Skip to main content
Annals of Laboratory Medicine logoLink to Annals of Laboratory Medicine
. 2013 Feb 21;33(2):121–124. doi: 10.3343/alm.2013.33.2.121

Comparison of 3 Automated Immunoassays for Detection of Anti-Hepatitis A Virus Immunoglobulin M in a Tertiary Care Hospital

Hyewon Park 1, Yu-joo Lee 1, Moon-Woo Seong 1, Do-Hoon Lee 2, Myoung Hee Park 1, Eun Young Song 1,
PMCID: PMC3589637  PMID: 23483019

Abstract

Three automated immunoassay kits for anti-Hepatitis A Virus (HAV) IgM-Architect, (Abbott Laboratories, USA), Elecsys (Roche Diagnostics, Germany), and ADVIA Centaur (Siemens Healthcare Diagnostics Inc., USA)-were compared. We included 178 consecutive samples, for which an anti-HAV IgM test was requested at Seoul National University Hospital from September 2009 to January 2010. Reviewing of medical records, reverse transcription (RT)-PCR for HAV RNA, or total anti-HAV assay were performed on 16 (9.0%) samples with discrepant results. The percent agreements (kappas) of the Architect and ADVIA Centaur, Architect and Elecsys, and ADVIA Centaur and Elecsys kits were 96.6% (0.91), 96.6% (0.92), and 97.8% (0.94), respectively. Eight out of 16 discrepant samples showed gray-zone values in Architect but were nonreactive in the others. Slightly earlier seroconversion was suspected in Elecsys. The 3 assays showed comparable performances with excellent agreements in a tertiary care hospital setting.

Keywords: Anti-hepatitis A virus immunoglobulin M, Immunoassay, Agreement


An anti-hepatitis A virus (HAV) IgM test is crucial to diagnose current HAV infection. Commercialized anti-HAV IgM chemiluminescence immunoassay has been widely used recently because of its significantly improved specificity and technical simplicity [1], although reports on performance are scarce [2, 3]. Performance of 3 anti-HAV IgM assays-Architect HAV Antibody (HAVAb)-IgM (Abbott Laboratories, Abbott Park, IL, USA), Elecsys Anti-HAV IgM (Roche Diagnostics, Mannheim, Germany), and ADVIA Centaur HAV IgM (Siemens Healthcare Diagnostics Inc., Deerfield, IL, USA)-was compared under routine conditions in the clinical laboratory of Seoul National University Hospital.

The study included 178 consecutive samples for immediate anti-HAV IgM testing using Architect HAVAb-IgM between September 2009 and January 2010. We collected the remaining sera as aliquots in 1.5 mL tubes immediately after the Architect HAVAB-IgM test and stored them at -80℃ until analysis. Elecsys and ADVIA Centaur HAV IgM were performed on the same day according to the manufacturers' instructions. For Architect, signal-to-cutoff (S/CO) values of 0.80-1.20 were considered gray-zone values. For ADVIA Centaur, an S/CO ≥0.80 and <1.20 was considered equivocal.

Medical records were reviewed, or reverse transcription (RT)-PCR for HAV and ADVIA Centaur total HAV were performed for 16 sera showing discrepant results. RNA was extracted using a Chemagic Viral DNA/RNA preparation kit (Chemagen, Baesweiler, Germany), and RT-PCR was performed using the AccuPower HAV Real-Time RT-PCR kit (Bioneer Corp., Daejeon, Korea). This study was approved by the Seoul National University Hospital Institutional Review Board (E-1110-046-381).

The agreements (kappas) between assays were calculated [4]. Correlations in S/CO values between assays were evaluated by a Spearman's test, excluding those results exceeding the measurable range using SPSS for Windows (version 12.0; SPSS Inc., Chicago, IL, USA).

Among 178 samples, 45 (25.3%) were reactive and 117 (65.7%) were nonreactive for all 3 kits. When the gray-zone results of Architect and ADVIA Centaur were interpreted as nonreactive, the percent agreements (kappas) between Architect and ADVIA Centaur, Architect and Elecsys, and ADVIA Centaur and Elecsys were 96.6% (0.91), 96.6% (0.92), and 97.8% (0.94), respectively. Among the 16 (9.0%) discrepant sera, 8 (case 1-8, Table 1) showed gray-zone values with Architect, but they were nonreactive with ADVIA Centaur and Elecsys. The negative anti-HAV IgM follow-up tests indicated that cases 1 and 2 were less likely to have HAV infection. For cases 3-8, HAV infection could not be ruled out from additional test results (HAV RT-PCR, negative; total anti-HAV, reactive). Case 9 (Architect, reactive; others, nonreactive) and Case 10 (ADVIA Centaur, reactive; others, nonreactive) were also less likely to have HAV infection considering the negative HAV RT-PCR, although very high levels of AST and ALT were seen.

Table 1.

Clinical characteristics of cases with discrepant results among Architect, ADVIA Centaur, and Elecsys Anti-HAV IgM assays (N=16)

graphic file with name alm-33-121-i001.jpg

*For Architect HAVAb-IgM, specimens with signal-to-cutoff (S/CO) values 0.80-1.20 were considered gray-zone. For ADVIA Centaur HAV IgM, S/CO values ≥0.80 and <1.20 were considered equivocal.

Abbreviations: T.bil, total bilirubin; D.bil, direct bilirubin; F/U, follow up; HAV, hepatitis A virus; N, nonreactive or negative; G, gray zone; R, reactive; P, positive; NT, not tested; S/CO, signal-to-cutoff; RT-PCR, reverse transcription-PCR.

Cases 11 and 12, confirmed as HAV+ (positive RT-PCR), were nonreactive with ADVIA Centaur but reactive with Elecsys. Cases 13 and 14, confirmed as HAV+ from reactive results with higher S/CO values of follow-up anti-HAV IgM tests in all 3 assays, showed gray-zone results with Architect and were reactive with Elecsys. Case 13 was nonreactive with ADVIA Centaur.

Cases 15 and 16, with infection history (7 and 8 months ago, respectively), (reactive anti-HAV IgM and clinical course consistent with HAV infection) were reactive with ADVIA Centaur and Elecsys and nonreactive and in the gray-zone with Architect, respectively.

Although, these assays were not quantitative, their S/CO values were moderately correlated with each other. Spearman's correlation coefficient (r) between Architect and the ADVIA Centaur HAV IgM was 0.757 (P<0.001); Architect and Elecsys, 0.732 (P<0.001); and Elecsys and ADVIA Centaur, 0.776 (P<0.001) (Fig. 1).

Fig. 1.

Fig. 1

Correlations of signal-to-cutoff (S/CO) values among Architect HAVAb-IgM, ADVIA Centaur HAV IgM, and Elecsys Anti-HAV IgM assays. (A) Scatter plot of S/CO values of Architect HAVAb-IgM and ADVIA Centaur HAV IgM assays, (B) Architect HAVAb-IgM and Elecsys Anti-HAV IgM assays, and (C) ADVIA Centaur HAV IgM and Elecsys Anti-HAV IgM assays.

Abbreviation: HAV, hepatitis A virus.

Here, 3 kits showed excellent overall agreement (kappas: 0.91-0.94) when the gray-zone values of Architect were considered nonreactive (ADVIA Centaur showed no equivocal results). Architect showed gray-zone results in 12 samples: HAV infections, 4; less-likely infections, 2; uncertain for infection, 6. The agreement was slightly lower (kappa values: Architect and ADVIA Centaur, 0.81; Architect and Elecsys, 0.87; data not shown) when the gray-zone values of Architect were considered reactive.

ELISAs can exhibit false-reactive results in various conditions, including autoimmune diseases or renal failure [5]. Rheumatoid factor or heterophilic antibodies can also interfere with immunoassay results [6, 7]. Nonspecific binding of serum IgM to the microparticle bead induces false reactivity in the Liaison system adopting chemiluminescence immunoassay, in the absence of rheumatoid factor or paraprotein; the use of chemical-blocking reagents eliminated this problem [8]. The Architect system adopts a different assay principle (direct coating of HAV antigens on a microparticle bead) from that of the other assays (using streptavidin-coated microparticles and biotinylated mouse anti-human IgM antibodies). Further investigations are needed to determine if gray-zone results, more frequently observed with Architect, could be partially explained by the nonspecific adsorption of proteins to the microparticle bead.

In cases 11-14, in the early phase of HAV infection, the ADVIA Centaur and Architect showed slightly later seroconversions compared to the Elecsys. Two cases with history of HAV infection (~7-8 months ago) were reactive with ADVIA Centaur and Elecsys with low S/CO values (1.09-1.66), whereas Architect showed nonreactive in one sample and gray-zone in another, suggesting a slight difference in the sensitivity for the detection of decreasing anti-HAV IgM in patients who had recovered from previous HAV infection.

Although all 3 kits are not quantitative tests, the S/CO values showed moderate correlations among them. For samples from patients with resolving HAV infection, S/CO values were low (1.09-1.66), suggesting very low anti-HAV IgM levels. Further development of quantitative tests for anti-HAV IgM may be helpful in patients showing atypical disease courses during HAV infection or HAV reactivation after transplantation [9].

In conclusion, 3 automated immunoassay kits showed comparable performances, with excellent overall agreement among them when performed on samples submitted to a tertiary care hospital and can be successfully applied in clinical laboratory practice.

Footnotes

No potential conflicts of interest relevant to this article were reported.

References

  • 1.Dufour DR, Talastas M, Fernandez MD, Hwang JH, Kim JW, Kim NY, et al. Chemiluminescence assay improves specificity of hepatitis C antibody detection. Clin Chem. 2003;49:940–944. doi: 10.1373/49.6.940. [DOI] [PubMed] [Google Scholar]
  • 2.Hess G, Faatz E, Melchior W, Bayer H. Analysis of immunoassays to detect antibodies to hepatitis A virus (anti-HAV) and anti-HAV immunoglobulin M. J Virol Methods. 1995;51:221–228. doi: 10.1016/0166-0934(94)00108-s. [DOI] [PubMed] [Google Scholar]
  • 3.Wiedmann M, Boehm S, Schumacher W, Swysen C, Zauke M. Evaluation of three commercial assays for the detection of hepatitis a virus. Eur J Clin Microbiol Infect Dis. 2003;22:129–130. doi: 10.1007/s10096-002-0861-7. [DOI] [PubMed] [Google Scholar]
  • 4.Clinical and Laboratory Standards Institute. CLSI document, GP29-A. 2nd ed. Wayne, PA: Clinical and Laboratory Standards Institute; 2008. Assessment of laboratory tests when proficiency testing is not available; Approved guideline. [Google Scholar]
  • 5.Mylonakis E, Paliou M, Lally M, Flanigan TP, Rich JD. Laboratory testing for infection with the human immunodeficiency virus: established and novel approaches. Am J Med. 2000;109:568–576. doi: 10.1016/s0002-9343(00)00583-0. [DOI] [PubMed] [Google Scholar]
  • 6.Cavalier E, Carlisi A, Chapelle JP, Delanaye P. False positive PTH results: an easy strategy to test and detect analytical interferences in routine practice. Clin Chim Acta. 2008;387:150–152. doi: 10.1016/j.cca.2007.08.019. [DOI] [PubMed] [Google Scholar]
  • 7.Kellar KL, Kalwar RR, Dubois KA, Crouse D, Chafin WD, Kane BE. Multiplexed fluorescent bead-based immunoassays for quantitation of human cytokines in serum and culture supernatants. Cytometry. 2001;45:27–36. doi: 10.1002/1097-0320(20010901)45:1<27::aid-cyto1141>3.0.co;2-i. [DOI] [PubMed] [Google Scholar]
  • 8.Berth M, Bosmans E. Prevention of assay interference in infectious-disease serology tests done on the liaison platform. Clin Vaccine Immunol. 2008;15:891–892. doi: 10.1128/CVI.00012-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Eisenbach C, Longerich T, Fickenscher H, Schalasta G, Stremmel W, Encke J. Recurrence of clinically significant hepatitis A following liver transplantation for fulminant hepatitis A. J Clin Virol. 2006;35:109–112. doi: 10.1016/j.jcv.2005.08.005. [DOI] [PubMed] [Google Scholar]

Articles from Annals of Laboratory Medicine are provided here courtesy of Korean Society for Laboratory Medicine

RESOURCES