Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2013 Oct 16.
Published in final edited form as: Environ Sci Technol. 2012 Sep 25;46(20):11213–11219. doi: 10.1021/es300253p

Evaluation of an Arsenic Test Kit for Rapid Well Screening in Bangladesh

Christine Marie George 1,*, Yan Zheng 2,**, Joseph H Graziano 1, Shahriar Bin Rasul 3, Zakir Hossain 4, Jacob L Mey 5, Alexander van Geen 5
PMCID: PMC3653972  NIHMSID: NIHMS410231  PMID: 22866936

Abstract

Exposure to arsenic in groundwater via drinking remains unabated for millions of villagers in Bangladesh. Since a blanket testing campaign using test kits almost a decade ago, millions of new wells have been installed but not tested, thus affordable testing is needed. The performance of the Arsenic Econo-Quick (EQ) kit was evaluated by blindly testing 123 wells in Bangladesh and comparing with laboratory measurements; 65 wells were tested twice. A subset of the same 123 wells was also tested using the Hach EZ kit in the field and the Digital Arsenator in the laboratory in Bangladesh. The EQ kit correctly determined the status of 110 (89%) and 113 (92%) out of 123 wells relative to the WHO guideline (10 μg/L) and the Bangladesh standard (50 μg/L), respectively. Relative to the WHO guideline, all misclassifications were underestimates for wells containing between >10 and 27 μg/L As. Relative to the Bangladesh As standard, over- and under-estimates were evenly distributed. Given its short reaction time of 10 min relative to the Hach EZ and its lower cost compared to the Arsenator, the EQ kit appears to have several advantages for well testing in Bangladesh and elsewhere.

Keywords: Arsenic, groundwater, test-kit, tubewell, screening, Bangladesh

INTRODUCTION

Concerns about elevated arsenic (As) concentrations in Bangladesh groundwater were first raised in the mid-1990s. As of 2009, an estimated 22 million people were still drinking water that does not meet the Bangladesh Arsenic (As) standard of 50 μg/L and 5.6 million were exposed to As above 200 μg/L 1. Exposure to elevated levels of inorganic As is associated with cancers of the skin, bladder, and lung 2-4, developmental effects in children 5; 6, cardiovascular disease 7; 8, and skin lesions 9; 10.

The most common action taken by villagers in Bangladesh to reduce As exposure over the past decade has been to switch to a neighboring well that is low in As. This was made possible by a combination of (a) blanket testing of close to 5 million wells with test kits, mostly the Hach EZ, throughout the affected regions between 2000 to 2005 and (b) the spatially heterogeneous distribution of As in groundwater at a spatial scale of a village 11. In Araihazar upazilla (sub-district), it has been shown that 90% of the residents lived within 100 meters of a low-As well even though close to 50% of the wells were high in As within the same area 12. The installation of deep tubewells is the second most common form of As mitigation in Bangladesh11; 13.

In many regions of Bangladesh it has been more than six years since the previous nationwide water As testing program was conducted under the Bangladesh Arsenic Mitigation and Water Supply Program (BAMWSP) 14 However, the pace of new well installations has not abated markedly and the proportion of untested wells has therefore been growing 15. A survey conducted in Araihazar in 2005 has shown that more recently installed wells were no more likely to be low in As than older wells 16. The Multiple Cluster Indicator Survey in 2006 has shown that the As status of 38% of wells was unknown in Bangladesh 17, and this has increased to 44% by 2009 18. In a survey conducted in Singair upazilla, the same district where the present study was conducted, 80% of the surveyed households reported in 2009-2010 that their wells had not been tested previously 15. There is a renewed and urgent need to redirect households from high- to low-As wells by testing these newly installed and untested wells which could number in the millions. An affordable and reliable test kit could also help establish a testing capacity locally available to community throughout the country to monitor periodically low-As water sources, including deep community wells.

The growing proportion of untested wells, and the exposure of villagers in Bangladesh to As resulting from continuing tubewell installations, motivated this evaluation of a field test kit, the Arsenic Econo-Quick (EQ) introduced by Industrial Test Systems Inc. (http://www.sensafe.com/). The new kit appeared promising because the prescribed reaction time of 10 min was short and the cost was low ($0.17/test for a large-quantity order by UNICEF in Bangladesh; $0.60/test list price in the US). Several studies have been conducted evaluating the effectiveness of As test kits yielding widely varying results19; 20. Ideally, a test kit for As should be light and compact, be easy to use, require a short reaction time, generate minimum quantities of chemical wastes, and be able to accurately measure As concentrations relative to the World Health Organization guideline of 10 ug/L as well as the higher Bangladesh standard for As in drinking water.20

METHODS

Recruitment and Sampling

Village workers were recruited by the Christian Commission for Development Bangladesh (CCDB), a local non-governmental organization, to sample tubewells and deploy the EQ and Hach EZ Arsenic (EZ) test kits. Their educational level ranged from completion of secondary school certificate to higher secondary school certificate (Grades 8-13). A total of 123 untested tubewells were randomly selected for testing (twice for a subset of 65 wells) with field kits in villages of Singair and Shibalaya upazilas, within the Manikhanj district of Bangladesh. When tubewells were tested in the field more than once, village workers were blinded to the previous results. Each well was tagged with a numbered metal placard for identification. Groundwater from all wells was collected in 20 mL scintillation vials for laboratory analysis. A subset of 60 wells was also tested using the EZ kit using a reaction time that was extended from 20 to 40 min, following the demonstration that this modification reduced the likelihood of classifying a well as meeting the Bangladesh standard for As in drinking water of 50 μg/L when it did not 21. In addition, a subset of 92 well water samples were tested in the laboratory with the Digital Arsenator (Wagtech).

Field Measurements

Although the principle of detection is the same for the three kits that were evaluated, the classic 19th century Gutzeit method, the procedures and reagents used differ (Table 1) 22. The first reagent of the EQ kit (Part no. 481298), added with a scoop to a 50 mL water sample, is tartaric acid amended with small amounts of iron and nickel sulfate, presumably to accelerate the reaction. A second reagent (potassium peroxymonosulfate) provided with the EQ kit to oxidize hydrogen sulfide that could potentially suppress the signal was not used. Only hydrogen sulfide at >10−6 M levels appears to interfere with the measurement and such levels can be ruled out by smell for the majority of groundwater pumped from tubewells in Bangladesh. Skipping this step reduces the total reaction time from 12 to 10 minutes. Unlike the EZ kit, the EQ kit includes a temporary cap for shaking the sample to ensure that the tartaric acid dissolves completely before the next reagent, Zn powder, is added with another scoop. The reference chart provided with the EQ kit displays the yellow to brown range of colors expected for As concentrations of 0, 10, 25, 50, 100, 200, 300, 500, and 1000 μg/L. Village workers were instructed not to interpolate their readings between categories but instead to select the As concentration on the chart that matched the color of the test strip most closely. In the few cases that the village workers did interpolate, the reading was converted to the closest reference concentration on the strip and, in the even fewer cases when the reported value was exactly midway between two reference concentrations, the reading was converted to the higher value.

Table 1.

Comparison of Arsenic Field Test Methods

Kits Tested in this Study Kits Tested by Others
ITS Econo Quick Hach EZ Wagtech
Digital Arsenator
Merck Sensitive Lyophilized Bioreporter
Bacteria
Volume (mL) 50 50 50 50 1
Reagents Tartaric acid amended with
small amounts of iron and
nickel sulfate, , Zn powder,
mercuric bromide strip,
potassium
peroxymonosulfate (optional)
Sulfamic acid crystals, Zn
powder, mercuric bromide
strip
Sulfamic acid, sodium
borohydride, Zn powder,
mercuric bromide strip
Sulfuric acid, Zn powder,
mercuric bromide strip
Luminescent whole cell
living
bacterial biosensor
Reaction Time
(minutes)
10-121 20-402 20-403 N/A 4.5 hours (150 samples run in
parallel )12
Cost per test
(USD)
~0.6 ~0.6 ~6.64 ~0.504 N/A
Arsenic Readings
(μg/L As)
0, 10, 25, 50, 100, 200, 300,
500, 1000
0, 10, 25, 50, 100, 250, 500 Continuous 20, 50, 100, 200, 500 Continuous
Performance5 Underestimates
(%)
Overestimates
(%)
Underestimates
(%)
Overestimates
(%)
Underestimates
(%)
Overestimates
(%)
Underestimates
(%)
Overestimates
(%)
Underestimates
(%)
Overestimates
(%)
Relative to 10 μg/L 11%6 0%6 3%6, 4%7,
5%8
2%6, 1%7,
2%8
3%6, 0%9,
10%10
7%6, 9%9,
7%10
1%11 3%11 0%12 8%12
Relative to 50 μg/L 4%6 4%6 2%6, 4%7,
1%8
3%6, 1%7,
6%8
4%6, 10%9,
1%10
2%6, 6%9,
4%10
1%11 3%11 4%12 4%12
1

Twelve minutes is necessary if the sulfur interference step is used.

2

Twenty minutes is recommended by the arsenic test kit manufacturer , however Van Geen et al. 2005 demonstrated that a 40 minute react ion period reduces in consistencies in the 50-100 μg/L range.

3

If the arsenic concentrat ion is below 80 μg/L then the react ion period is 20 minutes, otherwise the sample must be diluted and analyzed again result ing in a react ion t ime of 40 minutes.

4

Ahammadul Kabir. Rapid Review of Locally Available Arsenic Field Test ing Kits. DFID. The est imates for the Wagtech Digital Arsenator include equipment costs. Report

5

Comparison of well status relative to the WHO guideline of 10 μg/L and Bangladesh standard for As in drinking water of 50 μg/L assigned by the four kits. Percentage is of the total number of test kit samples collected. Studies:

6

Present Study

7

Van Geen et al 2005

8

Steinmaus et al 2006

9

Sankararakrishnan et al 2008.

10

Safrazadeh-Amiri et al 2011

11

Jakariya et al. 2007

12

Siegfried et al 2012

The Hach EZ kit (Part No. 2822800; current list price in the US of $0.60/test) was used for the majority of the tubewells tested under BAMWSP. The EZ kit relies on sulfamic acid crystals to acidify a 50 mL sample. A procedure intended to eliminate interference by hydrogen sulfide, in this case cotton impregnated with Pb acetate, was also eliminated in this study. Village workers were instructed to use a 40 min reaction time and reported the results as 0, 10, 25, 50, 100, 250, or 500 μg/L As. Here too, readings were converted to the nearest reference concentration on the strip when interpolated concentrations were reported.

Laboratory Measurements

Groundwater samples collected in 20 mL scintillation vials were acidified to 1% with high-purity Optima HCl at Lamont-Doherty Earth Observatory at least 48 hours before analysis. This has been shown to ensure re-dissolution of any As that could have adsorbed to precipitated Fe oxides 16. Water samples were then diluted 1:10 in a solution spiked with 73Ge for internal drift correction and analyzed for As by high-resolution inductively-coupled plasma mass spectrometry (HR ICP-MS), which eliminates the isobaric interference of ArCl. Further details are provided elsewhere 21; 23. The detection limit for As is typically <0.2 μg/L, estimated here by multiplying the As concentration corresponding to the blank by a factor of 3. The long-term reproducibility determined from consistency standards included with each run averaged 4% (1-sigma) in the 40-500 μg/L range. This is comparable to the previously reported error estimate for single measurements by HR ICP-MS of 4 μg/L augmented by 2% of the measured concentration.

Although it is designed to be deployed in the field, the Digital Arsenator (Wagtech Part No. WAG-WE10500) was used in the laboratory, as is typically the case in Bangladesh. A subset of 92 well water samples tested with at least one of two other kits were collected in plastic 60 mL bottles. Before analysis, the samples were acidified with 0.3-0.5 ml of 1:1 HCl to ensure redissolution of any precipitated Fe oxides. The Arsenator relies on additions of sulfamic acid and sodium borohydride to a 50 mL sample to generate AsH3 over a 20 min reaction time, but additional steps in the procedure increase total processing time to approximately 40 min if the water arsenic concentration is found to exceed 100 μg/L. If quantification above an As concentration of 100 μg/L is desired using the Arsenator, the sample is diluted and reanalyzed. In addition to its significantly higher purchase price ($1800 for the reading unit and $1/test for reagents), the Arsenator differs from the EQ and EZ kit in that the color of a test strip is measured with a digital reader instead of being estimated visually.

RESULTS AND DISCUSSION

ICP-MS Data

Concentrations of As measured in groundwater from 123 tubewells by HR ICP-MS ranged from 0.1 to 452 μg/L, with a mean of 60 μg/L. The set of samples was roughly evenly split between 51 (41%) tubewells containing 0.1-10 μg/L As (and meeting the WHO guideline for drinking water), 38 (31%) tubewells with 10-50 μg/L As that do not meet the WHO guideline but still meet the Bangladesh standard, and 34 (28%) tubewells with >50 μg/L As. In this analysis, ICP-MS data are used as the reference to compare the performance of the field kits.

Performance of the EQ Kit

Readings in the field using the EQ kit were identical for 47 out of 65 wells that were analyzed twice. For only one out of the 18 remaining duplicates did the readings differ by more than one interval. Relative to the WHO guideline of 10 μg/L, the EQ kit correctly determined the status of 110 (89%) out of 123 wells (Table 1). All 13 misclassifications relative to the WHO guideline were underestimates for wells containing between 10 and 27 μg/L range As, the only category (10-25 μg/L) that the EQ kit performed significantly worse than the EZ kit and the Arsenator (Fig. 1). The EQ kit correctly determined the status of 113 (92%) out of 123 wells relative to the Bangladesh standard of 50 μg/L. One of the misclassifications was an underestimate for a well containing 193 μg/L according to duplicate EQ kit measurements as well as EZ and Arsenator determinations. This suggests that the sample bottle collected from this well for ICP-MS analysis was most likely mislabeled in the field. Excluding this well, four of the EQ misclassifications were overestimates for wells containing 25-50 μg/L As and the other five were underestimates with three samples in the 50-100 μg/L range and two in the >100μg/L range. Relative to either 10 μg/L or 50 μg/L, the EQ kit correctly identified 91% to 100% of well waters for all five categories of As concentrations as measured by ICP-MS (Fig. 1).

Figure 1.

Figure 1

Comparison of ICP-MS and Field Kit Arsenic Results Relative to 50 μg/L and 10 μg/L.

Performance of the EZ Kit

Relative to 10 μg/L and 50 μg/L, the EZ kit underestimated the As content of only 2 wells (3%) and 1 well (2%), respectively, out of a total of 60 that were tested (Table 1). The number of wells for which the As content was overestimated using the EZ kit relative to either threshold was 1 (2%) and 2 (3%), respectively (Table 1). Relative to either 10 μg/L or 50 μg/L, the EZ kit correctly identified 97% to 100% of well waters for all five categories of As concentrations as measured by ICP-MS (Fig. 1).

Comparison with Laboratory Measurements

A different way to evaluate the performance of the EQ and EZ kits is to compare concentrations inferred from the visual readings across the entire range of ICP-MS measurements. Although such a comparison is informative, it is less relevant to public health and policy than a binary classification described above. For this comparison, the boundary between each range of As concentrations was set mid-way between each of the readings illustrated on the two kit’s reference charts (Figure 2). The rationale is that the actual As concentration is as likely to be slightly below or as slightly above the reported reading. In the case of the EQ kit, the resulting 8 categories are 0-5, 5-17.5, 17.5-37.5, 37.5-75, 75-150, 150-250, 250-350, and ≥500 ug/L. Even when considering these relatively wide ranges and 2-sigma error estimates for the ICP-MS measurements, the EQ kit consistently overestimates the As content of well water above 150 ug/L by about a factor two (Fig. 2a). For the EZ kit, the first 4 categories are the same as for the EQ kit and the next 3 are 75-175, 175-375 and ≥500 μg/L. Unlike the EQ kit, discrepancies between EZ kit readings and ICP-MS measurements are not systematically distributed relative to the line corresponding to an exact match (Fig. 2b).

Figure 2.

Figure 2

Comparison of As concentrations in water samples measured by inductively-coupled plasma mass spectrometry (ICP-MS) compared with the outcome of field and laboratory testing with three different kits. One tubewell with an ITS EQ reading of 1000 μg/L and an actual concentration of 395 μg/L is excluded from (a) for clarity. The data in (d) are an expanded version of the same data shown in (c). Horizontal bars indicate the estimated 2-sigma errors for HR ICP-MS measurements (22). Vertical error bars in (a) and (b) indicate the full range of As concentrations ranges for the ITS EQ and the EZ kit listed in the text, respectively. Vertical errors bars in (c) and (d) correspond to an estimated error of ±10% of the reported Arsenator readings. The one-to-one relationship indicating a perfect match is shown as a dotted line.

Performance of the Arsenator

Correspondence between the Arsenator and ICP-MS measurements (Fig. 2c) is improved relative to either of the kits used in the field, including for As concentrations in the 0-80 μg/L range (Fig. 1d). The 4 clear outliers, two of which stand out based on EQ and EZ kit readings as well, likely indicate mislabeling in the field, and possibly exchanged labels. The Arsenator was comparable with the EZ and EQ kits with respect to classify the status of wells relative to the 10 and 50 μg/L thresholds but has the worst performance (93%) for well water with < 10 μg/L [As] as determined by ICP-MS (Fig. 1).

Previous Studies

In Jakariya et al. 2007, an evaluation of the Merck sensitive kit (Table 1) in comparison to laboratory measurements was conducted for 12,532 tubewells in Matlab, Bangladesh24. The proportion of underestimates and overestimates during this survey were low relative to the WHO guideline, 1% and 3%, and Bangladesh arsenic standard, 1% and 3%, respectively. Two previous evaluations were conducted using Hach EZ test kit and had comparable findings to our present study, the first evaluation was conducted in Bangladesh 21, and the second was conducted in the United States in Fallon, Nevada 25. Percent underestimates relative to the WHO and Bangladesh As standards were 4% 21 and 5% 25, and 4% and 1%, respectively; and overestimates were 1% and 2%, and 1% and 6%, respectively. The Wagtech digital arsenator has been used in two previous studies in Bangladesh 26; 27. The proportions reported by Sankararakrishnan et al. were slightly higher at 10% for underestimates (compared to 4% found in the present study) and 6% for overestimates (compared to 2%)26. In contrast, Safarzadeh-Amiri et al reported a higher proportion of underestimates relative to 10 μg/L than found in the present study (10% vs. 3%) and a lower proportion of underestimates relative to 50 μg/L (1% vs. 4%)27. Using a novel form of lyophilized bioreporter bacteria, Siegfried et al. presented field kit results that were comparable in terms of performance to the three kits evaluated in the present study28. However, deploying this kit requires considerably more training and seems cumbersome to use in the field.

Practical Implications

Past debates over the usefulness of field kits for testing the As content of tubewell water in Bangladesh and other affected countries have been fraught in part with the notion that it is important to be able to distinguish concentrations around the Bangladesh standard of 50 μg/L. There is essentially no known threshold below which As exposure has no deleterious health effects and, without evidence to the contrary, the impact should be assumed to be proportional to dose. Nevertheless, correct classification relative to either the 10 μg/L or the 50 μg/L threshold has implications because test results have been shown to be dominant key factor determining whether a household switches to a different well 29; 30. Our results show that all methods evaluated here, the EQ, EZ kits and the Digital Arsenator, are quite similar with respect to underestimating or overestimating As concentrations relative the Bangladesh standard. The comparison shows that the Arsenator can provide a relatively inexpensive form of quality control for field kit measurements.

Relying on the EQ does increase the chance relative to the EZ kit and Arsenator that a well whose As content is marginally above the WHO guideline will be considered safe because of its singular tendency to under-estimate the 10-27 μg/L range as < 10 μg/L. Further, the EQ slightly increases the chance relative to the EZ kit and the Arsenator that a well whose As content is below the Bangladesh standard for As will be considered unsafe. This could be a serious shortcoming only in those villages where the proportion of unsafe wells is particularly high, as it would reduce the opportunity for switching among private wells. Although overestimation of As concentrations above 150 μg/L by the EQ kit has no implications for classifying wells relative to the WHO guideline or the Bangladesh standard, recalibration of the reference color strip is recommended. The EQ kit has the advantages of a shorter reaction time and a lower cost relative to the EZ kit and the Arsenator.

Significance

The Bangladesh Arsenic Mitigation Water Supply Project sponsored by the World Bank, UNICEF, and other organizations between 2001 and 2004 was the largest of its kind to test for As in well water in any country. A significant proportion of these tubewells were probably incorrectly classified as safe relative to the Bangladesh standard of 50 μg/L because the manufacturer’s recommended reaction time of 20 minutes was used 21. Considering that a 2009 national survey conducted by UNICEF and the Bangladesh Bureau of Statistics has found that nearly half of the wells in the country were untested 18, there is an urgent need for expanding the availability of well testing at the village level.

The Ministry of Local Government and Rural Development Cooperatives of Bangladesh, in collaboration with UNICEF and several other developmental agencies, recently piloted a pay-for-use (fee-based) well-testing program for As through the local government in 8 upazillas of Bangladesh. In an evaluation of the program conducted in Meherpur Sadar upazilla, it was found that a majority of households were switching to drinking water sources identified by pay-for-use testing to be safe with respect to As 31; 32. The advantage of pay-for-use testing is that it provides a financial incentive for the tester to seek out untested wells. An expansion of this testing program at the national scale is being planned. Another massive blanket testing campaign that is free of charge would likely again reduce As exposure, but would probably also delay the viability of commercial or subsidized testing for several years. As in the case of the choice of a threshold for distinguishing safe and unsafe wells, the pros and cons of testing-for-a-fee vs. free blanket testing need to be carefully weighed.

Acknowledgements

This work benefited from support by UNICEF Bangladesh and grant P42 ES 10349 from the U.S. National Institute of Environmental Health Sciences. CMG conducted the field work while a UNICEF consultant under the guidance of YZ. The views expressed here are those of the authors and not those of UNICEF. We would also like to acknowledge the hard work of our field team: Masud Noor (Study Coordinator); Rubina Akter; Nasrin Akter; Selina Akter; Anisur Rahaman Kan. This is Lamont-Doherty Earth Observatory contribution number 7575.

References

  • 1.UNICEF, Bangladesh. Bureau.of.Statistics Bangladesh National Drinking Water Quality Survey of 2009. 2009.
  • 2.Marshall G, Ferreccio C, Yuan Y, Bates MN, Steinmaus C, Selvin S, Liaw J, Smith AH. Fifty-year study of lung and bladder cancer mortality in Chile related to arsenic in drinking water. J Natl Cancer Inst. 2007;99:920–928. doi: 10.1093/jnci/djm004. [DOI] [PubMed] [Google Scholar]
  • 3.Chen Y, Ahsan H. Cancer burden from arsenic in drinking water in Bangladesh. Am J Public Health. 2004;94:741–744. doi: 10.2105/ajph.94.5.741. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Morales KH, Ryan L, Kuo TL, Wu MM, Chen CJ. Risk of internal cancers from arsenic in drinking water. Environ Health Perspect. 2000;108:655–661. doi: 10.1289/ehp.00108655. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Calderon J, Navarro ME, Jimenez-Capdeville ME, Santos-Diaz MA, Golden A, Rodriguez-Leyva I, Borja-Aburto V, Diaz-Barriga F. Exposure to arsenic and lead and neuropsychological development in Mexican children. Environ Res. 2001;85:69–76. doi: 10.1006/enrs.2000.4106. [DOI] [PubMed] [Google Scholar]
  • 6.Wasserman GA, Liu X, Parvez F, Factor-Litvak P, Ahsan H, Levy D, Kline J, van Geen A, Mey J, Slavkovich V, et al. Arsenic and manganese exposure and children’s intellectual function. Neurotoxicology. 2011;32:450–457. doi: 10.1016/j.neuro.2011.03.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Chen Y, Factor-Litvak P, Howe GR, Graziano JH, Brandt-Rauf P, Parvez F, van Geen A, Ahsan H. Arsenic exposure from drinking water, dietary intakes of B vitamins and folate, and risk of high blood pressure in Bangladesh: a population-based, cross-sectional study. Am J Epidemiol. 2007;165:541–552. doi: 10.1093/aje/kwk037. [DOI] [PubMed] [Google Scholar]
  • 8.Chen Y, Graziano JH, Parvez F, Liu M, Slavkovich V, Kalra T, Argos M, Islam T, Ahmed A, Rakibuz-Zaman M, et al. Arsenic exposure from drinking water and mortality from cardiovascular disease in Bangladesh: prospective cohort study. BMJ. 2011;342:d2431. doi: 10.1136/bmj.d2431. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Haque R, Mazumder DN, Samanta S, Ghosh N, Kalman D, Smith MM, Mitra S, Santra A, Lahiri S, Das S, et al. Arsenic in drinking water and skin lesions: dose-response data from West Bengal, India. Epidemiology. 2003;14:174–182. doi: 10.1097/01.EDE.0000040361.55051.54. [DOI] [PubMed] [Google Scholar]
  • 10.Ahsan H, Chen Y, Parvez F, Zablotska L, Argos M, Hussain I, Momotaj H, Levy D, Cheng Z, Slavkovich V, et al. Arsenic exposure from drinking water and risk of premalignant skin lesions in Bangladesh: baseline results from the Health Effects of Arsenic Longitudinal Study. Am J Epidemiol. 2006;163:1138–1148. doi: 10.1093/aje/kwj154. [DOI] [PubMed] [Google Scholar]
  • 11.Ahmed MF, Ahuja S, Alauddin M, Hug SJ, Lloyd JR, Pfaff A, Pichler T, Saltikov C, Stute M, van Geen A. Epidemiology. Ensuring safe drinking water in Bangladesh. Science. 2006;314:1687–1688. doi: 10.1126/science.1133146. [DOI] [PubMed] [Google Scholar]
  • 12.Van Geen A, Ahsan H, Horneman AH, Dhar RK, Zheng Y, Hussain I, Ahmed KM, Gelman A, Stute M, Simpson HJ, et al. Promotion of well-switching to mitigate the current arsenic crisis in Bangladesh. Bull World Health Organ. 2002;80:732–737. [PMC free article] [PubMed] [Google Scholar]
  • 13.Van Geen A, Ahmed KM, Seddique AA, Shamsudduha M. Community wells to mitigate the arsenic crisis in Bangladesh. Bulletin of the World Health Organization. 2003;81:632–638. [PMC free article] [PubMed] [Google Scholar]
  • 14.Bangladesh Arsenic Mitigation Water Supply Program. www.bamwsp.org/Survey%20Results.htm.
  • 15.George CM, Graziano JH, Mey JL, van Geen A. Impact on arsenic exposure of a growing proportion of untested wells in Bangladesh. Environmental Health. 2012;11:7. doi: 10.1186/1476-069X-11-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.van Geen A, Cheng Z, Jia Q, Seddique AA, Rahman MW, Rahman MM, Ahmed KM. Monitoring 51 community wells in Araihazar, Bangladesh, for up to 5 years: implications for arsenic mitigation. J Environ Sci Health A Tox Hazard Subst Environ Eng. 2007;42:1729–1740. doi: 10.1080/10934520701564236. [DOI] [PubMed] [Google Scholar]
  • 17.Statistics, U.N.C.s.F.a.t.B.B.o Bangladesh Multiple Indicator Cluster Survey 2006. 2006;Volume 1 [Google Scholar]
  • 18.Pathey P. Monitoring the Situation of Children and Women: Multiple Indicator Cluster Survey 2009. Bangladesh Bureau of Statistics and United Nations Children’s Fund (UNICEF) Volume 1: Technical Report. 2009 [Google Scholar]
  • 19.Feldmann J. Onsite Testing for Arsenic: Field Test Kits. Reviews of Environmental Contamination. 2008;Volume 197:61–75. doi: 10.1007/978-0-387-79284-2_3. [DOI] [PubMed] [Google Scholar]
  • 20.Arora M, Megharaj M, Naidu R. Arsenic testing field kits: some considerations and recommendations. Environmental Geochemistry and Health. 2009;31:45–48. doi: 10.1007/s10653-008-9231-4. [DOI] [PubMed] [Google Scholar]
  • 21.Van Geen A, Cheng Z, Seddique AA, Hoque MA, Gelman A, Graziano JH, Ahsan H, Parvez F, Ahmed KM. Reliability of a commercial kit to test groundwater for arsenic in Bangladesh. Environ Sci Technol. 2005;39:299–303. [PubMed] [Google Scholar]
  • 22.Gutzeit H. Pharm Zeitung. 1879;24:263. [Google Scholar]
  • 23.Cheng Z, Zheng Y, Mortlock R, Van Geen A. Rapid multi-element analysis of groundwater by high-resolution inductively coupled plasma mass spectrometry. Anal Bioanal Chem. 2004;379:512–518. doi: 10.1007/s00216-004-2618-x. [DOI] [PubMed] [Google Scholar]
  • 24.Jakariya M, Vahter M, Rahman M, Wahed MA, Hore SK, Bhattacharya P, Jacks G, Persson LÅ. Screening of arsenic in tubewell water with field test kits: evaluation of the method from public health perspective. Science of the Total Environment. 2007;379:167–175. doi: 10.1016/j.scitotenv.2006.11.053. [DOI] [PubMed] [Google Scholar]
  • 25.Steinmaus CM, George CM, Kalman DA, Smith AH. Evaluation of two new arsenic field test kits capable of detecting arsenic water concentrations close to 10 microg/L. Environ Sci Technol. 2006;40:3362–3366. doi: 10.1021/es060015i. [DOI] [PubMed] [Google Scholar]
  • 26.Sankararamakrishnan N, Chauhan D, Nickson R, Tripathi R, Iyengar L. Evaluation of two commercial field test kits used for screening of groundwater for arsenic in Northern India. Science of the Total Environment. 2008;401:162–167. doi: 10.1016/j.scitotenv.2008.03.042. [DOI] [PubMed] [Google Scholar]
  • 27.Safarzadeh-Amiri A, Fowlie P, Kazi A, Siraj S, Ahmed S, Akbor A. Validation of analysis of arsenic in water samples using Wagtech Digital Arsenator. Science of the Total Environment. 2011 doi: 10.1016/j.scitotenv.2011.03.016. [DOI] [PubMed] [Google Scholar]
  • 28.Siegfried K, Endes C, Bhuiyan AFMK, Kuppardt A, Mattusch J, van der Meer JR, Chatzinotas A, Harms H. Field testing of arsenic in groundwater samples of Bangladesh using a test kit based on lyophilized bioreporter bacteria. Environmental science & technology. 2012 doi: 10.1021/es203511k. [DOI] [PubMed] [Google Scholar]
  • 29.Schoenfeld A. Area, Village, and Household Response to Arsenic Testing and Labeling of Tubewells in Araihazar, Bangladesh. Columbia University; New York City: 2005. [Google Scholar]
  • 30.Opar A, Pfaff A, Seddique AA, Ahmed KM, Graziano JH, van Geen A. Responses of 6500 households to arsenic mitigation in Araihazar, Bangladesh. Health Place. 2007;13:164–172. doi: 10.1016/j.healthplace.2005.11.004. [DOI] [PubMed] [Google Scholar]
  • 31.Horizontal Learning Program Learning Note: Pay-for-use Testing, Marking & Switching of Wells. Department of Public Health Enginnering Bangladesh, Water Sanitation Programme of World Bank, and United Nation’s Children’s Fund; 2011. [Google Scholar]
  • 32.Zheng Y, Ravenscroft P, Rahman SM, Hakim SAI, Ng JC, Noller BC, Naidu R, Bundschudh J, Battacharya P. Understanding the Geological and Medical Interface of Arsenic - As 2012. CRC Press, Taylor & Francis; 2012. Pay-for-use arsenic testing: Promoting demand-driven mitigation and monitoring in Bangladesh; pp. 519–521. [Google Scholar]

RESOURCES