daytime and nocturnal AUC. Postprandial and diurnal glucose metabolism can be examined independently for its impact on fetal growth.

Analytic Approach to CGMS Data

In an effort to improve consistency in the interpretation and extraction of the glucose variables seen in Table 1, we have adopted rigorous procedures for handling our CGMS data.

Interpretation and identification of variables

First, data from each 72-h CGMS monitoring period are exported from product-specific software to Microsoft (Redmond, WA) Excel®. Although the device-specific software summarizes glucose variables for clinical use, these variables do not necessarily represent those specific to a research context. Exporting the data gives the option for use of a printed record from the entire monitoring period. The data are manually inspected for (1) the expected rise/fall after a meal (appropriate directionality), (2) values that are ≥2 SD from the immediate previous value, (3) relative correspondence with the meter glucose value used for calibration, and (4) the relationship between the glucose value and the signal (ISig) from the CGMS sensor (Medtronic MiniMed, Northridge, CA).39 Inconsistent data that do not meet these criteria are not used in the calculations. We exclude data that do not correspond to the meter preprandial or steady-state glucose value within 20 mg/dL within that timing vicinity or if they are outliers (defined as ≥2 SD from the immediate surrounding values). The variables outlined in Table 1 are then identified and marked using the patient's log, upon which meal start times, meter glucose values, and all events (including changes in physical activity) are noted. For quality control, we utilize a second investigator who subsequently performs the same procedure so that discrepancies are discussed and resolved.

Missing data

If data are missing because of sensor failure, calibration error, or suspected inaccuracy, this time period is not available for analysis. Only if the subject is receiving an identical diet in both calories and macronutrient composition and physical activity is highly controlled on a previous or subsequent day could the missing data be potentially replaced (with time-corresponding data). If small numbers of values are missing during a time of glycemic stability devoid of changes in caloric consumption or activity, the values immediately surrounding the missing period could be used as has been customary in seminal highly controlled studies.40,41 It is highly recommended that an a priori approach to missing or incongruous data be outlined in detail before any data extraction begins. The individual accuracy of sensors may also be assessed using the mean absolute difference (MAD) (calculated as MAD [mg/dL]=CGMS glucose÷meter glucose) or the mean absolute relative difference (MARD) (calculated as MARD [%]=[MAD÷meter BG]×100).42 However, use of MAD and MARD are only as accurate as the in vitro meter BG.

Extraction and analysis

Once the data have been interpreted and the variables have been identified, they are extracted from the exported format into a separate worksheet, such as in Microsoft Excel. In this worksheet, the variables are pooled, and mean, AUC, and calculated values are computed. From Excel, data can be transferred to a statistical analysis program of choice. Because of the highly interdependent nature of CGMS values, others caution that traditional statistics may be inappropriate and suggest that mathematical modeling techniques are more suitable.23

Average across a 72-h period

The goal is to use the mean of two or three consecutive variables across 2–3 days in the final analysis. For example, the mean of 2-h breakfast AUC calculations from 2 consecutive days serve as the overall average 2-h breakfast AUC. There should be three nocturnal periods in 72 h, so three values are averaged for the variable to be used in the analysis (i.e., lowest nocturnal BG, nocturnal BG).7

Application of Approach Using Unpublished Data in Pregnant Women at Risk for Excess Fetal Growth

Differences in populations of pregnant women at risk for fetal overgrowth without diabetes

We have utilized some of our original unpublished data to illustrate how these measures can be applied to detect differences in glycemia between groups of pregnant women. Using CGMS, it was possible to graphically portray a significantly higher 24-h mean glucose AUC in obese women, of which a minority were later diagnosed with gestational diabetes mellitus, versus normal-weight pregnant women (Fig. 1A) despite fixed diets. Figure 1B provides a more detailed, graphical view of the 24-h mean glucose AUC as distinct patterns between the obese and normal-weight women at 15–16 versus 27–28 weeks of gestation using CGMS® System Gold™ (Medtronic Minimed). In this study, both groups of women were placed on eucaloric diets with a macronutrient content of 50% carbohydrate, 35% fat, and 15% protein. All food was provided. Almost all measures of glycemia within a 24-h period were higher in obese compared with normal-weight women despite a fixed diet in which total calories and macronutrient composition were precisely matched. These general observations are supported by a previous study we published and by others7,10 in less controlled settings.

FIG. 1.

FIG. 1.

(A) Difference in mean 24-h glucose area under the curve (AUC) (2 days) by continuous glucose monitoring system at week 15–16 and 27–28 in pregnancy between normal-weight (NW) and obese (OB) groups. Data are mean±SEM values (n=13; six OB [three later diagnosed with gestational diabetes mellitus] and seven NW). Paired data were assessed with the repeated-measures analysis of variance model for 15–16 weeks versus 27–28 weeks: P=0.004 for week 27–28. (B) Pattern of 24-h glycemia at gestational weeks 15–16 and 27–28 by continuous glucose monitoring system.

Infant adiposity through a CGMS lens

We have observed strong positive associations between our identified CGMS glucose variables and neonatal adiposity across normal-weight and obese women without gestational diabetes.7,8 As shown in Figure 2 in this unpublished cohort of obese and normal-weight pregnant women in which dual-energy X-ray absorptiometry at 2 weeks of birth was used to measure infant adiposity, these associations were strongest at 27–28 weeks of gestation, compared with gestational week 15–16. In particular, the 1- and 2-h mean postprandial breakfast glucose responses at 27–28 weeks of gestation (including 2-h AUC) were associated with neonatal adiposity at 2 weeks of life across measurement methods (dual-energy X-ray absorptiometry, skinfold calipers; r value range, 0.699–0.867; P<0.02 for all correlations) (Fig. 2). In 2001, Sivan et al.43 reported that the response to breakfast in pregnant women with gestational diabetes resulted in 2.5-fold higher glucose concentrations. Our observations are consistent with the published literature where postprandial meter glucose was correlated with infant size at birth.44,45 Thus, across pregnant women, CGMS technology has revealed important differences and associations between maternal glycemia and neonatal adiposity using these data approaches in controlled8 and less controlled7 settings.

FIG. 2.

FIG. 2.

Across normal-weight and obese pregnant women, 1-h postprandial blood glucose level by continuous glucose monitoring system (at 27–28 weeks of gestation) is highly associated with infant's percentage body fat as measured by dual-energy X-ray absorptiometry at 2 weeks of life (n=11).

A salient finding in these pilot data was the continuous positive association between postprandial maternal glucose (including 2-h AUC) at gestational week 27–28 and neonatal adiposity across normal-weight and obese women with or without gestational diabetes mellitus.8,9 Recognizing the elusive nature of fetal fat accretion and the ability to observe full 24-h patterns of glycemia through CGMS, we further dichotomized women in our preliminary data based on the infant adiposity phenotype (total percentage body fat) instead of the maternal phenotype (body mass index). Figure 3 shows that despite being normal weight or obese, the women who had infants with excess adiposity (≥16%; black line) had higher patterns of glycemia over 24 h and particularly after meals. In fact, in the women who had infants with ≥16% body fat (2 NW, 1 OB, 1 GDM), CGMS revealed a higher fasting glucose, followed by an accentuated postprandial breakfast period, with the other meals following precedence. The illustration demonstrates a novel view of maternal glycemia by infant outcome through CGMS technology.

FIG. 3.

FIG. 3.

Comparison of 24-h maternal glycemia between women who had infants with either ≥16% body fat (BF) (n=4) versus <16% BF (n=7). Arrows denote postmeal excursions.

Conclusions

The capacity for CGMS to record consecutive measures throughout a 24-h period in fasting and postprandial states at different gestational ages, while manipulating diet and controlling physical activity, is a powerful tool to better understand glycemic patterns in pregnancy in relation to fetal growth. However, the data are only clinically valuable if the conditions under which they are used are clearly described, variables of interest are carefully defined, and methods to deal with incongruous data are established prospectively. Particularly in pregnancy, more uniform variable definition, handling, and reporting of CGMS glucose data are necessary to further scientific investigation and draw meaningful conclusions. There are likely to be other CGMS-derived glucose variables that are clinically relevant to specific outcomes of other studies in pregnant women, and we encourage clear description of them so that a working methodology might become available in the field. Our understanding of the contribution of glycemic patterns to fetal growth, compared with other nutrients, and how diet and physical activity can modify glucose availability to the fetus will surely increase our knowledge of why some pregnant women, with or without diabetes, deliver infants with excess adiposity who are at risk for neonatal hypoglycemia. A request by editorial review committees for researchers to clearly specify the CGMS measures and analytical approaches used will advance the understanding of both investigators and practitioners as to which glycemic patterns in pregnancy optimize both maternal and fetal outcomes.

Acknowledgments

We would like to acknowledge our support from grants RO1DK 078645 and R21 DK 088324 from the National Institutes of Health, as well as the support of Dr. Jacob Friedman. Further support by NIH/NCRR Colorado CTSI grant UL1 TR000154 is also acknowledged.

Author Disclosure Statement

No competing financial interests exist. T.L.H. and L.A.B. contributed equally to writing and editing this manuscript. The approach to CGMS data outlined in this manuscript was conceived by T.L.H. and mentored by L.A.B.

References

  • 1.Kitzmiller JL. Block JM. Brown FM. Catalano PM. Conway DL. Coustan DR. Gunderson EP. Herman WH. Hoffman LD. Inturrisi M. Jovanovic LB. Kjos SI. Knopp RH. Montoro MN. Ogata ES. Paramsothy P. Reader DM. Rosenn BM. Thomas AM. Kirkman MS. Managing preexisting diabetes for pregnancy: summary of evidence and consensus recommendations for care. Diabetes Care. 2008;31:1060–1079. doi: 10.2337/dc08-9020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.de Veciana M. Major CA. Morgan MA. Asrat T. Toohey JS. Lien JM. Evans AT. Postprandial versus preprandial blood glucose monitoring in women with gestational diabetes mellitus requiring insulin therapy. N Engl J Med. 1995;333:1237–1241. doi: 10.1056/NEJM199511093331901. [DOI] [PubMed] [Google Scholar]
  • 3.Hernandez TL. Friedman JE. van Pelt RE. Barbour LA. Patterns of glycemia in normal pregnancy: should the current therapeutic targets be challenged? Diabetes Care. 2011;34:1660–1668. doi: 10.2337/dc11-0241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.The HAPO Study Cooperative Research Group: Hyperglycemia and Adverse Pregnancy Outcome (HAPO) Study: associations with neonatal anthropometrics. Diabetes. 2009;58:453–459. doi: 10.2337/db08-1112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Metzger BE. Lowe LP. Dyer AR. Trimble ER. Chaovarindr U. Coustan DR. Hadden DR. McCance DR. Hod M. McIntyre HD. Oats JJ. Persson B. Rogers MS. Sacks DA. Hyperglycemia and adverse pregnancy outcomes. N Engl J Med. 2008;358:1991–2002. doi: 10.1056/NEJMoa0707943. [DOI] [PubMed] [Google Scholar]
  • 6.Boney CM. Verma A. Tucker R. Vohr BR. Metabolic syndrome in childhood: association with birth weight, maternal obesity, and gestational diabetes mellitus. Pediatrics. 2005;115:e290–e296. doi: 10.1542/peds.2004-1808. [DOI] [PubMed] [Google Scholar]
  • 7.Harmon KA. Gerard L. Jensen DR. Kealey EH. Hernandez TL. Reece MS. Barbour LA. Bessesen DH. Continuous glucose profiles in obese and normal-weight pregnant women on a controlled diet: metabolic determinants of fetal growth. Diabetes Care. 2011;34:2198–2204. doi: 10.2337/dc11-0723. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Hernandez TL. Patterns of Glycemia During the Insulin Resistance of Pregnancy: Obesity, Gestational Diabetes and Relationship to Neonatal Adiposity [Ph.D. thesis] Denver: University of Colorado; 2009. [Google Scholar]
  • 9.Hernandez TL. van Pelt RE. Jensen DR. Reece MS. Kahn B. Friedman JE. Barbour LA. Glycemic patterns that predict neonatal adiposity in lean and obese pregnant women on fixed diets [abstract] Diabetes. 2009;58:A640. [Google Scholar]
  • 10.Yogev Y. Ben Haroush A. Chen R. Rosenn B. Hod M. Langer O. Diurnal glycemic profile in obese and normal weight nondiabetic pregnant women. Am J Obstet Gynecol. 2004;191:949–953. doi: 10.1016/j.ajog.2004.06.059. [DOI] [PubMed] [Google Scholar]
  • 11.Rodbard D. Interpretation of continuous glucose monitoring data: glycemic variability and quality of glycemic control. Diabetes Technol Ther. 2009;11(Suppl 1):S55–S67. doi: 10.1089/dia.2008.0132. [DOI] [PubMed] [Google Scholar]
  • 12.Zisser HC. Biersmith MA. Jovanovic LB. Yogev Y. Hod M. Kovatchev BP. Fetal risk assessment in pregnancies complicated by diabetes mellitus. J Diabetes Sci Technol. 2010;4:1368–1373. doi: 10.1177/193229681000400610. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Clarke W. Kovatchev B. Statistical tools to analyze continuous glucose monitor data. Diabetes Technol Ther. 2009;11(Suppl 1):S45–S54. doi: 10.1089/dia.2008.0138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Rodbard D. Bailey T. Jovanovic L. Zisser H. Kaplan R. Garg SK. Improved quality of glycemic control and reduced glycemic variability with use of continuous glucose monitoring. Diabetes Technol Ther. 2009;11:717–723. doi: 10.1089/dia.2009.0077. [DOI] [PubMed] [Google Scholar]
  • 15.Voelmle M. Ritchie P. Naik R. Garg S. Improved A1C values with no increase in hypoglycemia in pregnant women with Type 1 diabetes using real-time continuous home glucose monitors [abstract] Diabetes. 2008;57:A117. [Google Scholar]
  • 16.Sacks DB. Bruns DE. Goldstein DE. Maclaren NK. McDonald JM. Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2002;48:436–472. [PubMed] [Google Scholar]
  • 17.D'Orazio P. Burnett RW. Fogh-Andersen N. Jacobs E. Kuwa K. Kulpmann WR. Larsson L. Lewenstam A. Maas AH. Mager G. Naskalski JW. Okorodudu AO. Approved IFCC recommendation on reporting results for blood glucose (abbreviated) Clin Chem. 2005;51:1573–1576. doi: 10.1373/clinchem.2005.051979. [DOI] [PubMed] [Google Scholar]
  • 18.Consensus statement on self-monitoring of blood glucose. Diabetes Care. 1987;10:95–99. [PubMed] [Google Scholar]
  • 19.American Diabetes Association: Standards of medical care in diabetes—2008. Diabetes Care. 2008;31(Suppl 1):S12–S54. doi: 10.2337/dc08-S012. [DOI] [PubMed] [Google Scholar]
  • 20.Kovatchev B. Anderson S. Heinemann L. Clarke W. Comparison of the numerical and clinical accuracy of four continuous glucose monitors. Diabetes Care. 2008;31:1160–1164. doi: 10.2337/dc07-2401. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Gross TM. Bode BW. Einhorn D. Kayne DM. Reed JH. White NH. Mastrototaro JJ. Performance evaluation of the MiniMed continuous glucose monitoring system during patient home use. Diabetes Technol Ther. 2000;2:49–56. doi: 10.1089/152091500316737. [DOI] [PubMed] [Google Scholar]
  • 22.Gross TM. Ter VA. Continuous glucose monitoring in previously unstudied population subgroups. Diabetes Technol Ther. 2000;2(Suppl 1):S27–S34. doi: 10.1089/15209150050214096. [DOI] [PubMed] [Google Scholar]
  • 23.Kovatchev B. Breton M. Clarke W. Analytical methods for the retrieval and interpretation of continuous glucose monitoring data in diabetes. Methods Enzymol. 2009;454:69–86. doi: 10.1016/S0076-6879(08)03803-2. [DOI] [PubMed] [Google Scholar]
  • 24.Chen R. Yogev Y. Ben Haroush A. Jovanovic L. Hod M. Phillip M. Continuous glucose monitoring for the evaluation and improved control of gestational diabetes mellitus. J Matern Fetal Neonatal Med. 2003;14:256–260. doi: 10.1080/jmf.14.4.256.260. [DOI] [PubMed] [Google Scholar]
  • 25.Siegmund T. Rad NT. Ritterath C. Siebert G. Henrich W. Buhling KJ. Longitudinal changes in the continuous glucose profile measured by the CGMS in healthy pregnant women and determination of cut-off values. Eur J Obstet Gynecol Reprod Biol. 2008;139:46–52. doi: 10.1016/j.ejogrb.2007.12.006. [DOI] [PubMed] [Google Scholar]
  • 26.Ben Haroush A. Yogev Y. Chen R. Rosenn B. Hod M. Langer O. The postprandial glucose profile in the diabetic pregnancy. Am J Obstet Gynecol. 2004;191:576–581. doi: 10.1016/j.ajog.2004.01.055. [DOI] [PubMed] [Google Scholar]
  • 27.Cypryk K. Pertynska-Marczewska M. Szymczak W. Wilcynski J. Lewinski A. Evaluation of metabolic control in women with gestational diabetes mellitus by the continuous glucose monitoring system: a pilot study. Endocr Pract. 2006;12:245–250. doi: 10.4158/EP.12.3.245. [DOI] [PubMed] [Google Scholar]
  • 28.Buhling KJ. Winkel T. Wolf C. Kurzidim B. Mahmoudi M. Wohlfarth K. Wascher C. Schink T. Dudenhausen JW. Optimal timing for postprandial glucose measurement in pregnant women with diabetes and a non-diabetic pregnant population evaluated by the Continuous Glucose Monitoring System (CGMS) J Perinat Med. 2005;33:125–131. doi: 10.1515/JPM.2005.024. [DOI] [PubMed] [Google Scholar]
  • 29.Porter H. Lookinland S. Belfort MA. Evaluation of a new real-time blood continuous glucose monitoring system in pregnant women without gestational diabetes. A pilot study. J Perinat Neonatal Nurs. 2004;18:93–102. doi: 10.1097/00005237-200404000-00004. [DOI] [PubMed] [Google Scholar]
  • 30.Buhling KJ. Kurzidim B. Wolf C. Wohlfarth K. Mahmoudi M. Wascher C. Siebert G. Dudenhausen JW. Introductory experience with the continuous glucose monitoring system (CGMS; Medtronic Minimed) in detecting hyperglycemia by comparing the self-monitoring of blood glucose (SMBG) in non-pregnant women and in pregnant women with impaired glucose tolerance and gestational diabetes. Exp Clin Endocrinol Diabetes. 2004;112:556–560. doi: 10.1055/s-2004-830399. [DOI] [PubMed] [Google Scholar]
  • 31.Adolfsson P. Nilsson S. Lindblad B. Continuous glucose monitoring system during physical exercise in adolescents with type 1 diabetes. Acta Paediatr. 2011;100:1603–1609. doi: 10.1111/j.1651-2227.2011.02390.x. [DOI] [PubMed] [Google Scholar]
  • 32.Willman SP. Leveno KJ. Guzick DS. Williams ML. Whalley PJ. Glucose threshold for macrosomia in pregnancy complicated by diabetes. Am J Obstet Gynecol. 1986;154:470–475. doi: 10.1016/0002-9378(86)90692-7. [DOI] [PubMed] [Google Scholar]
  • 33.Metzger BE. Buchanan TA. Coustan D. de Leiva A. Dunger DB. Hadden DR. Hod M. Kitzmiller JL. Kjos SL. Oats JN. Pettitt DJ. Sacks DA. Zoupas C. Summary, recommendations of the Fifth International Workshop-Conference on Gestational Diabetes Mellitus. Diabetes Care. 2007;30(Suppl 2):S251–S260. doi: 10.2337/dc07-s225. Erratum in Diabetes Care 2007;30:3154. [DOI] [PubMed] [Google Scholar]
  • 34.Langer O. Levy J. Brustman L. Anyaegbunam A. Merkatz R. Divon M. Glycemic control in gestational diabetes mellitus—how tight is tight enough: small for gestational age versus large for gestational age? Am J Obstet Gynecol. 1989;161:646–653. doi: 10.1016/0002-9378(89)90371-2. [DOI] [PubMed] [Google Scholar]
  • 35.Sacks DA. Chen W. Wolde-Tsadik G. Buchanan TA. When is fasting really fasting? The influence of time of day, interval after a meal, and maternal body mass on maternal glycemia in gestational diabetes. Am J Obstet Gynecol. 1999;181:904–911. doi: 10.1016/s0002-9378(99)70323-6. [DOI] [PubMed] [Google Scholar]
  • 36.Pedersen J. Course of diabetes during pregnancy. Acta Endocrinol (Copenh) 1952;9:342–364. doi: 10.1530/acta.0.0090342. [DOI] [PubMed] [Google Scholar]
  • 37.Hay WW., Jr Placental-fetal glucose exchange and fetal glucose metabolism. Trans Am Clin Climatol Assoc. 2006;117:321–339. [PMC free article] [PubMed] [Google Scholar]
  • 38.Potteiger JA. Jacobsen DJ. Donnelly JE. A comparison of methods for analyzing glucose and insulin areas under the curve following nine months of exercise in overweight adults. Int J Obes Relat Metab Disord. 2002;26:87–89. doi: 10.1038/sj.ijo.0801839. [DOI] [PubMed] [Google Scholar]
  • 39.McGarraugh G. The chemistry of commercial continuous glucose monitors. Diabetes Technol Ther. 2009;11(Suppl 1):S-17–S-24. doi: 10.1089/dia.2008.0133. [DOI] [PubMed] [Google Scholar]
  • 40.Metzger BE. Phelps RL. Freinkel N. Navickas IA. Effects of gestational diabetes on diurnal profiles of plasma glucose, lipids, and individual amino acids. Diabetes Care. 1980;3:402–409. doi: 10.2337/diacare.3.3.402. [DOI] [PubMed] [Google Scholar]
  • 41.Phelps RL. Metzger BE. Freinkel N. Carbohydrate metabolism in pregnancy. XVII. Diurnal profiles of plasma glucose, insulin, free fatty acids, triglycerides, cholesterol, and individual amino acids in late normal pregnancy. Am J Obstet Gynecol. 1981;140:730–736. [PubMed] [Google Scholar]
  • 42.Zijlstra E. Heise T. Nosek L. Heinemann L. Heckermann S. Continuous glucose monitoring: quality of hypoglycaemia detection. Diabetes Obes Metab. 2012 Sep 13; doi: 10.1111/dom.12001. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
  • 43.Sivan E. Weisz B. Homko CJ. Reece EA. Schiff E. One or two hours postprandial glucose measurements: are they the same? Am J Obstet Gynecol. 2001;185:604–607. doi: 10.1067/mob.2001.117184. [DOI] [PubMed] [Google Scholar]
  • 44.Jovanovic-Peterson L. Peterson CM. Reed GF. Metzger BE. Mills JL. Knopp RH. Aarons JH. Maternal postprandial glucose levels, infant birth weight: the Diabetes in Early Pregnancy Study. The National Institute of Child Health and Human Development—Diabetes in Early Pregnancy Study. Am J Obstet Gynecol. 1991;164:103–111. doi: 10.1016/0002-9378(91)90637-7. [DOI] [PubMed] [Google Scholar]
  • 45.Parfitt VJ. Clark JD. Turner GM. Hartog M. Maternal postprandial blood glucose levels influence infant birth weight in diabetic pregnancy. Diabetes Res. 1992;19:133–135. [PubMed] [Google Scholar]