Abstract
Objective
Patient-prosthesis mismatch (PPM) may affect clinical outcomes in patients with aortic valve replacement (AVR). We retrospectively examined the PPM in patients with isolated AVR in the Japan Adult Cardiovascular Surgery Database (JACVSD).
Methods
We examined all patients with isolated AVR between January 1, 2008 and December 31, 2009. The JACVSD data collection form has a total of 255 variables. We defined PPM as an effective orifice area index of ≤0.85 m2/cm2.
Results
PPM was observed in 306 of 3,609 cases analyzed, PPM rate was 8.5 %. Body surface area was larger and body mass index was higher in the PPM group than the non-PPM group (P < 0.001). Patients with PPM were older (P = 0.001) and had a higher prevalence of diabetes (P = 0.004), dyslipidemia (P < 0.001), hypertension (P < 0.001), cerebrovascular disease (P = 0.031), old myocardial infarction (P = 0.006), previous percutaneous coronary artery intervention (P = 0.001), coronary artery disease (P = 0.018), and aortic valve stenosis (P < 0.001). Perioperative blood transfusion (P < 0.001) and dialysis (P = 0.005) were more frequent in the PPM group. Postoperative ventilation (P = 0.004) and intensive care unit stay (P = 0.004) were significantly longer in the PPM group.
Conclusions
Age, aortic valve stenosis, dyslipidemia, hypertension, old myocardial infarction, previous percutaneous coronary artery intervention, diabetes mellitus, cerebrovascular disease, and high body mass index were the risk factors for PPM. PPM was not an independent risk factor for short-term mortality.
Keywords: Aortic valve replacement, Prosthesis, Mortality, Hemodynamics
Introduction
Patient-prosthesis mismatch (PPM) may affect clinical outcomes in patients with aortic valve replacement (AVR). Some clinical studies show differences in early and late outcomes in patients with PPM, but many are single-center studies. Prosthetic heart valves have a limited effective orifice area (EOA), although supra-annular type prostheses with a larger EOA are commercially available. People in the Eastern countries are physically smaller than those in the Western countries, and it is not always easy to implant a prosthetic valve of an adequate size in a smaller patient. Thus, the influence of PPM on clinical results after AVR should be clarified. We retrospectively examined PPM in patients with isolated AVR in Japan.
Methods
Study population
The Japan Adult Cardiovascular Surgery Database (JACVSD) was initiated in 2000 to evaluate surgical outcomes after cardiovascular procedures in many centers throughout Japan. The JACVSD currently captures clinical information from nearly half of all the Japanese hospitals performing cardiovascular surgery. The data collection form has a total of 255 variables (definitions are available online at http://www.jacvsd.umin.jp), and these are almost identical to those of the Society of Thoracic Surgeons (STS) National Database (definitions are available online at http://sts.org). The JACVSD has developed a software for the web-based data collection system that enables the data manager of each participating hospital to submit their data electronically to the central office. Although participation in the JACVSD is voluntary, data completeness is a high priority. Accuracy of submitted data was maintained by data auditing, which involved monthly visits by administrative staff to participating hospitals to check data against clinical records. Validity of data was further confirmed by an independent comparison of the volume of cardiac surgery at specific hospitals entered in the JACVSD, compared with that reported in the Japanese Association for Thoracic Surgery annual survey [1].
We examined all patients with isolated AVR between January 1, 2008 and December 31, 2009. First, the JACVSD records that had been obtained without informed consent from the patient were excluded. Records with missing age (or age out of range), missing sex, or missing 30-day status were also excluded. After this data cleaning, the population for this analysis consisted of 3,609 cases from 167 participating institutes throughout Japan. Short-term mortality was defined as death from any cause within 30 days after operation, if the patient was discharged from hospital, or death at any time if the patient was not discharged. Using the definition from a previous study [2, 3], major morbidity was defined as any of the following five postoperative in-hospital complications: stroke, re-operation for any reason, mechanical ventilation for more than 24 h after surgery, renal failure, or deep sternal wound infection.
Statistical analysis
The predicted EOA was obtained from the catalog of each model and size of prosthesis implanted (Table 1). The EOA index (EOAI) was determined by the EOA of the implanted prosthetic valve (cm2) divided by the patient’s preoperative body surface area (m2). PPM was defined as an EOAI of ≤0.85 as described and validated in recent studies [4–6].
Table 1.
n a | (%) | Prosthetic valve size (mm) | ||||||
---|---|---|---|---|---|---|---|---|
17 | 19 | 21 | 23 | 25 | 27 | |||
Stented bioprosthetic valves | ||||||||
Carpentier–Edwards PERIMOUNT pericardial Magna bioprosthesis | 1214 | 33.6 | 1.58 | 1.90 | 2.07 | 2.33 | – | |
Carpentier–Edwards PERIMOUNT pericardial bioprosthesis | 722 | 20 | 1.24 | 1.45 | 1.63 | – | – | |
Carpentier–Edwards standard porcine bioprosthesis | 67 | 1.9 | 0.85 | 1.48 | 1.69 | 1.94 | 2.25 | |
Carpentier–Edwards supra-annular aortic porcine bioprosthesis | 18 | 0.5 | 1.10 | 1.40 | 1.50 | 1.70 | 1.90 | |
Carpentier–Edwards PERIMOUNT plus pericardial bioprosthesis | 2 | 0.06 | 1.10 | 1.40 | 1.50 | 1.70 | 1.90 | |
Medtronic intact porcine bioprosthesis | 1 | 0.03 | 0.85 | 1.02 | 1.27 | 1.40 | 1.66 | |
Stentless bioprosthetic valves | ||||||||
Medtronic freestyle stentless porcine bioprosthesis—subcoronary | 7 | 0.2 | 1.20 | 1.42 | 1.70 | 2.15 | 2.49 | |
Mechanical prosthesis | ||||||||
St. Jude medical regent valve | 679 | 18.8 | 1.30 | 1.70 | 2.00 | 2.50 | 2.60 | 3.50 |
ATS mechanical prosthesisb | 313 | 8.6 | 1.20 | 1.50 | 1.70 | 2.10 | 2.50 | |
MCRI On-X mechanical prosthesis | 199 | 5.5 | 1.50 | 1.70 | 2.00 | 2.40 | 3.20 | |
St. Jude Medical mechanical heart valve | 130 | 3.6 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
St. Jude Medical Masters Series mechanical heart valve | 66 | 1.8 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
CarboMedics standard aortic valve | 51 | 1.4 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
CarboMedics reduced cuff aortic valve | 47 | 1.3 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
St. Jude Medical Mechanical Heart Valve Hemodynamic Plus Series | 23 | 0.64 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | 2.70 |
Sorin Bicarbon (Baxter Mira) mechanical prosthesis slimline | 22 | 0.61 | 1.01 | 1.50 | 1.90 | |||
CarboMedics Top-Hat supra-annular aortic valve | 21 | 0.58 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
CarboMedics mechanical prosthesis | 19 | 0.53 | 1.00 | 1.30 | 1.60 | 1.80 | 2.40 | |
CarboMedics small adult aortic and mitral valves | 7 | 0.19 | 1.00 | |||||
Edwards Tekna mechanical prosthesis | 1 | 0.03 | 0.90 | 1.20 | 1.30 | 1.40 | – |
aNumber of patients (%) in the present study
bATS prosthetic valve includes two series of AP and the Standard series. EOA of AP18 is equivalent to Standard 21, AP20 is equivalent to Standard 23
We examined the differences between PPM and non-PPM groups using bivariate tests: the Fisher’s exact test and Chi-squared test for categorical covariates, and the unpaired t test or Wilcoxon rank-sum test for continuous covariates. Descriptive data are expressed as mean ± 1 standard deviation. The level of statistical significance was set at 0.05 (two-sided).
To examine effects of PPM, we conducted multivariate logistic regression analysis for each outcome [7]. In this analysis, each outcome was entered as an independent variable (short-term mortality, major morbidity), and preoperative patient risk factors in the JACVSD valve risk models were entered as dependent variables [8].
Results
Based on preoperative and operative data and the aforementioned definitions, 8.5 % of patients (306/3,609) had PPM. Comparisons between the PPM and the non-PPM groups are presented in Table 2. Body surface area (BSA) was larger and body mass index (BMI) was higher in the PPM group than the non-PPM group (P < 0.001). Patients with PPM were older (P = 0.001) and had a higher prevalence of diabetes (P = 0.004), dyslipidemia (P < 0.001), hypertension (P < 0.001), cerebrovascular disease (CVD P = 0.031), old myocardial infarction (P = 0.006), previous percutaneous coronary artery intervention (P = 0.001), three-vessel coronary artery disease (P = 0.018), and predominant aortic valve stenosis (P < 0.001). Patients in the non-PPM group had a higher prevalence of moderate or severe aortic insufficiency (P < 0.001).
Table 2.
Variable | Non-PPM (n = 3,303) | PPM (n = 306) | p value |
---|---|---|---|
Preoperative data | |||
Age | 68 ± 12 | 70 ± 10 | 0.001* |
Female | 1505 (45.6) | 156 (51.0) | 0.072 |
Body surface area, m2 | 1.55 ± 0.19 | 1.61 ± 0.18 | <0.001* |
Body mass index | 22.7 ± 3.56 | 24.2 ± 3.64 | <0.001* |
Body mass index > 25 | 100 (3.0) | 20 (6.5) | <0.001* |
NHYA functional class ≧III | 638 (19.3) | 71 (23.2) | 0.114 |
Smoking | 1118 (33.8) | 107 (35.0) | 0.705 |
Hypertension | 2090 (63.3) | 224 (73.2) | <0.001* |
Dyslipidemia | 1192 (33.1) | 140 (45.8) | <0.001* |
Diabetes | 543 (16.4) | 71 (23.2) | 0.004* |
Renal failure | 347 (10.5) | 22 (7.2) | 0.075 |
Dialysis | 226 (6.8) | 13 (4.2) | 0.092 |
Cerebrovascular disease | 239 (7.2) | 33 (10.8) | 0.031* |
Infective endocarditis | 186 (5.6) | 19 (6.2) | 0.698 |
Chronic lung disease | 91 (2.8) | 9 (2.9) | 0.855 |
Peripheral arterial disease | 172 (5.2) | 13 (4.2) | 0.587 |
Thoracic aortic aneurysm | 91 (2.8) | 10 (3.3) | 0.585 |
Percutaneous coronary intervention | 141 (4.3) | 28 (9.2) | 0.001* |
Old myocardial infarction | 55 (1.7) | 13 (4.2) | 0.006* |
Angina pectoris | 185 (5.6) | 26 (8.5) | 0.055 |
Coronary 1 vessel disease | 145 (4.4) | 19 (6.2) | 0.15 |
Coronary 2 vessel disease | 63 (1.9) | 11 (3.6) | 0.056 |
Coronary 3 vessel disease | 26 (0.8) | 7 (2.3) | 0.018* |
Predominant aortic valve stenosis | 2062 (62.4) | 245 (80.1) | <0.001* |
Aortic valve insufficiency > 3 | 13074 (41.6) | 78 (25.5) | <0.001* |
LV ejection fraction < 30 % | 102 (3.1) | 11 (3.6) | 0.606 |
Operative data | |||
Emergent/salvage operation | 43 (1.3) | 6 (2.0) | 0.303 |
Blood transfusion | 2141 (64.8) | 233 (76.1) | <0.001* |
Re-do sternotomy | 146 (4.4) | 13 (4.2) | 1.00 |
Outcome | |||
Re-operation for bleeding | 105 (3.2) | 12 (3.9) | 0.498 |
Stroke | 40 (1.2) | 7 (2.3) | 0.113 |
Renal failure | 130 (3.9) | 20 (6.5) | 0.036* |
Dialysis | 43 (1.3) | 11 (3.6) | 0.005* |
Heart Block | 57 (1.7) | 1 (0.3) | 0.058 |
Prolonged ventilation | 189 (5.7) | 31 (10.1) | 0.004* |
ICU stay > 7 day | 157 (4.8) | 27 (8.8) | 0.004* |
Short-term mortality | 74 (2.2) | 12 (3.9) | 0.076 |
Data are no. of patients (%) or mean ± SD values
NYHA New York Heart Association, LV indicates left ventricular
* Significant (P < 0.05) difference
There were no significant differences between the groups with respect to most of the common operative data, except for perioperative blood transfusion (P < 0.001). Postoperative renal failure (P = 0.036) and dialysis (P = 0.005) were significantly more frequent in the PPM group. Postoperative ventilation (P = 0.004) and intensive care unit stay (P = 0.004) were significantly longer in the PPM group.
The short-term mortality rate was 3.9 % in the PPM group and 2.2 % in the non-PPM group. There was no significant difference between the groups in short-term mortality (P = 0.076).
Discussion
Many articles referring to PPM have been published. PPM is classified according to EOAI as mild-to-moderate (>0.65 to ≤0.85 cm2/m2) or severe (<0.65 cm2/m2). Some studies found that PPM had adverse effects on clinical outcomes, [9–11] and others did not [13–20]. Adverse effects include elevated N-terminal pro B-type natriuretic peptide levels, [9] early stenotic dysfunction of bioprosthetic valves, [10] and delayed left ventricular mass (LVM) regression [11]. Tomoeda et al. [12] found that EOAI should be greater than 0.77 cm2/m2 for adequate LVM regression, but an increasing number of studies have found that LVM regression was not related to EOAI [13–16]. Some investigators reported that postoperative survival was not significantly different between patients with and without PPM [17–20]. Jamieson et al. [19] analyzed 3,343 cases of AVR and found that the predictors of overall mortality were age, age category, New York Heart Association functional class III/IV, concomitant coronary artery bypass graft surgery, prosthesis type, preoperative congestive heart failure, diabetes mellitus, renal failure, and chronic obstructive pulmonary disease. Furthermore, they conclude that EOAI category was not predictive of overall mortality, early mortality, or late mortality.
Aortic root enlargement is sometimes performed to avoid the PPM, but clinical results including increased surgical risks are controversial [21, 22]. Newly developed prosthetic valves with smaller sewing rings and supra-annular implantation techniques contribute to avoiding the PPM. Not surprisingly, the incidence of mismatch also increases with diminishing prosthesis size, and it is widely recognized that patients with a valve size ≤21 mm tend to have much higher gradients. Nonetheless, it must be emphasized that severe mismatch can also occur in patients receiving a prosthesis size >21 mm and that, ultimately, it is always the relation between prosthesis size and body size, rather than each factor taken separately, that determines the final hemodynamic outcome [5].
PPM was observed in 8.5 % patients who underwent isolated AVR during 2008 and 2009 in Japan. Dumesni and PiBarot [11] reported that the rates of PPM were between 20 and 70 %. The low rate of PPM in this series might be due to the physical differences between Japanese people and other people. PPM may be very rare in patients undergoing AVR; therefore, its clinical significance may be less in Japan than previously hypothesized in the Western countries. Thus, aggressive over-sizing or root enlargement strategies may be unwarranted in Japan.
The fact that mismatch occurs more frequently in patients with stenotic native valves and in older patients is also consistent with this concept because patients with stenotic native valves generally have smaller valvular annuli than those with regurgitant valves, and calcific aortic stenosis is by far the most prevalent lesion in older patients undergoing aortic valve replacement. Elderly patients had a higher rate of PPM. PPM could be justified in some elderly patients with lower metabolic requirements and limited physical activity.
Preoperative risk factors for PPM are related to lifestyle-related diseases. The Japan Society for the Study of Obesity defines obesity as BMI ≥25 kg/m2. Our study shows that weight control is mandatory in obese patients to reduce the potential adverse effects of PPM. The prevalence of atherosclerotic aortic valve stenosis has increased over recent years. The risk factors for PPM in this study are consistent with the causes of aortic valve disease. Annular stiffness due to calcification and post-inflammatory changes in patients with aortic valve stenosis might cause an inappropriate valve choice followed by PPM.
The rate of PPM was significantly reduced in Japan during the 2 years studied. New generation prosthetic valves such as the Carpentier-Edwards Magna and the SJM Regent valves might have contributed to this reduction. As the JACVSD records do not include valve implantation techniques (supra-annular or intra-annular position), we cannot determine whether such technical differences contributed to the reduction in the rate of PPM.
There was no difference in early postoperative mortality rates between the two groups, but mechanical ventilation longer than 24 h, renal failure requiring dialysis, and intensive care unit stay longer than 7 days were significantly higher in the PPM group. Postoperative complications including early mortality and major morbidity were significantly more frequent in the PPM group than the non-PPM group, resulting in PPM patients requiring more health resources than non-PPM patients. The high morbidity rates may be caused by postoperative low cardiac output syndrome. As the JACVSD records do not include the treatment details such as catecholamine doses, we cannot determine the causes of the high morbidity rate in the PPM group.
Study limitations
This is a retrospective study, limited to the evaluation of early clinical outcomes. We have used in vitro manufactures’ EOA, which may overestimate in vivo echocardiographic EOA, and we have no data on postoperative transvalvular gradients. The impact of PPM on the functional outcome following AVR is difficult to evaluate because of the confounding effects of concomitant cardiovascular and non-cardiovascular disease. Perioperative cardiac function may be the most important factor for the outcome. We have neither follow-up data on patient functional status nor follow-up echocardiographic data on EOA or left ventricular mass regression. Further studies should be indicated specifically to examine the effect of mismatch on symptomatic improvement and exercise tolerance after aortic valve replacement.
Conclusions
This study demonstrates that age, aortic valve stenosis, dyslipidemia, hypertension, old myocardial infarction, history of percutaneous coronary artery intervention, diabetes mellitus, cerebrovascular disease, and high BMI are the risk factors for PPM in patients undergoing isolated AVR. Perioperative blood transfusion, dialysis, and intensive care unit stay longer than 7 days were more frequent in the PPM group. PPM was not an independent risk factor for short-term mortality in patients undergoing AVR in Japan.
Open Access
This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
References
- 1.Ueda Y, Fujii Y, Udagawa H. Thoracic and cardiovascular surgery in Japan during 2006. Gen Thorac Cardiovasc Surg. 2008;56:365–388. doi: 10.1007/s11748-008-0267-2. [DOI] [PubMed] [Google Scholar]
- 2.Grover FL, Shroyer AL, Edwards FH, Grover FL, Shroyer ALW, Edwards FH, et al. Data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1996;26:1229–1231. doi: 10.1016/0003-4975(96)00589-9. [DOI] [PubMed] [Google Scholar]
- 3.Shroyer AL, Edwards FH, Grover FL. Updates to the data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1998;65:1494–1497. doi: 10.1016/S0003-4975(98)00261-6. [DOI] [PubMed] [Google Scholar]
- 4.Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–14. doi: 10.1016/j.ejcts.2006.03.046. [DOI] [PubMed] [Google Scholar]
- 5.Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–1141. doi: 10.1016/S0735-1097(00)00859-7. [DOI] [PubMed] [Google Scholar]
- 6.Blais C, Dumensnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108:983–988. doi: 10.1161/01.CIR.0000085167.67105.32. [DOI] [PubMed] [Google Scholar]
- 7.Hosmer DW, Lemeshow S. Applied logistic regression. New York: Wiley; 1989. pp. 82–132. [Google Scholar]
- 8.Motomura N, Miyata H, Tsukihara H, Takamoto S, Japan Cardiovascular Surgery Database Organization. Risk Model of Valve Surgery in Japan Using the Japan Adult Cardiovascular Surgery Database. J Heart Valve Dis. 2010;19(6):684–691. [PubMed]
- 9.Melina G, Angeloni E, Benedetto U, Refice S, Miceli A, Miele C, et al. Relationship between prosthesis-patient mismatch and pro-brain natriuretic peptides after aortic valve replacement. J Heart Valve Dis. 2010;19:171–176. [PubMed] [Google Scholar]
- 10.Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121:2123–2129. doi: 10.1161/CIRCULATIONAHA.109.901272. [DOI] [PubMed] [Google Scholar]
- 11.Dumesnil JG, PiBarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13:250–257. doi: 10.1007/s11886-011-0172-7. [DOI] [PubMed] [Google Scholar]
- 12.Tomoeda H, Ueda T, Teshima H, Arinaga K, Tayama K, Fukunaga S, et al. Postoperative left ventricular mass regression after aortic valve replacement for aortic stenosis. Ann Thorac Surg. 2010;89:745–750. doi: 10.1016/j.athoracsur.2009.12.034. [DOI] [PubMed] [Google Scholar]
- 13.Pisano C, D’Amico T, Palmeri C, Franchino R, Fattouch K, Bianco G, et al. Valve prosthesis-patient mismatch: hemodynamic, echocardiographic and clinical consequences. Interact Cardiovasc Thorac Surg. 2011;13:606–610. doi: 10.1510/icvts.2011.283218. [DOI] [PubMed] [Google Scholar]
- 14.Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, et al. Aortic valve replacement with 17-mm mechanical prostheses: is patient-prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–77. doi: 10.1016/j.athoracsur.2010.08.036. [DOI] [PubMed] [Google Scholar]
- 15.Brown ML, Schaff HV, Suri RM, Li Z, Sundt TM, Dearani JA, et al. Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. J Thorac Cardiovasc Surg. 2011;142:e5–e9. doi: 10.1016/j.jtcvs.2010.08.043. [DOI] [PubMed] [Google Scholar]
- 16.Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-advanced performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–1199. doi: 10.1016/j.athoracsur.2009.12.074. [DOI] [PubMed] [Google Scholar]
- 17.Cotoni DA, Palac RT, Dacey LJ, O’Rourke DJ. Defining patient-prosthesis mismatch and its effect on survival in patients with impaired ejection fraction. Ann Thorac Surg. 2011;91:692–699. doi: 10.1016/j.athoracsur.2010.11.033. [DOI] [PubMed] [Google Scholar]
- 18.Howell NJ, Keogh BE, Ray D, Bonser RS, Graham TR, Mascaro J, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–64. doi: 10.1016/j.athoracsur.2009.07.037. [DOI] [PubMed] [Google Scholar]
- 19.Jamieson WR, Ye J, Higgins J, Cheung A, Fradet GJ, Skasgard P, et al. Effect of prosthesis-patient mismatch on long-term survival with aortic valve replacement: assessment to 15 years. Ann Thorac Surg. 2010;89:51–58. doi: 10.1016/j.athoracsur.2009.08.070. [DOI] [PubMed] [Google Scholar]
- 20.Okamura H, Yamaguchi A, Noguchi K, Naito K, Yuri K, Adachi H. Hemodynamics and outcomes of aortic valve replacement with a 17- or 19-mm valve. Asian Cardiovasc Thorac Ann. 2010;18:450–455. doi: 10.1177/0218492310381174. [DOI] [PubMed] [Google Scholar]
- 21.Coutinho GF, Correia PM, Paupério G, de Oliveira F, Antunes MJ. Aortic root enlargement does not increase the surgical risk and short-term patient outcome? Eur J Cardiothorac Surg. 2011;40:441–447. doi: 10.1016/j.ejcts.2010.11.064. [DOI] [PubMed] [Google Scholar]
- 22.Kulik A, Al-Saigh M, Chan V, Masters RG, Bédard P, Lam BK, et al. Enlargement of the small aortic root during aortic valve replacement: is there a benefit? Ann Thorac Surg. 2008;85:94–100. doi: 10.1016/j.athoracsur.2007.07.058. [DOI] [PubMed] [Google Scholar]