Skip to main content
Tissue Engineering. Part C, Methods logoLink to Tissue Engineering. Part C, Methods
. 2012 Dec 21;19(6):427–437. doi: 10.1089/ten.tec.2012.0021

A Novel Internal Fixator Device for Peripheral Nerve Regeneration

Ting-Hsien Chuang 1,2,*, Robin E Wilson 3,*, James M Love 1, John P Fisher 1, Sameer B Shah 1,2,
PMCID: PMC3629849  PMID: 23102114

Abstract

Recovery from peripheral nerve damage, especially for a transected nerve, is rarely complete, resulting in impaired motor function, sensory loss, and chronic pain with inappropriate autonomic responses that seriously impair quality of life. In consequence, strategies for enhancing peripheral nerve repair are of high clinical importance. Tension is a key determinant of neuronal growth and function. In vitro and in vivo experiments have shown that moderate levels of imposed tension (strain) can encourage axonal outgrowth; however, few strategies of peripheral nerve repair emphasize the mechanical environment of the injured nerve. Toward the development of more effective nerve regeneration strategies, we demonstrate the design, fabrication, and implementation of a novel, modular nerve-lengthening device, which allows the imposition of moderate tensile loads in parallel with existing scaffold-based tissue engineering strategies for nerve repair. This concept would enable nerve regeneration in two superposed regimes of nerve extension—traditional extension through axonal outgrowth into a scaffold and extension in intact regions of the proximal nerve, such as that occurring during growth or limb-lengthening. Self-sizing silicone nerve cuffs were fabricated to grip nerve stumps without slippage, and nerves were deformed by actuating a telescoping internal fixator. Poly(lactic co-glycolic) acid (PLGA) constructs mounted on the telescoping rods were apposed to the nerve stumps to guide axonal outgrowth. Neuronal cells were exposed to PLGA using direct contact and extract methods, and they exhibited no signs of cytotoxic effects in terms of cell morphology and viability. We confirmed the feasibility of implanting and actuating our device within a sciatic nerve gap and observed axonal outgrowth following device implantation. The successful fabrication and implementation of our device provides a novel method for examining mechanical influences on nerve regeneration.

Introduction

Peripheral nerve injury may result from trauma, cancer, or congenital defects,13 with the severity of injury categorized by the degree and reversibility of structural changes to the nerve.4,5 Recovery from peripheral nerve damage, especially for a transected nerve, is rarely complete, resulting in impaired motor function, sensory loss, and chronic pain with inappropriate autonomic responses that may seriously impair quality of life. It is estimated that more than 50,000 peripheral nerve repair procedures are annually performed in the United States alone, imposing a financial burden of 7 billion dollars per year.2 In consequence, strategies for enhancing peripheral nerve repair are of high clinical importance.

Short gaps (<10 mm) may be readily bridged by surgical reconnection of stumps6 or through a variety of autologous and nonautologous nerve guidance channels (NGCs).3,714 Autologous nerve grafts remain the gold standard for repairing longer nerve gaps, but are in limited supply, exhibit donor site morbidity, and may exhibit a size mismatch compared to the transected nerve. Nonautologous grafts can incorporate biological or synthetic components and have been designed with increasing complexity. Biological grafts include acellular nerve grafts or use skeletal muscle,15,16 vein1719, and tendon,20 while synthetic NGCs have been fabricated with a variety of geometries, porosities, and material properties.3,21 Grafts have also been designed to incorporate chemical and biological cues to encourage nerve regeneration, such as Schwann cells, stem cells, or neurotrophic factors.2225 Despite such developments, engineered NGCs do not perform as well as autografts, particularly for large defect sizes, when degeneration is more likely to outpace neuronal extension.26 Therefore, there is still a need to develop a strategy for repairing large gaps (>10 mm).

Tensile loading is an influence on nerve growth and function that has been underutilized in the context of nerve regeneration. Nerves bear tension under several physiological scenarios. Peripheral nerves exist under tension, releasing strains of up to 11% following transection,2729 and may deform additionally, in some cases upward of 20%, during joint movement.3032 During growth, axonal tension serves as a survival and stabilization signal for axons. Conversely, neurites under no tension retract.33 It has also been shown that moderate tensile loads can accelerate neuronal growth, both in vitro and in vivo.3439

Toward the development of more effective nerve regeneration strategies, in this study, we detail the design, fabrication, and preliminary implementation of a novel internal fixator device. This modular device, by imposing mechanical loads on regenerating nerves in parallel with existing tissue engineering strategies for nerve repair, facilitates two regimes of nerve extension—traditional extension through axonal outgrowth into an engineered scaffold and novel extension of intact regions of the proximal nerve (Fig. 1a).

FIG. 1.

FIG. 1.

Device design. The nerve stretching device is composed of spiral nerve cuffs, poly(lactic co-glycolic) acid (PLGA) nerve guidance channel (NGC), and stainless steel backbone (a) Two regions of regeneration, A: Enhanced axonal outgrowth into a tissue engineered nerve guide/scaffold; B: Lengthening of intact regions of nerve stumps–cf. limb lengthening. (b) Dimensions of the device. Color images available online at www.liebertpub.com/tec

Materials and Methods

Animals

We used adult 14-week-old male Sprague-Dawley rats (350–400 g), based on the use of this strain in a number of nerve regeneration studies and its well-characterized sciatic nerve architecture.40, 41 Animal use protocols were approved by the UMCP and UCSD Institutional Animal Care and Use Committees.

Device fabrication

The device is composed of three major components: self-sizing silicone spiral nerve cuffs, poly(lactic co-glycolic) acid (PLGA) NGCs, and a mechanical backbone to which the cuffs and NGCs are attached. Dimensions for the device used with rat sciatic nerves are provided in Figure 1b.

Spiral nerve cuffs

To impose tensile loading on the transected nerve, self-sizing spiral nerve cuffs were created to firmly grip the proximal and distal nerve stumps, without exerting excessive compression. We modified a protocol for the fabrication of spiral cuff electrodes42; however, no electrode was embedded in our cuffs (Fig. 2a). Two Silastic® silicone sheets (Dow Corning) of 0.005 inches in thickness were bound together using a silicone adhesive (biomedical grade Silastic® elastomer MDX4-4210; Dow Corning). The first sheet was unstretched, and placed adjacent to one stainless steel slab. The second sheet was stretched to a specified strain (50%, 60%, 70%, or 100% strain; cf. Fig. 2d), and attached to a second grooved slab via an additional layer of silicone adhesive, to create a textured surface on the interior surface of the cuff that would increase friction between the cuff and nerve stump. The sandwiched silicone composite was cured at 60°C for 2 h. After curing, the resultant spiral sheet was cut to a length of 0.5 cm, to enable self-sizing of approximately one and a quarter spirals around the sciatic nerve of adult male Sprague-Dawley rats (350–400 g).

FIG. 2.

FIG. 2.

Fabrication and characterization of spiral nerve cuff. (a) Two layers of silicone sheets (with one being stretched) are glued together; (c) the microgroove pattern on the slab is transferred to the cuff; (b) the resultant curled nerve cuff after curing; (d) inner diameter decreases with increasing % strain of prestretched silicone sheet; (e) sample ex vivo testing of trypan blue-labeled nerve in cuff revealed no appreciable compression or tethering. Color images available online at www.liebertpub.com/tec

Nerve guidance channel

Nerve guides were prepared using a modified phase inversion technique.43 PLGA (lactic to glycolic acid mol ratio of 75:25, Mw=66,000–107,000; Sigma) was dissolved in tetraglycol (Sigma) at 60°C (10 wt%) and then Pluronic® F-127 was added (3 wt%) to increase the hydrophilicity. Alginate hydrogel rods were formed by first dissolving alginate (Sigma) in water (4 wt%) and injecting the solution into 2% CaCl2 with a syringe (14-gauge needle). After saturation, the alginate hydrogel was immersed in the PLGA/Pluronic F-127 solution. Due to phase separation between polymer (PLGA/F127) and nonsolvent (water in alginate hydrogel), PLGA precipitated onto the alginate rod as water diffused out of the hydrogel. The construct was thoroughly washed in water for 24 h to remove excess tetraglycol. After retrieval of the alginate rod; the resultant PLGA tubes were dried at room temperature and cut to the desired length.

Mechanical backbone

A stainless steel rod of 1.57 mm in diameter (Component Supply Co.) was used as the backbone of the device. This material was selected due to its strength and previous applications in implanted devices. The end of the device to be affixed to the proximal stump consisted of one spiral nerve cuff and a PLGA NGC fixed onto a 14-gauge thin wall stainless steel hypotube (Component Supply Co.), designed to slide along the stainless steel rod. The hypotube had a 28-gauge guidewire attached, enabling us to stretch the proximal nerve stump. A second spiral nerve cuff and PLGA NGC were fixed to the distal terminal of the inner stainless steel rod. Medical grade n-butyl cyanoacrylate adhesive was used to affix the cuffs, guidance channel, and guidewire to the device backbone.

Device characterization

Spiral nerve cuffs

To find the correlation between preloaded strain in the silicone sheet and degree of spiral curl, the inner diameters of nerve cuffs of 50%, 60%, 70%, and 100% strains were measured and compared. Inner diameter was defined as the distance from the inner most tip of the curved cuff to the surface 180° from this starting point. We examined the capability of cuffs to grip nerves without slippage in rat cadavers (n=6) and in vivo (n=8). We exposed rat sciatic nerves and then deployed self-sizing spiral nerve cuffs (fabricated under 100% strain, Fig. 2d.) by unwinding them and letting them roll onto the nerves. The inner edge was gently pulled with forceps to ensure that there was no gap between cuff and nerve. In cadavers, we actuated the device the entire possible 6 mm (Fig. 1b). In vivo, we stretched nerves 20% beyond physiological strain (∼2 mm), and examined cuff positioning between 1 day and 3 weeks later.

We also assessed the possibility of nerve compression by the cuffs. During device implantations in situ and in vivo, the interface of the nerve and cuff was examined at 5×–20×magnification, to confirm that the nerve did not narrow upon entry to the cuff, indicated by a change in the diameter or trajectory of the nerve at the nerve–cuff interface. These effects were quantified ex vivo. We wrapped cuffs around nerves soaked in trypan blue (n=4), which provided contrast between nerve and cuff. Digital images (three images/nerve, from three different angles/nerve) were captured to compare nerve geometry at the nerve–cuff interface and away from the cuff. We first compared the angular trajectory of the nerve 0 mm (nerve-cuff border) to 0.25 mm from the cuff with the trajectory of the nerve in the adjacent nerve segment 0.25–0.5 mm from the cuff. In addition, the nerve diameter at the cuff–nerve interface (0 mm) was compared to the diameter 0.5 mm away from the cuff. Control trajectories and diameters were measured >3 mm from the cuff; angular trajectories in adjacent 0.25 mm segments and diameters flanking these segments were compared. A difference in the ratio of trajectories in adjacent regions, coupled with a difference in diameter, would suggest compression or tethering.

PLGA cytotoxicity

Both direct contact and test on extract methods of in vitro cytotoxicity testing were performed on the SH-SY5Y neuronal (neuroblastoma) cell line (ATCC # CRL-2266). The impact of leachable factors from PLGA on cell viability was performed as suggested in the ISO 10993-5 standard. Cell culture medium (90% MEM/F12, 10% fetal bovine serum) was incubated in PLGA-coated Petri dishes at 37°C for 24 h and then was used to feed SH-SY5Y cells. To test the effects of PLGA on cell growth, after 96 h of culture SH-SY5Y cells were collected by trypsinization of adherent cells. The numbers of total cells were estimated by cell counting using hemocytometer and statistically compared by Student's t-test (type I error α=0.05). Cell viability was tested using Live/Dead© fluorescence assay. Morphology of cells fed with PLGA-incubated medium was observed under an inverted light microscope. For the direct contact study, visualization of morphology of cells grown on PLGA using traditional transmitted light microscopy was difficult due to the opaque nature of the thick PLGA layer. Therefore, SH-SY5Y cells were first seeded onto discs coated with PLGA or PLGA followed by laminin in a Petri dish filled with culture medium. Discs were then inverted onto coverslips for imaging. For identifying the contours of the neuronal cells, they were stained with Alexa Fluor® 488 conjugated wheat germ agglutinin, which binds to sialic acid and N-acetylglucosaminyl sugar residues that reside on the cell membrane.44 Subsequent imaging was performed on an inverted widefield fluorescence microscope (Nikon TE-2000U) using filters appropriate for FITC visualization and a Leica SP5 confocal imaging system and a 63×objective at a resolution of 0.4805 μm/pix. For the latter, an argon laser enabled excitation at 488 nm and emission was captured between 500 and 550 nm.

Device implantation

Implantations were initially performed in cadavers, and then in vivo. For initial in situ studies, 14-week-old adult male Sprague-Dawley rats were sacrificed by CO2 euthanization. For in vivo studies, 14-week-old adult male Sprague-Dawley rats were anesthetized using 5% isoflurane inhalation anesthesia, followed by injection of analgesic (0.05 mg/kg buprenorphine) and antibiotics (5 mg/kg Baytril®). Anesthesia was maintained by 2% isoflurane inhalation all through the surgery. The surgical site was shaved and sterilized, and the sciatic nerve was exposed and severed as above. The device was implanted; the incision to the muscle was closed by 4-0 Vicryl® suture and the incision to the skin with 3-0 Prolene® monofilament suture. The rat was kept for up to 3 weeks with full access to food and water.

The sciatic nerve was exposed by separating branches of the hamstring muscles. A 10 mm segment of sciatic nerve proximal to the trifurcation was removed, and then the device was implanted. The backbone was oriented along the original axis of the nerve within the nerve bed. Proximal and distal nerve stumps were wrapped with the spiral nerve cuffs. Forceps were used to carefully grasp the epineurial sheath, and bring the tip of the nerve stump toward the open ends of the PLGA NGC. No more than 0.25 mm of the nerve was placed within the channel, to ensure guidance within the channel. The guidewire was then pulled to stretch the proximal stump of the sciatic nerve proximal stump to the desired length. Once the device was positioned appropriately, it was prevented from translating by fixing it to the underlying muscle bed with stainless steel anchors (28 gauge). A group of rats also underwent the same surgical protocol without device implantation. Contralateral nerves were harvested as controls.

Characterization of response to device

Following sacrifice, a gross assessment of device tolerance and immunohistochemistry were performed. The length of regenerating proximal nerve stump that extended beyond the cuff was measured and then rapidly frozen down in chilled isopentane for histology preparation. The regenerating nerve was embedded in Optimal Cutting Temperature (Sakura Fintek USA, Inc.) and cut into 10-μm sections for immunostaining. Primary antibodies including mouse anti-rat SMI-31 monoclonal antibody (Covance) diluted at 1:200 and rabbit anti-rat S-100 polyclonal antibody (Sigma-Aldrich) diluted at 1:200 along with secondary antibodies goat anti-mouse Alexa-Flour 488 conjugated and goat anti-rabbit Alexa-Flour 594 conjugated diluted at 1:200 (Invitrogen) were used and then visualized with confocol microscopy as above (Leica SP5), using filters appropriate for Texas Red visualization.

Results

Spiral nerve cuffs

Spiral nerve cuffs were successfully fabricated, and the microgroove pattern was successfully transferred to the inner surface of the cuff (Fig. 2b, c). Measurements of the inner diameter of the spiral nerve cuffs showed a decreasing trend of inner diameter with increasing strain in the stretched silicone sheet (Fig. 2d). Pilot testing of the efficacy of nerve gripping indicated that cuffs of 70% and 100% preloaded strain (average inner diameter of 1.35 and 0.85 mm, respectively) successfully held the nerve without slippage. We elected to use the 100% cuff for further characterization, as self-sizing was more consistent. For 14 out of 14 tests (100%) in cadavers and in vivo, there was no indication that nerves detached from the cuffs, or slid within the cuff, based on comparison of cuff position to reference marks at the proximal cuff–nerve boundary. The inner surface of spiral nerve cuff was in close contact with the nerve stump but caused minimal nerve compression. In all in situ and in vivo experiments (n=14), the trajectory and diameter of the nerve outside of the cuff was compared to its trajectory entering the cuff, and no indication of compression or tethering was observed in 100% (14/14) of these qualitative assessments. This was tested more formally ex vivo, by quantifying the geometry and trajectories of nerves entering the cuffs. Nerves were dyed with trypan blue, to enhance contrast with the optically clear cuff. Consistent with qualitative observations of a continuous nerve trajectory into the cuff (e.g., Fig. 2e), based on the ratio of trajectories and diameters in adjacent regions at and away from the cuff (control), no deviation in trajectory (control: 1.00±0.002 vs. cuff: 1.00±0.006, p=0.41, n=8 interfaces) or diameter (control: 0.99±0.01 vs. cuff: 0.98±0.01, p=0.67, n=8 interfaces) was observed at the interface of cuff entry.

PLGA nerve guidance channel

Due to phase separation, white solid state PLGA gradually precipitated onto the water-eluding alginate hydrogel. After retrieval of the alginate rod, a hollow PLGA tube was produced (Fig. 3a). The PLGA tube was then cut into small segments to serve as NGCs (Fig. 3b). Cross-sectional scanning electron microscopy images showed that the PLGA tubes had a porous structure (Fig. 3c) as reported previously.45

FIG. 3.

FIG. 3.

Fabrication of PLGA NGC. (a) Alginate rod was immersed in PLGA/Pluronic F-127 solution for 10 min. Tubular PLGA layer is formed due to phase separation when water diffuses out the alginate rod; coated alginate rod was thoroughly washed in water before alginate rod is removed (b) PLGA tube is cut into short segments to serve as NGC. (c) Cross section scanning electron microscopy picture of PLGA tube.

PLGA cytotoxicity

SH-SY5Y cell fed with medium preincubated with PLGA showed similar morphology (Fig. 4b) to that of control (Fig. 4a). Live/Dead® cell viability assay also showed that only very few dead cells were present (Fig. 4d; 4c as control). Quantification revealed no significant difference in viability between treated and control cells (94%±3.71% vs. 98%±0.27% (mean±standard error of the mean), respectively; p=0.30; Student's t-test, n=3). Thus, both morphologically and biochemically, no signs of cytotoxicity were observed. For the effects of PLGA on cell division and growth, the experimental group displayed no significant difference compared with the control group (Fig. 4e). For direct observation of morphology of cells grown on PLGA, cell membranes were labeled with fluorescent wheat germ agglutinin. Confocal imaging revealed that SH-SY5Y cells adhered to a laminin-coated PLGA substrate, spread, and extended outward (Fig. 4f). Outgrowth and cell density on a laminin-coated surface was superior to that on PLGA alone (data not shown).

FIG. 4.

FIG. 4.

PLGA cytotoxicity. (a–d) Cells were fed with PLGA-incubated culture medium to test whether leachable substances from a PLGA scaffold would have cytotoxic effects on the cells. (a, b) No signs of PLGA cytotoxicity was observed in terms of cell morphology; (c, d) Live/Dead® cell viability assay indicated very few dead cells; (e) No statistical difference was found between control and PLGA group, p=0.05. (f) Visualization of morphology of cells grown on PLGA by fluorescent wheat germ agglutinin staining of cell membranes. SH-SY5Y cells attached, spread, and proliferated on PLGA. Color images available online at www.liebertpub.com/tec

In situ implantation—feasibility of no-slip actuation

The assembled device is shown ex vivo under nonactuated and actuated configurations (Fig. 5a, b). Deployment of the device was first performed in a rat cadaver, to demonstrate the feasibility of actuation. The device was implanted within a nerve defect, with the nerve stumps wrapped by spiral cuffs (Fig. 5c). By pulling the guidewire attached to the hypotube sliding over the rod, the nerve stump of proximal end was successfully stretched and placed under tension (Fig. 5d). This configuration corresponds to the maximum possible one-time deformation (6 mm) of the nerve, which is limited by guidance channel dimensions. Though unlikely to be physiologically relevant, this cadaver experiment demonstrates no-slip gripping of nerve at substantial deformation, and thus opposing tension.

FIG. 5.

FIG. 5.

Demonstration of slip-free nerve deformation. (a, b) Device fully extended and maximally actuated, ex vivo. (c, d) Fully extended device and actuated device in a rat sciatic nerve defect. Note that implanted device can stretch the nerve stumps 6 mm without slippage. Color images available online at www.liebertpub.com/tec

In vivo implantation—feasibility of promoting regeneration

The device was successfully deployed across a sciatic nerve defect in anesthetized rats (Fig. 6a), and tensioned to physiological (∼10%; n=4) or super-physiological deformation (∼20%; n=4), demonstrating feasibility of device implantation and actuation in vivo. While a comparative, quantitative assessment of regeneration is underway, but outside of the scope of this study, several observations confirmed the feasibility of using our device to probe regeneration. All animals tolerated the device for up to 3 weeks without obvious signs of infection. Upon reopening the incision, as expected, fibrotic encapsulation of the device was observed (Fig. 6b). Nevertheless, the device was cleanly excised, with minimal connective tissue bound to either the stainless steel backbone (Fig. 6c) or the attached cuff/nerve complex (Fig. 6d). Moreover, both nerve stumps remained confined within the cuffs without apparent slippage, and the proximal stump extended about 6 mm beyond the cuff into the guidance channel (Fig 6d; guidance channel removed to visualize nerve). Nerve confinement and alignment by the device was in sharp contrast to injured nerves allowed to recover for 1 week in the absence of a guidance channel, which were misaligned and blocked from regeneration by fibrosis (Fig. 6e, f).

FIG. 6.

FIG. 6.

Response to 2-week implantation of device. (a) A nerve gap was created by removal of 10 mm segment from the rat sciatic nerve. (b) Following 2 weeks of implantation, minor fibrosis was observed at the implantation site. (c) The stainless steel backbone was cleanly extracted at 2 weeks. (d) The proximal stump was still securely held by the nerve cuff (red arrow), and the regenerating tip extended beyond the cuff ∼6 mm (blue arrow). (e, f) In the absence of device implantation the two stumps remained disconnected and misaligned. A bulge was observed at the proximal stump and the degenerating distal stump appeared fused with surrounding fatty/connective tissue. Color images available online at www.liebertpub.com/tec

Immunolabeling indicated the presence of axons in the regenerated region of the stump, indicated by positive staining of phosphorylated neurofilaments (SMI31 antibody) (Fig. 7b, c) and Schwann cell marker (S100 antibody) (Fig. 7e, f). The parallel alignment of axons and Schwann cells was similar to that of contralateral controls (Fig. 7d, g). Collectively, these preliminary in vivo data indicated the feasibility of device usage in vivo, and provide confidence in the deployment of such devices to test hypotheses regarding the role of tensile loading in nerve regeneration.

FIG. 7.

FIG. 7.

Evidence of regenerative neural outgrowth. (a) The regenerating nerve extended ∼6 mm beyond the cuff, and was stained with anti-SMI-31 (a–c) and anti-S100 (e, f) antibodies, which labeled phosphorylated neurofilaments and Schwann cells, respectively. (d, g) The contralateral sciatic nerve was stained with the same two markers and served as a control. Color images available online at www.liebertpub.com/tec

Discussion

Toward the implementation of a novel strategy for accelerating peripheral nerve regeneration, we have successfully designed, fabricated, and implanted across a rat sciatic nerve gap a modular device that enables simultaneous lengthening of the proximal nerve stump and axonal outgrowth into an engineered scaffold. The novelty of our approach lies in the application of tensile loading (stretch) as a strategy to accelerate peripheral nerve regeneration, and the seamless integration of this strategy with existing tissue engineering strategies for nerve repair.

A role for tension in nerve repair

Although several recent studies implicate tension as a key regulator of neuronal survival, the role of tension in peripheral nerve repair is debatable. In clinical practice, tension-free repair remains the preferred treatment.46,47 This is motivated by suggestions that excessive tension results in scar tissue formation and adhesion48,49 or impairs blood supply50,51; studies showed blood flow reduced ∼50% when nerves are stretched to 10%.52,53 On the other hand, Sunderland et al. showed that modest levels of tension were well tolerated in rat sciatic nerve regeneration model,54 and robust nerve regeneration was observed at 4 weeks in all except the 9 mm-repair group. Moreover, Smith et al. demonstrated the capacity of integrated axons to undergo substantial growth via continuous mechanical tension; indeed a stretch-induced axonal growth of 1 cm in length by 10 days of stretch was achieved.55 The most compelling evidence for a role of tension in nerve regeneration may be found in animal and human models of limb lengthening, where nerves tolerate substantial deformations during the lengthening process.34,5665

Therefore, the more appropriate debate should be on defining an appropriate threshold beyond which tension is detrimental. Our device design is motivated by the hypothesis that maintaining physiological strain during the lengthening process will accelerate nerve regeneration.

Novel device features and characterization

Our device consisted of three key design features: silicone spiral nerve cuffs to hold the nerve, a telescoping stainless steel backbone to enable nerve deformation, and PLGA NGCs to promote axonal outgrowth.

Self-sizing spiral cuffs, modeled after spiral cuff electrodes,42 were fabricated to a desired inner diameter (Fig. 2d). They were also patterned with microgrooves on the inner surface to increase friction between the cuff and underlying nerve. Self-sizing cuffs successfully held nerves during deformation without excessive compression (Fig. 2e). In contrast to sutures, which would concentrate stresses in their vicinity during stretch, cuffs also promoted the distribution of loads over a larger surface area. Mounting the cuffs on telescoping rods enabled actuation through a simple guidewire, intracorporeally. While there is room for innovation in controlling the degree of actuation, the fact that the guidewire is the only piece of the device protruding from the body is attractive, and a significant advantage over an external fixator design.

Though any scaffold may be integrated into our device, we initially selected tubular PLGA/Pluronic F-127 conduits as NGCs. This material was selected based on its hydrophilicity, our ability to carefully control its geometry, even in hydrous conditions, and its promise as a guidance channel.43 We confirmed that the resultant PLGA nerve guidance had a porous structure (Fig. 3c), which is likely to be permeable to nutrients. We extended the previous fabrication and characterization of PLGA NGCs by performing direct and indirect contact cytotoxicity tests. Though PLGA is a well-established biomaterial, our use of neuronal cells extends previous studies on fibroblasts,6668 and is directly relevant to our intended application for neuronal regeneration. For both direct contact and extracts testing, no signs of cytotoxicity were found in terms of cell proliferation and cell viability. With respect to morphology, SH-SY5Y cells on PLGA exhibited adherent cell bodies with projections extending outward, similar to neuronal cells cultured in tissue culture Petri dishes. A substantial improvement in neuronal adherence and neurite morphology was observed when neurons were plated on PLGA coated with laminin. Consequently, this minor, but beneficial change should be incorporated into implanted PLGA NGCs.

Device implantation and surgical implementation

We confirmed the feasibility of implanting our device across a sciatic nerve gap and lengthening the proximal nerve stump without slippage up to 6 mm (∼60%) in a rat cadaver (Fig. 5c, d). This excessive strain would likely be detrimental physiologically, but did confirm the ability to maintain a suture-free grip on a nerve despite considerable opposing tensile loads. Preliminary results in vivo were also promising. Nerves were successfully subject to a one-time stretch of ∼10% (physiological strain) and ∼20% (super physiological strain), and revealed that rats did not show signs of infection, and survived at least 3 weeks before sacrifice. Within this time frame, the proximal nerve stump extended about 6 mm beyond the cuff (Fig. 6d); we further identified regenerating axons as a part of this elongating tissue, based on positive, well aligned labeling of a marker for mature axons, phosphorylated neurofilaments (Fig. 7). Such regenerative growth was in contrast to nerve outgrowth in the absence of intervention, which appeared misaligned, and plagued by interactions with surrounding fatty and connective tissue (Fig 6e, f). These observations indicate the necessity of both NGC and an aligning backbone, to increase the probability of successfully reconnecting nerve stumps.

In practice, because the proximal and distal guidance channels will collide following sufficient lengthening (6 mm total, for our described prototype), an ideal scenario will be to have the channels meet shortly before axons exit the proximal channel. Based on literature values, dimensions of the prototype device assume lengthening over 5–6 days, at conservative lengthening rates of 1 mm/day and axonal outgrowth rates of 0.3 mm/day. However, these rates will eventually be optimized based on empirically determined thresholds for lengthening and axonal outgrowth rates into the scaffold.

Finally, we would be remiss if we did not point out some potential pitfalls of implanting our device in vivo. Most anticipated issues are likely to arise due to inflammatory and fibrotic processes. These include jamming of the telescoping mechanism of the hypotube and the inner rod owing to fibrous tissue infiltration, cell infiltration into the nerve gap, and scar formation. We speculate that due to actuation frequently, and primarily at early time points, the jamming of the telescoping mechanism is unlikely and fibrosis will be no more of an issue with our device than any other implanted scaffolds or devices, including an external nerve lengthening device.6971 We will test the utility of medical grade expanded polytetrafluoroethylene sleeves as a protective sheath surrounding the device. Another potential drawback is possible infection at the site where guidewire extends from the animal's body. Such a scenario would be treated with antibiotics; however, it should be noted that many peripheral nerve devices, including spiral cuff electrodes,42 have been successfully implanted with leads exiting the body. Therefore, we speculate that this risk is minimal. Finally, to successfully stretch the proximal nerve stump with the device, the mechanical backbone must be secured to a reference position. In our initial study, we have secured the device to the underlying muscle using simple stainless steel anchors. This is less invasive and appears adequate, based on preliminary short-term survival surgeries; however, should complications arise, anchorage to the femur provides a more invasive, but previously validated option.6971

Comparison to other strategies for nerve regeneration

Our lengthening device is intended to be used in parallel with other guidance strategies, including autologous grafts, synthetic grafts, acellular guidance matrices, and seeded matrices. Consequently, a direct comparison between a passive scaffold and the same scaffold incorporated within our device would be most appropriate. For the device introduced in this study, the appropriate comparisons would be an uncoated passive PLGA scaffold33 and a passive PLGA scaffold coated with laminin. Differential responses between our device and these scaffolds would reflect effects of mechanical extension of intact regions of the proximal nerve superposed on outgrowth into the scaffold (Fig. 1).

To our knowledge, only one other group has directly evaluated tensile loading as a regenerative strategy. The Ochiai group proposed an intriguing series of articles 6971 that implemented a direct lengthening device to lengthen proximal and distal nerve stumps. Regeneration rates were equal or greater than those observed with an autologous graft, lending strong support to the hypothesis that mechanical loading can be beneficial to outgrowth. However, three key issues detract from the translational viability of this technology. First, based on their proposed design, nerves are readily lengthened, but the device configuration is not amenable to reattachment without additional surgery (i.e., there is nowhere for outgrowing axons to go). Second, because this design requires a complex external fixator mounted within the adjacent femur, more invasive surgery is required. Finally, nerve ends are secured with suture for tensioning; such attachment is likely to impose compression or uneven strain distributions owing to the discrete sites of attachment, again diminishing the likelihood of successful axonal outgrowth. Through the use of an internal fixator and self-sizing cuffs, we believe that our design has accounted for the most pressing of these issues. Though yet at an early stage of development, the successful fabrication and implementation of our device provides considerable enthusiasm for this device and the broader strategy of mechanical influences on nerve regeneration.

Acknowledgments

This research was supported by grant W81XWH1010773 from the U.S. Army Medical Research and Materiel Command and grant CBET1042522 from the National Science Foundation. We also acknowledge helpful discussions with members of the Neuromuscular Bioengineering Laboratory. Scanning electron microscopy image was taken at the University of Maryland Nanoscale Imaging Spectroscopy and Properties Laboratory.

Disclosure Statement

No competing financial interests exist.

References

  • 1.Burnett M.G. Zager E.L. Pathophysiology of peripheral nerve injury: a brief review. Neurosurg Focus. 2004;16:E1. doi: 10.3171/foc.2004.16.5.2. [DOI] [PubMed] [Google Scholar]
  • 2.Evans G.R. Peripheral nerve injury: a review and approach to tissue engineered constructs. Anat Rec. 2001;263:396. doi: 10.1002/ar.1120. [DOI] [PubMed] [Google Scholar]
  • 3.Schmidt C.E. Leach J.B. Neural tissue engineering: strategies for repair and regeneration. Annu Rev Biomed Eng. 2003;5:293. doi: 10.1146/annurev.bioeng.5.011303.120731. [DOI] [PubMed] [Google Scholar]
  • 4.Seddon H.J. Three types of nerve injury. Brain. 1943;66:247. [Google Scholar]
  • 5.Sunderland S. A classification of peripheral nerve injury produced by loss of function. Brain. 1951;74:491. doi: 10.1093/brain/74.4.491. [DOI] [PubMed] [Google Scholar]
  • 6.Scherman P. Kanje M. Dahlin L.B. Bridging short nerve defects by direct repair under tension, nerve grafts or longitudinal sutures. Restor Neurol Neurosci. 2004;22:65. [PubMed] [Google Scholar]
  • 7.Dahlin L. Johansson F. Lindwall C. Kanje M. Chapter 28: Future perspective in peripheral nerve reconstruction. Int Rev Neurobiol. 2009;87:507. doi: 10.1016/S0074-7742(09)87028-1. [DOI] [PubMed] [Google Scholar]
  • 8.Terenghi G. Wiberg M. Kingham P.J. Chapter 21: Use of stem cells for improving nerve regeneration. Int Rev Neurobiol. 2009;87:393. doi: 10.1016/S0074-7742(09)87021-9. [DOI] [PubMed] [Google Scholar]
  • 9.Zacchigna S. Giacca M. Chapter 20: Gene therapy perspectives for nerve repair. Int Rev Neurobiol. 2009;87:381. doi: 10.1016/S0074-7742(09)87020-7. [DOI] [PubMed] [Google Scholar]
  • 10.Donat J.R. Wisniewski H.M. The spatio-temporal pattern of Wallerian degeneration in mammalian peripheral nerves. Brain Res. 1973;53:41. doi: 10.1016/0006-8993(73)90765-8. [DOI] [PubMed] [Google Scholar]
  • 11.Atchabahian A. Genden E.M. MacKinnon S.E. Doolabh V.B. Hunter D.A. Regeneration through long nerve grafts in the swine model. Microsurgery. 1998;18:379. doi: 10.1002/(sici)1098-2752(1998)18:6<379::aid-micr7>3.0.co;2-5. [DOI] [PubMed] [Google Scholar]
  • 12.Millesi H. Meissl G. Berger A. The interfascicular nerve-grafting of the median and ulnar nerves. J Bone Joint Surg Am. 1972;54:727. [PubMed] [Google Scholar]
  • 13.Strasberg S.R. Mackinnon S.E. Genden E.M. Bain J.R. Purcell C.M. Hunter D.A. Hay J.B. Long-segment nerve allograft regeneration in the sheep model: experimental study and review of the literature. J Reconstr Microsurg. 1996;12:529. doi: 10.1055/s-2007-1006625. [DOI] [PubMed] [Google Scholar]
  • 14.Bora F.W., Jr. Bednar J.M. Osterman A.L. Brown M.J. Sumner A.J. Prosthetic nerve grafts: a resorbable tube as an alternative to autogenous nerve grafting. J Hand Surg Am. 1987;12:685. doi: 10.1016/s0363-5023(87)80048-5. [DOI] [PubMed] [Google Scholar]
  • 15.Fawcett J.W. Keynes R.J. Muscle basal lamina: a new graft material for peripheral nerve repair. J Neurosurg. 1986;65:354. doi: 10.3171/jns.1986.65.3.0354. [DOI] [PubMed] [Google Scholar]
  • 16.Gattuso J.M. Glasby M.A. Gschmeissner S.E. Norris R.W. A comparison of immediate and delayed repair of peripheral nerves using freeze-thawed autologous skeletal muscle grafts—in the rat. Br J Plast Surg. 1989;42:306. doi: 10.1016/0007-1226(89)90152-5. [DOI] [PubMed] [Google Scholar]
  • 17.Chiu D.T. Janecka I. Krizek T.J. Wolff M. Lovelace R.E. Autogenous vein graft as a conduit for nerve regeneration. Surgery. 1982;91:226. [PubMed] [Google Scholar]
  • 18.Suematsu N. Atsuta Y. Hirayama T. Vein graft for repair of peripheral nerve gap. J Reconstr Microsurg. 1988;4:313. doi: 10.1055/s-2007-1006937. [DOI] [PubMed] [Google Scholar]
  • 19.Strauch B. Ferder M. Lovelle-Allen S. Moore K. Kim D.J. Llena J. Determining the maximal length of a vein conduit used as an interposition graft for nerve regeneration. J Reconstr Microsurg. 1996;12:521. doi: 10.1055/s-2007-1006624. [DOI] [PubMed] [Google Scholar]
  • 20.Brandt J. Dahlin L.B. Lundborg G. Autologous tendons used as grafts for bridging peripheral nerve defects. J Hand Surg Br. 1999;24:284. doi: 10.1054/jhsb.1999.0074. [DOI] [PubMed] [Google Scholar]
  • 21.Bell J. Haycock J.W. Next generation nerve guides—materials, fabrication, growth factors and cell delivery. Tissue Eng Part B Rev. 2012;8:116. doi: 10.1089/ten.TEB.2011.0498. [DOI] [PubMed] [Google Scholar]
  • 22.Ao Q. Fung C.K. Tsui A.Y. Cai S. Zuo H.C. Chan Y.S. Shum D.K. The regeneration of transected sciatic nerves of adult rats using chitosan nerve conduits seeded with bone marrow stromal cell-derived Schwann cells. Biomaterials. 2011;32:787. doi: 10.1016/j.biomaterials.2010.09.046. [DOI] [PubMed] [Google Scholar]
  • 23.Harvey A.R. Chen M. Plant G.W. Dyson S.E. Regrowth of axons within Schwann cell-filled polycarbonate tubes implanted into the damaged optic tract and cerebral cortex of rats. Restor Neurol Neurosci. 1994;6:221. doi: 10.3233/RNN-1994-6306. [DOI] [PubMed] [Google Scholar]
  • 24.Hsu S.H. Su C.H. Chiu I.M. A novel approach to align adult neural stem cells on micropatterned conduits for peripheral nerve regeneration: a feasibility study. Artif Organs. 2009;33:26. doi: 10.1111/j.1525-1594.2008.00671.x. [DOI] [PubMed] [Google Scholar]
  • 25.Marchesi C. Pluderi M. Colleoni F. Belicchi M. Meregalli M. Farini A. Parolini D. Draghi L. Fruguglietti M.E. Gavina M. Porretti L. Cattaneo A. Battistelli M. Prelle A. Moggio M. Borsa S. Bello L. Spagnoli D. Gaini S.M. Tanzi M.C. Bresolin N. Grimoldi N. Torrente Y. Skin-derived stem cells transplanted into resorbable guides provide functional nerve regeneration after sciatic nerve resection. Glia. 2007;55:425. doi: 10.1002/glia.20470. [DOI] [PubMed] [Google Scholar]
  • 26.Lundborg G. Dahlin L.B. Danielsen N. Gelberman R.H. Longo F.M. Powell H.C. Varon S. Nerve regeneration in silicone chambers: influence of gap length and of distal stump components. Exp Neurol. 1982;76:361. doi: 10.1016/0014-4886(82)90215-1. [DOI] [PubMed] [Google Scholar]
  • 27.Brown R. Pedowitz R. Rydevik B. Woo S. Hargens A. Massie J. Kwan M. Garfin S.R. Effects of acute graded strain on efferent conduction properties in the rabbit tibial nerve. Clin Orthop Relat Res. 1993;296:288. [PubMed] [Google Scholar]
  • 28.Kwan M.K. Wall E.J. Massie J. Garfin S.R. Strain, stress and stretch of peripheral nerve. Rabbit experiments in vitro and in vivo. Acta Orthop Scand. 1992;63:267. doi: 10.3109/17453679209154780. [DOI] [PubMed] [Google Scholar]
  • 29.Rydevik B.L. Kwan M.K. Myers R.R. Brown R.A. Triggs K.J. Woo S.L. Garfin S.R. An in vitro mechanical and histological study of acute stretching on rabbit tibial nerve. J Orthop Res. 1990;8:694. doi: 10.1002/jor.1100080511. [DOI] [PubMed] [Google Scholar]
  • 30.Phillips J.B. Smit X. De Zoysa N. Afoke A. Brown R.A. Peripheral nerves in the rat exhibit localized heterogeneity of tensile properties during limb movement. J Physiol. 2004;557:879. doi: 10.1113/jphysiol.2004.061804. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Wright T.W. Glowczewskie F. Wheeler D. Miller G. Cowin D. Excursion and strain of the median nerve. J Bone Joint Surg Am. 1996;78:1897. doi: 10.2106/00004623-199612000-00013. [DOI] [PubMed] [Google Scholar]
  • 32.Wright T.W. Glowczewskie F., Jr. Cowin D. Wheeler D.L. Ulnar nerve excursion and strain at the elbow and wrist associated with upper extremity motion. J Hand Surg Am. 2001;26:655. doi: 10.1053/jhsu.2001.26140. [DOI] [PubMed] [Google Scholar]
  • 33.Anava S. Greenbaum A. Ben Jacob E. Hanein Y. Ayali A. The regulative role of neurite mechanical tension in network development. Biophys J. 2009;96:1661. doi: 10.1016/j.bpj.2008.10.058. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Abe I. Ochiai N. Ichimura H. Tsujino A. Sun J. Hara Y. Internodes can nearly double in length with gradual elongation of the adult rat sciatic nerve. J Orthop Res. 2004;22:571. doi: 10.1016/j.orthres.2003.08.019. [DOI] [PubMed] [Google Scholar]
  • 35.Bray D. Axonal growth in response to experimentally applied mechanical tension. Dev Biol. 1984;102:379. doi: 10.1016/0012-1606(84)90202-1. [DOI] [PubMed] [Google Scholar]
  • 36.Bueno F.R. Shah S.B. Implications of tensile loading for the tissue engineering of nerves. Tissue Eng Part B Rev. 2008;14:219. doi: 10.1089/ten.teb.2008.0020. [DOI] [PubMed] [Google Scholar]
  • 37.Pfister B.J. Iwata A. Meaney D.F. Smith D.H. Extreme stretch growth of integrated axons. J Neurosci. 2004;24:7978. doi: 10.1523/JNEUROSCI.1974-04.2004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Topp K.S. Boyd B.S. Structure and biomechanics of peripheral nerves: nerve responses to physical stresses and implications for physical therapist practice. Phys Ther. 2006;86:92. doi: 10.1093/ptj/86.1.92. [DOI] [PubMed] [Google Scholar]
  • 39.Zheng J. Lamoureux P. Santiago V. Dennerll T. Buxbaum R.E. Heidemann S.R. Tensile regulation of axonal elongation and initiation. J Neurosci. 1991;11:1117. doi: 10.1523/JNEUROSCI.11-04-01117.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Rupp A. Schmahl W. Lederer W. Matiasek K. Strain differences in the branching of the sciatic nerve in rats. Anat Histol Embryol. 2007;36:202. doi: 10.1111/j.1439-0264.2007.00751.x. [DOI] [PubMed] [Google Scholar]
  • 41.Strasberg J.E. Strasberg S. Mackinnon S.E. Watanabe O. Hunter D.A. Tarasidis G. Strain differences in peripheral-nerve regeneration in rats. J Reconstr Microsurg. 1999;15:287. doi: 10.1055/s-2007-1000103. [DOI] [PubMed] [Google Scholar]
  • 42.Naples G.G. Mortimer J.T. Scheiner A. Sweeney J.D. A spiral nerve cuff electrode for peripheral nerve stimulation. IEEE Trans Biomed Eng. 1988;35:905. doi: 10.1109/10.8670. [DOI] [PubMed] [Google Scholar]
  • 43.Oh S.H. Kim J.H. Song K.S. Jeon B.H. Yoon J.H. Seo T.B. Namgung U. Lee I.W. Lee J.H. Peripheral nerve regeneration within an asymmetrically porous PLGA/Pluronic F127 nerve guide conduit. Biomaterials. 2008;29:1601. doi: 10.1016/j.biomaterials.2007.11.036. [DOI] [PubMed] [Google Scholar]
  • 44.Nagata Y. Burger M.M. Wheat germ agglutinin. Molecular characteristics and specificity for sugar binding. J Biol Chem. 1974;249:3116. [PubMed] [Google Scholar]
  • 45.Oh S.H. Lee J.H. Fabrication and characterization of hydrophilized porous PLGA nerve guide conduits by a modified immersion precipitation method. J Biomed Mater Res A. 2007;80:530. doi: 10.1002/jbm.a.30937. [DOI] [PubMed] [Google Scholar]
  • 46.McDonald D.S. Bell M.S. Peripheral nerve gap repair facilitated by a dynamic tension device. Can J Plast Surg. 2010;18:e17. [PMC free article] [PubMed] [Google Scholar]
  • 47.Rivlin M. Sheikh E. Isaac R. Beredjiklian P.K. The role of nerve allografts and conduits for nerve injuries. Hand Clin. 2010;26:435. doi: 10.1016/j.hcl.2010.04.010. [DOI] [PubMed] [Google Scholar]
  • 48.Millesi H. Reappraisal of nerve repair. Surg Clin North Am. 1981;61:321. doi: 10.1016/s0039-6109(16)42384-4. [DOI] [PubMed] [Google Scholar]
  • 49.Millesi H. The nerve gap. Theory and clinical practice. Hand Clin. 1986;2:651. [PubMed] [Google Scholar]
  • 50.Flores A.J. Lavernia C.J. Owens P.W. Anatomy and physiology of peripheral nerve injury and repair. Am J Orthop (Belle Mead NJ) 2000;29:167. [PubMed] [Google Scholar]
  • 51.Lundborg G. Intraneural microcirculation. Orthop Clin North Am. 1988;19:1. [PubMed] [Google Scholar]
  • 52.Clark W.L. Trumble T.E. Swiontkowski M.F. Tencer A.F. Nerve tension and blood flow in a rat model of immediate and delayed repairs. J Hand Surg Am. 1992;17:677. doi: 10.1016/0363-5023(92)90316-h. [DOI] [PubMed] [Google Scholar]
  • 53.Driscoll P.J. Glasby M.A. Lawson G.M. An in vivo study of peripheral nerves in continuity: biomechanical and physiological responses to elongation. J Orthop Res. 2002;20:370. doi: 10.1016/S0736-0266(01)00104-8. [DOI] [PubMed] [Google Scholar]
  • 54.Sunderland I.R. Brenner M.J. Singham J. Rickman S.R. Hunter D.A. Mackinnon S.E. Effect of tension on nerve regeneration in rat sciatic nerve transection model. Ann Plast Surg. 2004;53:382. doi: 10.1097/01.sap.0000125502.63302.47. [DOI] [PubMed] [Google Scholar]
  • 55.Smith D.H. Wolf J.A. Meaney D.F. A new strategy to produce sustained growth of central nervous system axons: continuous mechanical tension. Tissue Eng. 2001;7:131. doi: 10.1089/107632701300062714. [DOI] [PubMed] [Google Scholar]
  • 56.Abe I. Tsujino A. Hara Y. Ichimura H. Ochiai N. Paranodal demyelination by gradual nerve stretch can be repaired by elongation of internodes. Acta Neuropathol (Berl) 2002;104:505. doi: 10.1007/s00401-002-0584-y. [DOI] [PubMed] [Google Scholar]
  • 57.Abe I. Tsujino A. Hara Y. Ochiai N. Effect of the rate of prestretching a peripheral nerve on regeneration potential after transection and repair. J Orthop Sci. 2003;8:693. doi: 10.1007/s00776-003-0700-4. [DOI] [PubMed] [Google Scholar]
  • 58.Bora F.W., Jr. Richardson S. Black J. The biomechanical responses to tension in a peripheral nerve. J Hand Surg [Am] 1980;5:21. doi: 10.1016/s0363-5023(80)80037-2. [DOI] [PubMed] [Google Scholar]
  • 59.Ichimura H. Shiga T. Abe I. Hara Y. Terui N. Tsujino A. Ochiai N. Distribution of sodium channels during nerve elongation in rat peripheral nerve. J Orthop Sci. 2005;10:214. doi: 10.1007/s00776-004-0870-8. [DOI] [PubMed] [Google Scholar]
  • 60.Ikeda K. Tomita K. Tanaka S. Experimental study of peripheral nerve injury during gradual limb elongation. Hand Surg. 2000;5:41. doi: 10.1142/s0218810400000028. [DOI] [PubMed] [Google Scholar]
  • 61.Yokota A. Doi M. Ohtsuka H. Abe M. Nerve conduction and microanatomy in the rabbit sciatic nerve after gradual limb lengthening-distraction neurogenesis. J Orthop Res. 2003;21:36. doi: 10.1016/S0736-0266(02)00100-6. [DOI] [PubMed] [Google Scholar]
  • 62.Li J. Shi R. A device for the electrophysiological recording of peripheral nerves in response to stretch. J Neurosci Methods. 2006;154:102. doi: 10.1016/j.jneumeth.2005.12.007. [DOI] [PubMed] [Google Scholar]
  • 63.Jou I.M. Lai K.A. Shen C.L. Yamano Y. Changes in conduction, blood flow, histology, and neurological status following acute nerve-stretch injury induced by femoral lengthening. J Orthop Res. 2000;18:149. doi: 10.1002/jor.1100180121. [DOI] [PubMed] [Google Scholar]
  • 64.Lee C. Ma J. Deal D.N. Smith B.P. Koman L.A. Smith T.L. Shilt J.S. Neuromuscular recovery after distraction osteogenesis at different frequencies in a rabbit model. J Pediatr Orthop. 2006;26:628. doi: 10.1097/01.bpo.0000226283.39319.06. [DOI] [PubMed] [Google Scholar]
  • 65.Spiegel D.A. Seaber A.V. Chen L.E. Urbaniak J.R. Recovery following stretch injury to the sciatic nerve of the rat: an in vivo study. J Reconstr Microsurg. 1993;9:69. doi: 10.1055/s-2007-1006641. [DOI] [PubMed] [Google Scholar]
  • 66.Bruggeman J.P. de Bruin B.J. Bettinger C.J. Langer R. Biodegradable poly(polyol sebacate) polymers. Biomaterials. 2008;29:4726. doi: 10.1016/j.biomaterials.2008.08.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Hwang C.M. Khademhosseini A. Park Y. Sun K. Lee S.H. Microfluidic chip-based fabrication of PLGA microfiber scaffolds for tissue engineering. Langmuir. 2008;24:6845. doi: 10.1021/la800253b. [DOI] [PubMed] [Google Scholar]
  • 68.Zhao L. He C. Gao Y. Cen L. Cui L. Cao Y. Preparation and cytocompatibility of PLGA scaffolds with controllable fiber morphology and diameter using electrospinning method. J Biomed Mater Res B Appl Biomater. 2008;87:26. doi: 10.1002/jbm.b.31060. [DOI] [PubMed] [Google Scholar]
  • 69.Nishiura Y. Hara Y. Yoshii Y. Ochiai N. Simultaneous gradual lengthening of both proximal and distal nerve stumps for repair of peripheral nerve defect in rats. Muscle Nerve. 2008;38:1474. doi: 10.1002/mus.21147. [DOI] [PubMed] [Google Scholar]
  • 70.Sharula Hara Y. Nishiura Y. Saijilafu Kubota S. Ochiai N. Repair of the sciatic nerve defect with a direct gradual lengthening of proximal and distal nerve stumps in rabbits. Plast Reconstr Surg. 2010;125:846. doi: 10.1097/PRS.0b013e3181ccdbd4. [DOI] [PubMed] [Google Scholar]
  • 71.Yamada Y. Nishiura Y. Saijilafu Hara Y. Ichimura H. Yoshii Y. Ochiai N. Repair of peripheral nerve defect by direct gradual lengthening of the distal nerve stump in rats: cellular reaction. Scand J Plast Reconstr Surg Hand Surg. 2009;43:297. doi: 10.1080/02844310903052578. [DOI] [PubMed] [Google Scholar]

Articles from Tissue Engineering. Part C, Methods are provided here courtesy of Mary Ann Liebert, Inc.

RESOURCES