Abstract
Purpose
Temperature is a key measure in human red blood cell concentrate (RBC) quality control. A precise description of transient temperature distributions in RBC units removed from steady storage exposed to ambient temperature is at present unknown. Magnetic resonance thermometry was employed to visualize and analyse RBC warm up processes, to describe time courses of RBC mean, surface and core temperatures by an analytical model, and to determine and investigate corresponding model parameters.
Methods
Warm-up processes of 47 RBC units stored at 1–6°C and exposed to 21.25°C ambient temperature were investigated by proton resonance frequency thermometry. Temperature distributions were visualized and analysed with dedicated software allowing derivation of RBC mean, surface and core temperature-time courses during warm up. Time-dependence of mean temperature was assumed to fulfil a lumped capacitive model of heat transfer. Time courses of relative surface and core temperature changes to ambient temperature were similarly assumed to follow shifted exponential decays characterized by a time constant and a relative time shift, respectively.
Results
The lumped capacitive model of heat transfer and shifted exponential decays described time-dependence of mean, surface and core temperatures close to perfect (mean R2 were 0.999±0.001, 0.996±0.004 and 0.998±0.002, respectively). Mean time constants were τ mean = 55.3±3.7 min, τ surface = 41.4±2.9 min and τ core = 76.8±7.1 min, mean relative time shifts were Δsurface = 0.07±0.02 and Δcore = 0.04±0.01. None of the constants correlated significantly with temperature differences between ambient and storage temperature.
Conclusion
Lumped capacitive model of heat transfer and shifted exponential decays represent simple analytical formulas to describe transient mean, surface and core temperatures of RBC during warm up, which might be a helpful tool in RBC temperature monitoring and quality control. Independence of constants on differences between ambient and storage temperature suggests validity of models for arbitrary storage and ambient temperatures.
Introduction
Storage and transport of human red blood cell concentrate (RBC) is subject of a variety of laws and regulations whereas temperature is a key measure in RBC quality control. According to international guidelines, RBC should be stored at 1 to 6°C to mitigate bacterial contamination, slow metabolic processes but maintain RBC viability [1], [2], [3]. During transport RBC core temperature must not exceed the upper temperature limit of 10°C, else units must be used or discharged [4], [5], [6]. To comply with this standard, RBC transport temperature is commonly estimated either by the time a sample is outside of controlled refrigerated storage or via temperature sensitive labels attached to the surface of RBC pouch. Both methods however give only rather crude estimates of RBC temperature: According to the well-established “30 minutes rule in transfusion medicine” RBCs may be returned to 1 to 6°C storage within 30 min as general upper time limit, whereas neither RBC storage and ambient temperature nor sample volume is taken into account [7], [8], [9]. Controlling temperature by temperature indicators or time–temperature integrator units attached to the surface of the RBC pouch implies that surface temperature well represents sample’s core temperature, which can but doesn’t have to be true [10], [11], [12], [13], [14], [15]. Facing worldwide decreased availability of lifesaving RBC challenges a more precise evaluation of correlations between maximum (surface), mean and minimum (core) temperature in RBC units removed from stock to prevent unnecessary wastage of blood components and, if transfused, ensure minimal risk to patients [1], [2], [3], [16], [17].
Magnetic resonance (MR) thermometry represents a unique tool for non-invasive, contactless, three-dimensional assessment of thermal processes. Although temperature monitoring is feasible with various temperature sensitive MR parameters, proton resonance frequency (PRF) temperature mapping is the method of choice for fast, stable and accurate temperature measurements [18], [19], [20]. After calibration, discrete time-series of mean, surface and core temperatures of RBC units during warm up can be derived from PRF images, which allow analytical modelling of warm up processes. As RBC pouch’s surface resistance to heat transfer can be assumed to be larger than the thermal resistance within RBC, time-dependence of RBC units’ mean temperature during warm up should be well described by a lumped capacitance model of heat transfer, with relative mean to ambient temperature differences following an exponential decay [21], [22]. Additionally, similar time-shifted exponential decays might provide adequate descriptions of surface and core temperature time courses during warm up.
The purpose of the present study was 1) to visualize and analyse warm up processes of RBC withdrawn from steady storage conditions by PRF thermometry, 2) to investigate appropriateness of lumped capacitance model-like behaviour of mean, surface and core temperatures during warm up and 3) to determine and analyse corresponding RBC specific parameters with respect to temperature and geometry dependence.
Materials and Methods
RBC Collection
59 RBC units (12 units for calibration and 47 RBC units for warm up experiments), discharged from routine stock of the Department of Blood Group Serology and Transfusion Medicine of the Medical University of Graz due to leucocyte or neopterin elevation, were used within the study. Whole blood donations were obtained according to the Austrian regulations for blood donation [23] from healthy volunteer blood donors after informed written consent and collected into triple bags (LCR5 filtration set) containing 63 ml citrate phosphate dextrose (MacoPharma LAB. Pharmaceutiques, Tourcoing, France). After centrifugation at 4000·g for 10 min at 20°C, RBC and plasma were separated from buffy coat fraction and transferred into satellite containers using an automated separator (Compomat G4, NPBI, Amsterdam, Netherlands). Within 30 min after separation RBC was leukoreduced using the LCR5 leukoreduction filter (MacoPharma LAB. Pharmaceutiques, Tourcoing, France) and re-suspended in 100 ml of saline adenine glucose mannitol (SAGM containing 900 mg of glucose monohydrate, 877 mg of sodium chloride, 525 mg of mannitol and 16.9 mg of adenine).
Sample Storage and Temperature Monitoring
RBC units were stored at temperatures between 1 to 6°C (MobiCool W35 12/230V, Zhuhai, China). To exclude short term effects of the cooling system RBC unit’s storage temperature T storage was derived from a RFID temperature data logger (SensoTag Siemens HFST-T109, Vienna, Austria, overall accuracy of ±0.5°C) mounted in a reference RBC unit as internal probe [24].
Throughout experiments ambient temperature in the MR investigation room T ambient was stabilized by air conditioning to 21.25°C and controlled by RFID temperature data loggers in 5 min time intervals.
Determination of PRF Thermal Coefficient of RBC
To derive PRF thermal coefficient of RBC, a series of 12 calibration measurements was performed [25], [26]. A calibration unit (Fig. 1a) build-up of refrigerated RBC, a reference bulb thermometer (Ludwig Schneider Messtechnik, Wertheim, Germany, overall accuracy ±0.2°C) and a reference phantom (5 g agar solved in 500 ml water) kept at ambient temperature to measure non-temperature related phase drifts was positioned in a 1.5 T MR system (Siemens Magnetom Espree, Erlangen, Germany), centered in a 12 channel head coil. After acquisition of reference images, difference phase images covering the thermometer tip with 7 gapless transversal slices and thermometer reference temperatures were measured with 3 min pause time for RBC temperatures between 1 and 18°C. Parameters of the 2D multislice gradient echo (GRE) sequence were repetition time = 33.15 ms, echo time TE = 20 ms, flip angle = 14°, voxel size = 1.1×1.1×7.5 mm3, field of view = 170×170 mm2, bandwidth = 65 Hz/Pix and data acquisition time = 38 s. (Due to this protocol no radio frequency induced heating was noticed in bulb thermometer equipped phantoms at ambient temperature.).
Temperature induced phase shifts were obtained as differences between means of phase changes of RBC located around the thermometer tip during warm up and means of reference phantom phases [27]. A PRF thermal coefficient α RBC of RBC was determined for each calibration measurement series from linear fit of resultant phase differences φ as function of temperature T between 1 and 18°C according to α RBC = (1/2π⋅γ⋅B0⋅TE)⋅(dφ/dT), where γ is the gyromagnetic ratio of 1H nuclei, B 0 the main magnetic field strength and dφ/dT the slope of the respective fit [26]. α RBC mean value was used for calculation of temperature maps in RBC warm up experiments.
PRF Thermometry of RBC Warm Up
To prevent uncontrolled thermal transfer processes and motion as well as to ensure a fast and reproducible experimental onset, RBC units were stored, transported and investigated in upright position in a two-layer plastic holder (Lego®) with minimum distance of 1 mm between frame and RBC. Withdrawn from storage, RBC units were plugged into an adjustment unit (Fig. 1b) equipped with 2 reference phantoms (kept at T ambient throughout experiments) which was positioned in the MR system in the center of a 12 channel head coil. Immediately after exposition to ambient temperature (mean setup time = 16±3 s) reference and difference phase images were acquired with the multislice GRE sequence described for calibration measurements covering the entire RBC volume with equidistant sagittal slices (gap = 90–95%). After 20 measurements recorded with 3 min pause time (total investigation time = 69.7 min), turbo spin echo (TSE) images for segmentation were acquired in identical slice positions (repetition time = 3000 ms, echo time = 51 ms, flip angle = 170°, voxel size = 0.5×0.5×7.5 mm3, field of view = 170×170 mm2, bandwidth = 159 Hz/Pix, turbo factor = 5, number of slices = 7, data acquisition time = 3 min 17 s).
Image Analysis
Phase difference images were transformed to RBC temperature maps by dedicated software developed in Matlab (R2010b. The MathWorks Inc., Natick, Massachusetts). Evaluation flow chart, including RBC, SAGM and reference phantom segmentation, determination of non-temperature-related phase differences in the reference phantoms and correction of phase differences in RBC by the median of the phase differences in the reference phantoms is shown in Fig. 2. From corrected phase differences φ in RBC, corresponding temperatures T were calculated according to T = T storage+(φ/2π⋅α RBC ⋅γ⋅B0⋅TE) where α RBC is the mean PRF thermal coefficient of RBC derived from calibration measurements and T storage the temperature of RBC at onset of measurements [19]. RBC volume (as RBC area times slice distance), RBC pouch’s volume (as pouch’s area including RBC and SAGM times slice distance) height and width as well as its center position were derived from segmentation. Thermal RBC core position was defined as the position in the sample, where time-averaged temperature was minimal.
At each time temperature maps were employed to determine RBC mean temperature T mean (calculated as average temperature of all voxels in the RBC volume) as well as RBC volume fractions vol(T) exceeding a temperature T for all integer-valued T>T storage. Linear extrapolation at half maximum of volume fraction-time curve vol(T) was used to specify (1) the time when RBC volume starts to exceed T, allowing T to be interpreted as surface temperature T surface and (2) the time when 100% of RBC volume exceeds T, allowing T to be interpreted as core temperature T core. Evaluation of RBC volume fractions vol(T) for different temperatures T resulted in time courses of RBC surface and core temperatures (Fig. 3).
Modeling Time Dependence of Mean, Surface and Core Temperature during Warm Up
For the analytical description of RBC’s warm up processes, it was hypothesized, that mean temperature T mean during warm up is described by a lumped capacitance model of heat transfer [21], [22]. Consequently relative mean temperature difference θ mean = (T ambient–T mean)/(T ambient–T storage) to ambient temperature Tambient should fulfill
(1) |
where t is the time the RBC pouch is exposed to ambient temperature and τ mean is a RBC pouch specific time constant, not dependent on storage temperature T storage or difference between ambient and storage temperature T ambient–T storage, respectively. Moreover, for RBC surface and core temperatures it was hypothesized, that relative surface and core temperature differences θ surface and θ core to ambient temperature T ambient follow “shifted” exponential decays. Thus, θ surface and θ core should fulfill
(2) |
(3) |
with time constants τ surface and τ core and shifts Δsurface and Δcore. The shifts characterize (in units of time constants) how much earlier (or later in case of core) the decay of θ surface (and θ core) starts than the instantaneous decay of θ mean.
Relative temperature spread θ spread in RBC was introduced according to θ core–θ surface, becoming maximal at
(4) |
Moreover, by eliminating time, Eqs (1)-(3) can be transformed to
(5) |
(6) |
relating relative surface temperature difference with relative mean and core temperatures.
Statistical Analysis
Statistical analysis including linear fit of phase differences versus temperature for the calibration measurements as well as fitting of the exponential decays of the relative temperatures to the experimental RBC warm up data was performed using NCSS (Hintze J. (2007). NCSS. LLC. Kaysville, Utah). Mean values are given together with standard deviations. Standard deviations of fitted parameters were employed to calculate uncertainties of (relative) temperature (difference) courses via Gaussian law of error propagation. Dependency of time constants, shifts and (time of) maximal temperature spread on storage temperature and geometric parameters of the RBC pouches were investigated by means of correlation and linear regression analysis. p-values smaller 0.01 were considered as statistically significant.
Results
PRF Thermal Coefficient of RBC
Linear correlation between RBC phase differences and temperature changes in the calibration measurements was close to perfect (mean correlation coefficient 0.999±0.001). Mean PRF thermal coefficient of RBC was α RBC = 1.05⋅10−8±0.02⋅10−8°C−1.
Qualitative Description of RBC Warm Up Temperature Distributions
Similar qualitative warm up phenomena were observed in all RBC units irrespectively of storage temperatures ranging from 1 to 6°C (mean T storage = 3.6±1.4°C). A typical example of RBC temperature maps derived from PRF thermometry measurements is shown in Fig. 4.
Withdrawn from steady storage conditions, RBC’s temperature at the surface of the pouch and at SAGM border immediately started to rise. During warm up heat was continuously transferred from RBC surface to the adjacent inner layers, causing an increasingly non-uniform temperature distribution in the sample with isotherms located symmetrically around temperature core RBCcore in both transversal (horizontal and latitudinal) extensions of the unit. In vertical extension RBCcore was displaced by 13±4% from pouch center. Highest temperatures, the respective T surface, were reached at the surface at sections with smaller width at the top of the pouch.
Increases in mean, surface and core temperatures (or decreases of relative mean, surface and core temperature differences) decreased with increasing warm-up time.
Time and “Geometry” Dependence of Mean, Surface and Core Temperatures during Warm Up
Measured relative temperature differences θ mean, θ surface and θ core fulfilled lumped capacitance model of heat transfer [Eq. (1)] and “shifted” exponential decays [Eqs (2) and (3)] close to perfect. Mean R 2 were 0.999±0.001, 0.996±0.004 and 0.998±0.002, respectively. The resulting mean time constants were τ mean = 55.3±3.7 min, τ surface = 41.4±2.9 min and τ core = 76.8±7.1 min, resulting mean time shifts Δsurface = 0.07±0.02 and Δcore = 0.04±0.01. Average time courses of θ mean, θ surface and θ core together with corresponding uncertainties are shown in Fig. 5. Mean time when relative temperature spread is maximal [Eq. (4)] was t spread = 45.5±4.1 min, the corresponding mean maximal relative temperature spread was θ spread(t spread) = 0.26±0.04.
Neither time constants and shifts nor (time of) maximal relative temperature spread depended on storage temperature or T ambient–T storage. Variations in geometric parameters RBC volume (255±17 ml), pouch volume (287±26 ml), pouch height h (13.1±0.5 cm) and width w (3.4±0.2 cm) were small and shifts as well as maximal relative temperature spread did not depend on these parameters (except a weak but significant correlation of r = −0.47 between Δsurface and pouch volume). Correlations between time constants and time of maximal temperature spread with geometric parameters are summarized in Table 1. Linear regression results of τ mean, τ surface, τ core and t spread versus the width-height-ratio w/h are shown in Fig. 6.
Table 1. Correlation coefficients between time constants and time of maximal temperature spread with geometric RBC pouch parameters.
Geometric parameter | τ mean | τ surface | τ core | t spread |
RBC volume | 0.39* a | 0.17 | 0.40* | 0.41* |
Pouch volume | 0.59* | −0.02 | 0.46* | 0.55* |
RBC-to-pouch-volume-ratio | −0.38* | 0.18 | −0.23 | −0.34 |
Height h | −0.50* | −0.39* | −0.32 | −0.47* |
Width w | 0.68* | 0.38* | 0.59* | 0.68* |
Width-height-ratio w/h | 0.72* | 0.46* | 0.57* | 0.71* |
*indicates statistical significance.
Temperature Prediction from Surface Temperature
Mean quotients of time constants τ surface/τ mean and τ surface/τ core appearing in Eqs (5) and (6) were 0.75±0.06 and 0.54±0.06, respectively. Both did not depend significantly on storage temperature and T ambient–T storage. Significant correlations with geometric parameters of RBC pouches were found for pouch volume (r = −0.53 in case of τ surface/τ mean, r = −0.40 in case of τ surface/τ core).
Average dependencies of relative mean and core temperature differences θ mean and θ core on relative surface temperature difference θ surface together with corresponding uncertainties are shown in Fig. 7. The influence of storage temperature T storage on the prediction of absolute mean or core temperature from absolute surface temperature is indicated in Fig. 8.
Discussion
Time-resolved temperature distributions of RBC units removed from refrigerated storage between 1 and 6°C and exposed to ambient temperature of 21.25°C assessed non-invasively by PRF thermometry were visualized and analyzed by dedicated software. Major findings derived from this study were (1) RBC mean temperature time course during warm up almost perfectly followed a lumped capacitance model law, (2) surface and core temperature warm up time courses of RBC almost perfectly fulfilled time-shifted exponential models and (3) neither time constants nor shifts depended significantly on storage temperature or the difference between ambient and storage temperatures, which supports the general applicability of the analytical description and corresponding RBC constants for arbitrary storage and ambient temperatures.
MR thermometry is typically employed for in vivo monitoring of temperature during thermotherapy-heating of tissue [18], [20], [27], [28], [29], [30]. Calibration experiments confirmed that PRF method, implying linear relationship between temperature and phase changes, is applicable to measure RBC temperature in the range between 1 and 18°C. PRF thermal coefficient α RBC derived within the present study of RBC was close to the one of water [31] and in the range of other biological tissue [19], [27].
Removed from refrigerated storage to ambient temperature, RBC surface immediately starts to warm up while temperature in the core of the sample rises time delayed to peripheral layers causing non-homogeneous temperature distributions in RBC. As the average temperature spread in RBC reaches more than a quarter of the difference between ambient and storage temperature, which is consistent with results derived from invasive measurements [14], RBC’s resistance to heat transfer is not negligible and specification of the temperature assessed is inevitable.
Aside from mean temperature, which represents the temperature after “gentle mixing”, surface and core temperatures were introduced essentially noise and partial volume independent (Fig. 3) as maximal and minimal temperature in RBC. As highest temperatures appeared at the RBC surface (Fig. 4) the term surface temperature for maximal temperature is justified. Temperatures at RBC pouch’s surface were, however, inhomogeneously distributed as previously demonstrated by Johnson et al [10]. Surface temperature therefore represents the upper temperature bound reached at thinner layers of the pouch. RBC core temperature, required to maintain below 10°C during transport by current guidelines [5], [6], [8], was reached vertically displaced from the pouch center towards thicker sample layers. This might explain discrepancies in core temperature time courses described in previous studies applying thermocouple data loggers [1], [9], [14], [32].
Time dependence of relative mean, surface and core temperature differences to ambient temperatures of all samples were very well described by (shifted) exponential decays. Time constants and shifts did not depend significantly on the difference between ambient and storage temperature. Consequently lumped-capacitance-model-like formulas Eqs. (1)–(3), together with the mean values of the fitted constants (Fig. 5), should provide suitable descriptions of RBC warm up for wide ranges of not only storage but also ambient temperatures. Standard deviations of time constants and shifts as well as derived uncertainties of the relative temperature difference-time-curves shown in Fig. 5 are not solely caused by measurement errors, but result also from deviations in RBC unit’s geometry. Although variations in volume, height, width and width-height-ratio were small, all time constants correlated significantly with these geometric parameters. In accordance with Perry et al. [9] core (and mean) temperature time constants increased with RBC and pouch volume, confirming that more voluminous samples need longer to warm up. However, the correlation of all time constants with RBC width and width-height-ratio was stronger. This behavior is to be expected from lumped capacitive model of heat transfer, because time constants are proportional to body volume per heat transfer or pouch surface area [21], which should in turn be estimated from RBC pouch width, or even better, from unit’s width-height-ratio.
According to lumped-capacitance-model-like formulas Eqs. (1)–(3), times when absolute mean, surface or core temperatures of RBC during warm up exceed a temperature limit strongly depend on storage and ambient temperature. For example keeping RBC outside from refrigerated storage for less than 30 min thus ensures RBC core temperature to stay below 10°C only within certain ranges of storage and ambient temperatures: The mean relative core temperature difference to ambient temperature after 30 min was 0.70, which transforms to limit ambient temperatures of 31 and 20°C for storage temperatures of 1 and 6°C such that core temperature stays below 10°C. These limit temperatures support validity and applicability of the “30 minutes rule”. As for storage and ambient temperature of 1 and 20°C respectively, 10°C RBC core temperature is reached in mean only after 52 min, current guidelines could eventually allow longer “reuse-phase” of RBC units.
For the use of temperature sensitive labels attached to the surface of RBC pouch the problem transforms to the question if “inside” RBC core temperature can reliably be predicted from RBC surface temperature. Recently it could be shown that times when surface, mean and core temperatures reach 10°C during warm up tremendously differ but can be correlated with each other [15]. Eqs (5) and (6) as well as Fig. 7 provide the general relationship between mean, core and surface temperatures during RBC warm up. The sole appearance of quotients of time constants renders relationships to large extent independent from RBC geometry. However, prediction of mean and core temperature from surface temperature depends on storage and ambient temperature. Therefore precise temperature monitoring by temperature sensitive labels attached to the surface of RBC pouch is possible only if storage and ambient temperatures are known as well. The impact of the “RBC storage temperature memory effect” is shown in Fig. 8 by specifying mean and core temperature relations on surface temperature for recommended lowest (1°C) and highest (6°C) storage temperature limits for an ambient temperature of 21.25°C.
Regarding the present study, several limitations should be mentioned. RBC warm up during fast setup (16 s ±3 s) was not taken into account and mixing of RBC prior to MR measurement was not possible as residual motion would affect PRF thermometry. Furthermore storage temperature was not measured directly from entire samples but derived from RFID data logger measurements mounted in a reference RBC unit as internal probe continuously kept at storage temperature throughout experiments.
Limited measurement time (70 min) and time resolution (3 min) of PRF measurements were chosen to minimize B0 drifts and warming due to radio frequency absorption and to keep the total scan time reasonable. As temperature changes were rather slow an increase in time resolution should not add additional information, behavior of warm up temperatures above 70 min warm up time is, an extrapolation, which is not proven by the experiments.
Finally it has to be addressed, that there is an enormous manufacturer variability in geometry, material and properties of red blood cell containers and the present results apply a priori only to MacoPharma blood collection systems used in the study.
In conclusion, lumped capacitive model of heat transfer and shifted exponential decays represent simple analytical formulas to describe transient mean, surface and core temperatures of RBC during warm up, which might be a helpful tool in RBC temperature monitoring and quality control. Independence of constants on differences between ambient and storage temperature suggests validity of models for arbitrary storage and ambient temperatures.
Funding Statement
The authors have no support or funding to report.
References
- 1. Brunskill S, Thomas S, Whitmore E, McDonald CP, Dorée C, et al. (2012) What Is the maximum time that a unit of red blood cells can be safely left out of controlled temperature storage. Transfus Med Rev 26(3): 209–223. [DOI] [PubMed] [Google Scholar]
- 2. Campbell-Lee S, Shaz B, Arena R, Sloan S, Fung M, et al. (2012) Red blood cell products: consideration of the discrepant temperature ranges permitted for storage versus transport. Transfusion 52(1): 195–200. [DOI] [PubMed] [Google Scholar]
- 3. Hancock V, Cardigan R, Thomas S (2011) Red cell concentrate storage and transport temperature. Transfus Med 21(5): 325–329. [DOI] [PubMed] [Google Scholar]
- 4.Electronic Code of Federal Regulations, 21 CFR 640.11(a) (2010). Washington, DC. US Government Printing Office. Available: http://ecfr.gpoaccess.gov/. Accessed 2012 January 12.
- 5.Council of Europe (2009) Guide for the Preparation, Use and Quality Assurance of Blood Components, 15th ed. Strasbourg: Council of Europe Publishing.
- 6.American Association of Blood Banks (2009) Standards for blood banks and transfusion services, 26th ed. Bethesda: American Association of Blood Banks.
- 7. Pick P, Fabijanic J (1971) Temperature changes in donor blood under different storage conditions. Transfusion 11(4): 213–215. [DOI] [PubMed] [Google Scholar]
- 8.WHO Guidelines, Good Distribution Practices (GDP) for Pharmaceutical Products, Annex 5. WHO Technical Report Series 937. WHO website. Available: http://www.who.int/bloodsafety/Manual_on_Management,Mainten ance_and_Use_of_Blood_Cold_Chain_Equipment.pdf. Accessed 2012 Jan 12.
- 9. Perry HE, Prasad P, Kirwan S, Huang YQ (2010) Core temperature changes in resuspended red blood cells (RBCs) and pediatric RBCs removed from refrigerated storage. Transfusion 50(1): 174–177. [DOI] [PubMed] [Google Scholar]
- 10. Johnson V, Langeberg A, Taye-Makuria A, Sandler G (2006) Temperature-sensitive labels for containers of RBCs. Am J Clin Pathol 126(3): 406–410. [DOI] [PubMed] [Google Scholar]
- 11. Davis R, Geiger B, Gutierrez A, Heaser J, Veeramani D (2009) Tracking blood products in blood centres using radio frequency identification: a comprehensive assessment. Vox Sang 97(1): 50–60. [DOI] [PubMed] [Google Scholar]
- 12. Klose T, Borchert HH, Pruss A, Roth WK, Bohnen HR, et al. (2010) Current concepts for quality assured long-distance transport of temperature-sensitive red blood cell concentrates. Vox Sang 99(1): 44–53. [DOI] [PubMed] [Google Scholar]
- 13. Kozma N, Speletz H, Reiter U, Lanzer G, Wagner T (2011) Impact of 13.56 MHz radio frequency identification systems on the quality of stored red blood cells. Transfusion 51(11): 2384–2390. [DOI] [PubMed] [Google Scholar]
- 14. Reiter U, Wagner T, Kozma N, Reiter G, Lanzer G (2011) Core and surface temperatures in a red-blood-cell unit during storage and transport. Vox Sang 101(1): 10–15. [DOI] [PubMed] [Google Scholar]
- 15. Reiter U, Reiter G, Wagner T, Kozma N, Roland J, et al. (2013) Four-dimensional temperature distributions in red blood cells withdrawn from storage and exposed to ambient temperature: a magnetic resonance thermometry study. Transfusion 53(1): 167–173. [DOI] [PubMed] [Google Scholar]
- 16. Hamill TR (1990) The 30-minute rule for reissuing blood: are we needlessly discarding units. Transfusion 30(1): 58–62. [DOI] [PubMed] [Google Scholar]
- 17. Sharma RR, Subramanian PG, Kumar S, Singh M, Sharma M, et al. (2004) Evaluation of storage conditions and bacterial prolifernationin blood components. Lab Med 35(10): 616–619. [Google Scholar]
- 18. Rieke V, Butts-Pauly K (2008) MR thermometry. J Magn Reson Imaging 27(2): 376–390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. B. Quesson B, de Zwart JA, Moonen CT (2000) Magnetic resonance temperature imaging for guidance of thermotherapy. J Magn Reson Imaging 12(4): 525–533. [DOI] [PubMed] [Google Scholar]
- 20. Kickhefel A, Roland J, Weiss C, Schick F (2010) Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences. Phys Med 26(4): 192–201. [DOI] [PubMed] [Google Scholar]
- 21.Holman JP (2001) Heat Transfer, 9th ed. Berkshire: McGraw-Hill.
- 22.Granryd E (2009) Heat Transfer: Collection of formulas and Tables of Thermal Properties, 8th ed. Stockholm: Dept. of Energy Technology, The Royal Institute of Technology.
- 23.Federal Law Gazette of the Republic of Austria (1999) Bundesgesetz über die Festlegung von Qualitäts- und Sicherheitsstandards für die Gewinnung, Verarbeitung, Lagerung und Verteilung von menschlichen Zellen und Geweben zur Verwendung beim Menschen; BGBl. I Nr. 44/1999, §14 (in its currently valid version). Austrian Federal Chancellery website. Available: http://www.ris.bka.gv.at/. Accessed 2012 Jan 12.
- 24.Siemens Simatic Sensors RFID systems SLG D12S special version. Data sheet, 08/2008, J31069-D0188-U001-A2–7618; 08/2008. Siemens Website. Available: http://cache.automation.siemens.com/dnl/zU/zU1ODgzOQAA_315 39613_HB/SLG_D12S_special_version_en.pdf. Accessed: 2012 January 12.
- 25. McDannold N (2005) Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: review of validation studies. Int J Hyperthermia 21(6): 533–546. [DOI] [PubMed] [Google Scholar]
- 26. Peters RD, Hinks RS, Henkelman RM (2000) Ex vivo tissue-type independence in proton-resonance frequency shift MR thermometry. Magn Reson Med 40(3): 454–459. [DOI] [PubMed] [Google Scholar]
- 27. Peters RD, Hinks RS, Henkelman RM (1999) Heat-source orientation and geometry dependence in proton-resonance frequency shift magnetic resonance thermometry. Magn Reson Med 41(5): 909–918. [DOI] [PubMed] [Google Scholar]
- 28. Gellermann J, Weihrauch M, Cho CH, Wlodarczyk W, Fähling H, et al. (2006) Comparison of MR-thermography and planning calculations in phantoms. Med Phys 33(10): 3912–3920. [DOI] [PubMed] [Google Scholar]
- 29. Grissom WA, Rieke V, Holbrook AB, Medan Y, Lustig M, et al. (2010) Hybrid referenceless and multibaseline subtraction MR thermometry for monitoring thermal therapies in moving organs. Med Phys 37(9): 5014–5026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. de Senneville BD, Roujol S, Jaïs P, Moonen CT, Herigault G, et al. (2012) Feasibility of fast MR-thermometry during cardiac radiofrequency ablation,”. NMR Biomed 25(4): 556–562. [DOI] [PubMed] [Google Scholar]
- 31. Wu T, Kendell KR, Felmlee JP, Lewis BD, Ehman RL (2000) Reliability of water proton chemical shift temperature calibration for focused ultrasound ablation therapy. Med Phys 27(1): 221–224. [DOI] [PubMed] [Google Scholar]
- 32. Thomas S, Wiltshire M, Hancock V, Fletcher S, McDonald C, et al. (2011) Core temperature changes in red blood cells. Transfusion 51(2): 442–443. [DOI] [PubMed] [Google Scholar]