Abstract
Background
MRI is not able to directly measure the concentration of contrast agent. It is measured indirectly from the signal intensity (SI). It is very important to know how much contrast agent should be injected to receive a maximum SI in the region of interest (ROI).
Objectives
The aim of this study was to investigate the maximum relationship between contrast concentration and signal intensity (SI) on T1-weighted images using spin echo (SE), fast spin echo (FSE) and inversion recovery (IR) sequences.
Materials and Methods
To assess the relationship between SI and concentration, a water-filled phantom containing vials of different concentrations of gadolinium DTPA (Gd-DTPA) (0 to 19.77 mmol/L) or a constant concentration (1.2 mmol/L) of contrast agent was used. The vials of constant concentration were used to measure coil nonuniformity. The mean SI was obtained in the ROI using T1-weighted images. All studies were carried out using a 0.3 T clinical MR scanner with a standard head coil.
Results
This study shows that maximum SI will appear at different ranges in different sequences. The maximum SI can be seen at concentrations of 5.95, 4.96 and 3.98mmol/L for SE, FSE and IR, respectively.
Conclusion
Using standard imaging parameters, each MRI sequence reaches its maximum SI in a specific contrast concentration, which is highest in SE and least in IR in a comparison between SE, FSE and IR sequences.
Keywords: Magnetic Resonance Imaging, Relaxation, Gadolinium DTPA
1. Background
For improved contrast of MR images, paramagnetic-metal contrast agents are administered. While paramagnetic contrast medium passes through the tissue, it produces a local magnetic field inhomogeneity that leads to reduction in the transverse relaxation time (T2) and longitudinal relaxation time (T1) of the tissue. Decrease in T1 typically causes an increase in signal intensity (SI); whereas, decrease in T2 causes a decrease in SI (1). The T1-shortening effect is dominant at low concentrations of Gd-DTPA and the T2-shortening effect is dominant at high contrast concentrations and leads to a decrease in the SI. The net result of these effects on the MR signal intensity will depend on the image parameters and the type of imaging sequence (2, 3). The concentration of contrast agent in MRI is measured indirectly from SI. It is very important to know how much contrast agent should be injected to obtain a maximum SI in the region of interest (ROI). Increasing the concentration of contrast agent leads to a decrease in the SI; therefore, the obtained amount of optimal injected dose should be considered (1). Many studies have accepted the use of 0.1 mmol/kg of body weight injection in different T1-weighted sequences with different image parameters and MRI unit strength (4-10) Some other investigators performed different sequences with the use of high-dose contrast administration such as 0.2 mmol/kg of body weight (11, 12).
2. Objectives
The aim of this study was to investigate the concentration which leads to maximum signal intensity (SI) on T1-weighted images in spin-echo (SE), fast-spin-echo (FSE) and inversion recovery (IR) sequences with standard image parameters of the MR scanner.
3. Materials and Methods
3.1. Basics
Many factors such as the magnetic field strength, the pulse sequence parameters, the dose of the contrast agent, the injection rate and bolus volume, and the tissue topology may affect the SI (13-15). The MR sequence can influence the relationship between T1 and SI, which in turn is dependent on the contrast concentration (16). The relationship between SI and imaging parameters in the IR sequence can be written as:
Where S(t) is the signal intensity after administration of the contrast agent, and S0 is the observed signal intensity when no magnetization preparation pre-pulses are applied or there is no contrast agent. TI and TR denote the inversion time and the repetition time, respectively. T1, TE, and T2 denote the tissue longitudinal relaxation time, the echo time and transverse relaxation time, respectively (3, 17, 18). Equation (1) with a concentration of contrast agent at time t (C (t)) can be described as (3):
Equation 1.
Where T1 Pre is the longitudinal relaxation times before contrast application and K is a constant that depends on the contrast medium used (19). For saturation recovery (SE) sequences, the relationship between SI and image parameters and the concentration of contrast agent has been defined as:
Where TS is the saturation time. At lower concentrations of the contrast agent, the term exp[-TE/T2] can be ignored in equations number 1, 2 and 3 (16, 18).
Equation 2.
Equation 3.
3.2. Calculation of Injection Volume of Contrast Agent
The amount of the injected dose required to establish a known concentration of the contrast agent in the ROI of the brain has been reported by Moody et al. (20):
X is concentration of the contrast agent (mmol/L) in the ROI. BSA (m2) is the body surface area. BSA can be described as (21).
Equation 5.
3.3. Phantom
To find the relationship between the SI and concentration in different sequences, we designed a phantom to hold vials which contained either various or constant concentrations of the contrast agent (Figure 1). The vials of constant concentration (1.2 mmol/L) were used for measurement of coil non-uniformity (1, 3). Vials with different concentrations were used for measurement of the relationship between SI and concentration. A standard clinical head coil was used with the phantom. The vials were set vertically, and the axes of the vials were perpendicular to the image plane (Figure 2, coronal image). The phantom consisted of 25 vials containing 22 different contrast concentrations (glass tube, inner diameter approximately 15 mm filled with different concentrations of Gd-DTPA (Magnevist, Schering Health Care Ltd, West Sussex, UK). The concentration of Gd-DTPA ranged from 0 to 19.77 mmol/L (0.00, 0.30, 0.45, 0.60, 0.75, 0.90, 1.20, 1.50, 1.80, 2.10, 2.39, 2.69, 2.99, 3.28, 3.58, 3.98, 4.96, 5.95, 7.93, 9.90, 13.85 and 19.77 mmol/L).
Figure 1. The phantom.

Figure 2. Coronal image of the phantom containing vials with different contrast concentrations.
Two experiments were performed, one using vials with different concentrations and one using vials with constant concentration. The vials in the phantom with a constant concentration were placed in exactly the same positions of the vials with different concentrations.
The non-uniformity of the coil was calculated from the SI of each vial with constant concentration and it was normalized to give a correction factor. For calculating the corrected SI for different concentrations, the SI of each vial was multiplied by its correction factor (22).
3.4. Image Acquisition
All studies were carried out with use of a 0.3 T clinical MR scanner (Hitachi Medical Corporation, AIRIS). The standard imaging parameters for constant and different concentrations were TR = 411 ms, TE = 15 ms, pixel size = 2×2 mm 2, matrix size =128×128 , slice thickness = 10 mm, flip angle = 90° for SE; TR = 816 ms, TE = 15 ms, pixel size = 2×2 mm 2, matrix size = 128×128, slice thickness = 10 mm, flip angle = 90°, Echo Train Length = 2 for FSE; and TR = 506 ms, TE = 15 ms, TI = 20 ms, pixel size = 2×2 mm 2, matrix size 128×128 slice thickness = 10 mm 2, flip angle = 90° for IR.
3.5. Image Analysis
After transferring the image data from the MR scanner to a personal computer, the image processing software Interactive Data Language (IDL, Research Systems, Inc. NY, USA; http://www.rsinc.com) was used for processing. Special programs were written to find the following:
- To measure the mean SI of the nine innermost pixels out of the total number of 44 pixels of each vial, to avoid partial volume effects.
- The correction factors of the non-uniformity of the coils from the SI of the vials with constant concentration. Then the SI of the vials with different concentration was multiplied by these factors to find the corrected SI.
- To draw the concentration versus the SI curve to measure the maximum value of the concentration of the contrast agent that gives a maximum SI.
These programs could be run from either a UNIX workstation or a personal computer.
4. Results
Figures 3, 4 and 5 show the maximum relationship between SI and concentration in SE, FSE, and IR sequences. We detected maximum SI at concentrations of 5.95, 4.96 and 3.98 mmol/L for SE, FSE, and IR, respectively. The linear relationship between the concentrations and corrected SI that gave a 0.95 coefficient of determination (R2) was at 2.02, 1.18, and 1.99 mmol/L for SE, FSE, and IR, respectively. In addition, the figures show that at concentrations higher than the above, the SI will decrease. The error bars show the standard deviation of the SI from the nine innermost pixels in each vial. For example, 0.12, 0.10, and 0.08 mmol/kg of body weight (for an average body, i.e., height = 175 cm, weight = 85 kg) of contrast agent should be injected for SE, FSE, and IR sequences, respectively to give a maximum SI in the region of interest based on equation number 4. Higher doses of contrast will cause a decrease in SI.
Figure 3. Mean corrected (for non-uniformity of the coil) SI from the nine innermost pixels of the vials versus concentration of contrast agent using SE. The maximum SI (22169) can be seen at a concentration of 5.95 mmol/L. The linear relationship between concentrations and corrected SI that gave an R2 of 0.95 was at 2.02 mmol/L. The error bars show the standard deviation of the SI in each vial.
Figure 4. Mean corrected (for non-uniformity of the coil) SI from the nine innermost pixels of the vials versus concentration of contrast agent using FSE. The maximum SI (14218) can be seen at a concentration of 4.96 mmol/L. The figure illustrates that a linear relationship between SI and concentration reaches 1.18 mmol/L (R2 = 0.95). The error bars show the standard deviation of the SI in each vial.
Figure 5. Mean corrected (for non-uniformity of the coil) SI from the nine innermost pixels of the vials versus concentration of contrast agent using IR. The maximum SI (17038) can be seen at a concentration of 3.98 mmol/L. The linear relationship between concentration and corrected SI resulting in an R2 of 0.95 was at 1.99 mmol/L. The error bars show the standard deviation of the SI in each vial.
Equation 4.
5. Discussion
The concentration of a contrast agent such as Gd-DTPA in MRI has essentially no linear correlation with the SI. The contrast agent induces both T1- and T2- shortening effects, which lead to opposing effects on the SI. T1- shortening will increase the SI and T2- shortening will decrease it. As mentioned in equations number 2 and 3, both T1 and T2 may affect the SI at high concentrations. The T2-shortening effect becomes dominant at high concentrations and leads to a decreasing SI. The effect of T1-shortening is significant when the concentration of contrast agent is low (17-19). Melhem et al. (4) compared contrast-to-noise ratios between FSE (546/10 (TR/effective TE, echo-train length = 4) and SE (546/10, TR/TE) in T1-weighted MR sequences of 32 enhancing brain lesions. MR images were obtained at 1.5 T after IV administration of 0.10 mmol/kg gadopentetate dimeglumine. Sugahara et al. (5) evaluated the T1-FSE sequence (690/12 (TR/TE effective); echo train length = 3) to determine whether this technique can replace the conventional T1-weighted spin-echo sequence (690/14 (TR/TE)) for routine contrast-enhanced imaging. They obtained the images after intravenous administration of 0.1 mmol/kg gadopentetate dimeglumine with a 1.5 T magnet.
Qian et al. (6) used T1 IR and T1 SE sequences for detecting brain metastases. For contrast enhancement, they administered gadopentetate dimeglumine (Magnevist) with the standard dose of 0.1 mmol/kg of body weight. Kizildağ et al. (7) used T1-weighted (T1W), FSE T2-weighted (FSE T2W) and fluid-attenuated inversion recovery (FLAIR) sequences to assess the features of normal brain development in terms of myelination in infants and young children on a 1 Tesla MR unit with administration of 0.1 mg/Kg of Dormicum. Zhou et al. (8) injected 0.1 mmol/kg of body weight of contrast agent (Gd-DTPA) to evaluate intracranial tumors by SE (TR/TE = 440/14 ms) T1-weighted images, using a 1.5 MRI system.
Tomura et al. (9) used SE ((TR/TE) (300-540/8-14 ms, 1.5 T) and FSE FLAIR sequences to compare brain tumors by injecting 0.1 mmol/Kg of body weight of gadopentetate dimeglumine.
Alibek et al. (10) performed SE ((TR/TE) (470/17 ms, 1.5 T) and FLAIR (TI/TR (860/1720 ms)) sequences by injecting 0.1 mmol/Kg body weight of gadolinium contrast agent to detect brain lesions in an unsedated pediatric patient with a 1.5 T MRI system. Al-Saeed et al. (11) performed FLAIR (relaxation/repetition time/time interval (TI)/number of excitations, 1920/8.2/750/2) and FSE (relaxation/repetition time/echo time/number of excitations, 600/8.1/2) T1-weighted sequences for comparison of 20 patients with brain lesions. They administered 0.2 mmol/Kg of gadolinium chelate for each patient and used a 1.5 T MRI system. Kakeda et al. (12) compared SE (TR/TE = 520/9 ms), IR-FSE (TR/TE/TI = 2500/9.1/1000 ms, echo train length = 7), and 3D-gradient echo (GRE) sequences to detect brain metastases at 3 T scanner. They administered 0.2 mmol/kg of body weight gadoteridol for all patients. They reported that a 0.2 mmol/L of contrast agent concentration gave the maximum SI in SE and IR-FSE sequences in a phantom study using a 3 T MRI system. Our study showed that 5.95, 4.96, and 3.98 mmol/L of contrast agent concentration for SE, FSE, and IR sequences, respectively, gave a maximum SI in the region of interest. The difference between the results of this study and that of Kakeda et al. (12) may be due to differences in the use of different TR, TE, and MRI strengths. Different image parameters and MRI strengths can have an effect on the maximum relationship between SI and contrast agent concentration (1, 13, 14). Many studies have been performed without considering imaging parameters, imaging sequences and strength of the MRI system, with injection of 0.1 or 0.2 mmol/kg of body weight of contrast agent. Our study showed that the maximum concentration which leads to maximum SI depends on the image sequence. At higher doses as mentioned in the results, there was not only an increased SI, but also a decreased SI (Figures 3, 4 and 5).
This study shows that maximum SI will appear at different concentrations when different sequences are used with standard imaging parameters. A maximum SI can be obtained at high concentrations using the SE sequence, but not FSE and IR sequences. In addition, a maximum SI can be seen at higher concentrations for the FSE sequence and not in IR sequences. The results of this study can be used in clinical studies in the future.
Acknowledgments
We thank Tabriz University of Medical Sciences for their valuable support.
Footnotes
Implication for health policy/practice/research/medical education: The article may help radiologists and clinicians use the best sequences for clinical studies.
Please cite this paper as: Nazarpoor M, Poureisa M, Daghighi M.H. Comparison of Maximum Signal Intensity of Contrast Agent on T1-Weighted Images Using Spin Echo, Fast Spin Echo and Inversion Recovery Sequences. Iran J Radiol. 2013;10(01):27-32. DOI: 10.5812/iranjradiol.5452
Authors’ Contribution: None declared.
Financial Disclosure: None declared.
Funding/Support: None declared.
References
- 1.Nazarpoor M. The effect of repetition time on the maximum linear relationship between contrast agent concentration and signal intensity on T1 weighted image using inversion recovery (IR) sequence. IJR. 2009;6(4):247–52. [Google Scholar]
- 2.Martel AL, Moody AR, Allder SJ, Delay GS, Morgan PS. Extracting parametric images from dynamic contrast-enhanced MRI studies of the brain using factor analysis. Medical Image Analysis. 2001;5(1):29–39. doi: 10.1016/S1361-8415(00)00032-3. [DOI] [PubMed] [Google Scholar]
- 3.Nazarpoor M. Effects of inversion and saturation times on relationships between contrast agent concentrations and signal intensities of T1-weighted magnetic resonance images. Radiol Phys Technol. 2010;3(2):120–6. doi: 10.1007/s12194-010-0087-9. [DOI] [PubMed] [Google Scholar]
- 4.Melhem ER, Guidone PL, Jara H, Yucel EK. Improved contrast of enhancing brain lesions using contrast-enhanced T1-weighted fast spin-echo MR imaging. AJR Am J Roentgenol. 1997;168(4):1091–5. doi: 10.2214/ajr.168.4.9124121. [DOI] [PubMed] [Google Scholar]
- 5.Sugahara T, Korogi Y, Ge Y, Shigematsu Y, Liang L, Yoshizumi K, et al. Contrast enhancement of intracranial lesions: conventional T1-weighted spin-echo versus fast spin-echo MR imaging techniques. AJNR Am J Neuroradiol. 1999;20(8):1554–9. [PMC free article] [PubMed] [Google Scholar]
- 6.Qian YF, Yu CL, Zhang C, Yu YQ. MR T1-weighted inversion recovery imaging in detecting brain metastases: could it replace T1-weighted spin-echo imaging? AJNR Am J Neuroradiol. 2008;29(4):701–4. doi: 10.3174/ajnr.A0907. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kizildag B, Dusunceli E, Fitoz S, Erden I. The role of classic spin echo and FLAIR sequences for the evaluation of myelination in MR imaging. Diagn Interv Radiol. 2005;11(3):130–6. [PubMed] [Google Scholar]
- 8.Zhou ZR, Shen TZ, Chen XR, Peng WJ. Diagnostic value of contrast-enhanced fluid-attenuated inversion-recovery MRI for intracranial tumors in comparison with post-contrast T1W spin-echo MRI. Chin Med J (Engl). 2006;119(6):467–73. [PubMed] [Google Scholar]
- 9.Tomura N, Kato K, Takahashi S, Sashi R, Izumi J-i, Narita K, et al. Multi-shot echo-planar Flair imaging of brain tumors: comparison of spin-echo T1-weighted, fast spin-echo T2-weighted, and fast spin-echo Flair imaging. Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society. 2002;26(2):65–72. doi: 10.1016/s0895-6111(01)00039-8. [DOI] [PubMed] [Google Scholar]
- 10.Alibek S, Adamietz B, Cavallaro A, Stemmer A, Anders K, Kramer M, et al. Contrast-enhanced T1-weighted Fluid-attenuated Inversion-recovery BLADE Magnetic Resonance Imaging of the Brain: An Alternative to Spin-echo Technique for Detection of Brain Lesions in the Unsedated Pediatric Patient? Academic radiology. 2008;15(8):986–95. doi: 10.1016/j.acra.2008.03.009. [DOI] [PubMed] [Google Scholar]
- 11.Al-Saeed O, Ismail M, Athyal RP, Rudwan M, Khafajee S. T1-weighted fluid-attenuated inversion recovery and T1-weighted fast spin-echo contrast-enhanced imaging: a comparison in 20 patients with brain lesions. J Med Imaging Radiat Oncol. 2009;53(4):366–72. doi: 10.1111/j.1754-9485.2009.02093.x. [DOI] [PubMed] [Google Scholar]
- 12.Kakeda S, Korogi Y, Hiai Y, Ohnari N, Moriya J, Kamada K, et al. Detection of brain metastasis at 3T: comparison among SE, IR-FSE and 3D-GRE sequences. Eur Radiol. 2007;17(9):2345–51. doi: 10.1007/s00330-007-0599-9. [DOI] [PubMed] [Google Scholar]
- 13.Nazarpoor M, Moody AR, Martel AL, Morgan PS, editors. The relationship between contrast agent concentration and SI on T1 weighted images for measuring perfusion with MRI, 20th Annual Scientific Meeting of Society of Magnetic Resonance in Medicine and Biology;.; 2003; Rotterdam, NL. pp. S243–4. [Google Scholar]
- 14.Nazarpoor M, Mayabi Z, Shfaie A, Pesianian E, Aghaverdizadeh D. Maximum Relationship between Signal Intensity and Concentration of Contrast Agent in 0.3 T and 1.5 T using T1-weighted Spin Echo Sequence. Med J Tarbiat Univ Med Sci Health Serv. 2011;32(6):72–6. [Google Scholar]
- 15.Matt A. Bernstein PD, King KF, Zhou XJ. Handbook of MRI Pulse Sequences: A Guide for Scientists, Engineers, Radiologists, Technologists. Academic Press; 2004. [Google Scholar]
- 16.Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann HJ. Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest Radiol. 2005;40(11):715–24. doi: 10.1097/01.rli.0000184756.66360.d3. [DOI] [PubMed] [Google Scholar]
- 17.Kubota T, Yamada K, Kizu O, Hirota T, Ito H, Ishihara K, et al. Relationship between contrast enhancement on fluid-attenuated inversion recovery MR sequences and signal intensity on T2-weighted MR images: visual evaluation of brain tumors. J Magn Reson Imaging. 2005;21(6):694–700. doi: 10.1002/jmri.20331. [DOI] [PubMed] [Google Scholar]
- 18.McRobbie A. Vulnerability, violence and (cosmopolitan) ethics: Butler's Precarious Life. Br J Sociol. 2006;57(1):69–86. doi: 10.1111/j.1468-4446.2006.00094.x. [DOI] [PubMed] [Google Scholar]
- 19.Roberts TP. Physiologic measurements by contrast-enhanced MR imaging: expectations and limitations. J Magn Reson Imaging. 1997;7(1):82–90. doi: 10.1002/jmri.1880070112. [DOI] [PubMed] [Google Scholar]
- 20.Moody AR, Martel A, Kenton A, Allder S, Horsfield MA, Delay G, et al. Contrast-reduced imaging of tissue concentration and arterial level (CRITICAL) for assessment of cerebral hemodynamics in acute stroke by magnetic resonance. Invest Radiol. 2000;35(7):401–11. doi: 10.1097/00004424-200007000-00002. [DOI] [PubMed] [Google Scholar]
- 21.Mosteller RD. Simplified calculation of body-surface area. N Engl J Med. 1987;317(17):1098. doi: 10.1056/NEJM198710223171717. [DOI] [PubMed] [Google Scholar]
- 22.Nazarpoor M. Effect of concentration of contrast agent on the inflow effect for measuring absolute perfusion by use of inversion recovery T(1)-weighted TurboFLASH images. Radiol Phys Technol. 2012;5(1):86–91. doi: 10.1007/s12194-011-0140-3. [DOI] [PubMed] [Google Scholar]