Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2013 Apr 1.
Published in final edited form as: J Magn Reson. 2008 Jul 13;194(2):182–191. doi: 10.1016/j.jmr.2008.07.006

MR Imaging of Apparent 3He Gas Transport in Narrow Pipes and Rodent Airways

Kevin R Minard 1, Richard E Jacob 1, Gernot Laicher 2, Daniel R Einstein 1, Andrew P Kuprat 1, Richard A Corley 1
PMCID: PMC3613337  NIHMSID: NIHMS447213  PMID: 18667344

Abstract

High sensitivity makes hyperpolarized 3He an attractive signal source for visualizing gas flow with magnetic resonance (MR) imaging. Its rapid Brownian motion, however, can blur observed flow lamina and alter measured diffusion rates when excited nuclei traverse shear-induced velocity gradients during data acquisition. Here, both effects are described analytically, and predicted values for measured transport during laminar flow through a straight, 3.2-mm-diameter pipe are validated using two-dimensional (2D) constant-time images of different binary gas mixtures. Results show explicitly how measured transport in narrow conduits is characterized by apparent values that depend on underlying gas dynamics and imaging time. In ventilated rats, this is found to obscure acquired airflow images. Nevertheless, flow splitting at airway branches is still evident and use of 3D vector flow mapping is shown to reveal surprising detail that highlights the correlation between gas dynamics and lung structure.

Keywords: 3He, Gas, Flow, Diffusion, MRI

1. Introduction

Magnetic resonance (MR) imaging with hyperpolarized xenon (129Xe) or helium (3He) affords high sensitivity for measuring gas flow in straight pipes (12), porous foam (3), flexible tubes (4), human airway models (2, 5), and breathing patients (2). Flow quantification, however, is challenged by rapid gas diffusion. This is especially problematic in small conduits where diffusing gas can readily sample shear-induced velocity gradients over the imaging time (t). In phase-contrast MR images this has been observed to blur laminar flow in small capillaries (1). For a pipe of radius a, and gas molecules with diffusion coefficient D, results show that the degree of blurring can be significant - even over relatively short time scales (t < a2/D). Since axial diffusion is also found to be time-dependent (1), prior work suggests that MR measurements of gas transport in narrow conduits provide apparent values that depend on underlying flow structure, gas diffusion, and imaging time.

Recent work on gas-phase MR stresses its potential utility for visualizing gas dynamics in applications ranging from toxicology (6) and inhaled drug delivery (2) to catalytic (7) and aeronautic design (8). In related studies, most results have been acquired in larger conduits where the imaging time (t) is short relative to gas diffusion and flow structure. Blurring effects have therefore been ignored, and to a large extent, good agreement between measured flow and predictions based on computational fluid dynamics (CFD) has been observed (2, 5, 8). Even in these studies, however, subtle differences are apparent in regions where strong velocity gradients exist on diffusive length scales. In small capillaries, Monte Carlo simulations suggest that these differences might be attributable to diffusion-mediated sampling of multiple flow streams during image collection (1). Nevertheless, no analytical description of this phenomenon has yet been formulated. Furthermore, no measurements of 3He flow in narrow conduits of complex shape have yet been reported. It is therefore unclear what information might be available, or to what extent it might be limited by diffusion-mediated effects.

Prior work on gas transport in small capillaries employed one-dimensional (1D) flow imaging, axial diffusion was not resolved, and results were understood through Monte Carlo simulations (1). Here, imaging is extended to two- & three-dimensions to facilitate transport measurements in more complex systems, and apparent gas dynamics in straight pipes is described analytically to provide deeper physical insight. Analytical theory explicitly accounts for gas dynamics, incorporates the details of MR data acquisition, and is shown to accurately predict the combined effects of flow and diffusion over a range of experimental conditions. In live rats, these effects are found to obscure laminar flow in pulmonary airways, thereby, confounding any direct comparison with CFD predictions. Despite this limitation, however, 3D images of in-vivo airflow are shown to provide a detailed view of pulmonary airflow patterns not yet seen. Hallmarks include the visualization of flow streaming similar to that recently reported in humans (2), and detailed flow splitting at airway branches that may provide a quantitative basis for assessing ventilation to each of the rat’s five lung lobes.

2. Theory

2.1. Structure & Function of Constant-Time Imaging (CTI)

During MR imaging random Brownian motion causes irreversible signal loss that reduces sensitivity and fundamentally limits resolution (911). Constant-time imaging (CTI) minimizes both adverse effects (1213), and for this reason, is increasingly used for gas visualization (6, 8, 1415). Its simple structure and function is also easily understood, and here, this facilitates a rigorous description of measured gas transport.

During a 2D CTI experiment like that shown in Fig. 1, spatial information within the excited slice is phase-encoded along each planar axis. Repeated application of the pulse sequence using different phase encoding gradients is then utilized to collect a motion-sensitive image. Subsequent analysis of images acquired using different motion-sensitizing gradients then provides a basis for local transport measurements (1618).

Figure 1.

Figure 1

Timing diagram for 2D CTI. Since slice-select gradients (checkered) are velocity compensated, moving gas is refocused before motion-sensitizing gradients (solid-black) are applied to measure gas dynamics along the same axis. Motion sensitization, however, occurs after a 2D phase encoding gradient (shaded) is applied to resolve planar location at t = 0. Consequently, if gas molecules traverse multiple flow streams during the evolution time (t), direct correspondence is lost between planar location and measured transport.

As with all MR imaging methods, a finite time is typically required to encode location and motion. In the context of Fig. 1, this means that excited gas molecules can diffuse across shear-induced velocity gradients after their location is initially encoded, and motion-sensitizing gradients are applied. Direct correspondence between location and dynamics is therefore compromised, and this necessarily blurs measured transport. The following theory describes this loss of correspondence and its ensuing affects on measured transport in constant-time images.

2.2. Effects of Gas Motion on 2D CTI

In this study the slice-select gradient in Fig. 1 is aligned with the central axis of a straight, circular pipe, and 2D phase encoding is employed for visualizing 3He gas over its cross section. Axial flow and diffusion is then measured using the spatially resolved MR signal (S) from two different images - one acquired with motion sensitizing gradients applied along the slice-select axis (SG), and the other without (S0).

During data acquisition molecular motion through applied gradients alters the relative phase ϕ and attenuation β of the spatially resolved MR signal at each location r0 in a motion-sensitized image (1618). If 2D phase encoding pulses are short, and gas location is assumed to be instantaneously encoded at t=0 defined in Fig. 1, then relative signal changes can be conveniently expressed as an ensemble average (〈 〉) over all molecules initially encoded at r0, such that (16, 19) –

SGS0=eiϕβ=eiγ0τF(t)v(t)dt. [1]

Here, γ denotes the gyromagnetic ratio, and v(t) represents each molecule’s instantaneous axial velocity. Other parameters include the time (τ) when the NMR receiver is turned on, and the time-integral of motion-sensitizing gradients -

F(t)=0tG(t')dt', [2]

where G(t) is their time-dependent amplitude, and F(τ) is taken to be zero at the start of data acquisition (16, 2021).

Since a large number of molecules normally contribute to the spatially resolved MR signal it is useful to relate relative changes to their underlying statistical dynamics (16, 19). In practice, this is facilitated by exploiting Kubo’s generalized cumulant expansion theorem, such that (2021) –

iϕβ=λ=n=1(iγ)n0τdt10t1dt20tn1dtnv(t1)v(t2)v(tn)cF(t1)F(t2)F(tn). [3]

In this expression angular brackets 〈 〉c denote cumulant averages over the enclosed quantity. Generally, these completely define underlying statistics, are calculated in terms of ensemble averages (〈 〉), and are sometimes referred to as connected averages since they vanish if any component (i.e. velocity at different times) is uncorrelated with the others (2223). For convenience, the first three cumulants are defined below, and expressions for higher orders are found elsewhere (2223) -

v(t1)c=v(t1), [4]
v(t1)v(t2)c=v(t1)v(t2)v(t1)v(t2), [5]
v(t1)v(t2)v(t3)c=v(t1)v(t2)v(t3)v(t1)v(t2)v(t3)v(t2)v(t1)v(t3)v(t3)v(t1)v(t2)+2v(t1)v(t2)v(t3). [6]

Equation 3 shows that relative phase (ϕ) is determined by the real part of the cumulant expansion (Re λ), that relative attenuation (β) is determined by the imaginary part (-Im λ), and emphasizes how both are influenced by motion sensitizing gradients and the statistical dynamics of gas motion along their applied direction. In practice, this general description serves as a convenient theoretical construct that has previously been employed for other MR transport studies (2021). Here, integrals are explicitly time-ordered such that τ>t1>t2>…tn (2224). Generally, this facilitates the analysis of time-dependent processes (24). Utilization in the current study, however, is only valid for stationary transport processes since MR images are acquired over time using repeated data acquisitions. Physically, this means that underlying dynamics can only depend on the time intervals between successive observations - rather than the exact time each occurs (25). Use of Eq. 3 also assumes that relative changes in either image phase or attenuation are dominated by molecular displacements during motion sensitizing gradient pulses, and that the effects of motion during spatial encoding are relatively small. To ensure this as much as possible, slice-select gradients in Fig.1 are velocity compensated, and phase encoding is kept as short as possible (26).

2.3. The Statistical Dynamics of Laminar Pipe Flow

During constant laminar flow in a straight, circular pipe, moving material is directed down its central axis with a speed (V) that varies with radial distance (r), such that (2728) –

V(r)=Vmax(1r2a2). [7]

Here, a represents the pipe’s radius and Vmax is twice the average flow speed across the pipe (28).

Generally, Eq. 7 represents a specific solution to the Navier-Stokes equation that is valid when the Reynolds number is less than ~ 2100, gas density is constant, flow is steady, Newton’s law of viscosity applies, and effects near the pipe’s entrance and exit are neglected (28). In the current context, V(r0) describes the average axial velocity (〈v(t = 0)〉) when gas is initially encoded at some radius r0. At later times, 〈v(t)〉 changes as encoded gas diffuses down the pipe and across streamlines. To formulate a statistical description of this time-dependent motion it is helpful to think of Eq. 7 as a stationary flow field that is sampled by randomly diffusing gas molecules. If r(t) is a stationary Markov process describing each molecule’s random position in the absence of flow (25), v(t) may be thought of as having two distinct contributions. One arises from the flow field itself and is represented using the random function V(r(t)). The other describes each molecule’s random axial velocity as it diffuses down the pipe in the absence of flow. If this is represented by u(t), then net axial motion is expressed as a simple superposition -

v(t)=u(t)+V(r(t)). [8]

In the absence of flow (i.e. when V=0), Eq. 8 shows that axial motion is governed by the random velocity fluctuations u(t) that diffusing gas normally exhibits. During flow, diffusion mediated changes in location r determine the precise flow rate a diffusing gas molecule must experience. When diffusion is slow relative to flow structure and imaging time, V(r(t)) ~ V(r0), and Eq. 8 represents the traditional description of constant flow and diffusion that is widely used for liquid-state studies (1619). Here, a more general formulation is proposed by accounting for changes in flow velocity that occur as gas diffuses across streamlines on the imaging time scale. Generally, since a finite time is required to diffuse from one location r to the next, changes in flow velocity are mediated by the dynamics of Brownian displacements and don’t occur instantaneously. Rather, causality is ensured by an accurate description of gas diffusion and the likelihood that changes in location can occur over any given timescale.

Implicit in Eq. 8 is the assumption that axial velocity fluctuations from molecular diffusion down the pipe (u) are statistically independent of the axial flow (V). Physically, this is not unreasonable since the correlation time for u(t) is much shorter than time needed to diffuse across streamlines. One practical consequence of this is that cumulants for statistically independent processes are additive (22). As a result,

v(t1)v(t2)v(tn)c=u(t1)u(t2)u(tn)c+V(r(t1))V(r(t2))V(r(tn))c, [9]

and effects from axial diffusion (u) and flow (V) can be treated separately, such that –

λ=λu+λV. [10]

Since u(t) describes axial velocity fluctuations for diffusing gas in the absence of flow, and these are completely random on the imaging time scale, their average (〈u(t)〉) is zero (19). A statistical description of unrestricted diffusion as a Gaussian random process also shows that 〈u(t1)u(t2)〉 = 2Dζ(t2t1), where D is the free (unrestricted) diffusion coefficient and ζ is the dirac delta function (19). Since all cumulants other than second order (n=2) are zero for a Gaussian random process (23), substitution of previous results into Eq. 3 shows that (19) –

λu=Db, [11]

where the so-called gradient factor (b) is γ20τF2(t')dt'.

Equation 11 describes the contribution axial diffusion makes to the cumulant expansion (λu) in Eq. 10. The other contribution (λV) arises from flow and is mediated by Brownian displacements across streamlines. To describe this it is useful to express ensemble averages over the random function V(r(t)) in terms of the stationary flow field V and the random position r for diffusing gas at different times –

V(r(t1))V(r(t2))V(r(tn))=V(r1)V(r2)V(rn). [12]

Here, r1 is used as shorthand notation to represent the random position for a gas molecule at a time t1 (i.e. r(t1)) after its location is first encoded at ro. Values for r2rn are similarly defined.

To evaluate the right side of Eq. 12 it is generally necessary to define the joint probability density (Wn) for different space-time trajectories that start from the encoded image location r0 (25, 30). The ensemble average is then taken over all possible paths that originate from ro and pass through specified times.

Figure 2 shows a trajectory that is described by W2(r0|r2,t2;r1,t1)dr2 dr1. This utilizes time-ordered notation (t1>t2>t3…tn-1>tn) and describes the probability for finding a gas molecule within a small, elementary volume dr2 at (r2, t2), and then within dr1 of (r1, t1) - after location is first encoded at r0 (25). As a stationary Markov process, each leg in the trajectory is statistically independent and probabilities only depend on time intervals between successive points (25, 30). In this case, the joint probability density function for trajectories of any order (Wn) can be defined conveniently using the diffusion propagator P, such that (25, 30)

Wn(r0|rn,tn;rn1,tn1r1,t1)=P(r0|rn,tn)P(rn|rn1,tn1tn)P(r2|r1,t1t2). [13]

Generally, n propagators are used starting from the leftmost on the right side of the equality. Functionally, P(r0|r,t)dr describes the conditional probability that a gas molecule initially at r0 will be within dr of r after a time t (19, 25, 30). Other propagators in Eq. 13 have the same functional form, but initial and final locations differ in order to describe all possible trajectories.

Figure 2.

Figure 2

Space-time trajectories for the second order, joint probability density function (W2).

In practice, the above considerations provide a general theoretical framework for calculating ensemble averages over diffusing gas molecules in any stationary flow field (V). The general result for calculating the nth order ensemble average over any random function V(r) has been used previously by others (24) –

V(r(t1))V(r(t2))V(r(tn))=r1dr1r2dr2rndrnV(r1)V(r2)V(rn)Wn(r0|rn,tn;rn1,tn1;r1,t1). [14]

2.4. 2D Diffusion Propagator for A Cylindrical Pip

In a straight cylindrical pipe, diffusion over its cross-section determines how individual molecules sample the stationary flow field defined by Eq. 7. The propagator (P) is therefore governed by Fick’s law, D2P=Pt, where D is the ‘free’ (i.e. unrestricted) molecular diffusion coefficient. This is conveniently solved in two-dimensions (2D) using cylindrical-polar coordinates (r, θ), and the initial condition that P(r0, θ0|r, θ, t =0) = (1/r) ζ(r-r0) ζ(θ- θ0) where ζ is the Dirac delta function. To ensure probability is conserved it is also required that tP(r0,θ0|r,θ,t)rdrdθ is always zero. Utilization of Fick’s law and the divergence theorem then shows that a specific solution must satisfy Pr=0 at all points on the pipe’s surface. Under these conditions the diffusion propagator can be expressed in the following fashion (29) -

P(r0,θ0|r,θ,t)=j=0k=1Ajk2eβjk2Dt/a2Jj(βjkr/a)Jj(βjkr0/a)Cos(j(θθ0)). [15]

In this equation Jj denotes the jth order Bessel function, and βjk is the kth root of –

ddx(Jj(x))=0. [16]

In addition, for j≠0

Ajk2=(2πa2)βjk2Jj2(βjk)(βjk2j2), [17]

otherwise

A0k2=(1πa2)1J02(β0k). [18]

A 2D analogy for Eq. 14 is then used to express the nth-order ensemble average of axial flow velocity in a straight pipe –

V(r(t1))V(r(t2))V(r(tn))=r1θ1r1dr1dθ1r2θ2r2dr2dθ2rnθnrndrndθnV(r1)V(r2)V(rn)Wn(r0,θ0|rn,θn,tn;rn1,θn1,tn1r1,θ1,t1). [19]

Here, radial integrals vary from 0 to a, and integrals over θ are from 0 to 2π.

2.5. Defining Apparent Flow & Diffusion

In most liquid-state applications molecular diffusion is not rapid enough to move excited nuclei across different flow streams during the imaging time. If intra-voxel dephasing is neglected, the relative phase for the spatially resolved MR signal at location r0 in a flow-encoded image (ϕ(r0)) is just α (r0), where α=γ0τF(t')dt', (r0) and represents the average velocity over the resolved voxel. A plot of ϕ(r0) versus α at each location is therefore commonly employed for measuring regional fluid flow (1619). Here, the same approach is used to measure the apparent gas velocity (vapp) at each image location (r0). Predictions based on the numerical evaluation of the cumulant expansion (λ) are then formulated using the same definition, such that –

vapp(r0)=ϕ(r0)α=Imλα. [20]

Similar rationale is used for defining the apparent gas diffusion coefficient –

Dapp(r0)=β(r0)b=Reλb. [21]

Particularly, since Eq. 11 shows that this yields a value for the gas diffusion coefficient (D) in the absence of flow.

3. Materials & Methods

3.1. 3He Gas Generation, Delivery, & Characterization for Pipe Flow

Recent work describes our apparatus for generating hyperpolarized (HP) 3He (31). Immediately before MR experiments, 0.4–0.5 L of 3He gas with ~40% polarization was released into a 2-liter Tedlar bag (Jensen Inert Products, Coral Springs, FL) that resided inside an acrylic box located within the fringe field of the imaging magnet. During flow experiments, the box was pressurized to push HP gas out of the bag at a constant rate. Immediately after leaving, 3He gas was diluted to ~10-volume-percent by mixing it with another flow stream consisting of either N2 or C3F8 gas. The resulting mixture then traveled into a Tygon hose that was ~1 meter long, had an inner diameter of 3.2 mm, and carried gas straight through the magnet bore. Generally, use of different binary gas mixtures provided a convenient way of altering the rate of 3He diffusion so impact on acquired images could be examined.

During gas delivery the volumetric flow through the Tygon tube was measured to be ~240 cc/min. This was monitored at the tube’s outlet by bubbling each binary gas mixture into a (inverted) graduated cylinder filled with water. Measurements with and without 3He then allowed determination of delivered gas fractions, and facilitated fine adjustment to achieve described conditions.

At ~240 cc/min the average axial flow velocity in the Tygon tube ( = Vmax/2) was ~0.5 m/s. At standard temperature and pressure (STP), the kinematic viscosity (vk) of nitrogen is 1.4 × 10−4 m2/s, and vk= 1.5 × 10−5 m2/sec for C3F8 (32). The Reynold’s number (Re = 2av̄/vk) for dilute mixtures with 3He is therefore about 10 and 100 respectively. Since this is much lower than the transitional value to turbulance (~2100), flow under the specified conditions is expected to exhibit the radial velocity profile defined by Eq. 7.

Prior to imaging gas transport, a non-imaging gradient-echo experiment was employed to measure the diffusion coefficient for diluted 3He in N2 and C3F8 mixtures. To avoid flow-enhanced effects, motion-sensitizing gradients were applied transverse to the flow axis using a bipolar pulse pair (17) similar to that depicted in Fig. 1. The gradient pulse width (δ) was 0.5 ms and the leading-edge pulse-separation (Δ) was 0.6 ms. The measured 3He diffusion coefficient transverse to flow in a 10-volume-percent mix with N2 was 0.78 +/− 0.04 cm2/s. For the same fraction of 3He in C3F8 the measured value was 0.36 +/− 0.02 cm2/s. Since the diffusion time (Δ- δ/3) was only ~ 0.4 msec, the root-mean-square displacement (2Dt) for diffusing 3He gas was ~250 and 170 microns for mixtures with N2 and C3F8, respectively. Given that both were much smaller than the tube diameter (3.2 mm), potential effects stemming from restricted diffusion on the measurement time-scale were negligible. This was confirmed later using the Chapman-Enskog theory of binary gas diffusion - which predicts similar 3He diffusion coefficients (33).

3.2. 2D CTI of Axial Gas Transport in a Straight Pipe

Visualization of 3He axial gas transport was achieved using the 2D, CTI experiment depicted in Fig. 1. Imaging was performed at 2.0 Tesla using a UNITYPlus console (Varian, Palo Alto, CA), a home-built 7-cm-diameter radio-frequency volume coil, and a 30-cm-diameter horizontal-bore magnet (Oxford Instruments, Oxford, UK) equipped with actively shield gradients characterized by 60 µsec rise-times (Resonance Research Inc., Billerica, MA). During gas delivery, constant-time images were collected using a 1.0 cm × 1.0 cm field-of-view (FOV), and a 6.0-mm-thick slice that was transverse to the central axis of the Tygon tube. The imaging pulse sequence utilized a 20 ms repetition time (TR), a receiver bandwidth of 1 kHz, and 64 phase-encoding steps along each planar direction. In all cases, the bipolar gradient was characterized by a pulse-duration (δ) of 300 µs, and two images were acquired – one with no motion-sensitizing gradient and the other with a finite amplitude (G).

To examine the effects of encoding duration, imaging was repeated using two different leading-edge pulse-separations (Δ). For Δ = 1 ms, the gradient amplitudes were 0 and 4 Gauss/cm; and for Δ = 4 ms, they were 0 and 1 Gauss/cm. Generally, this ensured flow sensitivity (α) was the same. For each Δ-value, images were acquired in an interleaved fashion such that each point in k-space was sampled sequentially using both motion-sensitizing gradient amplitudes. The phase encode gradient was then incremented after 2 averages. Because two interleaved images were acquired, relative image attenuation was directly attributed to gas transport and was uninfluenced by either T1- (signal loss inside the Tedlar bag) or T2-relaxation.

After raw data collection, images were reconstructed on a 128 × 128 matrix using a phase-sensitive, 2D Fourier Transform. Analysis of images acquired without motion sensitizing gradients showed that the average signal-to-noise ratio (SNR) over the pipe was ~100. Apparent flow velocity (Eq. 20) was determined at each matrix location with an SNR ≥ 50 by measuring the phase difference between data acquired with and without motion sensitizing gradients. Apparent diffusion (Eq. 21) was determined by measuring relative attenuation. In both cases, gradient rise-time was ignored and motion-sensitizing gradient pulses in Fig. 1 were assumed to be square in shape. In this case α = γGΔδ (1718), b = γ2 G2 δ2 (Δ − δ/3) (19), and it is noted that the gyromagnetic ratio (γ) for 3He is 20376 rad/(Gauss·sec) (34). Since apparent diffusion is determined by measuring relative attenuation, imaging gradients make no contribution to the b-value used in Eq. 21 (35). Possible cross-terms with applied slice selection gradients are also zero because they refocus spins before motion-sensitizing gradients are applied (35).

3.3. Simulated Pipe Dynamics

Apparent transport in a straight pipe was predicted in Mathematica using Eqs. 2021. Related code generally involved the numerical evaluation of the cumulant expansion (λV) and multi-dimensional integrals over the diffusion propagator defined in Eq. 15. The Appendix shows that significant simplification is possible because only the j=0 terms contribute to angular integrals that define ensemble averages in Eq. 19, and that each of these is 2π.

To optimize the speed and accuracy of numerical calculations, apparent transport was evaluated multiple times using the first three terms in the cumulant expansion of Eq. 3 (n=3). Generally, each evaluation was performed using either – 1) a different support (i.e. discrete sampling) for numerical integration, or 2) a different number of k-terms in the expansion for the diffusion propagator (Eq. 15). The minimum values need for results to change less than a few percent after successive doublings were then chosen. This showed that only ~32 k-terms were needed to achieve computational convergence comparable to the uncertainty of experimental data.

3.4. 2D Airflow Imaging in Ventilated Rats

Pulmonary airflow was visualized in Male Sprague-Dawley rats (Charles River Laboratories, Wilmington, MA) using 2D CTI and protocols approved by the Institutional Animal Care and Use Committee at Pacific Northwest National Laboratory. Prior to experiments, rats were anesthetized using 3–4% isoflurane in air. A tracheotomy was then performed and a 14-gauge catheter tube was inserted into the exposed trachea. After securing with surgical string, each animal was placed in a supine position and its tracheal tube was attached to the mouthpiece of a home-built, computer-controlled, ventilator similar to described designs (3637). To facilitate animal positioning, the rat and mouthpiece were both secured on a wooden tray that also accommodated a NMR compatible ECG/temperature monitoring unit (SA Instruments Inc., Stony Brook, NY, Model 1025). The tray was then slid onto a mating rail that was located inside a double-tuned (1H/3He) RF coil (38).

During imaging, body temperature was maintained by circulated warm air inside the magnet bore. Body temperature and heart rate were monitoring using a rectal probe and ECG electrodes respectively. To avoid possible RF interference, physiological data were fed via fiber optic to a dedicated laptop. Typically, rectal temperature was about 35°C, and pulse rate was between 250–330 beats per minute.

Figure 3 shows a typical respiratory maneuver used for imaging pulmonary airflow. During imaging, the maneuver was repeated every 1.6 seconds, and the pulse-sequence was applied during inhalation with a stationary flow rate. Delivered gas mixtures and drive pressures are specified in the caption. The trace in Fig. 3 shows the flow velocity through a calibrated pneumotach that was placed between the ventilator’s mouthpiece and the attached rat (39). It shows that gas speed is constant over the imaging window. Generally, in vivo CTI acquisition is similar to that used for visualizing pipe flow. In vivo imaging differs in that – 1) a thinner slice is employed to reduce volume averaging, 2) a lower flip angle is used to avoid saturation of inflowing magnetization, and 3) a shorter TR is employed to allow 32 acquisitions during the constant flow window. Specific values for each are therefore specified in later figures.

Figure 3.

Figure 3

Typical respiratory maneuver for in vivo airflow imaging. Boxes along top show gas delivery events and the trace depicts the rate of flowing gas (cc/sec) when measured with a calibrated pneumotach. The O2/N2 is a 30:70 volume mix that is delivered using a drive pressure of ~ 11 cm H20, and the He/N2 is about a 70:30 volume mix (delivered at ~ 10 cm H20). Imaging occurs in the shaded region during a push of oxygen (O2). The O2 is delivered using a drive pressure of about 33 cm H20, and constant flow is achieved by forcing gas through a commercial restrictor (Airtronics, Bellevue, WA, 0.012” orifice diameter, model # F2815-121-B85,). Generally, differences in drive pressure and flow reflect the unique delivery route & plumbing for each gas mixture and don’t reflect actual pressure at the trachea.

3.5. 3D Imaging of Rat Pulmonary Airway Structure & Flow

Extending MR imaging to 3D for visualizing pulmonary structure and flow in ventilated rats exploited two recent technical developments. First, more 3He gas was generated by using larger oblong polarization cells and two parallel lasers. Two cells (each containing ~0.8 L) could then be pumped simultaneously (31). The second development was to employ a hybrid, radial imaging scheme for more efficient k-space sampling (31). The 3D pulse-sequence employs a low flip-angle (~10°) excitation that is followed by bi-polar motion-sensitizing gradient like that shown in Figure 1. Spatial location is then encoded using a 2D radial sampling scheme in conjunction with standard phase encoding along the orthogonal direction (31).

Unlike in 2D imaging studies, rats used for 3D visualization were orally intubated. Prior to intubation, rats were administered 0.02 mL/kg BW glycopyrrolate subcutaneously to inhibit oral secretions and facilitate insertion of an endotracheal tube. After 10–15 minutes, the animals were then anesthetized with isoflurane and placed supine on a tilting intubation table. Approximately 0.02 mL of lidocane was then applied directly to the larynx. Shortly thereafter, a 14 gauge catheter tube was orally inserted into the trachea. The tube was pre-cut to a length to assure that it would pass through the larynx but not come near the carina. Once inserted, the catheter tube was connected to the ventilator.

In the current study gas from two polarization cells was used for each rat. Gas from the first cell was used to acquire a high-resolution view of pulmonary airways. Raw data were collected using 128 phase encodes and 200 radial projections with 128 complex points. The FOV was 6.4 cm on a side and the acquisition bandwidth was 51 kHz. The hybrid pulse sequence utilized a 500 ms hard RF pulse together with an 8 ms TR and no motion encoding. The acquired signal consisted of an asymmetric gradient echo that refocused 1.085 ms after the center of the RF excitation. Image reconstruction was then done as before (31). During the breathing maneuver shown in Fig. 3, 32 phase-encoded projections were acquired during inhalation with a stationary flow rate. The maneuver was then repeated 800 times for a total acquisition length of 21 minutes. To reduce possible blurring, the O2 push was reduced by decreasing the drive pressure to ~ 4 cm H2O.

After gas in the first 3He cell was used for high-resolution imaging, O2 drive pressure was increased to the level specified in Fig. 3. Then, the second 3He cell was utilized for lower-resolution, 3D flow visualization in the same animal. Raw data were collected using 4 interleaved, 3D acquisitions that employed 32-phase-encodes and 128 radial projections comprised of 128 complex points. One 3D acquisition utilized 4 G/cm for motion sensing along the magnet’s x-axis, another used 4 G/cm along its y-axis, the third employed 3 G/cm along z, and finally, the fourth had none. With 32 acquisitions for every breathing maneuver, data collection required 512 repetitions for a total acquisition time of ~14 minutes. All 3D images used the same TR (8 ms), had a refocus time of 2.385 ms for the collected echo, and employed the same timing for motion-sensitizing gradients (δ = 0.3 / Δ = 1.0 ms).

4. Results & Discussion

4.1. Simulated Pipe Flow & The Gaussian Approximation

Equation 19 shows that the average axial flow velocity for an ensemble of molecules originally at r0 inside a straight cylindrical pipe varies with time, such that -

v(t)=r=0aϕ=02πP(r0|r,ϕ,t)V(r)rdrdϕ. [22]

Substituting with Eq. 15 and evaluating integrals then gives –

v(t)=Vmax2+4πVmaxa2k=2A0k2β0k2eβ0k2Dt/a2J0(β0kr0/a)J2(β0k). [23]

The time-dependence embodied by Eq. 23 shows that the average axial flow velocity for 3He gas originally encoded at ro (in a CTI image) approaches Vmax/2 at long times. At shorter times, the precise behavior depends on the initial radial position (ro), the 3He diffusion coefficient (D), and the time t after spatial encoding (see Fig. 1). This general behavior is illustrated in Figure 4 where 〈v(t)〉 is plotted as a function of time for molecules starting at the center of the pipe (dashed) and its edge (solid). Projected behavior shows that gas molecules near the edge accelerate to the average flow velocity (Vmax/2.0), whereas molecules at the center slow to the same value. Generally, the approach to average flow is more rapid at the edge because the radial derivative of V(r) in Eq. 7 is steeper at that location. Consequently, as gas diffuses across different flow streams, molecules initially at the edge experience a larger change in velocity - that then drive’s their more rapid approach to the average axial flow speed (Vmax/2).

Figure 4.

Figure 4

Predicted average axial flow velocity 〈v(t)〉 in cm/s for dilute 3He gas after location in a straight tube is initially encoded at t=0 using the CTI pulse sequence shown in Fig. 1. Flow dynamics were calculated for a 10% volume mixture of 3He in N2 using Eq. 23, a tube radius (a) of 0.16 cm, a maximum flow velocity (Vmax) of 94.4 cm/s, and 0.78 cm2/s for the 3He diffusion coefficient (D) in the binary gas mixture. The red line shows the average flow velocity (Vmax/2 = 47.2 cm/s) across the pipe, and this corresponds to a volume flow rate of approximately 228 cc/min. The dashed line shows the average axial velocity predicted for gas molecules starting from the middle of the tube (r = 0), and the solid line shows the average velocity for gas initially encoded at the tube’s edge (r = a). The shaded region highlights the sampling area of motion sensitizing gradients (i.e. non-zero F(t)) for conditions specified in Figure 5.

In practice, the temporal dependence of 〈v(t)〉 has a significant impact on the apparent axial velocity (vapp) calculated using Eq. 20. Mathematically, this is because 〈v(t)〉 defines the first order (n=1) term in the cumulant expansion of Eq. 3. Since this is imaginary it defines the lowest order contribution to apparent velocity. After integration (over F(t)) the result is used in Eq. 20 to formulate a first order approximation to the apparent velocity (1vapp). Since Eq. 20 shows that only imaginary terms contribute to vapp, and Eq. 3 shows these are all odd, the next highest order contribution comes from the third term in the cumulant expansion (3vapp), such that vapp = 1vapp + 3vapp + …

A plot of 1vapp is shown in Fig. 5 as a function of radial location in a 3.2-mm-diameter straight pipe. Predicted results are formulated using t=0.425 ms as the start of motion encoding, G=4 G/cm, Δ=1.0 ms, δ=0.3 ms, and the same conditions specified in Fig. 4. The axial velocity profile predicted by Eq. 7 is also shown in Fig. 5 for comparison. Differences illustrate how rapid gas diffusion blurs the apparent axial flow so it no longer resembles predictions based solely on the Navier-Stokes equation. By comparison, apparent axial flow is seen to be significantly slower than Eq. 7 predicts at the center of the pipe, and is faster at the pipe edge – as expected from the behavior of 〈v(t)〉 seen in Fig. 4.

Figure 5.

Figure 5

Predicted axial velocity for gas in a straight tube when bi-polar, motion-sensitizing gradients in Fig. 1 start at t=0.425 msecs, G=4 Gauss/cm, Δ=1.0 ms, and δ=0.3 ms. Gas properties, tube radius, and flow conditions are specified in Fig. 4. Predictions based on the solution to the Navier-Stokes equations (V(r) in Eq. 7) are plotted in black, the first order (n=1) approximation to apparent flow velocity (1vapp) in Eq. 20 is shown in bold, and 50 times the third order (n=3) contribution to apparent flow (3vapp) is shown in red.

Figure 5 also shows a plot for the third order contribution to the apparent flow velocity (3vapp). This is seen to be comparatively small and provides valuable insight into the relative influence of higher order terms in the cumulant expansion. From a practical standpoint, this is important for determining how many terms in Eq. 3 are necessary for making accurate predictions. If too few are used achieved accuracy will likely be insufficient, and if too many are employed unnecessary computations will be performed. In prior work, contributions from higher order terms were considered when the correlation time for molecular motion is much shorter than the imaging time (2223). In this study, Fig. 4 shows that motion is more slowly varying. Consequently, the relative importance of higher-order approximations needs reconsideration.

At the onset of this study the relative importance of higher order contributions to the cumulant expansion would be difficult to predict. A general formulation of motion effects is therefore formulated. The benefit of this is that simplifying assumptions about relative contributions can now be made and tested directly using 1vapp and 3vapp in Fig. 5.

During the pulse-sequence shown in Fig. 1, F(t) (from Eq. 2) quickly reaches a minimum value of F = −γGδ. It then remains at that value for a time Δ, after which, it returns to zero. Over the time Δ, when velocity is effectively sampled, gas molecules initially encoded at each location r0 will exhibit some average value that can be determined directly from 〈v(t)〉 in Fig. 4. Since the higher order cumulants describe the statistical deviation dv from this average, contributions in Eq. 3 can be expressed in powers of F dv (22). If dv is slowly varying over Δ, the nth term in Eq. 3 is then of the order (1/n!)(dv Δ F)n when n>1. To validate this scaling relationship, consider the shaded region in Fig. 4. There, it is seen that the flow velocity for diffusing gas at different locations changes by an amount (dv) that is ~20 cm/s when F(t) is non-zero. Since Δ=0.001 sec and F = −γGδ = 24.5 radian per cm, the third order (n=3) contribution to Eq. 3 is estimated (i.e. (1/3!)(dv Δ F)3) to be ~0.019 radian. Given α = γGΔ δ in Eq. 20, and this is 0.024 radian · s / cm under the specified conditions, n=3 contributions to the apparent axial velocity (3vapp) in Eq. 20 are then predicted to be ~0.8 cm/sec. Comparison with the actual third order term in Fig. 5 shows that this is very nearly what is calculated; thereby, validating proposed scaling.

Figure 5 shows that third order contributions to apparent flow are only about 2% of the first order term. Because this is smaller than experimental uncertainty, higher order contributions are safely neglected. Importantly, the same scaling arguments hold for estimating relative contributions to the apparent diffusion coefficient (Dapp) defined through Eq. 21. In this case (1/n!)(dv Δ F)n is about 0.12 for n=2 and ~0.002 for n=4. Approximating Dapp by only the second order (n=2) term in the cumulant expansion is therefore again accurate to about 2%.

The above considerations justify approximation of the cumulant expansion using the first and second terms for calculating apparent flow and diffusion respectively. In practice, this is termed a Gaussian approximation because all cumulants higher than 2nd order are exactly zero for a Gaussian random process (23). Since this can be achieved experimentally simply by adjusting Δ and/or F, its importance to apparent transport studies cannot be over emphasized (2223).

4.2. Apparent 3He Gas Transport & Size Effects in a Straight Pipe

Apparent 3He velocity measurements in a straight pipe are summarized in Fig. 6. Generally, excellent agreement is found between experimental data and predictions based on a first order (n=1) approximation to Eq. 20. Images acquired using different Δ values illustrate how blurring of laminar flow structure becomes more severe as time is lengthened to apply motion-sensitizing gradients. Nevertheless, flow rates (cc/sec) determined by integrating the apparent velocity (vapp) over the pipe cross-section remain constant and are generally within ~10% of experimental values. Measurements of apparent 3He velocity therefore provide a quantitative basis for measuring regional flow rates – even when flow structure is blurred in small conduits. This is predicted mathematically simply by integrating the first order approximation to Eq. 20 over the pipe’s cross-section and noting the diffusion propagator is normalized. The result is that flow rates measured from imaging data are always equal to the value predicted from the Navier-Stokes solution in Eq. 7.

Figure 6.

Figure 6

Apparent axial 3He gas flow in a straight, 3.2-millimeter-diameter tube. Predictions for apparent flow (vapp) are based on the first order (n=1) approximation and are formulated using a Vmax value derived from measured volume flow. In each image, the black plot shows the radial profile for axial flow across the tube diameter. Average uncertainty in measured velocity for all 4 experiments is estimated to be ± 1.3 cm/s using the known α-values and the measured signal-to-noise ratio in acquired images (40).

Laminar flow structure in Fig. 6 shows that the degree of blurring depends on the binary gas mixture – being more pronounced when 3He is diluted in N2 rather than with C3F8. Physically, this is expected since 3He diffusion in N2 (~0.78 cm2/s) is about twice as fast as for 3He in C3F8 (~0.36 cm2/s). Since the diffusion coefficient for HP 129Xe is ~10 times lower (1), it is potentially useful if the goal is to visualize more laminar flow structure in small conduits. Its lower sensitivity relative to 3He, however, would inevitably reduce achievable resolution so some compromises would be required.

Measured and predicted values for apparent axial 3He diffusion are summarized in Fig. 7. Like with apparent flow, close agreement is observed between experimental data and formulated predictions. Images acquired using different Δ values illustrate how axial diffusion is enhanced as more time is allowed for molecules to traverse multiple streamlines. Not surprisingly, the precise enhancement is seen to vary with gas mixture. Less expected is that the apparent diffusion exhibits a laminar-like structure that reaches a maximum about midway between the pipe’s center and wall. Generally, this reflects complex gas dynamics and the behavior of the 2nd order cumulant in Eq. 5.

Figure 7.

Figure 7

Apparent axial 3He gas diffusion during laminar flow in a straight, 3.2-millimeter-diameter tube. Flow conditions are the same as in Fig. 6, and predictions (for Dapp) are based on a second order (n=2) approximation. In each image, the black plot shows the radial profile for axial diffusion across the tube diameter. Uncertainty in measured diffusion is estimated from known b-values and the measured signal-to-noise ratio in acquired images (41). For Δ= 1.0 ms uncertainty is estimated to be ± 0.03 cm2/s, and for Δ= 4.0 ms it is ± 0.15 cm2/s. All plots include a contribution of D that represents the free gas diffusion coefficient in the absence of flow. For 3He in N2 D = 0.78 +/− 0.04 cm2/s, and for 3He in C3F8 D = 0.36 +/− 0.02 cm2/s.

Predicted images of apparent flow velocity in pipes of different size are shown in Fig. 8. These simulated results suggest that laminar structure should be clearly observed in larger pipes with sizes comparable to those in human pulmonary airways. Significant blurring of laminar flow is, however, expected in smaller airways - like those in the major airways of rodents.

Figure 8.

Figure 8

Predicted apparent axial 3He gas flow in tubes of different diameter (2a). Predictions are based on a first order (n=1) approximation (1vapp) using the gradient parameters specified in Fig. 5. Top row from left to right: Navier-Stokes prediction for a straight pipe of any radius, 1vapp for a = 11 mm, and a = 6 mm. Bottom row: 1vapp for a = 3, a = 1.5, and a = 0.5 mm. In each image, the black plot shows the radial profile for axial flow across the tube diameter.

4.3. 2D & 3D Pulmonary Airflow Imaging in Ventilated Rats

Figure 9 shows measured 3He gas flow in 2D constant-time images of rat pulmonary airways. Because a tracheotomy was employed, inhaled gas flowed straight into the rat trachea. Consequently, measured flow in the trachea was similar to that seen in a straight pipe of comparable size. Similarities illustrated in Fig. 9B include a non-zero axial velocity at the edge of the trachea, and a blurred laminar flow profile.

Figure 9.

Figure 9

2D CTI of pulmonary airflow in a 220 g rat. A) Sagittal proton image showing slice locations for 3He flow visualization. B) Left: measured flow speed normal to the slice plane in the rat trachea. B) Right: simulated apparent velocity in a straight pipe of comparable size and volume flow (~2 cc/s). White traces show flow profiles across the center and the bold scale bar represents 1 mm. C) Flow speed normal to the slice plane in the bronchi. All images were acquired using a 3 mm thick slice, a 10° flip angle, a square FOV 1.0 cm on a side, and an 8 ms TR. The color scale shows measured flow speed (cm/s) and uncertainty is estimated to be ± 10 cm/s using established methods (40).

Although some flow structure is evident in the rat trachea, Fig. 9C shows that nothing comparable is observed at the lung’s first bifurcation. Analysis of 3D flow data suggests that this is mainly due to not measuring all the vector flow components. Before this is examined, however, the more obvious is first noted. Namely, that flow at the bifurcation is significantly slower than in the trachea. Physically, this is because mass conservation requires flow to slow when it splits to fill the right and left bronchi. During inhalation, this is expected to occur at every successive bifurcation - as gas moves deeper down the airway tree. A high degree of correlation is therefore anticipated between measured gas speed and airway branching structure.

Figure 10A shows a maximum intensity projection of 3D rat pulmonary airways observed using radial-hybrid 3He imaging. Results provide a detailed view of branching topology that feeds inhaled air to each of the rat’s five lung lobes (42). More remarkable, however, is that all major airways are distinguished in Fig. 10B, where effects on the calculated velocity magnitude resulting from the combination of measured components are shown in full 3D. Flow data were acquired in the same rat; thereby, providing detailed information about both airway structure and inhaled gas dynamics.

Figure 10.

Figure 10

3D imaging of pulmonary structure and flow in a 322 g rat. A) Maximum intensity projection of 3D airway structure visualized with high-resolution, hybrid, radial imaging. Colorized airways highlight main branches feeding the rat lung’s five lobes. Raw image data is zero-filled once along each dimension. B) 3D rendering of velocity magnitude calculated from lower resolution, motion-sensitized, radial-hybrid images. The color scale shows calculated velocity magnitude (cm/s) and uncertainty is estimated to be ± 2 cm/s using established methods (40). Flow results are not rendered using the same software employed for A), and as a consequence, viewing angles are not strictly the same. No zero-fill is employed prior to reconstructing flow images. It is also noted that flow is not analyzed higher up in the trachea because signal drops off more severely due to the dilution of 3He from the extra O2 push used for flow measurements (see section 3.4). C) 3D flow viewed from the same perspective as the sagittal image shown in Fig. 9 A.

Besides flow splitting, gas streaming is also observed in 3D data. Generally, streaming is characterized by a narrow region of faster flow along the airway’s outside radius of curvature. In humans, this has been visualized in 2D images of 3He flow through the trachea (2). Figure 10C shows similar transport in the rat trachea. Presumably, this was not observed in initial 2D experiments because use of a tracheotomy ensured that gas flowed straight into the lung. With oral intubation and the rat in a supine position, gas must travel along a more curved path – as illustrated in Fig. 10C. Slice-by-slice inspection of 3D flow data shows that streaming in the trachea continues down the entire branching structure along the same outside edge. It was also found that streaming was most evident when all the vector components are acquired. Conceptually, this is not surprising since airflow in deeper branches is generally oblique relative to the laboratory frame in which motion-sensitizing gradients are applied. In this context, 3D vector mapping is critical to completely capture actual gas dynamics.

The measured flow speed shown in Figs. 10B & C is a scalar representation of the full vector flow field. Generally, this provides a quantitative view of pulmonary gas delivery not previously seen. Although measured velocity is inevitably blurred, significant slowing due to major airway branching is still observed. Like in straight pipes, it is also expected that measured flow rates in each branch are quantitative. To illustrate this, gas flow (cc/s) was measured in each of the five colored airways shown in Figure 10A. Flow was determined in two steps. First, data in Figure 10A was used for measuring the cross sectional area for each airway feeding a different lung lobe. Gas flow was then calculated using the vector dot product between the measured area, a unit vector normal to the airway’s cross section, and the average gas velocity vector that is measured at the same location from lower resolution data used for generating Figures 10B–C. The results are 0.86±0.04, 0.70±0.07, 0.18±0.01, 0.22±0.03, and 0.21±0.03 cc/s for flow in the left, right caudal, right middle, right cranial, and accessory lobes respectively. The total flow determined from imaging data is therefore 2.17±0.18 cc/s. Prior to imaging, the actual flow rate was measured to be 2.31±0.02 cc/s using a calibrated pneumotach. Total flow measured with MRI is therefore in good agreement (within 6%) with known ventilation settings. Results might therefore provide a quantitative basis for measuring lobar ventilation in pre-clinical animal models. If possible, it may then become feasible to monitor changes with disease, and to relate these to alterations in down stream lung compliance.

Observation of gas streaming in the rat airways generally highlights that some laminar flow structure is discernable. In animal models, this could be useful for understanding the deposition of inhaled materials. In the past, CFD modeling has been used to begin understanding related issues. A full CFD model of the rat respiratory system has, however, never been developed or tested. The work described here represents an integral part of our ongoing efforts to exploit MR imaging for developing and validating CFD models for the rat respiratory tract (6). Current work is aimed at exploiting the analytical techniques described here to predict apparent 3He flow from CFD predictions; thereby, providing a basis for future comparisons in live animals.

5. Conclusions

An analytical description of apparent gas transport accurately predicts measured 3He gas dynamics in straight pipes, and shows that diffusion-mediated effects blur laminar flow structure in rodent-sized airways. Nevertheless, 3D airflow imaging in rodents reveals a detailed picture of lobar airflow not previously seen. This includes flow splitting from airway branching as well as flow streaming along the outside radius of curvature.

Acknowledgments

Research was performed in the Environmental Molecular Sciences Laboratory (a national scientific user facility sponsored by the US Department of Energy’s Office of Biological and Environmental Research) located at Pacific Northwest National Laboratory (PNNL), and operated for DOE by Battelle. Financial support was provided by NIH NHLBI RO1 HL073598. Special thanks to Angie Woodstock (PNNL) for her help with animal handling, and Brian Saam (University of Utah) for 3He polarization cells.

Appendix

Angular contributions to Eq. 20 are generally of the following form –

θ1=02πdθ1θ2=02πdθ2θn=02πdθnCos(j(θ1θ2))Cos(j(θ2θ3))Cos(j(θnθ0)), [A.1]

where j, j′, and j″ represent summation indices from expansions like that defined in Eq. 16. Simplification is achieved by first noting –

Cos(j(θ1θ2))=Cos(jθ1)Cos(jθ2)+Sin(jθ1)Sin(jθ2). [A.2]

Integration of Eq. A.2 over θ1 then gives 2π for j = 0 and is zero otherwise. The series therefore reduces to –

2πθ2=02πdθ2θn=02πdθnCos(j(θ2θ3))Cos(j(θnθ0)). [A.3]

Expanding Cos(j′2 − θ3)) like in Eq. A.2 and integrating over θ2 is then again 2π for j′ = 0 and zero otherwise. Repeating the procedure for each successive Cos term in [A.1] therefore collapses the entire series to give (2π)n if all j’s are 0. For other j-values the integral series is zero.

References

  • 1.Kaiser LG, Logan JW, Meersmann T, Pines A. Dynamic NMR Microscopy of Gas Phase Poiseuille Flow. J. Magn. Reson. 2001;149:144–148. [Google Scholar]
  • 2.de Rochefort L, Maitre X, Fodil R, Vial L, Louis B, Isabey D, Croce C, Darrasse L, Apiou G, Caillibotte G, Bittoun J, Durrand E. Phase-Contrast Velocimetry With Hyperpolarized 3He for In Vitro and In Vivo Characterization of Airflow. Magn. Reson. Med. 2006;55:1318–25. doi: 10.1002/mrm.20899. [DOI] [PubMed] [Google Scholar]
  • 3.Brunner E, Haake M, Kaiser L, Pines A, Reimer JA. Gas Flow MRI Using Circulating Laser-Polarized 129Xe. J. Magn. Reson. 1999;138:155–159. doi: 10.1006/jmre.1998.1675. [DOI] [PubMed] [Google Scholar]
  • 4.Mair RW, Rosen MS, Wang R, Cory DG, Walsworth RL. Diffusion NMR methods applied to xenon gas for materials study. Magn. Reson. Chem. 2002;40:S29–S39. doi: 10.1002/mrc.1114. [DOI] [PubMed] [Google Scholar]
  • 5.de Rochefort L, Vial L, Fodil R, Maitre X, Louis B, Isabey D, Caillibotte G, Thiriet M, Bittoun J, Durand E, Sbirlea-Apiou G. In vitro validation of computational fluid dynamics simulation in human proximal airways with hyperpolarized 3He magnetic resonance phase-contrast velocimetry. J Appl Physiol. 2007;102:2012–2023. doi: 10.1152/japplphysiol.01610.2005. [DOI] [PubMed] [Google Scholar]
  • 6.Minard KR, Einstein DR, Jacob RE, Kabilan S, Kuprat A, Timchalk C, Trease L, Corley RA. Application of Magnetic Resonance (MR) Imaging for the Development and Validation of Computational Fluid Dynamic (CFD) Models of the Rat Respiratory System. Inhalation Toxicology. 2006;18:787–794. doi: 10.1080/08958370600748729. [DOI] [PubMed] [Google Scholar]
  • 7.Koptyug IV, Altobelli SA, Fukushima E, Matveev AV, Sagdeev RZ. Thermally Polarized 1H NMR Microimaging Studies of Liquid and Gas Flow in Monolithic Catalysts. J. Magn. Reson. 2000;147:36–42. doi: 10.1006/jmre.2000.2186. [DOI] [PubMed] [Google Scholar]
  • 8.Newling B, Poirier CC, Zhi Y, Rioux JA, Coristine AJ, Roach D, Balcom BJ. Velocity Imaging of Highly Turbulent Gas Flow. Phys. Rev. Letters. 2004;93(15) doi: 10.1103/PhysRevLett.93.154503. 154503-1 – 4. [DOI] [PubMed] [Google Scholar]
  • 9.Brandl M, Haase A. Molecular Diffusion in NMR Microscopy. J. Magn. Reson. 1994;103 B:162–167. [Google Scholar]
  • 10.Callaghan PT, Eccles CD. Diffusion-limited resolution in nuclear magnetic resonance microscopy. J. Magn. Reson. 1988;78:1–8. [Google Scholar]
  • 11.Ahn ZH, Cho CB. A generalized formulation of diffusion effects in micron resolution nuclear magnetic resonance imaging. Med. Phys. 1989;16(1):22–28. doi: 10.1118/1.596393. [DOI] [PubMed] [Google Scholar]
  • 12.Gravina S, Cory DG. Sensitivity and Resolution of Constant-Time Imaging. J. Magn. Reson. 1994;104, Series B:53–61. [Google Scholar]
  • 13.Choi S, Tang X-W, Cory DG. Constant Time Imaging Approaches to NMR Microscopy. Int. J. of Imaging Systems and Techn. 1997;8:263–276. [Google Scholar]
  • 14.Prado PJ, Balcom BJ, Mastikhin IV, Cross AR, Armstrong RL, Logan A. Magnetic Resonance Imaging of Gases: A Single-Point Ramped Imaging with T1 Enhancement (SPRITE) Study. J. Magn. Reson. 1999;137:324–332. doi: 10.1006/jmre.1998.1674. [DOI] [PubMed] [Google Scholar]
  • 15.Minard KR, Timchalk C, Corley RA. T2-Shortening of 3He gas by Magnetic Microspheres. J. Magn. Reson. 2005;173:90–96. doi: 10.1016/j.jmr.2004.11.026. [DOI] [PubMed] [Google Scholar]
  • 16.Stepisnik J. Measuring and imaging of flow by NMR. Progress in NMR Spectroscopy. 1985;17:187–209. [Google Scholar]
  • 17.Caprihan A, Fukushima E. Flow Measurements by NMR. Physics Reports. 1990;198(4):195–235. [Google Scholar]
  • 18.Pope JM, Yao S. Quantitative NMR of imaging of flow. Concepts Magn. Reson. 1993;5:381–302. [Google Scholar]
  • 19.Callaghan PT. Principles of Nuclear Magnetic Resonance Microscopy. New York: Oxford University Press Inc.; 1993. [Google Scholar]
  • 20.Stepisnik J. Validity limits of Gaussian approximation in cumulant expansion for diffusion attenuation of spin echo. Physica. 1999;B 270:110–117. [Google Scholar]
  • 21.Frøhlich AF, Østergaard L, Kiselev VG. Effect of impermeable boundaries on diffusion-attenuated MR signal. J. Magn. Reson. 2006;179:223–233. doi: 10.1016/j.jmr.2005.12.005. [DOI] [PubMed] [Google Scholar]
  • 22.Kubo R. Generalized Cumulant Expansion Method. Journal of the Physical Society of Japan. 1962;17:1100–1120. [Google Scholar]
  • 23.Kubo R. A Stochastic Theory of Line-Shape and Relaxation. In: ter Haar D, editor. Fluctuation, Relaxation, and Resonance in Magnetic Systems. London: Oliver and Boyd; 1962. pp. 23–68. [Google Scholar]
  • 24.Freed JH. Generalized Cumulant Expansions and Spin-Relaxation Theory. J. Chem. Phys. 1968;49:376–391. [Google Scholar]
  • 25.McLennan JA. Introduction to Non-Equilibrium Statistical Mechanics. New Jersey: Prentice Hall; 1989. [Google Scholar]
  • 26.Kuethe DO. Measuring distributions of diffusivity in turbulent fluids with magnetic resonance imaging. Phys Rev A. 1989;40(8):4542–4552. doi: 10.1103/physreva.40.4542. [DOI] [PubMed] [Google Scholar]
  • 27.Landau LD, Lifshitz EM. Fluid Mechanics. Pergamon Press; London: 1959. [Google Scholar]
  • 28.Bird RB, Stewart WE, Lightfoot EN. Transport Phenomena. Second Edition. New York: John Wiley & Sons Inc.; 2002. [Google Scholar]
  • 29.Sukstanskii AL, Yablonskiy DA. Effects of Restricted Diffusion on MR Signal Formation. J. Magn. Reson. 2002;157:92–105. doi: 10.1006/jmre.2002.2582. [DOI] [PubMed] [Google Scholar]
  • 30.Weng MC, Uhlenbeck GE. On the Theory of the Brownian Motion II. Reviews of Modern Physics. 1945;17:323–342. [Google Scholar]
  • 31.Jacob RE, Laicher G, Minard KR. 3D MRI of non-Gaussian 3He gas diffusion in the rat lung. J. Magn. Reson. 2007;188(2):357–366. doi: 10.1016/j.jmr.2007.08.014. [DOI] [PubMed] [Google Scholar]
  • 32. http://encyclopedia.airliquide.com/encyclopedia.asp.
  • 33.Chen XJ, Moller HE, Chawla MS, Cofer GP, Driehuys B, Hedlund LW, Johnson GA. Spatially Resolved Measurments of Hyperpolarized Gas Properties in the Lung In Vivo. Part I: Diffusion Coefficient. Magn. Reson. Med. 1999;42:721–728. doi: 10.1002/(sici)1522-2594(199910)42:4<721::aid-mrm14>3.0.co;2-d. [DOI] [PubMed] [Google Scholar]
  • 34.Weast RC, editor. CRC Handbook of Chemistry and Physics. Cleveland, Ohio: CRC Press; 1974. [Google Scholar]
  • 35.Kingsley PB. Introduction to Diffusion Tensor Imaging Mathematics: Part II. Anisotropy, Diffusion-Weighting Factors, and Gradient Encoding Schemes. Concepts in Magnetic Resonance Part A. 2006;28:123–154. [Google Scholar]
  • 36.Hedlund LW, Cofer GP, Owen SJ, Johnson GA. MR-compatible ventilator for small animals: computer-controlled ventilation for proton and noble gas imaging. Magn. Reson. Imaging. 2000;18:753–759. doi: 10.1016/s0730-725x(00)00154-5. [DOI] [PubMed] [Google Scholar]
  • 37.Dugas JP, Garbow JR, Kobayashi DK, Conradi MS. Hyperpolarized He-3 MRI of mouse lung. Magn. Reson. Med. 2004;52:1310–1317. doi: 10.1002/mrm.20300. [DOI] [PubMed] [Google Scholar]
  • 38.Lanz T, Ruff J, Weisser A, Haase A. Double tuned Na-23 H-1 nuclear magnetic resonance birdcage for application on mice in vivo. Rev. Sci. Instrum. 2001;72:2508–2510. [Google Scholar]
  • 39.Minard KR, Wind RA, Phelps RL. A Compact Respiratory-Triggering Device for Routine Microimaging of Laboratory Mice. JMRI. 1998;8:1343–1348. doi: 10.1002/jmri.1880080626. [DOI] [PubMed] [Google Scholar]
  • 40.Anderson AH, Kirsch JE. Analysis of noise in phase contrast MR imaging. Med. Phys. 1996;23:857–869. doi: 10.1118/1.597729. [DOI] [PubMed] [Google Scholar]
  • 41.O’Halloran RL, Holmes JH, Altes TA, Salerno M, Fain SB. The effects of SNR on ADC measurements in diffusion-weighted hyperpolarized He-3 MRI. J. Magn. Reson. 2007;185:42–49. doi: 10.1016/j.jmr.2006.11.006. [DOI] [PubMed] [Google Scholar]
  • 42.Popesko P, Rajtova V, Horak J. Anatomy of Small Laboratory Animals. Volume Two: Rat, Mouse, Hamster. Wolfe Publishing; London: 1992. [Google Scholar]

RESOURCES