Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2013 Mar 13;8(3):e58147. doi: 10.1371/journal.pone.0058147

Compound Ex Vivo and In Silico Method for Hemodynamic Analysis of Stented Arteries

Farhad Rikhtegar 1, Fernando Pacheco 2, Christophe Wyss 3, Kathryn S Stok 4, Heng Ge 3, Ryan J Choo 4, Aldo Ferrari 1, Dimos Poulikakos 1, Ralph Müller 4, Vartan Kurtcuoglu 1,5,*
Editor: Christian Schulz6
PMCID: PMC3596389  PMID: 23516442

Abstract

Hemodynamic factors such as low wall shear stress have been shown to influence endothelial healing and atherogenesis in stent-free vessels. However, in stented vessels, a reliable quantitative analysis of such relations has not been possible due to the lack of a suitable method for the accurate acquisition of blood flow. The objective of this work was to develop a method for the precise reconstruction of hemodynamics and quantification of wall shear stress in stented vessels. We have developed such a method that can be applied to vessels stented in or ex vivo and processed ex vivo. Here we stented the coronary arteries of ex vivo porcine hearts, performed vascular corrosion casting, acquired the vessel geometry using micro-computed tomography and reconstructed blood flow and shear stress using computational fluid dynamics. The method yields accurate local flow information through anatomic fidelity, capturing in detail the stent geometry, arterial tissue prolapse, radial and axial arterial deformation as well as strut malapposition. This novel compound method may serve as a unique tool for spatially resolved analysis of the relationship between hemodynamic factors and vascular biology. It can further be employed to optimize stent design and stenting strategies.

Introduction

Atherosclerosis is the leading cause of death in most developed countries, predominantly as a result of myocardial infarction due to coronary heart disease (CHD). Percutaneous coronary intervention (PCI) that generally involves the placement of a stent has become the primary mode of CHD treatment over the past 20 years [1].

CHD is characterized by progressive atherosclerotic plaques that narrow (stenose) the coronary artery lumen, thereby reducing blood flow to the myocardium. PCI is used to expand the lumen with a balloon catheter and to keep it open with a wire scaffold (stent).

Despite stent placement, incidence of renewed stenosis of the vessel can be as high as 30% [2], [3], most commonly due to neointimal hyperplasia (NIH) [4]. NIH is linked to both the injury or destruction of the endothelium [5], [6] and the loss of smooth muscle cells (SMC) due to stretching of the intima during stent deployment [7]. Expedient endothelial regeneration reduces NIH [8], [9], and endothelial regeneration itself is influenced by blood flow. Similarly, the distribution of atherosclerotic plaques is strongly influenced by the local wall shear stress (WSS) distribution [10], [11]. As WSS is proportional to the gradient of blood flow velocity at the endothelium, precise knowledge of hemodynamics is necessary to derive it. The required level of precision can currently not be achieved clinically using phase-contrast magnetic resonance imaging (PC-MRI) [12], Doppler ultrasound, or other flow measurement techniques [13−15]. For this reason, flow field reconstruction using computational fluid dynamics (CFD) based on medical image data has become the state-of-the-art for determining WSS in stent-free vessels [11], [16][19].

A prerequisite for deriving WSS in stented arteries using CFD is the precise definition of the stent geometry with feature sizes of the order of tens of microns. However, no current clinical imaging modality can yield a three-dimensional (3D) representation of a deployed stent with sufficient accuracy for reliable CFD calculations. Computed tomography (CT) [20, MRI [21], intravascular ultrasound [11], [18], [22] and digital angiography [23] do not offer sufficient spatial resolution to capture individual stent struts in detail, and optical coherence tomography is limited by the opacity of the struts to the emitted light.

To circumvent these limitations, hybrid approaches have been developed where the stent-free artery is acquired via CT, digital angiography or MRI, and a virtual stent is placed in the generated digital dataset prior to the calculation of WSS [24][31]. Other methods omit in vivo imaging completely [32][39], for example by performing image acquisition on explanted stented arteries using micro-computed tomography (µCT) [32], [33], or by placing stents in artificial artery models and then proceeding with µCT [33][39].

The individual methods have their respective strengths and weaknesses. While some optimize processing speed and cost by approximating the deployed stent in a computer aided design (CAD) environment [29], others opt for slower, more expensive but also more accurate approaches based on computational structural mechanics simulations of stent deployment [27]. Further methods give preference to actual rather than virtual stent deployment, thereby sacrificing the flexibility of computational techniques for the possibility to capture the expanded stent geometry with higher fidelity when real arteries are used [32], or for the possibility to investigate complex stenting procedures such as double stenting of main vessel and side branch [35]. Some approaches, finally, do not consider derivation of WSS [40][42].

In situations where destructive processing of a stented artery is not an issue, combination of vascular corrosion casting (VCC) with µCT and CFD may yield detailed reconstruction of WSS distribution. VCC, originally developed for producing anatomical specimens, can generate negatives of entire vascular trees with sub-micron accuracy, while µCT can be used to digitize the VCC cast with sufficient resolution to capture stent struts in detail. LaDisa and coworkers were the first to combine these methods by stenting rabbit iliac arteries in vivo, sacrificing the animals after two or three weeks, casting the artery lumen, macerating the surrounding tissue and removing the stent with a sanding pad before acquiring the lumen negative by µCT [43]. However, individual stent struts could not be resolved with their technique.

Here we present a method that combines VCC, µCT and CFD to a platform for the precise calculation of WSS in stented arteries. This method is able to accurately resolve both the macroscopic arrangement of stented vessels as well as the microscopic structure of the stent struts.

Methods

An expanded methods section is provided in the supporting information online (Appendix S1).

Heart Preparation and Stenting

Porcine hearts obtained with permission from the local slaughterhouse (Metzegerei Angst AG, Zurich, Switzerland) were cannulated in preparation for stenting of the LCA. Absorbable metal scaffolds of 10 mm length and 3 mm diameter (Biotronik AG, Bülach, Switzerland) were placed by an interventional cardiologist under angiographic guidance using the manufacturer-specified inflation pressure of 12 bar.

Vascular Corrosion Casting

A 1∶0.1225 by weight mixture of low-shrinkage epoxy-based Biodur E 20 (EP20–EP22) resin (Biodur Products GmbH, Heidelberg, Germany) and iodine-saturated methyl ethyl ketone solvent was used as a radio-opaque casting material [44]. The resin was injected into the stented coronary vascular tree under physiological pressure of 90 mmHg (120 mbar) [45]. After a setting period of 36 hours, the heart was macerated for 12 h at 55°C in a 7.5% w/v solution of potassium hydroxide.

µCT Imaging of Stented Casts

The stented coronary arteries were first imaged using micro-computed tomography (µCT 80, Scanco Medical AG, Brüttisellen, Switzerland) with an isotropic voxel size of 74 µm (energy 70 kVp, integration time 300 ms, tube current 114 µA, and two times frame averaging) to provide the image data of the overall coronary arterial tree geometry. Following this, the stented sections were removed from the artery tree and re-scanned (µCT 40, Scanco) with an isotropic voxel size of 6 µm (energy 70 kVp, integration time 300 ms, tube current 114 µA, and two times frame averaging) in order to resolve individual stent struts (Figure S1).

Image Processing

A constrained 3D Gauss filter was used to partly suppress noise in the raw µCT volumes (σ = 1.2, s = 1.0). The coronary artery lumen was segmented from both µCT datasets independently using a semi-automatic, intensity-based approach in Avizo 6.2 (Visualization Sciences Group SAS, Merignac, France). The resulting 3D geometries were registered and merged in Geomagic Studio 12 (Geomagic, Inc., Morrisville, NC, USA) to where the high resolution geometry represented the stented artery region and the lower resolution one the remainder of the arterial tree (Figure S2). The merged geometry was exported in STL format for subsequent computational grid generation.

CFD Calculations

A computational grid consisting of approximately 48 million tetrahedral elements was generated in the merged geometry in ANSYS ICEM CFD (ANSYS, Inc., Canonsburg, PA, USA); see Figure S3. To calculate flow velocity, pressure and WSS distribution, transient and steady-state computational flow analysis was carried out with the finite volume CFD code ANSYS CFX using a Newtonian fluid model with constant density of 1050 kg/m3 and dynamic viscosity of 0.0035 Pa.s [46]. Boundary conditions were chosen as follows: No slip at the vessel wall, blood inflow rate of 0.95 mL/s at the ostium for the steady-state calculations [16] and time-dependent flow rate for the transient case according to [17]. Zero relative pressure was set at the outlet with the largest diameter and outflow rates at the remaining outlets were determined according to Murray’s law [47]; see Figure S4. For the transient simulations, two cardiac cycles were calculated using a time step size of 0.01 seconds, but only the data of the second cycle were evaluated to obtain results independent of the initial conditions. With residual reduction to 10−8 of the initial value as convergence criterion, the steady-state calculations took approximately two hours on 32 AMD Opteron 6174 processor cores. The transient computations required 25 minutes per time step with a convergence criterion of 10−6 at each point in time. Grid independence studies were performed (Appendix S1).

Results

In the following we will show on ex vivo porcine hearts that the compound method presented herein ensures anatomic fidelity, capturing arterial tissue prolapse, radial and axial arterial deformation as well as stent malapposition. We will further show how this method yields detailed blood flow fields and wall shear stress maps in stented coronary arteries (see Video S1), noting that it can also be used in ex vivo human arteries with minimal change to the protocol.

Arterial Tissue Prolapse

The commonly used stent-to-artery diameter ratio of 1.1–1.2 [48] in coronary arteries can lead in conjunction with the elasticity of the vessel wall to tissue prolapse [49], [50]. We use the term ‘prolapse’ in accordance with the biomedical engineering literature to refer to any degree of tissue protrusion between the stent struts, noting that in the medical literature it is generally associated with protrusion of plaque or thrombus beyond the inner stent surface. Prolapse affects local hemodynamics, thereby altering WSS. Moreover, tissue prolapse is associated with increased incidence of acute and subacute thrombosis [50].

As illustrated in Fig. 1, the method captures arterial tissue prolapse. Panel A shows a representative µCT scan section of a stented porcine left coronary artery corrosion cast. Both the cured, contrast-enhanced resin in place of the artery lumen as well as cross-sections of stent struts are clearly visible. The dark areas between resin and struts are due to gas formed in the VCC process through the interaction of stent with resin solution. These areas are merged with the lumen representation in the image segmentation step. Fig. 1B shows the reconstructed surface of a stented porcine left coronary artery (LCA) lumen negative that was acquired using µCT of a corrosion cast. The white arrows between the imprints of the stent struts point to regions of prolapse. These occur most markedly in areas without strut connectors, indicating that arterial tissue prolapse is dependent on stent design.

Figure 1. Arterial tissue prolapse between stent struts.

Figure 1

(A) Representative micro-computed tomography (µCT) scan section of stented porcine left coronary artery corrosion cast. (B) Reconstructed surface of arterial lumen negative obtained by µCT of a corrosion cast. The white arrows point to prolapsed regions. (C) Wall shear stress (WSS) distribution in the same region shown in B. Higher WSS is evident in prolapsed regions compared to regions without prolapse. Blood flow is from left to right. (D) Velocity contour plots in prolapsed (top) and prolapse-free inter-strut sections. Regions of low velocity are evident near stent struts in both cases, with larger low-velocity regions in the prolapse-free segment. This segment also shows lower near-wall velocity.

Fig. 1C shows the corresponding WSS distribution. Higher WSS is evident in regions of prolapse owing to higher velocities near the wall compared to prolapse-free sections as illustrated by the velocity contours in Fig. 1D. Regions of low velocity are recognizable near the struts in both cases, but the size of these regions is clearly different for the two conditions. It is thus evident that arterial tissue prolapse influences hemodynamics and therewith the local WSS distribution with which predictions of plaque development can be made.

Radial Wall Deformation

Histological studies show that stent deployment changes the circular cross-sectional shape of the artery [51]. This method captures such deformations: Fig. 2A depicts a representative cross-section through a stented coronary artery from a µCT image of the lumen. The dotted circle serves as a reference for the local deformation caused by the struts. These deformations impact hemodynamics and WSS. Fig. 2B demonstrates the increase in lumen diameter from the stent-free section of the artery to its stented part, which can lead to a local decrease in both WSS and velocity. This is seen in detail in Fig. 2C: To ensure mass conservation, blood has to accelerate from the stented part of the artery with a larger diameter to the stent-free section with smaller diameter.

Figure 2. Figure 2. Radial wall deformation of stented artery.

Figure 2

(A) Representative micro-computed tomography cross-section of a stented porcine left coronary artery corrosion cast. The dotted circle shows the nominal circular cross-section (B) Radial arterial enlargement caused by stenting. The arrows indicate arterial diameter in the stented (left) and stent-free regions (C) Velocity vectors in the mid-longitudinal section plane of the stented artery. The velocity profile along the vessel centerline from Point I to Point II is shown in (D), where the increase in velocity in the stent-free section is visible. The vertical dashed lines mark the end of the stent and the axial locations of Points I and II shown in panel C.

Axial Arterial Deformation

Stenting results in substantial axial deformation, causing a straightening of the artery [52], which affects hemodynamics and WSS distribution substantially.

Fig. 3 shows that the herein presented compound method captures axial arterial deformation. The dashed line in Fig. 3A approximates the centerline of the stent-free artery. The solid line illustrates the change of the centerline shape in the stented region. Evaluated from left to right, the sudden straightening of the artery in the stented segment and the abrupt return in curvature to that of the stent-free region are evident. Fig. 3B shows that this leads to an extended area of low WSS at the outer vessel wall. Without the stent, the inner arterial wall would be the main location of low WSS, atherosclerotic plaque formation and neointimal hyperplasia (NIH) [22], [46]. Consequently, axial arterial deformation due to stent placement has to be accounted for in WSS derivations.

Figure 3. Axial arterial deformation due to stenting.

Figure 3

(A) Visualization of arterial centerline change in a stented section. The solid line shows the axis of the stent, while the dashed line approximates the centerline of the stent-free artery. (B) Wall shear stress (WSS) distribution in the same stented artery. An extended area of low WSS is seen immediately downstream of the stent at the outer artery wall due to the change in curvature.

Stent Malapposition

Stent malapposition alters in-stent hemodynamics, causes low WSS distally and is hypothesized to be a major factor in thrombosis [53]. The method presented here captures malapposed struts: Fig. 4A and 4B show the arterial lumen surface and WSS in malapposed and fully apposed stented regions, respectively. The malapposed strut shown in Figs. 4C and 4D causes tunneling of blood flow between the strut and the endothelium, leading to high WSS and perturbation of the local flow field. Such perturbed flow (see Video S2) is associated with increased risk of thrombosis [54].

Figure 4. Stent malapposition and its effect on local hemodynamics.

Figure 4

(A) Imprint of malapposed stent end section (arrow) in artery lumen negative (top) and corresponding wall shear stress (WSS) distribution (bottom). Higher WSS can be observed in the vicinity of the malapposed strut due to flow tunneling compared to (B), where a similar fully apposed stent end section is shown. (C) Velocity contour in axial cross-section of the stented artery near a malapposed strut. Changes in velocity and division of blood flow can be seen. (D) Velocity vector plot in the vicinity of the malapposed strut demonstrates the presence of vortices. These influence WSS distribution and may lead to thrombosis.

Reconstruction of Hemodynamic State

Both the local geometry at the vessel wall, as well as the large scale arterial anatomy, influence WSS distribution and can be accounted for with this method. This can be seen in Fig. 5 where regions of low, atheroprone shear stress are present at bifurcations and nearby individual stent struts. Interestingly, low WSS is not only present adjacent to struts arranged perpendicular to the flow direction, but also occurs in the vicinity of inter-strut connectors arranged parallel to the artery’s longitudinal axis.

Figure 5. Wall shear stress distribution in a porcine left coronary artery with two stents.

Figure 5

The bottom inset shows a magnified view of the stented segments. Wall shear stress (WSS) below 0.5 Pa is reported to correlate with sites of intimal thickening [51]. Such low WSS can be seen here to occur mainly in the vicinity of stent struts and at bifurcations. The left inset shows low and high wall shear stress regions occurring, respectively, at the outer and inner walls of the bifurcation (arrows).

Figure 6 shows the distribution of oscillatory shear index (OSI), which is a measure of temporal WSS change (see Appendix S1). High values of OSI have been shown to correlate with atheroprone regions of the vessel [17]. Increased values of OSI are seen here near strut junctions, which is in agreement with earlier observations [55].

Figure 6. Oscillatory shear index (OSI) distribution in a porcine left coronary artery with two stents.

Figure 6

The inset shows a magnified view of part of the second stented segment. Elevated values of OSI have been reported to correlate with atheroprone vessel regions [17]. Areas of increased OSI are visible near strut junctions. They contain small focal spots that reach values close close to the maximum of 0.5.

The flow structures that lead to low WSS are illustrated in Fig. 7. Panel A shows velocity contours and streamlines projected axially onto a cross-sectional plane immediately upstream of the stent. The velocity profile is near parabolic, and there are no recirculation areas discernible. Entering the stented vessel region in downstream direction (Panel B), flow disturbances begin to develop and quickly lead to the generation of recirculation zones (Panels C to I). In addition, the parabolic velocity profile is altered due to changes in shape of the arterial cross-section, as well as due to the stent struts’ influence on the near-wall flow. This can be seen in the offset of the velocity contours in Fig. 8.

Figure 7. Velocity contours and secondary flow around individual stent struts in a porcine left coronary artery.

Figure 7

Top: Overview and close-up of reconstructed surface of arterial lumen negative obtained by µCT of a corrosion cast. The labels A to I indicate the location of the cross-sections shown in the bottom panels. Bottom: Velocity contour plots at cross-sections A to I. To visualize secondary flow structures, streamlines are derived from velocity vectors projected onto the respective cross-section.

Figure 8. Velocity profiles in a porcine left coronary artery with two stents.

Figure 8

Results are shown for cross-sections upstream of the distal stent (A and B), within (C, D E and F) and downstream of the stent (G and H). Top: Velocity contour plots. Bottom: Velocity projections onto axial planes. The vertical axes are normalized to a common diameter.

Axial arterial shape change as a result of the stent placement further affect the velocity profile as illustrated in Fig. 8. The top panel shows an artery segment with two stents. Entering the second stent from the upstream direction at cross-section A, the location of peak velocity is deflected from the vessel centerline in plane D due to the increased curvature of the vessel at the stent edge. Effects of the stent strut as described above maintain the off-center profile throughout the length of the stent (planes E and F).

The combined effect of local flow disturbance by the stent struts, changes in cross-sectional area and flow deflection as a result of axial arterial deformation is quantified in Fig. 9. There the distribution of low (<0.5 Pa), intermediate (0.5 to 2.5 Pa) and high (>2.5 Pa) WSS [11], [56], [57] is shown relative to the surface area of selected arterial sections in a vessel with two stents. These sections correspond to the area immediately upstream of the first stent (labeled ‘proximal’ in the top panel), the first and second stent (‘proximal stent’, ‘distal stent’), and the region after the second stent (‘distal’). In the stented sections, more than 40% of the wall surface area is exposed to low shear stress. In comparison, virtually no low WSS is present in the sections upstream of the first stent and downstream of the second stent. The high WSS observed after the second stent is due to the narrowing of the artery in that area.

Figure 9. Distribution of relative vessel wall area exposed to different levels of shear stress.

Figure 9

Results are shown in the bottom panel as percentage of the respective segment’s surface area in a porcine left coronary artery with two stents. Low wall shear stress: <0.5 Pa. Moderate: 0.5<WSS<2.5). High: >2.5 Pa. Top: Corresponding reconstructed surface of the arterial lumen negative.

Discussion

As a result of the small feature sizes of stents and limited resolution of clinical imaging modalities, alternative methods have to be used to obtain the lumen geometry of stented arteries for calculation of WSS. It is accepted that wall shear stress affects vascular biology, influencing atherosclerotic plaque development and NIH. Here we have presented a method with which WSS can be determined by combining VCC, µCT and CFD. This method can be used in sacrificed animals or post mortem in humans after stenting has been performed either in vivo or ex vivo.

While similar approaches have been used before, this compound method removes some of the prior limitations: LaDisa et al. used VCC and µCT [43] with subsequent CFD modeling, but could not resolve individual stent struts. Morlacchi and co-workers stented pigs in vivo, excised the stented artery segments, embedded these in resin, acquired the deployed stent geometry with µCT and, in addition, performed histological analysis [32]. However, they could not acquire the lumen geometry, which had to be approximated for subsequent CFD analysis.

Since the presented method allows for processing stents deployed in vivo, it is expected to yield more accurate reconstruction of the in vivo WSS distribution than other methods that rely on µCT, but do not allow for in vivo stent deployment. Benndorf and co-workers observed marked differences between the WSS field obtained in a stented PTFE tube and an ex vivo stented preserved arterial segment [33]. This indicates that the choice of arterial wall representation can influence the derived hemodynamic parameters substantially. While one should expect that the preserved arterial segment mimics in vivo conditions better than the PTFE tube, the segment still showed artifacts in the form of circumferential creasing. Such creases are not observed in vivo and are presumably a result of the preservation process.

For comparison with virtual stent placement approaches, one has to consider whether these are capable of reproducing critical features observed in vivo. Our results show that neglecting tissue prolapse or stent malapposition may alter the reconstructed WSS field substantially. This is in accordance with the findings of Benndorf [33]. We further show that neglecting radial or axial arterial deformation will change WSS distribution as observed before by LaDisa [25], [26] and Murphy [46]. Virtual stent placement methods that cannot reproduce these critical features may have cost and processing time advantages, but are limited in the accuracy of WSS reconstruction. Assessment of virtual methods that do take into account critical features is less trivial. On the one hand, these methods are the first choice for predicting stent deployment and WSS distribution in live humans or animals. On the other hand, they idealize arterial wall mechanics, which is especially then critical when diseased arteries are investigated, as the presence of plaques and calcifications may change the wall properties substantially in an anisotropic and heterogeneous manner. Since the method introduced here can rely on the true geometry of vessels stented in vivo, we expect it to yield more accurate WSS readings than approaches based on virtual stent deployment. Of course, a suitable study is necessary to confirm this.

The presented method is based on three consecutively applied tools. These can each be optimized, but the interface between the methods has to be considered in the process. The optimal VCC resin has no shrinkage during curing and high stiffness thereafter. Biodur E 20 used here is a low shrinkage resin that produces rigid casts, maintaining the 3D configuration of the artery during maceration. The relative deviation of the cast lumen volume from the original one is less than 2%. There are other resins such as PU4ii that show lower shrinkage, but remain pliable after curing [58].

The resin should ideally have an X-ray attenuation behavior comparable to that of the stent to allow for optimal µCT imaging. Most VCC resins, however, have a very low attenuation coefficient, leading to low signal-to-noise ratio [59]. To increase resin opacity, we saturated the resin solvent, methyl ethyl ketone, with iodine [44]. This decreases viscosity, prolonging resin hardening time, and also increases shrinkage slightly. Alternatively, take-up of an attenuating compound after resin curing could be used, e.g. by bathing the cast in an aqueous solution of osmium tetroxide [59], [60]. However, this process is time consuming and only yields low penetration depths. In addition, OsO4 is very toxic.

The choice of scanner and acquisition settings has a great influence on the final results. The small size of the stent struts necessitates a high scan resolution, which in most scanners excludes the use of larger samples such as complete coronary artery tree casts. To circumvent this problem, we acquired the overall geometry and the stented section with two different scanners, introducing a time consuming registration step to merge the data. While a single step acquisition at high resolution would allow for a more automated work flow, it would result in very large datasets of over 30 gigabytes that are difficult to handle. In addition, scan time would increase by at least a factor of four, and scan cost would go up accordingly.

The quality of artery lumen reconstruction is the main determining factor for the accuracy of the WSS distribution calculations. Of similarly high importance is the choice of boundary conditions. Here we used a generic volumetric inflow rate or temporal profile, one zero pressure outlet and Murray’s law to set flow rates at the remaining outlets [47]. More accurate results could be obtained by applying a subject-specific inflow rate, which requires either phase contrast magnetic resonance imaging or invasive intravascular Doppler ultrasound measurements in vivo [21], [61], [62]. Boundary conditions at the outlets would ideally be determined by in vivo pressure or flow measurements as well. However, measurements in the distal artery segments are less accurate than in the larger parent vessels. Also, with increasing number of outlets, this approach becomes impractical. Applying Murray’s law, empirical variations thereof [63] or lower order models of the downstream vasculature to determine the boundary conditions at the outlets appear reasonable [29], [64].

Flow disturbances introduced by the stent lead to areas of low shear rate where blood displays non-Newtonian behavior. Consequently, a shear rate dependent rheology model should be used for best results. However, it is difficult to predict and to test which of the many existing models will give the most accurate results in stented arteries [23].

The main limitation of the presented method is that it entails destructive procedures. It can thus not be applied to live humans or to animals that should be kept alive. However, this does not exclude stenting in vivo and further processing ex vivo. When the method is applied to live animals, these can be treated without modifications to common protocols up to the point of sacrifice, after which VCC is started. If no in vivo acquisition of blood flow rates is foreseen in the original protocol, it should be added to derive realistic boundary conditions.

The method can also be used for post-mortem investigation of stented, atherosclerotic arteries in humans for research purposes. Processing of diseased vessels may require adaptation of individual process parameters. In particular, the possible entrapment of plaques and thrombi in the VCC resin and the presence of calcifications may render the image segmentation process more challenging [65], requiring changes in the concentration of the contrast agent and modification of µCT parameters. In addition, resin viscosity may need to be reduced if high grade stenoses are present [66]. Further studies are thus required to validate the performance of the method in diseased vessels.

Again due to the destructive nature of the method, histologic analysis and concurrent WSS derivation on the same artery are not possible. Consequently, a larger sample size is needed to statistically correlate vascular biology with WSS or other hemodynamic parameters. This adds to the comparably high cost of the method which derives from the large number of steps involved that each requires a high level of expertise. Next to in vivo stenting, the main cost factors are the high resolution µCT imaging and processing of the therewith associated large datasets.

Finally, unwanted interaction between the resin components and the stent may occur. In the current study such interaction resulted in gas bubbles, which were dealt with in the image segmentation process. It cannot be excluded that with other stents resin interaction may become a limitation.

In conclusion, the method presented herein constitutes a unique tool for accessing WSS in stented arteries. It can be employed to study the effects of hemodynamics on vascular biology, to develop stenting strategies that optimize hemodynamics and to design new stents that minimize regions of NIH promoting WSS.

Supporting Information

Figure S1

Cross-sections of stent struts acquired at (A) 6 µm, (B) 8 µm, (C) 10 µm and (D) 12 µm scan resolution. Using the manufacturer’s production specifications as reference, the 6 µm resolution scan was found to capture the stent geometry with sufficient accuracy.

(TIF)

Figure S2

Registration of the high resolution surface of a stented artery section with the lower resolution surface of the whole arterial geometry. (A) Low resolution surface of the whole artery (B) High resolution surface of the stented region (C) Combined surface.

(TIF)

Figure S3

Tetrahedral mesh cross-section at a stented artery section. The inset shows the refined computational grid at the artery wall.

(TIF)

Figure S4

Illustration of the bifurcation mass-flow conditions based on Murray’s law as given by Equations S1 and S2.

(TIF)

Appendix S1

Expanded Methods.

(DOCX)

Video S1

Reconstruction of stented LCA section and visualization of blood flow field.

(MPG)

Video S2

Visualization of flow disturbance introduced by malapposed stent strut.

(MPG)

Acknowledgments

We would like to thank Mr. Axel Lang of the Institute of Anatomy, University of Zurich, for his help with the vascular corrosion casting.

Funding Statement

This work was partially funded by the Swiss Federal Commission for Technology and Innovation (CTI) through the EnOp project, grant 9921.1. No additional external funding was received for this study. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1. Venkitachalam L, Kip KE, Selzer F, Wilensky RL, Slater J, et al. (2008) Twenty-year evolution of Percutaneous coronary intervention and its impact on clinical outcomes - a report from the NHLBI-sponsored, multicenter 1985–86 PTCA and 1997–2006 Dynamic Registries. Circulation: Cardiovascular Interventions 2: 6–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, et al. (2003) Sirolimus-Eluting Stents versus Standard Stents in Patients with Stenosis in a Native Coronary Artery. New England Journal of Medicine 349: 1315–1323. [DOI] [PubMed] [Google Scholar]
  • 3. Elezi S, Kastrati A, Neumann F-J, Hadamitzky M, Dirschinger J, et al. (1998) Vessel Size and Long-Term Outcome After Coronary Stent Placement. Circulation 98: 1875–1880. [DOI] [PubMed] [Google Scholar]
  • 4. Stone PH, Coskun AU, Kinlay S, Clark ME, Sonka M, et al. (2003) Effect of Endothelial Shear Stress on the Progression of Coronary Artery Disease, Vascular Remodeling, and In-Stent Restenosis in Humans. Circulation 108: 438–444. [DOI] [PubMed] [Google Scholar]
  • 5. Hoffmann R, Mintz GS (2000) Coronary in-stent restenosis–predictors, treatment and prevention. European Heart Journal 21: 1739–1749. [DOI] [PubMed] [Google Scholar]
  • 6. Grewe PH, Deneke T, Machraoui A, Barmeyer J, Müller K-M (2000) Acute and chronic tissue response to coronary stent implantation: pathologic findings in human specimen. Journal of the American College of Cardiology 35: 157–163. [DOI] [PubMed] [Google Scholar]
  • 7. Fingerle J, Au YP, Clowes AW, Reidy MA (1990) Intimal lesion formation in rat carotid arteries after endothelial denudation in absence of medial injury. Arteriosclerosis, Thrombosis, and Vascular Biology 10: 1082–1087. [DOI] [PubMed] [Google Scholar]
  • 8. Asahara T, Bauters C, Pastore C, Kearney M, Rossow S, et al. (1995) Local delivery of vascular endothelial growth factor accelerates reendothelialization and attenuates intimal hyperplasia in balloon-injured rat carotid artery. Circulation 91: 2793–2801. [DOI] [PubMed] [Google Scholar]
  • 9. Steinmetz M, Nickenig G, Werner N (2010) Endothelial-Regenerating Cells An Expanding Universe. Hypertension 55: 593–599. [DOI] [PubMed] [Google Scholar]
  • 10. Glagov S, Zarins C, Giddens DP, Ku DN (1988) Hemodynamics and atherosclerosis. Insights and perspectives gained from studies of human arteries. Arch Pathol Lab Med 112: 1018–1031. [PubMed] [Google Scholar]
  • 11. Samady H, Eshtehardi P, McDaniel MC, Suo J, Dhawan SS, et al. (2011) Coronary Artery Wall Shear Stress Is Associated With Progression and Transformation of Atherosclerotic Plaque and Arterial Remodeling in Patients With Coronary Artery Disease/Clinical Perspective. Circulation 124: 779–788. [DOI] [PubMed] [Google Scholar]
  • 12. Hollnagel DI, Summers PE, Poulikakos D, Kollias SS (2009) Comparative velocity investigations in cerebral arteries and aneurysms: 3D phase-contrast MR angiography, laser Doppler velocimetry and computational fluid dynamics. NMR Biomed 22: 795–808. [DOI] [PubMed] [Google Scholar]
  • 13. Kaufmann PA, Camici PG (2005) Myocardial blood flow measurement by PET: technical aspects and clinical applications. J Nucl Med 46: 75–88. [PubMed] [Google Scholar]
  • 14. Doriot P, Dorsaz P, Dorsaz L, Chatelain P (2000) Accuracy of coronary flow measurements performed by means of Doppler wires. Ultrasound Med Biol 26: 221–228. [DOI] [PubMed] [Google Scholar]
  • 15. Fearon WF, Farouque HM, Balsam LB, Caffarelli AD, Cooke DT, et al. (2003) Comparison of coronary thermodilution and Doppler velocity for assessing coronary flow reserve. Circulation 108: 2198–2200. [DOI] [PubMed] [Google Scholar]
  • 16.Berne R.M., Levy M.N., (1986). Cardiovascular Physiology 5th ed. Mosby, St. Louis. 312 p.
  • 17. Rikhtegar F, Knight JA, Olgac U, Saur SC, Poulikakos D, et al. (2012) Choosing the optimal wall shear parameter for the prediction of plaque location–A patient-specific computational study in human left coronary arteries. Atherosclerosis 221: 432–437. [DOI] [PubMed] [Google Scholar]
  • 18. Chiastra C, Morlacchi S, Pereira S, Dubini G, Migliavacca F (2012) Computational fluid dynamics of stented coronary bifurcations studied with a hybrid discretization method. European Journal of Mechanics - B/Fluids 35: 76–84. [Google Scholar]
  • 19. Stone PH, Saito S, Takahashi S, Makita Y, Nakamura S, et al. (2012) Prediction of progression of coronary artery disease and clinical outcomes using vascular profiling of endothelial shear stress and arterial plaque characteristics: the PREDICTION Study. Circulation 126: 172–181. [DOI] [PubMed] [Google Scholar]
  • 20. Ehara M, Kawai M, Surmely J-F, Matsubara T, Terashima M, et al. (2007) Diagnostic Accuracy of Coronary In-Stent Restenosis Using 64-Slice Computed Tomography: Comparison With Invasive Coronary Angiography. Journal of the American College of Cardiology 49: 951–959. [DOI] [PubMed] [Google Scholar]
  • 21. Furber AP, Lethimonnier F, Le Jeune J-J, Balzer P, Jallet P, et al. (1999) Noninvasive assessment of the infarct-related coronary artery blood flow velocity using phase-contrast magnetic resonance imaging after coronary angioplasty. The American journal of cardiology 84: 24–30. [DOI] [PubMed] [Google Scholar]
  • 22. Wentzel JJ, Krams R, Schuurbiers JCH, Oomen JA, Kloet J, et al. (2001) Relationship Between Neointimal Thickness and Shear Stress After Wallstent Implantation in Human Coronary Arteries. Circulation 103: 1740–1745. [DOI] [PubMed] [Google Scholar]
  • 23. Yilmaz F, Gundogdu MY (2008) A critical review on blood flow in large arteries; relevance to blood rheology, viscosity models, and physiologic conditions. Korea-Australia Rheology Journal 20: 197–211. [Google Scholar]
  • 24.LaDisa JF, Jr., Olson LE, Hettrick DA, Warltier DC, Kersten JR, et al. (2005) Axial stent strut angle influences wall shear stress after stent implantation: analysis using 3D computational fluid dynamics models of stent foreshortening. Biomedical engineering online 4: 59. Available: http://www.biomedical-engineering-online.com/content/4/1/59. Accessed 21 January 2012. [DOI] [PMC free article] [PubMed]
  • 25. LaDisa JF Jr, Olson LE, Guler I, Hettrick DA, Kersten JR, et al. (2005) Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with time-dependent 3D computational fluid dynamics models. Journal of applied physiology 98: 947–957. [DOI] [PubMed] [Google Scholar]
  • 26.LaDisa JF, Jr., Olson LE, Douglas HA, Warltier DC, Kersten JR, et al. (2006) Alterations in regional vascular geometry produced by theoretical stent implantation influence distributions of wall shear stress: analysis of a curved coronary artery using 3D computational fluid dynamics modeling. Biomedical engineering online 5: 40. Available: http://www.biomedical-engineering-online.com/content/5/1/40. Accessed 21 January 2012. [DOI] [PMC free article] [PubMed]
  • 27. Mortier P, Holzapfel G, De Beule M, Van Loo D, Taeymans Y, et al. (2010) A Novel Simulation Strategy for Stent Insertion and Deployment in Curved Coronary Bifurcations: Comparison of Three Drug-Eluting Stents. Annals of biomedical engineering 38: 88–99. [DOI] [PubMed] [Google Scholar]
  • 28. Larrabide I, Kim M, Augsburger L, Villa-Uriol MC, Rüfenacht D, et al. (2012) Fast virtual deployment of self-expandable stents: Method and in vitro evaluation for intracranial aneurysmal stenting. Medical Image Analysis 16: 721–730. [DOI] [PubMed] [Google Scholar]
  • 29. Gundert TJ, Shadden SC, Williams AR, Koo BK, Feinstein JA, et al. (2011) A rapid and computationally inexpensive method to virtually implant current and next-generation stents into subject-specific computational fluid dynamics models. Annals of biomedical engineering 39: 1423–1437. [DOI] [PubMed] [Google Scholar]
  • 30. Williams AR, Koo B-K, Gundert TJ, Fitzgerald PJ, LaDisa JF (2010) Local hemodynamic changes caused by main branch stent implantation and subsequent virtual side branch balloon angioplasty in a representative coronary bifurcation. Journal of applied physiology 109: 532–540. [DOI] [PubMed] [Google Scholar]
  • 31.De Santis G, Conti M, Trachet B, De Schryver T, De Beule M, et al.. (2011) Haemodynamic impact of stent–vessel (mal)apposition following carotid artery stenting: mind the gaps! Computer Methods in Biomechanics and Biomedical Engineering: 1–12. [DOI] [PubMed]
  • 32. Morlacchi S, Keller B, Arcangeli P, Balzan M, Migliavacca F, et al. (2011) Hemodynamics and In-stent Restenosis: Micro-CT Images, Histology, and Computer Simulations. Annals of biomedical engineering 39: 2615–2626. [DOI] [PubMed] [Google Scholar]
  • 33. Benndorf G, Ionescu M, Valdivia y Alvarado M, Biondi A, Hipp J, et al. (2010) Anomalous hemodynamic effects of a self-expanding intracranial stent: comparing in-vitro and ex-vivo models using ultra-high resolution microCT based CFD. Journal of biomechanics 43: 740–748. [DOI] [PubMed] [Google Scholar]
  • 34. Foin N, Secco GG, Ghilencea L, Kram R, Di Mario C (2011) Final proximal post-dilatation is necessary after kissing balloon in bifurcation stenting. Eurointervention 7: 597–604. [DOI] [PubMed] [Google Scholar]
  • 35. Hikichi Y, Inoue T, Node K (2009) Benefits and Limitations of Cypher Stent-Based Bifurcation Approaches: In Vitro Evaluation Using Micro-Focus CT Scan. Journal of Interventional Cardiology 22: 128–134. [DOI] [PubMed] [Google Scholar]
  • 36. Wang S, Ding G, Zhang Y, Yang X (2011) Computational haemodynamics in two idealised cerebral wide-necked aneurysms after stent placement. Computer Methods in Biomechanics and Biomedical Engineering 14: 927–937. [DOI] [PubMed] [Google Scholar]
  • 37. Ohta M, Chuan H, Nakayama T, Takahashi A, Rufenacht DA (2007) Three-Dimensional Reconstruction of a Cerebral Stent using Micro-CT for Computational Simulation. Journal of Intelligent Material Systems and Structures 19: 313–318. [Google Scholar]
  • 38. Connolley T, Nash D, Buffiere JY, Sharif F, McHugh PE (2007) X-ray micro-tomography of a coronary stent deployed in a model artery. Medical engineering & physics 29: 1132–1141. [DOI] [PubMed] [Google Scholar]
  • 39. Benndorf G, Ionescu M, MV YA, Hipp J, Metcalfe R (2009) Wall shear stress in intracranial self-expanding stents studied using ultra-high-resolution 3D reconstructions. AJNR American journal of neuroradiology 30: 479–486. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Yajima J (2007) Quantifizierung der Intimahyperplasie nach Ballonangioplastie und Stentimplantation in der Kaninchenaorta mittels 3D mikro-Computer Tomographie. Gießen: Universität Gießen. VVB Laufersweiler publishing, 40 p.
  • 41.Kralev S, Haag B, Spannenberger J, Lang S, Brockmann MA, et al. (2011) Expansion of the Multi-Link Frontier™ Coronary Bifurcation Stent: Micro-Computed Tomographic Assessment in Human Autopsy and Porcine Heart Samples. PloS one 6: e21778. Available: http://www.plosone.org/article/info 3Adoi 2F10.1371 2Fjournal.pone.0021778. Accessed 21 January 2012. [DOI] [PMC free article] [PubMed]
  • 42. Foerst J, Ball T, Kaplan AV (2010) Postmortem In Situ Micro-CT Evaluation of Coronary Stent Fracture. Catheterization and Cardiovascular Interventions 76: 527–531. [DOI] [PubMed] [Google Scholar]
  • 43. LaDisa JF Jr, Olson LE, Molthen RC, Hettrick DA, Pratt PF, et al. (2005) Alterations in wall shear stress predict sites of neointimal hyperplasia after stent implantation in rabbit iliac arteries. American journal of physiology Heart and circulatory physiology 288: H2465–2475. [DOI] [PubMed] [Google Scholar]
  • 44. Andermahr J, Helling HJ, Landwehr P, Fischbach R, Koebke J, et al. (1999) The lateral calcaneal artery. Surgical and Radiologic Anatomy 20: 419–423. [DOI] [PubMed] [Google Scholar]
  • 45. Myers JG, Moore JA, Ojha M, Johnston KW, Ethier CR (2001) Factors Influencing Blood Flow Patterns in the Human Right Coronary Artery. Annals of biomedical engineering 29: 109–120. [DOI] [PubMed] [Google Scholar]
  • 46. Murphy J, Boyle F (2010) Predicting neointimal hyperplasia in stented arteries using time-dependant computational fluid dynamics: a review. Computers in biology and medicine 40: 408–418. [DOI] [PubMed] [Google Scholar]
  • 47. Murray CD (1926) The Physiological Principle of Minimum Work: I. The Vascular System and the Cost of Blood Volume. Proceedings of the National Academy of Sciences of the United States of America 12: 207–214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Murphy JB, Boyle FJ (2010) A full-range, multi-variable, CFD-based methodology to identify abnormal near-wall hemodynamics in a stented coronary artery. Biorheology 47: 117–132. [DOI] [PubMed] [Google Scholar]
  • 49. Jang I-K, Tearney G, Bouma B (2001) Visualization of Tissue Prolapse Between Coronary Stent Struts by Optical Coherence Tomography. Circulation 104: 2754. [DOI] [PubMed] [Google Scholar]
  • 50.Hong YJ, Jeong MH, Choi YH, Song JA, Kim DH, et al. Impact of tissue prolapse after stent implantation on short- and long-term clinical outcomes in patients with acute myocardial infarction: An intravascular ultrasound analysis. International journal of cardiology, In press. [DOI] [PubMed]
  • 51. Garasic JM, Edelman ER, Squire JC, Seifert P, Williams MS, et al. (2000) Stent and Artery Geometry Determine Intimal Thickening Independent of Arterial Injury. Circulation 101: 812–818. [DOI] [PubMed] [Google Scholar]
  • 52. Wentzel JJ, M. Whelan D, van der Giessen WJ, van Beusekom HMM, Andhyiswara I, et al. (2000) Coronary stent implantation changes 3-D vessel geometry and 3-D shear stress distribution. Journal of Biomechanics 33: 1287–1295. [DOI] [PubMed] [Google Scholar]
  • 53. Castagna MT, Mintz GS, Leiboff BO, Ahmed JM, Mehran R, et al. (2001) The contribution of “mechanical” problems to in-stent restenosis: An intravascular ultrasonographic analysis of 1090 consecutive in-stent restenosis lesions. American Heart Journal 142: 970–974. [DOI] [PubMed] [Google Scholar]
  • 54. Napoli C, De Nigris F, Pignalosa O, Lerman A, Sica G, et al. (2006) In vivo veritas: Thrombosis mechanisms in animal models. Scandinavian Journal of Clinical & Laboratory Investigation 66: 407–428. [DOI] [PubMed] [Google Scholar]
  • 55. Katritsis DG, Theodorakakos A, Pantos I, Gavaises M, Karcanias N, et al. (2012) Flow Patterns at Stented Coronary Bifurcations. Circulation: Cardiovascular Interventions 5: 530–539. [DOI] [PubMed] [Google Scholar]
  • 56. Malek AM, Alper SL, Izumo S (1999) Hemodynamic shear stress and its role in atherosclerosis. Jama-Journal of the American Medical Association 282: 2035–2042. [DOI] [PubMed] [Google Scholar]
  • 57.Gimbrone MA, Topper JN, Nagel T, Anderson KR, Garcia-Cardena G (2000) Endothelial dysfunction, hemodynamic forces, and atherogenesis. In: Numano F, Gimbrone MA, editors. Atherosclerosis V: The Fifth Saratoga Conference. New York: New York Acad Sciences. 230–240. [DOI] [PubMed]
  • 58. Meyer EP, Beer GM, Lang A, Manestar M, Krucker T, et al. (2007) Polyurethane elastomer: a new material for the visualization of cadaveric blood vessels. Clin Anat 20: 448–454. [DOI] [PubMed] [Google Scholar]
  • 59. Krucker T, Lang A, Meyer EP (2006) New polyurethane-based material for vascular corrosion casting with improved physical and imaging characteristics. Microscopy research and technique 69: 138–147. [DOI] [PubMed] [Google Scholar]
  • 60. Riew CK, Smith RW (1971) Modified osmium tetroxide stain for the microscopy of rubber-toughened resins. Journal of Polymer Science Part A-1: Polymer Chemistry 9: 2739–2744. [Google Scholar]
  • 61. Johnson K, Sharma P, Oshinski J (2008) Coronary artery flow measurement using navigator echo gated phase contrast magnetic resonance velocity mapping at 3.0 T. Journal of Biomechanics. 41: 595–602. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62. Gatehouse PD, Keegan J, Crowe LA, Masood S, Mohiaddin RH, et al. (2005) Applications of phase-contrast flow and velocity imaging in cardiovascular MRI. Eur Radiol 15: 2172–2184. [DOI] [PubMed] [Google Scholar]
  • 63. van der Giessen AG, Groen HC, Doriot P-A, de Feyter PJ, van der Steen AFW, et al. (2011) The influence of boundary conditions on wall shear stress distribution in patients specific coronary trees. Journal of Biomechanics 44: 1089–1095. [DOI] [PubMed] [Google Scholar]
  • 64. Vignon-Clementel IE, Figueroa CA, Jansen KE, Taylor CA (2010) Outflow boundary conditions for 3D simulations of non-periodic blood flow and pressure fields in deformable arteries. Computer Methods in Biomechanics and Biomedical Engineering 13: 625–640. [DOI] [PubMed] [Google Scholar]
  • 65. Olgac U, Poulikakos D, Saur SC, Alkadhi H, Kurtcuoglu V (2009) Patient-specific three-dimensional simulation of LDL accumulation in a human left coronary artery in its healthy and atherosclerotic states. Am J Physiol Heart Circ Physiol 296: H1969–1982. [DOI] [PubMed] [Google Scholar]
  • 66. Verli FD, Rossi-Schneider TR, Schneider FL, Yurgel LS, de Souza MAL (2007) Vascular Corrosion Casting Technique Steps. Scanning 29: 128–132. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Figure S1

Cross-sections of stent struts acquired at (A) 6 µm, (B) 8 µm, (C) 10 µm and (D) 12 µm scan resolution. Using the manufacturer’s production specifications as reference, the 6 µm resolution scan was found to capture the stent geometry with sufficient accuracy.

(TIF)

Figure S2

Registration of the high resolution surface of a stented artery section with the lower resolution surface of the whole arterial geometry. (A) Low resolution surface of the whole artery (B) High resolution surface of the stented region (C) Combined surface.

(TIF)

Figure S3

Tetrahedral mesh cross-section at a stented artery section. The inset shows the refined computational grid at the artery wall.

(TIF)

Figure S4

Illustration of the bifurcation mass-flow conditions based on Murray’s law as given by Equations S1 and S2.

(TIF)

Appendix S1

Expanded Methods.

(DOCX)

Video S1

Reconstruction of stented LCA section and visualization of blood flow field.

(MPG)

Video S2

Visualization of flow disturbance introduced by malapposed stent strut.

(MPG)


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES