Abstract
Twelve laboratories evaluated candidate material for an Aspergillus DNA calibrator. The DNA material was quantified using limiting-dilution analysis; the mean concentration was determined to be 1.73 × 1010 units/ml. The calibrator can be used to standardize aspergillosis diagnostic assays which detect and/or quantify nucleic acid.
TEXT
Despite advances in the science of disease diagnosis, including that of infectious diseases, the diagnosis of invasive aspergillosis (IA) remains challenging. Molecular methods have not been widely used diagnostically due to the lack of standardization and validation of these tests (1). Without biologic standards, assay comparison and calibration become very difficult.
The Aspergillus Technology Consortium (AsTeC) is an NIH-contracted consortium which was established to develop and maintain a repository of prospectively collected clinical samples from patients at high risk for developing IA. In addition, AsTeC was also established to evaluate prospective diagnostic assays for IA. In order to appropriately evaluate and compare novel diagnostic assays, it was necessary to establish a reference standard. Herein, we describe a collaborative effort with the Invasive Aspergillosis Animal Models (IAAM) group to develop a nucleic-acid material which can be used to standardize diagnostic assays which target Aspergillus DNA.
Aspergillus DNA was prepared following modification of a previously published method (2). Conidia were collected from potato dextrose (PD) plates and then suspended in two 300-ml aliquots of half-strength PD broth. Following overnight shaking (225 rpm at 30°C), the cultures were centrifuged and pellets were washed with water. Each pellet was resuspended in 20 ml of spheroplasting buffer (1 M sorbitol, 0.1 M EDTA, pH 8.0, and 10 μl of beta-mercaptoethanol diluted to 10 ml with H2O) and incubated with zymolyase (0.1g) for 1 h at 30°C. Following centrifugation, the pellet was resuspended in lysis buffer (10 mM EDTA, 10 mM Tris, and 0.5% SDS) and incubated with 500 μl proteinase K at 50°C for 1 h. Following this incubation, a 1× volume of phenol was added, and incubation was continued at 50°C for 30 min. The aqueous layer was repeatedly extracted with hot phenol until the aqueous layer was clear. Chloroform was added to the supernatant; after separation, DNA was precipitated overnight at −20°C using a 2.5× volume of 100% ethanol and a 0.1× volume of sodium acetate, pelleted, washed twice with 70% ethanol, and air dried. The resuspended pellet was treated with RNase for 45 min and then reextracted with phenol-chloroform and precipitated using 100% ethanol and sodium acetate followed by 70%-ethanol washes. The quality of the purified DNA was assessed by spectrophotometry (expected A260/A280 = 1.7 to 1.9).
Twelve clinical and research laboratories in the United States and Europe were enlisted for determining the quantity of DNA. Each laboratory received a blinded panel of specimens consisting of 10 replicates of calibrator dilutions of 1 × 10−6, 1 × 10−7, 1 × 10−8, 3 × 10−8, 1 × 10−9, and 3 × 10−9 and 5 negative aliquots (Tris-EDTA [TE] buffer).
The concentration of the undiluted specimen was determined by limiting-dilution analysis. This method has been described previously for establishing international standard material for hepatitis C and B viruses (3, 4). The effects of repeated freeze-thaw cycles and prolonged −80°C storage on the integrity of the DNA were assessed at dilutions of 10−5 and 10−6. Ten aliquots were tested at each time point and each freeze-thaw cycle. Each aliquot was tested in a real-time PCR assay as previously described by the IAAM group (6).
The characteristics of the 12 real-time PCR assays used by the participating laboratories are shown in Table 1. Of the 60 negative samples (5 samples × 12 laboratories), there was one false-positive result. Table 1 shows the results of limiting-dilution analysis and calibrator concentration for each laboratory; the results are reported as units/ml and are not corrected for gene copy number. Neither the platform, the specific target, nor the volume of material added to the assay appeared to be correlated with apparent sensitivity of the assay. Table 2 shows the qualitative reporting of all participating centers. The concentration of the calibrator determined by the 12 laboratories ranged from 3.64 × 109 to 3.25 × 1010 U/ml, with a mean concentration of 1.73 × 1010 U/ml (standard deviation [SD], 0.78 × 1010 units/ml). This mean value was the assigned concentration for the undiluted calibrator material.
Table 1.
Real-time PCR assays and results for the 12 laboratories
| Testing center | Amplification method | Gene target | Vol of DNA added to reaction mixture (μl) | Limiting dilution | Calibrator concn (U/ml) |
|---|---|---|---|---|---|
| 1 | Real-time PCR, TaqMan | 18S rRNA | 5 | 1.58 × 10−8 | 1.27 × 1010 |
| 2 | Real-time PCR, TaqMan | 18S rRNA | 10 | 3.8 × 10−9 | 2.63 × 1010 |
| 3 | Real-time PCR, TaqMan | 18S rRNA | 5 | 9.06 ×10−9 | 2.21 × 1010 |
| 4 | Real-time PCR, TaqMan | 18S rRNA | 10 | 2.75 × 10−8 | 3.64 × 109 |
| 5 | Real-time PCR, TaqMan | 18S rRNA | 10 | 5.83 ×10−9 | 1.72 × 1010 |
| 6 | Real-time PCR, TaqMan | 28S rRNA | 25 | 3.00 × 10−9 | 1.33 × 1010 |
| 7 | Real-time PCR, FRETa | 28S rRNA | 10 | 5.96 × 10−9 | 1.68 × 1010 |
| 8 | Real-time PCR, FRET | 18S rRNA | 10 | 1.38 × 10−8 | 7.25 ×109 |
| 9 | Real-time PCR, molecular beacon probes | ITS1 region | 2 | 2.71 × 10−8 | 1.85 × 1010 |
| 10 | Real-time PCR, TaqMan | 28S rRNA | 5 | 1.00 × 10−8 | 2.00 × 1010 |
| 11 | Real-time PCR, TaqMan | ITS region | 10 | 3.08 × 10−9 | 3.25 × 1010 |
| 12 | Real-time PCR, TaqMan | 28s rRNA | 7.5 | 7.84 × 10−9 | 1.70 × 1010 |
FRET, fluorescent resonance energy transfer.
Table 2.
Qualitative results from the 12 participating testing centers
| Dilution | No. of replicates positive/total no. tested) for testing center: |
|||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | |
| 1 × 10−9 | 1/10 | 2/10 | 0/10 | 0/10 | 2/10 | 2/10 | 1/10 | 0/10 | 0/10 | 1/10 | 3/10 | 1/10 |
| 3 × 10−9 | 3/10 | 4/10 | 2/10 | 0/10 | 4/10 | 7/10 | 3/10 | 2/10 | 1/8 | 2/10 | 6/10 | 3/10 |
| 1 × 10−8 | 4/10 | 10/10 | 8/10 | 1/10 | 8/10 | 9/10 | 9/10 | 7/10 | 1/8 | 6/10 | 10/10 | 9/10 |
| 3 × 10−8 | 8/10 | 10/10 | 9/10 | 7/10 | 10/10 | 10/10 | 10/10 | 8/10 | 7/8 | 9/10 | 9/10 | 10/10 |
| 1 × 10−7 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 9/10 |
| 1 × 10−6 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 |
The results of the stability study performed at the central laboratory are shown in Tables 3 and 4. There was little change in crossing-threshold (CT) values over the 2-year storage period. The mean and SD CT values for the 10−5 dilution were 30.4 (0.2) for time zero and 30.5 (0.2) after 2 years of storage. These two values are within the variability of PCR assays. Similar results were seen for the more dilute 10−6 sample. Based on CT values, the DNA was stable for up to 2 years when storage was done at −80°C. Table 4 shows the results of testing after repeated freeze-thaw cycles. These data show that the calibrator material is also stable over at least 10 freeze-thaw cycles.
Table 3.
Stability of AF293 DNA calibrator when stored at −80°Ca
| Dilution and statistic |
CT value for storage time |
|||||
|---|---|---|---|---|---|---|
| Time zero | 1 mo | 3 mo | 6 mo | 1 yr | 2 yr | |
| 10−5 | ||||||
| Mean | 30.4 | 30.5 | 30.6 | 31.0 | 30.8 | 30.5 |
| SD | 0.2 | 0.3 | 0.3 | 0.3 | 0.5 | 0.2 |
| 10−6 | ||||||
| Mean | 34.3 | 34.5 | 34.8 | 34.4 | 34.3 | 34.2 |
| SD | 0.8 | 0.6 | 0.8 | 0.6 | 0.6 | 0.5 |
All values are the crossing-threshold (CT) values obtained by real-time PCR. For references purposes, a value of 3.3 CT is equivalent to a 1-log10 change in the quantity of DNA. All experiments were performed by the central laboratory.
Table 4.
Stability of AF293 DNA calibrator when taken through multiple freeze-thaw cyclesa
| Dilution and statistic |
CT value for freeze-thaw cycle: |
|||||
|---|---|---|---|---|---|---|
| 0 | 1 | 2 | 3 | 5 | 10 | |
| 10−5 | ||||||
| Mean | 30.4 | 30.5 | 30.6 | 30.6 | 31.1 | 30.8 |
| SD | 0.2 | 0.3 | 0.1 | 0.2 | 0.3 | 0.2 |
| 10−6 | ||||||
| Mean | 34.3 | 34.3 | 34.7 | 34.0 | 34.7 | 34.5 |
| SD | 0.8 | 0.8 | 0.7 | 0.4 | 0.4 | 0.6 |
See footnote a of Table 3.
A candidate material was created by purifying DNA from Aspergillus fumigatus strain AF293, and an arbitrary “unit” value was assigned to the material based on limiting-dilution studies performed in 12 laboratories in the United States and Europe. Despite apparent differences in qualitative sensitivities of assays, the analysis produced remarkably similar results, with a standard deviation of 0.78 × 10−10 log10 across the 12 participating laboratories. This work builds upon that of the EAPCRI (2, 5), who demonstrated that standardization of extraction techniques led to improved sensitivity performance of PCR for detecting Aspergillus conidia. The development of this DNA standard will allow improved interlaboratory comparison of techniques and fungal load values.
This study is limited in that all assays used target in some way the ribosomal DNA (rDNA) of Aspergillus. Clearly, this is an attractive target for developing sensitive PCR assays, since there are typically 30 to 90 copies of the rDNA present in most isolates of Aspergillus (6). However, while we can assign an arbitrary “unit” value to our material, it is difficult to extrapolate this to a genome equivalent given the variable number of copies of the DNA per genome. Since the candidate material is DNA, it could not be used to assess methods of extraction from clinical specimens, which could add considerable variability to quantitative results.
Individuals and institutions interested in obtaining calibrator material should contact the corresponding author.
ACKNOWLEDGMENTS
This work was supported by NIH-NIAID N01-AI70023/HHSN266200700023C (AsTeC) and N01-AI30041 (IAAM).
Special thanks go to Rory Duncan and Alec Richey.
AsTeC members are John R. Wingard and Michele Sugrue, University of Florida, Gainesville, FL; Barbara D. Alexander, Duke University Medical Center, Durham, NC (participating laboratory); Angela M. Caliendo (participating laboratory) and G. Marshall Lyon, Emory University Hospital, Atlanta, GA; Lindsey Baden and Francisco Marty, Harvard University and Brigham and Women's Hospital, Boston, MA; L. Joseph Wheat, MiraVista Diagnostics, Indianapolis, IN; David Denning, The University of Manchester, Manchester, United Kingdom; Ming-Hong Nguyen, and Cornelius J. Clancy, University of Pittsburgh Medical Center, Pittsburgh, PA. The Laboratory Working Group of EAPCRI members (all participating laboratories) are Juergen Loeffler, Universitätsklinikum Würzburg, Wuerzburg, Germany; Stéphane Bretagne, University Paris-Est, Créteil, France; Willem J. G. Melchers, Radboud University Nijmegen Medical Centre, Nijmegen, Netherlands; P. Lewis White, University Hospital of Wales, Cardiff, Wales, United Kingdom; Lena Klingspor, Karolinska Institutet, Karolinska University Hospital, Stockholm, Sweden; Elaine McCulloch, Royal Hospital for Sick Children, Glasgow, Scotland; and Manuel Cuenca-Estrella, Instituto de Salud Carlos III, Madrid, Spain. The Invasive Aspergillosis Animal Models group consists of Thomas F. Patterson (participating laboratory), Brian Wickes, William R. Kirkpatrick, Laura Navjar, Nathan P. Wiederhold, and Monica Herrera, University of Texas Health Science Center at San Antonio, San Antonio, TX; Scott Filler, Harbor-UCLA, Torrance, CA; Donald C. Sheppard, McGill University, Montreal, Quebec, Canada; and David Denning and Peter Warn, University of Manchester, Manchester, United Kingdom. Other participating laboratories are as follows: Mark Wessel and Steven B. Kleiboeker, Viracor-IBT Laboratories, Lee's Summit, MO; and Brent L. Seaton, Focus Diagnostics, Inc., Cypress, CA.
Footnotes
Published ahead of print 24 April 2013
REFERENCES
- 1. De Pauw B, Walsh TJ, Donnelly JP, Stevens DA, Edwards JE, Calandra T, Pappas PG, Maertens J, Lortholary O, Kauffman CA, Denning DW, Patterson TF, Maschmeyer G, Bille J, Dismukes WE, Herbrecht R, Hope WW, Kibbler CC, Kullberg BJ, Marr KA, Munoz P, Odds FC, Perfect JR, Restrepo A, Ruhnke M, Segal BH, Sobel JD, Sorrell TC, Viscoli C, Wingard JR, Zaoutis T, Bennett JE. 2008. Revised definitions of invasive fungal disease from the European Organization for Research and Treatment of Cancer/Invasive Fungal Infections Cooperative Group and the National Institute of Allergy and Infectious Diseases Mycoses Study Group (EORTC/MSG) Consensus Group. Clin. Infect. Dis. 46:1813–1821 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. White PL, Bretagne S, Klingspor L, Melchers WJ, McCulloch E, Schulz B, Finnstrom N, Mengoli C, Barnes RA, Donnelly JP, Loeffler J. 2010. Aspergillus PCR: one step closer to standardization. J. Clin. Microbiol. 48:1231–1240 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Saldanha J. 1999. Standardization: a progress report. Biologicals 27:285–289 [DOI] [PubMed] [Google Scholar]
- 4. Saldanha J, Gerlich W, Lelie N, Dawson P, Heermann K, Heath A. 2001. An international collaborative study to establish a World Health Organization international standard for hepatitis B virus DNA nucleic acid amplification techniques. Vox Sang. 80:63–71 [DOI] [PubMed] [Google Scholar]
- 5. White PL, Mengoli C, Bretagne S, Cuenca-Estrella M, Finnstrom N, Klingspor L, Melchers WJ, McCulloch E, Barnes RA, Donnelly JP, Loeffler J. 2011. Evaluation of Aspergillus PCR protocols for testing serum specimens. J. Clin. Microbiol. 49:3842–3848 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Herrera ML, Vallor AC, Gelfond JA, Patterson TF, Wickes BL. 2009. Strain-dependent variation in 18S ribosomal DNA copy numbers in Aspergillus fumigatus. J. Clin. Microbiol. 47:1325–1332 [DOI] [PMC free article] [PubMed] [Google Scholar]
