Abstract
Recently, an MRI method based on the Bloch-Siegert (BS) shift phenomenon has been proposed as a fast and precise way to map the radio-frequency (RF) transmit field (B1+ field). For MRI at high field, the mapping sensitivity of this approach is limited by tissue heating associated with the BS irradiation pulse. To mitigate this, we investigated the possibility of lowering the off-resonance frequency of this pulse, as theoretical analysis indicated that the sensitivity of Bloch-Siegert based B1+ mapping can be substantially improved when irradiating closer to resonance. Using optimized irradiation pulse shape and gradient crushers to minimize direct excitation effects, in vivo experiments on human brain at 7T confirmed the improved sensitivity available with this approach. This improved sensitivity translated into an 80% reduction in B1+ estimation errors, without increasing tissue heating.
Keywords: High field MRI, flip angle, wavelength effects, B1 mapping, transmit sense
INTRODUCTION
In MRI, spatial variations in the RF transmit field (the B1+ field, shortened to B1 in the following) can result in imprecise volume selection, and can affect contrast and sensitivity (1). B1 mapping techniques provide a way to quantify this inhomogeneity and take it into account or manipulate it using applications such as transmit calibration (2), RF shimming (1), and parallel transmission (3,4). Because of the continuing increase in the number of available transmit channels and the sophistication of their use for RF excitation, there is an ongoing need for B1 mapping methods with improved speed and accuracy. Especially, a recently proposed phase-based method by Sacolick, et al. (5), which is based on the Bloch-Siegert (BS) frequency shift (6), has been shown to provide robust performance over a range of experimental conditions. Because of their robustness and rapid mapping capability, BS based methods are finding increased use in various applications (7-12). Unfortunately, the speed and sensitivity of the BS methods are often limited by the amount of RF power deposited in the tissue, which may exceed the safety limits for allowable tissue heating (indicated by specific absorption rate - SAR) (13). For this reason, several BS variants (14-20) have been proposed to reduce SAR by changing the imaging approach and image readout scheme, by optimizing the BS irradiation pulse parameters, and by introducing gradient spoiling schemes to eliminate spurious signal excited by the BS pulse. These variants have improved mapping sensitivity, accuracy, and speed (14,18-21). Here, we attempt to further improve B1 mapping performance.
BS methods have traditionally used large frequency offsets for the BS irradiation pulse (relative to both the variation in resonance frequency γB0 over the object and the pulse amplitude γB1) for two reasons: to minimize sensitivity to off-resonance, and to minimize direct excitation of magnetization by this pulse (5). In addition, the use of large frequency offsets facilitates B1 calculation since it allows approximating the relationship between the square of B1 amplitude and the BS frequency shift by a linear function. However, it may be possible to improve on B1 mapping performance by reducing this frequency offset and correcting for the negative side effects mentioned above. To investigate this, we performed theoretical analysis of the BS mapping technique without restricting the irradiation offset. Based on this, we demonstrate that with some sequence modifications, the BS-based sequence can be applied far beyond the regime of offsets originally defined by (5) to achieve lower RF power deposition and better Signal-to-Noise Ratio (SNR) efficiency. Finally, for a gradient-echo (GRE) based BS mapping sequence, the effect of sequence parameters on SNR at a given SAR level is studied. The theoretical findings were then confirmed with studies of human brain at 7T.
METHODS
Basic Equations for BS Shift
BS B1 mapping generally involves subtraction of phase shifts induced by BS pulses at positive and negative off-resonance frequencies ωRF (5). These phase shifts arise from changes in the apparent precession frequency occurring during a BS pulse. As shown in the Appendix I, the frequency difference induced by two BS pulses (ω2BS) at off-resonance frequencies +/-ωRF (ωRF=2πΔfRF>0) in the presence of local static main field (B0) inhomogeneity ΔωBo=2πΔfBo is given by:
(1) |
Under the assumption |ωRF|>|ΔωBo| (Assumption 1, see Appendix I), the BS induced phase shift difference is:
(2) |
for any arbitrary pulse shape B1(t)=B1,peakB1,normalized(t)=B1pB1n(t). In analogy to (5), when |γB1p|<<ωRF , this can be approximated to first order by:
(3) |
SAR Consideration for Constant-Frequency BS Pulse
Because a typical BS pulse is a large flip-angle pulse (18) (generally several 2π rotations), it is usually the dominant contributor to SAR by far, especially for GRE-based sequences. Thus, reducing the energy of the BS pulse is an effective way to reduce the SAR of the B1 mapping sequence. For an arbitrarily shaped BS pulse, we have:
(4) |
where V(t) is the voltage applied at the RF amplifier on the system to generate the pulse.
Assuming ωRF is kept constant over the duration of the pulse T, as is generally done in practical applications, and Assumption 1 holds (|ωRF|>|ΔωBo|), Eq.3 can be further simplified to:
(5) |
leading to:
(6) |
This suggests that the only way to reduce SAR without affecting the SNR in B1 map (see Appendix I) is to reduce ωRF. A drawback however is that at low ωRF, the linear approximation may no longer be valid, and this needs to be accounted for in order to avoid systematic B1 estimation errors. Another important observation from Eq.6 is that under the linear approximation, the energy efficiency of the BS pulse (generated shift per unit SAR) (18) is independent of pulse shape, given:
(7) |
i.e. under linear approximation, the energy efficiency of the BS pulse only depends on the frequency offsets and is independent of the pulse shape.
In practice, potential problems with reducing ωRF are
an increased sensitivity to B0 inhomogeneity (reflected in ΔωBo);
the generation of artifacts originating from magnetization directly excited by the BS pulse;
a reduction in the range of B1 values that satisfies the linear approximation assumption (Eq. 3).
Although each of these may reduce mapping accuracy (5,18,20), this can be avoided by proper adjustments.
Specifically, to counteract the increased sensitivity to B0 inhomogeneity, the distribution of ΔωBo can be measured either simultaneously with an integrated sequence (e.g. (20) and Figure 1b) or with a separate sequence. Based on this, the effect of off-resonance can be accounted for through the use of Eq.2 or via pre-computed lookup tables (18).
Artifacts related to the direct excitation effect may be suppressed by crusher gradients (14,16,19-21). As it turns out, the crushers in BS sequences also help to reduce some of the non-linear effect associated with low frequency offsets due to averaging across magnetization phases. As shown in Appendix II, crushers of sufficient gradient moment reduce the dependency of the BS shift on the imaging flip angle, and this is particularly important when operating under the non-linear regime. As large gradient moments require long crusher durations, which lead to T2*-induced signal loss, crusher moment optimization needs to be performed.
Finally, the effects of operating the BS sequence outside the linear regime can potentially be accounted once the effects of the irradiation pulse have been accurately determined. This can be done through simulations, as will be shown in the following section.
Bloch-Siegert Shift Close to Resonance
When trying to increase mapping efficiency by reducing ωRF, fairly soon one starts violating the condition |γB1p|<<ωRF under which the linear approximation (5) holds, and the relationship between B1p and ϕBS becomes more complicated than Eq.3, even when ignoring the increased direct excitation effects. As illustration, if ΔfRF=500 Hz is used, for any B1p>12μT, we have γB1p>ωRF, and our linearity condition is violated. In the following, we will refer to this situation as “overdriven” Bloch Siegert mapping, or OBS mapping. The situation where the linear condition still holds will be referred to as the “regular” condition.
As shown in Eq. A2.5, even with sufficient crushers, the recovered BS phase shifts (Eq.2) from image phase difference will have a B1 dependency (shown in θ) for large B1. Thus in general, Eq.2 is no longer applicable under the overdriven condition. As a result, derivation of analytical equations becomes very hard, and if at all possible, these equations will be pulse shape dependent.
In the following, the signal evolution during OBS mapping was studied by simulations of the Bloch equations. For this purpose, a previously developed GRE-based sequence (20) (Figure 1b) was used, in which ΔfRF was set to 500Hz for demonstration purpose, given that this value, under practical conditions, can easily satisfy both Assumption 1 and the overdriven condition. Figure 2a shows how ϕ2BS changes as a function of B1p for the hard pulse, Fermi pulse (5), and our pulse (“QDAPX” as shown in Figure 1c, which was previously optimized for stop-band performance at ΔfRF=2kHz) (20). Solid lines represent the results from the Bloch simulation, whereas the dashed lines represent corresponding values calculated from Eq.2. The figure shows that over a large range of B1p levels, ϕ2BS continues to follow the general trend defined by Eq.2 quite well, although at high B1p values, depending on the pulse shape, different degrees of oscillation can be observed. The smoothness of this curve directly relates to the robustness in B1 estimation in the presence of noise in ϕ2BS, since a ϕ2BS-B1 curve with large oscillations has regions where small changes in ϕ2BS correspond to large changes in B1. In addition, the QDAPX pulse is the least sensitive to noise in B0 offsets (Figure 2c). Based on these to facts, the QDAPX pulse was chosen for further optimizations of the OBS sequence.
To optimize the sequence parameters (such as imaging flip angle and repetition time TR) at a constant SAR level, both the SNR and the SAR level acquired in a previous volunteer study (20) were used to calibrate the simulation parameters. In addition, for each simulation, the steady state magnetization level was computed by simulating the imaging sequence repeatedly for durations longer than five times T1 value (T1 was assumed to be 1.5 seconds). Angle-to-Noise Ratio (ANR) efficiency, i.e. mapping sensitivity, was determined from the calculated BS phase shift angle, divided by the thermal noise level and the square root of the scan time. Simulation results for various imaging flip angles at a given SAR level (100%) (Figure 2b) show that for a specific SAR level, optimal ANR is achieved at long TR; furthermore, at a specific TR value, ANR is optimal when using imaging flip angles close to the Ernst angle, (e.g. 30° for TR=200ms and 60° for TR=1000ms for T1=1.5s).
MRI Experimental Setup
The performance of the OBS sequence was demonstrated on three volunteers (approved by Institutional Review Board). All experiments were performed on a Siemens Magnetom 7T (Erlangen, Germany) whole-body scanner based on an Agilent 7T-830-AS (Oxford, UK) shielded magnet design, with a 32-channel receive with a volume-transmit head coil (Nova Medical Inc., Wilmington, MA, USA). Common imaging parameters were: field-of-view=256mm, image matrix size=64×64, slice thickness=5mm, and BS pulse duration T=8ms.
The following single slice experiments were performed on two volunteers:
A reference B1 map based on the original GRE sequence (Figure 1a) was used, with ΔfRF=8kHz Fermi pulse shape, nominal ϕ2BS=70°, γB1p/ωRF=0.068, TR=844ms, TE=11ms, 90° imaging flip angle, 10 averages (to increase SNR). SAR level was at 100% of maximal allowable value;
B1 mapping with the modified sequence (Figure 1b) using QDAPX pulse at ΔfRF=500Hz: OBS B1 mapping with γB1p/ωRF=1.54 and sufficient crusher moments, and regular Bloch-Siegert B1 mapping at γB1p/ωRF=0.37 with and without crusher. TR=500ms, TE=[15,16]ms, 45° imaging flip angle, single repetition;
ANR study with modified sequence on the second volunteer: TR was varied from 74ms to 1000ms (corresponding γB1p/ωRF ranged from 0.42 to 1.54, respectively via increasing B1p), SAR = 55% of maximum, imaging flip angles using Ernst angle calculated from each TR , with QDAPX pulse at ΔfRF=500Hz, with 10 averages to derive the ANR map, which was determined by dividing the mean of the averages by their standard deviation. For comparison, the original sequence using ΔfRF=8kHz Fermi pulse at TR=1000ms was also acquired with 10 averages at 98% maximum SAR. All TE values were kept the same (TE=[14.88ms, 15.88ms]).
In addition, the crusher moment was optimized on one of the volunteers by setting imaging flip angle to 0° and monitoring the signal amplitude while stepping through various crusher moment values, until the direct excitation equaled the noise floor. This ensured that artifacts from direct excitation by the BS pulse were minimized.
To investigate multi-slice BS mapping, which is more challenging than the single slice experiment due to a potential detrimental effect of a BS pulse for a specific slice affecting the magnetization of other slices, the following in vivo data were acquired as well:
Acquisition of a 16-slice B0 map covering the entire brain using double-echo GRE sequence (TE =[3.33, 4.33]ms) with 5mm slice thickness and 5mm gap, TR=500ms, 45° imaging flip angle, single repetition;
Acquisition of a reference B1 map with the original GRE sequence (Figure 1a), with ΔfRF=8kHz Fermi pulse shape, nominal ϕ2BS=33.6°, γB1p/ωRF=0.047, TR=2081ms, 75° imaging flip angle, 10 averages (to increase SNR and derive ANR). SAR level was at 98% of maximal allowable value. Due to the long scan time (4 minutes 26 seconds per repetition) for this experiment, only four 5mm slices were acquired, with 15mm gap between slices. Minimum TE values ([10.33ms 11.33ms]) were used for optimal SNR;
The modified sequence (Figure 1b) using the QDAPX pulse at ΔfRF=500Hz and γB1p/ωRF=1.37 with imaging parameters identical to previous experiment. Minimum TE values ([14.88ms 15.88ms]) were used for optimal SNR.
All B1 maps based on the overdriven condition were reconstructed using simulated ϕ2BS-B1 curves similar to Figure 2a, while including ΔfBo, which was derived from a simultaneously acquired B0 map. After this, the B1 maps were rescaled with respect to the nominal B1 value to generate a relative B1 map. Then the relative B1 estimation errors were computed by taking the difference between measured and the reference B1 maps, and dividing by the reference B1 map. The Root-Mean-Squared Error (RMSE) within brain was then computed for each B1 map to derive a quantitative metric for accuracy.
RESULTS
The advantage of OBS in terms of accuracy in B1 map is demonstrated in Figure 3 and Table 1. Under the overdriven condition with γB1p/ωRF=1.54 and sufficient crusher, the B1 map computed from the ϕ2BS-B1 curve has an RMSE of 0.57%. In fact, for most of the brain region within the slice, the relative B1 errors are below 1%. In contrast, the B1 map computed based on linear assumption under this condition underestimated the actual B1, and has an RMSE of 11.47%. When operating under regular condition (γB1p/ωRF=0.37), the B1 map acquired with crushers and computed based on linear assumption has a RMSE of 3.01%, which is comparable to that achieved in recent studies (e.g. (18,20)), but about 5-fold larger than the RMSE acquired under the overdriven condition. Without crushers, large errors (RMSE=12.57%) can be observed in the B1 map even when operating under regular condition, as a result of direct excitation from the BS pulse. This suggests that the use of a crusher is necessary when the frequency offset of the BS pulse is brought closer to the Lamor frequency (e.g. 500Hz in this case). For the results presented in the following crusher moments fifteen times larger than that of the slice rephasing gradient were used, amounting to 98.9 mT/m*ms.
Table 1.
The ANR advantage of OBS is illustrated by the results presented in Figure 4. As predicted by the simulations, the mapping efficiency increases with increasing TR. This increase results from fact that at a constant level of RF power deposition (i.e. SAR), increased TR allows an increase in BS pulse amplitude. The associated (approximately linear) increase in ϕ2BS with increase in TR translates into an efficiency gain of roughly sqrt(TR), due to reduced averaging per unit scan time associated with longer TR. Thus long TR sequences are preferred in terms of ANR efficiency at a given SAR level. Unlike predicted by simulation (Figure 2b), with the imaging flip angle properly adjusted with TR, this increase starts to saturate at longer TR, due to uncounted system instability. Nevertheless, comparison of an OBS sequence with a traditional BS sequence used at identical TR, TE, and imaging flip angle (Figure 4b) shows ANR efficiencies of 130.39 and 50.84 respectively (OBS: ΔfRF=500Hz, QDAPX pulse, 55% maximum SAR level, BS: ΔfRF=8kHz, Fermi pulse, 98% maximum SAR level). The OBS sequence achieved at least 2.5 times improvement in ANR with 44% lower RF energy deposition than the BS mapping under conventional conditions.
Figure 5 summarizes the results on multi-slice acquisition. Figure 5a and b show sagittal B0 maps and corresponding histogram covering the entire brain, indicating that the B0 offset is within the range from -500Hz to 500Hz. This suggests that the Assumption 1 is satisfied. The ANR advantage of OBS during multi-slice acquisition is demonstrated by Figure 5c-f. In fact, when comparing OBS with BS, the root-mean-squared ratio in ANR between these two sequences is 2.97, despite of a roughly 25% loss in imaging signal under the overdriven condition. The latter is attributed to the increase in TE necessitated by the crushers and a potential signal saturation associated with irradiating close to resonance.
DISCUSSION
In this study, it is shown that the BS-based B1 mapping sequences can be performed with γB1p/ωRF ratios far higher than those used previously, and well outside the regime for which the linear approximation holds. As predicted by the theoretical findings regarding the relationship between SAR and ωRF, significant gains in B1 mapping sensitivity can be achieved while maintaining or reducing the SAR level without sacrificing B1 estimation accuracy. In addition, the analytical equations for the BS shift in the presence of off-resonance irradiation were derived and simplified assuming upper limits for B0 inhomogeneity. These equations provide alternative formulae with better theoretical accuracy than the original linear approximation under both traditional and overdriven cases. Although Eq.2 is derived under a sufficient off-resonance assumption proposed by Ramsey (22), the actual BS phase shift under overdriven condition will oscillate around the value calculated by Eq.2 when a sufficiently large crusher pair is used. The amplitude of the oscillation is pulse-shape dependent. In general, Eq.2 provides a rough estimate of the BS shift without the need of actual Bloch simulations. In addition, since Eq.2 was derived without Taylor-series approximation, it could be used as a more accurate alternative to the linear approximation (5) even when |γB1p|<ωRF.
The introduction of crusher gradients into the BS sequences was found to have benefit additional to the reducing direct excitation artifacts suggested in previous reports (14,16,19-21). In fact, a sufficiently strong crusher pair removes the dependency on imaging flip angle and reduces the sensitivity to actual slice profiles and steady state effects. Thus, the flip angle of the imaging pulse can be simply set to the Ernst angle to further improve SNR. Crusher gradients stabilize the ϕ2BS at high B1 values (i.e. overdriven states), which makes the B1 estimation from ϕ2BS more robust to the noise. The benefits of crusher gradients are not limited to conditions specific to OBS, but apply to traditional BS mapping conditions as well. Therefore, crusher gradients are expected to improve the performance of the BS mapping in general. A drawback of using strong crushers is the associated increase in TE, which leads to some signal (and efficiency) loss. This suggests that the actual crusher moments should be optimized rather than using an arbitrarily large value.
It is found both in theory and in experiment that for GRE-based BS sequences, long TR is preferred for better ANR efficiency. It also suggested that the optimal imaging flip angle should be close to the Ernst angle. In this context, EPI-readout (19) is preferable over single line readout (as used here) to speed up the acquisition without sacrificing SNR.
The in vivo results suggest that the TR can be increased to improve ANR efficiency. However in practice, there is an upper limit on TR when no additional gain is achievable in ANR efficiency. Theoretically, when ΔωBo<<ωRF, the asymptotes for Eq.2. are
(7) |
for |γB1p|<<ωRF and:
(8) |
for |γB1p|>>ωRF. In the region when Eq.8 applies, ϕ2BS only increases linearly with B1p, thus the ANR efficiency starts to level off. A practical limit to TR may be dictated by system instabilities (such as B0 drifting or phase instability), which will eventually become dominant and reduce the ANR efficiency, as demonstrated by Figure 4a. The actual upper limit in practice depends on system stability and the level of physiologic noise and motion of the subject.
Both analytical equations and simulation show that the pulse shape has little effect on how ϕ2BS changes with pulse energy when the linear approximation applies (|γB1p|<<ωRF). In this regime, the energy efficiency only depends on ϕRF. However, under the overdriven condition, the relationship between ϕ2BS and pulse energy or B1 does become dependent on pulse shape. An extreme example is when |γB1p|>>ωRF, based on Eq. 8, the most energy efficient pulse will be the block pulse. As shown with the simulation, different pulse shapes have different smoothness of the ϕ2BS-B1 curve and different deviations from Eq.2. This translates into varying robustness to noise in the estimation of B1 from ϕ2BS. In addition, although not explored here, different pulses will also have different sensitivity to ΔωBo. In this paper, for demonstration purpose, a previously designed pulse (20) was chosen for its robustness to noise in the phase image and in the B0 map. It is possible to formulate a pulse optimization problem based on the findings of this paper to further improve the pulse, although the result will be system- and subject-dependent.
The range of the ΔωBo through the imaging volume determines lower limits of ωRF that can be used through Assumption 1. In the experiments presented above, 500Hz frequency offset was chosen for demonstration purpose, with |ωRF|>|ΔωBo| verified by B0 mapping in vivo. For other applications where ΔωBo could be potentially larger than 500Hz, BS pulse with higher frequency offset has to be used to satisfy Assumption 1. However, this does not prevent overdriving BS sequences to achieve better SNR efficiency. In addition, further reduction in ωRF in our experiments is limited by regions with large ΔωBo located at the boundary of large susceptibility changes, e.g. near the nasal cavity and other air-tissue interfaces. In GRE data, these regions generally also suffer from signal loss due to intra-voxel dephasing. Both these issues may be further improved by more advanced B0 shimming methods. In addition, SNR in B0 map also has some effects on the final B1 map. Although it has no effects on either the ANR or the mapping efficiency, it does affect the SNR in the final B1 map if a B0 correction is used. This is not only affecting B1 mapping under overdriven condition, but also applicable generally to any Bloch-Siegert based B1 mapping whenever B0 information is used. The question of how to minimize the effect of the B0 correction on the B1 map is beyond the scope of this paper, but generally speaking for Bloch-Siegert sequences, it depends on the echo spacing, the frequency offset of the Bloch-Siegert pulse, and the ΔωBo sensitivity for a particular pulse shape. In practice, the robustness to noise in the B0 map can be increased by optimizing the pulse shape as shown above, using a high SNR B0 map, or increasing ωRF which in turn reduces the ANR and thus the SNR in the B1 map as well. Since dual frequency acquisition is used in most of the Bloch-Siegert sequences, B0 dependency is in first order compensated for and therefore at a reasonable SNR level, the lower limit of ωRF shouldn't be affected much by the SNR in the B0 map.
One interesting thing to consider is the potentially adverse effect of the crushers on the imaging signal magnitude. Since the Bloch-Siegert pulse is a non-selective pulse, in theory there could be the chance that during multi-slice acquisition, some magnetization outside the current imaging slice is saturated by the Bloch-Siegert pulse and thus results in signal loss when that location is excited soon after. This, again, depends on the actual pulse shape, the subject, and the system. In addition, the use of crushers also prolongs the minimum TE that can be used, which will result in additional signal losses. In our study, after the optimization, a crusher moment about 99 mT/s*ms was used. However, as demonstrated by our experiments (Figure 5c-f), these two adverse effects on SNR are more than compensated for by the large sensitivity gain brought by operating BS sequences under overdriven condition. In addition, although not investigated by this study, SNR losses caused by these two aforementioned effects can be further suppressed by further minimizing the direct excitation effects of the pulse for low-frequency offset purpose, e.g. via similar methods used in (23,24).
It should also be mentioned that B1 mapping is an active area of research and that some of the novel approaches under development may prove to be preferable over BS methods. For example, methods like the one proposed recently by Bornert et al (25,26), may be more straightforward and less limited by SAR concerns. Such methods would obviate the need for the various optimization steps required for BS methods, and therefore be more desirable in practice.
CONCLUSION
Reducing the off-resonance frequency of the irradiation pulse in BS-based B1 mapping allows improved mapping efficiency and speed. Potential detrimental effects of this approach on B1 mapping accuracy can be mitigated by modifications to the acquisition technique and to the theoretical model used for the B1 calculation. As a result, improved B1 accuracy can be achieved for given scan time and SAR, as was demonstrated in human studies.
Acknowledgements
This research was supported by the Intramural Research Program of the National Institute of Neurological Disorders and Stroke. The authors would also like to thank Dr. Sacolick for providing the original Fermi pulse used in (5), and the reviewers of this and related work submitted by our group for their constructive comments.
Funding Source:
Intramural Research Program of the National Institute of Neurological Disorders and Stroke, National Institutes of Health
ABBREVIATION LIST
- RF
Radio-frequency
- SAR
Specific Absorption
- GRE
Gradient Echo
- BS
Bloch-Siegert
- OBS
Overdriven Bloch-Siegert
- ANR
Angle-to-Noise Ratio
- RMSE
Root-Mean-Squared Error
- SNR
Signal-to-Noise Ratio
APPENDIX I
Basic Equations for BS Shift
While equations for the linear approximation for the BS shift were given in (5), here we further examine the BS effect as well as some important details that were not emphasized in (5). We first start with the equation for the Bloch-Siegert frequency shift derived by Ramsey (22) (Note equation (3) in Ramsey's paper has a sufficient off-resonant assumption). In a static main magnetic field with strength B0 without application of RF, the magnetic resonance frequency will equal the Lamor frequency ω0=γB0. If, after excitation, an RF pulse with frequency ω2 is applied, and ωoff=ω0-ω2, the resonance frequency ω during this pulse is given by:
(A1.1) |
where ω1=γB1. This will introduce an additional frequency shift (i.e. the BS frequency shift ωBS) given as a function of ωoff:
(A1.2) |
As pointed out in (5), in practice, to remove undesired phase effects in the image, the phase difference between images acquired with both positive and negative frequency shifts is used in calculating the final B1 map. For convenience, let us define ωRF=2πΔfRF>0, and BS pulses with frequence offsets +ωRF and -ωRF symmetrically placed around ω0 are used. For completeness, we assume there is a local B0 inhomogeneity resulting in off-resonance precession at frequency ω0+ΔωBo rather than at ω0, with ΔωBo=2πΔfBo. In this case, the measurement with the BS pulse at +ωRF will have ωoff=ΔωBo+ωRF, whereas the measurement with the BS pulse at -ωRF will have ωoff=ΔωBo-ωRF. In this case the resulting BS frequency difference will be:
(A1.3) |
which yields Eq.1.
To facilitate further derivation, here we introduce the first practical assumption.
Assumption 1: The local B0 inhomogeneity is not larger than the frequency offset of the BS pulse, i.e. ΔωBo < ωRF. At a given level of inhomogeneity, this condition can be met by using sufficiently large frequency offsets for the BS pulses.
Under this assumption, Eq.1 can be rearranged as:
(A1.4) |
Since equation (A1.4) is symmetric regarding ±ΔωBo, for convenience, we can further assume ΔωBo≥0. In this context, the corresponding equation for the general BS phase shift, i.e. Eq.2, can be derived with the integration over the entire pulse duration. Since pulse parameters (frequency offset ωRF, normalized shape B1n(t), and pulse duration T) in Eq.2 can be determined before the experiment, and the local B0 inhomogeneity can be measured as well, Eq.2 forms a mapping from the local peak B1 (B1p) to the resulting BS phase shift, which generally can be solved numerically for any predefined pulse in mapping the local B1 from the BS phase shift.
Taylor Expansion of the BS Shift
For this purpose, we introduce an additional assumption.
Assumption 2: The pulse sequence operates in a low-SAR regime, so that |γB1p|≤ωRF. At given B1p, this condition can be met in practice by using BS pulses with sufficiently large frequency offsets. Under this assumption, based on the Taylor expansion, we have:
(A1.5) |
Note that the equations above also apply to conditions where ωRF and ΔωBo are time dependent.
When ΔωBo=0, and only first order Taylor approximation is used, we have:
(A1.6) |
Or:
(A1.7) |
Based on the theory of error propagation, it can show that SNRB1∝ SNRimageγ2SAR/ωRF, which suggests for a constant image SNR (e.g. fixed TE and TR), the SNR in B1 map is proportional to SAR/ωRF.
APPENDIX II
In general, with ignoring relaxation effects, the rotation of a magnetization vector under the effect of some pulse with B1 (ω1=γB1), duration t, and frequency offset ωoff=2πΔfoff is given by:
(A2.1) |
with M1 and M2 are the magnetization vector before and after the application of the pulse, respectively; θ=tan-1(ω1/ωoff), ; and rotation matrices are:
(A2.2) |
When sufficient bipolar crusher pair is applied around the pulse, Eq.A2.1 becomes:
(A2.3) |
In case of the BS-based sequence presented in this paper, let's further assume the flip angle of the imaging pulse is FA, and the normalized initial magnetization is M1 =[0,0,1]T. The integrand in (A2.3) becomes:
(A2.4) |
The final magnetization vector becomes:
(A2.5) |
with a significant reduction in complexity due to the periodicity of sinusoid functions. Since the ratio between x and y components determines the phase shift in the x-y plane, i.e. the phase shift in the signal, it is obvious that with the presence of sufficient bipolar crusher, the Bloch-Siegert phase shift (derived from φ) does not depend on the imaging flip angle FA, so long as FA>0. Another important point is that although Eq.A2.1 implicitly assumes a constant B1 (i.e. hard pulse), for an arbitrary pulse with time-varying B1 values, at any instance, the current magnetization has some equivalent flip angle, thus the conclusion from Eq.A2.5 does not change, i.e. the Bloch-Siegert phase shift does not depend on imaging flip angle for any pulse when a sufficient surrounding bipolar crusher pair presents. Nevertheless, since sin(FA) contributes the total magnitude of the transverse magnetization vector in the x-y plane, it does affect the SNR in the phase angle calculation when noise is present.
REFERENCES
- 1.Hoult DI, Phil D. Sensitivity and power deposition in a high-field imaging experiment. J Magn Reson Imaging. 2000;12(1):46–67. doi: 10.1002/1522-2586(200007)12:1<46::aid-jmri6>3.0.co;2-d. [DOI] [PubMed] [Google Scholar]
- 2.Sacolick LI, Sun L, Vogel MW, Dixon WT, Hancu I. Fast radiofrequency flip angle calibration by Bloch-Siegert shift. Magn Reson Med. 2011;66(5):1333–1338. doi: 10.1002/mrm.22902. [DOI] [PubMed] [Google Scholar]
- 3.Zhu Y. Parallel excitation with an array of transmit coils. Magnetic Resonance in Medicine. 2004;51(4):775–784. doi: 10.1002/mrm.20011. [DOI] [PubMed] [Google Scholar]
- 4.Katscher U, Bornert P, Leussler C, van den Brink JS. Transmit SENSE. Magn Reson Med. 2003;49(1):144–150. doi: 10.1002/mrm.10353. [DOI] [PubMed] [Google Scholar]
- 5.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63(5):1315–1322. doi: 10.1002/mrm.22357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bloch F, Siegert A. Magnetic resonance for nonrotating fields. Physical Review. 1940;57(6):522–527. [Google Scholar]
- 7.Schulte RF, Sacolick L, Deppe MH, Janich MA, Schwaiger M, Wild JM, Wiesinger F. Transmit gain calibration for nonproton MR using the Bloch-Siegert shift. NMR Biomed. 2011;24:1068–1072. doi: 10.1002/nbm.1657. [DOI] [PubMed] [Google Scholar]
- 8.Lau AZ, Chen AP, Cunningham CH. Integrated Bloch-Siegert B1 mapping and multislice imaging of hyperpolarized 13C pyruvate and bicarbonate in the heart. Magn Reson Med. 2012;67(1):62–71. doi: 10.1002/mrm.22977. [DOI] [PubMed] [Google Scholar]
- 9.Sacolick L, Grissom WA, Kudielka G, Loew W, Vogel MW. Interference Bloch-Siegert B1 Mapping for Parallel Transmit.. 19th Annual Meeting & Exhibition of ISMRM; Montreal, Canada. 2011.p. 2926. [Google Scholar]
- 10.Duan Q, van Gelderen P, Duyn J. Tailored excitation using nonlinear B0-shims. Magn Reson Med. 2012;67(3):601–608. doi: 10.1002/mrm.23278. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Geschewski FH, Brenner D, Felder J, Jon Shah N. Optimum coupling and multimode excitation of traveling-waves in a whole-body 9.4T scanner. Magnetic Resonance in Medicine. doi: 10.1002/mrm.24403. DOI: 10.1002/mrm.24403. [DOI] [PubMed] [Google Scholar]
- 12.Koning W, Bluemink JJ, Langenhuizen EAJ, Raaijmakers AJ, Andreychenko A, van den Berg CAT, Luijten PR, Zwanenburg JJM, Klomp DWJ. High-resolution MRI of the carotid arteries using a leaky waveguide transmitter and a high-density receive array at 7 T. Magnetic Resonance in Medicine. doi: 10.1002/mrm.24345. DOI: 10.1002/mrm.24345. [DOI] [PubMed] [Google Scholar]
- 13.FDA . Guidance for Industry and FDA Staff: Criteria for Significant Risk Investigations of Magnetic Resonance Diagnostic Devices. U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health, Radiological Devices Branch, Division of Reproductive, Abdominal, and Radiological Devices, Office of Device Evaluation; 2003. [Google Scholar]
- 14.Nehrke K, Bornert P. Fast B1 Mapping using a STEAM-based Bloch-Siegert Preparation Pulse.. 19th Annual Meeting & Exhibition of ISMRM; Montreal, CA. 2011.p. 4411. [Google Scholar]
- 15.Sacolick L, Lee S-K, Grissom WA, Vogel MW. Fast Spin Echo Bloch-Siegert B1 Mapping.. 19th Annual Meeting & Exhibition of ISMRM; Montreal, Canada. 2011.p. 2927. [Google Scholar]
- 16.Sturm VJ, Basse-Lusebrink TC, Kampf T, Stoll G, Jakob PM. Improved encoding strategy for CPMG-based Bloch-Siegert B 1+ mapping. Magn Reson Med. 2012;68(2):507–515. doi: 10.1002/mrm.23232. [DOI] [PubMed] [Google Scholar]
- 17.Basse-Lusebrink TC, Kampf T, Fischer A, Sturm VJ, Neumann D, Kostler H, Hahn D, Stoll G, Jakob PM. SAR-reduced spin-echo-based Bloch-Siegert B(1) (+) mapping: BS-SE-BURST. Magn Reson Med. 2012;68(2):529–536. doi: 10.1002/mrm.23259. [DOI] [PubMed] [Google Scholar]
- 18.Khalighi MM, Rutt BK, Kerr AB. RF pulse optimization for Bloch-Siegert B +1 mapping. Magn Reson Med. 2012;68(3):857–862. doi: 10.1002/mrm.23271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Duan Q, Inati S, van Gelderen P, Patil S, Duyn J. Implementation and Validation of Fast Whole-Brain B1 Mapping Based on Bloch-Siegert Shift and EPI Readout.. 20th Annual Meeting & Exhibition of ISMRM; Melbourne, Australia. 2012.p. 609. [Google Scholar]
- 20.Duan Q, van Gelderen P, Duyn J. Fast Simultaneous B0/B1 Mapping by Bloch-Siegert Shift with Improved Gradient Scheme and Pulse Design.. 20th Annual Meeting & Exhibition of ISMRM; Melbourne, Australia. 2012.p. 2504. [Google Scholar]
- 21.Basse-Lüsebrink TC, Sturm VJF, Kampf T, Stoll G, Jakob PM. Fast CPMG-based Bloch-Siegert B1+ mapping. Magnetic Resonance in Medicine. 2011;67(2):405–418. doi: 10.1002/mrm.23013. [DOI] [PubMed] [Google Scholar]
- 22.Ramsey NF. Resonance Transitions Induced by Perturbations at Two or More Different Frequencies. Physical Review. 1955;100(4):1191–1194. [Google Scholar]
- 23.Khalighi MM, Rutt BK, Kerr AB. Adiabatic Pulse Design for Bloch-Siegert B1+ Mapping.. 20th Annual Meeting & Exhibition of ISMRM; Melbourne, Australia. 2012.p. 607. [Google Scholar]
- 24.Jankiewicz M, Gore JC, Grissom WA. Saturation Pulse Design with Explicit Sensitivity Maximization for Bloch-Siegert B1+ Mapping.. 20th Annual Meeting & Exhibition of ISMRM; Melbourne, Australia. 2012.p. 3362. [Google Scholar]
- 25.Börnert P, Nehrke K, Versluis M, Webb A. Very fast volumetric B1+ mapping at 7 Tesla using DREAM.. 20th Annual Meeting & Exhibition of ISMRM; Melbourne, Australia. 2012.p. 606. [Google Scholar]
- 26.Nehrke K, Bornert P. DREAM-a novel approach for robust, ultrafast, multislice B(1) mapping. Magn Reson Med. 2012;68(5):1517–1526. doi: 10.1002/mrm.24158. [DOI] [PubMed] [Google Scholar]